The Interster Bowie & Dick Autoclave Test Pack aligns with the requirements of AS4187 as a steam penetration, air removal test. Result record sheet included.
PRODUCT FEATURES o 7kg mimic o Does not contain heavy metals o Kitemark - BS EN-ISO 11140-4 o Conforms to EN-ISO 11140-1 (Type 2) o Process indicator on pack conforms to
EN-ISO 11140-1 (Type 1) o Contains a specially developed indicator ink which
is affixed through a silkscreen printing process
SUITABLE FOR o Autoclave sterilisers
REORDER FEATURE PCS/UNIT UNITS/CTN
3FSKS610860 BSI Kitemark 40 1
INTERSTER BOWIE & DICK AUTOCLAVE TEST PACK - KITEMARK
SMHO
-All-
BRO
-014
2_RE
V3_0
6_20
All p
rodu
ct in
form
ation
is c
orre
ct a
t dat
e of
pub
licati
on
NONSTERILE
LATEX
-
Test results Bowie & Dick Type Autoclave Lean Pack:
Before use
Not properly changed sheetA clearly defined mark is visible.Explanation: presence of air or non-condensable gases. Possible causes: poor air removal, leakage or gas in the steam and sub-atmospheric pulsing.
Completely changed sheetMinimum even colour change that must be achieved.
Not properly changed sheetAn indistinct mark, almost always in the middle, is visible.Explanation: superheated steam. Possible causes: the test pack has been placed in the sterilizer without immediately starting the process(desiccation), too high steam velocities or steam supply pressure.
Not properly changed sheetPossible causes: a too short sterilization time or too low temperature.
Not properly changed sheetCompletely changed sheet except for the corners. Possible causes: wet steam.
Do not use the above shown pictures as a reference colour
Interster International B.V.Samsonweg 2, 1521 RC Wormerveer T: +31 75 627 22 41 - E: [email protected]
www.interster.nl/exportrange
Art. no. 3FSKS610860 - 3,5 min.