Calibration technology
Integrated calibration systemfor electric instrumentation Model CBS-SC5000
Calibration bench, model CBS-SC5000
Applications
■ Calibration and testing electrical equipment and electrical/electronic instrumentation
Main features
■ 19” modular construction for table operation; modules may be joined (or splitted) at any time with minor modifi cations
■ Aluminium structure with painted steel plates on sides, top and rear; front panels, are anodised aluminium with engraved mimics.
■ Air set for cleaning purpose. including pressure reducer, gauge and fl exible hose with air jet.
Description
General InformationThe calibration system CBS-SC5000 is suitable to check electric/electronic instruments and devices, providing the operator with a wide variety of AC and DC supplies and facilities to enable accurate measurements of the most common electric parameters. Instruments checked with system CBS-SC5000 includes DC powered electronic transmitters/converters, AC powered controllers, recorders and AC / DC devices like pressure and temperature switches, solenoid valves, relays, etc. The below table summarizes voltage outputs withtheir main features:
Air set for cleaning purpose. including pressure reducer, gauge and fl exible hose with air jet.Operator’s safety improved with:
■ earth leakage circuit breaker ■ protective covers over internal live connections ■ totally enclosed bottom
All components comply with international standards.
The left module contains most of the AC supply facilities. The upper part accommodates the main stabilizer, with 4.5 digits indicator for input/output voltages. The middle part is reserved to the variable AC supply, including the manual control, digital ammeter voltmeter and the automatic switch protected output socket. The lower part output sockets for isolated supplies. The base is equipped with an air circuit for cleaning purposes: pressure reducer, gauge and fl exible hose with air jet.
The right module contains in the upper part a 6.5 digits LCD multimeter to perform accurate measurements. The middle part provides low level supplies. AC and rectifi ed DC for those devices not requiring stabilised supply like relays, solenoid valves. etc. Adjustable DC stabilised supplies are located just below, together with the continuity check circuit suitable to test on/off equipment operation. The base plate includes four AC sockets connected to the main switch.
WIKA data sheet CT 92.34
Page 1 of 5WIKA data sheet CT 92.34 ∙ 03/2015
Range Power Variable NotesAC 0 ... 260 V 10 A Yes -
AC 230 V 2.5 kVA Yes StabilisedAC 220 V 2 A No IsolatedAC 115 V 2 A No Isolated
DC -30 ... 0 ... 30 V 3 A Yes StabilisedDC 12, 24, 48 V 150 VA 1) No Rectifi edDC 12, 24, 48 V 150 VA 1) No -
1) each output
The calibration system CBS-SC5000 is a 19” modular construction for table operation; modules may be joined (or splitted) at any time with minor modifi cations.
Page 2 of 5 WIKA data sheet CT 92.34 ∙ 03/2015
Specifications
Specifications
1) Stabilised AC supply
Input AC 230 V ±10 % 50 Hz
Output AC 230 V ±10 % adjustable
Stability ±1.5 % (0/100 % load)
Output power 2.5 KVA max.
Harmonic distortion 0.2% max.
Reading Digital indicator, range AC 0 ... 300 V, 4.5 digits LE display, accuracy ±0.1 % of reading ±0.5 V
2) Variable AC supply
Input AC 230 V ±10 % 50 Hz
Output Variable between AC 0 ... 260 V
Current 4A max.
Indication Two digital indicators, 3 digits display, voltmeter and ammeter
3) Isolated AC supply source
Input AC 230 V ±10% 50 Hz
Output 1 AC 115 V, 2A with socket to CEE standards
Output 2 AC 230 V, 2A with socket to DIN/VDE standards
4) Air panel
Input 7 bar
Output The circuit includes a reducing valve, a pressure gauge and a flexible hose with air jet for cleaning purposes
1
2
3
4
5
6
7
8
Page 3 of 5WIKA data sheet CT 92.34 ∙ 03/2015
5) Precision multimeter
Display 6.5-digit resolution VFD dot matrix
Output DC V, AC V, resistance, DC A, AC A, frequency/period
6) AC and DC power supply
Input From the variable AC supply
Rated outputs 1) AC 12 V / DC 12 V - 10A max.
AC 24 V / DC 24 V - 6A max.
AC 48 V / DC 48V - 3A max.1) With input voltage AC 260 V 50 Hz, smoothed but not stabilised outputs.
7) DC stabilised voltage source
Output voltage DC 0 ... 30 V CH1 / CH2
Output vurrent DC 0 ... 3 A CH1 / CH2
Tracking series voltage DC 0 ... 60 V
Tracking series current DC 0 ... 6 A.
Display 3-digit 0.5” LE display, accuracy ±0.5 % of rdg + 2 digitsSwitch check continuity circuit, including buzzer and terminals, for checking operation of pressure switches and
other equipment with NO/NC contacts
8) System mains switch and AC sockets
Input AC 230 V ±10 % 50 Hz
Switch rating AC 230 V - 32A
Output Four sockets connected to the main switch and available with socket to DIN/VDE standards
Power The total power is limited by the system main switch rating
Mechanical The system CBS-SC5000 housing is composed by two 19’’ module cabinet. Steel structure, finely finished In light grey RAL 7032; mimic panels engraved and colored on anodised aluminum plates
Dimensions Front 554 mm
■ Depth 450 mm ■ Height 711 mm
Electrical safety
The system CBS-SC5000 is featured for the electrical safety, including:
■ Main switch ■ Earth leakage circuit breaker to standard IEC-1010-
1 for personnel and fire protection by detecting an earth leakage fault current in excess of 30 mA and disconnecting the main supply within 200 ms
■ ‘Double isolation’ protection in connecting the external main supply to the system main switch, by completely enclosing live connections in an insulating case
■ Protective transparent covers over live connections inside the equipment
Air External “instrument air” (clean, dry and oil free) supply is required, at 7 bar pressure.The circuit includes a reducing valve, a pressure gauge and a flexible hose with air jet for cleaning purposes.
Power requirement
■ Main supply AC 230 V ±10 % 50/60 Hz ■ Main switch rating: 32A ■ System operation: 6.5 kVA (maximum) ■ AC sockets rating: 16 A
Operating Environment
■ Specified for 0°C ... 50°C ■ Specified to 80 % r.h. at 35 °C, not condensing ■ Storage environment: -40° ... 70 °C
Page 4 of 5 WIKA data sheet CT 92.34 ∙ 03/2015
Dimensions in mm
WIKA Italia Srl & C. Sas Via G. Marconi, 820020 Arese (Milano)/ItaliaTel. +39 02 93861-1Fax +39 02 [email protected]
© 2015 WIKA Alexander Wiegand SE & Co. KG, all rights reserved.The specifications given in this document represent the state of engineering at the time of publishing.We reserve the right to make modifications to the specifications and materials.
Page 5 of 5 WIKA data sheet CT 92.34 ∙ 03/2015
03/2
015
EN
Supply connections
■ Main Power cable gland 16 mm diameter for 3-wire 4 mm2 cross section cable. Located at the rear panel
■ Air ¼” BSP female for 10 bar maximum, located at the rear panel
Standard Certification
■ Certificate of conformity. ■ Calibration reports
Certification on request
■ Certificate of calibration, against reference masters, trace-able to National and International Standards (E.A.).
■ Official certificates issued by accredited laboratories of the Italian National Service (ACCREDIA) membership of the E.A. (European Accreditation Laboratory).
Standard supply
■ Leads for the multimeter ■ One plug for each socket ■ Air gun with flexible tube and quick-fit connection (code
5040) ■ Instruction manual in one copy ■ Kit of fuses flexible hose with air jet for cleaning purposes.
Technical documentation
Two sets of the Installation, operating and maintenance manual.