Initial Results from the Scintillator Fast Lost Ion Probe

28
Initial Results from the Scintillator Fast Lost Ion Probe D. Darrow NSTX Physics Meeting February 28, 2005

description

Initial Results from the Scintillator Fast Lost Ion Probe. D. Darrow NSTX Physics Meeting February 28, 2005. Goal & Motivations. Goal: Predict fast ion losses from ST plasmas Motivations: Dimensionless parameters of beam ions similar to 3.5 MeV a s in NSST (good model system) - PowerPoint PPT Presentation

Transcript of Initial Results from the Scintillator Fast Lost Ion Probe

Page 1: Initial Results from the Scintillator Fast Lost Ion Probe

Initial Results from the Scintillator Fast Lost Ion Probe

D. Darrow

NSTX Physics Meeting

February 28, 2005

Page 2: Initial Results from the Scintillator Fast Lost Ion Probe

Goal & Motivations

Goal:– Predict fast ion losses from ST plasmas

Motivations:– Dimensionless parameters of beam ions similar to

3.5 MeV s in NSST (good model system)– Lost beam ion characteristics can reveal internal

physics, esp. effects of MHD instabilities

Page 3: Initial Results from the Scintillator Fast Lost Ion Probe

Outline

• Loss mechanisms

• sFLIP diagnostic

• Example data

• Parametric dependence of loss

• Model of detector signal

Page 4: Initial Results from the Scintillator Fast Lost Ion Probe

Fast ion loss mechanisms

• Prompt orbit loss: fast ion born in loss cone• Radial transport to wall (P):

– MHD– TF ripple

• Pitch angle scattering into loss cone ():– Classical collisions– ICRF heating

Page 5: Initial Results from the Scintillator Fast Lost Ion Probe

This work: mainly prompt loss

• Prompt loss increases with:– decreasing Ip

– decreasing outer

– decreasing Rtan

• NSTX: 80–90 keV D NBI– A: Rtan = 69.4 cm

– B: Rtan = 59.2 cm

– C: Rtan = 48.7 cm

0

20

40

60

80

100

0 200 400 600 800 1000Plasma current (kA)

outer

=3.8 cm

outer

=14.4 cm

Page 6: Initial Results from the Scintillator Fast Lost Ion Probe

Scintillator fast lost ion (sFLIP) probe is magnetic spectrometer

• Combination of B and aperture geometry disperse different pitch angles and energies on scintillator plate

Scintillator detector: principle of operation

Bay J

Vessel & limiters

NSTX Midplane

ScintillatorDetectorBeam C footprint

Page 7: Initial Results from the Scintillator Fast Lost Ion Probe

Scintillator probe assembly

Aperture

Light shield

Graphitearmor

Base &Heat sink

Scintillator(inside)

Plasma

Vacuumwindow

Bay J

Incidentions

: 5–60 cm, : 10°–70° (typ.)

Page 8: Initial Results from the Scintillator Fast Lost Ion Probe

Typical orbit to detector

• Commonly only a few steps contribute in each orbit

• Model includes full 3D structure of vessel & beam deposition

Page 9: Initial Results from the Scintillator Fast Lost Ion Probe

& map can be applied to data

NSTX sFLIP diagnostic, shot 111192, frame 5, with grid for t=169 msB=0.2692 T, rho=22 cm, E_D=84 keV, chi=23°, 32°, & 60°

Page 10: Initial Results from the Scintillator Fast Lost Ion Probe

Fiber optic bundle limits resolution of fast ion parameters

• Limited resolution of bundle (50 x 50) causes discretization of image & uncertainty in scintillator position in camera field of view

CCD Camera

Scintillator

Fiber bundle

Single fiber

Position calibration image of scintillator

Page 11: Initial Results from the Scintillator Fast Lost Ion Probe

Instrumental “line widths” also set limit on resolution

• Example case: 80 keV (=24 cm) FWHM is =8 cm

• Pitch angle line width: 6° FWHM

Page 12: Initial Results from the Scintillator Fast Lost Ion Probe

Beam ion loss clearly seen

112132: 800 kA, 4 MW

QuickTime™ and a decompressor

are needed to see this picture.

Higher

Lower

Lower Higher

30 frames/s= pitch angle

= tan-1(v||/v)

Page 13: Initial Results from the Scintillator Fast Lost Ion Probe

Several general classes of loss seen

• Few cases analyzed so far, but all consistent loss at injection energy (prompt loss)

“Bar” loss: wide range

Typically early in NBI:

low ne & deeper dep’n

(113002, 330 ms)

High loss

Typ. later in NBI: high ne

Often modulated by MHD

(112232, 400 ms)

Multiple discrete s

(111130)

Page 14: Initial Results from the Scintillator Fast Lost Ion Probe

Methodology of prompt loss investigation

• Compare losses from 112164 (source A only) & 112166 (source C only) to determine effect of Rtan (nominally identical shots)

• Compare different time slices within each shot to determine effect of Ip on loss, since beam injection starts during Ip ramp up

Page 15: Initial Results from the Scintillator Fast Lost Ion Probe

Parameters for 112164, 112166

112164: A

112166: C

Page 16: Initial Results from the Scintillator Fast Lost Ion Probe

Measurements show loss decreases as Ip increases

99 ms, 500 kA

116 ms, 650 kA

149 ms, 750 kA

112164

Source A

90 keV

Page 17: Initial Results from the Scintillator Fast Lost Ion Probe

More loss seen from source C than A under same conditions

112164 (A)–top vs 112166 (C )–bottom

100 ms 115 ms 150 ms

Page 18: Initial Results from the Scintillator Fast Lost Ion Probe

Are these prompt losses?

• If so, then:– Detected energy must equal injection energy– Detected pitch angle must correspond to an orbit

populated directly by the beam deposition

Page 19: Initial Results from the Scintillator Fast Lost Ion Probe

Gyroradius range appears consistent with loss at Einj

• 90 keV D, 0.25 T => =25 cm• Scintillator image position calib. injects uncertainty

10°

20°30°

40°50° 60°

5

10

15

20

Pitch Angle ()

Gyr

orad

ius

cent

roid

(cm

)

112164, 100 ms 112166, 150 ms

10°

20°30°

50°60°

5

10

15

20

Pitch Angle ()

Gyr

orad

ius

cent

roid

(cm

)

40° 70°

Page 20: Initial Results from the Scintillator Fast Lost Ion Probe

Detector signal modeling for range of detected

• Need efficient method to compare volume of phase space sampled by detector with volumes populated through beam injection

• “Constants of Motion” (COM) approach: orbit fully characterized by E, (=mvperp

2/2B), & P (=mvR+qpol)

• For prompt loss, where E does not change, problem is 2D: plot beam deposition & detected orbits in (P, ) and look for overlap

• But, conservation marginal in STs!

Page 21: Initial Results from the Scintillator Fast Lost Ion Probe

COM model (cont’d)

• Treat beam as ensemble of test particles deposited in 3D volume where beam passes through plasma – all velocities parallel to beam axis– ~100,000 particles typically

• Model detected ions as 2D fan of velocities at detector entrance aperture– ~100 velocities, ~1° steps in

• Plot both sets in same (P, ) space for Einj, look for overlap

Page 22: Initial Results from the Scintillator Fast Lost Ion Probe

Example case

• Clear overlap seen between deposited beam orbits and orbits sampled by sFLIP

• Predicts loss at detector, =20° to 54°

P (10-20 kg m2/s)

(1

0-14

J/T

)

112166, 100 ms, 500 kA, source A, 90 keV

Beam ions

sFLIP

Ra

nge

of

p

red

icte

d a

t d

ete

cto

r

Page 23: Initial Results from the Scintillator Fast Lost Ion Probe

Model in reasonable agreement with measured range

• Model predicts =22.7 cm, 10°≤≤35°

• Measured spot is extended due to finite aperture size, but is consistent with model &

10°

20°30°

40°50°60°

70°

5

10

15

20

Pitch Angle ()

Gyr

orad

ius

cent

roid

(cm

)

Model

112166, 150 ms, source C

Page 24: Initial Results from the Scintillator Fast Lost Ion Probe

Model reproduces observed differences between A & C

• C fills low orbits at detector (t>100 ms); A does not

112164: source A

97 ms 139 ms 169 ms

100 ms 140 ms 170 ms

112166: source C P

Page 25: Initial Results from the Scintillator Fast Lost Ion Probe

Bright, high loss often observed during MHD

• Lost at injection energy, =64°

• Prompt loss model: 48°–63°

• Loss appears too localized in to be consistent with prompt loss

10°

20°30°40°50° 60°

5

10

15

20

Pitch Angle ()

Gyr

orad

ius

cent

roid

(cm

)

Model

70° 80°

112074, 400 ms, sources A, B, & C

(with Fredrickson, Medley)

Page 26: Initial Results from the Scintillator Fast Lost Ion Probe

MHD-lost ions are banana orbits, near P/T boundary

• P/T boundary at 60°

• Bounce frequency changes rapidly with here–87 kHz for this orbit

PPPL Lorentz ORBIT v205 for NSTX 02/08/16 11:10:00ORBIT(S) CALCULATED ON 02/03/05Comment Orbits to sFLIP in 112074, t=403 msORBIT STRUCK LIMITER AT R= 1.502 Ph= -83.4 Z= .465

3.1416 Init gyro ang.0000 Init pitch ang1.60 Detector R-.14 Detector Z

1.33046 Detector phi1.3744 Det RZ ang

-1.5708 Det tor ang1.00 Ion charge2.00 Ion mass

9.000E-02 Init energy0.000E+00 Init mu

0 Time dir'n60/ 81 IX/NX1/ 1 IY/NY.01 Step len (m)

100.00 Orbit len.0616 Aper sep (m).0060 Col hwidth.0010 Col hheight.0463 Det hwidth.0010 Det hheight

GYRO = .3749CDPTCH= .0000RIPTCH= 115.80RMN = .00EFFIC =9.757E-09SUMEFF=3.590E-08

0.5 1.0 1.5

-1.5

-1.0

-0.5

0.0

0.5

1.0

1.5

-1 01

-1

0

1

Page 27: Initial Results from the Scintillator Fast Lost Ion Probe

Summary

• sFLIP diagnostic now measuring beam ion loss routinely

• Beam ion loss parametric dependence, gyroradius, & pitch angles match prompt orbit loss

• (P, ) mapping provides fast calculation of prompt loss pitch angles at detector

• MHD-induced loss seen near P/T boundary

Page 28: Initial Results from the Scintillator Fast Lost Ion Probe

Future plans

• Make absolute calibration of loss rate with internal Faraday cups

• Higher resolution fiber bundle (?)

• Augment model to include orbit class boundaries, loss boundary

• Investigate loss at high rotation speed