Index of Microvascular Resistance (IMR) and Absolute...
Transcript of Index of Microvascular Resistance (IMR) and Absolute...
![Page 1: Index of Microvascular Resistance (IMR) and Absolute ...assets.escardio.org/assets/presentations/OTHER2013/...C Proximal Sensor (Shaft of the Wire) Distal Sensor T 0 1/2 T i 1/2 T](https://reader031.fdocuments.in/reader031/viewer/2022012006/6102a3e361953b73803d9b33/html5/thumbnails/1.jpg)
Bernard De Bruyne, MD, PhD
Cardiovascular Center Aalst
OLV-Clinic Aalst, Belgium
ww
w.c
ard
io-a
als
t.b
e
IMR and ACF
ETP, Sofia Antipolis, April 2013
Index of Microvascular Resistance (IMR)
and
Absolute Coronary Blood Flow (ACF)
![Page 2: Index of Microvascular Resistance (IMR) and Absolute ...assets.escardio.org/assets/presentations/OTHER2013/...C Proximal Sensor (Shaft of the Wire) Distal Sensor T 0 1/2 T i 1/2 T](https://reader031.fdocuments.in/reader031/viewer/2022012006/6102a3e361953b73803d9b33/html5/thumbnails/2.jpg)
ww
w.c
ard
io-a
als
t.b
e
IMR and ACF
ETP, Sofia Antipolis, April 2013
1. Concept(s) of dilution methods
2. Bolus injection: IMR
3. Continuous Infusion: ACF
![Page 3: Index of Microvascular Resistance (IMR) and Absolute ...assets.escardio.org/assets/presentations/OTHER2013/...C Proximal Sensor (Shaft of the Wire) Distal Sensor T 0 1/2 T i 1/2 T](https://reader031.fdocuments.in/reader031/viewer/2022012006/6102a3e361953b73803d9b33/html5/thumbnails/3.jpg)
Time, s
Indicators
Dye
Contrast Intensity
Viscosity
Hemoglobin
pH
Crypton
Temperature
...
Co
nc
en
tra
tio
n
ww
w.c
ard
io-a
als
t.b
e
IMR and ACF
ETP, Sofia Antipolis, April 2013
Concept(s) of dilution methods
![Page 4: Index of Microvascular Resistance (IMR) and Absolute ...assets.escardio.org/assets/presentations/OTHER2013/...C Proximal Sensor (Shaft of the Wire) Distal Sensor T 0 1/2 T i 1/2 T](https://reader031.fdocuments.in/reader031/viewer/2022012006/6102a3e361953b73803d9b33/html5/thumbnails/4.jpg)
If 1. Mixing of the indicator is complete2. Heat exchange with adjacent tissue is negligible
3. Flow is constant during the measurement
4. No recirculation of the indicator during the measurement
5. The injection of the indicator does not cause any change to the flow
K. Pávek et al. Circ Res 1964
ww
w.c
ard
io-a
als
t.b
e
IMR and ACF
ETP, Sofia Antipolis, April 2013
Concept(s) of dilution methods
![Page 5: Index of Microvascular Resistance (IMR) and Absolute ...assets.escardio.org/assets/presentations/OTHER2013/...C Proximal Sensor (Shaft of the Wire) Distal Sensor T 0 1/2 T i 1/2 T](https://reader031.fdocuments.in/reader031/viewer/2022012006/6102a3e361953b73803d9b33/html5/thumbnails/5.jpg)
Temperature
Time
F =M c(t)dt
0
• Amount ‘M’ of indicator should be known
• Vascular volume should not be known
ww
w.c
ard
io-a
als
t.b
e
IMR and ACF
ETP, Sofia Antipolis, April 2013
Concept(s) of dilution methods
![Page 6: Index of Microvascular Resistance (IMR) and Absolute ...assets.escardio.org/assets/presentations/OTHER2013/...C Proximal Sensor (Shaft of the Wire) Distal Sensor T 0 1/2 T i 1/2 T](https://reader031.fdocuments.in/reader031/viewer/2022012006/6102a3e361953b73803d9b33/html5/thumbnails/6.jpg)
RestCO=4.8 L/min
DobutamineCO=8.7 L/min
ww
w.c
ard
io-a
als
t.b
e
IMR and ACF
ETP, Sofia Antipolis, April 2013
Concept(s) of dilution methods
![Page 7: Index of Microvascular Resistance (IMR) and Absolute ...assets.escardio.org/assets/presentations/OTHER2013/...C Proximal Sensor (Shaft of the Wire) Distal Sensor T 0 1/2 T i 1/2 T](https://reader031.fdocuments.in/reader031/viewer/2022012006/6102a3e361953b73803d9b33/html5/thumbnails/7.jpg)
Temperature
Time
F =M c(t)dt
0
• Amount ‘M’ of indicator should be known
• Vascular volume should not be known
• Amount of ‘indicator’ should not be known
• Vascular volume ‘V’ should be known
ww
w.c
ard
io-a
als
t.b
e
IMR and ACF
ETP, Sofia Antipolis, April 2013
F = V
Tmn
Concept(s) of dilution methods
![Page 8: Index of Microvascular Resistance (IMR) and Absolute ...assets.escardio.org/assets/presentations/OTHER2013/...C Proximal Sensor (Shaft of the Wire) Distal Sensor T 0 1/2 T i 1/2 T](https://reader031.fdocuments.in/reader031/viewer/2022012006/6102a3e361953b73803d9b33/html5/thumbnails/8.jpg)
ww
w.c
ard
io-a
als
t.b
e
IMR and ACF
ETP, Sofia Antipolis, April 2013
Concept(s) of dilution methods
![Page 9: Index of Microvascular Resistance (IMR) and Absolute ...assets.escardio.org/assets/presentations/OTHER2013/...C Proximal Sensor (Shaft of the Wire) Distal Sensor T 0 1/2 T i 1/2 T](https://reader031.fdocuments.in/reader031/viewer/2022012006/6102a3e361953b73803d9b33/html5/thumbnails/9.jpg)
ww
w.c
ard
io-a
als
t.b
e
IMR and ACF
ETP, Sofia Antipolis, April 2013
Concept(s) of dilution methods
![Page 10: Index of Microvascular Resistance (IMR) and Absolute ...assets.escardio.org/assets/presentations/OTHER2013/...C Proximal Sensor (Shaft of the Wire) Distal Sensor T 0 1/2 T i 1/2 T](https://reader031.fdocuments.in/reader031/viewer/2022012006/6102a3e361953b73803d9b33/html5/thumbnails/10.jpg)
Temperature of the shaft of the wire
Temperature of the sensor
ww
w.c
ard
io-a
als
t.b
e
IMR and ACF
ETP, Sofia Antipolis, April 2013
Concept(s) of dilution methods
![Page 11: Index of Microvascular Resistance (IMR) and Absolute ...assets.escardio.org/assets/presentations/OTHER2013/...C Proximal Sensor (Shaft of the Wire) Distal Sensor T 0 1/2 T i 1/2 T](https://reader031.fdocuments.in/reader031/viewer/2022012006/6102a3e361953b73803d9b33/html5/thumbnails/11.jpg)
ww
w.c
ard
io-a
als
t.b
e
IMR and ACF
ETP, Sofia Antipolis, April 2013
Concept(s) of dilution methods
![Page 12: Index of Microvascular Resistance (IMR) and Absolute ...assets.escardio.org/assets/presentations/OTHER2013/...C Proximal Sensor (Shaft of the Wire) Distal Sensor T 0 1/2 T i 1/2 T](https://reader031.fdocuments.in/reader031/viewer/2022012006/6102a3e361953b73803d9b33/html5/thumbnails/12.jpg)
Mean Transit Time, Tmn
= Mean time needed for the indicator to travel from the guiding to the sensor
° C
Proximal Sensor(Shaft of the Wire)
Distal Sensor
T01/2 Ti 1/2 Ti
Tmn
De Bruyne B et al Circulation 2001;104:2003
Pijls NHJ et al Circulation 2002 105:2482
ww
w.c
ard
io-a
als
t.b
e
IMR and ACF
ETP, Sofia Antipolis, April 2013
![Page 13: Index of Microvascular Resistance (IMR) and Absolute ...assets.escardio.org/assets/presentations/OTHER2013/...C Proximal Sensor (Shaft of the Wire) Distal Sensor T 0 1/2 T i 1/2 T](https://reader031.fdocuments.in/reader031/viewer/2022012006/6102a3e361953b73803d9b33/html5/thumbnails/13.jpg)
General Principle of Coronary Thermodilution
F = V
Tmn
Tmn (s) = V (mL)
F (mL/s)
Tmn (s) = an index of flow
ww
w.c
ard
io-a
als
t.b
e
IMR and ACF
ETP, Sofia Antipolis, April 2013
![Page 14: Index of Microvascular Resistance (IMR) and Absolute ...assets.escardio.org/assets/presentations/OTHER2013/...C Proximal Sensor (Shaft of the Wire) Distal Sensor T 0 1/2 T i 1/2 T](https://reader031.fdocuments.in/reader031/viewer/2022012006/6102a3e361953b73803d9b33/html5/thumbnails/14.jpg)
Tmn (s) = an index of flow
CFR IMR
ww
w.c
ard
io-a
als
t.b
e
IMR and ACF
ETP, Sofia Antipolis, April 2013
![Page 15: Index of Microvascular Resistance (IMR) and Absolute ...assets.escardio.org/assets/presentations/OTHER2013/...C Proximal Sensor (Shaft of the Wire) Distal Sensor T 0 1/2 T i 1/2 T](https://reader031.fdocuments.in/reader031/viewer/2022012006/6102a3e361953b73803d9b33/html5/thumbnails/15.jpg)
Then,
IMR = Pd x Tmn
Myocardial Resistance (Rmyo) = Pd - Pv
F(mm Hg.mL-1.min-1)
IMR: Definition
Since: (1) F 1 / Tmn
(2) Pv 0 mm Hg
= Pd
1 / Tmn
IMR
Fearon WF et al Circulation 2003;107:3129
ww
w.c
ard
io-a
als
t.b
e
IMR and ACF
ETP, Sofia Antipolis, April 2013
![Page 16: Index of Microvascular Resistance (IMR) and Absolute ...assets.escardio.org/assets/presentations/OTHER2013/...C Proximal Sensor (Shaft of the Wire) Distal Sensor T 0 1/2 T i 1/2 T](https://reader031.fdocuments.in/reader031/viewer/2022012006/6102a3e361953b73803d9b33/html5/thumbnails/16.jpg)
IMR = 79 x 0.17 = 13.3
ww
w.c
ard
io-a
als
t.b
e
IMR and ACF
ETP, Sofia Antipolis, April 2013
IMR in Practice
![Page 17: Index of Microvascular Resistance (IMR) and Absolute ...assets.escardio.org/assets/presentations/OTHER2013/...C Proximal Sensor (Shaft of the Wire) Distal Sensor T 0 1/2 T i 1/2 T](https://reader031.fdocuments.in/reader031/viewer/2022012006/6102a3e361953b73803d9b33/html5/thumbnails/17.jpg)
• Location of the sensor too close to guiding
• Injection time (Tinj) too long
• ‘Wedging’ of the guiding in the ostium
• Tmn too short (< 0.1 s)
• Side branches (especially when far from the guiding)
Bolus Coronary Thermodilution
Pitfalls and Limitations
ww
w.c
ard
io-a
als
t.b
e
IMR and ACF
ETP, Sofia Antipolis, April 2013
![Page 18: Index of Microvascular Resistance (IMR) and Absolute ...assets.escardio.org/assets/presentations/OTHER2013/...C Proximal Sensor (Shaft of the Wire) Distal Sensor T 0 1/2 T i 1/2 T](https://reader031.fdocuments.in/reader031/viewer/2022012006/6102a3e361953b73803d9b33/html5/thumbnails/18.jpg)
IMR = Pa . Tmn . Pd - Pw
Pa - Pw
To take into account the
effect of collaterals
IMR = Pd.Tmn
Driving Pressure Across the Myocardium
Myocardial Flow
ww
w.c
ard
io-a
als
t.b
e
IMR and ACF
ETP, Sofia Antipolis, April 2013
IMR: Definition
![Page 19: Index of Microvascular Resistance (IMR) and Absolute ...assets.escardio.org/assets/presentations/OTHER2013/...C Proximal Sensor (Shaft of the Wire) Distal Sensor T 0 1/2 T i 1/2 T](https://reader031.fdocuments.in/reader031/viewer/2022012006/6102a3e361953b73803d9b33/html5/thumbnails/19.jpg)
ww
w.c
ard
io-a
als
t.b
e
IMR and ACF
ETP, Sofia Antipolis, April 2013
1. Concept(s) of dilution methods
2. Bolus injection: IMR
3. Continuous Infusion: ACF
![Page 20: Index of Microvascular Resistance (IMR) and Absolute ...assets.escardio.org/assets/presentations/OTHER2013/...C Proximal Sensor (Shaft of the Wire) Distal Sensor T 0 1/2 T i 1/2 T](https://reader031.fdocuments.in/reader031/viewer/2022012006/6102a3e361953b73803d9b33/html5/thumbnails/20.jpg)
Constant Flow of injectate
(mL)
Concentration (temperature)
over time (g/mL/min)
ww
w.c
ard
io-a
als
t.b
e
IMR and ACF
ETP, Sofia Antipolis, April 2013
To be known: Flow of the pump (20 mL/min)
Proximal Temperature
Distal Temperature
Continuous Infusion: ACF
Q = Qi x x 1.08 TiT
Aarnoudse W et al J Am Coll Cardiol 2007;50:2294
![Page 21: Index of Microvascular Resistance (IMR) and Absolute ...assets.escardio.org/assets/presentations/OTHER2013/...C Proximal Sensor (Shaft of the Wire) Distal Sensor T 0 1/2 T i 1/2 T](https://reader031.fdocuments.in/reader031/viewer/2022012006/6102a3e361953b73803d9b33/html5/thumbnails/21.jpg)
... ....
Constant Flow of injectate
(mL)
Concentration over time
(g/mL/min)
ww
w.c
ard
io-a
als
t.b
e
IMR and ACF
ETP, Sofia Antipolis, April 2013
To be known: Flow of the pump (20 mL/min)
Proximal Temperature
Distal Temperature
Continuous Infusion: ACF
Q = Qi x x 1.08 TiT
Aarnoudse W et al J Am Coll Cardiol 2007;50:2294
![Page 22: Index of Microvascular Resistance (IMR) and Absolute ...assets.escardio.org/assets/presentations/OTHER2013/...C Proximal Sensor (Shaft of the Wire) Distal Sensor T 0 1/2 T i 1/2 T](https://reader031.fdocuments.in/reader031/viewer/2022012006/6102a3e361953b73803d9b33/html5/thumbnails/22.jpg)
Catheter for continuous saline infusion
•4 side holes allowing optimal mixing of saline
•Minimal or no saline dripping through the distal port
ww
w.c
ard
io-a
als
t.b
e
IMR and ACF
ETP, Sofia Antipolis, April 2013
![Page 23: Index of Microvascular Resistance (IMR) and Absolute ...assets.escardio.org/assets/presentations/OTHER2013/...C Proximal Sensor (Shaft of the Wire) Distal Sensor T 0 1/2 T i 1/2 T](https://reader031.fdocuments.in/reader031/viewer/2022012006/6102a3e361953b73803d9b33/html5/thumbnails/23.jpg)
Catheter for continuous saline infusion
•4 side holes allowing optimal mixing of saline
•Minimal or no saline dripping through the distal port
ww
w.c
ard
io-a
als
t.b
e
IMR and ACF
ETP, Sofia Antipolis, April 2013
![Page 24: Index of Microvascular Resistance (IMR) and Absolute ...assets.escardio.org/assets/presentations/OTHER2013/...C Proximal Sensor (Shaft of the Wire) Distal Sensor T 0 1/2 T i 1/2 T](https://reader031.fdocuments.in/reader031/viewer/2022012006/6102a3e361953b73803d9b33/html5/thumbnails/24.jpg)
TIME
Temperature
Q = Qi x x 1.08 Tb - Ti
Tb - T
Q = Qi x x 1.08 Ti
T
0
Tb
T
Ti
Start infusion (20 mL/min)
ww
w.c
ard
io-a
als
t.b
e
IMR and ACF
ETP, Sofia Antipolis, April 2013
Continuous Infusion: ACF
Aarnoudse W et al J Am Coll Cardiol 2007;50:2294
![Page 25: Index of Microvascular Resistance (IMR) and Absolute ...assets.escardio.org/assets/presentations/OTHER2013/...C Proximal Sensor (Shaft of the Wire) Distal Sensor T 0 1/2 T i 1/2 T](https://reader031.fdocuments.in/reader031/viewer/2022012006/6102a3e361953b73803d9b33/html5/thumbnails/25.jpg)
Tip guiding catheter
Infusion catheter
sensor
ww
w.c
ard
io-a
als
t.b
e
IMR and ACF
ETP, Sofia Antipolis, April 2013
Continuous Infusion: ACF
![Page 26: Index of Microvascular Resistance (IMR) and Absolute ...assets.escardio.org/assets/presentations/OTHER2013/...C Proximal Sensor (Shaft of the Wire) Distal Sensor T 0 1/2 T i 1/2 T](https://reader031.fdocuments.in/reader031/viewer/2022012006/6102a3e361953b73803d9b33/html5/thumbnails/26.jpg)
Q = Qi x x 1.08 Tb - Ti
Tb - T
Q = Qi x x 1.08 Ti
T
ww
w.c
ard
io-a
als
t.b
e
IMR and ACF
ETP, Sofia Antipolis, April 2013
![Page 27: Index of Microvascular Resistance (IMR) and Absolute ...assets.escardio.org/assets/presentations/OTHER2013/...C Proximal Sensor (Shaft of the Wire) Distal Sensor T 0 1/2 T i 1/2 T](https://reader031.fdocuments.in/reader031/viewer/2022012006/6102a3e361953b73803d9b33/html5/thumbnails/27.jpg)
hyperemia
steady state infusion
sensor back into infusion catheter
end of procedure
Tb=0 T=-1.1
Ti=-7.1
Q = 25 x (7.1 / 0.97) x 1.08 = 173 mL/min
Continuous Infusion: ACF
ww
w.c
ard
io-a
als
t.b
e
IMR and ACF
ETP, Sofia Antipolis, April 2013
T=0 Tb=-0.97Tk=-7.1
![Page 28: Index of Microvascular Resistance (IMR) and Absolute ...assets.escardio.org/assets/presentations/OTHER2013/...C Proximal Sensor (Shaft of the Wire) Distal Sensor T 0 1/2 T i 1/2 T](https://reader031.fdocuments.in/reader031/viewer/2022012006/6102a3e361953b73803d9b33/html5/thumbnails/28.jpg)
Guiding catheter
Infusion catheter
Sensor
ww
w.c
ard
io-a
als
t.b
e
IMR and ACF
ETP, Sofia Antipolis, April 2013
![Page 29: Index of Microvascular Resistance (IMR) and Absolute ...assets.escardio.org/assets/presentations/OTHER2013/...C Proximal Sensor (Shaft of the Wire) Distal Sensor T 0 1/2 T i 1/2 T](https://reader031.fdocuments.in/reader031/viewer/2022012006/6102a3e361953b73803d9b33/html5/thumbnails/29.jpg)
ww
w.c
ard
io-a
als
t.b
e
IMR and ACF
ETP, Sofia Antipolis, April 2013
![Page 30: Index of Microvascular Resistance (IMR) and Absolute ...assets.escardio.org/assets/presentations/OTHER2013/...C Proximal Sensor (Shaft of the Wire) Distal Sensor T 0 1/2 T i 1/2 T](https://reader031.fdocuments.in/reader031/viewer/2022012006/6102a3e361953b73803d9b33/html5/thumbnails/30.jpg)
ww
w.c
ard
io-a
als
t.b
e
IMR and ACF
ETP, Sofia Antipolis, April 2013
![Page 31: Index of Microvascular Resistance (IMR) and Absolute ...assets.escardio.org/assets/presentations/OTHER2013/...C Proximal Sensor (Shaft of the Wire) Distal Sensor T 0 1/2 T i 1/2 T](https://reader031.fdocuments.in/reader031/viewer/2022012006/6102a3e361953b73803d9b33/html5/thumbnails/31.jpg)
Absolute hyperemic coronary flow: 126 mL/min
Minimal myocardial resistance: 0.436 mm Hg/mL/min
Infusion catheter
Sensor
ww
w.c
ard
io-a
als
t.b
e
IMR and ACF
ETP, Sofia Antipolis, April 2013
![Page 32: Index of Microvascular Resistance (IMR) and Absolute ...assets.escardio.org/assets/presentations/OTHER2013/...C Proximal Sensor (Shaft of the Wire) Distal Sensor T 0 1/2 T i 1/2 T](https://reader031.fdocuments.in/reader031/viewer/2022012006/6102a3e361953b73803d9b33/html5/thumbnails/32.jpg)
ww
w.c
ard
io-a
als
t.b
e
IMR and ACF
ETP, Sofia Antipolis, April 2013
DOGS MAN
ACF: Validation Studies
Aarnoudse W et al J Am Coll Cardiol 2007;50:2294
![Page 33: Index of Microvascular Resistance (IMR) and Absolute ...assets.escardio.org/assets/presentations/OTHER2013/...C Proximal Sensor (Shaft of the Wire) Distal Sensor T 0 1/2 T i 1/2 T](https://reader031.fdocuments.in/reader031/viewer/2022012006/6102a3e361953b73803d9b33/html5/thumbnails/33.jpg)
ww
w.c
ard
io-a
als
t.b
e
IMR and ACF
ETP, Sofia Antipolis, April 2013
Conclusion
Coronary thermodilution using a bolus injection
allows the mesurements of the mean transit time
(Tmn, an index of coronary flow) and of IMR
Coronary thermodilution by continuous infusion allows
the measurement of absolute coronary blood flow