Incidental G-6-PD Identification
Using the PerkinElmer GSP Neonatal GALT Kit
APHL NBS> Symposium October 28, 2014
![Page 2: Incidental G-6-PD Identification Using the PerkinElmer GSP ...PerkinElmer Neonatal GALT kit . 4 . Beutler Assay to GSP GALT . Screening for Galactosemia • 1964 to February 2012 –](https://reader033.fdocuments.in/reader033/viewer/2022053007/5f0ae4c97e708231d42ddcb2/html5/thumbnails/2.jpg)
Project Initiation
![Page 3: Incidental G-6-PD Identification Using the PerkinElmer GSP ...PerkinElmer Neonatal GALT kit . 4 . Beutler Assay to GSP GALT . Screening for Galactosemia • 1964 to February 2012 –](https://reader033.fdocuments.in/reader033/viewer/2022053007/5f0ae4c97e708231d42ddcb2/html5/thumbnails/3.jpg)
Instrumentation
3
AutoDelfia
GSP
![Page 4: Incidental G-6-PD Identification Using the PerkinElmer GSP ...PerkinElmer Neonatal GALT kit . 4 . Beutler Assay to GSP GALT . Screening for Galactosemia • 1964 to February 2012 –](https://reader033.fdocuments.in/reader033/viewer/2022053007/5f0ae4c97e708231d42ddcb2/html5/thumbnails/4.jpg)
Instrumentation
• PerkinElmer AutoDelfia » T4, TSH, 17-OHP and IRT
• Introduction of Genetic Specimen Processor (GSP) » Transfer of AutoDelfia tests » Expansion capability to add GALT and
Biotinidase • Decision made to add GALT to GSP using the
PerkinElmer Neonatal GALT kit
4
![Page 5: Incidental G-6-PD Identification Using the PerkinElmer GSP ...PerkinElmer Neonatal GALT kit . 4 . Beutler Assay to GSP GALT . Screening for Galactosemia • 1964 to February 2012 –](https://reader033.fdocuments.in/reader033/viewer/2022053007/5f0ae4c97e708231d42ddcb2/html5/thumbnails/5.jpg)
Beutler Assay to GSP GALT
![Page 6: Incidental G-6-PD Identification Using the PerkinElmer GSP ...PerkinElmer Neonatal GALT kit . 4 . Beutler Assay to GSP GALT . Screening for Galactosemia • 1964 to February 2012 –](https://reader033.fdocuments.in/reader033/viewer/2022053007/5f0ae4c97e708231d42ddcb2/html5/thumbnails/6.jpg)
Screening for Galactosemia
• 1964 to February 2012 – Beutler assay • Detection of galactose-1-uridyltransferase
(GALT) enzyme • Presence or absence of fluorescence • Results read visually using a UV light box
6
Abnormal
Normal
![Page 7: Incidental G-6-PD Identification Using the PerkinElmer GSP ...PerkinElmer Neonatal GALT kit . 4 . Beutler Assay to GSP GALT . Screening for Galactosemia • 1964 to February 2012 –](https://reader033.fdocuments.in/reader033/viewer/2022053007/5f0ae4c97e708231d42ddcb2/html5/thumbnails/7.jpg)
Beutler Assay
7
• Manual method • In-house reagents • Visually read • Qualitative • Subjective
![Page 8: Incidental G-6-PD Identification Using the PerkinElmer GSP ...PerkinElmer Neonatal GALT kit . 4 . Beutler Assay to GSP GALT . Screening for Galactosemia • 1964 to February 2012 –](https://reader033.fdocuments.in/reader033/viewer/2022053007/5f0ae4c97e708231d42ddcb2/html5/thumbnails/8.jpg)
PE GSP GALT Assay
• Automated method • Manufactured reagents
» Adaptation of Beutler/Baluda assay
• Instrument read • Quantitative • Objective
8
![Page 9: Incidental G-6-PD Identification Using the PerkinElmer GSP ...PerkinElmer Neonatal GALT kit . 4 . Beutler Assay to GSP GALT . Screening for Galactosemia • 1964 to February 2012 –](https://reader033.fdocuments.in/reader033/viewer/2022053007/5f0ae4c97e708231d42ddcb2/html5/thumbnails/9.jpg)
Validation
• Beutler to GSP Neonatal GALT Kit comparisons
• 2,800 blood spot specimens » 1,411 first screen specimens (zero to 7 days of
age) » 1,359 second screen specimens (>7 days of age) » All Beutler normal
• CDC PT samples » Beutler normal, reduced or no fluorescence
9
![Page 10: Incidental G-6-PD Identification Using the PerkinElmer GSP ...PerkinElmer Neonatal GALT kit . 4 . Beutler Assay to GSP GALT . Screening for Galactosemia • 1964 to February 2012 –](https://reader033.fdocuments.in/reader033/viewer/2022053007/5f0ae4c97e708231d42ddcb2/html5/thumbnails/10.jpg)
Validation Results
• First specimens - zero through 7 days of age » Mean 12.58 U/dL » Standard deviation 3.92
• Second specimens - > 7 days of age » Mean of 12.43 U/dL » Standard deviation of 11.30 » Higher variability
• CDC specimens – Interpretation matched Beutler GALT test results
10
![Page 11: Incidental G-6-PD Identification Using the PerkinElmer GSP ...PerkinElmer Neonatal GALT kit . 4 . Beutler Assay to GSP GALT . Screening for Galactosemia • 1964 to February 2012 –](https://reader033.fdocuments.in/reader033/viewer/2022053007/5f0ae4c97e708231d42ddcb2/html5/thumbnails/11.jpg)
Validation Results
• Classic galactosemia specimens » GSP GALT values < 1.5 U/dL
• Reduced fluorescence specimens » GSP GALT values < 5.50 U/dL
• Proposed cutoffs » Abnormal < 2.5 U/dL » Borderline 2.5 to 3.49 U/dL » Normal > 3.50 U/dL
11
![Page 12: Incidental G-6-PD Identification Using the PerkinElmer GSP ...PerkinElmer Neonatal GALT kit . 4 . Beutler Assay to GSP GALT . Screening for Galactosemia • 1964 to February 2012 –](https://reader033.fdocuments.in/reader033/viewer/2022053007/5f0ae4c97e708231d42ddcb2/html5/thumbnails/12.jpg)
Validation Results
• False Positive Predictions » < 2.50 U/dL would produce 11 false positives in
addition to the true positives
• Borderline range »< 3.00 U/dL – 15 false positives »< 4.00 U/dL – 1 false positive »Selected cutoff of < 3.50 U/dL – 8 more
borderline positives
12
![Page 13: Incidental G-6-PD Identification Using the PerkinElmer GSP ...PerkinElmer Neonatal GALT kit . 4 . Beutler Assay to GSP GALT . Screening for Galactosemia • 1964 to February 2012 –](https://reader033.fdocuments.in/reader033/viewer/2022053007/5f0ae4c97e708231d42ddcb2/html5/thumbnails/13.jpg)
Assay Implementation
![Page 14: Incidental G-6-PD Identification Using the PerkinElmer GSP ...PerkinElmer Neonatal GALT kit . 4 . Beutler Assay to GSP GALT . Screening for Galactosemia • 1964 to February 2012 –](https://reader033.fdocuments.in/reader033/viewer/2022053007/5f0ae4c97e708231d42ddcb2/html5/thumbnails/14.jpg)
Screening Results
• Increased number of presumptive positives with the GSP GALT » 120 in 2011 (Beutler assay) » 681 in 2012
• Mainly reduced fluorescence values • Majority from Hawaii • Asian and Pacific Islander population
14
![Page 15: Incidental G-6-PD Identification Using the PerkinElmer GSP ...PerkinElmer Neonatal GALT kit . 4 . Beutler Assay to GSP GALT . Screening for Galactosemia • 1964 to February 2012 –](https://reader033.fdocuments.in/reader033/viewer/2022053007/5f0ae4c97e708231d42ddcb2/html5/thumbnails/15.jpg)
Diagnostic Results
• Galactosemia false positives • Confirmation of glucose-6-phosphate
dehydrogenase deficiency (G-6-PD) • Already well known to be common in the
Hawaiian population • Testing routinely performed in the state
15
![Page 16: Incidental G-6-PD Identification Using the PerkinElmer GSP ...PerkinElmer Neonatal GALT kit . 4 . Beutler Assay to GSP GALT . Screening for Galactosemia • 1964 to February 2012 –](https://reader033.fdocuments.in/reader033/viewer/2022053007/5f0ae4c97e708231d42ddcb2/html5/thumbnails/16.jpg)
• Deficiency in glucose-6-phosphate dehydrogenase
• Insufficient amounts – red blood cell destruction
• Triggered by infections, severe stress, certain foods or drugs
16
G-6-PD
![Page 17: Incidental G-6-PD Identification Using the PerkinElmer GSP ...PerkinElmer Neonatal GALT kit . 4 . Beutler Assay to GSP GALT . Screening for Galactosemia • 1964 to February 2012 –](https://reader033.fdocuments.in/reader033/viewer/2022053007/5f0ae4c97e708231d42ddcb2/html5/thumbnails/17.jpg)
Assay Evaluation
![Page 18: Incidental G-6-PD Identification Using the PerkinElmer GSP ...PerkinElmer Neonatal GALT kit . 4 . Beutler Assay to GSP GALT . Screening for Galactosemia • 1964 to February 2012 –](https://reader033.fdocuments.in/reader033/viewer/2022053007/5f0ae4c97e708231d42ddcb2/html5/thumbnails/18.jpg)
Cutoff Analysis and Revision
• 6 month data review • Evaluation of confirmed cases
» GALT deficient » D/G » Duartes » Carrier » Other variant
18
![Page 19: Incidental G-6-PD Identification Using the PerkinElmer GSP ...PerkinElmer Neonatal GALT kit . 4 . Beutler Assay to GSP GALT . Screening for Galactosemia • 1964 to February 2012 –](https://reader033.fdocuments.in/reader033/viewer/2022053007/5f0ae4c97e708231d42ddcb2/html5/thumbnails/19.jpg)
Cutoff Analysis and Revision
• Beutler assay classic Galactosemia specimens » GSP GALT values < 1.5 U/dL
• Beutler assay reduced fluorescence specimens » GSP GALT values < 5.50 U/dL
• Revised cutoffs » Abnormal < 2.5 U/dL < 2.0 U/dL » Borderline 2.5 to 3.49 U/dL 2.0 to 3.49 U/dL » Normal > 3.50 U/dL No change
19
![Page 20: Incidental G-6-PD Identification Using the PerkinElmer GSP ...PerkinElmer Neonatal GALT kit . 4 . Beutler Assay to GSP GALT . Screening for Galactosemia • 1964 to February 2012 –](https://reader033.fdocuments.in/reader033/viewer/2022053007/5f0ae4c97e708231d42ddcb2/html5/thumbnails/20.jpg)
Beutler Assay to GSP GALT
Statistics
![Page 21: Incidental G-6-PD Identification Using the PerkinElmer GSP ...PerkinElmer Neonatal GALT kit . 4 . Beutler Assay to GSP GALT . Screening for Galactosemia • 1964 to February 2012 –](https://reader033.fdocuments.in/reader033/viewer/2022053007/5f0ae4c97e708231d42ddcb2/html5/thumbnails/21.jpg)
Beutler vs. GSP GALT Cases
Beutler GSP GALT 2009 2010 2011 2012 2013 As of
10/2014 Presumptive Positive 189 108 120 681 922 660
Classic Galactosemia 0 3 2 1 3 1 G-6-PD 0 0 0 0 5 46 Duarte Variant 43 14 9 12 30 14 G/N Heterozygote 8 2 1 0 12 4
Specimens Screened 315,670 301,897 293,405 292,372 291,946 216,480
Average PP Frequency 1:1,094 1:197 21
![Page 22: Incidental G-6-PD Identification Using the PerkinElmer GSP ...PerkinElmer Neonatal GALT kit . 4 . Beutler Assay to GSP GALT . Screening for Galactosemia • 1964 to February 2012 –](https://reader033.fdocuments.in/reader033/viewer/2022053007/5f0ae4c97e708231d42ddcb2/html5/thumbnails/22.jpg)
Summary
![Page 23: Incidental G-6-PD Identification Using the PerkinElmer GSP ...PerkinElmer Neonatal GALT kit . 4 . Beutler Assay to GSP GALT . Screening for Galactosemia • 1964 to February 2012 –](https://reader033.fdocuments.in/reader033/viewer/2022053007/5f0ae4c97e708231d42ddcb2/html5/thumbnails/23.jpg)
Summary
• GSP Neonatal GALT Kit »Greater sensitivity than Beutler in G-6-PD detection »Unexpected increase in presumptive
positives with population bias »Increased short-term follow-up »Increased diagnostic costs – GALT and
G-6-PD analysis vs. GALT alone
23
![Page 24: Incidental G-6-PD Identification Using the PerkinElmer GSP ...PerkinElmer Neonatal GALT kit . 4 . Beutler Assay to GSP GALT . Screening for Galactosemia • 1964 to February 2012 –](https://reader033.fdocuments.in/reader033/viewer/2022053007/5f0ae4c97e708231d42ddcb2/html5/thumbnails/24.jpg)
Incidental vs. Supplemental
Findings
![Page 25: Incidental G-6-PD Identification Using the PerkinElmer GSP ...PerkinElmer Neonatal GALT kit . 4 . Beutler Assay to GSP GALT . Screening for Galactosemia • 1964 to February 2012 –](https://reader033.fdocuments.in/reader033/viewer/2022053007/5f0ae4c97e708231d42ddcb2/html5/thumbnails/25.jpg)
Definitions
• in·ci·den·tal » adjective \ˌin(t)-sə-ˈden-təl\ : happening as
a minor part or result of something else
• sup·ple·men·tal » adjective \ˌsə-plə-ˈmen-təl\ : added to
something else to make it complete
25
![Page 26: Incidental G-6-PD Identification Using the PerkinElmer GSP ...PerkinElmer Neonatal GALT kit . 4 . Beutler Assay to GSP GALT . Screening for Galactosemia • 1964 to February 2012 –](https://reader033.fdocuments.in/reader033/viewer/2022053007/5f0ae4c97e708231d42ddcb2/html5/thumbnails/26.jpg)
26