Isodose linesdefinition, influencing factors,
clinical application
LIBIN SCARIAResident Medical Physicist
![Page 2: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/2.jpg)
Beams of ionising radiation have characteristic
process of energy deposition, in
order to represent volumetric &
planar variations in absorbed dose are
depicted by isodose curves
![Page 3: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/3.jpg)
Dose distribution
![Page 4: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/4.jpg)
PERCENT DEPTH DOSE
Quotient expressed as percentage, of the absorbed dose at any depth d to the absorbed dose at dmax along the central axis of the beam
![Page 5: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/5.jpg)
Isodose curves– Lines joining the points of equal percentage depth
dose(PDD)– Curves are usually drawn at regular intervals of
absorbed dose & expressed as a percentage of dose
• Isodose charts– Family of isodose curves– PDD values are normalised at Dmax or reference
depth
![Page 6: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/6.jpg)
![Page 7: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/7.jpg)
Field size (dosimetrical):lateral distance between the 50% isodose lines at a reference depth
Field size (geometrical) : field defining light is made to
coincide with 50% isodose lines of the radiation beam projected on a
plane perpendicular to the beam axis and at standard SSD or SAD
![Page 8: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/8.jpg)
![Page 9: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/9.jpg)
Beam profile• Falloff of the beam– By the reduced side scatter– Physical penumbra width
• High dose or ‘horns’ near the surface in the periphery of the field
• Outside the geometric limits of the beam and the penumbra, the dose variation is the result of– side scatter from the field– both leakage and scatter from the collimator system
![Page 10: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/10.jpg)
![Page 11: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/11.jpg)
PENUMBRA• Dose transitions near the boarders of the
field• Region at the radiation beam over which
the dose rate changes rapidly as a function of distance from the central axis
Geometric penumbraTransmission penumbraPhysical penumbraScatter penumbra
![Page 12: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/12.jpg)
For cobalt-60 teletherapy: P ~ 1cmFor linacs : P~ 3mm
![Page 13: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/13.jpg)
![Page 14: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/14.jpg)
![Page 15: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/15.jpg)
Measurement of isodose curves• Ion chambers• Solid state detectors
– TLD,Silicon diodes,• Radiographic films• Computer driven devices
Ion chamber is the most reliable method due to its relatively flat energy response and precision
![Page 16: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/16.jpg)
![Page 17: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/17.jpg)
Beam Analysing System3D water phantom- IBA Wellhofer Blue Phantom
• Two ion chambers :– Detector A: to move in the tank
of water to sample the dose rate– Detector B: fixed at some point
in the field to monitor the beam intensity with time
• Final response A/B is independent of fluctuation in the output
![Page 18: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/18.jpg)
Sources of isodose chartAtlases of premeasured isodose
chartsGenerated by calculations using
different algorithmsManufacturers of radiation generators
![Page 19: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/19.jpg)
Parameters of isodose curvesBeam quality
Source size, SSD, SDD Penumbra
Collimation and flattening filterField size
![Page 20: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/20.jpg)
Beam Quality• Depth of a given isodose curve increases with
beam quality• Greater lateral scatter associated with lower-
energy beams• For megavoltage beams, the scatter outside the
field is minimized as a result of forward scattering – becomes more a function of collimation than energy
![Page 21: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/21.jpg)
![Page 22: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/22.jpg)
4 MV 6MV 15MV
![Page 23: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/23.jpg)
Field size One of the most important parameters in
treatment planning PDD increases as field size increases Field size dependence of PDD is less
pronounced for higher energy than for lower energies
Field size smaller than 6 cmRelative large penumbra regionBell shapeTPS should be mandatory for small field size
![Page 24: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/24.jpg)
5*5 cm2 10*10 cm2 15*15cm2
![Page 25: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/25.jpg)
![Page 26: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/26.jpg)
SSD affects the PDD and the depth of the isodose curvesPDD icreases with SSD
![Page 27: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/27.jpg)
Beam fattening filter• Intensity is more at central axis and
decreases as we move away• Non-uniform dose at any given depth• FF is used to uniform it• Usually made up of Al or Brass
![Page 28: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/28.jpg)
Wedge filters Wedge shaped absorber causes a progressive
decrease in intensity across the beam Results in tilt of the isodose curve & degree of
the tilt depends upon the slope of the wedge filter
Material: tungsten, brass. Lead or steel
![Page 29: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/29.jpg)
15o Wedge
![Page 30: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/30.jpg)
30o Wedge
![Page 31: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/31.jpg)
45o Wedge
![Page 32: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/32.jpg)
60o Wedge
![Page 33: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/33.jpg)
Wedge isodose angle (θ) is the complement of the angle through which the isodose curve is tilted with respect to the central ray of the beam at any specified depth
This depth is important because the angle will decrease with increasing depth.
Angle of isodose tilt to decrease with increasing depth in the phantom.
The choice of the reference depth varies:
10 centimeters1/2 - 2/3rd of the beam widthAt the 50% isodose curve
![Page 34: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/34.jpg)
Hinge angle,φIt is the angle between central axes of two
beams passing through the wedge
Relationship b/w φ & θ
Wedge angle,θ= 90 – φ/2
![Page 35: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/35.jpg)
![Page 36: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/36.jpg)
![Page 37: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/37.jpg)
Bolus• A tissue equivalent material
used to reduce the depth of the maximum dose (Dmax)
• In megavoltage radiation bolus is primarily used to bring up the buildup zone near the skin in treating superficial lesions.
• The thickness is usually 0.5 cm to 1.5 cm
![Page 38: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/38.jpg)
![Page 39: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/39.jpg)
![Page 40: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/40.jpg)
Combined Open Field Technique
• Criteria:– The dose distribution within the tumor volume is
reasonably uniform (±5%).– The maximum dose to the tissue in the beam is not
excessive (not more than 110% of the prescribed dose)
– Normal critical structures in the beam do not receive doses near or beyond tolerance
![Page 41: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/41.jpg)
![Page 42: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/42.jpg)
Parallel Opposed Fields
• Advantages– The simplicity and reproducibility of setup– Homogeneous dose to the tumour– Less chances of geometrical miss
• Disadvantage– Excessive dose to normal tissues and critical
organs above and below the tumour
![Page 43: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/43.jpg)
Multiple fields To deliver maximum dose to the
tumour and minimum dose to the surrounding tissues
• Using fields of appropriate size• Increasing the number of fields or portals• Selecting appropriate beam directions• Adjusting beam weights• Using appropriate beam energy• Using beam modifiers
![Page 44: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/44.jpg)
Multiple fields Certain beam angles are prohibited
• Presence of critical organs in those directions• Setup accuracy of a treatment may be better with
parallel opposed beam arrangement
The acceptability of a treatment plan depends not only on the dose distribution but also on
• Practical feasibility• Setup accuracy• Reproducibility of the treatment technique
![Page 45: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/45.jpg)
![Page 46: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/46.jpg)
![Page 47: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/47.jpg)
![Page 48: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/48.jpg)
![Page 49: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/49.jpg)
Isodose lines of Electrons
![Page 50: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/50.jpg)
![Page 51: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/51.jpg)
Isodose curves• Scattering of electrons
determines shapes of isodose curves– Expansion– Lateral constriction
• Larger field size required at surface
![Page 52: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/52.jpg)
![Page 53: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/53.jpg)
![Page 54: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/54.jpg)
Proton isoeffective dose
![Page 55: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/55.jpg)
“Charged Particle Therapy” to better preserve healthy tissues and organs at risk
![Page 56: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/56.jpg)
![Page 57: Isodose lines](https://reader034.fdocuments.in/reader034/viewer/2022050614/588096481a28abd8158b49a9/html5/thumbnails/57.jpg)
Medicine is
a science of uncertainity
and an art of
probabilityDr. William Osler