Intox 8000 Summit
-
Upload
rakesh-srivastava -
Category
Documents
-
view
226 -
download
0
Transcript of Intox 8000 Summit
-
7/29/2019 Intox 8000 Summit
1/39
Both instruments
in Oklahoma test
Breath using
INFRAREDSPECTROSCOPY
-
7/29/2019 Intox 8000 Summit
2/39
INFRARED
SPECTROSCOPY The ethanol molecule has two ends, the
hydrogen end and the hydroxide (carbon-
oxygen bond) end.
NOW HOLD THAT THOUGHT . . . .
-
7/29/2019 Intox 8000 Summit
3/39
In the visual spectrum:We can tell how much fog or smoke is in the air by
how much light is blocked from a light source.
-
7/29/2019 Intox 8000 Summit
4/39
Ethanol Infrared Spectra with notation for I8000
IR filters
0%
10%
20%
30%
40%
50%
60%
70%
80%
90%
100%
3 4 5 6 7 8 9 10 11
Wavelength(microns)
PercentAbsorbtio
3.4 micron filter
9.4 micron filter
Thus, each molecule has its own infrared fingerprint.
Ethanol molecules absorb infrared light at wavelengths of approximately 3.00, 3.39,
7.25, 9.18, 9.50 and 11.5 microns.
No other compound absorbs infrared light at all of these wavelengths exclusively.
Each molecule absorbs infrared light waves in a unique way.
BEER-LAMBERT LAW(Provides the theoretical basis for IR breath testing.)
-
7/29/2019 Intox 8000 Summit
5/39
In the Infrared spectrum, the
hydrogen endand the hydroxide
endact like the fog to block the the
light BUT ONLY in the 3.4 uM
and9.4 uM wavelengths,exclusively.
3.4hhhhhhh9.4hhhhhhhhhhh
-
7/29/2019 Intox 8000 Summit
6/39
EVERYTHING ELSE PASSES
THROUGH
hhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhh
hhhhhhhhhhhhhhhhhhhhhhhhh3.4hhhhhh
9.4hhhhhhhhhhh
-
7/29/2019 Intox 8000 Summit
7/39
THE INSTRUMENT
MEASURES THE INTENSITY
OF THE INFRARED LIGHTWITH CLEAN ROOM AIR.
hhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhh
hhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhh
hhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhh
hhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhh
hhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhhh
-
7/29/2019 Intox 8000 Summit
8/39
The Instrument then compares
the breath sample to determine
how much of the 3.4 & 9.4uM aremissing (blocked by ethanol)
hhhhhhhhhhhhhhhhhhhhhhhhhh
hhhhhhhhhhhhhhhhhhhhhhhhhh
-
7/29/2019 Intox 8000 Summit
9/39
THE DIFFERENCE, OR MISSING IR
WAVE INTENSITY CAN BEMEASURED AND IS DUE TO THE
BREATH ALCOHOL CONTENT
-
7/29/2019 Intox 8000 Summit
10/39
Ethanol Infrared Spectra with notation for I8000
IR filters
0%
10%
20%
30%
40%
50%
60%
70%
80%
90%
100%
3 4 5 6 7 8 9 10 11
Wavelength(microns)
PercentAbsorbtio
3.4 micron filter
9.4 micron filter
Thus, each molecule has its own infrared fingerprint.
Ethanol molecules absorb infrared light at wavelengths of approximately 3.00, 3.39,
7.25, 9.18, 9.50 and 11.5 microns.
No other compound absorbs infrared light at all of these wavelengths exclusively.
Each molecule absorbs infrared light waves in a unique way.
BEER-LAMBERT LAW(Provides the theoretical basis for IR breath testing.)
-
7/29/2019 Intox 8000 Summit
11/39
IFthey are not the same, by a 5%
difference,THENthe instrument willdetermine a substance other than
ethanol, an interferent, is in thesample. The measurements of the
3.4uM and 9.4uM wavelength filters
must match or the substance is an
interferent.
-
7/29/2019 Intox 8000 Summit
12/39
SEE?
-
7/29/2019 Intox 8000 Summit
13/39
Intoxilyzer 8000
The 8000 operation is similar to the 5000with the following enhancements
ACCESS CARD with bar code - used to
scan in operator informationAble to scan the bar code on DL for subject
information
No Specialist for the 8000 No Log Of Tests now on each Affidavit
-
7/29/2019 Intox 8000 Summit
14/39
Intoxilyzer 8000
Uses dry gas instead of simulator solutionAble to run approx 325 test before dry gas
cylinder needs to be changed or approx 2
years before cylinder expiration date Retains complete Affidavit information
Sends Affidavit information to BOT and
LEADRS
-
7/29/2019 Intox 8000 Summit
15/39
Intoxilyzer 8000
To perform a test on the Intoxilyzer 8000 anofficer must have a valid permit.
The first thing he will do is scan his ACCESS cardand put in his PIN or manually type in the
information if he does not have his ACCESScard.
The 8000 will check an internal data base toverify if your permit is valid, if not you will not be
able to operate the instrument. It takes about fifteen minutes to delete an
operator from the database across the state if heis disqualified for any reason.
-
7/29/2019 Intox 8000 Summit
16/39
Access Card information The officers nolonger have to type this information.
LAST, FIRST, MIDDLE NAME
BADGE NUMBER PERMIT NUMBER
AGENCY
AGENCY ADDRESS, ZIP, PHONE PERSONAL IDENTIFICATION NUMBER (PIN)
(verify correct operator for digital signature)
INTOXILYZER 8000
-
7/29/2019 Intox 8000 Summit
17/39
The officers nolonger type in thefollowinginformation
LAST NAME FIRST NAME
MIDDLE NAME
DATE OF BIRTH SEX M/F
ADDRESS
CITY
INTOXILYZER 8000
STATE
ZIP
D.L. NUMBER
STATE OF ISSUE
REVIEW Y/N
-
7/29/2019 Intox 8000 Summit
18/39
-
7/29/2019 Intox 8000 Summit
19/39
BREATH TEST9 STEP SEQUENCE
AIR BLANK
BREATH SAMPLE
AIR BLANK
WAIT PERIODAIR BLANK
BREATH SAMPLE
AIR BLANKCONTROL CHECK
AIR BLANK
BREATH TEST SEQUENCE
-
7/29/2019 Intox 8000 Summit
20/39
BREATH TEST SEQUENCE
IF THE INTOXILYZER DETECTS RADIO
FREQUENCY IN CLOSE PROXIMITY
Cell phones within 6 feet
THE SCREEN WILL DISPLAY AND YOU
WILL GET A PRINTOUT STATING
-
7/29/2019 Intox 8000 Summit
21/39
BREATH TEST SEQUENCE IF
YOU PUSH THE START BUTTON AGAIN
THE SCREEN WILL DISPLAY AND YOUWILL GET A PRINTOUT STATING
SEQUENCE ABORTED
-
7/29/2019 Intox 8000 Summit
22/39
BREATH TEST SEQUENCE IF
THE SUBJECT BLOWS DURING AN AIR BLANK
THE SCREEN WILL DISPLAY AND YOUWILL GET A PRINTOUT STATING
IMPROPER SAMPLE
-
7/29/2019 Intox 8000 Summit
23/39
BREATH TEST SEQUENCE IF the subject has substantial mouth alcohol or
THE SUBJECT BLOWS A BREATH SAMPLE THATIS ABOVE THE RANGE OF THE INTOXILYZER, 0.60
THE SCREEN WILL DISPLAY AND YOUWILL GET A PRINTOUT STATING
RANGE EXCEEDED
-
7/29/2019 Intox 8000 Summit
24/39
BREATH TEST SEQUENCE IF THE SUBJECT DOESNT BLOW AN
ADEQUATE SAMPLE, STOPS TOO SOON,STARTS AND STOPS, OR BLOWS AND SUCKS
THE SCREEN WILL CONTINUE TO PROMPT PLEASEBLOW FOR 3 MINUTES AFTER WHICH YOU WILL GET
A PRINTOUT STATING
DEFICIENT SAMPLE
-
7/29/2019 Intox 8000 Summit
25/39
BREATH TEST SEQUENCE
IF
THE AIR BLANK IS CONTAMINATED
THE SCREEN WILL DISPLAY AND YOU
WILL GET A PRINTOUT STATING
AMBIENT FAIL (PURGE FAIL)
BREATH TEST SEQUENCE
-
7/29/2019 Intox 8000 Summit
26/39
BREATH TEST SEQUENCE
IF THE BREATH SAMPLE
IS CONTAMINATED ORTHE SUBJECT PUFFSRAPIDLY
THE SCREEN WILL WILL
DISPLAY
SAMPLE REJECTED
THE SCREEN WILL THEN PROMPT
RETRY Y/N
SELECTING Y WILL START A 16 MINUTE
DEPRIVATION CLOCK.
-
7/29/2019 Intox 8000 Summit
27/39
BREATH TEST SEQUENCE IF THE BREATH SAMPLES ARE
MORE THAN .03 APART, THETEST IS NOT VALID AND
THE INTOXILYZER WILL PROMPT RETRY Y/N
YOU CAN RETRY UP TO 3 TIMES WITHOUT HAVING
TO RE-ENTER THE SUBJECT INFORMATION.
THIS IS ANOTHER PROTECTION AGAINST TESTING
MOUTH ALCOHOL.
-
7/29/2019 Intox 8000 Summit
28/39
BREATH TEST SEQUENCE IF THE CONTROL TEST IS BEYOND THE
.01 TOLERANCE
THE INTOXILYZER WILL DISPLAY
Calibration Check Out of Tolerance THE PRINTOUT WILL SHOW THE RESULTS
WITH THE CONTROL TEST RESULT FLAGGED
INVALID AND
-
7/29/2019 Intox 8000 Summit
29/39
BREATH TEST SEQUENCE
THE INTRUMENT WILL BEDISABLED
THE TEST MUST BE ADMINISTERED
ON A DIFFERENT INTOXILYZER ORCONDUCT A BLOOD TEST
-
7/29/2019 Intox 8000 Summit
30/39
BREATH TEST SEQUENCE
IF THE TESTING SEQUENCE FAILS, ANYPRINTOUT IS EVIDENCE OF WHATHAPPENED DURING THE TEST.
THE SUBJECT SHOULD RECEIVE A COPY
A COPY SHOULD GO TO EVERYONE WHONORMALLY GETS A COPY OF THE THEAFFIDAVIT
THE PRINTOUTS ARE SAVED IN THEMEMORY OF THE INSTRUMENT AND AREIMMEDIATELY UPLOADED TO THE BOARDCOMPUTER IN OKLAHOMA CITY. YOU WILLBE ABLE TO GET THEM.
-
7/29/2019 Intox 8000 Summit
31/39
BREATH TEST SEQUENCE
IF THE TESTING SEQUENCECOMPLETES SUCCESSFULLY THEOFFICERS AFFIDAVIT AND NOTICE OFREVOCATION/DISQUALIFICATION ISSAVED THE SAME WAY.
THEY ARE NEVER PURGED FROM THE
MAIN COMPUTER. YOU WOULD BEABLE TO GET A PRINTOUT TEN YEARSFROM NOW. (It may cost you.)
-
7/29/2019 Intox 8000 Summit
32/39
The 8000
Intoxilyzer
Temporary
Driver License
-
7/29/2019 Intox 8000 Summit
33/39
OFFICERS
AFFIDAVITDISTRIBUTION
On the Officer/DPSAffidavit the officer will
complete SECTIONS 2,3 & 4. SECTION 4 doesNOT require anotary.
-
7/29/2019 Intox 8000 Summit
34/39
The 8000
Intoxilyzer Log
of Tests is the
Officer
Affidavit
-
7/29/2019 Intox 8000 Summit
35/39
IMPLIED CONSENTCIVIL ACTION NOT CRIMINAL BURDEN OF PROOF
Breath Test Operating Procedure Push the GREEN button andfollow the prompts.
If an Affidavit prints, the Officerfollowed the procedure approvedby the Board of Test. If a testresult appears, it is a valid,
states test even if the result is0.00.
-
7/29/2019 Intox 8000 Summit
36/39
ADMISSIBILITY 47 O.S. sec. 756 A (brutally edited)
Upon the trial of any proceeding out ofacts committed by any person driving a
motor vehicle while under the influence
of alcohol, evidence of the alcoholconcentration in the breath of the
person as shown by analysis of the
breath performed in accordance with
Sections 752 and 759 of [title 47] IS
admissible.
-
7/29/2019 Intox 8000 Summit
37/39
COMPLIANCE WITH SS 752 & 759
47 O.S. 752 H (brutally edited)
Tests breath for determining the alcohol
concentration, under the provisions of
[O.S. Title 47], to be admissible in evidenceunder the provisions of this [O.S. Title 47],
shall have been administered in accordance
with the rules and regulations of the Board.
-
7/29/2019 Intox 8000 Summit
38/39
The breath tube on the I8000 is shorterthan the 5000D and the heater may have
failed. Something in the sample chamber could
increase the reading by blocking the infra-red light.
Mouth alcohol was measured in the breathsample.
COMMON RED HERRINGS
-
7/29/2019 Intox 8000 Summit
39/39
No one can review the source code.
The particulate screen can clog.
The driver can affect the test result byhyperventilating.
RFI may have affected the result.
Another gaseous substance in the breath may
raise the result interferents.
COMMON RED HERRINGS