HYDROGEN PEROXIDE STERILISATION IN A ISO EN 13485 ENVIRONMENT Wouter Meert, manager CSSD, UZ Leuven.
-
Upload
tabitha-armstrong -
Category
Documents
-
view
227 -
download
0
Transcript of HYDROGEN PEROXIDE STERILISATION IN A ISO EN 13485 ENVIRONMENT Wouter Meert, manager CSSD, UZ Leuven.
![Page 1: HYDROGEN PEROXIDE STERILISATION IN A ISO EN 13485 ENVIRONMENT Wouter Meert, manager CSSD, UZ Leuven.](https://reader035.fdocuments.in/reader035/viewer/2022062305/5697bfd81a28abf838caec0f/html5/thumbnails/1.jpg)
HYDROGEN PEROXIDE STERILISATION IN A ISO EN
13485 ENVIRONMENT
Wouter Meert, manager CSSD, UZ Leuven
![Page 2: HYDROGEN PEROXIDE STERILISATION IN A ISO EN 13485 ENVIRONMENT Wouter Meert, manager CSSD, UZ Leuven.](https://reader035.fdocuments.in/reader035/viewer/2022062305/5697bfd81a28abf838caec0f/html5/thumbnails/2.jpg)
UZ Leuven
![Page 3: HYDROGEN PEROXIDE STERILISATION IN A ISO EN 13485 ENVIRONMENT Wouter Meert, manager CSSD, UZ Leuven.](https://reader035.fdocuments.in/reader035/viewer/2022062305/5697bfd81a28abf838caec0f/html5/thumbnails/3.jpg)
UZ Leuven Healthcare campus
![Page 4: HYDROGEN PEROXIDE STERILISATION IN A ISO EN 13485 ENVIRONMENT Wouter Meert, manager CSSD, UZ Leuven.](https://reader035.fdocuments.in/reader035/viewer/2022062305/5697bfd81a28abf838caec0f/html5/thumbnails/4.jpg)
![Page 5: HYDROGEN PEROXIDE STERILISATION IN A ISO EN 13485 ENVIRONMENT Wouter Meert, manager CSSD, UZ Leuven.](https://reader035.fdocuments.in/reader035/viewer/2022062305/5697bfd81a28abf838caec0f/html5/thumbnails/5.jpg)
UZ Leuven
• 1900 beds• 9 038 employees• 56 658 hospitalizations• 684 199 consults (2 748/day)• 100 005 one day
hospitalizations• 53 544 surgical procedures
(215/day – 45 operating rooms)
![Page 6: HYDROGEN PEROXIDE STERILISATION IN A ISO EN 13485 ENVIRONMENT Wouter Meert, manager CSSD, UZ Leuven.](https://reader035.fdocuments.in/reader035/viewer/2022062305/5697bfd81a28abf838caec0f/html5/thumbnails/6.jpg)
CSSD UZ Leuven
![Page 7: HYDROGEN PEROXIDE STERILISATION IN A ISO EN 13485 ENVIRONMENT Wouter Meert, manager CSSD, UZ Leuven.](https://reader035.fdocuments.in/reader035/viewer/2022062305/5697bfd81a28abf838caec0f/html5/thumbnails/7.jpg)
CSSD UZ Leuven
• ISO EN 13485 (since 2012, successfull recertification 2015)
• 54,90 FTE• 1900 m2• 3 batch and 3 tunnel WD• 5 floor load autoclaves• 2 Sterrad 100 NX
![Page 8: HYDROGEN PEROXIDE STERILISATION IN A ISO EN 13485 ENVIRONMENT Wouter Meert, manager CSSD, UZ Leuven.](https://reader035.fdocuments.in/reader035/viewer/2022062305/5697bfd81a28abf838caec0f/html5/thumbnails/8.jpg)
ISO EN 13485
• Organisational structure• Infrastructure• Structure QMS• Process controle• Customer feedback• Staf training
![Page 9: HYDROGEN PEROXIDE STERILISATION IN A ISO EN 13485 ENVIRONMENT Wouter Meert, manager CSSD, UZ Leuven.](https://reader035.fdocuments.in/reader035/viewer/2022062305/5697bfd81a28abf838caec0f/html5/thumbnails/9.jpg)
ISO EN 13485
![Page 10: HYDROGEN PEROXIDE STERILISATION IN A ISO EN 13485 ENVIRONMENT Wouter Meert, manager CSSD, UZ Leuven.](https://reader035.fdocuments.in/reader035/viewer/2022062305/5697bfd81a28abf838caec0f/html5/thumbnails/10.jpg)
Hydrogen peroxide sterilizers
![Page 11: HYDROGEN PEROXIDE STERILISATION IN A ISO EN 13485 ENVIRONMENT Wouter Meert, manager CSSD, UZ Leuven.](https://reader035.fdocuments.in/reader035/viewer/2022062305/5697bfd81a28abf838caec0f/html5/thumbnails/11.jpg)
Hydrogen peroxide UZ Leuven
• First Sterrad 100 S bought in 1998• Second in 2000• Half december 2013, 2 new
Sterrad 100 NX • Spring 2015, installation double
door• Current use 10 – 15
cycle/day/machine• 221 different items
![Page 12: HYDROGEN PEROXIDE STERILISATION IN A ISO EN 13485 ENVIRONMENT Wouter Meert, manager CSSD, UZ Leuven.](https://reader035.fdocuments.in/reader035/viewer/2022062305/5697bfd81a28abf838caec0f/html5/thumbnails/12.jpg)
Infrastructure
Before double door installation
![Page 13: HYDROGEN PEROXIDE STERILISATION IN A ISO EN 13485 ENVIRONMENT Wouter Meert, manager CSSD, UZ Leuven.](https://reader035.fdocuments.in/reader035/viewer/2022062305/5697bfd81a28abf838caec0f/html5/thumbnails/13.jpg)
Building plans
![Page 14: HYDROGEN PEROXIDE STERILISATION IN A ISO EN 13485 ENVIRONMENT Wouter Meert, manager CSSD, UZ Leuven.](https://reader035.fdocuments.in/reader035/viewer/2022062305/5697bfd81a28abf838caec0f/html5/thumbnails/14.jpg)
Current setting
![Page 15: HYDROGEN PEROXIDE STERILISATION IN A ISO EN 13485 ENVIRONMENT Wouter Meert, manager CSSD, UZ Leuven.](https://reader035.fdocuments.in/reader035/viewer/2022062305/5697bfd81a28abf838caec0f/html5/thumbnails/15.jpg)
Medical instruments• Optics• HD camera’s• Light cables• endoscopes• Da Vinci
instruments• Foetal surgery• Brachytherapi
e• Echoprobes
• Hipec • Sentinel probe• Diamant drills• Electrodes/
kirschner
80 %
![Page 16: HYDROGEN PEROXIDE STERILISATION IN A ISO EN 13485 ENVIRONMENT Wouter Meert, manager CSSD, UZ Leuven.](https://reader035.fdocuments.in/reader035/viewer/2022062305/5697bfd81a28abf838caec0f/html5/thumbnails/16.jpg)
Temperature probe Hipec
![Page 17: HYDROGEN PEROXIDE STERILISATION IN A ISO EN 13485 ENVIRONMENT Wouter Meert, manager CSSD, UZ Leuven.](https://reader035.fdocuments.in/reader035/viewer/2022062305/5697bfd81a28abf838caec0f/html5/thumbnails/17.jpg)
Echo probe
![Page 18: HYDROGEN PEROXIDE STERILISATION IN A ISO EN 13485 ENVIRONMENT Wouter Meert, manager CSSD, UZ Leuven.](https://reader035.fdocuments.in/reader035/viewer/2022062305/5697bfd81a28abf838caec0f/html5/thumbnails/18.jpg)
Utrecht applicator brachy
![Page 19: HYDROGEN PEROXIDE STERILISATION IN A ISO EN 13485 ENVIRONMENT Wouter Meert, manager CSSD, UZ Leuven.](https://reader035.fdocuments.in/reader035/viewer/2022062305/5697bfd81a28abf838caec0f/html5/thumbnails/19.jpg)
Da Vinci XI
![Page 20: HYDROGEN PEROXIDE STERILISATION IN A ISO EN 13485 ENVIRONMENT Wouter Meert, manager CSSD, UZ Leuven.](https://reader035.fdocuments.in/reader035/viewer/2022062305/5697bfd81a28abf838caec0f/html5/thumbnails/20.jpg)
Foetal surgery
![Page 21: HYDROGEN PEROXIDE STERILISATION IN A ISO EN 13485 ENVIRONMENT Wouter Meert, manager CSSD, UZ Leuven.](https://reader035.fdocuments.in/reader035/viewer/2022062305/5697bfd81a28abf838caec0f/html5/thumbnails/21.jpg)
Process switch
1 proces
• Standaard
• Duo• Flex• Express
![Page 22: HYDROGEN PEROXIDE STERILISATION IN A ISO EN 13485 ENVIRONMENT Wouter Meert, manager CSSD, UZ Leuven.](https://reader035.fdocuments.in/reader035/viewer/2022062305/5697bfd81a28abf838caec0f/html5/thumbnails/22.jpg)
Defining process compatibility
• Registration conform ISO EN 13485
• Listing instruments• Sterilisation Program selection:
– ASP Sterrad Sterility guide– User manual instrument
manufacturer– Sterility experts ASP
• Registration within traceability system and inventory
• Training employees
![Page 23: HYDROGEN PEROXIDE STERILISATION IN A ISO EN 13485 ENVIRONMENT Wouter Meert, manager CSSD, UZ Leuven.](https://reader035.fdocuments.in/reader035/viewer/2022062305/5697bfd81a28abf838caec0f/html5/thumbnails/23.jpg)
![Page 24: HYDROGEN PEROXIDE STERILISATION IN A ISO EN 13485 ENVIRONMENT Wouter Meert, manager CSSD, UZ Leuven.](https://reader035.fdocuments.in/reader035/viewer/2022062305/5697bfd81a28abf838caec0f/html5/thumbnails/24.jpg)
![Page 25: HYDROGEN PEROXIDE STERILISATION IN A ISO EN 13485 ENVIRONMENT Wouter Meert, manager CSSD, UZ Leuven.](https://reader035.fdocuments.in/reader035/viewer/2022062305/5697bfd81a28abf838caec0f/html5/thumbnails/25.jpg)
![Page 26: HYDROGEN PEROXIDE STERILISATION IN A ISO EN 13485 ENVIRONMENT Wouter Meert, manager CSSD, UZ Leuven.](https://reader035.fdocuments.in/reader035/viewer/2022062305/5697bfd81a28abf838caec0f/html5/thumbnails/26.jpg)
NBN EN ISO 14937
• IQ, OQ, PQ• Routine controle:
– Chemical indicators/ biological indicators
• Periodically maintenance• Parametric release
![Page 27: HYDROGEN PEROXIDE STERILISATION IN A ISO EN 13485 ENVIRONMENT Wouter Meert, manager CSSD, UZ Leuven.](https://reader035.fdocuments.in/reader035/viewer/2022062305/5697bfd81a28abf838caec0f/html5/thumbnails/27.jpg)
Parametric release
![Page 28: HYDROGEN PEROXIDE STERILISATION IN A ISO EN 13485 ENVIRONMENT Wouter Meert, manager CSSD, UZ Leuven.](https://reader035.fdocuments.in/reader035/viewer/2022062305/5697bfd81a28abf838caec0f/html5/thumbnails/28.jpg)
Conclusion• Need for hydrogen peroxide LTS
within our setting because of specific surgical instruments
• Important for ISO EN 13485 :– Sterrad Sterility Guide – Support by sterility advisors– Process control en parametric release
• More research is needed on effects of hydrogen peroxide on instruments