Instructions for Horiba Raman confocal imaging microscope · Raman confocal imaging microscope...

17
Dianwen Zhang (Email: [email protected]) Raman confocal imaging microscope Instructions for Horiba Raman confocal imaging microscope Imaging Technology Group Beckman Institute for Advanced Science and Technology University of Illinois at Urbana‐Champaign

Transcript of Instructions for Horiba Raman confocal imaging microscope · Raman confocal imaging microscope...

Page 1: Instructions for Horiba Raman confocal imaging microscope · Raman confocal imaging microscope Instructions for Horiba Raman confocal imaging microscope ... Close the Labspec software

DianwenZhang(Email:[email protected])

Raman confocal imaging microscope

Instructions for Horiba Raman confocal imaging microscope

ImagingTechnologyGroupBeckmanInstituteforAdvancedScienceandTechnology

UniversityofIllinoisatUrbana‐Champaign

Page 2: Instructions for Horiba Raman confocal imaging microscope · Raman confocal imaging microscope Instructions for Horiba Raman confocal imaging microscope ... Close the Labspec software

MicroscopySuite,ImagingTechnologyGroupBeckmanInstituteforAdvancedScienceandTechnologyUniversityofIllinoisatUrbana‐Champaign

OperationmanualofRamanconfocal

405NorthMathews,UrbanaIL61801Tel:217‐244‐0170;FAX:217‐244‐6219

http://itg.beckman.illinois.edu/

RevisionHistoryAuthorsandEditors RevisionDateDianwenZhang 4/20/2012DianwenZhang 12/22/2014DianwenZhang 06/28/2016DianwenZhang 05/10/2017

Page 3: Instructions for Horiba Raman confocal imaging microscope · Raman confocal imaging microscope Instructions for Horiba Raman confocal imaging microscope ... Close the Labspec software

MicroscopySuite,ImagingTechnologyGroupBeckmanInstituteforAdvancedScienceandTechnologyUniversityofIllinoisatUrbana‐Champaign

OperationmanualofRamanconfocal

405NorthMathews,UrbanaIL61801Tel:217‐244‐0170;FAX:217‐244‐6219

http://itg.beckman.illinois.edu/

TableofContents

TheRamanConfocalMicroscopeSystem:Overview..................................................................1 

StandardOperatingProcedure(SOP)forLasers.........................................................................2 

Laserspecifications.............................................................................................................................................2 

DuringnormaloperationsoftheLaser......................................................................................................3 

Laserstart‐upprocedure..................................................................................................................................4 

Lasershutdownprocedure..............................................................................................................................5 

Emergencyresponse..........................................................................................................................................5 

StartUpTheMicroscopeSystem.......................................................................................................6 

ShutDowntheMicroscopeSystem...................................................................................................8 

BasicLabspecSoftwareSettings........................................................................................................9 

GeneralSpectralAcquisitionProcedure......................................................................................10 

DefaultSystemConfigurationforGeneralUsers......................................................................12 

Troubleshooting...................................................................................................................................13 

Page 4: Instructions for Horiba Raman confocal imaging microscope · Raman confocal imaging microscope Instructions for Horiba Raman confocal imaging microscope ... Close the Labspec software

MicroscopySuite,ImagingTechnologyGroupBeckmanInstituteforAdvancedScienceandTechnologyUniversityofIllinoisatUrbana‐Champaign

OperationmanualofRamanconfocal

405NorthMathews,UrbanaIL61801Tel:217‐244‐0170;FAX:217‐244‐6219

http://itg.beckman.illinois.edu/

TheRamanConfocalMicroscopeSystem:Overview

Theleftrearview

Thefrontview

Synapsecamera

AndorNewtonEMCCDcamera

Halogenlampforepi‐illumination

Coarsefocusing

Finefocusing

Laserbutton

Objectiveturret

633nmlaser

830nmlaser 785 nmlaser

405 nmlaser

532 nmlaser

Page 5: Instructions for Horiba Raman confocal imaging microscope · Raman confocal imaging microscope Instructions for Horiba Raman confocal imaging microscope ... Close the Labspec software

MicroscopySuite,ImagingTechnologyGroupBeckmanInstituteforAdvancedScienceandTechnologyUniversityofIllinoisatUrbana‐Champaign

OperationmanualofRamanconfocal

405NorthMathews,UrbanaIL61801Tel:217‐244‐0170;FAX:217‐244‐6219

http://itg.beckman.illinois.edu/ ‐2‐

StandardOperatingProcedure(SOP)forLasers

Laserspecifications

Wavelength  405 nm (blue)  Max power  40 mW 

Laser Class  Class 3B  Laser Type  Continuous wave 

Manufacturer  ONDAX  Serial #  X6000 

Model  LM‐405‐PLR‐40‐2 

Wavelength  532 nm (green)  Max power  50 mW 

Laser Class  Class 3B  Laser Type  Continuous wave 

Manufacturer  Laser Quantum  Serial #  0420506‐82664 

Model  Ventus 532 50mW 

Wavelength  633 nm (red)  Max power 35 mW 

Laser Class  Class 3B  Laser Type  Continuous wave 

Manufacturer  Melles Griot  Serial #   

Model  05‐LHP‐928 

Wavelength  785 nm (invisible)  Max power  575 mW (set to 330 mW) 

Laser Class  Class 4  Laser Type  Continuous wave 

Manufacturer  Toptica Photonics AG  Serial #  XTRA_3V0_01113 

Model  XTRA II 

Wavelength  830 nm (invisible)  Max power  100 mW 

Laser Class  Class 3B  Laser Type  Continuous wave 

Manufacturer  Sacher Lasertechnik group  Serial #  RL‐830‐0711‐0322 

Model  N/A 

Page 6: Instructions for Horiba Raman confocal imaging microscope · Raman confocal imaging microscope Instructions for Horiba Raman confocal imaging microscope ... Close the Labspec software

MicroscopySuite,ImagingTechnologyGroupBeckmanInstituteforAdvancedScienceandTechnologyUniversityofIllinoisatUrbana‐Champaign

OperationmanualofRamanconfocal

405NorthMathews,UrbanaIL61801Tel:217‐244‐0170;FAX:217‐244‐6219

http://itg.beckman.illinois.edu/ ‐3‐

DuringnormaloperationsoftheLaserTheRamanconfocalisaHoribaLABRAMproductandaclass3Blaserproduct,whichmeans

Laserbeamsandreflectedbeamscouldbedangerous.Diffusedreflectedbeamsinmostcasesdonotpresentanyproblem.

Lasersarehighintensitylightsourcesproducingvisibleorinvisiblelight(forexample,785nmand830nmlasersinthesystem)atspecificwavelengths.Thisconcentratedenergyinanarrowlaserbeammaycausedamagetobiologicaltissues,especiallytoeyes.Allbelowlabelsonthemicroscopeindicateahazardoussituationwherethereisariskofseriousinjurytobecausedbylaserradiation.Specialattentionmustbepaid.

Assuredoorremainsclosedandlockedfromtheoutsidewhenlaserisinuse.Theexterior“InUse”lightmustbeturnedonwheneverthelaserisinoperation

Beforeremovingorchangingobjectives,thelaserbuttononthemicroscopefrontpanelmustbein“off”position.Incaseofremovingobjectives,theemptyobjectivepositionsontheobjectiveturretmustbecoveredwithobjectivecaps.

Page 7: Instructions for Horiba Raman confocal imaging microscope · Raman confocal imaging microscope Instructions for Horiba Raman confocal imaging microscope ... Close the Labspec software

MicroscopySuite,ImagingTechnologyGroupBeckmanInstituteforAdvancedScienceandTechnologyUniversityofIllinoisatUrbana‐Champaign

OperationmanualofRamanconfocal

405NorthMathews,UrbanaIL61801Tel:217‐244‐0170;FAX:217‐244‐6219

http://itg.beckman.illinois.edu/ ‐4‐

Whenusing785nmand830nmlasers(invisible),allusersarerequiredtowearlasergogglesthroughoutexperiments.

Usershouldavoidwearingreflectiveaccessories/jewelry(e.g.watches).

Laserstart‐upprocedureThelasercontrollersareshowninFigure1.Onlyonelaserissupposedtobeusedatatime;

DoNOTadjustthesettingsonthelasercontrollers.

Forthe405nmlaser(Figure1(a)),plugthepowersupplytothenearbypoweroutlet.Forthe532nmlaser(Figure1(b)),turnonthekeyswitch,andpressthe“Enable”buttonwhenitstartsflashing.Forthe633nmlaser(Figure1(b)),turnonthekeyswitchonthelasercontroller.Forthe785nmlaser(Figure1(c)),turnonthekeyswitchlocatedontheopticaltableinfrontofthe830nmlasercontroller.Forthe830nmlaser(Figure1(b)),pressonthepowerbutton,waitfor30secondsandpresstheshutter“on/off”buttonon.

Figure1 (a)405nmlaserpowersupply;(b)532,633and830nmlasercontrollers;(c)785laser

Shutter

633nmlasercontroller

532nmlasercontroller

830nmlasercontrollerPower

(b)(a) 405nmlaserpowersupply

785nmlaser“on/off”switch

(c)

Page 8: Instructions for Horiba Raman confocal imaging microscope · Raman confocal imaging microscope Instructions for Horiba Raman confocal imaging microscope ... Close the Labspec software

MicroscopySuite,ImagingTechnologyGroupBeckmanInstituteforAdvancedScienceandTechnologyUniversityofIllinoisatUrbana‐Champaign

OperationmanualofRamanconfocal

405NorthMathews,UrbanaIL61801Tel:217‐244‐0170;FAX:217‐244‐6219

http://itg.beckman.illinois.edu/ ‐5‐

LasershutdownprocedureForthe405nmlaser(Figure1(a)),unplugthepowersupply.Forthe532nmlaser(Figure1(b)),turnoffthekeyswitchonthelasercontroller.Forthe633nmlaser(Figure1(b)),turnoffthekeyswitchonthelasercontroller.Forthe785nmlaser(Figure1(c)),turnoffthekeyswitchlocatedontheopticaltableinfrontofthe830nmlasercontroller.Forthe830nmlaser(Figure1(b)),pressoffthepowerbuttononthelasercontroller.

EmergencyresponseINEVENTOFEMERGENCY:

‐ Turnoffalllasersimmediately.‐ Dial911.‐ Ifalasereyeinjuryissuspected,havetheinjuredpersonkeeptheirheaduprightand

motionlesstorestrictbleedingintheeye.Aphysicianshouldevaluatelasereyeinjuriesassoonaspossible[1].

‐ Ifthereisafire,leavethearea,pullthefirealarm,andcontactthefiredepartmentbycalling911.Donotfightthefireunlessitisverysmallandyouhavebeentrainedinfire‐fightingtechniques[1].

‐ InformthePI,thegroupsafetyofficer,andthebuildingmanagerassoonaspossible.Ifthereisaninjury,thePIneedstosubmitareportofinjurytoRiskManagement.Provideasmanydetailsaspossible[1].

‐ InformtheDRSat217‐333‐2755assoonaspractical.Provideasmanydetailsaspossible[1].

References

[1]UIUCDRSLaserSafetyManual.http://www.drs.illinois.edu/site‐documents/LaserSafetyManual2014.pdf

Page 9: Instructions for Horiba Raman confocal imaging microscope · Raman confocal imaging microscope Instructions for Horiba Raman confocal imaging microscope ... Close the Labspec software

MicroscopySuite,ImagingTechnologyGroupBeckmanInstituteforAdvancedScienceandTechnologyUniversityofIllinoisatUrbana‐Champaign

OperationmanualofRamanconfocal

405NorthMathews,UrbanaIL61801Tel:217‐244‐0170;FAX:217‐244‐6219

http://itg.beckman.illinois.edu/ ‐6‐

StartUpTheMicroscopeSystem

1. Turnonthelaser.

2. SetupthelaserpathtomicroscopeFollowingthenoteontheyellowlabelbesideeachplunger(#1‐4inFigure2),setupthefourfrontplungerstoguidethelasertothelaserentranceportofthemicroscope.

Fortheuserswhouse633nmlaseranddon’tuseSuperhead,pleasealsocheckthefifthplunger(Figure3)inthesystem:

3. Checkthetwoopticalfiltersinsidetheslidingchamber(Figure4)ontopofthemicroscope.Ifthelabelsoftheinstalledfiltersdoesn’tmatchthelaserwavelength,changethem.

Plunger1

Laserentrance(ahole)tothemicroscope

Plunger3

Plunger2

Plunger4

Plunger5needstobepulledupfortheuseof633nmlaserwhenusingthemicroscope.

Figure2 Thefrontplungers

Figure3 Therearplunger

Page 10: Instructions for Horiba Raman confocal imaging microscope · Raman confocal imaging microscope Instructions for Horiba Raman confocal imaging microscope ... Close the Labspec software

MicroscopySuite,ImagingTechnologyGroupBeckmanInstituteforAdvancedScienceandTechnologyUniversityofIllinoisatUrbana‐Champaign

OperationmanualofRamanconfocal

405NorthMathews,UrbanaIL61801Tel:217‐244‐0170;FAX:217‐244‐6219

http://itg.beckman.illinois.edu/ ‐7‐

4. TurnontheLEDlampforepi‐illumination.

5. Openthecontrolsoftware–Labspecandloadsampleformeasurements.IfLabspecfreezesorgivesanerrormessage“ConnectionwithS3000fails,”followtheinstructioninthesectionofTroubleshootingtorestartthemicroscopesystem.TheLabspecsoftwaremanualcanbefoundinthefolderof“Manuals”onthewindowsdesktop.

Openthechambercarefully Carefullychangethesetwofiltersifnecessary.

Figure4Filterschamber

Figure5LEDlightforepi‐illuminationwhitelight

Page 11: Instructions for Horiba Raman confocal imaging microscope · Raman confocal imaging microscope Instructions for Horiba Raman confocal imaging microscope ... Close the Labspec software

MicroscopySuite,ImagingTechnologyGroupBeckmanInstituteforAdvancedScienceandTechnologyUniversityofIllinoisatUrbana‐Champaign

OperationmanualofRamanconfocal

405NorthMathews,UrbanaIL61801Tel:217‐244‐0170;FAX:217‐244‐6219

http://itg.beckman.illinois.edu/ ‐8‐

ShutDowntheMicroscopeSystem1. ClosetheLabspecsoftware,andlogoffthecomputer.2. DecreasetheLEDlightintensitytominimumandthenturnofftheLEDlamp(The

powerbuttonattherearofthelamp).3. Turnthe“Laser”buttononthemicroscopefrontpaneltoOFFposition.4. Turnoffthelasersthatyouhaveused.5. Cleanuptheworkingplace(Takeawayallyourstuffs;usedglassslidesorcoverslips

shouldbethrownintosharpswastebin;etc.).

Page 12: Instructions for Horiba Raman confocal imaging microscope · Raman confocal imaging microscope Instructions for Horiba Raman confocal imaging microscope ... Close the Labspec software

MicroscopySuite,ImagingTechnologyGroupBeckmanInstituteforAdvancedScienceandTechnologyUniversityofIllinoisatUrbana‐Champaign

OperationmanualofRamanconfocal

405NorthMathews,UrbanaIL61801Tel:217‐244‐0170;FAX:217‐244‐6219

http://itg.beckman.illinois.edu/ ‐9‐

BasicLabspecSoftwareSettings

1. Selecttherightlaserline.2. Selecttherightdensityfilter.ThenumbersareopticaldensityOD=Log(Power

transmissionfactor).Thedashlinesgivenodensityfilterinuse.3. Theconfocalpinhole(Hole)controlsmicroscopespatialresolution.Forbestoptical

resolution,setitto200‐300μm,orotherwiseusethemaximumvalue950μm.Fillinadesiredvalue,click“enter”onkeyboard.

4. Clicktheredthunderbuttontoinitializethe“Hole”.5. Theslitsize(Slit)determinesspectralresolution.Fillinadesiredvalue,click“enter”on

keyboard.6. Clicktheredthunderbuttontoinitializethe“Slit”.7. Selectthedesiredgrating.Bydefault,only300g/mmand1800g/mmareavailable.300

g/mmforbroadspectralrange,butpoorspectralresolution,while1800fornarrowspectralrange,butgoodspectralresolution.

8. Payattentiontotheunit.Followthebelowpicturetochangeitifnecessary.

9. Setthecentralvalueofthex‐scaleofthespectrum.Changeitforvaryingspectralrange

under“singlewindow”modespectralacquisition.Fillinadesiredvalue,click“enter”onkeyboard.

10. Clicktheredthunderbuttontoinitializethe“Spectrometer”.11. Selecttheobjectivelenstobeused.Itmustbesetrightbeforeyourmappingacquisition.

Alldefaultobjectivesarelongworkingdistance(LWD)lens.12. Theprefixofyourdatafilename(optional).13. Makesurethisis“XYZMotorized”.14. Thesethreeredthunderbuttonsallowyoutodefinethecurrentsamplepositiontobe

theoriginofcoordinatesinthesystem.Usuallyitisneededtoclickthebuttonforz‐axisbeforeyoudoz‐axisscanningmapping.

15. Theexposuretimeforrealtimespectralacquisition.Theunitisseconds.

1 2

3

74 6 10 1315

11 14

5 89171216

Page 13: Instructions for Horiba Raman confocal imaging microscope · Raman confocal imaging microscope Instructions for Horiba Raman confocal imaging microscope ... Close the Labspec software

MicroscopySuite,ImagingTechnologyGroupBeckmanInstituteforAdvancedScienceandTechnologyUniversityofIllinoisatUrbana‐Champaign

OperationmanualofRamanconfocal

405NorthMathews,UrbanaIL61801Tel:217‐244‐0170;FAX:217‐244‐6219

http://itg.beckman.illinois.edu/ ‐10‐

16. Theexposuretimeforsinglespectrumacquisitionormappingacquisition.Theunitisseconds.

17. Setthenumberofaccumulationsperspectralacquisition.

GeneralSpectralAcquisitionProcedure

1. Checktheabovementionedsettingsatthebottompanelofthesoftwareandmakesuretheyareallcorrect.

2. Checkthe“Laser”buttononthefrontpanelofthemicroscopeandmakesureitisonthe"OFF"position.

3. Startvideobyclicking .4. YoushouldbeabletoseetheLEDwhitelightshiningonyoursamplethroughthe

objecivelens.Adjustthemicroscopefocustogetyourinterestedsampleareaintothe

objectivefocalplaneandthenstopthevideobyclicking .Youcansavetheimageinthevideowindowtoa*.ngvorotherimageformatfile.

5. Turnoffthefluorescenceroomlight,butyoumaykeeptheHalogenceilinglighton.Nowyouarereadyforspectralacquisition.

a. Ifyouaregoingtodoreal‐timespectralacquisition ,turnonthe“Laser”buttononthemicroscopefrontpanel,click tostartitandclick tostopit.Youcansavetheacquiredspectrumtoa*.ngsfileandconvertittoa*.txtfile.

b. Ifyouaregoingtodosinglespectrumacquisition ,followthebelowsteps:1) Ifyouhavecheckedthe“extendedrangeacquisition ”andknownthere

isnothingneededtochange,skipthisstep.Orotherwiseclick ,selecteither“Singlewindow”or“Multiplewindows”mode.Ifyouaregoingtousethe“Multiplewindows”mode,youalsoneedtosetforaspecificspectralrange.

2) Turnonthe“Laser”buttononthemicroscopefrontpanel,click togetasinglespectrum.Everytimeyouclick ,yougetonemorespectrum.

3) Usingthemenu“File→SaveAll”,youcansaveallacquiredspectratoasingle*.ngs ile,oryoucanalsousethemenu“File→SaveAs”tosaveasingleselectedspectrumtoa*.ngsfileorconvertittoa*.txtfile.

c. Ifyouaregoingtodomappingacquisition ,followthebelowsteps:1) Settouse“SingleWindow”modeinthe“extendedrangeacquisition ”.2) Setthemappingproperty .3) Turnonthe“Laser”buttononthemicroscopefrontpanel,click tostart

mappingacquisition.

Page 14: Instructions for Horiba Raman confocal imaging microscope · Raman confocal imaging microscope Instructions for Horiba Raman confocal imaging microscope ... Close the Labspec software

MicroscopySuite,ImagingTechnologyGroupBeckmanInstituteforAdvancedScienceandTechnologyUniversityofIllinoisatUrbana‐Champaign

OperationmanualofRamanconfocal

405NorthMathews,UrbanaIL61801Tel:217‐244‐0170;FAX:217‐244‐6219

http://itg.beckman.illinois.edu/ ‐11‐

4) Yougetthreenewwindowsforthemappingacquisition.Youonlyneedtosavetheallspectradatawindow“SpIm”–clicktoactivethewindow,andsaveittoa*.ngcfile.

Page 15: Instructions for Horiba Raman confocal imaging microscope · Raman confocal imaging microscope Instructions for Horiba Raman confocal imaging microscope ... Close the Labspec software

MicroscopySuite,ImagingTechnologyGroupBeckmanInstituteforAdvancedScienceandTechnologyUniversityofIllinoisatUrbana‐Champaign

OperationmanualofRamanconfocal

405NorthMathews,UrbanaIL61801Tel:217‐244‐0170;FAX:217‐244‐6219

http://itg.beckman.illinois.edu/ ‐12‐

DefaultSystemConfigurationforGeneralUsers

PushintheDualScan/Standardplunger

Pullout–onlySynapseforgeneralusers

Pushinformicroscopemode

Usemicroscopemode

Optional

Turnitononlywhenlaserisneeded UseonlySynapse

forgeneralusers

UseonlyXYstageforgeneralusers

Optional

Page 16: Instructions for Horiba Raman confocal imaging microscope · Raman confocal imaging microscope Instructions for Horiba Raman confocal imaging microscope ... Close the Labspec software

MicroscopySuite,ImagingTechnologyGroupBeckmanInstituteforAdvancedScienceandTechnologyUniversityofIllinoisatUrbana‐Champaign

OperationmanualofRamanconfocal

405NorthMathews,UrbanaIL61801Tel:217‐244‐0170;FAX:217‐244‐6219

http://itg.beckman.illinois.edu/ ‐13‐

Troubleshooting1. Problem:WhenstartingLabspec,itbecomesfrozenorshowsthebelowerrormessage

beforethefinalmainwindowcanappear.

Solution:Followthebelowstepstorestartandresetthemicroscope.

1) CloseLabspec.SometimesyoualsoneedchecktheWindowsTaskManagertomakesurethatthereisnoaprocesscalled“NGSLabspec.exe”runninginbackground.

2) Restartthemicroscope.Findthebelowsocketandthemicroscopepowerboxbesidethelasercontrollersontheopticaltable.Turnoffthesocketswitchpointedbytheyellowarrow.Waitmorethan1minuteandthenturnitbackon.Nowthemicroscopeisrestarted.

3) StartLabspec.Labspecshouldrunnormally,andafterafewcommondevice

initialization,itwillshowthebelowstagecalibrationmessage.Beforeyouclick“OK”,makesurethatnosampleisloadedortheloadedsamplewon’thittheobjectiveswhenthestagemovinginitsfullrange.Click“OK”tostartthestagecalibration.

4) Initializethethreekeymicroscopecomponents‐“Hole,”“Slit”and

“Spectrometer”.Seethebelowimage.Clicktheredthunderbuttonatthetop‐rightonthe“Hole”controlpanel,waituntiltheinitializationprocessiscomplete,andthenseparatelyrepeatthisprocedurefor“Slit”and“Spectrometer”.

Page 17: Instructions for Horiba Raman confocal imaging microscope · Raman confocal imaging microscope Instructions for Horiba Raman confocal imaging microscope ... Close the Labspec software

MicroscopySuite,ImagingTechnologyGroupBeckmanInstituteforAdvancedScienceandTechnologyUniversityofIllinoisatUrbana‐Champaign

OperationmanualofRamanconfocal

405NorthMathews,UrbanaIL61801Tel:217‐244‐0170;FAX:217‐244‐6219

http://itg.beckman.illinois.edu/ ‐14‐