Instrument Calibration Record for Fluke Thermocouple Probe, … · 2012. 11. 30. · t. SOUTHWEST...

2
t. SOUTHWEST RESEARCH INSTITUTE@ 6220 Culebra Road, P.O. Drawer 28510 Institute Quality Systems Institute Calibration Laboratory Phone: 210-522-5215 Fax 210-522-4834 Certificate of Calibration Submitted By: DIV20 Address: B57 Contact: DON BANNON Manufacturer I Model: FLUKE J 80PK-2A Serial No: 10637 Asset No: 010637 Description: THERMOCOUPLE PROBE Procedure: TEMPERATURE PROBES - 5 JUN 06 Calibration Laboratory Certificate #0972-01 Work Order: 303078039 Date Issued: Jan 11,2008 Calibration Date: Jan 11,2008 *Calibration Due: Jan 11,2009 Environment: Temp. 73.O"F Hum. 40 %RH **Data Type: FOUND-LEFT Calibration Location: Bldg. 64 DivIDLocation: N/A This certificate documents traceability to the National Institute of Standards and Technology (NIST) and the International System of Units (SI). The Laboratory quality system conforms to ISO/IEC 17025,2005,ANSWCSL 2540-1-1994 and relevant requirements of the IS0 9000-2000 standard. This certificate shall not be reproduced, except in full, without the written approval of the Southwest Research Institute Calibration Laboratory. This certificate shall not be used to claim product endorsement by Southwest Research Institute, American Association for Laboratory Accreditation (A2LA) or any agency of the U. S. Government. Results ofthis calibration relate only to the instrument described above at the time of calibration and does not imply any long term stability of the instrument. *Determined by the customer, does not imply the instrument will remain within tolerance as any number of factors may cause an out-of-tolerance condition before this date. **Found/Left = adjustment and/or repair was not required, As Left = adjusted and/or repaired was required. The client has sole responsibility for determination of in-/out-of-tolerance or compliance/noncompliance. See Remarks or attached Measurement Report with the same Work Order number for data. Reported uncertainty calculated in accordance with the IS0 "Guide to the Expression of Uncertainty in Measurement" (GUM) and represents an expanded uncertainty with a coverage factor of k=2 to approximate a 95% confidence level. Remarks: Cal -20" to 150°C Standards Used 4sset No. Serial No. Manufacturer Model Description Cal Due 109137 A21208 HART SCIENTIFIC 1575 SUPER THERMOMETER Jun l7,08 110814 A44625 HART SCIENTIFIC 1529 THERMOCOUPLE THERMOMETER Oct 16,08 )I1161 0202 HART SCIENTIFIC 5699 SPRT Sep 13,08 -1 , ( ) wgh (d srk ( ) jrg ( ) blt ( ) pwc +& \ ; , & , leviewed by. detrology Technician nb2lal rpt Rev date August 15,2005 - Measurements by: Bob Trollinger // I Y Metrology Technician Paee 1 of 1

Transcript of Instrument Calibration Record for Fluke Thermocouple Probe, … · 2012. 11. 30. · t. SOUTHWEST...

Page 1: Instrument Calibration Record for Fluke Thermocouple Probe, … · 2012. 11. 30. · t. SOUTHWEST RESEARCH INSTITUTE@ 6220 Culebra Road, P.O. Drawer 28510 Institute Quality Systems

t .

S O U T H W E S T R E S E A R C H I N S T I T U T E @ 6220 Cu leb ra R o a d , P.O. Drawer 2 8 5 1 0

Insti tute Qual i ty Sys t ems Insti tute Cal ibrat ion Laboratory

Phone : 210-522-5215 Fax 210-522-4834

Certificate of Calibration

Submitted By: DIV20 Address: B57 Contact: DON BANNON

Manufacturer I Model: FLUKE J 80PK-2A

Serial No: 10637 Asset No: 010637

Description: THERMOCOUPLE PROBE

Procedure: TEMPERATURE PROBES - 5 JUN 06

Calibration Laboratory Certificate #0972-01

Work Order: 303078039 Date Issued: Jan 1 1,2008

Calibration Date: Jan 1 1,2008 *Calibration Due: Jan 11,2009

Environment: Temp. 73.O"F Hum. 40 %RH **Data Type: FOUND-LEFT

Calibration Location: Bldg. 64

DivIDLocation: N/A

This certificate documents traceability to the National Institute of Standards and Technology (NIST) and the International System of Units (SI). The Laboratory quality system conforms to ISO/IEC 17025,2005, ANSWCSL 2540-1-1994 and relevant requirements of the IS0 9000-2000 standard. This certificate shall not be reproduced, except in full, without the written approval of the Southwest Research Institute Calibration Laboratory. This certificate shall not be used to claim product endorsement by Southwest Research Institute, American Association for Laboratory Accreditation (A2LA) or any agency of the U. S. Government. Results ofthis calibration relate only to the instrument described above at the time of calibration and does not imply any long term stability of the instrument.

*Determined by the customer, does not imply the instrument will remain within tolerance as any number of factors may cause an out-of-tolerance condition before this date. **Found/Left = adjustment and/or repair was not required, As Left = adjusted and/or repaired was required. The client has sole responsibility for determination of in-/out-of-tolerance or compliance/noncompliance. See Remarks or attached Measurement Report with the same Work Order number for data.

Reported uncertainty calculated in accordance with the IS0 "Guide to the Expression of Uncertainty in Measurement" (GUM) and represents an expanded uncertainty with a coverage factor of k=2 to approximate a 95% confidence level. Remarks: Cal -20" to 150°C

Standards Used 4sset No. Serial No. Manufacturer Model Description Cal Due

109137 A21208 HART SCIENTIFIC 1575 SUPER THERMOMETER Jun l 7 , 0 8

110814 A44625 HART SCIENTIFIC 1529 THERMOCOUPLE THERMOMETER Oct 16,08

)I1161 0202 HART SCIENTIFIC 5699 SPRT Sep 13,08

-1 ,

( ) wgh (d srk ( ) jrg ( ) blt ( ) pwc +& \;,&,

leviewed by. detrology Technician nb2lal rpt Rev date August 15,2005

- Measurements by: Bob Trollinger //

IY Metrology Technician Paee 1 of 1

Page 2: Instrument Calibration Record for Fluke Thermocouple Probe, … · 2012. 11. 30. · t. SOUTHWEST RESEARCH INSTITUTE@ 6220 Culebra Road, P.O. Drawer 28510 Institute Quality Systems

Southwest Research Institute Calibration Laboratory Measurement Report

Limited Calibration -20 to 150 OC Type K

FunctionIRange Test Point TI Reading Difference +/-Limit +/-Uncertainty FoundlLeft

Tem~erature "C " C "C "C "C Result -1 9.751 -19.38 0.37 2.2 0.47 Pass 0.145 0.09 -0.05 2.2 0.47 Pass 50.073 50.54 0.46 2.2 0.47 Pass 100.068 100.34 0.27 2.2 0.47 Pass 150.200 150.77 0.57 2.2 0.47 Pass

END OF REPORT

Technician: blt

Cal Date: 14-Jan-08

Work Order: 303078039 Asset No: 01 0637 Serial No: 10637

Rev, Jan108

Remarks:

Mfr: Fluke Model: 80PK-2A Type: Immersion Probe

Page 1 of 1 Fluke 80PK-2A Immersion Probe.xls]