Enric Roche MD , PHD IS IT POSSIBLE TO IMPROVE TESSARI´S METHOD ? ENRIC ROCHE, MD PhD. Director CLINICA VASCULAR BARCELONA Chief Vascular service HOSPITAL UNIVERSITARI SAGRAT COR & HOSPITAL GENERAL DE CATALUNYA RAMON PONS, QD Director INSTITUT QUÍMICA AVANÇADA , CONSEJO SUPERIOR DE INVESTIGACIONES CIENTÍFICAS ANTONI PUIG, MD CLINICA VASCULAR BARCELONA

Transcript of Ii foamer device presentation site 2015

Page 1: Ii foamer device presentation site 2015

Enric Roche MD , PHD

IS IT POSSIBLE TO IMPROVE TESSARI´S METHOD ?

ENRIC ROCHE, MD PhD. Director CLINICA VASCULAR BARCELONAChief Vascular service HOSPITAL UNIVERSITARI SAGRAT COR & HOSPITAL GENERAL DE CATALUNYARAMON PONS, QD Director INSTITUT QUÍMICA AVANÇADA , CONSEJO SUPERIOR DE INVESTIGACIONES CIENTÍFICAS ANTONI PUIG, MDCLINICA VASCULAR BARCELONA

Page 2: Ii foamer device presentation site 2015

Enric Roche MD , PHD

I declare I haven´t any conflict of interest related to this presentation.

Page 3: Ii foamer device presentation site 2015

Enric Roche MD , PHD

Several strategies to ablate Major and minor saphenas

TO STRIP BURN

ELECTRIFYBOIL

STICKEMBOLIZE

In your office,

Which % of Major/Minor saphenas are involved ?

Page 4: Ii foamer device presentation site 2015

Enric Roche MD , PHD

Main concerns…

Foam sclerotherapy has great variability

about syringes material, three way stopcock, type of gas, ratio sclerosant /gas, fabrication system, injection system, strategy , temperature, patient position,

Safety is importantPhysiological gases are safer than air, specially when a large volume of foam is used Cabrera J, Patología Vascular. 1995; Wright D, Phlebology 2006; Morrison N. J Vasc Surg, 2008, Beckitt T, , Beckitt T, Eur J Vasc Endovasc Surg 2011., Hesse G, Phlebology 2012,

Versatility is a plusEasy to obtain a 1-2% stable foam …How many cases could be treated with high concentrations sclerosants ? What about foams less concentrated ( < 0,5% …)

Page 5: Ii foamer device presentation site 2015

Enric Roche MD , PHD

FROTHER SYSTEM

Page 6: Ii foamer device presentation site 2015

Enric Roche MD , PHD

OBJECTIVE:

To demonstrate the feasibility of a new system of standardized foam fabrication and its superiority in front of Tessari´s method in terms of stability and versatility.

MATERIAL AND METHOD

TESSARI´S METHOD -0,5% POLIDOCANOL Foam with AIR - 2% POL Foam with AIR - 2% POL Foam with O2/CO2 (50%/50%)

FROTHER-0,5% POLIDOCANOL Foam with AIR - 2% POL Foam with AIR - 2% POL Foam with O2/CO2 (50%/50%)

Page 7: Ii foamer device presentation site 2015

Enric Roche MD , PHD

Tessari´s method: Two 10 cc (BD Brand – latex free ) with A three-way stopcock key ( B/ Braun) . A default 1/3 ratio was decided using 3 cc of POL / 9 cc of AIR or O2/CO2. ( O2/CO2 Experiments container was filled with gas bottle) After 20 passes Small amount In a slide for

measurements

Syringe left in a stand position for half-life calculation

MATERIAL AND METHOD

Foam measurements : Diameter of bubbles & Total foam volume A single layer of bubbles covered by a coverslip. Zeiss microscopy with a 2,5x with a Canon Power Shot 90S was used to take pictures. ( final magnification was 2428x1821 μm2 ) .Program used: ImageJ. Threshold particles analysis ( Size 150Infinity, Circularity 0,75-1)

Half –time life calculation: Was defined as Time necessary to get the half amount of sclerosant ( 1,5 ml) in a liquid state. (Total volume of sclerosant : 3 ml)Every 30 seconds was obtained a picture of the syringes in vertically placed position with a camera ( in a fixed position ) on a dark background.

Page 8: Ii foamer device presentation site 2015

Enric Roche MD , PHD

PURESPUME or FROTHER system: Because the ratio liquid /gas is unknow is necesary to calculate previosly the sclerosant volume.Weight the syringe with 10 cc of foam / without foam in a precision scale3 ml of POL was frothed during 2 minutes ( 15sec-300rpm, 15 sec-700rpm,90 sec-1600 rpm) ( O2/CO2 Experiments container was filled with gas bottle)

After 2 min. in a slide

A sample was deposited In a slide for measurements

A stand position for half-life calculation

MATERIAL AND METHOD

Page 9: Ii foamer device presentation site 2015

Enric Roche MD , PHD

RESULTSFrother system foam 2% POL AIR

Frother system foam 2% POL O2/CO2

Tessari´s foam 2% POL AIR

Tessari´s foam 2% POL O2/CO2

Page 10: Ii foamer device presentation site 2015

Enric Roche MD , PHD

RESULTS 1. FOAM MEASUREMENTS - Diameter of bubbles- Total foam volume

% POL FROTHER SYSTEM TESSARI AIR O2/CO2 AIR O2/CO2

Diameter ED- σ Diameter ED-σ

0,5 %134±33μm 109±21μm 38±17μm 25±15μm

2 %104±22μm 74±22μm 65+/-20μm 51±20μm 40±20μm 40±15 μm

IC = 95%POL= PolidocanolED= Estandar deviation

Page 11: Ii foamer device presentation site 2015

Enric Roche MD , PHD

RESULTS 1. FOAM MEASUREMENTS

% POL FROTHER SYSTEM TESSARI AIR O2/CO2 AIR O2/CO2

Diameter ED- σ Diameter ED-σ

0,5 %134±33μm 109±21μm 38±17μm 25±15μm

2 %104±22μm 74±22μm 65+/-20μm 51±20μm 40±20μm 40±15 μm

IC = 95%POL= PolidocanolED= Estandar deviation

- Diameter of bubbles- Total foam volume

Page 12: Ii foamer device presentation site 2015

Enric Roche MD , PHD

RESULTS

FROTHER SYSTEM 2% AIR

Diameter range of FROTHER SYSTEM is lower than TESSARI’S FOAM

TESSARI´S 2% AIR Bubble Range: 25-150 µm Bubble Range: 25-245 µm

Page 13: Ii foamer device presentation site 2015

Enric Roche MD , PHD

RESULTS

2. FOAM STABILITY: Half-time life

% POL FROTHER SYSTEM TESSARI´S FOAMAIR

minutesratio O2/CO2

minutesAIR minutes

O2/CO2 minutes

0,5% 6,5±3m 0,093 1/9 1,1±0,5m

2% 10,6±3m 0,089 1/10

2,2± 0,5m 2,0±0,5m 1,0±05m

IC = 95%

Statistical differences POL= PolidocanolRatio: Proportion of sclerosant/gas

Page 14: Ii foamer device presentation site 2015

Enric Roche MD , PHD

RESULTS

2. FOAM STABILITY: Ratio sclerosant /gas

% POL FROTHER SYSTEM TESSARI´S FOAM

AIR O2/CO2 AIR or O2/CO2

0,5% 0,093 - 1/9 0,25 - 1/3

2% 0,089 - 1/10 0,14 - 1/6 0,25 -1/3

IC = 95%Statistical differences POL= Polidocanol

Ratio: Proportion of sclerosant/gas

Page 15: Ii foamer device presentation site 2015

Enric Roche MD , PHD

RESULTS 2. FOAM STABILITY: Half-time life

FROTHER SYSTEM foam 2% AIR

Tessari´s foam 2% AIR

Page 16: Ii foamer device presentation site 2015

Enric Roche MD , PHD

CONCLUSIONS

2. Tessari´s foam is more heterogeneous : although there are a very little bubbles

3. In our experiment frother´s foam is more stable than Tessari´s foam. (five times more stable compared to Tessari´s method when using air and twice when using a mixture of O2/CO2 )

4. 2% concentration of Polidocanol foam is more stable than 0,5% POL.

1. This system has demonstrated that it is useful in manufacturing foam with air as well as with a gas mixture of O2/CO2.

Page 17: Ii foamer device presentation site 2015

Enric Roche MD , PHD

Comments

Our Tessari´s foam could be different than others foams

The standardization of parameters can be useful in offices where sclerotherapy is a common practice

Versatility is interesting in the office

Some Improvements have been necessary in order to request Health Authorities Approval