Humidity Calibator
-
Upload
darshan-patel -
Category
Documents
-
view
228 -
download
0
Transcript of Humidity Calibator
-
8/3/2019 Humidity Calibator
1/23
Humidity CalibratorHMK15
USER'SGUIDEM210185en-B
September 2002
-
8/3/2019 Humidity Calibator
2/23
M210185en-A User's Guide
ii
-
8/3/2019 Humidity Calibator
3/23
M210185en-A User's Guide
iii
PUBLISHED BY
Vaisala OyjPhone (int.): (+358 9) 894 91
P.O. Box 26 Fax: (+358 9) 8949 2227FIN-00421 Helsinki
Finland
Visit our Internet pages at http://www.vaisala.com/
Vaisala 2002
No part of this manual may be reproduced in any form or by any
means, electronic or mechanical (including photocopying), nor may its
contents be communicated to a third party without prior written
permission of the copyright holder.
The contents are subject to change without prior notice.
http://www.vaisala.com/http://www.vaisala.com/ -
8/3/2019 Humidity Calibator
4/23
M210185en-A User's Guide
iv
Contents
1. PRODUCT DESCRIPTION................................................................................................... 1
1.1. General .................................................................................................................... 1
1.2. Calibration certificate of the salt packages........................................................... 2
2. PREPARING THE SALT SOLUTIONS ................................................................................. 2
2.1. General instructions ............................................................................................... 2
2.2. Preparing the solutions .......................................................................................... 3
3. INSTRUCTIONS FOR THE CALIBRATION OF HUMIDITY INSTRUMENTS ....................... 7
3.1. General instructions ............................................................................................... 7
3.1.1. Thermometer............................................................................................. 8
3.2. Calibration............................................................................................................... 9
3.3. Greenspan's calibration table1............................................................................ 11
3.4. On-site calibrations and transportation............................................................... 11
4. MAINTENANCE ................................................................................................................. 15
5. SPARE PARTS AND ACCESSORIES ................................................................................ 15
6. TECHNICAL DATA ............................................................................................................ 16
WARRANTY
-
8/3/2019 Humidity Calibator
5/23
M210185en-A User's Guide
v
This page intentionally left blank.
-
8/3/2019 Humidity Calibator
6/23
-
8/3/2019 Humidity Calibator
7/23
1. PRODUCT DESCRIPTION
1.1. General
The HMK15 humidity calibrator has been developed for calibration and checking
of humidity probes and transmitters. The functioning of the calibrator is based on
the fact that certain salt solutions generate a certain relative humidity in the air
above them.
The four holes in the chamber cover are designed for Vaisala probes and
transmitters with 12 mm, 13.5 mm (2 holes) and 18.5 mm diameters. The saltsolutions suitable for the HMK15 calibrator are e.g. lithium chloride LiCl
(11 %RH), magnesium chloride MgCl2 (33 %RH), sodium chloride NaCl(75 %RH) and potassium sulphate K
2SO
4(97 %RH). In calibration, the sensor
head is inserted into a salt chamber containing a saturated salt solution. The
probe/transmitter reading is then adjusted to the correct value, i.e. to the
humidity that the specific salt solution generates at that particular temperature.
Calibration is usually performed at least at two different humidities to ensure
the sensor accuracy over the entire humidity range (0 - 100 %RH).
The HMK15 calibrator is suitable for both laboratory and field usage. During
transportation, the chambers can be tightly closed with specially designed
transit covers. The optional transit bag (order code HM27032) helps to
transport the calibrator in the vertical position. Furthermore, the bag can be
used for housing the calibrator also during calibration.
The following photo shows the HMK15 calibrator with some accessories
(marked with *):
Figure 1.1 The HMK15 calibrator with some accessories (*)
Salt chamber
Thermometer
Adapter fitting
Plugs
*Ready dosed
salt packagewith calibrationcertificate
Calibration
certificate forthermometer
*Ion exchanged water
Measurement cup
Mixing spoon
Base plate
-
8/3/2019 Humidity Calibator
8/23
User's Guide M210185en-B
2
Accessories include additional salt chambers, ion exchanged water, transit bag and
ready-dosed salt packages (LiCl 11%RH, MgCl2 33 %RH, NaCl 75 %RH,
K2SO4 97 %RH).
1.2. Calibration certificate of the salt packages
Vaisala's ready-dosed salt packages are delivered with calibration certificates
given on the basis of a batch check. In other words, a sample of packages are
taken from a batch and salts are prepared according to the instructions given in
this manual. These salt solutions are then checked at Vaisala Oy's
Measurement Standards Laboratory (this calibration laboratory is accredited
by FINAS, a member of the European Cooperation for Accreditation of
Laboratories). The calibration certificate certifies that the equilibrium
humidities generated by these salt solutions correspond to Greenspan's
calibration table within the specified accuracy (see Chapter6).
2. PREPARING THE SALT SOLUTIONS
2.1. General instructions
Prepare the salts using the equipment provided with the calibrator or
make sure that all accessories used are suitable for preparing salt
solutions and are absolutely clean. If necessary, wash them carefully
and rinse several times before preparing the salt solutions. The last
rinse must be made with distilled or ion exchanged water.
The salts are quick and easy to prepare with Vaisala's ready-dosed salt
packages. If you do not use ready-dosed salt packages, measure the
salts using the measurement cup provided. Make sure that the cup is
clean before measuring the salts. In the following instructions, the
required amounts are given in grammes and millilitres. The salt
solutions must be prepared from salts of pro analysis grade; these are
available in agencies selling industrial chemicals. Note that the water
must be distilled or ion exchanged (conductivity < 0.25 S/cm). Ion
exchanged water can also be ordered as an accessory for the HMK15
calibrator (order code 19767HM).
Handle the salts and equipment carefully and keep them absolutely
clean so that the salts are not contaminated. Make sure that the salts
do not mix.
At the bottom of the salt chamber there must not be more than 1 cm
of undissolved salt and liquid. Otherwise, the sensor to be calibrated
can be immersed into the solution. If necessary, verify the amount e.g.
with the measurement spoon.
Various standards (ASTM E104-85, DIN 50008, JIS Z8806) include
recommendations and instructions on how to prepare and store salt
solutions.
-
8/3/2019 Humidity Calibator
9/23
User's Guide M210185en-B
3
2.2. Preparing the solutions
Lithium chloride generates a reference humidity of approximately 11 %RH.
It is normally used as the dry end (offset) reference.
NOTE
Never add water to dry LiCl salt; the salt may heat up
so rapidly that it splashes out of the chamber.
NOTE
LiCl is harmful when swallowed; the solution is also
corrosive.
NOTE
If the LiCl solution is used or stored in temperatures
below +18C, its equilibrium humidity changes
permanently.
Magnesium chloride generates a reference humidity of approximately33 %RH. It is often used as a check point if calibration is performed at more
than two points.
Sodium chloride generates a reference humidity of approximately 75 %RH. It
is used as the wet end reference (gain) for probes measuring in applications
with normal humidities.
Potassium sulphate generates a reference humidity of approximately
97 %RH. Potassium sulphate is used as the wet end reference (gain) for probes
measuring in applications with very high humidities (90 - 100 %RH, e.g.
outdoor and concrete measurements).
Prepare the salt solutions according to the following instructions. Measure the
salts carefully if you do not use the ready-dosed packages.
-
8/3/2019 Humidity Calibator
10/23
User's Guide M210185en-B
4
1. Take the calibrator out of the box. Open the transit cover of the
chamber. Remove the measurement cover from the chamber holder
and press the transit cover on the holder (see Figure 2.1).
Figure 2.1 Press the transit cover on a chamber holder
2. Pour ion exchanged water into the chamber; the required amounts are
given in the table:
LiCl 12 ml of water
MgCl2 3 ml of water
NaCl 10 ml of water
K2SO4 10 ml of water
Figure 2.2 Pour a correct dose of ion exchanged water into the
chamber; dry the measurement cup
2. Sprinkle the contents of a salt package (or measure the salt according
to the table) in small quantities into the chamber, stirring constantly.
When measuring with the measurement cup, make sure that the cup is
clean and dry. Rinse and dry the cup after every use.
-
8/3/2019 Humidity Calibator
11/23
User's Guide M210185en-B
5
LiCl 15 g or 18 ml
MgCl2 30 g or 30 ml
NaCl 20 g or 15 mlK2SO4 30 g or 20 ml
Figure 2.3 Sprinkle a correct dose of salt into the chamber
stirring constantly
4. When all salt has been sprinkled into the chamber, the saturated salt
solution should have the ratio of 60 - 90 % undissolved salt to10 - 40 % liquid.
5. Close the chamber with the chamber cover (Figure 2.4).
Figure 2.4 Close the chamber with the chamber cover
-
8/3/2019 Humidity Calibator
12/23
User's Guide M210185en-B
6
6. Fasten the salt chamber on the holder in the base plate and close the
measurement holes with rubber plugs. The chambers can also be used
as individual checkers without the base plate. NOTE: the plugs have
three steps, each of which is suitable for a certain hole diameter: the
first step for the 12 mm, the second step for the 13.5 mm and the thirdstep for the 18.5 mm hole. Keep the holes always closed when not
calibrating.
7. Write the preparation date on a sticker and mark the chamber with it.
If you use a ready-dosed salt package, use the sticker with the batch
code. Mark all parts of the salt chamber (chamber, chamber cover and
transit cover) with stickers. This way, various covers and different
salts do not get mixed.
Figure 2.5 Mark the chamber and covers with stickers
8. Allow approximately 24 hours for stabilization before use so that the
salt solution reaches the equilibrium humidity.
9. See Chapter4 on how to maintain the salt solutions.
NOTE
If you do not use the instrument for a longer period of
time, close the chambers with transit covers.
-
8/3/2019 Humidity Calibator
13/23
User's Guide M210185en-B
7
3. INSTRUCTIONS FOR THE CALIBRATION OF HUMIDITY
INSTRUMENTS
3.1. General instructions
Usually, the errors during humidity calibration are due to temperature
differences. A temperature difference of 1C at +20C between the air in the
chamber and the sensor causes an error of 3 %RH at 50 %RH and an error of
6 %RH at 97 %RH. The more the transportation or process temperatures
differ from the temperature at the calibration site, the longer the stabilization
time required. In laboratory use, the calibrator should be stored in that part of
the room where the temperature is most stable. The calibrator must be kept out
of direct sunlight and away from localized heat sources, such as spot lights,
heaters and soldering irons. If the probe/transmitter is checked against several
humidity references, the checking must first be made at the dry end. Moredetailed instructions are given in individual manuals of transmitters and
probes.
Handle the probe as little as possible. Do not hold the salt chamber or other
parts of the calibrator in hand during calibration as they warm up and cause
errors in the readings.
NOTE: even the smallest water drop on the probe near the sensors distorts the
readings. Make sure that chamber covers and plugs are carefully closed.
0 0
1 0
2 0
3 0
4 0
5 0
C
HMP233probe
O-rings
sensors
max.1 cm
thermometer
saturated salsolution
undissolvedsalt
Figure 3.1 Salt chamber
-
8/3/2019 Humidity Calibator
14/23
User's Guide M210185en-B
8
3.1.1. Thermometer
The calibrator includes a thermometer which can contain either mercury or red
capillary liquid. If the thermometer contains mercury, it is delivered with a
calibration certificate. The accuracy of the mercury thermometer is sufficientfor checking the transmitter's temperature channel if desired. The mercury
thermometer has been calibrated at five points and the result is traceable to the
Measurement Standards Laboratory of the Centre for Metrology and
Accreditation in Finland. For more accurate results, use the corrections given
in the calibration certificate or their interpolated values. Consider also the
parallax error which may occur while reading the results. The mercury
thermometer should be recalibrated within three years in order to maintain the
traceability.
The manufacturer of the thermometer with red capillary liquid provides it with
a factory calibration certificate. As the accuracy and stability of this kind of athermometer are inferior to those of a mercury thermometer, it is not
recommended to be used for checking the temperature channel of the
transmitter.
The sleeve at the end of the thermometer's protective tube can be used in two
different positions. The thermometer is delivered with the sleeve in such a
position that it protects the part containing the capillary liquid. When turned
the other way round, it acts as an adapter when the thermometer is inserted
into the 13.5 mm hole (see Figure 3.2).
Figure 3.2 The sleeve protects the part with capillary liquid (on the
left) or acts as an adapter (on the right)
During calibration, the thermometer is inserted into the 13.5 mm hole of a salt
chamber. Press it downwards until it passes the O-rings. The thermometer is
correctly in place when you can feel a resistance while pressing it downwards.
-
8/3/2019 Humidity Calibator
15/23
User's Guide M210185en-B
9
Figure 3.3 During calibration, the thermometer is inserted into the
13.5 mm hole
When the thermometer is not in use or the calibrator is transferred from one
place to another, place the thermometer in holders (Figure 3.4).
Figure 3.4 Place the thermometer in holders when not in use
3.2. Calibration
Perform the calibration according to the following instructions:
Leave the HMK15 calibrator and the probe at the calibration site for
at least 30 minutes before starting the calibration in order to let the
probe temperature stabilize to the room temperature.
With lithium chloride it is not necessary to use the thermometer as the
humidity reading changes only very slightly in the temperature range
of +25...+30 C. However, if you use it make sure that the sleeve is in
the correct position and insert it into the 13.5 mm hole of the LiCl salt
chamber. NOTE: handle the thermometer as little as possible and do
not handle it by the measuring end. Press the thermometer downwards
until it passes the O-rings. The thermometer is correctly in place when
you feel a resistance while pressing it downwards.
-
8/3/2019 Humidity Calibator
16/23
User's Guide M210185en-B
10
Take off the grid or filter protecting the sensor. Take care not to
damage the sensor.
Some 12 mm probes need a adapter fitting to fit into the 13,5 mm
hole. Adapter fitting prevents sensor from soaking in the salt liquid.In case you need an adapter fitting, take off the grid or filter and
replace it with the adapter fitting. Two adapter fittings are attached on
the base plate of the HMK15, see picture on page 1.
CAUTION Remember to check from the user's guide of your device if theadapter fitting is needed. If sensor accidently soaks in salt liquid,
remove it quickly and rinse with clean water. Let it dry before taking
into use again.
Insert the probe into a suitable hole of the LiCl salt chamber. Press it
downwards until it passes the O-rings. The shorter the time the hole
stays open before inserting the probe, the shorter the stabilization time
required.
Wait until the humidity reading stabilizes; this takes about 10 - 30
minutes.
Read the salt chamber temperature from the thermometer and then
read the closest humidity value from the calibration table
(Greenspan's calibration table, LiCl solution).
Adjust the dry end (DRY, offset) to correspond to the value given in
the calibration table.
Insert the thermometer into the 13.5 mm hole of the NaCl salt
chamber. NOTE: when calibrating probes that are being used for long
term measurements (over 1 hour) in high humidities (90 - 100 %RH),
use the K2SO4 salt as the high end reference.
Insert the probe into a suitable hole of the NaCl (or K2SO4) salt
chamber. The shorter the time the hole stays open before inserting the
probe, the shorter the stabilization time required.
Wait until the humidity reading stabilizes; this takes about 10 - 30
minutes. Note that in high humidities the risk for errors increases.
Therefore, the stabilization time should be longer (approximately
20 - 40 minutes).
Read the salt chamber temperature from the thermometer and then
read the closest humidity value from the calibration table
(Greenspan's calibration table, NaCl or K2SO4 solution).
Adjust the wet end (WET, gain) to correspond to the value given in
the calibration table.
-
8/3/2019 Humidity Calibator
17/23
User's Guide M210185en-B
11
3.3. Greenspan's calibration table1
C LiCl MgCl2 NaCl K 2SO4
0 * 33.7 0.3 75.5 0.3 98.8 1.1
5 * 33.6 0.3 75.7 0.3 98.5 0.910 * 33.5 0.2 75.7 0.2 98.2 0.8
15 * 33.3 0.2 75.6 0.2 97.9 0.6
20 11.3 0.3 33.1 0.2 75.5 0.1 97.6 0.5
25 11.3 0.3 32.8 0.2 75.3 0.1 97.3 0.5
30 11.3 0.2 32.4 0.1 75.1 0.1 97.0 0.4
35 11.3 0.2 32.1 0.1 74.9 0.1 96.7 0.4
40 11.2 0.2 31.6 0.1 74.7 0.1 96.4 0.4
45 11.2 0.2 31.1 0.1 74.5 0.2 96.1 0.4
50 11.1 0.2 30.5 0.1 74.4 0.2 95.8 0.5
* If the LiCl solution is used or stored in temperatures below +18C, its equilibriumhumidity changes permanently
The numbers on the right in each column indicate the uncertainty of the
reference humidity of the salt in that temperature.
1Greenspan, L.: Journal of Research of the National Bureau of Standards - A
Physis and Chemistry Vol. 81A, No. 1 January-February 1977, pp. 89-95
3.4. On-site calibrations and transportation
The HMK15 calibrator is easily trasferred from one place to another. Two O-
rings seal the transit cover on the chamber. The optional transit bag helps to
transport the calibrator so that chambers stay in the vertical position.
Furthermore, the bag can be used for housing the calibrator during calibration.
When transferring the calibrator from one place to
another, follow the instructions below:
Turn the protective sleeve on the
thermometer and place the
thermometer in the holder.
Replace chamber covers withtransit covers. Press chamber covers
on vacant chamber holders for
transportation.
During transportation, keep the
calibrator so that the chambers stay as
upright as possible. This way, very
little salt solution gets on transit
covers and cleaning is easier. The
closer the transportation temperature is to the temperature of the
calibration site, the shorter the time required for temperaturestabilization. If the transportation temperature is below +18 C, the
-
8/3/2019 Humidity Calibator
18/23
User's Guide M210185en-B
12
LiCl salt chamber should be transported separately to keep the
solution warm.
When the calibrator is at the calibration site, remove transit covers
and fasten chamber covers on salt chambers. Close the holes with
plugs. NOTE: the plugs have three steps each of which is suitable for
a certain hole diameter: the first step for the 12 mm, the second step
for the 13.5 mm and the third step for the 18.5 mm hole.
Clean the transit covers with a damp cloth and press them on vacant
chamber holders.
Take the thermometer off the holder, turn the sleeve the other way
round and perform the calibration according to Chapter3.2.
In on-site calibrations it is important to allow enough time for the calibratorand probe temperatures to stabilize. Depending on the temperature differences
between the transportation and the calibration site or between the probe
removed from the process and the calibration site, a two point calibration
takes about 0.5 - 2 hours. If there are several frequently calibrated instruments
at the same site, knowing the stabilization times is useful.
The following figure illustrates an example of temperature and humidity
stabilization. In this example, an HMP233 probe has been transferred from a
temperature of 75C (oven) to a suitable hole of the NaCl salt chamber; the
calibrator is at room temperature. The figure shows that after 40 minutes, the
humidity reading differs 0.2 %RH from the final reading.
Time [min]
Hum
idity[%RH
70
71
72
73
74
75
76
77
78
0 5 10 15 20 25 30 35 40 45 50 55 60 65 70
10
20
30
40
50
60
70
80
90
Temp
erature[C]
HUMIDITY
TEMPERATURE
Figure 3.5 An example of RH and T stabilization when the probe is
removed from a process
-
8/3/2019 Humidity Calibator
19/23
User's Guide M210185en-B
13
Time [min.]
Humidity[%RH
60
62
64
66
68
70
72
74
0 10 20 30 40 50 60 70 80 90 100 110 120
0
4
8
12
16
20
24
28
Temperature[C]
HUMIDITY
TEMPERATURE
Figure 3.6 An example of the RH and T stabilization when the
transportation and calibration site temperatures are
different
-
8/3/2019 Humidity Calibator
20/23
User's Guide M210185en-B
14
In the following example, the transportation temperature of the calibrator
(without the LiCl) is +5C; the calibrator is then transferred into room
temperature. After this, an HMP233 probe that has been stored at room
temperature, is inserted into a suitable hole of the NaCl salt chamber. The
figure shows that after 40 minutes, the reading differs 1.4 %RH from the finalreading.
Time [min.]
Humidity[%RH]
60
62
64
66
68
70
72
74
0 10 20 30 40 50 60 70 80 90 100 110 120
0
4
8
12
16
20
24
28
Tempe
rature[C]
HUMIDITY
TEMPERATURE
Figure 3.7 An example of the RH and T stabilization when the
transportation and calibration site temperatures are
different
-
8/3/2019 Humidity Calibator
21/23
User's Guide M210185en-B
15
4. MAINTENANCE
Depending on the frequency of use and the general operating conditions, the
salt solutions maintain their characteristics for 6 - 12 months. After that, theymust be replaced. A visual check should be performed at intervals of 2 - 3
months. There must be a minimum of approximately 10% of undissolved salt
at the bottom of the chamber (max. 90%) and the salt must be clean, otherwise
it must be reprepared. Please, note that the LiCl solution may crystallize on
the surface. However, there may still be solution under the surface. Therefore,
in such cases stir the solution and check it again the next day. It is advisable to
keep a maintenance log which shows the dates of salt replacements and other
maintenance procedures.
For correct calibration it is essential that salt chambers are tightly closed.
Check the O-rings at each salt replacement. If they are damaged, replace themwith new ones. The O-rings used are of the following types:
O-ring location dimensions type
measurement hole 12 mm 12 x 2.5 mm NBR70
measurement hole 13.5 mm 13.5 x 2.5 mm NBR70
measurement hole 18.5 mm 18.3 x 2.4 mm NBR70
transit cover 41.2 x 3.0 mm NBR70
chamber 50.0 x 2.0 mm NBR70
5. SPARE PARTS AND ACCESSORIES
Order code Description
19728HM Mercury thermometer with calibration certificate
25130HM Thermometer with red capillary liquid
19729HM Ready-dosed LiCl salt package with calibration certificate
19730HM Ready-dosed MgCl2 salt package with calibration certificate
19731HM Ready-dosed NaCl salt package with calibration certificate
19732HM Ready-dosed K 2SO4 salt package with calibration certificate
19746HM Plug set
19766HM Salt chamber
19767HM Ion exchanged water
211011 Adapter fitting for 12 mm probes
HM27032 Transit bag
-
8/3/2019 Humidity Calibator
22/23
User's Guide M210185en-B
16
6. TECHNICAL DATA
Operating temperature range +0 ...+50 C
NOTE
If the LiCl solution is used or stored in
temperatures below +18C, its equilibrium humidity
changes permanently.
Accuracy:
lithium chloride LiCl 1.0 %RH + Greenspan's uncertainty*
magnesium chloride MgCl2 1.0 %RH + Greenspan's uncertainty*
natrium chloride NaCl 1.4 %RH + Greenspan's uncertainty*
potassium sulphate K2SO4 1.5 %RH + Greenspan's uncertainty*
* = The uncertainty given in Greenspan's calibration table (see Chapter 3.3) at the calibration
temperature, e.g. the accuracy of LiCl salt at +20 C is (1.0 + 0.3) %RH = 1.3 %RH
Response time (probe and calibrator
at the same temperature) with Vaisala sensors typically 10 min.
(deviation of the final value < 1%RH)
Accuracy of the thermometer:
with mercury 0.3 C
with red capillary liquid 1 C
Dimensions 20 x 9 x 23 cm
Materials metal parts: anodised aluminium
Weight 1 kg without salt solutions
-
8/3/2019 Humidity Calibator
23/23
User's Guide M210185en-B
Warranty
Vaisala hereby represents and warrants all Products manufactured by Vaisala and sold hereunder to be free from
defects in workmanship or material during a period of twelve (12) months from the date of delivery save for
products for which a special warranty is given. If any Product proves however to be defective in workmanship ormaterial within the period herein provided Vaisala undertakes to the exclusion of any other remedy to repair or at
its own option replace the defective Product or part thereof free of charge and otherwise on the same conditions
as for the original Product or part without extension to original warranty time. Defective parts replaced in
accordance with this clause shall be placed at the disposal of Vaisala.
Vaisala also warrants the quality of all repair and service works performed by its employees to products sold by
it. In case the repair or service works should appear inadequate or faulty and should this cause malfunction or
nonfunction of the product to which the service was performed Vaisala shall at its free option either repair or
have repaired or replace the product in question. The working hours used by employees of Vaisala for such repair
or replacement shall be free of charge to the client. This service warranty shall be valid for a period of six (6)
months from the date the service measures were completed.
This warranty is however subject to following conditions:
a) A substantiated written claim as to any alleged defects shall have been received by Vaisala within thirty (30)
days after the defect or fault became known or occurred, and
b) The allegedly defective Product or part shall, should Vaisala so require, be sent to the works of Vaisala or to
such other place as Vaisala may indicate in writing, freight and insurance prepaid and properly packed and
labelled, unless Vaisala agrees to inspect and repair the Product or replace it on site.
This warranty does not however apply when the defect has been caused through
a) normal wear and tear or accident;
b) misuse or other unsuitable or unauthorized use of the Product or negligence or error in storing, maintaining or
in handling the Product or any equipment thereof;
c) wrong installation or assembly or failure to service the Product or otherwise follow Vaisala's service
instructions including any repairs or installation or assembly or service made by unauthorized personnel not
approved by Vaisala or replacements with parts not manufactured or supplied by Vaisala;
d) modifications or changes of the Product as well as any adding to it without Vaisala's prior authorization;
e) other factors depending on the Customer or a third party.
Notwithstanding the aforesaid Vaisala's liability under this clause shall not apply to any defects arising out of
materials, designs or instructions provided by the Customer.
This warranty is expressly in lieu of and excludes all other conditions, warranties and liabilities, express or
implied, whether under law, statute or otherwise, including without limitation any implied warranties of
merchantability or of fitness for a particular purpose and all other obligations and liabilities of Vaisala or its
representatives with respect to any defect or deficiency applicable to or resulting directly or indirectly from the
Products supplied hereunder, which obligations and liabilities are hereby expressly cancelled and waived.
Vaisala's liability shall under no circumstances exceed the invoice price of any Product for which a warranty
claim is made, nor shall Vaisala in any circumstances be liable for lost profits or other consequential loss whether
direct or indirect or for special damages.