ICP MAKING STANDARDS. NUMBER OF STANDARDS Determine the number of standards necessary for your...

8
ICP MAKING STANDARDS

Transcript of ICP MAKING STANDARDS. NUMBER OF STANDARDS Determine the number of standards necessary for your...

Page 1: ICP MAKING STANDARDS. NUMBER OF STANDARDS Determine the number of standards necessary for your measurements (Blank, 100ppb, 500ppb, 1000ppb, etc)

ICPMAKING STANDARDS

Page 2: ICP MAKING STANDARDS. NUMBER OF STANDARDS Determine the number of standards necessary for your measurements (Blank, 100ppb, 500ppb, 1000ppb, etc)

NUMBER OF STANDARDS

Determine the number of standards necessary for your measurements (Blank, 100ppb, 500ppb, 1000ppb, etc)

Page 3: ICP MAKING STANDARDS. NUMBER OF STANDARDS Determine the number of standards necessary for your measurements (Blank, 100ppb, 500ppb, 1000ppb, etc)

BLANK

Blanks are prepared from reagent grade water with 3% trace metal grade nitric acid.

Blanks are typically prepared in 2L lots by adding 60 mL of concentrated HNO3 to 1940 mL of water (using a pre-marked bottle available in the PQ lab).

Page 4: ICP MAKING STANDARDS. NUMBER OF STANDARDS Determine the number of standards necessary for your measurements (Blank, 100ppb, 500ppb, 1000ppb, etc)

IV-ICPMS-71A Standard

• Typically, the ICP standard used IV-ICPMS-71A, which contains of 10 ppm Ag, Al, As, B, Ba, Be, Ca, Cd, Ce, Co, Cr3, Cs, Cu, Dy, Er, Eu, Fe, Ga, Gd, Ho, K, La, Lu, Mg, Mn, Na, Nd, Ni, P, Pb, Pr, Rb, S, Se, Sm, Sr, Th, Tl, Tm, U, V, Yb, Zn.

• This solution is diluted with the blank solution to prepare working ICP standards.

Page 5: ICP MAKING STANDARDS. NUMBER OF STANDARDS Determine the number of standards necessary for your measurements (Blank, 100ppb, 500ppb, 1000ppb, etc)

100 ppb

Dilute the stock 100 times.

For a 200 ml sample pipette 2 ml of the ICP Stock and dilute to 200 g with the blank solution.

Page 6: ICP MAKING STANDARDS. NUMBER OF STANDARDS Determine the number of standards necessary for your measurements (Blank, 100ppb, 500ppb, 1000ppb, etc)

500 ppb

Dilute the stock 500 times.

For a 200 ml sample pipette 10 ml of the ICP Stock and dilute to 200 g with the blank solution.

Page 7: ICP MAKING STANDARDS. NUMBER OF STANDARDS Determine the number of standards necessary for your measurements (Blank, 100ppb, 500ppb, 1000ppb, etc)

1000 ppb

Dilute the stock 1000 times.

For a 200 ml sample pipette 20 ml of the ICP Stock and dilute to 200 g with the blank solution.

Page 8: ICP MAKING STANDARDS. NUMBER OF STANDARDS Determine the number of standards necessary for your measurements (Blank, 100ppb, 500ppb, 1000ppb, etc)

Hints on Preparing Solutions

• All bottles must be very clean.

• Use gravimetric measurements whenever possible.

• Ie. Weigh solutions instead of pipetting.

• Do not pipette directly out of ICP stock solutions; pour instead into a falcon tube. This avoids contamination.

• Check the calibrates on pipettes when making up standards.