Infusion Kit compatibility James Baker - CHAOTIC 2 nd November 2011 1.

16
Infusion Kit compatibility James Baker - CHAOTIC 2 nd November 2011 1

description

Infusion Kits |3

Transcript of Infusion Kit compatibility James Baker - CHAOTIC 2 nd November 2011 1.

Page 1: Infusion Kit compatibility James Baker - CHAOTIC 2 nd November 2011 1.

Infusion Kit compatibility

James Baker - CHAOTIC2nd November 2011

1

Page 2: Infusion Kit compatibility James Baker - CHAOTIC 2 nd November 2011 1.

What do we do

Kit compatibility is performed to check that the formulation does not adsorb to the infusion apparatus.

For AZ and contracted out AZ studies we use the lowest flow rate, for the duration of the infusion on the study, using the lowest concentration in the formulation method. This is used for all dog, monkey and rat infusion systems studies. (Worst case scenario)

Study conducted using the same infusion kit as on the study.

If a maintenance flush is used this is performed first, prior to the infusion starting.

Samples are collected at appropriate intervals and analysed for loss of concentration.

Compatibility confirmed if the initial formulation is ±10% of theoretical and the % difference after infusion is ≤5%.

2

Page 3: Infusion Kit compatibility James Baker - CHAOTIC 2 nd November 2011 1.

Infusion Kits

|3

Page 4: Infusion Kit compatibility James Baker - CHAOTIC 2 nd November 2011 1.

Example infusion problem

We had a new formulation ( 10% Kleptose® , 4.5% w/v sodium chloride in WFI) and AZDXXXX.

We performed a 2 hour infusion, 2mL/kg on a 0.05mg/mL formulation for use on a MTD dog study

Hour 0-165% initialHour 1-257% initial

The infusion system consisted of a PEGA bag, PEGA tube, interlink and catheter. Infusion bag contents was then analysed and found to be 39% of initial. This was after about an hour left in the lab at room temperature after the 2 hour infusion.

Investigated, and was not an analytical dilution problem or recovery during analysis.

4

Page 5: Infusion Kit compatibility James Baker - CHAOTIC 2 nd November 2011 1.

Interim conclusion

Formulation was adhering to the infusion system.

Assumed to be the bag, could it be the material?

Change the bag (not PVC)? The bags for our infusion systems are not available in different materials (e.g. polyurethane).

Does this happen for a 0.5mg/mL concentration (the top concentration for the method)?

Repeat the compatibility in duplicate and also singularly on the infusion line using a syringe to investigate if the formulation adheres to the infusion line.

5

Page 6: Infusion Kit compatibility James Baker - CHAOTIC 2 nd November 2011 1.

Compatibility with 100mL bags

Result for syringe line at 120mins was a sample taken from the remainder in the line

6

Page 7: Infusion Kit compatibility James Baker - CHAOTIC 2 nd November 2011 1.

Results from infusion

Appears that sticking is occurring in the infusion line and the 100mL bag.

Study director contacted.

Results low but after delivering 7mLs the results became stable.

Dose to be altered by flow rate, 20 min flush to waste in the lines prior to dosing starting.

Experiment to be re-run at 2 flow rates (2mL/kg/hr & 10mL/kg/hr).

7

Page 8: Infusion Kit compatibility James Baker - CHAOTIC 2 nd November 2011 1.

Flow rates

8

Page 9: Infusion Kit compatibility James Baker - CHAOTIC 2 nd November 2011 1.

Results from flow rates

Results were satisfactory, doses were low but consistent at the 2mL flow rate.

Noted that at the higher flow rate less compound was being adsorbed.

Suggestion to use a larger bag volume to reduce the surface area/mL ratio.

9

Page 10: Infusion Kit compatibility James Baker - CHAOTIC 2 nd November 2011 1.

Compatibility performed on 300ml bag

10

Page 11: Infusion Kit compatibility James Baker - CHAOTIC 2 nd November 2011 1.

Results from 300mL bag

300mL bag gave better results.

Stability in the bag to be evaluated for 100mL and 300mL bags.

11

Page 12: Infusion Kit compatibility James Baker - CHAOTIC 2 nd November 2011 1.

Stability in 100 and 300mL bags

12

Page 13: Infusion Kit compatibility James Baker - CHAOTIC 2 nd November 2011 1.

Conclusion on analytical investigation

Changing to a 300mL bag has improved the compatibility result by approximately 2%.

Stability also improved by 1 day for RT.

Stability improved by using large bags - 3% loss for 2-8°C compared to 5% loss at 2-8°C for 100mL bags.

For the study samples to be taken after the flush and again after the 2 hour infusion for analysis and TK’s corrected for actual concentration delivered. (no significant change was noted).

13

Page 14: Infusion Kit compatibility James Baker - CHAOTIC 2 nd November 2011 1.

Pre and post analysis samples (% of initial) 300mL bag

Animal 103          93-102 pre to 95-100 post (max 3% change)Animal 104          96-100 pre to 95-98 post (max 3% change)Animal 105          96-99 pre to 96-99 post (max 2% change)Animal 106          96-101 pre to 95-99 post  (max 2% change)

14

Page 15: Infusion Kit compatibility James Baker - CHAOTIC 2 nd November 2011 1.

Questions

?15

Confidentiality Notice This file is private and may contain confidential and proprietary information. If you have received this file in error, please notify us and remove it from your system and note that you must not copy, distribute or take any action in reliance on it. Any unauthorized use or disclosure of the contents of this file is not permitted and may be unlawful. AstraZeneca PLC, 2 Kingdom Street, London, W2 6BD, UK, T: +44(0)20 7604 8000, F: +44 (0)20 7604 8151, www.astrazeneca.com

Page 16: Infusion Kit compatibility James Baker - CHAOTIC 2 nd November 2011 1.

Can of worms?

Do we need extra room temperature stability.

Stability required in bags/syringes at 2-8C as a real time stability

e.g. cycle from RT (formulation) to 2-8C (storage) to RT(Acclimatising before dosing) to 37C (in jacket)?

Stability at body temperature if an extended study e.g. 4 hrs?

How long to leave a formulation in a dosing line before dosing?

Saline precipitation / precipitation with flush vehicles? Use of Heparin.

Are inline filters are present in the infusion system?

16