Ic Principle

download Ic Principle

of 13

Transcript of Ic Principle

  • 8/2/2019 Ic Principle

    1/13

    AN INTRODUCTION

    TO

    ION CHROMATOGRAPHY

    Basic Principles of Chromatography

    Chromatography is an experimental technique used for analyzing and/or

    separating mixtures of chemical substances. There are many different

    kinds of chromatography, among them paper chromatography, gas

    chromatography, liquid chromatography, and ion-exchange

    chromatography. An ion chromatograph is an example of the latter kind of

    chromatography.All chromatographic methods share the same basic principles and

    mode of operation. In every case, a sample of the mixture to be analyzed

    (the analyte) is applied to some stationary fixed material (the adsorbent)

    and then a second material (the eluent) is passed through or over the

    stationary phase. The compounds contained in the analyte are then

    partitioned between the stationary adsorbent and the moving eluent. The

    success of the method depends on the fact that different materials adhere

    to the adsorbent with different forces. Some adhere to the adsorbent more

    strongly than others and are therefore moved through the adsorbent more

    slowly as the eluent flows over them. Other components of the analyte are

    less strongly adsorbed on the stationary phase and are moved along more

  • 8/2/2019 Ic Principle

    2/13

  • 8/2/2019 Ic Principle

    3/13

    The number of peaks corresponds to the minimum number of different

    substances (compounds or ions) contained in the analyte. If the analyte is

    found to display only a single peak, it is an indication that it is composed of

    only a single component, i.e., it is pure, although rigorous confirmation of

    purity may require additional testing.Ion ChromatographyA. The Injector and the Eluent

    Commercial ion chromatography systems have become available

    within the past ten to fifteen years. The basic components of such a system

    are shown in the cartoon below:

    For anion chromatography the eluent is a dilute aqueous solution of

    sodium bicarbonate and sodium carbonate, each about 1-4 mM in

    concentration. The water used for the preparation of the eluent is of the

    highest purity - Type I water which is virtually free from any contaminating

    ions. The sample is injected into the instrument through the injection

    system. The eluent, which the pump draws from the eluent jar and forces

    through the system, then carries the analyte onto the ion exchange column

    under a pressure of about 1000 psi (ca 65 atm).

  • 8/2/2019 Ic Principle

    4/13

    B. The Ion Exchange ColumnThe ion exchange column (sometimes called the analytical column)

    is tightly packed with the stationary adsorbent. This adsorbent is usually

    composed of tiny polymer beads which have positive charge centers

    (actually ammonium type cations) attached to their surfaces with covalent

    chemical bonds. If we could magnify the column packing, we imagine that

    it would look something like the cartoon which follows:

    When just the pure eluent is flowing through the column these

    positive centers on the polymer surface attract the negatively charged

    bicarbonate and carbonate ions in the eluent. But as the anions contained

    in the analyte sample begin to enter the column, these anions, which are

    also attracted to the positive centers on the polymer surface, may replace

    (i.e., they exchange with) the bicarbonate and carbonate ions stuck on the

    polymer surface. The analyte anions and the eluent anions compete with

    each other for the positive centers on the polymer surface. As a result of

    these interactions a rather complex equilibrium is set up as the analyte is

    carried through the column by the eluent. This equilibrium can be

    described by the cartoon below:

  • 8/2/2019 Ic Principle

    5/13

    How tightly the competing anions stick (i.e., how strongly they are

    attracted) to the surface of the polymer beads depends on the size of the

    anion and its charge. Usually, the greater the charge on the anion the more

    strongly it is attracted to the surface of the polymer beads (so, for example,

    sulfate ions are more tightly held to the polymer beads than chloride ions).

    Also, larger anions generally move more slowly through the column than

    smaller anions (e.g., bromide ions usually are more tightly bound to the

    polymer beads than chloride ions).

    So now we have a situation in which the eluent is trying to move the

    anions of the analyte forward through the column, but the polymer beads

    are trying to hold the anions in place by attracting them to the positive

    charge centers. Obviously, anions can only move forward through the

    column when they are in solution. So the speed at which a given type of

    anion moves through the column is dependent on the relative amount of

    time the ion is in solution. If an ion is only weakly attracted to the positive

    centers on the surface of the polymer beads, it will spend relatively more of

    its time in the solution phase than an anion that is strongly attracted to the

    polymer surface. So these anions move through the column more quickly.

    The more strongly the anion is attracted to the surface of the polymer

    beads, the more slowly it moves through the column.

  • 8/2/2019 Ic Principle

    6/13

    The result is that what was injected as a homogeneous solution of analyte

    anions at the top of the column, separates into bands of different kinds of

    ions as they travel through the column. For example, if we could see what

    happened to a solution of chloride, bromide and sulfate ions as they

    traveled through the ion exchange column, after sometime the column

    might look some thing like this:

    While the details of what goes on inside the column is more

    complicated than described above, perhaps this description will provide

    you with a qualitative understanding of how different anions can be

    separated on such a column.

  • 8/2/2019 Ic Principle

    7/13

    Detection and the Suppresser ColumnThe detector that is most commonly used in ion chromatography is

    a conductivity cell which measures the conductance of the solution

    passing through it. The conductance of a solution is proportional to the

    concentration of the ions dissolved in the solution. Because we use

    conductivity measurements to detect the different anions as they emerge

    from the column, it is essential to pass the eluent through a suppresser

    column before entering the detector. Heres why. The eluent contains

    relatively high concentrations (1-5 mM) of sodium bicarbonate and sodium

    carbonate - both strong electrolytes. If such a solution were passed directly

    into a conductivity detector (which measures the electrical conductance of

    the eluent solution), a large signal would be obtained just from the

    presence of these eluent ions. The actual concentration of the analyte

    anions of interest is usually only a tiny fraction of the concentration of the

    bicarbonate and carbonate ions contributed by the eluent itself. Usually the

    analyte anions of interest are present in the parts per million level or less;

    however the bicarbonate and carbonate ions on the eluent are present in

    much greater concentration approximately 100 times as large as the

    anions of interest. Thus the bicarbonate and carbonate ions impart a large

    conductivity to the solution entering the detector and the additional

    conductivity imparted to the solution of the eluent by the presence of a tiny

    amount of an anion of interest coming from the sample is extremely small.

    That is, we would have a very hard time measuring the difference in the

    conductivity of the pure eluent itself and the conductivity of the eluentcontaining a tiny amount of an anion of interest. In other words, the

    sensitivity of the measurement would be too low to be of value.

  • 8/2/2019 Ic Principle

    8/13

    To get around this problem, after emerging from the column and before

    entering the conductivity detector, we pass the solution through a

    suppresser column which effectively removes all the bicarbonate and

    carbonate ions from the solution, leaving only the analyte ions of interest

    dissolved in the solution. When the resulting solution is now passed

    through the conductivity cell, we can readily detect the signal due to the

    presence of a small amount of the analyte ion, because in this case the

    reference signal we are comparing it to is that of pure water which has an

    exceedingly low conductivity. So by removing the large quantities of the

    eluent anions, bicarbonate and carbonate, we greatly increase thesensitivity of the measurement. Indeed, using suppression, it is possible to

    detect ion concentrations as low as parts per billion in the analyte sample. The suppresser column essentially is another ion exchange column.

    Although there are different designs for such columns, we may think of

    these columns also as being packed with small polymer beads, but these

    beads carry acidic protons (H+1) on their surface:

  • 8/2/2019 Ic Principle

    9/13

    As the solution from the analytical column flows into the suppresser

    column, the carbonate and bicarbonate ions, both of which are basic,

    combine chemically with the protons on the polymer surface forming

    carbonic acid which , being unstable in aqueous solution, decomposes to

    carbon dioxide gas and water. In this way the carbonate and bicarbonate

    ions are removed from the solution:HCO3

    -1 + H+1 ----> [H2CO3] ---> H2O + CO2

    CO3-2 + 2 H+1 ----> [H2CO3] ---> H2O + CO2

    The spaces that are left on the polymer surface by the departure of the

    protons are filled by sodium ions provided by the eluent solution (i.e.,

    sodium ions undergo exchange with protons on the surface of the polymer

    beads). This process essentially removes all of the anions (bicarbonate and

    carbonate) that were present in the eluent solution but leaves the anions

    that were contained in the analyte dissolved in the solution. The reason

    that the analyte anions are not removed from the solution by the

    suppresser column is that anions like chloride, bromide, nitrate, sulfate,

    etc., are non-basic and do not combine with protons as they pass through

    the suppresser column.When only pure eluent is entering the suppresser column, the

    bicarbonate and carbonate anions are removed, the sodium ions are

    retained on the column, and what emerges from the end of the suppresser

    column is essentially pure water which has exceedingly low conductivity.

    In this condition the conductivity detector is zeroed. Then when eluent

    carrying a band of analyte anions enters the suppresser column, the

    bicarbonate and carbonate anions are again removed as before, but the

    analyte anions emerge from the suppresser column unscathed. When they

    enter the conductivity cell of the detector, the conductance of the solution

  • 8/2/2019 Ic Principle

    10/13

    increases and this increase is measured. The greater the concentration of

    the analyte anions the greater the conductivity and the larger is the peak

    that is recorded in the printout.Qualitative Identification of Anions

    The identification of the anions present in an analyte mixture is

    achieved by comparing the results of the analysis with results obtained by

    analyzing samples of known substances. The key to such identification

    rests on the fact that under a fixed set of analytical conditions (type of

    analytical column packing, concentration of eluent electrolytes, etc.) a

    given anion will always take the same amount of time to travel through the

    analytical column. This time is called the retention time (RT) see diagram

    following:

    . Each anion has its own characteristic retention time under a given

    set of conditions. Thus, for example, if we find that a known solution

    containing only sodium chloride, takes 2.76 min to emerge from the

    column, then finding a peak in the chromatogram of an unknown analyte

    which contains a peak at 2.76 minutes allows us to conclude that the

    sample contains chloride ion. In all cases identification of an anion needs

    to be done by comparison of the sample results with the results obtained

    by analysis of known solutions.

  • 8/2/2019 Ic Principle

    11/13

    Quantitative Identification of AnionsOne can also quite readily obtain a measurement of the

    concentration of a given anion in an unknown sample. This measurement

    rests on the fact that the conductivity that a solution displays is directly

    proportional to the concentration of ions it contains. So the larger the peak

    for a given ion in a chromatogram the greater is its concentration, that is,

    the area of the chromatogram peak for a given ion is proportional to the

    concentration of the ion in solution. To determine the concentration of an

    ion in an unknown solution one needs to compare the peak area of the ion

    in the unknown solution with the peak area of that ion is some solution(s)

    in which the concentration of the ion is known, i.e., a standard solution.

    Typically quantitative results are achieved by using external or internal

    standards. The use of external standards is more common.

    Suppose one wants to determine the precise concentration of

    chloride in an unknown solution. First, a set of standard solutions

    (prepared by dissolving pure sodium chloride in pure water) is made in

    which the precise concentration of the chloride ion is known. Typically a

    minimum of three standards are used - say a 100 ppm, 50 ppm and 10 ppm

    solution (ppm = parts per million). Then each of these solutions is analyzed

    on the ion chromatograph and from the results a standard curve is

    obtained. Such a curve typically is obtained by plotting peak height (or

    peak area) in the chromatogram Vs concentration of chloride ion which

    usually produces close to a straight-line graph. A typical result would look

    something like that shown below:

  • 8/2/2019 Ic Principle

    12/13

    Finally, the unknown solution is analyzed and, by measuring

    the peak area for the chloride ion (identified by retention time), and

    comparing it with the standard curve, the concentration of chloride ion in

    the unknown solution can be calculated.The Ion Chromatography ExperimentA. Sample Preparation and Analysis

    The actual experiment is very simple. All that needs to be done is to

    prepare the sample by gravity filtering a small portion (10 mL) of the water

    sample into an ion chromatography analysis tube, label the tube and placeit into the ion chromatography auto-sampler. An identifying name for the

    sample needs to be entered on the computer. After all samples are loaded

    into the auto-sampler and entered into the computer, the analysis begins

    and needs no further attention. The results will be distributed next week

  • 8/2/2019 Ic Principle

    13/13

    B. Analysis of ResultsTo analyze the results of the ion chromatography experiment, we

    will pool the results that were obtained for water samples that were

    collected from the same source. So there will be different sets of results -

    one each for samples collected from each of the different ponds. For each

    of these locations we will prepare a spreadsheet using Excel or some other

    spreadsheet program. Each student should enter his or her data on the

    relevant spreadsheet. The spreadsheet should contain columns for theconcentrations of each of the ions analyzed for - chloride, nitrate,

    phosphate, and sulfate and for the Acid Neutralizing Capacity (ANC) of the

    water sample (you measured this by titration last week in lab). After all the

    individual results are entered, we will calculate an average concentration

    for each ion and a standard deviation for the measurement of each ion

    concentration.This data will then become part of the data entered into the Calvin

    Environmental Assessment Program (CEAP), a program which seeks to

    monitor how human activities on the campus and surrounding

    neighborhoods impact the natural environment