HPT001.306 Rev. 3 Page 1 of 33 TP-1 TVAN Technical Training Health Physics (RADCON) Initial Training...

33
HPT001.306 Rev. 3 Page 1 of 33 TP-1 TVAN Technical Training Health Physics (RADCON) Initial Training Program Response Checks of Portable Radiation Detection Instruments HPT 001.306

Transcript of HPT001.306 Rev. 3 Page 1 of 33 TP-1 TVAN Technical Training Health Physics (RADCON) Initial Training...

HPT001.306Rev. 3Page 1 of 33

TP-1 TVAN Technical TrainingHealth Physics (RADCON) Initial Training Program

Response Checks of PortableRadiation Detection Instruments

HPT 001.306

HPT001.306Rev. 3Page 2 of 33

TP-2 TVAN Technical TrainingHealth Physics (RADCON) Initial Training Program

Enabling Objectives

Name Instrument Control Procedure

State Calibration Organization

State Frequency of Response Checks

Identify Response Check Equipment

Identify Shepherd Components

HPT001.306Rev. 3Page 3 of 33

TP-3 TVAN Technical TrainingHealth Physics (RADCON) Initial Training Program

Enabling Objectives

State Isotopes & Activity for Shepherd Sources

Identify Response Window Criteria Identify Instruments Response

Checked with Plutonium State Purpose of Scintrex Model

309A

HPT001.306Rev. 3Page 4 of 33

TP-4 TVAN Technical TrainingHealth Physics (RADCON) Initial Training Program

Instrumentation Controls

TVAN RCDP-8,

“Radiological Instrument/ Equipment Controls”

HPT001.306Rev. 3Page 5 of 33

TP-5 TVAN Technical TrainingHealth Physics (RADCON) Initial Training Program

Instrument Calibration

Portable Radiation Detection Instruments Calibrated by:

Environmental RadiologicalMonitoring & Instrumentation (ERM&I)

Western Area Radiological Laboratory,(WARL) Muscle Shoals, AL

HPT001.306Rev. 3Page 6 of 33

TP-6 TVAN Technical TrainingHealth Physics (RADCON) Initial Training Program

Response Check Frequency

Teletector & RSO-7 Daily Beta-Gamma Instruments Weekly Alpha & Beta Friskers Weekly Neutron Dose Instruments Weekly On-Site REP Instruments Weekly Electronic Dosimeters

Quarterly

HPT001.306Rev. 3Page 7 of 33

TP-7 TVAN Technical TrainingHealth Physics (RADCON) Initial Training Program

Response Checks ofBeta-Gamma Dose Instruments

Shepherd Model 89 Irradiator

- Source Rod Assembly

- Shielded Calibration Range

Assembly

- Attenuator Assembly

- Table Assembly

HPT001.306Rev. 3Page 8 of 33

TP-8 TVAN Technical TrainingHealth Physics (RADCON) Initial Training Program

Shepherd Source Rod Assembly

Cs-137130 mCi400 Ci

HPT001.306Rev. 3Page 9 of 33

TP-9 TVAN Technical TrainingHealth Physics (RADCON) Initial Training Program

Shepherd Shielded Calibration Range Assembly

HPT001.306Rev. 3Page 10 of 33

TP-10 TVAN Technical TrainingHealth Physics (RADCON) Initial Training Program

Shepherd Attenuator Assembly

HPT001.306Rev. 3Page 11 of 33

TP-11 TVAN Technical TrainingHealth Physics (RADCON) Initial Training Program

Shepherd Table Assembly

HPT001.306Rev. 3Page 12 of 33

TP-12 TVAN Technical TrainingHealth Physics (RADCON) Initial Training Program

Shepherd Precautions

Plugs in Unused Access Ports Hands & Fingers Clear of Door Never Lubricate Source Rod Don’t Twist Attenuator Handles Move Table Slowly Utilize Survey Meter

HPT001.306Rev. 3Page 13 of 33

TP-13 TVAN Technical TrainingHealth Physics (RADCON) Initial Training Program

Shepherd Performance Checks

Test the Interlock System- Attempt to Raise Source or Open Door with Power Off- Attempt to Raise Source with Door Open- Attempt to Open Door with Source Raised

HPT001.306Rev. 3Page 14 of 33

TP-14 TVAN Technical TrainingHealth Physics (RADCON) Initial Training Program

Establishing Response Windows

Expose at Least Four Instruments

Exposure Values Near Midscale

Set Window for Each Scale

HPT001.306Rev. 3Page 15 of 33

TP-15 TVAN Technical TrainingHealth Physics (RADCON) Initial Training Program

Response Window Criteria

Arithmetic Mean of all Instruments Calculate Highest Percent Difference Calculate Lowest Percent Difference Discard Highest or Lowest Percent

Difference if Exceed 10 Percent Response Window = Mean ± 20 % Expressed as Range

HPT001.306Rev. 3Page 16 of 33

TP-16 TVAN Technical TrainingHealth Physics (RADCON) Initial Training Program

Instrument Pre-Operational Checks

Inspect for Damage Inspect Switches, Alarms, Controls Inspect Cables & Power Cords Check Battery Strength Instrument in Calibration

HPT001.306Rev. 3Page 17 of 33

TP-17 TVAN Technical TrainingHealth Physics (RADCON) Initial Training Program

Performing Response Checks with Shepherd Irradiator

Position Instrument on Guide

- Beta Gamma Survey Meter

- Teletector

- Electronic Dosimeter

HPT001.306Rev. 3Page 18 of 33

TP-18 TVAN Technical TrainingHealth Physics (RADCON) Initial Training Program

Beta-Gamma Survey Meter

HPT001.306Rev. 3Page 19 of 33

TP-19 TVAN Technical TrainingHealth Physics (RADCON) Initial Training Program

Teletector

HPT001.306Rev. 3Page 20 of 33

TP-20 TVAN Technical TrainingHealth Physics (RADCON) Initial Training Program

Electronic Dosimeter

HPT001.306Rev. 3Page 21 of 33

TP-21 TVAN Technical TrainingHealth Physics (RADCON) Initial Training Program

Performing Response Checks with Shepherd Irradiator Position Attenuators and Platform Close Door Raise Source to make Exposure Lower Source to “Off” Repeat for Each Scale Document Results

HPT001.306Rev. 3Page 22 of 33

TP-22 TVAN Technical TrainingHealth Physics (RADCON) Initial Training Program

Observation Window

HPT001.306Rev. 3Page 23 of 33

TP-23 TVAN Technical TrainingHealth Physics (RADCON) Initial Training Program

Response Check of AlphaSurvey Instruments Alpha Sources, ie Pu-239 Inspect Mylar Windows Perform Light Test Expose Detector Compare Results with Response Log Check Each Scale or Decade Document Results

HPT001.306Rev. 3Page 24 of 33

TP-24 TVAN Technical TrainingHealth Physics (RADCON) Initial Training Program

Alpha Survey Instrument

HPT001.306Rev. 3Page 25 of 33

TP-25 TVAN Technical TrainingHealth Physics (RADCON) Initial Training Program

Neutron Survey Instrument

HPT001.306Rev. 3Page 26 of 33

TP-26 TVAN Technical TrainingHealth Physics (RADCON) Initial Training Program

Response Check ofNeutron Instruments

Set up the Instrument Scale Instrument Over Neutron Source Check Other Scales/Decades with

Pulse Generator Compare Results with Response

Log Document Results

HPT001.306Rev. 3Page 27 of 33

TP-27 TVAN Technical TrainingHealth Physics (RADCON) Initial Training Program

Frisker

HPT001.306Rev. 3Page 28 of 33

TP-28 TVAN Technical TrainingHealth Physics (RADCON) Initial Training Program

Response Check ofFriskers

Sources Identified in Response Log Set Frisker to Appropriate

Range/Scale Position Probe over Source Repeat for Other Scales Compare Results with Response Log Document Results

HPT001.306Rev. 3Page 29 of 33

TP-29 TVAN Technical TrainingHealth Physics (RADCON) Initial Training Program

Tritium in Air Monitor

HPT001.306Rev. 3Page 30 of 33

TP-30 TVAN Technical TrainingHealth Physics (RADCON) Initial Training Program

Response Check ofScintrex Model 309A

Inspect Filter Paper Ensure Battery >3.0 volts Set Function Switch to ‘Display On’ Press ‘Test’ Button Observe Display < 50

HPT001.306Rev. 3Page 31 of 33

TP-31 TVAN Technical TrainingHealth Physics (RADCON) Initial Training Program

Response Check ofScintrex Model 309A

Position Gamma Source on Right Side

Wait 1-2 Minutes Observe Reading Between 8 & 24 Depress ‘Test’ Button Document Results

HPT001.306Rev. 3Page 32 of 33

TP-32 TVAN Technical TrainingHealth Physics (RADCON) Initial Training Program

Summary Instrument Calibration Critical Response Checks Confirm Operation Shepherd Model 89 Irradiator Other Response Checks:

- Alpha Survey Instruments- Neutron Instruments- Friskers- Tritium-in-Air

HPT001.306Rev. 3Page 33 of 33

TP-33 TVAN Technical TrainingHealth Physics (RADCON) Initial Training Program

Prevent Errors -Reduce Contamination