Introduction to electron microscopy-day2 · Introduction to Electron Microscopy-II Prof. David...
Transcript of Introduction to electron microscopy-day2 · Introduction to Electron Microscopy-II Prof. David...
![Page 1: Introduction to electron microscopy-day2 · Introduction to Electron Microscopy-II Prof. David Muller, dm24@cornell.edu Rm 274 Clark Hall, 255-4065 Ernst Ruska and Max Knoll built](https://reader031.fdocuments.in/reader031/viewer/2022013114/5cc63cc488c993ca238c691d/html5/thumbnails/1.jpg)
David Muller 2008
Introduction to Electron Microscopy-IIProf. David Muller, [email protected] 274 Clark Hall, 255-4065
Ernst Ruska and Max Knoll builtthe first electron microscope in 1931
(Nobel Prize to Ruska in 1986)T4 Bacteriophage
Electron Microscopy bridges the 1 nm – 1 μm gap between x-ray diffraction and optical microscopy
![Page 2: Introduction to electron microscopy-day2 · Introduction to Electron Microscopy-II Prof. David Muller, dm24@cornell.edu Rm 274 Clark Hall, 255-4065 Ernst Ruska and Max Knoll built](https://reader031.fdocuments.in/reader031/viewer/2022013114/5cc63cc488c993ca238c691d/html5/thumbnails/2.jpg)
David Muller 2008
Lenses in a Transmission Microscope(and deflection coils to correct their alignment)
http://www.rodenburg.org/RODENBURG.pdf
Condensor: uniformly illuminate the sampleIf misaligned, you will lose the beam when changing magnification
Gun: electron sourceIf misaligned, low intensity & other alignments may also be out
Objective: image sample – determines resolution. If misaligned, the image will be distorted, blurry.
projector: magnifies image/ forms diffraction pattern – should not alter resolution. If misaligned, the image will be distorted, diffraction pattern may be blurry.
![Page 3: Introduction to electron microscopy-day2 · Introduction to Electron Microscopy-II Prof. David Muller, dm24@cornell.edu Rm 274 Clark Hall, 255-4065 Ernst Ruska and Max Knoll built](https://reader031.fdocuments.in/reader031/viewer/2022013114/5cc63cc488c993ca238c691d/html5/thumbnails/3.jpg)
Reciprocity
From L. Reimer, Transmission Electron Microscopy
Reciprocity: Electron intensities and ray paths in the microscope remain the same if (i) the direction of rays is reversed, and (ii) the source and detector are interchanged.
Proof follows from time-reversal symmetry of the electron trajectories and elastic scattering (to all orders).
Reciprocity does not hold for inelastic scattering:Sample is after probe forming optics in STEM - energy losses in sample do not cause chromatic blurring in the image
Sample is before the imaging optics in TEM – energy losses in the sample do cause chromatic blurring in the image. Imaging thick samples in TEM can be improved by energy filtering (so on the zero-loss image is recorded). This is not needed for STEM.
![Page 4: Introduction to electron microscopy-day2 · Introduction to Electron Microscopy-II Prof. David Muller, dm24@cornell.edu Rm 274 Clark Hall, 255-4065 Ernst Ruska and Max Knoll built](https://reader031.fdocuments.in/reader031/viewer/2022013114/5cc63cc488c993ca238c691d/html5/thumbnails/4.jpg)
Reciprocity
From L. Reimer, Transmission Electron Microscopy
Image recordedIn parallel
Image recorded serially by scanning the source
Condensoraperture(before sample)
Objective aperture
(after sample)
Controls coherence
Controls resolution
Collectoraperture
Condensoraperture
STEMTEM
![Page 5: Introduction to electron microscopy-day2 · Introduction to Electron Microscopy-II Prof. David Muller, dm24@cornell.edu Rm 274 Clark Hall, 255-4065 Ernst Ruska and Max Knoll built](https://reader031.fdocuments.in/reader031/viewer/2022013114/5cc63cc488c993ca238c691d/html5/thumbnails/5.jpg)
ReciprocityReciprocity (or STEM vs. CTEM)(or STEM vs. CTEM)CTEM STEM
Reciprocity (for zero-loss images):A hollow-cone image in CTEM an annular-dark field image in STEM.
Specimen
Illuminationangle α Collector
angle
β
Specimen
Objective Aperture
Objective Aperture
ViewingScreen Gun
Gun Detectors
However: In STEM, energy losses in the sample do not contribute to chromatic aberrations (Strong advantage for STEM in thick specimens)
![Page 6: Introduction to electron microscopy-day2 · Introduction to Electron Microscopy-II Prof. David Muller, dm24@cornell.edu Rm 274 Clark Hall, 255-4065 Ernst Ruska and Max Knoll built](https://reader031.fdocuments.in/reader031/viewer/2022013114/5cc63cc488c993ca238c691d/html5/thumbnails/6.jpg)
David Muller 2008Single atom
Sensitivity:
Electron Energy Loss Spectrometer
Annular Dark Field (ADF) detector
yx
200 kV IncidentElectron Beam
(ΔE=1 eV)
Incr
easi
ngen
ergy
loss
1 atom wide (0.2 nm) beam is scannedacross the sample to form a 2-D image
Elastic Scattering ~ "Z contrast"
Scanning Transmission Electron Microscopy
0 0.5 1 1.
ADF Signal Er M4 Edge
Distance (nm)
3 Å
P. Voyles, D. Muller, J. Grazul, P. Citrin, H. Gossmann, Nature 416 826 (2002)U. Kaiser, D. Muller, J. Grazul, M. Kawasaki, Nature Materials, 1 102 (2002)
![Page 7: Introduction to electron microscopy-day2 · Introduction to Electron Microscopy-II Prof. David Muller, dm24@cornell.edu Rm 274 Clark Hall, 255-4065 Ernst Ruska and Max Knoll built](https://reader031.fdocuments.in/reader031/viewer/2022013114/5cc63cc488c993ca238c691d/html5/thumbnails/7.jpg)
David Muller 2008
Brightness:Brightness: Not how many electronshow many electrons, but how many electrons go where we want them tohow many electrons go where we want them to
Thermionic Sources:• Large Source• Weak Electric Field• Electrons are “boiled off” in all
directions• Low brightness
Field Emission Guns:• Small Source• Strong Electric Field• Electrons tunnel out along field
lines• High brightness
ElectricField Lines
10 μm 0.1 μm
![Page 8: Introduction to electron microscopy-day2 · Introduction to Electron Microscopy-II Prof. David Muller, dm24@cornell.edu Rm 274 Clark Hall, 255-4065 Ernst Ruska and Max Knoll built](https://reader031.fdocuments.in/reader031/viewer/2022013114/5cc63cc488c993ca238c691d/html5/thumbnails/8.jpg)
David Muller 2008
The Brightness Equation
• Brightness is the current density (J=ΔI/ΔS) per solid angle (ΔΩ=πα2) where α is the half angle
2παβ J
SI
=ΔΩΔ
Δ= -(1)
• For a thermal source, the electron current has a maxwellian momentum distribution.
• Transverse momentum Normal to the Field
• Root mean square angle subtended:
• hence ⎟⎠⎞
⎜⎝⎛≈⎟
⎠⎞
⎜⎝⎛ +=
kTeVJ
kTeVJ
ππβ 1 Since eV>>kT
mkTpt 22 = meVmkTpn 222 +=
( )kTeVpp nt +== 11222α
α
E
pt
pn
Brightness increase with beam voltage, decrease with tip temperature
![Page 9: Introduction to electron microscopy-day2 · Introduction to Electron Microscopy-II Prof. David Muller, dm24@cornell.edu Rm 274 Clark Hall, 255-4065 Ernst Ruska and Max Knoll built](https://reader031.fdocuments.in/reader031/viewer/2022013114/5cc63cc488c993ca238c691d/html5/thumbnails/9.jpg)
David Muller 2008
• For current density J and probe diameter d0, the probe current is
( ) JdI p2
0 2π= -(2)
A small spot size implies a small beam current or large illumination angle
Brightness as a limit to Spatial Resolution
• From the Brightness equation (1) we also know
• Hence the probe current depends on
2βπα=J
220
2
4αβπ dI p = -(3)
• Rearranging, we find the probe size
αβπ14 21
20 ⎟⎟⎠
⎞⎜⎜⎝
⎛= pI
d -(4)
![Page 10: Introduction to electron microscopy-day2 · Introduction to Electron Microscopy-II Prof. David Muller, dm24@cornell.edu Rm 274 Clark Hall, 255-4065 Ernst Ruska and Max Knoll built](https://reader031.fdocuments.in/reader031/viewer/2022013114/5cc63cc488c993ca238c691d/html5/thumbnails/10.jpg)
David Muller 2008
10-10 - 10-1110-8 - 10-910-5 - 10-610-4 - 10-5Vacuum (Torr)
>5 years3-5 years150-300 hrs25-100 hrsLifetime
3001400-18001400-20002500-3000Operating Temp (K)
0.30.522Energy SpreadFWHM (eV)
1091085 x 1065 x 105Brightness at 100 kV
(A/cm2/Sr)
~0.03 nm~0.1 nm~ 2 nm~6 nmSource Size at 10 pA, 100 keV
WCold FEG
ZrO2/WSchottky FEG
LaB6Thermionic
TungstenThermionic
Properties of Electron Sources
![Page 11: Introduction to electron microscopy-day2 · Introduction to Electron Microscopy-II Prof. David Muller, dm24@cornell.edu Rm 274 Clark Hall, 255-4065 Ernst Ruska and Max Knoll built](https://reader031.fdocuments.in/reader031/viewer/2022013114/5cc63cc488c993ca238c691d/html5/thumbnails/11.jpg)
David Muller 2008
Tungsten Filament
•Cheap (~$20)•Short lifetime•Low brightness•Works in poor vacuum
![Page 12: Introduction to electron microscopy-day2 · Introduction to Electron Microscopy-II Prof. David Muller, dm24@cornell.edu Rm 274 Clark Hall, 255-4065 Ernst Ruska and Max Knoll built](https://reader031.fdocuments.in/reader031/viewer/2022013114/5cc63cc488c993ca238c691d/html5/thumbnails/12.jpg)
David Muller 2008 http://www.kimphys.com/cathode/catalog_PDF/LaB6_cathode_ES423.pdf
Thermionic LaB6 Filament
•~$1000•Longer life, but can be damaged by thermal shock•Medium brightness•Needs moderate vacuum
![Page 13: Introduction to electron microscopy-day2 · Introduction to Electron Microscopy-II Prof. David Muller, dm24@cornell.edu Rm 274 Clark Hall, 255-4065 Ernst Ruska and Max Knoll built](https://reader031.fdocuments.in/reader031/viewer/2022013114/5cc63cc488c993ca238c691d/html5/thumbnails/13.jpg)
David Muller 2008
Schottky Tip design
ZrZr reservoirreservoir
TipTip
Tungsten wire Tungsten wire filamentfilament
•Expensive (~$20,000?)•Long lifetime (years)•High brightness•Needs good vacuum
![Page 14: Introduction to electron microscopy-day2 · Introduction to Electron Microscopy-II Prof. David Muller, dm24@cornell.edu Rm 274 Clark Hall, 255-4065 Ernst Ruska and Max Knoll built](https://reader031.fdocuments.in/reader031/viewer/2022013114/5cc63cc488c993ca238c691d/html5/thumbnails/14.jpg)
David Muller 2008
Electron Scattering
Incident electron
Elastic scattering(large angle, α Z2)
Inelastic scattering(small angle, α Z)
•Elastic scattering can only change the direction of the electron,•Only inelastic scattering can slow it down (by losing energy in the sample)
Valence electrons can be excited(detected as secondary electrons)
![Page 15: Introduction to electron microscopy-day2 · Introduction to Electron Microscopy-II Prof. David Muller, dm24@cornell.edu Rm 274 Clark Hall, 255-4065 Ernst Ruska and Max Knoll built](https://reader031.fdocuments.in/reader031/viewer/2022013114/5cc63cc488c993ca238c691d/html5/thumbnails/15.jpg)
David Muller 2008
Beam Spreading
E0=200 keV E0=20 keV
1 μm of Carbon
Electron Range (in μm):
5.10
064.0 ERρ
≈
(density ρ in g/cm3, E0 in keV)
R~ 100 μm at 200 keV
David Joy’s simulation code is available at http://web.utk.edu/~srcutk/htm/simulati.htmA more detailed simulator can be found at http://www.gel.usherbrooke.ca/casino/What.html
![Page 16: Introduction to electron microscopy-day2 · Introduction to Electron Microscopy-II Prof. David Muller, dm24@cornell.edu Rm 274 Clark Hall, 255-4065 Ernst Ruska and Max Knoll built](https://reader031.fdocuments.in/reader031/viewer/2022013114/5cc63cc488c993ca238c691d/html5/thumbnails/16.jpg)
David Muller 2008
Beam Spreading
For thin films:
5.1
0%90
625 tZAE
b ρ=
At 100 kV:0.16 nm for 10 nm thick C1.8 nm for 50 nm thick C
E0 in kVt in cm, ρ in g/cm2
A in g/mol
0
0.5
1
1.5
2
0 1 2 3 4 5 6
FWHM of feature90% beam radius
Res
olut
ion
(μm
)
Si thickness (μm)
200 keV Electrons in Silicon
![Page 17: Introduction to electron microscopy-day2 · Introduction to Electron Microscopy-II Prof. David Muller, dm24@cornell.edu Rm 274 Clark Hall, 255-4065 Ernst Ruska and Max Knoll built](https://reader031.fdocuments.in/reader031/viewer/2022013114/5cc63cc488c993ca238c691d/html5/thumbnails/17.jpg)
David Muller 2008
Imaging Buried Atoms (Sb in Si)
Sb source turned on here No Sb in substrate
1 nm
Null test:No Sb in substrate
P. Voyles, D. Muller, J. Grazul, P. Citrin, H. Gossmann, Nature 416 826 (2002)
160
140
120
100
80
60
40
20
0
num
ber o
f ato
mic
col
umns
1.41.31.21.11.00.90.8normalized intensity
Si columns Si/Sb columns singly-occupied doubly-occupied
![Page 18: Introduction to electron microscopy-day2 · Introduction to Electron Microscopy-II Prof. David Muller, dm24@cornell.edu Rm 274 Clark Hall, 255-4065 Ernst Ruska and Max Knoll built](https://reader031.fdocuments.in/reader031/viewer/2022013114/5cc63cc488c993ca238c691d/html5/thumbnails/18.jpg)
David Muller 2008
CTF
PSF
(200 kV, C3=1.2 mm)
Effect of defocus and aperture size on an ADF-STEM image
![Page 19: Introduction to electron microscopy-day2 · Introduction to Electron Microscopy-II Prof. David Muller, dm24@cornell.edu Rm 274 Clark Hall, 255-4065 Ernst Ruska and Max Knoll built](https://reader031.fdocuments.in/reader031/viewer/2022013114/5cc63cc488c993ca238c691d/html5/thumbnails/19.jpg)
David Muller 2008
ADF of [110] Si at 13 mr, C3=1mm
Strong {111} fringes Strong {311} fringes
2 clicks overfocus
Best 111 and 311 fringes occur at different focus settingsIf the aperture is too large
What happens with a too-large aperture?
![Page 20: Introduction to electron microscopy-day2 · Introduction to Electron Microscopy-II Prof. David Muller, dm24@cornell.edu Rm 274 Clark Hall, 255-4065 Ernst Ruska and Max Knoll built](https://reader031.fdocuments.in/reader031/viewer/2022013114/5cc63cc488c993ca238c691d/html5/thumbnails/20.jpg)
David Muller 2008
Aperture Size is Critical(200 kV C3=1mm)
0
0.2
0.4
0.6
0.8
1
0 0.2 0.4 0.6 0.8
13 mrad10 mrad
CTF
k (1/A){111}
Si
{220}Si
30% increase in aperture size ~50% decrease in contrast for Si {111} fringes
![Page 21: Introduction to electron microscopy-day2 · Introduction to Electron Microscopy-II Prof. David Muller, dm24@cornell.edu Rm 274 Clark Hall, 255-4065 Ernst Ruska and Max Knoll built](https://reader031.fdocuments.in/reader031/viewer/2022013114/5cc63cc488c993ca238c691d/html5/thumbnails/21.jpg)
David Muller 2008
Aperture Size is Critical
0
5
10
15
20
25
30
0 0.2 0.4 0.6 0.8 1
10 mrad13 mrad
Bea
m C
urre
nt E
nclo
sed
(pA
)
Radius (nm)
r80%
= 0.25 nm
r80%
= 0.8 nm
r64%
= 0.1 nm
All the extra probe current falls into the tails of the probe – reduces SNR
(200 kV C3=1mm)
![Page 22: Introduction to electron microscopy-day2 · Introduction to Electron Microscopy-II Prof. David Muller, dm24@cornell.edu Rm 274 Clark Hall, 255-4065 Ernst Ruska and Max Knoll built](https://reader031.fdocuments.in/reader031/viewer/2022013114/5cc63cc488c993ca238c691d/html5/thumbnails/22.jpg)
David Muller 2008
Finding the Aperture with the smallest probe tails
(Kirkland, Fig 3.11)
4/1
3
22.1 ⎟⎟⎠
⎞⎜⎜⎝
⎛=
Coptλα
( ) 21
min 8.0 λSCf =Δ
![Page 23: Introduction to electron microscopy-day2 · Introduction to Electron Microscopy-II Prof. David Muller, dm24@cornell.edu Rm 274 Clark Hall, 255-4065 Ernst Ruska and Max Knoll built](https://reader031.fdocuments.in/reader031/viewer/2022013114/5cc63cc488c993ca238c691d/html5/thumbnails/23.jpg)
David Muller 2008
Summary
Contrast Transfer Functions: Coherent:
Lower resolution, higher contrastEasy to get contrast reversals with defocusAperture size only affects cutoff in CTF
Incoherent:
Higher resolution, lower contrastHarder to get contrast reversals with defocusAperture size is critical – affects CTF at all frequencies
4/34/13min 43.0 λCd =
4/1
3
4⎟⎟⎠
⎞⎜⎜⎝
⎛=
Coptλα
4/34/13min 77.0 λCd =
4/1
3
6⎟⎟⎠
⎞⎜⎜⎝
⎛=
Coptλα
![Page 24: Introduction to electron microscopy-day2 · Introduction to Electron Microscopy-II Prof. David Muller, dm24@cornell.edu Rm 274 Clark Hall, 255-4065 Ernst Ruska and Max Knoll built](https://reader031.fdocuments.in/reader031/viewer/2022013114/5cc63cc488c993ca238c691d/html5/thumbnails/24.jpg)
David Muller 2008
3D Tomography of Nanostructures?3D Tomography of Nanostructures?
J. Harms et al., Structure Fold Des 7, 931-41. (1999)
• 3D: 1 nm resolution• 2D: 0.4 nm•Low dose imaging of individual particles
•Don’t need crystals!
Kirrmoycin stalled ribosome at 13Å
Herpes simplex
Z.H. Zhou et al, J. Mol. Bio.
(Biologists have been doing this for 30 years with TEM)
![Page 25: Introduction to electron microscopy-day2 · Introduction to Electron Microscopy-II Prof. David Muller, dm24@cornell.edu Rm 274 Clark Hall, 255-4065 Ernst Ruska and Max Knoll built](https://reader031.fdocuments.in/reader031/viewer/2022013114/5cc63cc488c993ca238c691d/html5/thumbnails/25.jpg)
David Muller 2008
Tomography in Materials Science
Ta/TaN liner around a copper interconnect
Si nanocrystals in SiO2 Au clusters on Carbon Nanotubes
(Examples from researchers in my group)
M. Weyland, A. Yurtserver, P. Ercius, J. Cha 2005
![Page 26: Introduction to electron microscopy-day2 · Introduction to Electron Microscopy-II Prof. David Muller, dm24@cornell.edu Rm 274 Clark Hall, 255-4065 Ernst Ruska and Max Knoll built](https://reader031.fdocuments.in/reader031/viewer/2022013114/5cc63cc488c993ca238c691d/html5/thumbnails/26.jpg)
Duffield: John Grazul150 Duffield(TEM+STEM)
Clark: Mick ThomasF3 Clark (STEM+EDX)
•1 nm (polymers) –> atomic resolution of crystals in thin samples•X-ray mapping at 1 nm•EELS at < 1 nm•Requires sample thinning (except for nanoparticles)
Transmission Electron Microscopy
•Clark: Mick ThomasF3 Clark Hall
•Bard/Snee:John Hunt SB56 Bard/1149 Snee
•Dr. Jonathan ShuD-22 Clark Hall
•Prof. Kit UmbachSB-60C Bard Hall
•CNF Clean Room
Location•Topographic Imaging on wafers•Accurate height measurements on flat surfaces (~ 0.5 nm vertical)•Lateral Resolution 10-20 nm•In-situ – no vacuum required
Atomic Force Microscopy
•Imaging of complex structures at 1-20 nm resolution•X-ray mapping at 100-500 nm •In-vacuum•Clark: High spatial resolution•Snee/Bard: best x-ray mapping, OIM
Scanning Electron Microscopy
ApplicationsType
Materials Microscopy Resources on Campus(http://www.ccmr.cornell.edu/facilities/)