Liquid chromatography (LC) is a physical separation technique conducted in the liquid phase. A sample is separated into its constituent components by distributing between the mobile phase (a flowing liquid) and a stationary phase (sorbents packed inside a column). HPLC HPLC is a modern form of LC that uses small- particle columns through which the mobile phase is pumped at high pressure.

Transcript of HPLC

Page 1: HPLC

Liquid chromatography (LC) is a physical separation technique conducted in the liquid phase.

A sample is separated into its constituent components by distributing between the mobile phase (a flowing liquid) and a stationary phase (sorbents packed inside a column).

HPLC

HPLC is a modern form of LC that uses small-particle columns through which the mobile phase is pumped at high pressure.

Page 2: HPLC

1. high resolving power;

2. speed of separation;

3. continuous monitoring of the column effluent;

4. accurate quantitative measurement;

5. repetitive and reproducible analysis using the same column; and

6. automation of the analytical procedure and data handling.

High Performance?

Page 3: HPLC

High Pressure Liquid Chromatography (HPLC)

A representation of the dynamic partitioning process of the analytes between the flowing liquid and a spherical packing particle. The movement of component B is retarded in the column because each B molecule has stronger affinity for the stationary phase than the A molecule.

A schematic of the chromatographic process, where a mixture of analytes A and B are separated into two distinct bands as they migrate down the column filled with packing (stationary phase).

An in-line detector monitors the concentration of each separated component band in the effluent and generates a trace called the “chromatogram

Page 4: HPLC

Classical LC was first discovered by Mikhail Tswett in 1903 The invention of gas chromatography (GC) by A. J. P. Martin in 1952

The first generation of high-performance liquid chromatographs was in the 1960s

A Brief History

Page 5: HPLC

A Brief Introduction

Page 6: HPLC

East West University’s HPLC

Page 7: HPLC

Advantages and Limitations of HPLC

Page 8: HPLC

A Schematic Diagram of HPLC

Page 9: HPLC

Accurate, precise and robust method for quantitative analysis of pharmaceutical products.

Monitoring the stability of drugs in formulations with quantitation of any degradation products.

Measurement of drugs and their metabolites in biological fluids.

Determination of partition coefficients and pKa values of drugs and of drug protein binding.

Purification of drugs

Applications of HPLC

Page 10: HPLC

HPLC Separation of Benzodiazepines

Peaks:1 = bromazepam; 2 = nitrazepam; 3 = clonazepam; 4 = oxazepam; 5 = flunitrazepam;6 = hydroxydiazepam (temazepam); 7 = desmethyldiazepam (nordazepam);8 = diazepam (valium).

stationary phase: ChromSphere C18, 1.5 mm (non-porous); mobile phase: 3.5 ml/min; water–acetonitrile (85 : 15); temperature: 350C; UV detector 254 nm.

Conditions: sample: 40 ng each; column: 3 cm x 4.6 mm i.d.;

Page 11: HPLC

Column: Zorbax Eclipse-XDB C18; Solvent: acetonitrile/water mixture at pH = 3

pind

olol

met

opro

lol

labe

talo

l prop

anol

olal

pren

olol

o-ch

loro

anili

ne

Typical modern chromatogram of a mixture of b-blockers

Page 12: HPLC

HPLC Chromatogram

Page 13: HPLC

Instrumentation

Recriprocating

SyringeManual Injection

Automated Injection

Packings

Measure column performance

Column care and use

SolventDelivery

SampleIntroduction

Column Packingsand harware

Detectors

UV-Vis

Fluorescence

Refractive Index

ConductivityVolatametry/Amperometry

HPLC Instrumentation and TechniquesMobile phase reservoirsConstruction pump material

Pumps

Page 14: HPLC

The reagent bottle that holds HPLC solvent is used as a reservoir.

Solvent is delivered from the reservoir to the pump by means of Teflon tubing--called the "inlet line" to the pump

Bottle capped on outside with a UV- absorbing plastic

He degasser

Solvent Delivery System in HPLC

Mobile Phase Resorvoir

The reservoir that holds the mobile phase is often no more than a glass bottle.

Page 15: HPLC

Push filter (10 mm) prevent any dust or particulate matter to enter the pump.

This is because particulate matter can interfere with pumping action, damage valves and seals. In addition, it may also damage the column (by collecting at the top of the column).

Purpose of a sinker frit

Page 16: HPLC

Filtration of Mobile Phase of HPLC

Page 17: HPLC
Page 18: HPLC

DEGASSING

The practice of removing air from the mobile phase; degassing can be achieved by bubbling He gas into the M.P

He degassing removes dissolve oxygen from the M.P.

The presence of oxygen in mobile phase causes bubble formation resulting in air in the flow system and pump pressure will change causing spike in the chromatogram (due to air bubble formation in the detector cell)

Degassing of the liquid mobile phase

Why is required?

Instead of He degassing vaccum degassing method can also be used

Page 19: HPLC

The vessel should have some sort of cap to prevent particulate matter from contaminating the mobile phase. If you are using a solvent bottle as a reservoir, the top of the bottle can be wrapped in aluminum foil to keep dust out or the bottle cap can be drilled to allow inserting the inlet line through the cap.

Don't close the bottle too tightly or removal of mobile phase by the pump will create a vacuum.

This prevents mobile phase from flowing the pump, creating a "vapor lock" within the pump.

Requirements for a solvent reservoir

The reservoir and its attachment to the pump should be made of materials that will not contaminate the mobile phase: Teflon, glass, or stainless steel.

Page 20: HPLC

Filtration of Sample of HPLC

Page 21: HPLC

The actual cross-sectional distance across the open interior of a tube through which mobile phase flows.

This is an important parameter in at least three components: connecting tubing, column tubing, injector tubing.

Inner Diameter (ID) of Tubing of HPLC

The volume-length relationship is shown in the table below.

Page 22: HPLC

Pumps of HPLC

Page 23: HPLC

Reciprocating PumpPumping of Mobile Phase in HPLC System

Page 24: HPLC

Elution of Mobile Phase in HPLC System

Page 25: HPLC

Isocratic elution

An HPLC system generating a separation under conditions in which composition of the mobile phase does not change with time.

Gradient elution

A gradient denotes that there is a parameter in the separation that is changed with time in order to affect the elution.

In HPLC a gradient refers to a change in the mobile-phase composition with time.

Gradient profiles can be linear, step, nonlinear, convex and concave.

Elution of Mobile Phase in HPLC System

composition of the solvent is changed continuously or in a series of steps.

Page 26: HPLC

Gradient vs. Isocratic Elution in HPLC System

Page 27: HPLC

Low Pressure Gradient Elution in HPLC System

Page 28: HPLC

High Pressure Gradient Elution in HPLC System

Page 29: HPLC

Injector of HPLC System

Interchangeable loops are available to provide a choice of sample sizes ranging from 5 to 2000 /mL.

Page 30: HPLC

Load/Inject Position of Injector

Load position Inject position

Page 31: HPLC

Load position

Inject position

Load/Inject Position of Injector

Page 32: HPLC

Load/Inject Position of Injector

Load position Inject position

Page 33: HPLC

Sample Injecting Syringes of HPLC

Page 34: HPLC

Injector and Loop of HPLC

Page 35: HPLC

35

Filling the Loop by Autosampler

Filling the loop by three different autosampler designs:

A, pull-loop; B, push-loop; C, integral-loop injection.

1 = sample vial, 2 = syringe with stepper motor, 3 = loop, 4 = from pump

5 = to column, 6 = to waste, 7 = lowpressure seal, 8 = high-pressure seal,

9 = position of the movable loop when the valve is switched and the sample transferred to the column.

Page 36: HPLC

Sample Rack Control Rack

Injection of Sample by Auto Sampler

Page 37: HPLC

Injection of Sample Mechanism by Auto Sampler

Many HPLC instruments incorporate an autosampler with an automatic injector. These can inject continuously variable volumes.

Page 38: HPLC

Injection of Sample Mechanism by Auto Sampler

Page 39: HPLC

HPLC Detector

The function of the detector in HPLC is to monitor the

mobile phase as it emerges from the column.

Page 40: HPLC

Cells of HPLC Detector

The detector should be small and compatible with liquid flow.

No highly sensitive, universal detector system is available for high-performance liquid chromatography.

The detector used will depend on the nature of the sample.

Page 41: HPLC

The UV absorption detector is the most widely used in HPLC. It is based on the principle of absorption of UV/visible light as the effluent from the column is passed through a small flow cell held in the radiation beam.

It is characterised by high sensitivity (detection limit of about 1 x 10-9 g/mL - for highly absorbing compounds) and it is relatively insensitive to changes of temperature and flow rate.

The detector is generally suitable for gradient elution work since many of the solvents used in HPLC do not absorb to any significant extent at the wavelengths used for monitoring the column effluent.

The presence of air bubbles in the mobile phase can greatly impair the detector signal, causing spikes on the chromatogram.

Both single and double beam instruments are commercially available. Although the original detectors were single- or dual-wavelength instruments (254 and/or 280 nm),some manufacturers now supply variable-wavelength detectors covering the range 210-800 nm so that more selective detection is possible.

Ultraviolet Detectors

Page 42: HPLC

UV-Vis PDA

HPLC Detectors

Page 43: HPLC

PDA Detectors

Page 44: HPLC

Photodiode array detectors scan a range of

wavelengths every few milliseconds and

continually generate spectral information.

Wavelength, RT and absorbance can all be

plotted.

PDA detector provide three dimensional

information that allows an accurate

assessment of peak identity, purity and

quantitation in a single run.

Photo Diode Array Detector (PDA)

Page 45: HPLC

RT

Absorbance

Wavelength

RT

Absorbance

UV-Vis Detector PDA

Three Dimensional Chromatogram by PDA

Page 46: HPLC

Three Dimensional Chromatogram by PDA

Page 47: HPLC

Three Dimensional Chromatogram by PDA

Page 48: HPLC

It is difficult to determine component purity from a chromatogram.

A PDA detector can analyze peak purity by comparing spectra within a peak.

A pure peak has matching spectra throughout the peak (at all wavelengths)

Peak purity analysis by 3D Chromatogram of PDA

Page 49: HPLC

These detectors are based on the change of refractive index of the eluant from the column with respect to pure mobile phase.

Although they are widely used, the refractive index detectors suffer from several disadvantages -

Refractive Index Detectors

lack of high sensitivity,

lack of suitability for gradient elution, and

the need for strict temperature control ( +/- 0.001oC) to operate at

their highest sensitivity

Page 50: HPLC

Refractive Index Detectors

In this detector, both the column mobile phase and a reference flow of solvent are passed through small cells on the back surface of a prism.

When the two liquids are identical, there is no difference between the two beams reaching the photocell, but when the mobile phase containing solute passes through the cell, there is a change in the amount of light transmitted to the photocell, and a signal is produced.

Page 51: HPLC

These Detectors enable fluorescent compounds present in the mobile phase to be detected by passing the column effluent through a cell irradiated with ultraviolet light and measuring any resultant fluorescent radiation.

Although only a small proportion of inorganic and organic compounds are naturally fluorescent, many biologically active compounds (e.g. drugs) and environmental contaminants (e.g. polycyclic aromatic hydrocarbons) are fluorescent.

These Detectors have very high sensitivity.

Because both the excitation wavelength and the detected wavelength can be varied, the detector can be made selective.

An widespread used detector.

Fluorescence Detectors

Page 52: HPLC

Columns for HPLC

They are generally made from stainless steel tubing, fittings, and frits.

Most HPLC columns are made of 316-grade stainless steel, which is chromium–nickel–molybdenum steel, resistant to the usual HPLC pressure and also relatively inert to chemical corrosion.

Columns, of i.d. 2–5mm are generally used for analytical purposes.

Wider columns of i.d. between 10 mm and 25.4 mm may be used for preparative work.

Some companies even market preparative columns of i.d. 30 cm and more. Columns 5, 10, 15 or 25 cm long are

common

Page 53: HPLC

Effect on chromatography

• Short (30-150mm) - short run times, low backpressure• Long (250-300mm) - higher resolution, long run times

Column length

Page 54: HPLC

Narrow bore ( 2.1 mm) - higher detector sensitivity, Sharp peak Analytical – 4.6 mm Preparative (10-22 mm) - high sample loading

Column ID

Page 55: HPLC

The retention time of an unretained component (often marked by the first baseline disturbance caused by the elution of the sample solvent) is termed void time (to).

Retention Time

The time between the sample injection point and the analyte reaching a detector is called the retention time (tR).

Void Time (to).

The volume of solvent eluted during void time is known as the column void volume (V0).

Void Volume (V0).

Chromatographic terminology and Definition

Rt, T0, V0

Page 56: HPLC

56

Even if the analyte does not interact with the stationary phase, it will notappear in the detector immediately after injection.

An HPLC column is filled with small particles of porous material which have a significant volume of the liquid phase between the particles and inside their porous space, so the non-interacting analyte still has to travel through this volume before it enters the detector.

The volume of the liquid phase in the column is called “void volume” (V0).

Several other names are also used in the chromatographic literature: “deadvolume,” “hold-up volume,”

If a particular HPLC system provides constant and stable mobile-phase flow (F), one can convert retention volume (VR) and void volume (V0) into the retention time (tR) and a void time (t0).

Void Volume

Page 57: HPLC

Note that the void volume (V0) is equal to the void time (t0) multiplied by the flow rate (F). Vo = t0 • F

Or

Void volume V0 = 0.65 • π r2 Lwhere r is the inner radius of the column and L is the

length of the column.

Void Volume (V0)The concept of column void volume (V0) is important for several reasons.

Void volume is the volume of the empty column minus the volume occupied by the solid packing materials. It is the liquid holdup volume of the column that each analyte must elute from.

Chromatographic terminology and Definition

V0

Page 58: HPLC

1. operating flow-rate range, 2. sample loading capacity, and 3. mass sensitivity (the minimum detectable amount of the assay).

Note that V0 is proportional to the square of the inner radius of the column.

It is important to have a rough idea of the void volume of the column since it often dictates the

For example, a typical analytical column (150 mm • 4.6 mm i.d.) has a V0 of about 1.5 mL and is operated at ~l.0 mL/min.

In contrast, by reducing the inner diameter to 2.0 mm, a typical LC/MS column (150 mm • 2.0 mm i.d.) has a V0 of about 0.3 mL and is operated at ~0.2 mL/min.

Column void volumes also control the volumes of the eluting peaks.

Smaller column void volumes lead to smaller peak volumes and, therefore, higher analyte concentration. As a result, if the same mass of analyte is injected, small-diameter columns lead to higher sensitivity.

Chromatographic terminology and Definition

V0

Page 59: HPLC

The peak of HPLC chromatogram has both width (Wb) and height (h).

The subscript b denotes that the width was measured at the base line.

Sometimes the width halfway up the peak (W1/2) or at 5% of peak height (W0.05) is used to meet the method or compendial requirements.

Chromatographic terminology and Definition

Wb and h

Page 60: HPLC

60

Void time can be interpreted as part of the total analyte retention time that the analyte actually spends in the mobile phase moving through the column, and for the rest of the retention time the analyte sits on the stationary phase surface.

The void volume or void time is a very important parameter, and its correct determination could be critical for the interpretation of the experimental results.

V0

Page 61: HPLC

Four major descriptors are commonly used to report characteristics of the chromatographic column, system, and particular separation:

1. Retention factor (k)

2. Efficiency (N)

3. Selectivity ()

4. Resolution (R)

BASIC CHROMATOGRAPHIC DESCRIPTORS

Page 62: HPLC

62

The analyte retention consists of two parts: (1) the time the component resides in the mobile phase actually moving through

the column and (2) the time the analyte is retained on the stationary phase.

The difference between the total retention time (tR) and the hold-up time is called the reduced retention time (t’R), and corresponding difference between the analyte retention volume and the void volume is called the reduced retention volume, V’R.

The ratio of the reduced retention volume to the void volume is a widely used dimensionless parameter called retention factor, k.

Retention Factor (k)

Page 63: HPLC

A Quantitative Description of Column Efficiency

Two related terms are widely used as quantitative measures of chromatographic column efficiency:

(1) plate height (H)(2) plate count or number of theoretical plates (N).

The efficiency of chromatographic columns increases as the plate count (N) becomes greater and as the plate height (H) becomes smaller.

Enormous differences in efficiencies are encountered in columns as a result of differences in column type and in mobile and stationary phases.

Efficiencies in terms of plate numbers can vary from a few hundred to several hundred thousand.

Page 64: HPLC

Plate theory is used for the evaluation of the column efficiency.

Plate theory assumes that the analyte is in the instant equilibrium with the stationary phase and the column is considered to be divided into a number of hypothetical plates.

Each plate has a finite height (height of effective theoretical plate, HETP), and an analyte spends a finite time in this plate. This time is considered to be sufficient to achieve quilibrium.

The smaller the plate height or the greater the number of plates, the more efficient the analyte exchange is between two phases, and the better the separation.

That is why column efficiency is measured in number of theoretical plates.

Plate Theory

Page 65: HPLC

Xs = K . Xm

Where Xm is the concentration of solute in the mobile phase, Xs is the concentration of solute in the stationary phase, and K is the distribution coefficient of the solute between the two phases.

It should be noted that K is defined with reference to the stationary phase (i.e., K =Xs/Xm), thus the larger the distribution coefficient, the more the solute is distributed in the stationary phase.

Consider three consecutive plates in a column, the p -1, the p, and the p =1 plates and let there be a total of n plates in the column.

The three plates are depicted in Figure.

Experimental Determination of the Number of Plates in a Column

Page 66: HPLC

In most chromatograms, peaks tend to be Gaussian in shape and broaden with time, where Wb becomes larger with longer tR. This is caused by

band-broadening effects inside the column, and is fundamental to all chromatographic processes.

The term, plate number (N), is a quantitative measure of the efficiency of the column, and is related to the ratio of the retention time (tR) and

the standard deviation of the peak width (Wb).

Column Efficiency (N)

theoretical plates

Experimental Determination of the Number of Plates in a Column

Page 67: HPLC

Number of theoretical plates,

Following chromatogram shows a peak width (wb) of 10 units and a tR of 135 units. The column efficiency (N) can therefore be calculated as follows:

Experimental Determination of the Number of Plates in a Column

Page 68: HPLC

68

The plate theory shows that the peak width (the dispersion or peak spreading) is inversely proportional to the square root of the efficiency and, thus, the higher the efficiency, the narrower the peak.

Experimental Determination of the Number of Plates in a Column

Page 69: HPLC

Selectivity

The ability of the chromatographic system to discriminate different analytes is called selectivity (a).

Selectivity is determined as the ratio of the retention factors of two analytes, or the ratio of the reduced retention times

The increase of the selectivity in the development of the separation of a complex mixture is the primary goal of any chromatographer,

because if the selectivity for the pair of analytes is equal to 1, then it does not matter how narrow your peaks or how fast your separation—you will not be able to separate these components until you increase the selectivity.

Page 70: HPLC

70

I. Peaks are narrow and far from each other, simple decrease of the column length or flow rate can significantly shorten the runtime without the loss of separation quality.

II: Acceptable separation, method may not be rugged.

III: Acceptable separation, quantitation reproducibility could be low.

IV: Bad separation.

Selectivity vs. Eficency

Page 71: HPLC

71

?

A chromatographic analysis for the chlorinated pesticide Dieldrin gives a peak with a retention time of 8.68 min and a baseline width of 0.29 min. How many theoretical plates are involved in this separation? Given that the column used in this analysis is 2.0 meters long, what is the

height of a theoretical plate?

the number of theoretical plates is

Solving equation for H gives the average height of a theoretical plate as

Page 72: HPLC

72

Resolution

Resolution, R, is defined as the ratio of the distance between two peaks to the average width of these peaks at baseline.

This descriptor encompasses both the efficiency and selectivity.

The resolution (R) of a column tells how far apart two bands are relative to their widths.

The resolution provides a quantitative measure of the ability of the columnto separate two analytes.

Page 73: HPLC

Parameters affecting efficiency: Flow rate Column length Particle diameter Particle size distribution

Parameters affecting retention factor: Eluent type Eluent composition Stationary phase type Analyte nature

Parameters affecting selectivity: Stationary phase type Analyte nature Eluent additives Temperature Eluent composition (ionizable analytes)

Factors Influencing HPLC Separation

Page 74: HPLC

• Silanol interactions

• Degradation of stationary phase

• Unswept void volume, or void formation at head of column,

• Co-eluting material

• POOR MATCH BETWEEN ANALYTE, MOBILE PHASE, AND

COLUMN POLARITIES

Peak Tailing

Page 75: HPLC

Tailing Factor (Tf)

Front width

Peak width

5% height

• Tf = (peak width) / 2 x (front’s half width)• All widths measured at 5% peak height.• Values greater than 1.5 generally indicate that unwanted interactions are occurring.

Page 76: HPLC

• Overloaded column.• Channels in the solid phase.

Peak Fronting

Page 77: HPLC

1. Partition2. Adsorption 3. Ion exchange and 4. Exclusion

Four retention mechanisms are applicable to HPLC:

Retention Mechanisms Applicable to HPLC

Page 78: HPLC

Which Methods of HPLC

Page 79: HPLC

Modes of HPLC

Normal-Phase Chromatography (NPC)

Reversed-Phase Chromatography (RPC)

Ion-Exchange Chromatography (IEC)

Size-Exclusion Chromatography (SEC)

Affinity chromatography

Chiral chromatography

Hydrophilic interaction chromatography (HILIC)

Electrochromatography:

Supercritical fluid chromatography (SFC)

Page 80: HPLC

80

Schematics showing the basis of separation (a) adsorption chromatography,(b) partition chromatography, (c) ion-exchange chromatography, (d) Size exclusion chromatography.

The solute represented by the solid circle (•) is the more strongly retained.

Page 81: HPLC

Silica gel (Adsorption Modes of HPLC)

Page 82: HPLC

Progress of a chromatographic separation

Page 83: HPLC

Graphical Representation of the Separation Process

(b) The component resides for preference in the stationary phase and the component more in the mobile phase .1b). Here k = 5/2 = 2:5 and k = 2/5 = 0:4.

(c) A new equilibrium follows the addition of fresh eluent: sample molecules in the mobile phase are partly adsorbed by the ‘naked’ stationary phase surface, in accordance with their distribution coefficients, whereas those molecules that have previously been adsorbed appear again in the mobile phase.

(d) After repeating this process many times, the two components are finally separated. The component prefers the mobile phase and migrates more quickly than the component, which tends to ‘stick’ in the stationary phase.

(a) A mixture of two components, and is applied to the chromatographic bed.

Page 84: HPLC

Interaction of Normal Phase

SiOH

SiOHHOSi

HO

HO

start

2 5 9 min0

Page 85: HPLC

Stationary Phase for Normal Phase of HPLC

Si-OH

Si-CH2-CH2-CH2-NH2

Si

Si-CH2-CH2-CH2-CN

Si

Si

Si Si-CH2-CH2-CH2-OCHCH2

OH

OH

Unmodified Silica

Amino

Cyano

Diol

Page 86: HPLC

Primary solvents (non-polar) – Hydrocarbons (Pentane, Hexane, Heptane, Octane)– Carbon tetrachloride– Aromatic Hydrocarbons (Benzene, Toluene, Xylene)– Methylene chloride– Chloroform

Secondary solvents (polar)

– Ethyl acetate, Acetone– 2-propaol, ethanol, methanol, Acetonitrile– THF, Methyl-t-butyl ether (MTBE), Diethyl ether, – Dioxane, Pyridine

Mobile Phase Phase for Normal Phase of HPLC

A primary solvent is used as mobile phase. Addition of secondary solvents is to adjust retention time.

Page 87: HPLC

In liquid-liquid partition chromatography, a solute distributes itself between two immiscible liquid phases in a manner analogous to the partition that occurs in liquid-liquid extraction.

One liquid phase (the stationary phase) forms a thin film over the surface of an inert support, while the other liquid phase (the mobile phase) passes over the first.

Thus, the retention mechanism is the partitioning of solutes between mobile liquid phase and stationary liquid phase.

Partition Modes of HPLC

Page 88: HPLC

Stationary Phase for Reversed Phase of HPLC

Page 89: HPLC

Stationary Phase for Reversed Phase of HPLC

Page 90: HPLC

Column for Reverse Phase

Si-CH2-CH2CH2-----CH3 (C18)

Si-CH2-CH2-CH2-CH2-CH2-CH2-CH2-CH3

Si

Si-CH2-CH2-CH2-CH3

Si

Si

Si Si-CH3

ODS

C8

C4

C1CH3

CH3

Stationary Phase for Reversed Phase of HPLC

Page 91: HPLC

Stationary Phase for Reversed Phase of HPLC

Si-CH2-CH2CH2NH2Si Amino

Si-CH2-CH2CH2CNSi Cyano

Si Si-CH2-CH2CH2- Phenyl

Page 92: HPLC

Endcapping Capping” of exposed silanols with short hydrocarbon chains after the primary bonding step

Stationary Phase for Reversed Phase of HPLC

Page 93: HPLC

• Water + Organic Solvents - When buffer is used, the concentration and pH

are important factors

- Methanol, Acetonitrile or THF are common organic solvents for RP HPLC

Mobile Phase for RP

Page 94: HPLC

OH

start

2 5 9 min0

HO

Mobile Phase for RP

Page 95: HPLC

Increase of Solvent Polarity

CH3

COOH

OHCH2

COOH

OH

CH3 CH2

COOH

OH

H2C

CH3 CH2

COOH

OH

H2C

CH2

CH3

Page 96: HPLC

Normal Phase vs. Reversed Phase

SiOH

SiOHHOSi

HO

HO

start

2 5 9 min0

OH

start

2 5 9 min

Str

on

g

0

HO

Page 97: HPLC

Interaction of analytes with stationary phaseMP vs. RP

Page 98: HPLC

Particle Shape

Most modern HPLC packings materials have spherical particles, but some are irregular in shape.

Irregular particles have larger surface areas, and are relatively inexpensive.

Spherical particles offer lower backpressure, and higher performance, stability, and reproducibility than irregular particles.

Page 99: HPLC

Smaller particle sizes give higher efficiency and higher resolution than larger particle sizes.

Larger particle sizes offer faster flow rates and lower back-pressure.

Typical particle sizes range from 3 mm to 20 mm, and new 1.5 mm particle sizes are available to maximize resolution on short columns. A 5 mm particle size represents the best compromise between efficiency and back-pressure.

Particle Shape

Page 100: HPLC

For silica-based reversed-phase packings, carbon load indicates the amount of functional bonded phase attached to the base material.

Phases with lower carbon loads are more weakly hydrophobic, which may significantly reduce retention times over phases with higher carbon loads.

However, a higher carbon load will give higher capacity and often greater resolution, especially for compounds of similar hydrophobicity.

Particle Shape

Page 101: HPLC

Effect on chromatography

High surface area generally provides greater retention, capacity and resolution for separating complex, multi-component samples.

Low surface area packings generally equilibrate quickly, especially important in gradient analyses.

Surface Area

Page 102: HPLC

Larger pores allow larger solute molecules to be retained longer through maximum exposure to the surface area of the particles.

Choose a pore size of 150Å or less for sample MW 2000. Choose a pore size of 300Å or greater for sample MW > 2000.

Pore Size

Effect on chromatography

Page 103: HPLC

A Note

The larger the pore diameter, the smaller the surface area.

The larger the surface area the greater the retention.

The smaller the pore diameter the greater the steric hindrance effect.

Page 104: HPLC

• The main parameter affecting pH stability of packing material is Bonding Density

• Low pH (<2.5) causes hydrolysis of the siloxane bonds destroying bonded layer

– The higher the bonding density the lower hydrolysis effect.

• High pH (>8.5) causes silica dissolution– High bonding density shield silica surface which makes it stable

up to pH 13.

pH stability

Page 105: HPLC

Endcapping reduces peak-tailing of polar solutes that interact excessively with the otherwise exposed, mostly acidic silanols.

Non-endcapped packing provide a different selectivity than do endcapped packing, especially for such polar samples.

Endcapping

Effect on chromatography

Page 106: HPLC
Page 107: HPLC

HPLC Method Development

Page 108: HPLC

Introduction

• Common Mistakes in Method Development:• Inadequate Formulation of Method Goals• Little Knowledge of Chemistry of Analyte Mixture• Use of the First Reversed Phase C18 Column Available• Trial and Error with Different Columns and Mobile Phases

• These Mistakes Result In:• Laborious, Time-consuming Development Projects• Methods that Fail to Meet the Needs of the Analyst

Page 109: HPLC

HPLC Method Development - A Proposed Procedure

At Your Desk• Define your knowledge of the sample• Define your goals for the separation method• Choose the columns to be considered

In the Laboratory• Choose the initial mobile phase chemistry• Choose the detector type and starting parameters• Evaluate the potential columns for the sample• Optimize the separation conditions (isocratic or gradient) for the

chosen column• Validate the method for release to routine laboratories

Page 110: HPLC

Choosing the Appropriate HPLC Column Should Be Based Both Upon Knowledge of the

Sample and Goals for the Separation

Benefits of this Approach Include:• Small initial time investment• Big time savings in the HPLC laboratory• More “informed” approach to column selection• More efficient than “trial and error” approach

Page 111: HPLC

Knowledge of the Sample Influences the Choice of Column Bonded Phase Characteristics

• Knowledge of the Sample

• Structure of sample components?• Number of compounds present?• Sample matrix?• pKa values of sample components?• Concentration range?• Molecular weight range?• Solubility?• Other pertinent data?

Column Chemistry(bonded phase, bonding type, endcapping, carbon load)}

Page 112: HPLC

Goals for the Separation Influence the Choice of Column Particle Physical Characteristics

• Goals for the Separation• Max. resolution of all components?• Partial resolution?• Fast analysis?• Economy (low solvent usage)?• Column stability and lifetime?• Preparative method?• High sensitivity?• Other goals?

Column Physics(particle bed dimensions, particle shape, particle size, surface area, pore size)}

Page 113: HPLC

Column Selection Chart

Meth

od

G

oals

Hig

h E

fficie

ncy

Hig

h C

ap

acit

y

Low

B

ackp

ressu

re

Hig

h R

esolu

tion

Hig

h S

am

ple

Load

ab

ilit

y

Su

itab

le f

or

MW

>

2000

Hig

h S

tab

ilit

y

Hig

h S

en

sit

ivit

y

Fast

An

aly

sis

Low

Mob

ile

Ph

ase

Con

su

mp

tion

Sta

bilit

y a

t p

H

Extr

em

es

Fast

Eq

ilib

rati

on

Defa

ult

Colu

mn

(Good

for

most

Ap

plicati

on

s)

Particle Sizesmall (3µm) • •medium (5µm) •large (10µm) •

Column Lengthshort (30mm) • • • • •medium (150mm) •long (300mm) •

Column IDnarrow (2.1mm) • •medium (4.6mm) •wide (22.5mm) •

Surface Arealow (200m2/g) • • •high (300m2/g) • • •

Pore Sizesmall (60Å) • •medium (100Å) •large (300Å) •

Carbon Loadlow (3%) •medium (10%) •high (20%) • • •

Bonding Type monomeric • •polymeric • • • •

Particle Shapespherical • • • •irregular •

Page 114: HPLC

Choosing the Bonded Phase• Draw the molecular structures for all known components of the

mixture. Identify the two compounds whose structures are the most similar.

• e.g.:

O

OH

O

OH O H H

O

OH

O

O O

Prednisolone Prednisone

Page 115: HPLC

• For these two molecules, circle the structural features that differ. It is these differences that should be exploited to optimize the separation.

• e.g.:

Choosing the Bonded Phase

Prednisolone Prednisone

O

OH

O

OH O H H

O

OH

O

O O

Page 116: HPLC

Choosing the Bonded Phase• Use the results of the structural comparison to select a bonded phase

showing optimal selectivity for these two molecules. In this case consider using a silica column (no bonded phase) for its ability to retain polar solutes through hydrogen bonding.

O

OH

O

OH O H H

O

OH

O

O O

Prednisolone Prednisone

Page 117: HPLC

Functional Group Polarity Comparisons

Polarity Functional Group Structure Bonding Types Intermolecular Forces Displayed

Low Methylene s London

Phenyl s , p London

Halide s London, Dipole-Dipole

Ether s London, Dipole-Dipole, H-bonding

Nitro s , p London, Dipole-Dipole, H-bonding

Ester s , p London, Dipole-Dipole, H-bonding

Aldehyde s , p London, Dipole-Dipole, H-bonding

Ketone s , p London, Dipole-Dipole, H-bonding

Amino s , p London, Dipole-Dipole, H-bonding, Acid-base chemistry

Hydroxyl s London, Dipole-Dipole, H-bonding

High Carboxylic Acid s , p London, Dipole-Dipole, H-bonding, Acid-base chemistry

R (CH2)2

R

R F, Cl, Br, I

R O

R

N+

O

O-

R

O

O RR

O

R H

O

RR

R NH2

R OH

O

OHR

Page 118: HPLC

Choosing the Bonded Phase

• C18 or Octadecylsilane (ODS)• Very nonpolar - Retention is based on London (dispersion) interactions

with hydrophobic compounds. • Example : Alltima™ C18

Examples of bonded phases used for HPLC packing media:

Si

R

R

(CH2)17CH3

Page 119: HPLC

Choosing the Bonded Phase• Phenyl• Nonpolar - Retention is a mixed mechanism of hydrophobic and p - p

interactions. • Example : Platinum™ Phenyl

Si

R

R

(CH2)3 C

C

C

C

C

C

H H

H

HH

Page 120: HPLC

Choosing the Bonded Phase• Cyanopropyl• Intermediate polarity - Retention is a mixed mechanism of

hydrophobic, dipole-dipole, and p - p interactions.• Example : Alltima™ CN

Si

R

R

(CH2)3 C N

Page 121: HPLC

Choosing the Bonded Phase• Each bonded phase has unique selectivity for certain sample types.

As a practical example, to separate toluene and ethyl benzene:• Note a difference of one -CH2- unit

• Choose a C18 bonded phase for retention by hydrophobicity• Maximize hydrophobic selectivity with a high silica surface area, high

carbon load material like Alltima C18

Toluene Ethyl Benzene

Page 122: HPLC

Choosing the Particle Physical Characteristics

Use the Column Selection Chart• Use “default” column as starting point• Match up method goals with individual particle physical

characteristics• Change only those particle parameters that affect the method

goals• Recognize the “optimum” column as a possible compromiseExample: Sample Type: hydrophobic compounds Method Goal: highest resolution

Page 123: HPLC

Choosing the Particle Physical Characteristics

Column Selection Chart

Default Column Optimum Column†

Column Bed Dimensions 150 x 4.6mm 250 x 4.6mmParticle Size 5µm 3* or 5µmSurface Area 200m2/g >200m2/g Pore Size 100Å 100ÅCarbon Load 10% 16 - 20%Bonding Type Monomeric Mono- or PolymericBase Material Silica SilicaParticle Shape Spherical Spherical

* mobile phase backpressure may be excessive

† Optimum Column: Alltima C18™, 5µm, 250 x 4.6mm

*Note that the choice may represent a compromise. Here, the “optimum” column for resolution sacrifices speed.

Example: Sample Type: hydrophobic compounds Method Goal: highest resolution

Page 124: HPLC

Choosing the Particle Physical Characteristics

Column Dimensions • Length and internal diameter of packing bed

Particle Shape• Spherical or irregular

Particle Size • The average particle diameter, typically 3-20µm

Surface Area • Sum of particle outer surface and interior pore surface, in

m2/gram

Page 125: HPLC

Pore Size • Average size of pores or cavities in particles, ranging from

60-10,000Å

Bonding Type • Monomeric - single-point attachment of bonded phase molecule• Polymeric - multi-point attachment of bonded phase molecule

Carbon Load • Amount of bonded phase attached to base material, expressed as %C

Endcapping • “Capping” of exposed silanols with short hydrocarbon chains after the

primary bonding step

Choosing the Particle Physical Characteristics

Page 126: HPLC

Column DimensionsEffect on chromatography

Column Dimension• Short (30-50mm) - short run times, low backpressure• Long (250-300mm) - higher resolution, long run times• Narrow ( 2.1mm) - higher detector sensitivity• Wide (10-22mm) - high sample loading

Page 127: HPLC

Particle Shape• Effect on chromatography • Spherical particles offer reduced back pressures and longer column

life when using viscous mobile phases like 50:50 MeOH:H2O.

Page 128: HPLC

Particle Size• Effect on chromatography • Smaller particles offer higher efficiency, but also cause higher

backpressure. Choose 3µm particles for resolving complex, multi-component samples. Otherwise, choose 5 or 10µm packings.

Page 129: HPLC

Surface Area• Effect on chromatography • High surface area generally provides greater retention, capacity and

resolution for separating complex, multi-component samples. Low surface area packings generally equilibrate quickly, especially important in gradient analyses.

• High surface area silicas are used in Alltima™, Adsorbospherel® HS, and Adsorbosphere® UHS packings. Low surface area silicas are used in Alltech’s Platinum™, Econosphere™, and Brava™ packings.

Page 130: HPLC

Pore Size• Effect on chromatography • Larger pores allow larger solute molecules to be retained longer

through maximum exposure to the surface area of the particles. Choose a pore size of 150Å or less for sample MW 2000. Choose a pore size of 300Å or greater for sample MW > 2000.

Page 131: HPLC

Bonding Type• Effect on chromatography • Monomeric bonding offers increased mass transfer rates, higher column efficiency,

and faster column equilibration.

Si

R

R

(CH2)17CH3Si

CH3

CH3

(CH2)17CH3XOH +monomeric

bonding

Si

CH3

X

(CH2)17CH3X+polymeric

bonding

OH

OH O

O

Si

CH3

(CH2)17CH3

Polymeric bonding offers increased column stability, particularly when highly aqueous mobile phases are used. Polymeric bonding also enables the column to accept higher sample loading.

Page 132: HPLC

Carbon Load• Effect on chromatography • Higher carbon loads generally offer greater resolution and longer

run times. Low carbon loads shorten run times and many show a different selectivity.

Page 133: HPLC

EndcappingEffect on chromatography • Endcapping reduces peak-tailing of polar solutes that interact excessively

with the otherwise exposed, mostly acidic silanols. Non-endcapped packings provide a different selectivity than do endcapped packings, especially for such polar samples.

• Platinum™ EPS packings are non-endcapped to offer enhanced polar selectivity.

Page 134: HPLC

Conclusion

• In this approach to HPLC column selection, the bonded phase chemistry of the column is chosen on the basis of an analysis of the sample component structures.

• The physics of the column is chosen according to an analysis of the goals for the separation method.

• This approach succeeds in predicting unique, optimum bonded phase chemistries and particle bed physical characteristics that are likely to meet the goals for the separation method.

Page 135: HPLC

Column Selection Example #1

What goals do I have for the method?Maximum resolution of all components?Best Peak Shape for difficult samples? Fast analysis? Economy (low solvent consumption)? Column stability-long lifetime? Purify one or more unknown components for characterization?High sample loadability?High sensitivity?…Other (High Sample Throughput--Quick Equilibration)

Number of compounds present 4 Sample matrix --pKa values of compounds? --UV spectral information about compounds? Concentration range of compounds Molecular weight range of compounds 94 - 323

What do I know about the sample?

Page 136: HPLC

Column Selection Example #1

Structures of Compounds

OH

(CH2)5CH3

N

(CH2)3CH3

Phenol 3-Butylpyridine

Anthracene 3-Hexylanthracene

Page 137: HPLC

Column Selection Example #1

Which two sample components have the most similar structures? Draw them, then circle the structural differences between them.

Normal phase silica NH2 CNReversed phase C18 C8 Ph CN

(CH2)5CH3

Anthracene

3-Hexylanthracene

Note: The structural difference between these two compounds is the hydrophobic hexyl side chain. This suggests a non-polar C18 or C8 column would interact with this area of difference to help provide separation of these two compounds.

Recommended bonded phase (silica based materials only) – mark one

Page 138: HPLC

Column Selection Example #1

Column physical characteristics – use Column Selection Chart and Method Goals

Default Column Ideal Column Column bed dimensions (mm) 150 x 4.6 100 x 2.1Particle Size (µm) 5 5Surface area (m2/g) 200 <200Pore Size (Å) 100 100Carbon Load (%) 10 10Bonding type Monomeric Monomeric Particle shape spherical spherical

Page 139: HPLC

Available packing alternatives meeting the above criteria:

Packing Base Particle Particle Carbon Pore Surface Bonding End- Material Shape Size Load Size Area Type cap’d

(µm) (%) (Å) (m2/g)

silica Sph. 3, 5, 10 12 80 220 Mono. Yes

silica Sph. 3, 5 8.5 145 185 Mono. Yes

silica Sph. 3, 5, 10 10 80 200 Mono. Yes

silica Sph. 3, 5, 10 6 100 200 Mono. Yes

Column Selection Example #1

Allsphere ODS-2

Brava BDS C18

Econosphere C18

Platinum C18

Page 140: HPLC

Available packing alternatives meeting the above criteria:

Packing Base Particle Particle Carbon Pore Surface Bonding End- Material Shape Size Load Size Area Type cap’d

(µm) (%) (Å) (m2/g)

silica Sph. 3, 5, 10 12 80 220 Mono. Yes

silica Sph. 3, 5 8.5 145 185 Mono. Yes

silica Sph. 3, 5, 10 10 80 200 Mono. Yes

silica Sph. 3, 5, 10 6 100 200 Mono. Yes

Column of choice: Brava BDS C18, 100x2.1, 5µm (Spherical , 185m2/g, monomeric)

Column Selection Example #1

Increased Sensitivity, Low Solvent Consumption,

Fast AnalysisQuick Equilibration

Good balance of efficiency & backpressure

Reduced backpressure

Best Peak Shape

Allsphere ODS-2

Brava BDS C18

Econosphere C18

Platinum C18

Page 141: HPLC

Column Selection Example #2

What goals do I have for the method?Maximum resolution of all components? Partial resolution, resolving only select components?Fast analysis?Economy (low solvent consumption)?Column stability-long lifetime? Purify one or more unknown components for characterization?High sample loadability?High sensitivity? …Other

Number of compounds present 6+Sample matrix --pKa values of compounds? --UV spectral information about compounds? UV -254Concentration range of compounds --Molecular weight range of compounds 349 - 645

What do I know about the sample?

Page 142: HPLC

Column Selection Example #2

Structures of Compounds

N

NH S

O

O

O

NH2

OH

HH

NN

NH SS

N

O

O

OO

O

NH2

OH

O

HH

N

S

N

S

NN

N

NN

NH

OS

O

O

OH

HHN

NHS

N

OO

O

O O

ONH2OH

OH H

N

NN

N N

NNH

SNH

N

S

O

OO

O O

OOH

OH

HH

N

SNH

OS

O

O

O

O

OH

HH

Page 143: HPLC

Column Selection Example #2

Which two sample components have the most similar structures? Draw them, then circle the structural differences between them.

Notes: both structures very polar, with amine and pi bond functions--a RP CN column may give good separation by mixed-mode retention of hydrophobic, CN---H---NR2 hydrogen bonding and - interactions with double bonds.

Normal phase silica NH2 CN Reversed phase C18 C8 Ph CN

N

NH S

O

O

O

NH2

OH

HH

N

SNH

OS

O

O

O

O

OH

HH

Recommended bonded phase (silica based materials only) – mark one

Page 144: HPLC

Column Selection Example #2

Column physical characteristics – use Column Selection Chart and Method Goals

Default Column Ideal Column Column bed dimensions (mm) 150 x 4.6 250 x 2.1Particle Size (µm) 5 5Surface area (m2/g) 200 200 +Pore Size (Å) 100 Not criticalCarbon Load (%) 10 --Bonding type Monomeric PolymericParticle shape spherical spherical

Page 145: HPLC

Column Selection Example #2

Available packing alternatives meeting the above criteria:

Packing Base Particle Particle Carbon Pore Surface Bonding End-Material Shape Size Load Size Area Type cap’d

(µm) (%) (Å) (m2/g)

Adsorbosil CN silica Irreg. 5, 10 -- 60 450 Poly. Yes

Alltima CN silica Sph. 3, 5 -- 100 350 Poly. Yes

Allsphere CN silica Sph. 3, 5, 10 -- 80 220 Mono. No

Platinum CN silica Sph. 3, 5, 10 -- 100 200 Mono. No

Column of choice: Alltima CN, 250 x 2.1 , 5µm ( Spherical , 350 m2/g , polymeric)

High resolution,High sensitivity

High res.

Good balance of efficiency & backpressure

Reduced backpressure

Robust