Intraskandigi Manual
-
Upload
amar-raveendran -
Category
Documents
-
view
85 -
download
10
Transcript of Intraskandigi Manual
Company Confidential Page 1 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
Document#UMN-000089-0-20
Document Revision B
Publication Language Document Number:
ENGLISH UMN-000089-0-20
INTRA ORAL DIGITAL X-RAY SYSTEM
USER MANUAL Revision B
Company Confidential Page 2 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
TABLE OF CONTENT
INTRODUCTION .......................................................................................... 5
1. INDICATION FOR USE ................................................................... 6
2. CONTRAINDICATIONS ................................................................... 6
3. SAFETY REQUIREMENTS ............................................................. 6
3.1. GENERAL SAFETY TIPS ......................................................................... 6 3.2. SAFETY SYMBOLS ................................................................................. 7 3.3. IMPORTANT INFORMATION.................................................................... 8 3.4. COMPLIANCE TO STANDARDS ............................................................... 8 3.5. WARNING AND CAUTION STATEMENTS ................................................ 9
4. PRE-INSTALLATION INFORMATION AND RECOMMENDATIONS .............................................................................. 12
4.1. PURPOSE OF THE MANUAL ................................................................... 12 4.2. UNPACKING THE UNIT ........................................................................ 12 4.3. PHYSICAL DESCRIPTION OF THE COMPONENT ..................................... 13 4.4. HARDWARE SETUP AND CONNECTIONS .............................................. 15 4.5. COMPUTER REQUIREMENTS ................................................................ 16 4.6. INSTALLATION .................................................................................... 16 4.7. SITE SELECTION .................................................................................. 17 4.8. NETWORK INSTALLATION ................................................................... 17
5. SYSTEM OPERATING INSTRUCTIONS .................................... 17
5.1. SOFTWARE REQUIREMENTS: ............................................................... 17 5.2. HOW TO LOGIN INTO THE SOFTWARE? ................................................ 18 5.3. MENU OPTIONS ................................................................................... 20 5.3.1 TOP ICON BAR .................................................................................................................... 20 5.3.2 QUICK CAPTURE MODE ................................................................................................... 21 5.3.3 DETAILED CAPTURE MODE ............................................................................................. 23
5.4. IMAGE DISPLAY SECTION ................................................................... 39 5.3.4 Pan ........................................................................................................................................ 39 5.3.5 CLOSE .................................................................................................................................. 40 5.3.6 SET UP TOOL BAR .............................................................................................................. 40 5.3.7 SEARCH ................................................................................................................................ 46
6. PERIODIC DATA BACK UP .......................................................... 48
7. MAINTENANCE PROCEDURES .................................................. 48
7.1. CLEANING THE DAS SYSTEM ............................................................. 49
Company Confidential Page 4 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
7.2. CLEANING THE SENSORS ..................................................................... 49
8. WARRANTY STATEMENT ............................................................ 49
9. GLOSSARY ....................................................................................... 50
9.1. ABBREVIATIONS AND EXPLANATIONS OF SYMBOLS ........................... 50
10. TROUBLE SHOOTING. .................................................................. 51
10.1. MY SQL ERRORS .................................................................................. 51 10.2. PROBLEM WITH DAS ............................................................................ 51 10.3. PROBLEM WITH THE SENSOR .............................................................. 52
11. APPENDIX ....................................................................................... 53
11.1. APPENDIX A: CALIBRATION .................................................................. 53 11.2. APPENDIX B: TECHNICAL SPECIFICATION ............................................. 57 11.3. APPENDIX C: IF YOU NEED ASSISTANCE ................................................ 61
Company Confidential Page 5 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
INTRODUCTION
The IntraSkanDigi designed by Skanray Technologies is an innovative product to completely replace traditional X-ray film based intra oral system and film processor. This system offers following benefits:
v Diagnostic quality images – every time. v Lowers the cost of dental imaging by eliminating the
need for a costly film, chemistry and film processor.
v Saves critical time using advanced processing and display systems. Image capture and display takes less than 5 seconds. v IntraSkanDigi is DICOM compliant.
v IntraSkanDigi is intra-net enabled with multi viewing
capability. Technical Support
For technical support please call the local Skanray dealer. To directly call our customer support, please refer to Section 11.3 (Appendix C).
Supplies and Replacement parts To order supplies or replacement parts for IntraSkanDigi, contact your local Skanray dealer.
Company Confidential Page 6 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
1. Indication for use
The IntraSkanDigi is intended for the use as a digital dental radiography using X-ray for radiographic diagnostic intra-oral exposure. System also provides capturing, viewing and processing of the digital X-ray images. The system includes Digital X-ray sensor, Data Acquisition System, communication electronics, software, and various peripheral accessories. 2. Contraindications
There are no contraindications for use. 3. Safety Requirements
Read this section carefully. It contains important safety related instructions.
3.1. General Safety Tips
ELECTRICAL SAFETY
• Always switch off the unit and remove the mains plug when cleaning and disinfecting the unit.
• Do not attempt to open covers or repair the unit yourself or using
non certified service personnel.
INSTALLATION AND SERVICE
• Ensure that the x-ray unit is assembled and installed by qualified service personnel. Consult the factory or Dealer for installation of the unit.
Take the services of qualified personnel when relocating the unit.
CAUTION
Company Confidential Page 7 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
3.2. Safety Symbols
The following safety related symbols are found on the equipment. Symbol
symbol name Description
Caution symbol
This symbol indicates the user to be cautious and refer to the user manual for safe operating instructions.
Protective Earth
Mains Earth is required for continued protection against shock hazards.
High Voltage
Dangerous voltages present.
Caution: X-Ray
X-Ray Source Assembly / Tube Head capable of generating x-rays.
Small Focal Spot
Small Focal Spot
WEEE Symbol
Follow proper procedures for disposing this equipment. Cannot be disposed as general waste
Company Confidential Page 8 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
3.3. Important Information
It is important that all the personnel who will operate this system read and understand this manual before operating the device. All personnel should follow all warnings and cautions as outlined below, for their safety, safety of the patient and safety of others around them.
3.4. Compliance to Standards
This product is designed and manufactured to meet the following Standards and Regulations:
Standards Title
UL 60601-1 Medical Electrical Equipment, Part1: General Requirements for safety
21CFR Chapter1 Subchapter J Performance Standards for Light-Emitting Products
MDD93/42/ECC European Medical Device Directive (CE Mark)
ISO 13485 Medical Devices– Quality management systems
ISO 14971 Risk Analysis
Company Confidential Page 9 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
3.5. Warning and Caution Statements
In this manual, the following definitions apply for all WARNINGS and CAUTION statements. Warnings: Any operation, procedure or practice, which , if not strictly observed, may result in injury or long term health hazards to personnel or the patient. WARNINGS
Should be used by only trained professionals This device should be used by dentists and dental professionals only. Use of this device in procedures other than those described in this manual may result in injury.
Do not open the device to service it. None of the internal parts are user serviceable. If there is a internal problem, call the representative Skanray dealer nearest to you.
Use only grounded electrical connections. Connect the device to a PC connected to a grounded electrical outlet between 100-240 V ac, 50-60 Hz
Sensor should be covered by plastic sleeve. Sensor should not be placed inside the mouth directly. It will have risk of
chemical hazard, contamination, etc to the patient. A plastic sleeve should be used to cover the sensor properly.
Do not reuse the plastic sleeve. The plastic sleeve should be used only once. Once used within the oral cavity, the plastic sleeve becomes bio-contaminated and should NEVER be reused.
Company Confidential Page 10 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
User is instructed to insert sensor smoothly into patient’s mouth. Care should be taken while placing the sensor inside the patient’s mouth. Sensor should be placed in such a manner that it doesn’t cause any discomfort or injury to the patient.
Place the sensor in the holder provided with the equipment To avoid damage to the sensor it is advised to cover the sensor with a plastic cover and place it on the holder provided when not in use.
Users are instructed to use CE marked PC
Use only UL OR CE marked PC to avoid electric leakage
Users are instructed to verify the system date Users are instructed to check the system date. The system date (computer date) and the current date should be same. Any anomaly will affect the study date of the patient data.
Calibration procedure should be done periodically. Sensor is calibrated by service engineer at the time of installation. However, user is advised to run calibration once in 90 days to ensure the image quality.
Earthing of the electric point should be checked Check the earthing of the electric point to which the PC is connected. If earthing is not proper then there is a chance of either user or the patient getting an electric shock.
Check for the live part exposed. The system should be thoroughly checked for any electrical/electronic part or the wire intended to be concealed, being exposed. If any part is exposed
Company Confidential Page 11 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
or the casing is not proper, it can damage the system or cause injury to the patient or user.
Caution Statements
Do not place device on or next to heat source such as oven or water heater, Or radiation sources such as MRI or CT scanner.
Do not position the device in a dusty environment. Excessive dust levels can result in additional service, beyond normal service intervals.
Do not autoclave the sensors.
Do not scratch the sensor.
Company Confidential Page 12 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
4. Pre-installation information and recommendations
4.1. Purpose of the manual
The instructions contained in this manual should be carefully followed for safe, trouble free and effective utilization of the equipment . This manual provides the essential information for the installation, operation and routine care of the device.
4.2. Unpacking the Unit
The IntraSkanDigi system is shipped in a box. Completely unpack the box and store it in a safe, dry location. You may need to repack the unit for shipping if you ever encounter a maintenance outage or while relocating your office. Check whether the contents of the box are according to figures given in 4.2.1-4.2.6. Detailed description of each component is given in section 4.3.
Figure 4.2.1 Figure 4.2.2 DAS Sensor
Figure 4.2.3 Figure 4.2.4 Power Chord User Manual
Company Confidential Page 13 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
If any items are damaged or missing, contact your dealer immediately.
Figure 4.2.5 Figure 4.2.6 USB Type AB CD
4.3. Physical Description of the Component
Ø USB Type A B Cable
Type A-B USB cable is as shown in figure 4.3.1 This type of USB consists of two types of plugs Type A and Type B.
Figure 4.3.1
Type A Plug: These are flat form, commonly used at the end of the cable that plugs into computers USB hub or USB port. Type B Plug: These are square form, commonly used at the end of the cable that plugs into USB peripheral or sensors. Type A and Type B plugs are shown in figure 4.3.2
Figure 4.3.2
Company Confidential Page 14 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
Ø DAS(Data Acquisition System) DAS (Part No: 142-000001-0) is shown in figure 4.3.3. This device connects the sensor to the PC and acquires data from the sensor and sends to PC for display.
Figure 4.3.3
Ø Sensor:
The image sensor (Part No: 134-000002-0) is as shown in figure 4.3.4
is a CCD image sensor ideal for intra-oral X-ray imaging in
Dental diagnosis. This has 500x1000 Pixels each of which is 20x20 micro meter resolution. The fiber optic plate used as input window is as thin as 1.5 mm but highly resistant to X-ray irradiation, making it very reliable even over a long term operation. This is adaptable to both AC and DC X-ray sources.
Figure 4.3.4
Company Confidential Page 15 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
4.4. Hardware Setup and Connections
NOTE: Before you start the hardware setup, ensure that you have an acceptable computer as outlined in Section 4.5. The computer must have an USB port available. Step 1: Pick a location using the guidelines from Section 4.7 Step 2: Set up the computer and monitor as per the manufacturer’s recommendations. Step 3: The Type B plug of USB should be plugged in to the rear panel of the DAS and Type A plug should be plugged into the USB port of the PC. Step 4: Sensor should be connected to the front panel of DSA. The connection is shown schematically in Figure 4.4.1. & 4.4.2 Step 5: Connect the computer to the network if desired. Step 6: Turn the power ON of the computer.
Figure 4.4.1
Figure 4.4.2
Company Confidential Page 16 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
CUSTOMER SUPPLIED COMPONENTS Ø Computer System as described in section 4.5 Ø Monitor for viewing
4.5. Computer Requirements
Minimum Computer requirements Recommended
CPU Speed 800 MHz Pentium 4 2.4GHz
Operating System Microsoft XP SP2 Microsoft XP SP2
RAM 512 MB 2 GB
Hard Drive 80 GB 200 GB
USB port USB 2.0 USB 2.0
UPS 500 VA 800VA
Monitor 15 inch S-VGA 17 inch S-VGA
Video Display Adapter 4 MB RAM, 800 x 600 display, true color
8 MB RAM, 1024 x 768 display, true color
Keyboard, Mouse Standard Standard
Backup Device (CD, DVD, Thumbnail, HARD DISK etc.)
Highly Recommended Highly Recommended
4.6. Installation
This device is designed to be installed only by authorized professional/(s) by Skanray.
Company Confidential Page 17 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
4.7. Site Selection
The IntraSkanDigi can be installed almost anywhere in the dental office. The site should have:
Ø Access to standard grounded electric outlet. Ø Enough room to allow the operator to mount the sensors and use the computer
effectively.
4.8. Network Installation
Refer to your network professional for network installation. 5. System Operating Instructions
5.1. Software Requirements:
• Microsoft operating system (Windows-XP)
• MYSQL version 4.0.23
• JRE (Java Runtime Environment) version 1.5.0.08 or above version
• Nero (Nero 5 is recommended)
Company Confidential Page 18 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
5.2. How to Login into the Software?
As soon as the installation is successfully completed an icon (DENTAL X-RAY) will be created on the desktop (figure 5.2.1). Double click on the icon. When User double clicks on the icon, a Login window is displayed as shown in the figure 5.2.2
Figure: 5.2.1
Figure 5.2.2
Initially the user is provided with the Administrator User Name and Password.
Ø Enter the valid User Name and Password.
Ø Click the Login button.
Ø It will login to the Application.
Ø If you click the Reset button, it will clear the username and password fields.
Ø If user wants software to remember the user name, user should click on the
Remember username check box, so that whenever user opens the software
thereafter, the user name will be present in the field hence no need to reenter.
Company Confidential Page 19 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
Ø When user enters the invalid
User Name or invalid
Password, a popup message
window will appear as shown
in the figure: 5.2.3
Ø After entering the correct
username and password the
main window will appear as
shown in figure 5.2.4
Figure 5.2.3
Figure: 5.2.4
Company Confidential Page 20 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
5.3. Menu Options
There are mainly three different components in IntraSkanDigi’s main window as shown in Figure 5.3.1 The three main components are
1. Top Icon Bar 2. Image Display section 3. Tool section
Each of these components are explained in detail in the following sections.
Figure 5.3.1
5.3.1 TOP ICON BAR There are 4 icons in the left hand top corner.
Ø REGISTER
Ø IMAGE CAPTURE
Ø SETUP TOOLBAR
Ø SEARCH
Company Confidential Page 21 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
5.3.2 QUICK CAPTURE MODE –Without Patient Registration
Ø Click on capture button , If no previously registered patient is available the capture button will automatically forward you to a patient registration window as shown in figure 5.3.2.1.
Ø The mandatory fields are PATIENT
ID, PATIENT NAME and GENDER.
Ø Click on SAVE to save the patient
details. After saving, a tooth image display window will be displayed as shown in figure 5.3.2.2
Ø Select the tooth of which the image
is to be captured. Ø Click on CAPTURE IMAGE, a
message window with message “DEVICE READY-Waiting for X-ray” will be displayed as shown in figure 5.3.2.3
Ø While DEVICE READY-Waiting message is displaying, give the X-ray exposure and image capture will happen automatically.
Ø Once X-ray exposure is given the captured image will be loaded on to the Display panel for review.
Ø Image capture process will continue
until the CANCEL button is clicked
Figure 5.3.2.1
Figure 5.3.2.2
Company Confidential Page 22 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
or TIME OUT occurs.
Ø The image displayed in the image display window will be the last captured image.
Ø To capture a new series or
new patient Again Click on CAPTURE, a window as shown in figure 5.3.2.4 will be displayed.
Ø A message “CONTINUE WITH “CURRENT PATIENT OR NEW PATIENT” will be displayed.
Ø If the user wants to capture the
image of same tooth, click on “CURRENT SERIES” and continue capturing image.
Ø If the user wants to capture the
image of different tooth then click on “NEW SERIES”. Tooth image display window (Figure 5.3.2.2) will be displayed.
Ø Select the tooth image which
has to be captured and continue capturing.
Ø If user wishes to capture the
image for a new patient then click on “NEW PATIENT” and follow the capturing procedure mentioned above.
Figure 5.3.2.3
Figure 5.3.2.4
Company Confidential Page 23 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
5.3.3 DETAILED CAPTURE MODE Here how to go for step by step procedure of registering a patient and capturing the image is elaborated in detail.
5.3.3.1 REGISTER To REGISTER new patient:
Ø When you click on REGISTER icon (as in figure 5.3.3.1) A window will be displayed.(as in figure 5.3.2.1)
Ø Enter the mandatory fields, PATIENT ID,
PATIENT NAME and GENDER. Ø In this window other patient details such as
address, phone number, date of birth etc can also be entered, though they are not mandatory.
Ø There is another tab button called ‘optional’
available for entering other optional patient details such as patient’s profession, height/weight, education qualification, emergency contact etc.
Ø After entering all the necessary details, click on
SAVE button to proceed.
Ø Tooth display image window will be displayed, (as in Figure 5.3.2.2) select the tooth image which has to be captured and click on CAPTURE IMAGE button and the image will be captured.
Figure 5.3.3.1
Company Confidential Page 24 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
5.3.3.2 IMAGE CAPTURE
Ø By clicking the CAPTURE icon(as in Figure: 5.3.3.2), the screen shown in figure 5.3.2.4 will appear.
Ø The window displayed will have three options,
“NEW PATIENT”, “NEW SERIES” and “CURRENT SERIES”.
Ø Clicking on current series will capture the image of
the tooth and load it to the current series.
Ø If user wants to capture the images of a different tooth then click on “NEW SERIES”.
Ø The tooth display image will be displayed. Ø Select the desired tooth and click on CAPTURE
IMAGE.
Ø These images will be loaded to the NEW SERIES.
Ø If user wants to capture image of the new patient then click on “NEW PATIENT”.
Ø Clicking on NEW PATIENT will prompt user to
follow the registration step and then capture the image.
Ø Now the image displayed in the image display
section is the last image captured.
Ø There are many buttons in the tool section which can be used for better viewing of the captured image. The different buttons and their functions are as follows,
Figure: 5.3.3.2
Company Confidential Page 25 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
5.3.3.3 ZOOM PROCESS
Ø When one clicks on ZOOM PROCESS icon
(as in Figure: 5.3.3.3 ) It will display the list
of sub options such as Zoom Glass and Zoom
Glass2.
Zoom Glass:
Ø Click on ZOOM GLASS icon (as in Figure:
5.3.3.4) to enlarge the portion of the image where mouse pointer points.
Zoom Glass2:
Ø Clicking on ZOOM GLASS2 icon (as in Figure: 5.3.3.5), makes the image equalized and gives better contrast over the selected region.
Figure: 5.3.3.3
Figure: 5.3.3.4
Figure: 5.3.3.5
5.3.3.4 IMAGE PROCESS IMAGE Process icon (as in Figure: 5.3.3.6) when clicked will display the list of sub options such as Image Enhancement and Invert. Enhancement:
Ø When you click on IMAGE ENHANCEMENT icon (as in Figure: 5.3.3.7) it improves the interpretability or perception of information in images for human viewers, or it provides `better' input for other automated image processing techniques.
Ø This allows the areas of lower local contrast to gain a higher contrast without affecting the global
Figure: 5.3.3.6
Figure: 5.3.3.7
Company Confidential Page 26 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
contrast.
Invert: Ø clicking the Invert icon (as in Figure: 5.3.3.8)
inverts the grayscale value of the image. (i.e. Black
vs. White)
For example if the grayscale value is ‘0’, click the invert button the value will be changed to ‘255’ (Black vs. White).
Figure: 5.3.3.8
5.3.3.5 IMAGE ROTATION Ø IMAGE ROTATION icon (as in Figure: 5.3.3.9 )
display’s the list of sub options as Horizontal Flip, Vertical Flip, Rotate 90 and Rotate to a chosen angle when clicked.
Horizontal Flip:
Ø HORIZONTAL FLIP icon when clicked (as in
Figure: 5.3.3.10), flips the image horizontally. Vertical Flip:
Ø VERTICAL FLIP icon when clicked (as in Figure:
5.3.3.11), flips the image vertically.
Rotate 90:
Ø To tilt the image by 90° click on the ROTATE
90° icon (as in Figure 5.3.3.12).
Figure: 5.3.3.9
Figure: 5.3.3.10
Figure: 5.3.3.11
Figure: 5.3.3.12
Company Confidential Page 27 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
Rotate Angle:
Ø When you click the ROTATE ANGLE icon(as
in Figure 5.3.3.13) , a popup window opens
with the message “Please Enter the Angle” as
shown in figure 5.3.3.14
Ø One needs to enter the desired angle and click
the OK button then the image will be
rotated/titled by the Corresponding angle.
Figure: 5.3.3.13
Figure 5.3.3.14
5.3.3.6 MULTIPLE SERIES
Ø MULTIPLE SERIES icon when clicked (as in Figure 5.3.3.15) will display the list of sub options like single panel and multiple panel display.
1*1 Series
Ø When you click on the 1*1 SERIES icon (as in Figure 5.3.3.16) , the image will be displayed in a single panel as shown in figure 5.3.3.17
Figure: 5.3.3.15
Figure: 5.3.3.16
Company Confidential Page 28 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
Figure 5.3.3.17 1*4 Series
Ø When you click on the 1*4 SERIES icon (as in
Figure 5.3.3.18), a window is displayed with the
first 4 images in the series in different panels.(as in
Figure 5.3.3.19)
Figure: 5.3.3.18
Company Confidential Page 29 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
Figure 5.3.3.19
5.3.3.7 SERIES OPERATION
Ø When you click on the SERIES OPERATION icon
(as in Figure 5.3.3.20) it displays different sub
icons such as Series Minus, Series Plus, Series
First and Series Last. SERIES Minus:
Ø When you click on the SERIES MINUS icon (as in
Figure 5.3.3.21) it will display the image of the
previous series.
Ø If the loaded series is the first series then a warning
message will be displayed (as in figure 5.3.3.22).
Figure: 5.3.3.20
Figure: 5.3.3.21
Company Confidential Page 30 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
Ø For example, the current image series is in the 3rd
series, if the user clicks this button, this function
will move the image series from 3rd to the 2nd
series. The 2nd image series will be displayed on
the image viewer display panel. If clicked again 1st
series will be displayed.
SERIES Plus:
Ø When you click on the SERIES PLUS icon (as in
Figure 5.3.3.23) it will load the image in the next
series to the image view display section.
Ø If the loaded series is the last series then a
window will be displayed with warning message
“NO MORE SERIES TO LOAD” (as in figure
5.3.3.24).
Ø You can close the dialog box by clicking the Ok
button.
Ø For example, the current image series is in the 3rd
series, if the user clicks this button, this function
will move the image series from 3rd to the 4th
series. The 4th image series will be displayed on
the image viewer display panel
SERIES First:
Ø When you click the SERIES FIRST icon (as in
Figure 5.3.3.25) it will move the series
immediately to the first series viewing position. If
at all it is in First series it will pop up a dialog box
(as in figure: 5.3.3.22)
Ø You need to click the Ok button to close the dialog
box.
Figure: 5.3.3.22
Figure: 5.3.3.23
Figure: 5.3.3.24
Figure: 5.3.3.25
Company Confidential Page 31 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
SERIES Last:
Ø When you click the SERIES LAST icon, it
displays the image from the last series.
Ø If at all the current image displayed is of the last
series, a message will be displayed .(as in figure
5.3.3.27)
Figure: 5.3.3.26
Figure: 5.3.3.27
5.3.3.8 RESET
Ø When you click on the RESET icon (as in Figure 5.3.3.28), it will display the sub options such as Reset W/L, Reset Image and Line Reset.
RESET W/L:
Ø To understand the RESET W/L functionality lets
apply the contrast by pressing the right mouse
button and dragging on the image from right to
left or vice versa or from top to bottom or bottom
to top, then the contrast will be applied.
Ø Now when you click the RESET WL icon (as in
Figure 5.3.3.29), the original image without the
applied contrast would be displayed.
Figure: 5.3.3.28
Figure: 5.3.3.29
Company Confidential Page 32 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
IMAGE RESET:
Ø To understand the IMAGE RESET functionality,
select any of the option vertical flip or horizontal
flip or rotate angle on an image.
Ø Now when you click the IMAGE RESET icon
(as in Figure 5.3.3.30) it will reset the image to
its original position.
LINE RESET:
Ø To understand Line Reset Functionality you need to draw some lines,curves,points on the image (refer section 5.3.3.9 to understand drawing lines,curves,points on the image)
Ø Now when you click on LINE RESET (as in
Figure 5.3.3.31) all lines or points or curved lines on the image would be erased.
Figure: 5.3.3.30
Figure: 5.3.3.31
5.3.3.9 MEASUREMENT
Ø When you click on the MEASUREMENT icon (as
in Figure 5.3.3.32) it displays different sub options
such as LINE, POINT and CURVE measurement.
Figure: 5.3.3.32
Company Confidential Page 33 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
LINE MEASUREMENT:
Ø To draw a line on the image and measure it you
need click on the LINE measurement icon (as in
Figure 5.3.3.33), and place the mouse pointer on
the image, and then draw line by pressing and
holding the left mouse button and drag the mouse
till the destination point and release the mouse
button
Ø Once the user releases the mouse button a line will
be drawn on the image along with the
corresponding distance.
POINT MEASUREMENT:
Ø To draw a point on the image and to know its
value, click the POINT measurement icon (as in
Figure 5.3.3.34) and place the mouse pointer on
the image and press the left mouse button.
Ø Then the point will be placed on the image and the
corresponding value is displayed.
CURVE MEASUREMENT:
Ø To draw a curve on the image and get the distance
between the start point and end point of the curve,
click the CURVE icon (as in Figure 5.3.3.35),
place the mouse pointer on the image, press the left
mouse button and drag the mouse in the desired
pattern.
Figure: 5.3.3.33
Figure: 5.3.3.34
Figure: 5.3.3.35
Company Confidential Page 34 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
Once you release the mouse button the curve will be drawn on the image with the Curve measurement displayed.
5.3.3.10 REPORT
Ø The Functionality Report is used to generate reports by Doctor and store it.
Ø you need to click on the Report icon (as in Figure
5.3.3.39a) then it display’s a dialog box (as shown in Figure 5.3.3.39b ).one need to select the template based on requirement ,then input the details which is required in the report and that finishes the report, a message window will be displayed with message whether the report is final.
Ø If the report is final then choose YES, which will
not allow the user to edit the report in future. Ø If chosen NO, user can edit the report in future.
(refer section 5.3.7.5 in this document for setting up report templates)
Figure: 5.3.3.39a
Figure: 5.3.3.39b
5.3.3.11 PRINT
Ø The print functionality helps us to store the images. When you click on the PRINT icon (as in Figure 5.3.3.40 ) it displays sub options such as CD, LOCAL HDD,SCREEN SAVE icons.
C D:
Ø When you click on CD icon (as in figure 5.3.3.41) , it will check whether the CD is inserted and software for burning CD is present.
Figure: 5.3.3.40
Company Confidential Page 35 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
Ø If both the above checks are satisfied, then the data will be written onto the CD.
Ø If any one is not present, then a error message will
be displayed
v Nero 5 has to be installed for CD write to work.
LOCAL HDD:
Ø When you click on the LOCAL HDD icon(as in
Figure 5.3.3.42) , it will display a Save Location
window .(as in the figure 5.3.3.43)
Ø You need to select the path and file name of the
image and click the save button.
Ø Data will be saved in the selected path.
SCREEN SAVE
Ø When you click the SCREEN SAVE icon (as in
Figure 5.3.3.44) a popup message window will
be displayed (as in the figure 5.3.3.45) the image
screen will be saved in that particular
Patient’s information list.
Figure: 5.3.3.41
Figure: 5.3.3.42
Figure: 5.3.3.43
Figure: 5.3.3.44
Figure 5.3.3.45
Company Confidential Page 36 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
5.3.3.12 DELETE
Ø To delete current image that is displayed from the series click on the DELETE icon (as in figure 5.3.3.46).
Ø After clicking a confirmation window
will be displayed .( as in figure 5.3.3.47)
Ø To delete the image click on YES. Again a window will be displayed (as in figure 5.3.3.48)
Ø Clicking on OK button will delete the image
permanently. Ø Clicking on Cancel will not delete the image.
Figure: 5.3.3.46
Figure 5.3.3.47
Figure 5.3.3.48
Company Confidential Page 37 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
5.3.3.13 COLOR
Ø Color function is mainly used to select the color
between Gray Scale and Color Model display of
the images.
Ø When you click the COLOR icon(as in figure
5.3.3.49), a Color Model window is displayed.( as
in the figure:5.3.3.50)
Ø Select any one of the Color Model option, for
example select Gray color model radio button as
shown in the figure: 5.3.3.50, click Ok button.
Ø The selected color model will be applied on the
image.
If Cancel button is clicked, the selected color model will not be applied and will close the Color Model window
Figure: 5.3.3.49
Figure: 5.3.3.50
Company Confidential Page 38 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
5.3.3.14 HEADER INFORMATION The Header Information option is mainly used to view the header information such as Tag number, Tag Name and Tag value of the dicom image loaded in the image canvas.
Ø When you click on the HEADER icon (as in
Figure 5.3.3.51), it display’s the header
information of the dicom image which is loaded
in the image viewer display window.
Ø A Dicom Tag Information window is displayed
with the tag information (as in the figure:
5.3.3.52)
Figure: 5.3.3.51
Figure: 5.3.3.52
Company Confidential Page 39 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
5.4. Image Display Section
The images will be displayed in the IMAGE DISPLAY SELECTION. Figure 5.3.4.1 shows the image display section with a image in it.
Figure: 5.3.4.1
5.3.4 Pan
Ø To understand the Pan functionality Click the
check box in the PAN icon (as in Figure 5.3.5.1 )
Ø Now press and hold left mouse button and move
the mouse pointer in the image display section,
then image also moves according to the movement
of the mouse pointer.
Ø Deselect the Pan then the image will not be moved
by the motion of the mouse pointer.
Figure: 5.3.5.1
Company Confidential Page 40 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
5.3.5 CLOSE
Ø The CLOSE icon (as in Figure 5.3.6.1) is
positioned on the top right hand corner of the
screen.
Ø When you click on it, a window is displayed.(
as in figure 5.3.6.2 )
Ø Clicking “Yes" will close the application.
Figure 5.3.6.1
Figure 5.3.6.2
5.3.6 SET UP TOOL BAR
Ø When you click on the icon (as in figure 5.3.7.1) it
will give the option to change the settings. Ø A window will be displayed as shown in Figure:
5.3.7.2
Figure: 5.3.7.1
Figure 5.3.7.2
There are different setups. They are 1) Display Setup 2) Create Database 3) Sensor Parameter 4) Browser Sort Setup 5) Report Template Setup 6) Calibrate 7) User Login Setup
Company Confidential Page 41 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
5.3.6.1 DISPLAY SETUP The Display setup provides an option to set various Application parameters and also the parameters which are related with the Look and Feel. Briefly summarizing those options below.
Measurement:
Ø There are two options Single Mode Measurement and Multi Mode
Measurement.
Ø To choose Single Mode Measurement:
o Select the Single Measurement Mode radio button in the Measurement
window.
o Click the Ok button.
Ø To choose Multiple Mode Measurement:
o Select the Multiple Measurement Mode radio button in the Measurement
window.
o Click the Ok button
Ø Close the Measurement Mode window.
Flip Enable:
Ø Here you are provided with two options, one is flip enable and other is flip
disable.
Enable Calibration:
Ø Here you are provided with two options, one is enable Calibration and other disable Calibration.
Company Confidential Page 42 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
Thumbnail: Ø Here you are provided with two options, one is Enable Thumbnail and other
Disable Thumbnail. Text Color: Ø Here you are provided with the option to change the Display text color. Measurement Color: Ø Here you are provided with the option to change the distance color, angel color,
polygon color, and square color and point color display. Browse Color: Ø Here you are provided with the option for the user to change the background and
foreground color of patient list in browser. Filter: Ø Here you are provided with three options. MEAN MEDIAN and NO FILTER. SMOOTH TAB: Ø It’s mainly used for setting the image options i.e. smooth and Real Image. Ø When you click the Setup Tools option, a list of sub options will be displayed,
select the SMOOTH option and then click the Smooth tab, a window is displayed.
Ø Select the Smooth radio button in the Smooth window.
Ø Click the Ok button, a message window is popped up with the message “Values
Inserted Successfully”, click the ok button, the message window will be closed.
Ø Close the Smooth window by using the Cancel button.
Company Confidential Page 43 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
5.3.6.2 CREATE DATABASE The Create Database Option is mainly used to update the Images into the MYSQL Database. After loading the Images into the Image folder, the User can click this option for Image updating.
Ø When you click the SETUP tool bar icon , a
list of options will be displayed, select the
Create Database option, a window is displayed
shown in the figure: 5.3.7.2.1
Ø Click the Yes button, a progress bar window is
popped up, above the progress bar Total no of
images loading into the database are displayed.
Ø After loading all the images into the database, a
window is popped up with the message
“Database Created Successfully”. Click Ok
button, window will be closed.
Figure: 5.3.7.2.1
Company Confidential Page 44 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
5.3.6.3 SENSOR PARAMETER
Ø When you click the Setup tool bar , a list of
options will be displayed, select the SENSOR
PARAMETER option, a window is displayed
shown in the figure: 5.3.7.3.1
Ø There are five parameters to be set. o FILTERING: Two options are provided
here, on and off. If the value is off filtering is disabled, that is filtering is not done before the image is stored as Dicom image. If this is on filtering is enabled. This will filter the image.
o BINNING: The type of binning done
here is a 2X2 binning. In this the image size will be reduced to ¼ the size of the original image. Here four pixels forming a square box is taken, averaged and then considered as a single pixel, this single pixel is then written to a new file having a new dimension.
o WAIT TIME OUT: Time for which the
application waits for X-ray to happen o GAIN: This is gain of the sensor and
ranges from 1-10. o OFFSET: Range is from 0-255.
Ø After entering all the parameters with desired
values click on update.
After clicking on update a window will appear with
message “successfully saved”. Click on OK, window
will be closed.
Figure: 5.3.7.3.1
Company Confidential Page 45 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
5.3.6.4 BROWSER SORT SETUP The “Browse Sort Setup” Option is mainly used for sorting the Patient Information list based on Patient Name or by Study Date.
Ø When you click the Setup Tools button, a list
of sub options will be displayed, select the
Browse Sort Setup option and click the
selected option, a window is displayed as
shown in the figure: 5.3.7.4.1 Ø Select the desired option and click OK Ø After clicking OK a window will pop up with
message “UPDATE SUCCESSFUL” Ø Again click on OK to close the window.
Figure: 5.3.7.4.1
5.3.6.5 REPORT TEMPLATE SETUP
Ø When you click the Setup tool bar , a list of
options will be displayed, select the REPORT
TEMPLATE SETUP option, a window is
displayed as shown in the figure: 5.3.7.5.1
Here specify the path where the reports of the patients are to be stored.
Ø Clicking on REPORT DOC SETUP tab will
allow adding, deleting and updating the report.
Ø After path is set up click on OK, a window
with message VALUE STORED
SUCCESSFULLY. Just click on OK that
window will be closed.
Figure: 5.3.7.5.1
Company Confidential Page 46 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
5.3.6.6 CALIBRATE Calibration is performed on the system to make sure that the image obtained is free from error due to small flaws in the sensor
Ø When you click the Setup tool bar, a list of
options will be displayed, select the
CALIBRATE option, a window is displayed as
shown in the figure: 5.3.7.6.1 Ø Enter the password and follow the calibration
steps. Ø Please refer the Service Manual for Calibration
procedure or Appendix A in this document.
Figure: 5.3.7.6.1
5.3.7 SEARCH The “Browser” option is mainly used for browsing the image from the patient’s information list.
Ø When you click on the Browser icon (as in
figure: 5.3.8.1), a window is displayed as shown
in the figure: 5.3.8.2
Ø Click the Search button, the list of patient’s
information, will be displayed in the patient’s
information panel, i.e. on the left side of the
Patients Information window as shown in the
figure: 5.3.8.2
Ø User can search for the patient list in different
form
Figure: 5.3.8.1
Company Confidential Page 47 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
Ø Enter the patient name and click on search, list
of the patient name entered will be displayed.
Ø Enter patient ID, list corresponding to that
particular ID will be displayed.
Ø Same way there are options like FROM STUDY
DATE, TO STUDY DATE, TODAY ,
YESTERDAY, 7 DAYS and 30 DAYS.
Ø In Update database panel there are two options,
DELETE and REFRESH DATABASE.
Ø Select a data and click on DELETE, it will
display a message “Are You Sure You Want to
Delete the Data?”
Ø Click on Yes the data will be deleted.
Ø REFRESH DATABASE is used when the data
is loaded from the external device.
Ø Only DICOM data is accepted.
Figure: 5.3.8.3
Company Confidential Page 48 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
Figure: 5.3.8.2 6. PERIODIC DATA BACK UP
User should do the data backup periodically. However reminder shall be given by the application. User manually has to back-up two folders Images and Report backup from Intraskandigi folder, in user preferred devices (Thumb Drive, CD, Floppy, DVD etc). 7. Maintenance Procedures
The IntraSkanDigi system is designed for many years of trouble free operation. It’s manufactured from the highest quality components to ensure excellent performance. Maintenance from the operator’s end is minimal.
Company Confidential Page 49 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
7.1. Cleaning the DAS System
Turn off the IntraSkanDigi before cleaning. Wipe the outside surfaces with the paper towel dampened with a cold sterile solution or household, non-abrasive cleaner. Do not spray or soak the system. Be careful not to allow running or dripping solvents into the DAS. This could cause damage to the components inside. Allow to air dry before plugging in or turning back on.
7.2. Cleaning the Sensors
Sensors should be handled carefully. For best images, ensure no scratches on them and keep them dust-free. Use the following procedure to clean them:
1. Use lint-free, 100% cotton gauze. Gently wipe the cotton gauze over the dry sensor surface. Wipe back and forth and then in circular motion.
2. To clean any remaining stains, dampen gauze in anhydrous (water-free) isopropyl alcohol and wipe them using the same motion as above.
3. Completely dry the surface by wiping with another piece of cotton gauze. Ensure that the sensor is completely dry before use.
8. Warranty Statement
The IntraSkanDigi is a device designed exclusively for use in the dental office environment and this warranty is not applicable to other uses. The device is warranted against the defects arising from faulty materials or workmanship for one (1) year from the date of purchase. Any item returned to our factory, during the warranty period, through an authorized dealer, will be repaired or replaced at our option at no charge, provided that our inspection shall indicate it to have been defective. Dealer labor, shipping and handling charges are not covered by this warranty.
Company Confidential Page 50 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
9. Glossary
9.1. Abbreviations and Explanations of Symbols
AC Alternate current DC Direct current Hz Hertz, cycle per second MHz Mega hertz LED Light Emitting Diode kVp Peak voltage in thousands of volts CD-ROM Compact disk- read only CPU Central Processing Unit RAM Random Access Memory MB Mega Byte GB Giga Byte USB Universal Serial Bus HDD Hard Disk Drive
Company Confidential Page 51 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
10. Trouble Shooting.
In the event you experience a problem with the INTRA SKAN digi, Refer to the troubleshooting tips below.
10.1. My SQL errors
Ø One may get errors when MySQL database is
not responding or not started in this scenario
the system pop’s up the error message (as in
figure 10.1)
Ø Trouble shooting can be carried by following
these steps.
Step 1: Restart the application
If the problem still persists continue with step 2
Step 2:Restart the System
If the problem still persists continue with step 3
Step 3: go to start>control panel >Administrative
tools>Services
check whether mysql service is stopped ,if so
try starting it and then proceed with the
application.
Figure 10.1
10.2. Problem with DAS
Ø One may get errors when DAS is not
connected or not properly connected in this
scenario the system pop’s up the error message
Company Confidential Page 52 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
(as in figure 10.2). it also pops’ up a message
telling what to do as in figure 10.3
Ø Trouble shooting can be carried by following
these steps.
Step 1: Restart the application
If the problem still persists continue with step 2
Step 2:close the application, Connect the DAS again
and start the application
If the problem still persists continue with step 3
Step 3: Restart the System and then start the
application
Figure 10.2
Figure 10.3
10.3. Problem with the SENSOR
Ø If SENSOR is not connected properly, then the
image viewer display panel will display white
grainy plain image as shown in figure10.4).
Ø Trouble shooting can be carried by following
these steps.
Step 1: Close the application
Step 2:check the Sensor connection, connect it
Properly and start the application.
Figure 10.4
For any further assistance contact our nearest Dealer/Skanray Office. for details refer to
Section 11.3 (Appendix C).
Company Confidential Page 53 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
11. Appendix
11.1. Appendix A: Calibration
Calibration is done to minimize the dark and white noise effect of the sensor on the image. Calibration is done:
Ø When the instrument is new Ø When specified time period is elapsed Ø When a specified usage (operating hours) has elapsed
Steps for Calibration
Connect the DAS and sensor to the PC using the guidelines from section 4.4
Ø After connection is done start the application. Ø Place the sensor such that the active region should
face the X-ray source.
Ø Also the sensor should be in the field of view of the X-ray source.
Ø Click on SET UP TOOL BAR a window as shown
in figure 11.1.1.1 will appear.
Figure 11.1.1.1
Company Confidential Page 54 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
Ø Select calibrate radio button it will prompt the
password window as shown in figure 11.1.1.2
Ø Enter the service password. If the service password entered is wrong it will display a error message is displayed as shown in figure 11.1.1.3
Ø Click on ok and enter the correct password
Ø Now the device is ready and it will display the
message as “DEVICE READY WAITING FOR X-RAY” as shown in figure 11.1.1.4
Ø Now give the X-ray shot. For calibrating
IntraSkanDigi, 32 X-ray shots should be given. Ø The calculations will be performed by the system. Ø White images will be captured first and then Dark
images will be captured. Ø The derived calibration data will be stored in the
application folder. Ø A message will be displayed “CALIBRATION
SUCCESSFULLY COMPLETED”
Figure 11.1.1.2
Figure 11.1.1.3
Figure 11.1.1.4
Company Confidential Page 55 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
Verification of Calibration
Ø Now RESTART the application
Ø Place test phantom on the sensor.
Ø Now click on SET UP TOOLBAR select DISPLAY SETUP radio button, click on ENABLE CALIBRATION button and select CORRECTION DISABLED button.(as in Figure 11.1.2.1)
Ø Click on Apply, a success message will be displayed as shown in figure11.1.2.2
Ø Click on Ok. Ø Capture the image.
Ø Now again click on SET UP TOOLBAR, select DISPLAY SETUP radio button, click on ENABLE CALIBRATION button and select CORRECTION ENABLED button(as in Figure 11.1.2.3)
Figure 11.1.2.1
Figure 11.1.2.2
Figure 11.1.2.3
Company Confidential Page 56 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
Ø Click on Apply, a success message will be displayed as shown in figure 11.1.2.4
Ø Click on Ok.
Ø Capture the image of the same phantom.
Ø Now analyze the image captured.
Ø The image captured with correction disabled will
have some difference in the pixel values from one end to the other
Ø The image captured with the correction enabled
will have almost same pixel values at both the end. This shows that the calibration process was successful.
Figure 11.1.2.4
Company Confidential Page 57 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
11.2. Appendix B: Technical Specification
DAS (DATA ACQUISATION SYSTEM)
Figure 11.2.1
Company Confidential Page 58 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
PARAMETER CONDITION SPECIFICATION UNIT
Scanning 1 MHz
Frame readout time 1.62 s
Data Transfer time 1.61 s
Total Transfer time 3.2 s
AD conversion resolution 4095ADU 12 bit
Conversion gain 48.8 e/ADU
Readout noise 7 ADU
Dynamic Range 585
Memory Size 4 MByte
Power Consumption 350max mA
Storage Temperature -20 to +70 °C
Operating Temperature 0 to +50 °C
Operating Humidity No condensation 70% Max
Dimension 32(H)x115(W)x125(D) mm
Weight 200 g
Company Confidential Page 59 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
SENSOR
Figure 11.2.2
Company Confidential Page 60 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
§ SENSOR DETAILS
Type No Cooling No of total Pixels
No of Active Pixels
Active Area [mm(H) x mm(V)]
S8981-02 Non-cooled 1508 x 1002 1500 x 1000 30 x 20
§ GENERAL RATINGS
Parameter Specification
CCD Structure Full frame transfer
Fill factor 100%
Number of Active Pixels 1500 (H) x 1000 (V)
Pixel Size 20 (H) x 20 (V) µm
Active Area 30 (H) x 20 (V) µm
Vertical Clock Phase 2 phase
Horizontal Clock Phase 2 phase
Output Circuit Emitter follower without resistance
Dimensional Outline 38.7 (H) x 26.3 (V) mm
Reliability 100,000 shots at 60 kVp, 30 m Roentgen
Window Scintillator on 1.5mm FOP
Storage Temperature -20 to +70 °C
Operating Temperature 0 to +40 °C
Company Confidential Page 61 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].
11.3. Appendix C: If you need Assistance
The IntraSkanDigi is designed to provide years of trouble-free service, and carries a limited warranty. If you need assistance, you can contact the Skanray office nearest to you.
Mysore Skanray Technologies Pvt Ltd., #78 - A & B, Hootagalli Industrial Area, Mysore - 570 018. India. E-Mail: [email protected] Phone: +91 821 241 5559 Fax: +91 821 240 3344 E-Mail: For products: [email protected] For services: [email protected] For contacting the office: [email protected] Overseas Contacts United States: 4 Lakeview, Irvine, CA92604, USA. Phone: 1-401-854-6174 E-Mail: [email protected]
Company Confidential Page 62 of 62 All information contained in this document and accompanying documents is confidential and is the property of SKANRAY TECHNOLOGIES and cannot be divulged without the prior written consent from an authorized representative of SKANRAY TECHNOLOGIES. If this document was not intended for you please inform the company at [email protected].