Humidity Control

2
How chambers lower humidity Humidity test chambers typically employ one of three dehumidification techniques to achieve lower humidity conditions. These methods may also be used in temperature-only applications, when risk of condensation high, usually due to massive test loads and/or very fast temperature cycling rates. Standard method: Refrigeration coil Nearly every humidity chamber should have refrigeration or some cooling method to ensure humidity control. Relative humidity varies with temperature, so a stable temperature is required. The refrigeration in most humidity chambers also have a separate dehumidification coil to attract moisture. The moisture collects on the coil, where it condenses and drains out of the chamber, thus dehumidifying the chamber. The coil is set to stay above freezing to avoid becoming covered with ice. De-humidification via refrigeration is standard on test chambers and can get relative humidity down to 10% at higher temperatures. For situations where even lower humidity is needed, there are three typical methods: 1. Nitrogen purge Nitrogen gas is does not hold moisture, so it is naturally dry. By slowly bleeding nitrogen gas into the chamber, moisture-laden air is pushed, or purged, out of the chamber. Carbon dioxide is not a dry gas and isn't suitable for this application. Nitrogen should not be used in chambers where people may go inside. 2. Dry-air purge: A dry-air purge system is more common than nitrogen. External compressed air is fed into a desiccant dryer system before being used to purge the chamber air. (A desiccant is a substance that easily absorbs moisture, like the silica gel crystal packets you find packaged with some products.) The dry-air purge system doesn't need much maintenance. The system uses two desiccant “towers”. One dries the compressed air while the other tower is vented to the room to allow it to regenerate. You will often hear these systems “pop” a little compressed air while switching between towers. The desiccant can be used for years before replacement desiccant is needed. 3. Recirculating desiccant air dryer Larger chambers are impractical to dry with nitrogen or compressed air. Instead, the chamber air is circulated to a dessicant for drying, before being returned to the chamber. The desiccant is in the form of a honeycomb wheel, where part of the wheel is being vented to the room to release moisture. These systems typically heat the air to make it easier to capture moisture, and will also have a cooler so the air will return at a controlled temperature that doesn't impact the chamber operation.

description

describe tecnicas para el control de humedad

Transcript of Humidity Control

Page 1: Humidity Control

How chambers lower humidity

Humidity test chambers typically employ one of three dehumidification techniques to achieve lower

humidity conditions. These methods may also be used in temperature-only applications, when risk of

condensation high, usually due to massive test loads and/or very fast temperature cycling rates.

Standard method: Refrigeration coil

Nearly every humidity chamber should have refrigeration or some cooling

method to ensure humidity control. Relative humidity varies with

temperature, so a stable temperature is required. The refrigeration in

most humidity chambers also have a separate dehumidification coil to

attract moisture. The moisture collects on the coil, where it condenses and drains out of the

chamber, thus dehumidifying the chamber. The coil is set to stay above freezing to avoid becoming

covered with ice.

De-humidification via refrigeration is standard on test chambers and can get relative humidity down

to 10% at higher temperatures. For situations where even lower humidity is needed, there are three

typical methods:

1. Nitrogen purge

Nitrogen gas is does not hold moisture, so it is naturally dry. By slowly

bleeding nitrogen gas into the chamber, moisture-laden air is pushed, or

purged, out of the chamber. Carbon dioxide is not a dry gas and isn't

suitable for this application. Nitrogen should not be used in chambers

where people may go inside.

2. Dry-air purge:

A dry-air purge system is more common than nitrogen. External

compressed air is fed into a desiccant dryer system before being used to

purge the chamber air.

(A desiccant is a substance that easily absorbs moisture, like the silica gel

crystal packets you find packaged with some products.)

The dry-air purge system doesn't need much maintenance. The system

uses two desiccant “towers”. One dries the compressed air while the other tower is vented to the

room to allow it to regenerate. You will often hear these systems “pop” a little compressed air while

switching between towers. The desiccant can be used for years before replacement desiccant is

needed.

3. Recirculating desiccant air dryer

Larger chambers are impractical to dry with nitrogen or compressed air. Instead, the chamber air is

circulated to a dessicant for drying, before being returned to the chamber. The desiccant is in the

form of a honeycomb wheel, where part of the wheel is being vented to the room to release

moisture. These systems typically heat the air to make it easier to capture moisture, and will also

have a cooler so the air will return at a controlled temperature that doesn't impact the chamber

operation.

Page 2: Humidity Control

Latest News

See popular models currently in stock

Monday, March 17 2014

ESPEC has benchtop, reach-ins, and thermal shock chambers in stock. See our Quick Ship page

for a current list.

Continue reading...

Updated SU/SH benchtops

Tuesday, November 19 2013

Our popular benchtop chambers have been updated.

Continue reading...

ESPEC NORTH AMERICA, INC.4141 Central ParkwayHudsonville, MI 49426Phone: 616-896-6100FAX: 616-896-6150

©2014 ESPEC NORTH AMERICA

A Division of ESPEC Corp, Japan

Terms of Use

Privacy Policy

Contact Us

Call us at 877-GO-ESPEC (877-463-7732 for US/Canada, EST)

or click here to find your local sales rep

Chamber Resources | How chambers lower humidity | ESPEC North Am... http://www.espec.com/na/chamber_faq/answer/dehumidification

2 of 2 02/07/2014 12:39 p.m.