Humidity Calibator

download Humidity Calibator

of 23

Transcript of Humidity Calibator

  • 8/3/2019 Humidity Calibator

    1/23

    Humidity CalibratorHMK15

    USER'SGUIDEM210185en-B

    September 2002

  • 8/3/2019 Humidity Calibator

    2/23

    M210185en-A User's Guide

    ii

  • 8/3/2019 Humidity Calibator

    3/23

    M210185en-A User's Guide

    iii

    PUBLISHED BY

    Vaisala OyjPhone (int.): (+358 9) 894 91

    P.O. Box 26 Fax: (+358 9) 8949 2227FIN-00421 Helsinki

    Finland

    Visit our Internet pages at http://www.vaisala.com/

    Vaisala 2002

    No part of this manual may be reproduced in any form or by any

    means, electronic or mechanical (including photocopying), nor may its

    contents be communicated to a third party without prior written

    permission of the copyright holder.

    The contents are subject to change without prior notice.

    http://www.vaisala.com/http://www.vaisala.com/
  • 8/3/2019 Humidity Calibator

    4/23

    M210185en-A User's Guide

    iv

    Contents

    1. PRODUCT DESCRIPTION................................................................................................... 1

    1.1. General .................................................................................................................... 1

    1.2. Calibration certificate of the salt packages........................................................... 2

    2. PREPARING THE SALT SOLUTIONS ................................................................................. 2

    2.1. General instructions ............................................................................................... 2

    2.2. Preparing the solutions .......................................................................................... 3

    3. INSTRUCTIONS FOR THE CALIBRATION OF HUMIDITY INSTRUMENTS ....................... 7

    3.1. General instructions ............................................................................................... 7

    3.1.1. Thermometer............................................................................................. 8

    3.2. Calibration............................................................................................................... 9

    3.3. Greenspan's calibration table1............................................................................ 11

    3.4. On-site calibrations and transportation............................................................... 11

    4. MAINTENANCE ................................................................................................................. 15

    5. SPARE PARTS AND ACCESSORIES ................................................................................ 15

    6. TECHNICAL DATA ............................................................................................................ 16

    WARRANTY

  • 8/3/2019 Humidity Calibator

    5/23

    M210185en-A User's Guide

    v

    This page intentionally left blank.

  • 8/3/2019 Humidity Calibator

    6/23

  • 8/3/2019 Humidity Calibator

    7/23

    1. PRODUCT DESCRIPTION

    1.1. General

    The HMK15 humidity calibrator has been developed for calibration and checking

    of humidity probes and transmitters. The functioning of the calibrator is based on

    the fact that certain salt solutions generate a certain relative humidity in the air

    above them.

    The four holes in the chamber cover are designed for Vaisala probes and

    transmitters with 12 mm, 13.5 mm (2 holes) and 18.5 mm diameters. The saltsolutions suitable for the HMK15 calibrator are e.g. lithium chloride LiCl

    (11 %RH), magnesium chloride MgCl2 (33 %RH), sodium chloride NaCl(75 %RH) and potassium sulphate K

    2SO

    4(97 %RH). In calibration, the sensor

    head is inserted into a salt chamber containing a saturated salt solution. The

    probe/transmitter reading is then adjusted to the correct value, i.e. to the

    humidity that the specific salt solution generates at that particular temperature.

    Calibration is usually performed at least at two different humidities to ensure

    the sensor accuracy over the entire humidity range (0 - 100 %RH).

    The HMK15 calibrator is suitable for both laboratory and field usage. During

    transportation, the chambers can be tightly closed with specially designed

    transit covers. The optional transit bag (order code HM27032) helps to

    transport the calibrator in the vertical position. Furthermore, the bag can be

    used for housing the calibrator also during calibration.

    The following photo shows the HMK15 calibrator with some accessories

    (marked with *):

    Figure 1.1 The HMK15 calibrator with some accessories (*)

    Salt chamber

    Thermometer

    Adapter fitting

    Plugs

    *Ready dosed

    salt packagewith calibrationcertificate

    Calibration

    certificate forthermometer

    *Ion exchanged water

    Measurement cup

    Mixing spoon

    Base plate

  • 8/3/2019 Humidity Calibator

    8/23

    User's Guide M210185en-B

    2

    Accessories include additional salt chambers, ion exchanged water, transit bag and

    ready-dosed salt packages (LiCl 11%RH, MgCl2 33 %RH, NaCl 75 %RH,

    K2SO4 97 %RH).

    1.2. Calibration certificate of the salt packages

    Vaisala's ready-dosed salt packages are delivered with calibration certificates

    given on the basis of a batch check. In other words, a sample of packages are

    taken from a batch and salts are prepared according to the instructions given in

    this manual. These salt solutions are then checked at Vaisala Oy's

    Measurement Standards Laboratory (this calibration laboratory is accredited

    by FINAS, a member of the European Cooperation for Accreditation of

    Laboratories). The calibration certificate certifies that the equilibrium

    humidities generated by these salt solutions correspond to Greenspan's

    calibration table within the specified accuracy (see Chapter6).

    2. PREPARING THE SALT SOLUTIONS

    2.1. General instructions

    Prepare the salts using the equipment provided with the calibrator or

    make sure that all accessories used are suitable for preparing salt

    solutions and are absolutely clean. If necessary, wash them carefully

    and rinse several times before preparing the salt solutions. The last

    rinse must be made with distilled or ion exchanged water.

    The salts are quick and easy to prepare with Vaisala's ready-dosed salt

    packages. If you do not use ready-dosed salt packages, measure the

    salts using the measurement cup provided. Make sure that the cup is

    clean before measuring the salts. In the following instructions, the

    required amounts are given in grammes and millilitres. The salt

    solutions must be prepared from salts of pro analysis grade; these are

    available in agencies selling industrial chemicals. Note that the water

    must be distilled or ion exchanged (conductivity < 0.25 S/cm). Ion

    exchanged water can also be ordered as an accessory for the HMK15

    calibrator (order code 19767HM).

    Handle the salts and equipment carefully and keep them absolutely

    clean so that the salts are not contaminated. Make sure that the salts

    do not mix.

    At the bottom of the salt chamber there must not be more than 1 cm

    of undissolved salt and liquid. Otherwise, the sensor to be calibrated

    can be immersed into the solution. If necessary, verify the amount e.g.

    with the measurement spoon.

    Various standards (ASTM E104-85, DIN 50008, JIS Z8806) include

    recommendations and instructions on how to prepare and store salt

    solutions.

  • 8/3/2019 Humidity Calibator

    9/23

    User's Guide M210185en-B

    3

    2.2. Preparing the solutions

    Lithium chloride generates a reference humidity of approximately 11 %RH.

    It is normally used as the dry end (offset) reference.

    NOTE

    Never add water to dry LiCl salt; the salt may heat up

    so rapidly that it splashes out of the chamber.

    NOTE

    LiCl is harmful when swallowed; the solution is also

    corrosive.

    NOTE

    If the LiCl solution is used or stored in temperatures

    below +18C, its equilibrium humidity changes

    permanently.

    Magnesium chloride generates a reference humidity of approximately33 %RH. It is often used as a check point if calibration is performed at more

    than two points.

    Sodium chloride generates a reference humidity of approximately 75 %RH. It

    is used as the wet end reference (gain) for probes measuring in applications

    with normal humidities.

    Potassium sulphate generates a reference humidity of approximately

    97 %RH. Potassium sulphate is used as the wet end reference (gain) for probes

    measuring in applications with very high humidities (90 - 100 %RH, e.g.

    outdoor and concrete measurements).

    Prepare the salt solutions according to the following instructions. Measure the

    salts carefully if you do not use the ready-dosed packages.

  • 8/3/2019 Humidity Calibator

    10/23

    User's Guide M210185en-B

    4

    1. Take the calibrator out of the box. Open the transit cover of the

    chamber. Remove the measurement cover from the chamber holder

    and press the transit cover on the holder (see Figure 2.1).

    Figure 2.1 Press the transit cover on a chamber holder

    2. Pour ion exchanged water into the chamber; the required amounts are

    given in the table:

    LiCl 12 ml of water

    MgCl2 3 ml of water

    NaCl 10 ml of water

    K2SO4 10 ml of water

    Figure 2.2 Pour a correct dose of ion exchanged water into the

    chamber; dry the measurement cup

    2. Sprinkle the contents of a salt package (or measure the salt according

    to the table) in small quantities into the chamber, stirring constantly.

    When measuring with the measurement cup, make sure that the cup is

    clean and dry. Rinse and dry the cup after every use.

  • 8/3/2019 Humidity Calibator

    11/23

    User's Guide M210185en-B

    5

    LiCl 15 g or 18 ml

    MgCl2 30 g or 30 ml

    NaCl 20 g or 15 mlK2SO4 30 g or 20 ml

    Figure 2.3 Sprinkle a correct dose of salt into the chamber

    stirring constantly

    4. When all salt has been sprinkled into the chamber, the saturated salt

    solution should have the ratio of 60 - 90 % undissolved salt to10 - 40 % liquid.

    5. Close the chamber with the chamber cover (Figure 2.4).

    Figure 2.4 Close the chamber with the chamber cover

  • 8/3/2019 Humidity Calibator

    12/23

    User's Guide M210185en-B

    6

    6. Fasten the salt chamber on the holder in the base plate and close the

    measurement holes with rubber plugs. The chambers can also be used

    as individual checkers without the base plate. NOTE: the plugs have

    three steps, each of which is suitable for a certain hole diameter: the

    first step for the 12 mm, the second step for the 13.5 mm and the thirdstep for the 18.5 mm hole. Keep the holes always closed when not

    calibrating.

    7. Write the preparation date on a sticker and mark the chamber with it.

    If you use a ready-dosed salt package, use the sticker with the batch

    code. Mark all parts of the salt chamber (chamber, chamber cover and

    transit cover) with stickers. This way, various covers and different

    salts do not get mixed.

    Figure 2.5 Mark the chamber and covers with stickers

    8. Allow approximately 24 hours for stabilization before use so that the

    salt solution reaches the equilibrium humidity.

    9. See Chapter4 on how to maintain the salt solutions.

    NOTE

    If you do not use the instrument for a longer period of

    time, close the chambers with transit covers.

  • 8/3/2019 Humidity Calibator

    13/23

    User's Guide M210185en-B

    7

    3. INSTRUCTIONS FOR THE CALIBRATION OF HUMIDITY

    INSTRUMENTS

    3.1. General instructions

    Usually, the errors during humidity calibration are due to temperature

    differences. A temperature difference of 1C at +20C between the air in the

    chamber and the sensor causes an error of 3 %RH at 50 %RH and an error of

    6 %RH at 97 %RH. The more the transportation or process temperatures

    differ from the temperature at the calibration site, the longer the stabilization

    time required. In laboratory use, the calibrator should be stored in that part of

    the room where the temperature is most stable. The calibrator must be kept out

    of direct sunlight and away from localized heat sources, such as spot lights,

    heaters and soldering irons. If the probe/transmitter is checked against several

    humidity references, the checking must first be made at the dry end. Moredetailed instructions are given in individual manuals of transmitters and

    probes.

    Handle the probe as little as possible. Do not hold the salt chamber or other

    parts of the calibrator in hand during calibration as they warm up and cause

    errors in the readings.

    NOTE: even the smallest water drop on the probe near the sensors distorts the

    readings. Make sure that chamber covers and plugs are carefully closed.

    0 0

    1 0

    2 0

    3 0

    4 0

    5 0

    C

    HMP233probe

    O-rings

    sensors

    max.1 cm

    thermometer

    saturated salsolution

    undissolvedsalt

    Figure 3.1 Salt chamber

  • 8/3/2019 Humidity Calibator

    14/23

    User's Guide M210185en-B

    8

    3.1.1. Thermometer

    The calibrator includes a thermometer which can contain either mercury or red

    capillary liquid. If the thermometer contains mercury, it is delivered with a

    calibration certificate. The accuracy of the mercury thermometer is sufficientfor checking the transmitter's temperature channel if desired. The mercury

    thermometer has been calibrated at five points and the result is traceable to the

    Measurement Standards Laboratory of the Centre for Metrology and

    Accreditation in Finland. For more accurate results, use the corrections given

    in the calibration certificate or their interpolated values. Consider also the

    parallax error which may occur while reading the results. The mercury

    thermometer should be recalibrated within three years in order to maintain the

    traceability.

    The manufacturer of the thermometer with red capillary liquid provides it with

    a factory calibration certificate. As the accuracy and stability of this kind of athermometer are inferior to those of a mercury thermometer, it is not

    recommended to be used for checking the temperature channel of the

    transmitter.

    The sleeve at the end of the thermometer's protective tube can be used in two

    different positions. The thermometer is delivered with the sleeve in such a

    position that it protects the part containing the capillary liquid. When turned

    the other way round, it acts as an adapter when the thermometer is inserted

    into the 13.5 mm hole (see Figure 3.2).

    Figure 3.2 The sleeve protects the part with capillary liquid (on the

    left) or acts as an adapter (on the right)

    During calibration, the thermometer is inserted into the 13.5 mm hole of a salt

    chamber. Press it downwards until it passes the O-rings. The thermometer is

    correctly in place when you can feel a resistance while pressing it downwards.

  • 8/3/2019 Humidity Calibator

    15/23

    User's Guide M210185en-B

    9

    Figure 3.3 During calibration, the thermometer is inserted into the

    13.5 mm hole

    When the thermometer is not in use or the calibrator is transferred from one

    place to another, place the thermometer in holders (Figure 3.4).

    Figure 3.4 Place the thermometer in holders when not in use

    3.2. Calibration

    Perform the calibration according to the following instructions:

    Leave the HMK15 calibrator and the probe at the calibration site for

    at least 30 minutes before starting the calibration in order to let the

    probe temperature stabilize to the room temperature.

    With lithium chloride it is not necessary to use the thermometer as the

    humidity reading changes only very slightly in the temperature range

    of +25...+30 C. However, if you use it make sure that the sleeve is in

    the correct position and insert it into the 13.5 mm hole of the LiCl salt

    chamber. NOTE: handle the thermometer as little as possible and do

    not handle it by the measuring end. Press the thermometer downwards

    until it passes the O-rings. The thermometer is correctly in place when

    you feel a resistance while pressing it downwards.

  • 8/3/2019 Humidity Calibator

    16/23

    User's Guide M210185en-B

    10

    Take off the grid or filter protecting the sensor. Take care not to

    damage the sensor.

    Some 12 mm probes need a adapter fitting to fit into the 13,5 mm

    hole. Adapter fitting prevents sensor from soaking in the salt liquid.In case you need an adapter fitting, take off the grid or filter and

    replace it with the adapter fitting. Two adapter fittings are attached on

    the base plate of the HMK15, see picture on page 1.

    CAUTION Remember to check from the user's guide of your device if theadapter fitting is needed. If sensor accidently soaks in salt liquid,

    remove it quickly and rinse with clean water. Let it dry before taking

    into use again.

    Insert the probe into a suitable hole of the LiCl salt chamber. Press it

    downwards until it passes the O-rings. The shorter the time the hole

    stays open before inserting the probe, the shorter the stabilization time

    required.

    Wait until the humidity reading stabilizes; this takes about 10 - 30

    minutes.

    Read the salt chamber temperature from the thermometer and then

    read the closest humidity value from the calibration table

    (Greenspan's calibration table, LiCl solution).

    Adjust the dry end (DRY, offset) to correspond to the value given in

    the calibration table.

    Insert the thermometer into the 13.5 mm hole of the NaCl salt

    chamber. NOTE: when calibrating probes that are being used for long

    term measurements (over 1 hour) in high humidities (90 - 100 %RH),

    use the K2SO4 salt as the high end reference.

    Insert the probe into a suitable hole of the NaCl (or K2SO4) salt

    chamber. The shorter the time the hole stays open before inserting the

    probe, the shorter the stabilization time required.

    Wait until the humidity reading stabilizes; this takes about 10 - 30

    minutes. Note that in high humidities the risk for errors increases.

    Therefore, the stabilization time should be longer (approximately

    20 - 40 minutes).

    Read the salt chamber temperature from the thermometer and then

    read the closest humidity value from the calibration table

    (Greenspan's calibration table, NaCl or K2SO4 solution).

    Adjust the wet end (WET, gain) to correspond to the value given in

    the calibration table.

  • 8/3/2019 Humidity Calibator

    17/23

    User's Guide M210185en-B

    11

    3.3. Greenspan's calibration table1

    C LiCl MgCl2 NaCl K 2SO4

    0 * 33.7 0.3 75.5 0.3 98.8 1.1

    5 * 33.6 0.3 75.7 0.3 98.5 0.910 * 33.5 0.2 75.7 0.2 98.2 0.8

    15 * 33.3 0.2 75.6 0.2 97.9 0.6

    20 11.3 0.3 33.1 0.2 75.5 0.1 97.6 0.5

    25 11.3 0.3 32.8 0.2 75.3 0.1 97.3 0.5

    30 11.3 0.2 32.4 0.1 75.1 0.1 97.0 0.4

    35 11.3 0.2 32.1 0.1 74.9 0.1 96.7 0.4

    40 11.2 0.2 31.6 0.1 74.7 0.1 96.4 0.4

    45 11.2 0.2 31.1 0.1 74.5 0.2 96.1 0.4

    50 11.1 0.2 30.5 0.1 74.4 0.2 95.8 0.5

    * If the LiCl solution is used or stored in temperatures below +18C, its equilibriumhumidity changes permanently

    The numbers on the right in each column indicate the uncertainty of the

    reference humidity of the salt in that temperature.

    1Greenspan, L.: Journal of Research of the National Bureau of Standards - A

    Physis and Chemistry Vol. 81A, No. 1 January-February 1977, pp. 89-95

    3.4. On-site calibrations and transportation

    The HMK15 calibrator is easily trasferred from one place to another. Two O-

    rings seal the transit cover on the chamber. The optional transit bag helps to

    transport the calibrator so that chambers stay in the vertical position.

    Furthermore, the bag can be used for housing the calibrator during calibration.

    When transferring the calibrator from one place to

    another, follow the instructions below:

    Turn the protective sleeve on the

    thermometer and place the

    thermometer in the holder.

    Replace chamber covers withtransit covers. Press chamber covers

    on vacant chamber holders for

    transportation.

    During transportation, keep the

    calibrator so that the chambers stay as

    upright as possible. This way, very

    little salt solution gets on transit

    covers and cleaning is easier. The

    closer the transportation temperature is to the temperature of the

    calibration site, the shorter the time required for temperaturestabilization. If the transportation temperature is below +18 C, the

  • 8/3/2019 Humidity Calibator

    18/23

    User's Guide M210185en-B

    12

    LiCl salt chamber should be transported separately to keep the

    solution warm.

    When the calibrator is at the calibration site, remove transit covers

    and fasten chamber covers on salt chambers. Close the holes with

    plugs. NOTE: the plugs have three steps each of which is suitable for

    a certain hole diameter: the first step for the 12 mm, the second step

    for the 13.5 mm and the third step for the 18.5 mm hole.

    Clean the transit covers with a damp cloth and press them on vacant

    chamber holders.

    Take the thermometer off the holder, turn the sleeve the other way

    round and perform the calibration according to Chapter3.2.

    In on-site calibrations it is important to allow enough time for the calibratorand probe temperatures to stabilize. Depending on the temperature differences

    between the transportation and the calibration site or between the probe

    removed from the process and the calibration site, a two point calibration

    takes about 0.5 - 2 hours. If there are several frequently calibrated instruments

    at the same site, knowing the stabilization times is useful.

    The following figure illustrates an example of temperature and humidity

    stabilization. In this example, an HMP233 probe has been transferred from a

    temperature of 75C (oven) to a suitable hole of the NaCl salt chamber; the

    calibrator is at room temperature. The figure shows that after 40 minutes, the

    humidity reading differs 0.2 %RH from the final reading.

    Time [min]

    Hum

    idity[%RH

    70

    71

    72

    73

    74

    75

    76

    77

    78

    0 5 10 15 20 25 30 35 40 45 50 55 60 65 70

    10

    20

    30

    40

    50

    60

    70

    80

    90

    Temp

    erature[C]

    HUMIDITY

    TEMPERATURE

    Figure 3.5 An example of RH and T stabilization when the probe is

    removed from a process

  • 8/3/2019 Humidity Calibator

    19/23

    User's Guide M210185en-B

    13

    Time [min.]

    Humidity[%RH

    60

    62

    64

    66

    68

    70

    72

    74

    0 10 20 30 40 50 60 70 80 90 100 110 120

    0

    4

    8

    12

    16

    20

    24

    28

    Temperature[C]

    HUMIDITY

    TEMPERATURE

    Figure 3.6 An example of the RH and T stabilization when the

    transportation and calibration site temperatures are

    different

  • 8/3/2019 Humidity Calibator

    20/23

    User's Guide M210185en-B

    14

    In the following example, the transportation temperature of the calibrator

    (without the LiCl) is +5C; the calibrator is then transferred into room

    temperature. After this, an HMP233 probe that has been stored at room

    temperature, is inserted into a suitable hole of the NaCl salt chamber. The

    figure shows that after 40 minutes, the reading differs 1.4 %RH from the finalreading.

    Time [min.]

    Humidity[%RH]

    60

    62

    64

    66

    68

    70

    72

    74

    0 10 20 30 40 50 60 70 80 90 100 110 120

    0

    4

    8

    12

    16

    20

    24

    28

    Tempe

    rature[C]

    HUMIDITY

    TEMPERATURE

    Figure 3.7 An example of the RH and T stabilization when the

    transportation and calibration site temperatures are

    different

  • 8/3/2019 Humidity Calibator

    21/23

    User's Guide M210185en-B

    15

    4. MAINTENANCE

    Depending on the frequency of use and the general operating conditions, the

    salt solutions maintain their characteristics for 6 - 12 months. After that, theymust be replaced. A visual check should be performed at intervals of 2 - 3

    months. There must be a minimum of approximately 10% of undissolved salt

    at the bottom of the chamber (max. 90%) and the salt must be clean, otherwise

    it must be reprepared. Please, note that the LiCl solution may crystallize on

    the surface. However, there may still be solution under the surface. Therefore,

    in such cases stir the solution and check it again the next day. It is advisable to

    keep a maintenance log which shows the dates of salt replacements and other

    maintenance procedures.

    For correct calibration it is essential that salt chambers are tightly closed.

    Check the O-rings at each salt replacement. If they are damaged, replace themwith new ones. The O-rings used are of the following types:

    O-ring location dimensions type

    measurement hole 12 mm 12 x 2.5 mm NBR70

    measurement hole 13.5 mm 13.5 x 2.5 mm NBR70

    measurement hole 18.5 mm 18.3 x 2.4 mm NBR70

    transit cover 41.2 x 3.0 mm NBR70

    chamber 50.0 x 2.0 mm NBR70

    5. SPARE PARTS AND ACCESSORIES

    Order code Description

    19728HM Mercury thermometer with calibration certificate

    25130HM Thermometer with red capillary liquid

    19729HM Ready-dosed LiCl salt package with calibration certificate

    19730HM Ready-dosed MgCl2 salt package with calibration certificate

    19731HM Ready-dosed NaCl salt package with calibration certificate

    19732HM Ready-dosed K 2SO4 salt package with calibration certificate

    19746HM Plug set

    19766HM Salt chamber

    19767HM Ion exchanged water

    211011 Adapter fitting for 12 mm probes

    HM27032 Transit bag

  • 8/3/2019 Humidity Calibator

    22/23

    User's Guide M210185en-B

    16

    6. TECHNICAL DATA

    Operating temperature range +0 ...+50 C

    NOTE

    If the LiCl solution is used or stored in

    temperatures below +18C, its equilibrium humidity

    changes permanently.

    Accuracy:

    lithium chloride LiCl 1.0 %RH + Greenspan's uncertainty*

    magnesium chloride MgCl2 1.0 %RH + Greenspan's uncertainty*

    natrium chloride NaCl 1.4 %RH + Greenspan's uncertainty*

    potassium sulphate K2SO4 1.5 %RH + Greenspan's uncertainty*

    * = The uncertainty given in Greenspan's calibration table (see Chapter 3.3) at the calibration

    temperature, e.g. the accuracy of LiCl salt at +20 C is (1.0 + 0.3) %RH = 1.3 %RH

    Response time (probe and calibrator

    at the same temperature) with Vaisala sensors typically 10 min.

    (deviation of the final value < 1%RH)

    Accuracy of the thermometer:

    with mercury 0.3 C

    with red capillary liquid 1 C

    Dimensions 20 x 9 x 23 cm

    Materials metal parts: anodised aluminium

    Weight 1 kg without salt solutions

  • 8/3/2019 Humidity Calibator

    23/23

    User's Guide M210185en-B

    Warranty

    Vaisala hereby represents and warrants all Products manufactured by Vaisala and sold hereunder to be free from

    defects in workmanship or material during a period of twelve (12) months from the date of delivery save for

    products for which a special warranty is given. If any Product proves however to be defective in workmanship ormaterial within the period herein provided Vaisala undertakes to the exclusion of any other remedy to repair or at

    its own option replace the defective Product or part thereof free of charge and otherwise on the same conditions

    as for the original Product or part without extension to original warranty time. Defective parts replaced in

    accordance with this clause shall be placed at the disposal of Vaisala.

    Vaisala also warrants the quality of all repair and service works performed by its employees to products sold by

    it. In case the repair or service works should appear inadequate or faulty and should this cause malfunction or

    nonfunction of the product to which the service was performed Vaisala shall at its free option either repair or

    have repaired or replace the product in question. The working hours used by employees of Vaisala for such repair

    or replacement shall be free of charge to the client. This service warranty shall be valid for a period of six (6)

    months from the date the service measures were completed.

    This warranty is however subject to following conditions:

    a) A substantiated written claim as to any alleged defects shall have been received by Vaisala within thirty (30)

    days after the defect or fault became known or occurred, and

    b) The allegedly defective Product or part shall, should Vaisala so require, be sent to the works of Vaisala or to

    such other place as Vaisala may indicate in writing, freight and insurance prepaid and properly packed and

    labelled, unless Vaisala agrees to inspect and repair the Product or replace it on site.

    This warranty does not however apply when the defect has been caused through

    a) normal wear and tear or accident;

    b) misuse or other unsuitable or unauthorized use of the Product or negligence or error in storing, maintaining or

    in handling the Product or any equipment thereof;

    c) wrong installation or assembly or failure to service the Product or otherwise follow Vaisala's service

    instructions including any repairs or installation or assembly or service made by unauthorized personnel not

    approved by Vaisala or replacements with parts not manufactured or supplied by Vaisala;

    d) modifications or changes of the Product as well as any adding to it without Vaisala's prior authorization;

    e) other factors depending on the Customer or a third party.

    Notwithstanding the aforesaid Vaisala's liability under this clause shall not apply to any defects arising out of

    materials, designs or instructions provided by the Customer.

    This warranty is expressly in lieu of and excludes all other conditions, warranties and liabilities, express or

    implied, whether under law, statute or otherwise, including without limitation any implied warranties of

    merchantability or of fitness for a particular purpose and all other obligations and liabilities of Vaisala or its

    representatives with respect to any defect or deficiency applicable to or resulting directly or indirectly from the

    Products supplied hereunder, which obligations and liabilities are hereby expressly cancelled and waived.

    Vaisala's liability shall under no circumstances exceed the invoice price of any Product for which a warranty

    claim is made, nor shall Vaisala in any circumstances be liable for lost profits or other consequential loss whether

    direct or indirect or for special damages.