INNOVIVE PERFORMANCE REPORT Innorack 3.0 Acoustic Noise Levels Report · 2020. 3. 13. · Innovive...

6
Innorack ® 3.0 Acoustic Noise Levels Report (page 1 of 6) Innovive Inc Test Location I. INTRODUCTION In an animal laboratory environment, noise generated by equipment may have adverse effects on laboratory mice and can ultimately compromise the validity of an experiment. Hearing sensitivity at a range of frequencies is different across species. Mouse hearing ranges from 1 to 100 kHz while human hearing ranges 20 Hz to about 20 kHz. FIGURE 1 shows the human and mouse audiograms with decibels plotted against increasing frequencies. 1 The audiogram shows low hearing sensitivity at low frequencies, with sensitivity increasing as frequency increases until hearing again becomes less sensitive with further frequency increase. Hearing is most sensitive for humans at approximately 1 to 4 kHz and16 kHz for mice. This study was performed to evaluate the sound levels produced by the Innorack ® 3.0 IVC Mouse and to determine if mice are exposed to prolonged high-decibel sound within their hearing range. II. EXPERIMENTAL PROCEDURE Test Location and Equipment This test was performed at the Brüel & Kjær Acoustic Testing Facility in Canton, MI inside a hemi-anechoic chamber. Sound measurements were taken with a B.K.S.V. ¼ inch high frequency microphone. B.K.S.V. Pulse Analyzer software was used to capture and process test data. The test parameters are shown in FIGURE 2. A single-sided, 8-column, 96-cage position Innorack ® 3.0 served as the test IVC rack (see IMAGE 1). The rack was fully populated with cages. Each test cage contained ½ inch of corn cob bedding and a water bottle. Sampling Rate Lines dF Frequency Range Weighting 3.815 us 800 128 Hz 0 – 100 kHz linear FIGURE 1. Human and Mouse Audiograms HUMAN AND MOUSE AUDIOGRAMS Sound Pressure Level (dB/20μPa) 0 10,000 20,000 30,000 40,000 Hz 50,000 60,000 70,000 80,000 90,000 160 140 120 80 100 60 20 40 0 Human Mouse FIGURE 2. Signal Analyzer Test Settings IMAGE 1. Innorack ® 3.0 Mouse in a hemi-anechoic chamber at the Brüel & Kjær Acoustic Teting Facility. INNOVIVE PERFORMANCE REPORT Innorack ® 3.0 Acoustic Noise Levels Report

Transcript of INNOVIVE PERFORMANCE REPORT Innorack 3.0 Acoustic Noise Levels Report · 2020. 3. 13. · Innovive...

  • Innorack® 3.0 Acoustic Noise Levels Report (page 1 of 6)Innovive Inc

    Test Location

    I. INTRODUCTION

    In an animal laboratory environment, noise generated by equipment may have adverse effects on laboratory mice and can ultimately

    compromise the validity of an experiment. Hearing sensitivity at a range of frequencies is different across species. Mouse hearing ranges

    from 1 to 100 kHz while human hearing ranges 20 Hz to about 20 kHz. FIGURE 1 shows the human and mouse audiograms with decibels

    plotted against increasing frequencies.1

    The audiogram shows low hearing sensitivity at low frequencies, with sensitivity increasing as frequency increases until hearing again

    becomes less sensitive with further frequency increase. Hearing is most sensitive for humans at approximately 1 to 4 kHz and16 kHz

    for mice. This study was performed to evaluate the sound levels produced by the Innorack® 3.0 IVC Mouse and to determine if mice are

    exposed to prolonged high-decibel sound within their hearing range.

    II. EXPERIMENTAL PROCEDURE

    Test Location and Equipment

    This test was performed at the Brüel & Kjær Acoustic Testing

    Facility in Canton, MI inside a hemi-anechoic chamber.

    Sound measurements were taken with a B.K.S.V. ¼ inch high

    frequency microphone. B.K.S.V. Pulse Analyzer software was

    used to capture and process test data. The test parameters are

    shown in FIGURE 2. A single-sided, 8-column, 96-cage position

    Innorack® 3.0 served as the test IVC rack (see IMAGE 1). The rack

    was fully populated with cages. Each test cage contained ½ inch

    of corn cob bedding and a water bottle.

    Sampling Rate

    Lines dFFrequency

    RangeWeighting

    3.815 us 800 128 Hz 0 – 100 kHz linear

    FIGURE 1. Human and Mouse Audiograms

    HUMAN AND MOUSE AUDIOGRAMS

    So

    und

    Pre

    ssur

    e Le

    vel (

    dB

    /20µ

    Pa)

    0 10,000 20,000 30,000 40,000

    Hz

    50,000 60,000 70,000 80,000 90,000

    160

    140

    120

    80

    100

    60

    20

    40

    0

    Human

    Mouse

    FIGURE 2. Signal Analyzer Test Settings

    IMAGE 1. Innorack® 3.0 Mouse in a hemi-anechoic chamber at the Brüel &

    Kjær Acoustic Teting Facility.

    INNOVIVE PERFORMANCE REPORT

    Innorack® 3.0Acoustic Noise Levels Report

  • Innorack® 3.0 Acoustic Noise Levels Report (page 2 of 6)Innovive Inc

    III. RESULTS

    FIGURES 4-10 show test results for each ACH/DP setting used in the test method above.

    Test Method

    1. Place the microphone inside a disposable mouse cage.

    Locate the tip of the microphone in the center of the

    cage approximately 1” above the bedding material.

    2. Acquire a sound reading with the system off for 30

    seconds to get a baseline reading.

    3. Insert the cage to rack position A1 (see FIGURE 3).

    4. Set the Innorack system to 50 ACH, POS differential

    pressure

    5. Acquire a sound reading for 30 seconds.

    6. Set the Innorack system to 60 ACH, POS differential

    pressure.

    7. Acquire a sound reading for 30 seconds.

    8. Set the Innorack system to 70 ACH, POS differential

    pressure.

    9. Acquire a sound reading for 30 seconds.

    10. Repeat steps (4) through (9) with the cage at positions

    A12, D6, D7, E6, E7, H1, and H12 (see FIGURE 3).

    11. Repeat steps (3) through (9) with the Innorack system

    at 50 ACH, 60 ACH, and 70 ACH at NEG differential

    pressure.

    1k 1.5k 2k 2.5k 3k 3.5k 4k 4.5k 5k[Hz]

    0

    4

    8

    12

    16

    20

    24

    28

    [dB(A)/200µ Pa]

    Background Noise, Rack Interior

    A1 - BLUE

    A12 - PINK

    D6 - GREEN

    D7 - YELLOW

    E6 - RED

    E7 - ORANGE

    H1 - LIGHT GREEN

    H12 - BLACK

    RO

    WS

    SUPPLY

    BLOWER

    EXHAUST

    BLOWER

    COLUMN A B C D E F G H

    1 A1 H1

    2

    3

    4

    5

    6 D6 E6

    7 D7 E7

    8

    9

    10

    11

    12 A12 H12

    FIGURE 3. Test Shelf Positions

    A1, A12, D6, D7, E6, E7, H1, and H12 represent cage positions

    used for testing.

    FIGURE 4. Background Noise, Rack Interior

  • Innorack® 3.0 Acoustic Noise Levels Report (page 3 of 6)Innovive Inc

    1k 1.5k 2k 2.5k 3k 3.5k 4k 4.5k 5k[Hz]

    0

    4

    8

    12

    16

    20

    24

    28

    [dB/200µ Pa]

    50 ACH, NEG D.P., Rack Interior

    1k 1.5k 2k 2.5k 3k 3.5k 4k 4.5k 5k[Hz]

    0

    4

    8

    12

    16

    20

    24

    28

    [dB/200µ Pa]

    50 ACH, POS D.P., Rack Interior

    A1 - BLUE

    A12 - PINK

    D6 - GREEN

    D7 - YELLOW

    E6 - RED

    E7 - ORANGE

    H1 - LIGHT GREEN

    H12 - BLACK

    A1 - BLUE

    A12 - PINK

    D6 - GREEN

    D7 - YELLOW

    E6 - RED

    E7 - ORANGE

    H1 - LIGHT GREEN

    H12 - BLACK

    FIGURE 5. 50 ACH, POS DP, Rack Interior

    FIGURE 6. 50 ACH, NEG DP, Rack Interior

  • Innorack® 3.0 Acoustic Noise Levels Report (page 4 of 6)Innovive Inc

    1k 1.5k 2k 2.5k 3k 3.5k 4k 4.5k 5k[Hz]

    0

    4

    8

    12

    16

    20

    24

    28

    [dB/200µ Pa]

    60 ACH, POS D.P., Rack Interior

    1k 1.5k 2k 2.5k 3k 3.5k 4k 4.5k 5k[Hz]

    0

    4

    8

    12

    16

    20

    24

    28

    [dB/200µ Pa]

    60 ACH, NEG D.P., Rack Interior

    A1 - BLUE

    A12 - PINK

    D6 - GREEN

    D7 - YELLOW

    E6 - RED

    E7 - ORANGE

    H1 - LIGHT GREEN

    H12 - BLACK

    A1 - BLUE

    A12 - PINK

    D6 - GREEN

    D7 - YELLOW

    E6 - RED

    E7 - ORANGE

    H1 - LIGHT GREEN

    H12 - BLACK

    FIGURE 7. 60 ACH, POS DP, Rack Interior

    FIGURE 8. 60 ACH, NEG DP, Rack Interior

  • Innorack® 3.0 Acoustic Noise Levels Report (page 5 of 6)Innovive Inc

    1k 1.5k 2k 2.5k 3k 3.5k 4k 4.5k 5k[Hz]

    0

    4

    8

    12

    16

    20

    24

    28

    32

    3640

    [dB/200µ Pa]

    70 ACH, POS D.P., Rack Interior

    1k 1.5k 2k 2.5k 3k 3.5k 4k 4.5k 5k[Hz]

    0

    4

    8

    12

    16

    20

    24

    28

    [dB/200µ Pa]

    70 ACH, NEG D.P., Rack Interior

    A1 - BLUE

    A12 - PINK

    D6 - GREEN

    D7 - YELLOW

    E6 - RED

    E7 - ORANGE

    H1 - LIGHT GREEN

    H12 - BLACK

    A1 - BLUE

    A12 - PINK

    D6 - GREEN

    D7 - YELLOW

    E6 - RED

    E7 - ORANGE

    H1 - LIGHT GREEN

    H12 - BLACK

    FIGURE 10. 70 ACH, NEG DP, Rack Interior

    FIGURE 9. 70 ACH, POS DP, Rack Interior

  • Innorack® 3.0 Acoustic Noise Levels Report (page 6 of 6)Innovive Inc

    IV. DISCUSSION

    Plots were scaled to account for noise occurring only within the mouse hearing range, 1kHz – 100kHz. Cages docked in the top row

    recorded highest noise levels, though none exceeded 40dB and occurred at the bottom end of the mouse hearing range. Cages docked

    in all other positions recorded 12dB less on average than those docked in the top row.

    For all test positions, less than 15dB was present past 2000 Hz with no noise present past 5000 Hz. It can be concluded that acoustic

    noise produced by the Innorack® IVC Mouse 3.0 system is not harmful to the well-being of mice inside the cages.

    REFERENCES

    1. Randall R, Kinard W, Degraff J, Leverage N, Norton J. 2010. Noise in a Laboratory Animal Facility from Human and Mouse

    Perspectives. Journal of the American Association for Laboratory Animal Science.

    Test ParameterHighest dB

    Level*

    Frequency

    at Highest dB

    Frequency Range

    of Noise Content

    Mic Position

    at Highest dB Level

    50 ACH, POS DP 23 1000 Hz 1000 – 4600 Hz A1

    50 ACH, NEG DP 29 1150 Hz 1000 – 3800 Hz A1

    60 ACH, POS DP 27 1150 Hz 1000 – 4750 Hz A1

    60 ACH, NEG DP 23 1150 Hz 1000 – 4750 Hz A1

    70 ACH, POS DP 38 1024 Hz 1000 – 5000 Hz H1

    70 ACH, NEG DP 29 1408 Hz 1000 – 4750 Hz A1

    *dB Reference - Background Noise - dB/200µ Pa

    FIGURE 11. Summary of Acoustic Data Source: Brüel & Kjær North America Inc.