HP 7673B - S service manual

248
" B Troubleshooting Section B-Contents Overview of Failure Indicators - B.S Basic Strategy of Repair : B-4 'l'ypes of Failures B-4 lIijector Errors B.6 'I'ray Errors ................... ............................. B-12 Controller Errors ............. •• .............................. B.-22 Miscellaneous Errors B-24 HP-IB/Controller Board Errors B-27 Printed Error Messages B-27 8-1

description

Trouble shooting and service manual for 7673B Autosampler by HP (Agilent)

Transcript of HP 7673B - S service manual

Page 1: HP 7673B - S service manual

"

BTroubleshooting

Section B-Contents

Overview of Failure Indicators - B.S

Basic Strategy of Repair : B-4'l'ypes of Failures B-4lIijector Errors B.6

'I'ray Errors ...................•............................. B-12Controller Errors .............••.............................. B.-22

Miscellaneous Errors B-24HP-IB/Controller Board Errors B-27Printed Error Messages B-27

8-1

Page 2: HP 7673B - S service manual

i ~'.I7·n;:'~}~.\·,; ',f,.'ft,' '" .":, '!..(~~~~::",\ ;,~'J:;t.~',,',I~'~·

,":""

This section is intended to aid a trained service engineer in the troubleshooting process.Included in this chapter is an overview of the failure indicators, a basic strategy of repair,and a detailed action list for each type offailure.

WAHNING

HAZARDOUS VOLTAGES ARE PRESENT IN THE ELECTRICAL PORTIONS OFTHIS INSTRUMENT WHEN THE POWER CORD IS CONNECTED•. TO AVOID APOTENTIALLY DANGEROUS ELECTRICAL SHOCK, BE CERTAIN THE POWERCORD IS UNPLUGGED FROM ITS POWER SOURCE BEFORE REMOVING ANYACCESS PANELS.

WAHNING

MEASUREMENTS AND/OR TESTS THAT NEED TO BE MADE ON ELECTRICALLYENERGIZED PORTIONS OF THE INSTRUMENT SHOULD BE PERFORM!D ONLYBY SERVICE-TRAINED PERSONNEL WHO ARE AWARE OF ALL INVOLVEDHAZARDS.

B-2

Page 3: HP 7673B - S service manual

~' -'", "

Overview of Failure Indicators

Failures are indicated by the light patterns of the three LEOs on the front of the inJectormodule and the two LEOs on the front of the controller module (Figure B-1).

When the sampler is turned on, the two green LEOs should be on and the two red LEOsshould be off. The yellow LED should only be on during a run.

Ifone or both modules are not displaying the proper LED pattern, look at the chart onpage fl..5 for the LED lighting pattern and refer to the corresponding page in this section.

If the sampler is controlled by an external device such as an HP integrator or an HPChemStation, the sampler will send a printed error message to the external device.

-

Red

Yelow

Green

Figure 11·1. Location of Sampler LED••

Page 4: HP 7673B - S service manual

Basic Strategy of Repair

The troubleshooting strategy is to identify the failure, isolate the defective part, and replaceit.

Carefully observe the operation of the falling module. Since the operations are so inte­grated, much can be learned by observing not only what is nonoperational, but also by whatareas are working correctly. The sequence of normal functions for each module is describedin other sections of this manual.

For further help in identifYingthe problem, there are diagnostics modes for both the it\iec­tor and tray modules. These are described at the end of sections C and E.

Types of Failures

Throurhout this section, the following symbols identify the LEOs.

Injector

00 ControUer

I Dol0Immn1W.LL11J

Some error conditions can have more than one cause. The most frequent/simplest cause islisted first, with other possibilities listed second or third. Ifpossible, use the diagnosticmode to isolate the problem to one of the causes and follow the steps for troubleshootingthat cause.

Eaoh action list contains several things for you to check. Go through each step in order,and stop when you have found the failure. You may not necessarily complete ell the actionsin the list. After you have repaired the failure, retest by rpnning a normal sequence ofmovement or using the diagnostics mode.

8·4

Page 5: HP 7673B - S service manual

B-5

PageB-22"8-22B-22B-23

PageB-24B-25B-26

PlgeB-12B-13B-15B-17B-19B-21B-21

PageB-6B-7B-8B-9B-11B-11

DllcrlpUon

Injector Processor Error18~Volt Supply ErrorPower Supply Error

DllcrlpUon

Tray Not ConnectedVIal LosVLeftln GripperTheta Axis ErrorRadial Axis ErrorGripper AxIs ErrorTray Processor ErrorBottle Not Found

D"crlpUon

ControDer Processor ErrorController RAM ErrorConlroler ROM ErrorConfiguration Error

DllcrlpUon

Safety ErrorInjector Syringe ErrorInjector Turret ErrorInjector Plunger ErrorIncomplete Injection ErrorInjector Memory Error

Controller L'd

Green OnRedanNone

CONTROLLER ERRORS

Controller Led

Red 2Double RashesRed 3 Double RashesRed 4Double RashesRed 5 Double Rashes

MISCELLANEOUS ERRORS

InJector Led

Red, Yenow, Green OnRed, Yellow, Green OnNone

Table B·1. LED Error Code.

* The HP.JB/controler board also uses Algie red 1IaIhn toCOIIVTUlIcIU error cordIons flllJld durilgIs powllr-up seIf-test. See-Hp..fB/Cortrolllr BOlI"d Errors.-

TRAY ERRORS

Controller Led

Red Steady OnRed 2Single Rashes*Red 3 Single Flashes*Red 4Single Rashes*Red 5 Single Aashes*Red 6 Single Aashes*

. Red 7 Single Aashes*

INJECTOR ERRORS

InJector Led

Red Steady OnRed 2RashesRed 3 RashesRed 4 RashesRed 5RashesRed 6Rashes

Page 6: HP 7673B - S service manual

," ';

Injector Errors

•0 Red LED: On Steady

0-IDmn

Injector Door Open

1. Close injector door.

2. VerifY door switch is working properly.

Safety Error

• Injector door open.• Injector improperly mounted.

3. Press on door at switch. If this clears the problem, replace the front housing.

4. Check the switch connection at the injector logicboard.

6. Replace door switch ifneeesaary.

Injector Improperly Mounted

1. Remount the injector.

2. Ensure the mounting post is making contact with the mounting switch

8. Check the switch connection at the injector logicboard.

4. Replace the mounting switch if necessary.

8·8

Page 7: HP 7673B - S service manual

v.a Red LED: 2Flashes

~IIIIIll

Injector Syringe Error

• Syringe carriage movementobstructed.

• Syringe motor inoperable.• Syringe carriage sensor

inoperable.

Using the diagnostics mode, determine if the syringe carriage is moving properly. If it is,suspect sensor failure; otherwise, check for obstructions Of motor failure.

Syringe Carriage Movement Obstructed

1. Check to see if the syringe carriage is moving freely on the guide rods. Ifnot, removethe syringe carriage and clean the rods and/or remove any obstructions.

2. With the syrinp carriage removed, ensure that the syringe motor/drive belt combina­tion is moving freely. If not, remove the belt and check movement of both the idlerpulley and the motor. If either piece is not moving, replace it.

Syr,nge Motor Inoperable

1. Using the diagnostics mode, determine if the syringe motor is rotating. Ifnot, checkthe PC board connection for the motor. Rerun the test, observing the LEOs that indi­cate the motor windings are energized.

2. Remove the motor and test it for opens and shorts. Replace motor if necessary.

8. Check for the presence of 18-V supplies at the iI\jector logic board. If they are not atthe proper voltage levels, proceed to diagnose the controller module and cables.

4. Although the LEDs indicate the motor windings are energized, they do not verifY thatthe proper drive current is being supplied to the windings. Replace inJector logic boardand retest.

Syringe Carriage Sensor Inoperable

1. Start diagnostics and move the syringe carriage to the home position. Check that thegreen LED of the irdector module is lit.

2. Check the syringe carriage for a magnet. Using a known good magnet, test to see if themagnet in the carriage is installed properly (magnet faces should repel each other). Ifthe magnet is missing or installed improperly, replace the syringe carriage.

8. VerifYthe syringe sensor is connected properly to the i~ector logic board. Retest thesensor with the syringe carriage in the home position. Ifnecessary, replace the syringecarriage sensor (note that thls in one of a pair of sensors and both must be replaced).

Page 8: HP 7673B - S service manual

Ire1\'

0 Red LED: 3 FlashesO·

IITIIID

Injector Turret Error

• Turret movement obstructed.• Turret motor inoperable.• Turret sensor inoperable.

Using the diagnostic modes, determine if injector turret is moving properly. If it is, suspectsensor failure; otherwise, check for obstructions or motor failure.

Turret Movement Obstructed

1. Manually check to see if the injector turret is able to rotate freely through a full revolu­tion. If it is not rotating freely, continue with next step.

2. Open the uyector door and check the bottle stripper assembly. It should swinr freely inand out. Ifnot, remove the turret and check the arm for interference. Replace the bot­tle stripper assembly, spring or mounting sleeve if necessary.

Turret Motor Inoperable

1. U8ing diagnostics mode, determine if the turret motor is rotating freely and not lockingat some positions. Replace motor if necessary.

2. Check the connection of the turret motor to the injector logic board and retestoperation.

8. Check the irUector turret motor for shorts or opens. Replace motor if necessary.

4. Replace irUector board and retest.

Turret Sensor Inoperable

1. Check the hijector turret for a magnet. Using a known good magnet, test to see if themagnet is installed properly (magnet faces should repel each other). Replace the turretif necessary.

2. Manually verify that the sensor is working, using the diagnostics mode. The green LEDshould light when the sensor is tripped.

3. VerifYthat the turret sensor is connected properly to the injector logic board. Ifneces­sary, replace the sensor (note that this is one of a pair of sensors and both must bereplaced). .

8 ..8

Page 9: HP 7673B - S service manual

~.

0 Red LED: 4 Flashes

~mnlD

Injector Plunier Error

• Plunger movement obstructed.• Plunger motor inoperable.• Plunger carriage sensors

inoperable.

Using the diagnostic mode, determine if the plunger carrier is moving properly. If it is not,suspect sensor failure; otherwise, check for obstructions or motor failure.

Plunger Movement Obstructed

1. Remove syringe and check it for plunger stickiness or binding. Replace syringe ifnecessary,

2. Check the viscosity of the sample. Extremely viscous samples may have to be dilutedwith the proper solvent or heated using the tray's circulating water bath to allow propermovement of the plunger.

8. With the syringe removed, run the diagnostics on the plunger carrier. Look for inter­ference between the syringe carriage and the plunger carriage. Replace the syringecarriage or plunger carriage as necessary,

4. Check the belt tension on the plunger carriage. If it is too tight or too loose, adjust thetension by adjusting the position of the plunger motor wedge.

6. Check the solenoid lever arm and return spring to verify proper installation andoperation of the stop mechanism.

Plunger Motor Inoperable

1. Check to see if the motor is connected to the sensor board.

2. Check the ribbon cable connections at the sensor board and the iIUector logic board. Ifconnections are good, check the ribbon cable for open runs. Replace ifnecessary.

8. Check the DC motor operation using the diagnostic mode. If it not rotating, remove themotor and check to see if the pulley is loose. Replace the motor if necessary.

4. Check the motor winding resistance. If the motor is either open or shorted, replace it.

5. Replace the injector logic board and retest.

B-9

Page 10: HP 7673B - S service manual

Plunger Carriage 'sensorts) Inoperable

1. Start diagnostic mode and manually move the plunger carriage to the top of itsmovement. Check that the green LED is lit. Move the plunger to the bottom of themovement and check the green LED again.

2. With a known good magnet (faces should repel), test the plunger carrier magnet. If it isnot good, replace the plunger carrier,

3. Check the ribbon cable connections at the sensor board and the iIijector loric board. Ifconnections are good, check the ribbon cable for open runs. Replace if necesBaIy.

4. Replace the plunger sensor board by replacing the syringe carriage and retest.

5. Replace the injector logic board and retest.

8-10

Page 11: HP 7673B - S service manual

•a Red LED: 5 Flashes

~llnnD

Incomplete lDJection Error• Plunger not sensed at bottom of

movement during injection.

1. Remove the syringe from the injector end check it for stickiness and/or binding.Replace ifnecessary.

2. Run the diagnostics mode for the plunger carrier. Replace plunger carrier and/orsyringe carriage if necessary.

8. Replace ixijector logic board end retest.

II.a Red LED: 6 Flashes

aIITITnL....-

lDJector Memory Error• Processor on injector board is

malfunctioning.

1. Verify that the injector processor (and the EPROM if present) is positioned properlyand seated in the PC board socket.

2. Replace the injector processor (and/or EPROM) if necessary.

8. Replace the injector logic board if necessary.

8·11

Page 12: HP 7673B - S service manual

Tray Errors

_01 Red LED: On Steady

Tray Cable Not Installed

Tray Not Connected• Tray cable not installed.• Tray board or controller board

inoperable.

1. Verify that the tray cable is plugged into the controller module and the fastening screwsare seated.

2. VerifY that the connection of the cable inside the tray is secure.

8. Check to see if the tray cable has opens or shorts.

Tray Board or Controller Board Inoperable

1. Verify that the tray board and controller board are properly installed in the controllermodule and the connections are fully seated.

2. Replace the tray board and then the controller board ifnecessary.

Page 13: HP 7673B - S service manual

(_) 01 Red LED: 2 Single Ras""

Tray Failed to Deliver Vial

Vial LostlLeft in Gripper

• Tray failed to deliver vial.• Vial switch ofT when er!pper is

empty.• Gripper dropped vial.

1. If there is a vial in the gripper, remove it and return the vial to its position in the tray.

2. Check the "deliver to" location for the vial and verifY that the location is empty and freeof obstructions.

S. Ensure that the tray is installed correctly on the mounting bracket.

4. Rerun the system to observe how the eITOr may have occurred.

VI~I Switch Off When Gripper Is Empty

1. Remove the cover from the controller module and turn on the power. Make sure thegripper is in the home position and the Z.LED is lit. Insert a bottle into the gripper andcheck that the B--LED goes off. Remove the bottle and check that the BoLED goes onagain. If the BoLED is not lighting properly, continue.

2. Check the tray cable connection inside the bottom cover of the tray. Make sure theconnector is seated.

3. Remove the tray ann from the tray turret. Remove the tray arm covers and check thegripper assembly. The lever should be holding a vial magnet and it should be aligned

.with the lower magnetic sensor on the sensor board. Press the gripper paddle. If thevial magnet does not move away from the sensor, replace the gripper assembly.

4. Ensure that the connections on both the R~flex and the theta-flex cables are installedproperly.

6. Replace the ann sensor board and retest.

6. Replace the tray board and retest.

Page 14: HP 7673B - S service manual

Gripper Dropped Vial

1. Remove any obstructions that may have dislodged the vial from the gripper during trayarm movement.

2, Make sure sample vials meet HP specifications (see HP 7673 Operating Manual Appen­dix A, Consumables). If the vials are too tall, the vial in the gripper may be dislodged bythe ones in the traY.

3. Check the gripper jaws for smooth spring action movement. If either gripper jaw doesnot close properly, replace gripper assembly.

Page 15: HP 7673B - S service manual

I <_) C11 Red LED: 3Single Rash"

Theta Axis Error

• Tray rotation obstructed.• Theta-motor inoperable.• Theta-sensor inoperable.

Using diagnostic mode, determine if the tray arm is rotating properly. IT the arm movessmoothly or if the theta motor fails to stop when the arm does, suspect sensor failure.Otherwise, check for obstructions or motor failure.

Tray Rotation Obstructed

1. Remove any obstructions from the tray which may keep the arm from moving in its fullrotation.

2. Slowlyrotate the arm about its angular axis. It should move smoothly-do not forcethe arm. The arm should move approximately 856 degrees from one stop to the other.

8. Remove the tray turret from the base. Check the theta-motor gear and the turretgear for missing or broken teeth. Replace the motor assembly and/or turret assemblyas needed.

I CAUTION IImprop.r mounting of turr.t ....mbly will damage the th.ta fllx elreutt. Seef1gur. E·S for ••••mbly In.tructlon•.

Theta Motor Inoperable

1. Remove the bottom cover and unplug the theta motor. Check the winding resistanceof the theta-motor. Replace motor ifnecessary.

2. Check the theta-motor gear for looseness. Replace the motor assembly if necessary.

8. Make sure that the screws holding the motor in place are present and tight. IT thescrews are loose, reposition the motor using the motor locating tool and tighten thescrews.

4. VerifY the continuity of the tray cable lines which carty the theta-motor signals. Checkthe connections of the tray cable to the tray board.

5. Replace the tray board ifnecessary.

8-15

Page 16: HP 7673B - S service manual

Theta Sensor Inoperable

1. Using the diagnostics mode, check the theta axis movement. Watch the T-LED on thetray board. It should be lit for half of the rotation and ofT for the other half.

2. Check the theta-sensor connection at the termination board. Check the tray cableconnections at both ends.

3. Remove the tray turret from the base and replace the theta-sensor.. Before reassem­bling, test the operation of the theta-sensor by interrupting the sensor with a thinobject.

4. Replace the tray board and retest.

B-16

Page 17: HP 7673B - S service manual

<_) 01 Rod LED: 4SJng~ Rash..

Radial Axil Error

• Tray radial motion obstructed.• R motor inoperable.• Radial sensor inoperable.

Using diagnostic mode, determine if the tra,y arm is moving forward and backward properly.If the ann moves smoothly or if the R~motor falls to stop when the arm does, suspect sensorfailure; otherwise, check for obstructions or motor failure.

Tray Radial Motion Obstructed

1. Remove any objects that obstruct the movement of the tray arm.

2. Remove the cover from the controller module and turn the power on. Hold the arm inthe middle and gently move the arm in its radial axis to the full back (home) position.Do not force the arm to move. The R~LED should be lit.

8. Check arm rods and bushings for binding. Replace ifnecessary.

4. Open tray base and remove the large R~drive gear. Move the arm radially-the armshould move smoothly and freely.

6. Check the R~motorpinion gear and the ann rack for missing or broken teeth. Checkthat the pinion is free to turn in the turret. Replace as needed.

R Motor Inoperable

1. Remove the bottom cover from the tray and unplur the radial motor. Check theresistance of the motor windings. Replace the motor if any windings are open orshorted.

2. Check the radial motor gear for looseness. Replace the motor if necessary.

8. Check the tray cable for continuity and proper connections.

4. Replace tra,y board and retest.

Page 18: HP 7673B - S service manual

Radial Sensor Inoperable

1. Start the diagnostics mode. The R-LED on the tray board should be lit when the arm isall the wayback (home position).

2. Gently move the arm forward and check that the R-LED goes ofT.

8. Check the tray cable connections at both ends.

4. Remove the arm assembly and verifY the presence and proper installation of the magnetin the turret, Check the continuity of the sensor lines through the R-flex circuit. Checkthe connection of the R-flex circuit to the theta-flex circuit on the turret.

5. With the arm removed, verify the continuity of the radial axis sensor's lines through thetheta-flex circuit and the tray cable. Replace the tray cable or the theta-flex cable ifnecessary.

6. Replace arm sensor board and retest.

7. Replace the tray board and retest.

B-18

Page 19: HP 7673B - S service manual

(_> 01 Red LED: 5Singi. Flash"

I,~:'!"r''''''.'!'

Gripper Axle Error

• Gripper (Z) motion obstructed.• Z-motor inoperable.• Z-sensor inoperable.• Tray firmware incompatibility.

Using the diagnostic mode, determine if the gripper is moving up and down properly. If thegripper moves smoothly or if the Z-motor fails to stop when the gripper does, suspect sensorfailure; otherwise, check for obstructions or motor failure.

Gripper (Z) Motion Obstructed

1. Remove the tray arm and the right arm cover and verify free movement of the gripperand drive assembly by moving the drive belt manually.

2. Check the belt tensioner assembly to ensure that the Z-drive belt has adequate tension.

S. Check the gripper drive for obstructions.

4. Remove the belt from the drive pulley and verify that the gripper is able to move on thegripper rod without sticking. Replace gripper rod and/or gripper assembly ifnecessary.

Z Motor Inoperable

1. With the right half of the arm removed, run diagnostics. Watch the drive pulley on theZ-motor for movement. The motion should be smooth until the gripper reaches the topor the bottom of the stroke.

2. Check the connections for the Z-motor through the cables and reconnect or replace asnecessary.

S. Check the Z-motor pulley for looseness. Replace the arm core/motor assembly ifnecessary,

4. Replace the tray board and retest.

8-19

Page 20: HP 7673B - S service manual

Z Sensor Inoperable

1. Using diagnostic mode, run the Z-axis test and check for sensor operation. The Z.LEDshouldbe on when the gripper is at the top of its movement.

2. Check the connections for the Z·axis sensor back through the flex circuits and traycable. Replace parts as needed.

3. Using a known good magnet, verify that the sensor is operating. Check the gripperassembly for the presence of a magnet. Replace the gripper assembly if necessary.

4. Replace the tray board and retest.

Tray Flrrt'aware Incompatibility

1. Check Eprom on tray board. If the Eprom part number is 18594-80130or less, than itshould be used with an 18596Atray. If the Eprom is 18594-80235 or greater, than itshould be used with an 18596B tray.

8-20

Page 21: HP 7673B - S service manual

<-J01 Rod LED: 6Singlo Rashes

Replace tray board's processor.

<_)01 Rod LED: 7Snglo Rashes

Tray Procel.or Error

• Tray board processor failure.

Bottle Not Found• Tray gripper failed to pick up

sample bottle from injector.

1. Check the injector for proper installation on the mounting bracket. The right front footpin should be inserted into the hole in the bracket disk. The other three iIijector feetshould be in contact with the mounting bracket.

2. Check the tray for proper installation. The tray locks on the tray mounting bracketshould be inserted into the tray base.

3. Check the standoffs connecting the tray bracket to the injector bracket. The "A" posi­tions are for HP 6890Aand HP 5700 GCs.· The "B" positions are for the HP 5880AGC.

4. Make sure the sample vials meet HP specifications.

5. Make sure the needle support assembly has properly enpged the injector turret duringbottle pickup. Ifnot, troubleshoot the injector for turret errors.

6. Run a sequence and watch the tray gripper as it attempts to retrieve a bottle from theinjector turret. The gripper should be aligned with the "neck" of the sample bottle. Ifthe gripper is misaligned with the bottle, troubleshoot the tray for the misaligned axis(R, Z, or theta).

Page 22: HP 7673B - S service manual

Controller Errors

5::::::::;=====1 Red LED: 2 Double Rashes((_))0

Replace the controller board.

«-»)CJI Red LED: 3 Double Roshos

Replace the controller board.

1. Replace the controller board EPROM.

2. Replace the controller board.

8-22

Controller Processor Error

• Controller board processorfailure.

Controller RAM Error

• Controller board RAM failure.

Controller ROM Error"

• Controller board ROM failure.

Page 23: HP 7673B - S service manual

«->)CI1 Red LED: 5 Doub~ Flashes

Confi~ration Error• Option boards not installed in

the correct position.• Option board failure.

Option Boards Not Installed In the Correct Position

1. Check that all cables are connected properly.

2. VerifY that all option boards are installed properly.

Option Board Failure

1. Suspect !NET or tray board failure. Replace ifnecessary,

Page 24: HP 7673B - S service manual

Miscellaneous Errors

Injector

Red - OnYellow - OnGreen - On

Controller

Red - OffGreen - On

Injector Processor Error

• Processor on injector boardmalfunctioning.

• 18-V supplies not available on injectorboard.

Processor on Injector Board Malfunctioning

1. VerifY that the injector processor (and the EPROM if present) is positioned properlyand seated in the PC board socket.

2. Replace the injector processor (and/or EPROM) if necessary.

8. Replace the injector logic board if necessary.

1S-V Supplies Not Available on Injector Board

1. Check for + 18 V and -18 V at the injector logic board.

2. Check the cable and connections. Replace cable ifnecessary.

B-24

Page 25: HP 7673B - S service manual

IS-Volt Supply Error

+18·V or -18-V supply inoperable on powersupply board

Controller

Red OnGreen Off

Injector

Red - OnYellow - OnGreen - On

1. Check + 18-V and -18-V fuses on power supply. Replace if necessary.

2. Unload the system by pulling out all the option boards plugged into the main powersupply board. If the + 18 V and/or -18 V works, plug in one option board at a time todetermine the defective board. Replace it.

8. If system is not responding with all the boards pulled out, replace the main powersupply board.

Page 26: HP 7673B - S service manual

L 'P-'-',,'::, :'.•'~ .,

InJector

Red - OffYellow - OffGreen- - Off

Controller

Red - OffGreen - Off

Power Supply Error

No line voltage available+5-V supply on the main powersupply board inoperable

No Line Voltage Available

1. Check to ensure that bench power is available, power cord is plugged in, and powerswitch is on.

2. Check the line fuse located inside the line module of controller. Replace fuse ifnecessary.

8. Check that the cables are connected.

+5~V Supply on the MaIn Power Supply Board Inoperable

1. Check the power supply main board for the presence of +5 V.

2. Unload the system by pulling out all the option boards plugged in the main powersupply board. If the + 5 V works, plug in one option board at a time to determine thedefective board. Replace it.

8. Replace the main board if necessary.

Page 27: HP 7673B - S service manual

HP-IB/Controller Board Errors

When a controller module containing an HP-IB/controIler board is turned on, the HP-IBboard executes a self-test. Ifa problem is found, one of the followingerror messages will becommunicated through the controller's red LED:

Red LED

2 flashes

3f1ashes

4 flashes

5 flashes

6 flashes

7 flashes

Error Message

Defective RAM on HP-IB/oontroIler board

Defective ROM on HP-IB/controller board

Defective RAM on bar code reader board

Defective ROM on bar code reader board

Bar code reader cable not attached

Bar code reader board set to test mode

Printed Error Messages

The following printed error messages may be sent to an !NET integrator or ChemStation.There is no corresponding flashing LED pattern.

lDJector Communication' Error-A communication error occurred between the injectorand the controller.

Tray Communications Error-A communication error occurred between the tray boardand the controller.

lDJector Re,et Error-The + lS-volt supply from the controller was interrupted.

Tray Reset Error-The + lS-volt supply from the controller was interrupted.

B-27

Page 28: HP 7673B - S service manual
Page 29: HP 7673B - S service manual

-, I,' ::"I"'·:'~ ,:r'~' ,

cInjector Module

Section C-Contents

Il\iector Operation C~2

I1\Jector Replacement Parts C~6

Injector Errors ..........•........•........................... C~14

Injector Diagnostics C~15

Measuring Iniector Alignment C~18

Printed Circuit Board Component Overlays C~21

Iniector Pin Out Diagrams C~28

Page 30: HP 7673B - S service manual

Injector Operation

The HP 18593B Injector Module loads a syringe with a sample and injects the sample intoan HP gas chromatograph. The syringe and the sample vial are controlled by three keyassemblies: the syringe carriage, the plunger carrier, and the injector turret.

Syringe Carriage Movement

The syringe carriage assembly moves the syringe up and down within the injector tower(figure C4). The syringe carriage is driven by the syringe motor (figure C-7). The syringemotor is a two-phase stepper motor with bidirectional currents in each phase. In addition,it is driven at two power levels. The high power level is used during the injection, when hightorque and fast operation are required. The low power level is used during all other move­ments of the syringe.

For on-column injections, an on-column syringe must be used, and the on-column injectionparameter set to "yes." During on-column injection, the syringe is inserted 19 mm deeperinto the injection port and at a much slower speed.

The needle support assembly is attached to the syringe carriage (figure C-4). It supportsthe syringe needle when the needle enters a vial or the injection port. The needle supportusembly also stabilizes the injector turret whenever the tray module arm is loading orremoving sample vials.

When the syringe needle exits a vial, the bottle stripper assembly prevents the needle fromlifting a vial out of the turret (figure C-8). The bottle stripper rides along the turret androtates away from the syringe needle during iJijection.

Syringe Carriage Homing

The home position of the syringe assembly is the full up position. The home position ischecked on power-up by sensing a magnet at the bottom rear of the syringe cani.age using aJll88Iletic (hall effect) sensor located on the back plate (figure C-4). If the syringe carriagemagnet is not on the sensor, the assembly moves up a few motor steps, down a few motorsteps, and then up until the magnet is sensed. If the magnet was already on the sensor, thesyringe carriagewill check its position by moving a few steps down and up. These move­ments should be enough to turn the sensor off and then on again.

C-2

Page 31: HP 7673B - S service manual

Plunger Carrier Movement

The plunger pulls sample material into the syringe, measures the volume, and forces it outof the syringe.

The plunger is held by the plunger carrier, which is mounted on the syringe carriage. Theplunger motor is a speed-controlled DC servomotor with tachometer feedback. With thisfeedback, the motor is driven with whatever torque is necessary to achieve the desired speedto drive the plunger. The plunger motor and tachometer are located at the top of thesyringe caniage assembly (figure C4).

Depending on the injector parameters, the plunpr will stop at one of five "steps" to meas­ure iIijection volume. The five steps correspond to l()-()O% of the syringe volume. Duringsample and solvent washes, the plunger fills 80% of the syringe volume.

Plung"er Carrier Homing

The plunger's home position is the full down position and the magnet on the plunger carrierassembly is sensed by the magnetic sensor on the sensor board (figure C4). There is anadditional sensor at the full up position of the plunger.

Turret Movement

The iJUector turret contains five positions for holding sample vials and wash/waste bottles(figure C-5). The turret is mounted directly on the turret motor. The turret motor is afour-phase stepper motor, with two common leads that are tied to +12 V. A phase is ener­gized by driving it low. The turret rotates. the vials and bottles into the it\iector tower toallow access by the syringe.

When operating with a tray, the turret rotates to a specif1c position for the tray gripperto pick up or deliver sample vials. The syringe carriage then engages the needle supportassembly into the turret to prevent any turret rotation during vial pickup or delivery. Afterreceiving the sample vial, the needle support is raised and the injector turret rotates to startthe iIijection sequence.

C-3

Page 32: HP 7673B - S service manual

Turret Homing

The turret has a small magnet pressed into its lower, exterior surface. A magnetic sensormounted behind the turret senses this magnet to find the turret home position (figure C·7).During homing, the turret motor rotates the turret until the magnet is sensed. The turretmotor then rotates the turret a few steps in both directions to determine the center of themagnet,

If the sampler system is operating without a tray and is using stand-alone control, a mag-.netic bottle insert may be used to indicate the "last bottle" during a sequence. The turretsensor also will home on this magnetic insert during its homing routines.

Safety Switches

There is a safety mechanism on the injector tower to protect the operator during injection.When the plastic door covering the syringe assembly is opened, all motor movement in theinjector module is suspended. This is accomplished through a switch located inside the topright corner of the door (flgure C·6).

A second switch is used to determine if the injector is mounted correctly. The mountingswitch is located above the mounting sleeve and is closed by the mounting post when theiI\jector is seated on the mounting bracket (figure C-B>'

Injector Electronics

All injector functions are controlled by a microprocessor on the iDJectormain board(figure C-g). All motors, sensors, and switches connect directly or indirectly to the iDJectormain board (figure Col). The iDJector draws its power supplies from the controller modulethrough the hijector cable. The cable is also used to transmit TTL and serial communica­tion between the injector and the controller modules.

C·4

Page 33: HP 7673B - S service manual

InJectorLoalc Board

Serial OutRlldy OutStart Out

J2 [ 1 I- ... TurretJ

.MOlor

Controll.r I fow.r SUDpn..> J1Modul. - Senior BoardMOlor.

S.rlalln Sol.nold~Ready In -

Start In J& J1 J2

- ~Plung.rI.nlorl

- LEOI.Front Pan.1 - J8

J4 [h~ ~ Syringe I

Slart, SlOP. - J. Motor

ParameterSwitch..

Syring. S.nlor

Fan - [~ J7 - Turr.t S.nlor

J3 - Mountlng IwUch

Fan r~ JBSwitch - l - I Door Switch

Fiaun, C·l. Injector block diagr8.Q1.

Plung.rMotor

PlungerSolenoid

c-e

Page 34: HP 7673B - S service manual

Injector Replacement Parts

Fan

Elletronlal houllng

Start/ltop buttonl

Fault LEDI

Door toIyrlngl Glrrlagl ....mbly

Turret

B..e

'..s.9o,

Figure C·2. 'HP 18593B InJector Module.

e.G

Page 35: HP 7673B - S service manual

',' .,:~.~.,!~~.,::~:""

Syringe Carrlag,An,mbly

Screw0515-052.

,AONTVIEW

RIAR VIIW

Figure C-3. Initalllni the I)'rinie carriaie asllembly.

Page 36: HP 7673B - S service manual

'I,;",;

SiniorBOlrd Conn'cUonl

Noll: P1ungl' Darrllr with ninoIllerIdl er-071173·80Q30.

Plunger ClrrllrAuembly

07&73·110880

rPlungerSorlw

07873·20li70

NeedleBuppon

Auembly07673·80Q7li

Plunge Molor AUlmbly07873·60820

/Mlgnel Preu.d

Inlo Back

Blak 01 Plunglr CarrierShowing How 10 POllllonMillor Clip Assembly

Sollnold ConnectorCln Go On Elth.r Way

Figure C-4: Syrinie carriage 8.11embly.

c-a

Page 37: HP 7673B - S service manual

Turret A..embly07&13·&0505

JD :J.(';.,

- T....

e! e!!=:a.. ' f:

V"""'IIJ:

r - a - &Ie"

c .. ... ......... 1; ! I ..., ! 1.00'-

White Samplelnllrl -----...:.~~;;r07673-40226

Turret MagnetLocatldon \

Bottom Edgl ,

Figure e-G., Injector turret.

c-s

Page 38: HP 7673B - S service manual

Back Platll/BlIti Altlmbly07&73-150851

Front Hauling A..lmbly(IncIudII Door)07e73-80ae&

NOTE: Plaltlc tabl InapInto blckpllte.

Figure CoS. Injector coven.

BlckCovlr07673-400711

Door Switch ............81U'1. 07&73-15079 ........

0&24·06112 ........¥Fan A...mbly K ........35900·601120 r .........

, "...~=-='-

,J!)'

~~ Screw

0515-0810

C·10

Page 39: HP 7673B - S service manual

Ribbon Clbl.Cllpi

07173-40170

-L- Br.cketWllb.r

21IlO-04011

Syring. MotorA...mbly

0787a.80580

Blck PIII.IBI"A...mbly07173-&0855

\Syring.C.rrllg.S.nlor

-L- Br.cket-~~~1400-02110

Scr.w0151G-0637

Turr.t S.nlor

a:~!~~I1~Hom. Senior

Scr.w Turr.t0824-0851 "'" "'lliID';:O"......-s.nlor

Syring. MotorScr••1 '"\=:::----Ifl.-~.

0I51G-OII10

Idl.r Pull.y Sh.n07173·2015415

Hom. S.nlors.t(Includll Both S.nlorl)

07873·105110

FilfUre C-7. InJector tower-rear view.

Page 40: HP 7673B - S service manual

Bonle Stripper07673-81005

Spring07673-2067$

TurretMotor

AI..mbly07173-10810

Sleeve Brlcklt07673·00185

/'

TurrltMotorScriwi

015111·1380

Berl.0515·13D1

Blck PlltlIB..IAnembly

07873-808155

Fipre e·s. injector tower-front view.

Page 41: HP 7673B - S service manual

Electronici Hauling AaaemblyOncludel Both PinIII, HlnOl, and Hlngl Screwl)

07873-80728

InJlctorClble Aaaembly <'

07873-80648

Screw0518-0883

Mlln Baird Aaaembly07873-80025

or07873-89025

".80rlw tit;

051a.QU24 .~

\ :... J ~~ .

Ground Clbll I07873-80838

Front PlnllBaird AUlmbly

07&73·10035

..

l1I;L~

~j.1Y 'F"atP...IA Door0787....0215

Screw0381).2283

Fipioe C·B. InJeetor electronics.

Page 42: HP 7673B - S service manual

Injector Errors

An error condition is indicated if the red LED on the front panel is lit or flashing. Errorscan be caused by:

• Objects obstructing the movement of any injector mechanism.

• Sensors not tripped when expected.

• Safety door open.

• Il\fector installed improperly.

• Motor failed.

• Wiring damaged.

Refer to Section B, "Troubleshooting", to see a full list of error conditions and courses ofaction. When the fault is corrected, clear the error by pressing the STOP button on theinjector. ~en restart the sequence.

C-14

Page 43: HP 7673B - S service manual

Injector Diagnostics

The Injector Diagnostic Mode allows the service engineer to observe the operation ofinjector motors and sensors during the troubleshooting process.

I CAUTION IThe diagnostic mod. overrides the door and mount safety switches. Be sure totake the syringe out of the Injector before operating It In diagnostic mod•. Thismode should only be u..d by a trained service technician.

Starting Ol,agnostlcs

1. Turn off the controller.

2. Simultaneously press the STOP and START buttons on the front of the injector. Keeppressing;while you turn on the controller.

8. The Red LED should start flashing a double flash, indicating the first test in diagnosticmode.

LED Diagnostic Functions

Each of the LEDs on the front of the ifUeotor has a special function for each of the threetests available in the diagnostic mode.

Telt Red LED

Syringe Can1age Two flashes

Turret Three Rashes

Plunger Four Rashes

Yellow LED

On when syrlnge motor Isenergized

On when turret motor Isenergized

On when plunger motor Isenergized

OrlinLED

On when syringe carriage magnet Isdetected by syringe carriage sensor

On when turret orbotUe Insertmagnet Is detected by the turretsensor

On when plunger magnet Is detectedby upper orlower plunger sensor.

Table c-i, InJector Diaenostic LED.

Page 44: HP 7673B - S service manual

.. " i'f\~'T,'" ~I'., I':~;r

Syringe Carriage Test

1. Press START on the front of the injector. The turret and syringe carnage will move totheir home positions. The syringe motor will remain energized for a few moments. Thegreen LED will light whenever the syringe carriage sensor detects the syringe carriagemagnet.

2. Press START to drive the syringe motor one step (2 phases energized).

8. Hold START down to run the syringe motor continuously. Wait about 10 seconds for aresponse. Note the LEDs on the injector main board (figure C-ll). They should light inthe following sequence:

DSl

ONONOFFOFF

DS2

OFFOFFONON

DS3

OFFONONOFF

D84

ONOFFOFFON

Table C-2. LED Pattern durin, Syringe Motor Movement

4. Press STOP to advance to the turret test.

Turret Test

1. Press START. The turret will go through its homing routine. The motor remains enar­pzed for about 10 seconds. The green LED willllght when the turret sensor detects theturret magnet.

2. Press START again. The turret will rotate and home on the next magnet. If there isno magnetic bottle insert, the turret will find the original magnet again.

S. Move the turret manually to verifY the sensor trips when the magnet goes by.

4. Press STOP to advance to the plunger carrier test.

C·18

Page 45: HP 7673B - S service manual

Plunger Carrier Test

1. Press START. The plunger will go to the top of its movement while energizing thesolenoid to catch the plunger carrier at each stop on the wo:yup. The plunger motorthen drives the plunger carrier to the bottom of its movement. The green LED willlight when the plunger carrier is at the top or bottom of its travel.

Note the LEOs on the injector main board. DSB lights when the I?rocessor is sending asignal to the plunger motor drive to enable the plunger motor.

DS5 is on except when the plunger motor is stalled. The plunger motor stalls briefly atthe top and bottom of the plunger carrier travel and during the setting of the samplevolume.

2. Manually move the plunger carrier up and down and verify that each sensor trips whenthe magnet moves by.

3. Press STOP to return to the syringe carriage test.

C·17

Page 46: HP 7673B - S service manual

Measuring Injector Alignment

Misalignment between key assemblies within the injector could cause errors and/or bendsyringe needles. The following tools can be used to determine if the injector is properlyaligned:

07673-20810 False Syringe07673-20800 Turret Sleeve07673-20820 Base Plug

These tools are part of the Alignment Tool Kit <HP part number 07673-60950).

Measuring Alignment Between the Syringe Carriage and the Turret

1. Remove the syringe and the needle support assembly from the syringe carnage.2. Install the false syringe in the syringe caniage (figure C-IO).

3. Remove the turret, swing out the bottle stripper and replace the turret.

4. Place a vial insert into the .am:nn.d turret position.

5. Place the turret sleeve into the vial insert.

6. Rotate the turret sleeve under the false syringe.

7. Manually move the syringe carriage downward. The false syringe should slide cleanlythrough the turret sleeve. If it does not, do the following:

• Check the alignment tools for proper installation.

• Check the syringe carriage, the injector turret, and the turret motor for properinstallation.

• Replace the syringe carriage and check alignment again. Repeat this process withthe injector turret and the injector back plate/base assembly ifnecessary.

C-18

Page 47: HP 7673B - S service manual

Measuring Alignment between the Syringe Carriage and the InjectorBase .

1. Move the syringe carnage back up but keep the false syringe installed.

2. Remove the turret.

3. Remove the injector from the GC..Stand it on a flat surface.

4. Place the base plug into the iniector base hole. The narrow portion of the plug shouldface downward.

5. Manually move the syringe carriage downward. The false syringe should slide cleanlythrough the base plug. If it does not, do the following:

• Check the alignment tools for proper installation.

• Check the syringe carnage for proper installation.

• Replace the syringe carriage and check alignment again.

• Replace back plate/base assembly and check alignment again.

C-19

Page 48: HP 7673B - S service manual

•...!

•...i~ ~

Baae Plug07673·20820

Injector BI..

Turrlt Sleevl07873·20800

Turrlt In.ert07873-40160

T'"·~.~_lll""ro---

Fipre C-l0. Installing injector alianment tools.

- - - - - - ""M=GlJIi-A'l

FII.. Syrlngl07873-20810

c-eo

Page 49: HP 7673B - S service manual

Print Circuit Boards Component Overlays

Fisure C·tt: InJector main board (07678·60025).

C-21

Page 50: HP 7673B - S service manual

0 0

a0840

J1 IndJ2 Ire ontheopposite sid••

I~

1 I,

I J1 I I J2 I @!I CR3l _____1 L_____J

0oc:::J. R3

REV @!I CR2B 0c:::Je R2

[ill E!lCR1

0r:::=J. R1

a0 830

07673-60010 07673·60035

0

1:::9$:::1

I:::!!:::I~ ~

0 .....-

J1

'---

0 0 0

Fieure C-12. Injector BenBorboard (07873·60010) andinjecfor control. board (07673-60035).

C-22

Page 51: HP 7673B - S service manual

Injector Pin Out Diagrams

TOU23

TOU32

+5V

+TURRETI~O+-__+-+HOME

READY IN

FROMCONTROllE

MODULE

Fiaure e-1S. "Injector main board wirinl diagram.

C·23

Page 52: HP 7673B - S service manual

J2

+12V

34

12IIII

TURRET MOTORWINDING RESISTANCE

78 0 ..t115% EACH

J4

2BLACK

1 RED/WHITE

a RED

TO 7 WHITESYRINGEDRIVER

a GREEN

II

4

3

SYRINGE MOTORWINDING R£8tBTANCE0.340 j-10% EACH

Figure C-14. Installlni main board wiring dialJ"am.

C-24

Page 53: HP 7673B - S service manual

+18V

PLUNGERSTOP

PLUNGERMOTOR (+)

TACH (+)

+12V

PLUNGERHOME

J5ON INJECTOR

MAIN BD

SYRINGE CARRIAGE SENSOR BDr---------------~I J1 J2 I

I

TACH17en+12%-25%

PLUNGERSTOP

SOLENOID

+ + PLUNGERMOTORun+8%-25%

Figure C-llS. Injector main board/sensor board wlr1ni diairani.

C-26

Page 54: HP 7673B - S service manual
Page 55: HP 7673B - S service manual

,DController Module

Section D-Contents

Controller Module Description D-2Common Configurations D-5

Controller Module Block Diagrams D-6Control Signals D-9Fuses D-9Power Supply Usages ................................•........ D·10Controller Replacement Parts D-11Printed Circuit Board-Component Overlays D-12

Controller Main Board D·12HP-IL Board .....................•....................... D-18T'rL Board 0·14HP 5880A Interface Board D.15Standard Controller Board •.......................... ~ D-16TraY Controller Board D-17RS-232 Board D-18Loop Board ..••••.•.••.•....•••.••....•....•............. D-19Bar Code Reader Board ....................•............... D-20HP·ffi/Controller Board 0-21HP5880A Mainfuune Board D-22

HP 5880A Dlagnostic Mode 0·23

0·1

Page 56: HP 7673B - S service manual

"I '''':I'\'''J:'l'~I':'''ll'''I~!''''~'1~I;W:~~~iM\!$Jl'Hnt~~!~~t'''··7IIlW''~~~!!II,.

Controller Module Description

The HP 18594B Controller Module houses most of the electronics for an HP 7678 Auto­matic Sampler system and serves as an interface between the automatic sampler and otheranalytical instruments.

The automatic sampler can be configured in several different ways. Depending upon theconfiguration, the controller will have different option boards installed in it.

Each controller contains the power supply board (HP part number 18594-60125) whichprovides the necessary voltages to operate the sampler modules. The power supply boardalso provides a data bus for data communication between the option boards.

Depending upon the configuration of the automatic sampler, up to four boards may bemounted on the power supply board. The slots are labeled TRAY, AUX, COMM, andCONTROLLER (figure D-1 ).

Controller Slot

The following boards may be put into the CONTROLLER slot:

Standard Controller(18594-60060)(Rebuilt 18594-69060)

HP 5880A Interface(18594-60060)

Hp-m/Controller(18594-60110)(Rebuilt 18594-69110)

0-2

Coordinates the tray, the bar code reader, one or two injectorsfor INET, stand-alone, "loop," or RS-282 control (see figure D-2).

Provides an interface specific for HP 6880A GC control. Itprovides two-way communication between the HP 5880A GC(or the HP 8888A Integrator) and the controller module (seefigure D-8).

Allows HP-IB GEEE-488) communications between the HP 7678and HP ChemStations or "A" series 8850 LAS. It also acts as acontroller board for these specific configurations. For U A" series8860 LAS control, an RS-282 board is required in the COMM slot.The HP-IB/controller board does not support INET or stand-alonecontrol (see figure D-4).

Page 57: HP 7673B - S service manual

Communication Slot

The following boards may be put into the COMM slot:

HP-IL(18594-60030)

TTL(18594-60040)

RS-232 .(18694-60080)

Loop(18694-60090)

. Auxiliary Slot

INET communication board. Although the hardware on thisboard is !NET compatible, the software to control it is on thecontroller board. Therefore, it can only function with a standardcontroller board in the CONTROLLER slot.

Provides remote and BCD information to the integrator. If theTTL board is in this slot, the injector should be in the front posi­tion. It performs two distinct functions. The first function is theexternal interface for the READY IN/OUT and START IN/OUTsignals. The second function provides a BCD bottle number toexternal data handling devices.

Allows HP 7678 control through RS-232 communications. Formore information see, the "RS-232-C Serial Interface Manual,"HP part number 18694-90800.

Allows an HP 3350A/8367A LAS system to send the next bottlenumber to the 7678. Valid bottle numbers with the tray moduleare 01-00 (where "00" is read as "100") or for an injector-onlysystem, 01-03. Once a valid bottle number is sent to the HP 7673,the sampler begins its tray and injection sequences.

The following boards may be put into the AUX slot:

TTL(18594-60040)

Bar Code Reader(18594-60100)<Rebuilt: 18594-69100)

Provides remote and BCD information to the integrators. If theTTL board is in this slot, then the injector should be in the rearposition.

Controls the bar code reader (BCR) module. It also acts as aTTL board, so it can replace a TTL board in this slot with no lossof functionality. The BCR board contains a bank of eight dipswitches to specifY beeper operation, self-test mode, the positionof the BCR on the tray, and the bar code label type.

0-3

Page 58: HP 7673B - S service manual

Tray Slot

If there is a tray module, the tray board (18594·60075or rebuilt 18594~69075) must be inthe TRAYslot to control tray operation. Also, when a tray board is installed, you must havethe tray cable installed.

The main functions of the tray board are reading the status of the. three homing sensors andthe bottle sensor, driving the three stepper motors (Z, R, Theta) in the tray, and communi­eating with the controller board.

Page 59: HP 7673B - S service manual

Common Configurations

~UUUUUu Illij ~UUUUUu IlIDI·· III

il' I U..-,.

I· i.- J .am. ,. t:;

! • i!!

til)iii ~ wi~ltu i· ~gi f

~I!!: !I!I iII I I 8

1!I .I ...

G>

1~uuuuuu 100 ~uuuuuu IlliJs:I

II I i..-,. .... iMIlIli '2 30

8I.· ..... _1·1 III

£:

.i)t ~

~ .~~

I !ti i I ~I~ ...1:10Ij it!:

Vol

I~:=~.!)

0-5

Page 60: HP 7673B - S service manual

,iL,..

Controller Module Block Diagrams

IL~ I

II I I--f-

0 I--I----:~ I

::::::II

Illi IIi i ~----I1f. 1

18- 1

1II 1

1QI. IQ.L - - ....- - - - - - - -- - - - -1-- -

I

II

il ItilI

II 1.-- ..I

L I_/

lill[:;.. u..

I~I Ifii 'I

J-•l- i I...... ..-- i'- ......

I ~I

[]I

II I I f~ l ~ f i

IS Iu ..I II II 1.--.

II

iI,... Ir- ..I :~:l I0 I

:::: -:: 11E :1: HI IE' Ig Ii Il • I~

.. I ..--I

0·8

Page 61: HP 7673B - S service manual

, " ...·'.J,'l:~".·: ,"'''I:J'lr: ,:,",,~.~:, ,J'~'.~'l(-~",~~~"~,~~,,,:,I ':,'i:\r,~~':'t:"".' " .

IIil. L _

I I -

0I-- -

.....-I--

f---

1111.; IIiII 10lio----111111

I II&; 1

II1

I I 1;: ...... --------------- ,.-~---·1 l

III 111~ 11l!tI II ..--II III

!I'I.It

(]II 14-- j III

II I

1II I~ IIII 1I

II ~--II iI

t: · I:::1:; II

~ • l El1 I! \\1:: IIIj .. Il-l' ~--- • I.... I ;;::;::::::::.

0·7

Page 62: HP 7673B - S service manual

I I r---I--

0 r---r---.... r---

I!I IIII i f--------IIIIIII

'"I

.---

II

D~ ---------------- ~~---I ......•........ ' .........

I

II aIj 11!! II~

if I : I"-

II !! -I IIi I :{~~~~~~t:~~~~r~: .-...I

~I I If! I

III

~ !I

iilili J! I Ic I) IIIc .II

311"0 I I•c I l "..;. :-

! ! I II :f{i)r

~_ ...

I I!I! I •.L I I «II

B I I[] Ii l ali1,1III I

I

II "'"

~--II II

E:' · I .,::,,:.:,:':

II :'1• I.::;::" .> 11! i I II I! iiL • I~

II n! 1.---•IL~___________________

Page 63: HP 7673B - S service manual

I;':' I~ "

Control Signals

The control signals used between the controller and the injector are:

READY OUT

READY IN

START IN

START OUT

SERIALCOMM

A logic high indicates injector readiness.

A logic high indicates system outside of the injector is ready.

A negative going logic pulse tells the injector to start an injectionsequence.

The injector emits a negative going logic pulse at the time ofinjection, intended to start the system GCand integrator.

A two-way serial communication link from the injector processor'sUART.

All these signal lines are RS-232 compatible, and they are logically inverted. A logic high atthe injector is seen at the RS-232 line as -12 volts, and a logic low at the injector is seen atthe RS-282 line as +12 volts.

Fuses

Fuses have been installed at several locations within the instrument for the protection ofmajor power circuits. They are designed to open as quickly as possible to prevent damage toother components within the circuit. Occasionally, an opened fuse may have been caused bya short one-time surge;' however, it is far more common that a component within the pro­tected circuit has failed. When an open fuse is noticed, replace it. If the replacement blows,suspect a component failure.

The following is a list of the fuses for the automatic sampler.

Power Supply Board

Una Module

Reference Designator

F1, F2

100 Une

120 nne

220/240 nna

Fuse Type

15A

3.0 A (NTO)

2.5A (NTO)

2A (TO)

HPP/N

2110-0054

2110-0003

2110-0083

2110-0303

Table b-i. Controller Module Fuses

0-9

Page 64: HP 7673B - S service manual

Power Supply Usages

The main board in the controller module provides various voltages to the option boards andthe sampler modules. The figure below shows where these voltages are used.

POWER WHERESUPPLIES USED

,+5V~+5V

.,ALL m SIGNALSALL LED.

+18V ., +18V _SYRINGE MOTOR RMOTOR BCRMOTORPLUNGER MOTOR ZMOTOR BCR SOLENOIDPLUNGER SOLENOID THETA MOTOR

+12V _TURRET MOTOR THETA SENSOR SERIAL COMMUNICATIONS..PLUNGER MOTOR CTLS RSENSOR

Z8ENSOR, ·18V ..

-18'1 ~, . SYRINGE MOTORPLUNGER MOTOR

·12V .. PLUNGER MOTOR CONTROLSERIAL COMMUNICATIONS

-23V _SYRINGE MOTOR CONTROL~

Fip.re D-5. Power supply usage••

0·10

Page 65: HP 7673B - S service manual

Controller Replacement Parts

PLASTIOOOVER

01Q40-40041

NAMEPLATE

1MG4-001OS

•~" .

POWER 8UPF'LY BOARD1!8g.(-eo12a

1815g.(-ee125(REBUILT)

OPTION BOARDS

I SORew

05111011il

Figure D·G. Controller module replacement parts.

0-11

Page 66: HP 7673B - S service manual

Printed Circuit Boards-Component Overlays

o 8 0

a=

@l-O_.J--.......@~

o'

o'

o fI

-o.

"o.

Figure D-7. Controller main board (18594-80125).

0-12

Page 67: HP 7673B - S service manual

~ ~T2

ocmDO I T1 Ioamo 11

O[]IJO

oamo

O[]DOo []DO0C!!!J0O[][]O.

.O(][]O 01I~

°ODOC1 0[]i]0O[]i]O

L.....1 O[][]O oCllJOOUiJOOCiLJO

1 II

J1U1

L.....-

10 II14 ,.o[J!Jo

OCDJO

FiJUre D-8. HP-IL board (18394-60030).

0-13

Page 68: HP 7673B - S service manual

J2

~ J3 0

II K2 II IB~~~~

K1

R2G -c:J-R21 -e:J-R22 -c:J-

B BRl1-=-BCRB -c:J- U4R11 -c:J-CR10-c::::J-

R20-c::J- B ~IC2 -e:::J-

C11 r+R24 -c::::J-

- R23 -e:::J-C3 -e:::J-

J1 B BoC4-c::::r

LJ

Figure D-9: TI'L board (18594a60040).

Page 69: HP 7673B - S service manual

~ J3 ~B

E~OTP1

I-- OTP2

U1 OTP3OTP4·OTPII

J1 1 d~:o Tl'1IOTP7

-e::J- C. OlPI-1 r CII OTPII

OTP10I-- -e:J- CI OTP11

~R1B OTP12

I----c::r R2 U2 0-120+120+1

i-e:::r C7 0+1rCB OGND-e:J- CII OOND

J2 -e:J- C10

I---

Fiaure D-I0. lIP 8880A interface board (18594-60050).

Page 70: HP 7673B - S service manual

<>

Ull III

~i § ~+.tll~

U17

D~18

Figure D~l1. Standard controller board (18594~60060).

Page 71: HP 7673B - S service manual

Figure D-12•. Tray controller board (18394·60073).

0-17

Page 72: HP 7673B - S service manual

---lI I

J2

~::J3• 12

~r1 11

BG~·~ '~D--<=>- CIUK

UK ~IIIUK1100IDOIlIO

.8 ,B .B .B§E1 C4--<=>-

CI--<=>-

B1 C1 ~U1

L.....l.ce--<=>-

1.CI--<=>-

J1

'iD"ii

Fii\U'6 D~1S: RS~2S2board (l8594~60080).

D~18

Page 73: HP 7673B - S service manual

o[l[Jo

,1"'"'' . '·':~~'II""r',' '~.":'1"

J2111111111 It 110000000

C8 1-0oc

.B: ,:0 ,8

o (liJO

,8

Figure-n·14. Loop board (18594·60090).

0-19

Page 74: HP 7673B - S service manual

Figure D-13. Bar code reader board (183940060100).

Page 75: HP 7673B - S service manual

-i: "J

J3 o

Figure D·16. HP·m/controller board (18594·60110).

0-21

Page 76: HP 7673B - S service manual

-\

+IVD

NOlIlIo

D-11V

DJ:.!IY

~UniT

WID

~ <> t+Q TEST MODE INweADONSFAULT LAMP B10NUI'tmBPBETA-

. cz ~ -) TEST oleFAULTWrrtl

l- < ON> - ) 80411

011 ~ 01. I 014 <OFF> 1111110 COMM.11)011 01 n u n - 1111110 COMM. 2)

C4 +ClO 7U 7U:: : _ r~=Cii:lPABALCOMM.

w

n0,,~..;J~~~ ,:~ ,~

~:l.O'" .. ~~ ~ ~:' rv-J

-.II

--I

<>Figure D-17. HP 5880A mainframe board (19321-80010).

Page 77: HP 7673B - S service manual

HP 5880A Diagnostic Mode

The interface provides two-way communication between the HP 5880A GC (or theHP 3388A Integrator) and the HP 7673A controller module. The system consists of fourassemblies:

HP 5880A mainframe boardHP 18594B interface boardCable (HP 5880A internal)Cable (HP 5880AlHP 18594B)

Part No. 19321-60010Part No. 18594-60050Part No. 19321-60500Part No. 19321-60510

On the HP 5880A end of the 19321-60010 mainframe board, the communication is doneover the standard MC5 (or MC2) bus. Communications between the HP 5880A controlboard and the automatic sampler controller is encoded using + 12V and -12V.

The U44 code includes a system self-test program. The test covers the HP 5880A main­frame board, the cables, and the automatic sampler control board. For the purpose of thistest, the system is divided into 6 sections:

HP 6880A communicationsTRAY communicationsINJN1 communicationsINJJII2 communicationsPARALLEL communications

0-23

Page 78: HP 7673B - S service manual

Each section is tested in sequence. If the test of the first section is successful, the test pro­gram proceeds with the second section, and so on. After the fifth section is tested success­fully, the test LED (DS-1 on the control board) is turned ON for one second, and then thecomplete test is repeated. Ifany of the sections fails the test, a corresponding light patternis displayed by the test LED and the failed section is tested again.

To get into the test mode, the followingprocedure needs to be followed.

1. Turn OFF power to both the HP 5880Aand the HP 7673Ainstruments.

2. Move the "NORMAL/ TEST" Jumper on the HP 5880Amainframe board to the"TEST" position.

3. Set all switches of Sl on the HP 5880Ainterface board (HP part number 18594.60050)toTSTMODE.

4. Verity that the cable (HP part number 19321-60510) is connected properly between thecontroller and the HP 5880A.

5. Turn the HP 5880A power ON.

6. Check. the test LED on the mainframe board for one of the flashing patterns listedunder "symptom" in Table D·2. Follow the corresponding "Action" instroctions totroubleshoot the failure.

Page 79: HP 7673B - S service manual

-r ,'"r,., ,. ,:,~"v..

SYMPTOM ACTION

HP 5880A LED Signal: HP 5880A mainframe board (19321-60010)1 sec ON and HP 5880A Interface board (18594w60050)

1 sec OFF and cables between them are aU okay. Check Injector1 sec ON and controller LEOs forproblem indications. Trouble-

etc. shoot non-HP 7673A related aspects ofthe HP 5880A.

Means everything okay. No problem In HP 5880NHP 7673A control circuits.

SYMPTOM ACTION

HP 5880A LED Signal: Replace HP 5880A mainframe boardON Continuously (HP part number 19321-60010)

Means defective microprocessor (8049) orassociated components on themainframe board.

AmONSYMPTOM

HP 5880A LED Signal:. OFF Continuously

1. Replace HP 5880A mainframe board(HP part number 19321-60010)

2. Replace HP 5880A ROM, RAM and processor boards.

Means defective communications between HP 5880A processor and the mainframe board.

ACTIONSYMPTOM

HP 5880A LED Signal:1SHORT FLASH

1. Replace HP 5880A mainframe board(HP part number 19321-60010).

2. Replace HP 5880A ROM, RAM and processor boards.

Means defective convnunlcatlon between HP 5880A processor and the mainframe board (bad data).

ACTIONSYMPTOM

HP 5880A LED Signal:2SHORT FLASHES

1. Replace HP 5880A mainframe board(HP part number 19321-60010).

2. Replace HP 5880A Interface board(HP part I'lJmber 18594-60050)

3. Check Interconnecting cables.

Means defective convnunlcatlon In the tray path betWeen HP 5880A Interface board and the HP 5880Amainframe board.

Table D-2. HP 5880A Dia~o.tic. Mode <continued next pale)

0·25

Page 80: HP 7673B - S service manual

SYMPTOM ACTION

HP 5880A LED Signal: 1. Replace HP 5880A mainframe board3SHORT FLASHES (Part No. 19321-60010).

2. Replace HP 5880A Interface board(Part No. 18594-60050).

3. Check lntercomeCUng cables.

Means defective communication In the Injector#! path between HP 5880A Interface board and theHP 5880A mainframe board.

SYMPTOM ACTION

HP 5880A LED Signal: 1. Replace HP 5880A mailframe board4SHORT FLASHES. (Part No. 19321-60010)

2. Replace HP 5880A Interface board(Part No. 18594-60050).

3. Check IntercolVleeting cables.

Means defective communication In the InjectorN2 path between HP 5880A Interlace board and theHP 5880A mainframe board.

SYMPTOM ACTION

HP 5880A LED Signal: 1. Replace HP 5880A mainframe board5SHORT FLASHES (Part No. 19321-60010).

2. Replace HP 5880A Interface board3. Check Intercoooeetlng cables.

Means defective communication In the Injector's paraDel paths between HP 5880A Interface board and theHP 5880A mainframe board.

Table D-2. lIP G880A Diagno.tic.

Test points for all the key signals are provided throughout the board (resets, clocks, RJWstrobes, inputs, and outputs). The "ALE" <Address Latch Enable) test point carries a signalthat is 1/15th of U44 clock frequency. This signal indicates the U44 is alive.

0-28

Page 81: HP 7673B - S service manual

ETray Module

;

Section E-Contents

Tray Operation .............•................................. E-2

Vial Movement Sequence ...•................................... E-5Tray Replacement Parts ~ E-6Theta-Flex Assembly Procedures E-12Tray Errors ....•............................................ E-18Tray Diagnostics .....•....................................... E-14Printed Circuit Board Component Overlays E-15'!'ray Pinout Diagrams E-16

E-1

Page 82: HP 7673B - S service manual

Tray Operation

The HP 18596B Tray Module includes four tray quadrants that hold sample vials and arobotic manipulator that moves the sample vials to and from the injector module and thebar code reader module.

There are three axes of movement for the tray module:

1. The Gripper (Z)Axis represents the vertical movement of the gripper.'

2. The Radial (R) Axis represents the inward and outward movement of the arm.

8. The Angular (Theta) Axis represents the circular movement or the arm.

Gripper (Z) Axis Movement/Homing

The gripper picks up and releases the vials and moves up and down along the Gripper orZ-axis. The gripper assembly (figure E-4) consists of the housing; the gripper jaws, a seriesof levers, and magnets for bottle sensing and homing. When a bottle is in the gripper, itpushes a lever that moves a small arm containing a magnet in the tray arm away from itsmagnetic (hall effect) sensor.

The Z~motor, is a stepper motor and is located in the arm. It drives the Z~beltwhich, inturn, moves the gripper in the Z~axis.

The gripper's home position is a specific position at the top of the Z~axi8. To find the homeposition, the gripper is moved up until the Z-magnet is sensed by the magnetic sensormounted on the sensor board in the arm (figure E-U. It continues upward until the Z-homemagnet passes the sensor. It then moves back down to the center of the sensor whichbecomes the home reference position. The gripper will then move up eight steps to positionthe bottle magnet in front of the bottle sensor. It is only in this position that the grippercan sense whether it is holding a bottle.

Radial (R) Axis Movement/Homing

The arm itself has two axes of movement, the radial (R) and the angular (theta).

The arm is driven in the R-axis by a stepper motor located in the tray base (figure E-G).The motor has a pressed-on gear that meshes with a larger drive gear. The drive gear isattached to a pinion which, in turn, meshes with the rack rear on the left arm cover.

The radial (R) home position is sensed when the arm is fully back. To find home, the armmoves tothe full back position until the R-magnet in the turret assembly (figure E~5) issensed by the R-sensor mounted at the bottom of the arm sensor board. It continues mov­ing back in small steps to ensure it hits the fixed stop. It then moves small steps forwardand backward to find the edge of the'sensor. This position becomes the R-home position.

Page 83: HP 7673B - S service manual

Angular (Theta) Axis Movement/Homing

The rotational axis of the arm about the center. of the tray is called the angular (theta) axis.The arm is driven on this axis by a stepper motor located in the tray base (figure E·6). Themotor rotates the turret directly through a circular gear rack molded into the tray turret,

The theta home position is determined by a photo-interrupter (emitter/collector pair) beingtriggered by a plastic vane molded into the turret, The vane covers half of a full rotation ofthe turret; thus, the interrupter is on for halfa revolution and off for half a revolution. Dur­ing homing, the turret rotates until a change of state (on/off) is sensed. The direction ofrotation is then reversed and the turretrotates a fixed number of steps. The direction isreversed again and the turret rotates one small step at a time until the edge of the vane issensed. This position becomes the theta-home position.

Tray Quadrants

The tray has four removable quadrants, each of which accommodates 25 sample vials.Quadrants can be temperature controlled through the use of a circulating water bath. Seethe HP 7678 Automatic Sampler Operating Manual for temperature and inlet pressurespecifications.

Tray Electronics

The tray module is controlled by the tray board located in the controller module. The signalpaths for the tray sensors and motors travel through the tray cable and two flex circuits.The theta-flex circuit rotates within the tray turret. The R-flex circuit moves within thearm in the R·8Jds (figure E-l).

Page 84: HP 7673B - S service manual

ARMSENSORBOARD

J2Z-MOTOR

,-------1r--------------------...// - I -~~~~~1

TRAY ARM 1 serneI SENSOR11 ~I l-t;;J--------J.~rr R·HOME1_____ _ _ _ ____ _ _ _ ...J IENSOR

R.fLEX-----------.,

ITRAYTUMET I-----1 __1

--II

THETA·SENSOR II

TRAY BASE THETA·MOTOR I

I R·MOTOR J4 J1 I(.-~~~~------------------ --~===-(

TRAY CABLE

CONTROLLERMODULE

Fla;ure E·l. Tray module block dlalram.

E·4

Page 85: HP 7673B - S service manual

J:I;T'I;.' ~', , ,11,,0'" I

Vial Movement Sequence

StartOperation

HomingRoutine

Vial Check

Pickupa Vial

Deliver theVial to theInjector

Finish

The sequence starts when the START button on the front panel is pressed,or via remote signal from an external controller such as INET.

The arm finds its home position for the Z-axis, the R-axis, and the theta­axis.

If this is a start-up after a sequence interruption, the tray arm checks forthe presence of a sample vial in the gripper, the injector turret, and the barcode reader (if installed). Ifa vial is found, it is returned to its proper posi­tion in the tray. The arm then returns to the home position to begin thesequence.

The arm moves on the R-axis to the center row and then rotates on thetheta-axis to the proper angular position. It next moves on the R-axis toalign with the vial. The gripper moves down to the proper height androtates to grip the vial's neck. The arm makes small radial adjustmentsduring pickup to optimize alignment with the vial. The gripper moves up tothe Z-home position to confirm the presence of a vial.

The arm moves to the middle row on the R-axis then rotates on the theta­axis to align radially with the iJijector. It moves on the R-axis to the turretlocation then rotates on the theta-axis slightly to move the vial over thesample position in the hUeetor turret. The gripper lowers the vial into theturret, helped by the cone shape of the sample position. The gripper rotatesto disengage the vial, while the injector turret is stabilized by the needlesupport assembly. The motions are then reversed to return the arm to thehome position.

All tray module movement is completed before the Injector Module isactivated.

E-8

Page 86: HP 7673B - S service manual

Tray Replacement Parts

QUldrlnt Turret

F1lure E-1. HP 18596B Tray Module.

E-6

Arm

Grlpplr

Gripper Ilwl

Slmpllvlll

TrIYBI..

Page 87: HP 7673B - S service manual

ArmAllembly18686·60855

-,

Serl..0516-0886 Rod Holding PIli'

11586-00015

I CAUTION IWhen unplugging flex elreu",

bl elr,Mor elreu" wllltllr.

Do not pull on circuit.

Tee MlnlfoldSNR 100361

Trey Via' Quadrant,(SII of.)

185n-.uJ016

Figure E-3. Removing the tray arm assembly.

E-7

Page 88: HP 7673B - S service manual

Left Arm HIli18&116-4004&

R·Flex Circuli18&118-80038

Senior BOlrd1811111-110035

Arm Core/MotorAllY

181596-110905

E-8

Smooth Pulley185118-401011

Screw051&-0886

Right Arm Hllf1811116-40055

Fiaure E·4. Tray arm assembly (18596-60855).

Screw01115-08117

Screw0515·01lU

Page 89: HP 7673B - S service manual

JCAl1TlON IDo not remove bUlhlngl from turrltanlmbly orthl arm will become

mllallgnld.

Rack Pinion Any~18&Q8·IOU& Turret Anembly

18598·60515(Includ81 BUlhlngl)

Radial AlIIHoming Magnll

Connector from Thltaoflel Anlmbly

Rod/Arm1859&·20500

<II.

~ CoRing0510·121&

Flaure E-5. Tray turret.

E·g

Page 90: HP 7673B - S service manual

t ,',I,",1~ {;'~j

T",

Tray Termination Board18511&-80075

1lI111-aanlor185M-B0830

Screw0515-1823

~ "". Tray Cabll CovlrScrew - ~·f 18511540235

0515·0886 a J8

r. M Sari.Trey Cabll -- ¥ 0515-0888

18511&-&0715 fBottom Covlr11511&.20745

Cotter Pin1480-0858

Tbeta-Motor A..ambly1859&-8OSQ5

I CAUTION IUlllnitanatlontooll

In IIrvlal klL

Tray Ball18511s-&01575

Sari.OS15-0894 ..-,

Drain -.dl1859a-C0240 - ~

O-ring01101-0583~

FiaiJre E-6. Tray base.

E-10

Page 91: HP 7673B - S service manual

TrlY Base Anembly18598·8087&

Thltl.flll Cover18598-40170

Thet•.fI81 An.mbly18&9&-80040

Thrult Walher1410·1229 "

EYIlet0381·0014

Thetl·Slnlor

~~~~~~!~~~~~~~~~ 18598-80830

I CAUTION'

~..._ '.nlor round hollrnl on pell.T.ot-I~..,::o;;;.., 0,,1 holelalor IGrIW.

•~ Flat Hlad Screw0&1&-0077

Turret~18&98·80&1$

I CAUTION IOrllntltlonor thlta-nelIxtrlmely Import.ntl

Tbilihouid only be.nembledby tr.lnld service perlonnel.

(III Inembly D9url E-8.)

Turret Van. lorSlnllnll Thlt. POlltlon

FI,ure E·7. Tray base/turret.

Page 92: HP 7673B - S service manual

Theta·Flex Assembly Procedures

. ,,"r-: '

Fiaure E-8. Theta-flex assembly procedures.

J1 Connector

Drain Hola

Turn the tray base so that thedrain hole Isat the front.

Thread the J2 connector throughthe tray base.

Turn the turret so that the J1connecter Ison the left.

Unwind the theta-flex-3/4 ofaturn.

Place the square tab of the theta­flex within the ridges Inside thebase.

3

2

1

E·12

Page 93: HP 7673B - S service manual

Tray Errors

If the red LED on the front panel of the controller is lit or flashing, an error condition isindicated. Errors can be caused by:

• Objects obstructing the movement of the arm or gripper.

• Sensors not being tripped when expected.

• The tray cable not plugged in when the tray board is installed.

• Motor failure.

• Damaged wirlni or flex circuits.

Refer to Section B, "Troubleshooting,II to.888a fullllst of error conditions and courses. ofaction. When the fault is corrected, press the STOP button on the injector to clear theerror. Then restart the sequence.

E-13

Page 94: HP 7673B - S service manual

Tray Diagnostics

The Tray Diagnostics Mode allows the service engineer to observe the operation of traymotors and sensors during the troubleshooting process.

Follow these steps to take the Tray Module through its Diagnostic tests.

1. Tum the controller ofT.

2. Move the jumper on the tray board (figure D-12) to the TEST position. Note the fouryellow LEDs (R,Z,T,B) on the tray board.

8. Turn the controller on. The tray automatically starts the diagnostic tests.

4. Setup-The gripper moves to the home position (full up) and the Z-LED lights. The B(Bottle Sensor) LED will also light. The B LED Is lit only when the ¢pper I.empty and the ¢pper Is In the home position.

5. R-Cycle-The arm moves to the R-home position (full hack) and the R-LED lights.The arm moves one full cycle (out and back) and the R-LED lights againat home.

6. Theta Cycle-The arm moves out to the center row and rotates counterclockwise tothe stop pin on the theta-axis, then reverses direction and rotates clockwise, again tothe stop pin. The T-LED lights for half of the revolution and is ofT for half the revolu­tion. The arm then rotates back to the theta home position. The T-LED remains litthroughout the other tests.

7. Z-Cycle-The arm moves forward on the R-axis beyond the edge of the tray. The grip­per moves one full cycle (down and up). The Z-LED lights at the top (home position) ofthe cycle.

The tray repeats this pattern until the controller is turned ofT. To keep the tray in the sameaxis test, insert a vial into the gripper during the desired test. The tray continues to cyclethrough the selected axis until the vial is removed or the power is turned ofT. Note thatthe B-LED Is not lit while there Ie a vialln the gripper.

E-14

Page 95: HP 7673B - S service manual

Printed Circuit Board Component Overlays

.- 18598·B0075

oo

o o 0 I J1 I

J4 C::'j

f185116-80035

o 0

J4 ON REVERSE BIDE

Fllure E-9. Tray termination board (18596-60075) and. arm sensor board (18596-60035).

E-16

Page 96: HP 7673B - S service manual

Tray Pinout Diagrams

-eOPTICALSENSOR

JII

J4

J3

TRAYCABLE

1,14 GND 1,22,15 +12V 3,43 RHOME 154 -& HOME 715 -& SENSE 18 ZC 117 Z03 138 201 1151 RD1 1710 R04 1111 -&C 21

. 12 &2 23 J113 &4 2518 OND 817 ZHOME a18 BOTTLE 1011 Z04 1220 202 1421 RC 1822 R01 1823 Roa 2024 -&01 2225 &03 24 3 t-...IILlI------IIILIIr.IJ.IJ.I.-----J

r--.--_. ,

i r-~- --R- - IARM~ENSOR!O. 1~~, FLEX FLEX 3~~

I 4 3 3 J2 15~;.;-e2 15 2~~

, 1 7 4 12a~o.:-__-'l

I 1 J1 8 12a""""IIQ.III.I:I.-/'

'I J2:: - - - -J; - - -;- ·l10 4 3 3 I1 8

, 7 88 11""'-+----.

! .15 10 10 i,

I~:rr-"'"

'_._._._. _2. J.

Fllure E-IO. Tray pinout diagram.

E·18

Page 97: HP 7673B - S service manual

, I', ~:~: ,I - ~'.. ,

I RMOTOR

2 CONN

J1R

FLEX

11"""1-__-----,11 [!J[!]~~[!] II12 1!ll!l1!ll!ll!l10I J2I RI FLlX1-------',--------_...

DD

1 2

28

DC

DD

DD

DID

DD

lID tal

DD

IlIID

J5TERM

BOARD

r----------------I 21 12 I 21DD I ~~ DD 1 ~~

J2 DD I ~~ J2 J1 I~-.II(!)(!)TERM I~~I"" mm & eo l!lm

BOARD D a:D I I ~~ FLEX FLlX l!l [!]D 1m I 1 l!ll!l 1 [!][!lg 10 I I 10 II 10 1 10 III L. ,

I1 ]:: 21 2 l!ll!l 1 eo 1-------

J3 -- ~ [!][!l MOTOa! II a a a a a 1TERM 1m 1m I [!][!l cONN'10 a a a a a 2

BOARD7 D D 8 I 8 [!]~ 7 1 J1I '"""------' I SENSOR

[mI ,BOARD

J4 1 D D 2 L 2 ~~ 1 R I J2TERM l1:lil1:li ~~~ MOTOR ARM

BOARD5 D D 811

I [!]~ I CONNI SENSORI BOARD

iI~r !•[mDD I.-J......II

[!][!J& , DD ~~ l!l~

~ SENSOR I 2 D D 1 I 1 ~ [!]

I [!) CONN L J

r-­I 'III 1IIIIII1J1

~:~

25

Figure E-fl. Tray connector pinout diagrams,

E-17

Page 98: HP 7673B - S service manual
Page 99: HP 7673B - S service manual

Bar Code Reader Module

The next revision of the Bar Code Reader Service Manual will be in this smaller format.The current revision is available as HP part number 18587-90120.

F

F·1

Page 100: HP 7673B - S service manual

r . ~,J~ .', T :: .:; ,

Page 101: HP 7673B - S service manual

GCable Diagrams

SectIon G-Contents

Remote Cables .....•..•..•.................................... G-2HP 8896A <O3396-60550} G~2HP 8892A/8898A (18694-60500) G-8HP 5890A (18594-60570) G-4HP 5890A/8890A <l8594-60590} G-5HP 5700A SeriesJ8890A (18594-60600) G-6HP 5890Al18652A (18594-60610) G-7HP 5700A Seriesl18662A (18694-60620) G-8HP 5700A Series <l8594.6Q680} ....................•......... G-9HP 5890Al85900AIB/C (85900-60820) G·lO

BCD Cables .......••••••....••••....•..•.................... G·11HP 8896A (08896-60560) e-nHP 8392A/3393A (18594·60510) ..•.......................... G·12General Purpose <l8594-60520} G-18Nelson (19164-60550) G-14HP 85900C (85900·60850) •................................. G-15

Instrument Cables ...••••••.•........•.....•.....•............ G·16Injector Cable (07678-60645) G-16Tray Cable (1859B-G07l5) G·17HP 5880A Control Cable (19821·6051O) G-18

G-1

Page 102: HP 7673B - S service manual

",,' -.~

Remote Cables

HP 33Q8AENDHP 7873AEND

111r~g 2mr i m!!=!O!lIO 13- - - - - - ""----J

PIN# SIGNAL NAME WIRE COLOR1 START IN ORANGE2 GROUND BLACK3 JMPRTOtM BLUE4 JMPRTON3 BLUE& READY OUT REDe" JMPRTO#10 BROWN7 START OUT 1 GREENa GROUND BROWNg

10 JMPRT018 BROWN1112 READY IN WHITE

PIN # SIGNAL NAME

1 START OUT 03 START IN& READY INg GROUND

13 START OUT 114 READYOUT15 GROUND

WIRE COLORBLACKGREENREDBROWNORANGEWHITEDRAIN

Fipre 0-1. Remote cable. lIP SS96A (03396-80550).

G-2

Page 103: HP 7673B - S service manual

HP 7873A END HP 3392/93 END 12

'J1:"'~i=m..-J-l~------L~_

PINN 81GNALNAME WIRE COLOR

1 START IN 2 ORANGE2 START IN 1 BLACK3 UNUSED BLUE

• GROUND BLUE15 READY OUT RED8 JMPRTON10 BROWN7 8TARTOUT1 GREEN8 START OUT 0 BROWNg

10 JMPRTOft BROWN1112 READY IN WHITE

PINN SIGNAL NAME WIRE COLOR

15 GROUND DRAiNII STARTOUT1 BLACK3 GROUND/START IN0 BROWN

10 READY IN RED2 8TARTOUT2 ORANGE

11 START IN GREEN9 READY OUT WHITE

Fipre 0-2. Remote cable, lIP 8892A13393A (18394·60500).

Page 104: HP 7673B - S service manual

HP GaiDA END

~ ~2.1

~121111

HP 7&73A END

111'~I = 11St4-1ll570I 2-

NOTE:Splde lug connection mUlt beInechedtoground It the HP 71'173A for properoperltlon.

PINN 81GNALNAME WIRE COLOR

1 START IN ORANGE2 GROUND BLACK3 JMPRTOlll4 BLUE4 JMPRTON3 BLUEIi READY OUT 18 JMPR TONtO BROWN7 START OUT 1 GREEN8 STARIOUTO BROWNI READY OUT1

10. JMPRTOlte BROWN11 GROUND12 READY IN WHITE

PINN 81GNALNAME WIRE COLOR

1 STARliN GREEN2 GROUND BROWN3 JMPRTO*, BLUE4 JMPRTON3 BLUEIi READY OUT 1 WHITEI JMPRTONtO BROWN7 START OUT 1 ORANGE

• START OUT 0 BLACKI READY OUT 1

10 JMPRTO'" BROWN11 GROUND12 READY IN

Figure G·S. Remote cable, HP 5890A (18594.&0570).

Page 105: HP 7673B - S service manual

PIN#, SIONALNAME WiRE COLOR

1 START IN ORANGE2 GROUND BLACK3 JMPRTO*, BLUE4 JMPRTON3 . BLUE5 READYOUT1I JMPRTON10 BROWN7 8TARTOUT1 GREENa START OUT 0 BROWNII READY OUT 1

10 JMPRTOA4\ BROWN11 GROUND12 READY IN WHITE

PIN#, SIGNAL NAME WIRE COLOR

1 START IN BLUE2 GROUND RED3 JMPRTON4 BLUE4 JMPRTO-.. BLUE5 READY OUT 1 . WHITE• JMPRTON10 BROWN7 BTART OUT 1 ORANGE

• START OUT 0 BLACKII READY OUT 1

10 JMPRTOfllB BROWN11 GROUND12 IlEADY IN

HP IiSIlOA ENDHP 7873A END

'lgf~'-liL......:.:==-,J-Its- _..... _ -:..~_~ 2 111 'g2~ »> ~,211"

I

-, --s OO=m§J~.2

HP 33110AEND

PIN#, 8iONALNAME WIRE COLOR

1 STARTOUl BLUE2 ITARTIN1 GREEN4 GROUND RED7 START IN 0 BROWN

Fisure 0-4. "Y" remote cable, HP 5890AJ8S90A (18594-80590).

G-6

Page 106: HP 7673B - S service manual

ORANGEBLACK PIN# SIGNAL NAME WIRE COLOR

A READY OUT 1 REDDRAIN E READY OUT 0 BROWNRED E START IN 0 BROWN

E START IN 1 GREEN

PIN # SIBNAL NAME WIRE COLOR

1 START IN 22 OROUND BROWN3 UNUSED• GROUND5 READYOUT16 READY OUT 07 START OUT 16 SlART OUT 0II READY OUT1

10 START OUT 011 GROUND12 READY IN

I

-; - - -s OO=I[g1]~.2

111l1UOlOO

HP 7B73A END HP 57XXEND... .

'I:~HI,,"-===-J-'ts--....-__....--::.._iC'·L--.....~....,~ ~

-HP USDA END

PIN* SIGNAL NAME WIRE COLOR1 START OUT 02 STARTING• START OUT 17 START IN 1

8HOWNBLACKGREENORANGE

Q-8

Fiaure G-~. "Y" remote cable, HP ~70OASeriell/3390A (18~94-80600).

Page 107: HP 7673B - S service manual

I"":.,,";''''·''" ',"\"

PINN SIGNAL NAME WIRE COLOR

1 START IN ORANGE2 aROUND BLACK3 JMPRTO#4 BLUE4 JMPRTON3 BLUEII READY OUT18 JMPRTON10 BROWN7 START OUT 1 GREEN8 B!ABTOUIO BROWNg READY OUT 1

10 JMPRTOfMI BROWN11 GROUND12 READY IN WHITE

PINN SIGNAL NAME WIRE COLOR

1 START IN BLUE2 GROUND RED3 JMPRTO#4 BLUE4 JMPRTON3 BLUEri READY OUT 1 WHITE6 JMPRTON10 BROWN7 START OUT 1 ORANGEI START OUT 0 BLACKg READYOUT1

10 JMPRTO_ BROWN11 GBOUND12 READY IN

---

HP SlgOA END

I\

-------~- 21'_o....:.:;=;';':"'J-# ,... _ :

,... - -'2 "NOTE:

Spade lug connection mUltba attachedto ground It the HP 7873A lorproperoperlllon.

HP 7673AEND

,,l1If § GW!J.2 =

HP 18652A END

PINN SIGNAL NAME WIRE COLOR

B 8TARTOUT1 GREENB 8TARTIN BLUEC 8TART OUT 0 BROWNC OROUND REO

Figure 0-6. ''Y" reinote cable. lIP 3890A/18652A (18594-60610).

G-7

Page 108: HP 7673B - S service manual

PINN SIGNAL NAME WIRE COLOR

1 START IN 1 GREEN2 OROUND BROWN3 BLUE5 READy OUT1 GRAYe VIOLET7 START OUT 1 ORANGEe START OUT 0 BLACK9 READY OUT1 YELLOW

11 DRAIN12 READY IN RED

PIN N SIGNAL NAME

B STARliN 1A READYOUTE GROUNDE START IN 0

WIRE COLOR

ORANOEREDBROWNBLACK

11'~: I2 -

HP 7873A END

HP87XXEND

19t1=!O!2O ts-- - - - - - - - - - _ ,I--­.".~.""".~

""".~---- ,..r \--CDl-_--liS=.tr~- r@\ ....\ ~

,----- I!:: ~ = r@...,

HP 1ae152A END

PINN SIGNAL NAME

C GROUNDB START IN

WIRE COLOR

BROWNORANGE

c-a

Fisure G-7. "Y" remote cable, HP G700A Seriel/18652A (18594·60820).

Page 109: HP 7673B - S service manual

HP 1673AEND

PINN SIGNAL NAME WIRE COLOR PINN SIGNAL NAME WIRE COLOR

2 GROUND/START IN BROWN A READY OUT RED7 START OUT 1 ORANGE E START OUT BROWNa START OUT 0 BLACK E START IN 0 BLACK

11 GROUND DRAIN B START IN ORANGE12 READY IN RED

F1aure G·8. Remote cable, HP 5700A Series (18594.f108SO).

G-9

.'":.' II"" ";".'

Page 110: HP 7673B - S service manual

PINN SIGNAL NAME WIRE COLOR

1 GROUND BROWN/RED5 .lUITON I/O GREEN/BLUE8 paN ORANGE7 ~ WHITE/BLACK

alOlOa12 2•

HP 5890 END

PINN SIGNAL NAME WIRE COLOR1 START IN ORANGE2 GROUND BLACK3 JMPRTO#4 BLUE4 JMPRTO#3 BLUE5 READY OUT18 JMPR T0610 BROWN7 START OUT 1 GREEN8 START OUT 0 BROWN9 READY OUT 1

10 JMPRTOltl8 BROWN11 GROUND12 READY IN WHITE

PINN . SIGNAL NAME WIRE COLOR1 START IN BLUE2 GROUND RED3 JMPRTO#4 BLUE4 JMPRTO#3 BLUEG READY OUT 1 WHITE8 JMPRTOH10 BROWN7 8TARTOUT1 ORANGEI BTABT OUT 0 BLACK9 READY OUT 1

10 JMPRTO_ BROWN11 GROUND12 READY IN

HP 7673 END

lB'~:: . i moo.tOI20 ts - - - -, .... _ - --2 - \ /

, I

IIIIII

I

Figure G-9. lOY" remote cable, HP 5890A/35900A/B (35900-80820).

G-10

Page 111: HP 7673B - S service manual

BCD Cables

o ~ HP 7873A END HP 338SA END,

"A (~I:::T--- - - ..

i °0 .~ .'~: ~j-- .'.'.'.'....Q... ( e zr' ( .:i 0

PINN BIGNALNAME WIRE COLOR PINN SIGNAL NAME' WIRE COLOR

1 BCD fi GREEN 1 BCD fi GREEN2 BCD 7 VIOLET 2 BCD 7 VIOLET3 BCD II BLUE 3 BCD II BLUE4 BCD4 YELLOW 4 BCD4 YELLOWfi BCOO BLACK 8 BCDO BLACK8 BCD 3 ORANGE II BCD 3 ORANGE7 BCD2 RED 7 BCD2 RED

• BCOt BROWN 8 BCOt BROWNI GROUND GRAY I GROUND GRAY

Figure 0-10; BCD cable, lIP 3396A (03396-60560).

0-11

Page 112: HP 7673B - S service manual

HP 71173A END HP 3392/3393A END

0

~~~j12

qp----~~-1I5M-t0510

16

0

PINN SIGNAL NAME WIRE COLOR

5 BCOD BLACKa BC01 BROWN7 BC02 REOII Bcoa ORANGE4 BCD4 YELLOW1 BCD 5 GREENa BCD II BLUE2 BCD 7 VIOLETa GROUND GRAY

PINN SIGNAL NAME WIRE COLOR7 8L1 BLACK4 al2 BROWN

12 8U RED5 ILl ORANGE9 8M1 YELLOW

1D 8M2 GREEN3 SM4 BLUE

11 8MB VIOLET1 GROUND GRAY

G-12

Figure 0·11. BCD cable. lIP 3392A!3393A (18594·60510).

Page 113: HP 7673B - S service manual

HP lOl3AEND

0

:[{~j.- 1m15

0

PINN SIGNAL NAME WIRE COLOR

1 BCD I GREEN2 BCDl VIOLET3 BCD 0 BLUE4 BCD4 YELLOWI BCD 0 BLACK0 BCD 3 ORANGE1 BCD 2 RED8 BCD 1 BROWNI GROUND GRAY

15 +5VDC WHITE

Figure 0-12. BCD cable, aeneral purpose (18594-60520).

G-13

Page 114: HP 7673B - S service manual

HP 7673A END NELSON END11

0

:[~j,..- ~QI~ s15

0 - . 1

'....,

PINN SIGNAL NAME WIRE COLOR& BCD 0 BLACK8 BCD 1 BROWN7 BCD 2 REDI BCD 3 ORANGE4 BCD4 YELLOW1 BCD II GREEN3 BCD 8 BLUE2 BCD7 VIOLET, GROUND GRAY

PINN SIGNAL NAME WIRE COLOR1 BCD 0 BLACK2 BCD 1 BROWN3 BCD 2 RED4 BCDs ORANGE& BCD 4 YELLOWI BCDS GREEN7 BCD I BLUE8 BCD 7 VIOLET

27 GROUND GRAY

G-14

FiJUl'e 0-13. BCD cable, Nelson (19164-60330).

Page 115: HP 7673B - S service manual

-:T'""T""" , ~,!0~~:'!~::'I';P'~"~ , "-'~,,"~,.;~,~:,~\ ' ",,'., ""~"",",'i~"""~~""""~~''''"",·"q,\\,,,'1',· ..,,,,,'"'~ 1'l"'¥'fIiF.'J!,!"~~ d~,,~'l""~~",,!,,":,'I-~,~r' ';. ~ J":?' :t~: ,';":J"~I't~-<~I..r:rJ,"fi, (,"~'(\~

HP 7673 END HP 351100C END'()" Ie~:>~~~~o~---- <~):

...--15 ri!'••

~o· 600 ·0•• 00

g~ :0 g

0 '---

PIN# SIONALNAME WIRE COLOR PIN # SIONALNAME WIRE COLOR1 BCDli GREEN 1 GROUND GRAY2 BCD7 VIOLET 2 BCDO BLACK3 BCDS BLUE 3 BCD1 BROWN4 BCD4 YELLOW 4 BCD2 RED15 BCDO BLACK II BCD3 ORANGE8 BCD3 ORANGE S BCD7 VIOLET7 BCD2 RED 7 BCDI5 GREEN• BCD1 BROWN 8 BCDS BLUE

.lJc GROUND GRAY.lJc BCD4 YELLOW

NIC WHITE N/C WHITE

Figure G-14.· BCD cable, lIP 35900C (35900-60850).

Q·15

Page 116: HP 7673B - S service manual

.1,~,..:'-.':iB";~:~·I",~.I:\;::'\"" -.'I "

Instrument Cables

211

20 10

CONTROLlER END

PIN/f SIGNAL NAME WIRE COLOR

1 GROUND BLACK2 +IIV SUPPLY WHITE3 +18 VSUPPLY BROWN• +18 VSUPPLY WHITEII' 18VGROUND RED• 18VGROUND WHITE7 -18 VSUPPLY ORANGE8 -18 VSUPPLY WHITE8 POP YELLOW

10 POP WHITE11 SERIAL COMM OUT GREEN12 SERIAL COMM OUT WHITE13 5VGROUND BLUE14 START IN WHITE111 IIVGROUND VIOLET115 START OUT WHITE17 IIVGROUND GRAY18 READY IN WHITE18 IIVGROUND BLACK20 READY OUT BROWN

INJECTOR MAIN BOARD END

PINN SIGNAL NAME WIRE COLOR

1 GROUND BLACK2 +11 VSUPPLY WHITE3 +11 V8UPPLY BROWN• +18VSUPPLY WHITEIS 11VGROUND RED1 18VOROUND WHITE7 -11 VSUPPLY ORANGE8 -11 VSUPPLY WHITE8 POP YELLOW

10 POP WHITE11 SERIAL COMM OUT GREEN12 SERIAL COMM OUT WHITE13 ISVGROUND BLUE14 START IN WHITE15 ISVGROUND VIOLET11 START OUT WHITE17 SVGROUNO GRAY18 READY IN WHITE111 eVGROUND BLACK20 READY OUT BROWN

Fipre G·l~.· Injector cable (o7878·8064~).

G-18

Page 117: HP 7673B - S service manual

t-..cml

PINfi SIGNAL NAME

1 GROUND2 +12VSUPPLY3 R-HOME4 mETA-HOME15 THETA-SENSEa Z-MOTOR COMMON7 Z-MOTOR038 Z-MOTOR 011 R-MOTOR02

10 R-MOTOR0411 mETA-MOTOR12 mETA-MOTOR 0213 mETA-MOTOR 0414 GROUND15 +12VSUPPLY18 INTERLOCK17 Z-HOME18 BomE8ENSOR11 Z-MOTOR0420 Z·MOTOR0221 R-MOTOR COMMON22 RoMOTOR 0123 R·MOTOR 0324 THETA·MOTOR 01215 mETA-MOTOR 03

PlNfi SIGNAL NAME

1 GROUND2 GROUND3 +12VSUPPLY4 +12VSUPPLY5 R·HOMEa GROUND7 THETA·HOMES Z·HOME.. mETA-sENSE

10 BOTTLE SENSOR11 Z-MOTOR COMMON12 . I-MOTOR0413 Z-MOTOR0314 Z·MOTOR 0215 Z-MOTOR0118 R·MOTOR COMMON17 RoMOTOR 0218 R-MOTOR 0111 R·MOTOR 0420 R·MOTOR 0321 mETA-MOTOR COMMON22 mETA-MOTOR 0123 THETA-MOTOR 0224 mETA·MOTOR 03215 THETA·MOTOR 04

Page 118: HP 7673B - S service manual

HP 7673A END

:~~J""-" C~]:15

1::3 SHi

0 0

PINN SIGNAL NAME WIRE COLOR1 GROUND BROWN2 GROUND RED3 TRAYXMIT ORANGE4 INJ 1XMIT YEllOW5 SYS READY 2 GREENIi INJ 1 RCV BLUE7 INJ READY 1 VIOLET8 INJ READY 2 GRAYII GROUND WHITE

10 aROUND BLACK11 GROUND WHITE/BROWN12 INJ2XMIT WHITE/RED13 SYS READY 1 WHITE/ORANGE14 INJ 2RCV WHITE/YELLOW11 TRAYRCV WH1TE/BLACK

PINN SIGNAL NAME WIRE COLOR1 GROUND BROWN2 GROUND RED3 TRAYXMIT ORANGE4 INJ 1XMIT YELLOW1 8Y8READY2 GREENIi INJ1 RCV BLUE7 INJ RUDY 1 VIOLETI INJ READY 2 GRAYII GROUND WHITE

10 GROUND BLACK11 GROUND WHIMROWN12 INJ 2XMIT WHITElRED13 SYS READY 1 WHITE/ORANGE14 INJ 2RCV WHITE/YELLOW111 TRAYRCV WH1TE/BLACK

Q·18

Figure 0-17. HP 5880A control cable (19321·80GIO).

Page 119: HP 7673B - S service manual

lhR)SALES and SUPPORT OFFICESArranged alphabetically by country

HEADQUARTERS OFFICES ARGENTINA BARBADOS/WESTH.wl.tt-P.ekard Argentln. INDIES

If there la no ..lea offloe lIated 'or your are., S.A. Computer and Control.contaot on. of theae headquartera offlo.a: Mont.ne••• 2140/60

~B'dcIJ L1d.1428 ~ENOS AIRES ult. ,l.t Floor6:.': 1-11-1441 ~'deJ 8hotng Plaz.AI"'RICAAND UNITED KINGDOM ~A~~~ . ICHA

T.I: (eoe~ -72&4MIDDLE EAST Hewl.tt-P.ckard Ltd. T.I.x: 38 2~Hewl.tt-Packard S. A. , Nine Mil. Ride COMPUTROLSouth Il••t lurope 81.1•• WOKlNQHAMAUSTRALIAR~onBr~ BIIrklhlre, RGl13LL

MI I.E.£r~ T.,: 0344 n3100 ~~Ide, South Au.tr.lI.BELGIUM~, rue du , T.'p: 848805/8488141

,o.~x~ 848lil12 Hewl.tt-P.ckard Au.tr.lI. Hewl.tt-P.ekard BIIllillumCH-121 In 1, Ltd. 8.A./N.V.G"""'., Sw mrland 1~=lIlRoad 81vd de I. Weluwa, 100t.lp: 481110 me. eh

ELSEWHERE IN TH!~ KS ,S.A. 5OlI3

t.QllOO1~1~bJim~1A12 .1: 111=2-6811 1200 USS!LS

WORLD b·,ax i .1: (~32-2-7111-31-11able: WPARD

ASIA Hewlett-P.okard AdeI.1de Telaxl 114 HEWPAOInteroontlnent.1Hewl.tt-P.ckard All. Ltd. 34lil5 Deer Creek Road Brl.bane, Q~"and30th FlOCII' PALO ALTO, CA ll4304 OftlQe BERMUDABond Center W..t Tower USA Hewl.tt-Packard Austr.lI. Applied CQrnputerlllil Queentw.~ T.I: (416) 867-1601 L~dfentr.'~ 1 P~ROad Techl'lCll~

al: 6-648- T.,.x: 034-8300 Atlantlo .. BuildingT.lax: 7117lil3 HPA HX Oabl.: H!WPAOK THE , Q..-n.'.nd P.r.......-VlIl. Re.dO.bl.: HEWPAOK 40111 P.O. ~x HM 20(11T.,: 111-7-3Q0..4133~~1I87 6.'.x: 42133 HAMIL ON II

UNITED STATES 01"' T.': ~1I:Ueabl.: HEWPARD T.,p: II/ACT BACANADA AMERICA Brllbane

Hewlett-Packard (OM.d.) Ouatomer information Melbourne, Vlotori. Offlr.kA~'7 gr~ DrIVe

Center ~tt..p.ckardAu.tr.1 • BRAZIl.~800) 71!2..()ll(lO td. Hewlett-Packard do Br••11MISSI A , Ontario :00 AM te 6100 PM ~1-41a Str..t B.A .L4V 1M8 P.o1110Time .0. 1 Alarned. RIQI\IeljlroT.I: (4~-ll43O i~O , Vlotori.T.,p: 711O-l.AND.ALPHAVlLLE

EASTERN EUROPE EASTERN USA : 111-3-l1l111-21l1l011400 garuer~ S.o P.ulo6·laxI31~4

Hewlett-Packard Hewl.tt-Packard OClmpany able: HEwPARD T.I: ( 11\ 4 1-1311G••. m.b.h. 4 Ohok. C~Road Melbourne ~~Pl (0 1) 71361 HPBR

~Iablgc" ! ROOKVILLE 2OlI8O~, New South W.I•• ~abl.: HEWPAOK S.o. 0. x 7 T.,: (301) ~70 auloA-1222 VIE! NA AuI1rI.

T.,: (~~~ Hewl.tt-P.ckard Aultr.lla Hewl.tt-P.ck.rd do Br..11L~d.T.I.x: 44 H!PA A

MIDW~ST~RN USA 1 -23 T.ta.,.,. RO.d ~A.al. de 1!I0t.f~,NORTH~AN !UROPE Hewlatt-P.ekard Oornpeny NOFITH RVOE, N.S.W. 228-A-II14. ANDH.wlett-Packard 8.A. 11201 Tollvlew Drive 2113

.~~NJ.1I01v.g. ~'aan 241 ROLLING MEADOWB, a. T.l: 01-2 ... ~~H Ed t 010 Ar~tln.-AI. AP.. x 0118 lloooa T.lp:. U\16l11

~ - RI OE JANEIRONL-11~~t T.l: (312)~ O.bl.: IiWPARO SydMy~ST L T.I: 6lI2-l1422haNat~= b.l.x: ~HPB\~Rl.1i 20 7 lil SOUTHI!RN USA AUSTRIA .bl.: AOK 0 de.'p: 1l1l1111 hpner Hewl.tt-Packard J.nelro

SOUTHEA8THewl.tt-Pack.rd Company Ge•. m.b.h. Van Derl Clentllie. Ltd•.2000 South Park "'.(1& Llebllil·..·lEUROPE ATLANTA, QA 3033lil Poa. da. Vlel.t.., 100

P.O. Boxl2 Edlflelo ConvexHawl.tt-P.ckard 8. A. T.,: (404) lilM-1sao A-1222 VI NNA OliP 011400 O.ntroWgrld Tr.de Cent.r T.I:~~ ~cl.' de Alph.vlll.11 Avenu. LouI.-o...1 Telp: 1 PA A' .........1 SaQ Paul~razll1215 COlntrtn, GENEVA WESTERN USA .1: (011)421·1Swltzerl.nd

T.II (OfilJl lIlI 61 Hewl.tt-Packard Comp.ny van Den Olentltlo. Ltd•.T.,p: ~_ 61111 L.nkerlhlm Blvd. Ru. ~a Bonlfaele, 468Mall Addr."l .0. Box NOATHHOLLYWOOD, CA TodcI • S.nto.CH-1217 =,n 1, lill1101 20771 RIO DE JANEIRO AJGENEVA u.rland T.I: (818)~ T.I: (om6113~

T.,.x: 87! L BR

Page 120: HP 7673B - S service manual

P£)SALES and SUPPORT OFFICESArranged alphabetically by countryCANADA CHINA FRANC!!! GERMAN ...eDi!!RALAI~rt. Chin. Hewl.tt-P.okard Co. H_l.tt-P.ok.rd France REPUBLICH.wlett-P.okard (C.n.d.) ~td. Zone lndu.trl.lI. de Hewl.tt-P.ckard GmbH~' .O. I!lox 1MI10 Court.boeut G••Oh.tt..t~11eo 3rd Avenue N. E. 4th Floor, 2nd W.toh Avenu. de. Trr~. ~..eraohm tt.tr.... 7¥ALGARV, Albert. F.otory M.ln Q1Q47

MESlAJC .x 0- ;10 NI!!U ULM2A lST7 ShU~ Vu Shu, B.I 8an tOr•• t.l: 4Q~731-7O-73-0T.I: (403) 235-3100 Hu.n o.d .1: -6-Q077ll26 .I.x: 0 12 llS HP ULM-OH.I Ol.n Ol.trlot T.I.x: 8000481"

,.Ijl'lo Hewl.tt-Paok.rd GmbHHewlett-P.ok.rd (C.n.d.) .1: 231471 Hewlett-P.ok.rd France V.rtrl.b.zentrum A.tlngenLtd. T.I.x: 22601 CTSHP ON Paro d' .otlvlt.. de ,.

~~.fl."~~&1J~~b~8th8tr~ PoterleON, Al • Chin. Hewl.tt-P.ok.rd 00. ~"i!I~rt~~utel-l!lotmel Poatf~h 31 12

T581~ Ltd. Tel: (210264Q4-oT.l: ( ) 4M-eeee CHP Ih.n;h.1 I!lranoh T.I: ~'~244 T.lex: 58Q 70 hpr.d

23/1" h.ngh.1 Union T.I.x: 7 12FB~dlng Hewl.tt-Paokard tmbH

erltl.h COlumbl. 1 V.n An Rd Eut Hewlett-P.tard France Vertrlebazentrum uohen.fIClClUVer Are. ¥HANGHAl Eaohen.tr.... IIHewl.tt-P.okard (O.n.d.) .1: 285550 L. "'erlpol.

~lS TAUFKJACHENLtd. Tel.x: 33571 OHPSB ON 3, Chemin du Plgeonnl... de .1: 4Q-8Q-lS1-2070'~~Shellbrl~ W.y C.bl.: 34111 Shangh.1 I.0l'fl.... .Iexl 052411115 hpmohAl OND, ISrtl.h 3108 TOULOU8~ Oe<MxColumbl. T.I: ~'~o-' 12 Hewl.tt-P.ck.rd GmbHVlSX2W8 COLOMBIA T.I.x: 183QI"

Onoh.tt..tllll.T.I: (lI04) 270-.22n C... Clentl110aBlanoo y fr~l.all..

gl•. Sen C GERMAN FED!!!AAL 81 W~AONN2.11. 30A Nll-.22 Of. ao-oa A!!!PUBLIC Poattaoh 1

~Itoba BOGOTAHewl.tt-Paokard tmbH Tel: (O~~802-o

I.tt-Paokard (Oan.da) T.1: 281-7368, 232~llSVertrlebazentrum Itt. T.I.XI 7 hepk

Ltd. . T.I.x: 44QlI2B CXOBOOOHewl.tt-p:r~[Mcr~M"Blvd. GREAT SAITAIN, Manitoba Sol.r I!!leotronIO.~•• ~J172~OO-O S•• Unlt.d KIngdom

~ '~) lSfI4-27n ~all. )f5A No. Telex: 410 hpbhgOGO A GREECET.l: 2.63-QlSQQ

T.I..: 44Ql0 OXUBO 00 Hewl.tt-P.okard GmbH Hewl.tt-P.okard A.I!!.Hewl.tt-P.okard (Oan.d.) G.aoh.tt..t.lI. 32\ Klfl..,•• Avenue

k11~ld 8trHt~~t~r"WLt-430 1~Marou"l

DENMARK A NS, Gr_~ N, Ontario Hewl.tt-P.ck.rd A/B T.I: (~~~04-0 T.I: (~1) 118211811

~v~en25T.lex: 1 hpbIn d Te'-x: 1 218 HPHL

Tell (51Q) lI88-Ql111 OK lS ~IAKI!!PlO~tel: 46 2"'1-8l14 Hewlett-P.okatdQm~ Keatu Kar~r,::I. B.A.

.1..: 3740Q hp•• elkVllrtrl=~ "~,!{~?btTorontQ Are. ~~ar8~&a NHewl.tt-Paokard Oanad. tell 32=R5J7 371

~d.,.INLAND Poatt.m 142 elex: 21 GA

77 y Drive Hewl.tt-Paokard 9Y T.I: (7031) lI4S-o8S1=<u', Ontario ~&ank.lllonl.17 T.lex: 72e5 743 hep Hell.mooL4V 1MB E8P~0 PM'o ~x 111628tel: (4~ 818-l143O T.l: (QO~ 7-21 HewI.tt-PaQk.rd GmbH

1 7 RAI!!US.1..: -83e44 T.Iex: 1 15113 HEWPA 8F Tell 4'104QGeach.tt..t.U. T.lex: 41441

8ohIHf.~

Hewl.tt-P.okard (Oan.d.) PRANCE D-4eOO 1 GUAT!!!MALA~d.

HewI.tt-P.Okfd Fr.nce Tel: (om~1PE8A~Gfc.tem~l.

70 Qu..n.vlew Drlv. Z.1. Merour. T.l.x:Avenld. • erm. -48

OTTAWA, Ont.rlo A~ Berthelot

~~~'l~ iW"e63K2B 81(1 13 1I3L•• MUU. Oedex Hewl.tt-P.okard GmbHT.I: (1113) lI2O-&4l13 AlX-EN-PAOVENCE Vertrleb~trumNord

t.l: 33-4a~Q-4102 ~dt~ ~11!Zoa.I.x: 41 OF URGlIO 11 -2-318827Quebec/Montr••1Ar•• t.l: 4Q-4~ Tel..: 30657l16 IPE8A GUHewl.tt-Paokard (O.n.da) Hew1.tt-P.ok.rd I"ranee .I.x: 021 032 hphh dLtd. ~mln dee MouIIlQ HONG KONG17600 Tran.-oanad. Hwy Ite Po.t.l. 1112 Hewlett-Paokard GmbH Hewlett-Packard A.la Ltd.South 8erv1~ 8;131 I!!CULLV Cedex Geaoh.tt..t~ ~F" W..tT_KJA~, t~k-7~-3W12lS Helderl~ CentreHQJ X8 D-3OOO YER 81 ~QUMnaWayTel: (5~ IIQ7-4232 T.lex: 31 l1F T.I: (~) 570lI-o tri7nHong KongT.'-x: 41621 T.I..: 32lSQ hphan 5-&48

Hewlett-P.ok.rd Fr.nceICELAND

CHILE Bur••u vert du Bol. Brl.nd Hewl.tt-P.ckr.d GmbHgheman de I. Garde Geaoh.tt..t.1 • =tett-Paokfd Ioel.nd

~Ing Y CI•. '5d. P 212 212~~~

Id.b.kk.H anoa 1373- 1I0ina 1311 44~ NANTES Cedex 1~ 'ID'KJAVlK

'~?a~, 72-1716T.: (40) ~-22 el: 4 1-70-0&-0 T: -1"'7-1000Telex: 711 .I.x: ~10l1 hpmhm T.I.x: 3140Q

2

Page 121: HP 7673B - S service manual

"", I~ ' "I~ .'~;II

Page 122: HP 7673B - S service manual

[h£)SALES and SUPPORT OFFICESArranged alphabetically by countryPHILIPPINI!!:S "8Q30 L.. Ar~a. UNITED ARABAnalytloal T.chnologla. & aueoho (Vlmay.) EMIRATES~vda. daJ~arta, 8 UNITED STATESR"'~OhCorP.

~ dltlolo • "As ~mltao Ltd.ACC A Bldg.• Room 201al: (lol<I) ,,&4 32 '''&4 2\) .0. Box 1&41 HeWlatt-Paokarcl Compd::-Ytaamboa Street, COrner

8~AH Ou.t~ Intorm. t lon t.,.Saloado 8tr"t 33Tal: 01181 Tal: ( 110 62-0000Leg••pl VlIlaoa M.katl • Talax: 33032Talax: ea~ ~~..MJM Hourlll~: am - 6:00 pmMatro MANILA

SW!DEN Oabla: E A AH Paello lmeTel: 8180103, 8183873,81&4230 aot~buri Ara.

UNITED KINGDOM ALABAMATel~ll: 23tee PAW PN Hawtatt-P okard !I_lga A8Hawle~l-Paokard ~anv8305 PAW PN T~"Batan 1A GREAT BRITAIN roo ~ C'1.a Cenl«P.. oXJl: Hawlett-Paokard Ltd. UI~~ uttel1SAUDI ARABIA ~2123 ASTRA

~ar Houle lR ~36a44thenbu~ RI~ B~KNELL al: (205) 7Modem I!leotronlo el: (031 8Q 0 00~. R 121Elt.bll.=t telex: ( ) 1~1l88 (vi.

al: ('1re~ "2"au ARI~ONA~Iett 'f,ard Olvl.lan Spang. onlool) Telax: 833lett-P~~~~21 SWITZERLAND o~ta a=x a.t

AL;:<1~:'1'I31Gfl2 Hewlatt-P.ok.rd (Sohwalz) Hewlatt-P'ok.r~Ltd.el: eozl:f3-8000Tal: -17r& 805-17&4 Aa He.th.lda Parte cad

Telex: Vl ¢ HPMEEK 8J glar..tr.... 12 Cheadle He.thOAlJl"ORNlAOable: LE TA H-<IOl57 BASEL gTOCKPORT=-~""'ckard=anvAL-KHOBAR Tal: 081-881-::-13 helhlr~ 1:; . MarteetPax: 081-e81

E~~0828 ItlJ:lA,AO AMn~ OA G683<I

Modem ElectronloHewlett-Paokard (Sohwall) alex:

el: (018 7'222E.tabll~t Aa~'ett .trd Olvl.lon Allmend 2 Hewlett-P.okard Ltd.

Hewlatt~~~.0. ~x t

fH-8G67 WlOg';N lu.tomer Sart Cent. ~ 800tt levardR~ laza, 8th Floorell (OO<Iill7 3121 11 .kd.le Ro ANTA ,A ll8O&4fiO AH

Telex: W 'l!l'O oh Wlnner.h, WOKlNQHAM t:l~:("i,bdel: $4<1 Gt21Cablal AQ H

trk'hlr13..R~Telall: 1027 ~ ~~NAS 8Jel: (0 ~Oable: LEO A J DOAH

Telex: ...OOLOR~TAIWANHewlett aokard 0c:!nPanYModem llleotronio Hewlett-Paokard T.lwan SCOTLAND ~Cent« Green 00urtE.tabll.hmant ktd. h ulte A

~tt-l"ack~H~ R 0080301el: (=\ "lIQ-UU

Hew~'ok'rdOM.lon th Floor Hawlett-P.okarclSO H iP.O. 22016

~"dl~ wrrJf~n, ~ aRIY . l1"Gfl

~u .Ing North RoadTal: ""1-07 18, "1l1~ flJ T I te: 1331 1

~tt-fDack'rd~Telex:2020040 MEER T.I: (,) 712-G40<1 e ex: 72882/3

~~cJ,6lJ&J '\2Telex: ""30 HEWPACK

SINGAPORe Cable: Hl!WPACK Taipei URUGUAY Tell 3(3) &4Pablo~~ 8.A.O.e.l.Hawtett-P.okard Slng.pora

~ L1hTr'fl~ ~pany Avenlda • la 2an OONNEOTIOUT~8~pte. Ltd. 7t ~Ioor, 1 . 1

~~370 ~.tt-l"'akard Clompany1 gu Road Chu~ H~ao ••t Ro.d

~~~t'tR=d'fl ; O<ItO fA 11 Tel: 8 -aMTal: (203) 7fril

al: "rtt 18 el: (on 3lol<I-8101 Telex:Telex: ~HPSa80 RS Talax: 8lol<1 SANKWANaO.ble: ACK,THAILAND VEN!ZU!LASlng.pore

UNITeD STAT!'~Ima•• co. Ltd. An'~IO~ liJupplleI O.A. ~ORIOASPAIN PatllOllll Avenue A~ da I centro~ttt-P.okard~

g~~I0500 aunt. 103~1:. ake~~ve!ll23O La. R~"~'drIdJoo La. OholTa..rret•• N- . 18

Tel: 7~710 23<1-oGlll/3 ~ARACAS Tel~85Tel: (01) 837 bO tl1837 81Telaxl ~l!llmonooTH ell ~QO<I,~781 Cable: SA Bangkok Telex: 7" 0 VO

~tt..p:.1ard CompanyT.lex: 23818

.. B.~ IYd.TRINIDAD. TOBAGO V!!NEZUI!LA j:0nt.) ~11d~~~802\) 8aroalona Oomputer and control. Ltd. N A OLA ~Hawlett-P~kardTel: ( ) "tfS-&4roantenz. 1 W~ook

V~'•.A.el: (U3)~ 2" 61132173 P.O. !il3A r.n.var..1Lee Ruloat

QEORQIA64 1 T.~or Str"t

=1o~&3Telex: 52e03 PaR -oF.~PAlN

~'okard~T.I: 822-7 10 82-27085 Apart.do 2016 h Park l.oaTelex: 387227b8 COMOONA A GA~"1008 thvIU. CARA= 1tal: (~l~l

WQTel: ( )~~iUI'1I Mor.i..Telex: 25t AOK elex: 1 88--48GOIN ( dltlolg Forum) TURKEYTell ("'~17' 00 I,.t lno.YUGOSLAVIA HAWAIITelax:

~tepe. aueteoll« Sit,.I Hewlett-P.okard OompanyKe.kln K.lamDo~' P~III;T_"8~ V,lenol. Sakak 8/3 a.yrettape aener.1 de\"8t... 1 1 ~ 8treetYU-11 QRAe

~I.. I. CnOIlG' 8 ISTANB~ •

O~U HI He13T": (O~ "" Tal: 17 1328, t73 33<1<1 T'l: (011) 3<12-&41Tel: ( )1s2i-1N!iTelex: Telex: "2"go Telall: 11<133

4

Page 123: HP 7673B - S service manual

[hPlSALES and SUPPORT OFFICESArranged alphabetically by countryILLINOIS MASSACHUSeTT~ Hewlett-Packard ComParlY ~~l..paokard CompanyHewla,t-Paokard Company 'i1711~-Paokard '1any 3 ~IW~ Park Wa~t i 0 r.ator. ~ow2206 • EmfJ~ StrMt 1 76 lnut~ ROt W 8 ~NY 1187 1 or. u aP.O. ox 1 7 ~DOVER, 01 10 Tall (6ltl) -7800 ~=IS~38131BLOOMIN~TON, L Ta: (817) 862-1l5OO al: (1101) 34ll-a370~1702-180

~leHIQAN ~~AROUNAal: (308) 682-8411 tt ackard ComparlY Hewlatt-Paokard Company_Iatt-Paokard :rJj;ParlY 44 Vantao- Way

ILUNOIS 38l5l5O Orohard HII _¥ie~sF~%J 27420

SUlta 180Dr. ~AS~E~ 37226Hewla,tt-Paokard CompanyNOVl, MI 4= a: (1 ) M,H800 al: (81) 127tl5201 011Yl_ Drl~

ROLLINQ MEADOWS, IL Tal: (313) 200TEXAS80008

~NES2TAOHO

Tal: (312~ ~t-PaokardCompany Hewlatt-Paokard CompanyTal.1 81 -1088 latt aokard Company

1 ~Road 1~ Harwlq{rl~2026 W. LarPAnttr, Avanua , OH 44138 ~ ~ON, 7038INDIANA ,T. PAULI MN lt3 a: 1) 2~7300 al: ( 13) 778-&400

al: (812 144-1100Hawlatt-Paokard OOrnParlY

Hawlatt-Paokard~ Hawlatt-Packard~cy11811 N. Marld~tr..t MISSOURI~ARM!L IN 4 Hewlatt-Paokard eomParlY 0010 S~~o Pika ~JAA~~I~O' 1al: (317) 844-.tOO 1001 E. 10t.t Tarraca ~18 ~6342 all (214) 23\-810t

SUltalf all (613)IOWA KANSA CITY, MO Hawlatt-Paokard CompanyHawlatt-Paokard Company r.131~ OKLAHOMA 1~ ~ral P.rkwalo' 80.~ft:'wataDrl~ al: ( 18) 841-0.11 Haw\att-Paokard ComParlY , ONIO, TX 76232

S MOINES, IAHawlatt-Paokard ComParlY ee66 South L_I. all (612) 484-8338

SUltll06Tal: (616) 224-1436 ~~~~~~~tD~ TlA. A ~. 74138 urAH

Tal: (314) 344-6100 Tall (gl ) 461-8700 Hawlatt-Paokard Com8anyK.l!NTUCKY ~~OO~h 14~Iatt-Paokard~ NI!W JERSIY OR!GON AL C ur 118

N. Hur8tbourna =t-PaOkard company ~att-pa~dOOrn&::'JTall (601) 874-1700

)Sulta 100 120 • CanEd g. W. 10M« VIRGINIAtOUISVlLLE. KY 40223 ~ARAM~S. 07662 • ~OR87070

Haw"tt-P~okard Companyal: (602) 428-0100 al: ( 1) 2 at: ( ) -8000 4306 Cox cadLOUISIANA Hawlatt-Paokard ComWy ~NAU.£N v~rP6N'oISYLVANIA al: (804) 7~b-Hawlatt-Pa~ard ~pany ~~ E~land Ava. a.t

Haw!ltt..p~ok.ardcompany160 ~ma. rl~ a.t SCA A AY NJ 06854 WASHINGTON~TO OSE~ LA 70087 Tal: (201) 682-8100 111 ata rl~

~8BURQ~ ~6238 Hewlatt..paokrd Company•• Box 448N!WYORK al: (412) iI2 166115 8.!. 3 th St~t

~NNeR, LA 70083,e~, WA 88OOl1al: (1504) 4(17-.100 Hawlatt..paokard company

~ar Drl". South~~f~~

at: ( ) ll43-o4OOo4

MARYLAND al: (61i)~aa!~ Wl8T VlRQINIAHawl~=~ Company

V~F~llM82 Hawlatt..packard Company3701 • str..t Hawlatt..paokard OompanJk T: 16)

mJ3t~~r..tBALTIM RE~..ie&227 rn era..~ attica P L S ON WV 25304Tal: (301~ ~RT, 144l5O T!NNeSS!E Tall (304) 826--0482Talex: 71 -1843 al: ( 18) 223-0860 Haw~tt..paokard Ollmpany

0". ~ Cantar W180ON~INHawlatt-Paok&r'9 Company Hawlatt-Paokard Company t1ta~ it!att aokard Company2ehokae~R~ rm~e~Blvd. lIWrfl'k~ . 6N.~ata~

OOKVlLLe RPO L, 13011 'r.f: 371132 00 lD WI~al: (301)~ 0 Tal: (316) 461-1820 : (8 ) eee.:.747 Tal: (414) 7e.:i18OO

5

Page 124: HP 7673B - S service manual
Page 125: HP 7673B - S service manual

Fli;' HEWLETTa:~ PACKARD

Hewlett-Packard Company,Avondale Division, Avondale, PA 19311·0901)

Manual Update: Installing a syringe

This update provides new information for installing syringesinto theHP 185938Injector Module. Metal syringecUps enable you to installa syringefrom the front, similar to the A modules.

Front~Ing eyrtnge oap..

To update your manual, follow thlJ Instruction.

• Replace these pages in the AutomaticSampler Operating Manual(manualldt part number 07673-6(995).

TItle page, Print History, RFI and Sound Emission Certification,2-7,2-8,3-29,3-30,3-31, and 3-32

Printed in USA 4/9107673-90335

Page 126: HP 7673B - S service manual
Page 127: HP 7673B - S service manual

HP 7673 Automatic SamplerOperating Manual

r"3 HBWLETT~PACKARD

Manual Part No.07073-10111

Edition 2, AprIl 1"1PrIm.d In USA

Page 128: HP 7673B - S service manual

Printing History

The information contained in this document may be revised without notice.

HewldtwPackard makes no warranty of tury kind with ngard to thismaterial, including, but not limited to, the impliedwarrantiu ofmerchantability and fitness for a particularpurpose. Hewlett-Packard shall not beliable for errors contained herein or for incidental, or consequential damagesin connection with the furnishing, performance, or use of this material.

No part of this document may be photocopied or reproduced, or translated toanother program language without the prior written consent-ofHewlett-Packard Company.

First edition-October 1989Printed in USA

Second edition-April 1991Printed in USA

C Copyright till, 11 by Hewlett-Pickard CompanyAll Rightl Re..Ned

Page 129: HP 7673B - S service manual

RFI and Sound EmissionCertification for theFederal Republic of Germany

Manufacturer'8 Declaration

This is to certify that the equipment lIP 7673 is in accordance with the RadioInterference Requirements of Directive FI'Z 1046/1984. The GermanBundespost was notified that thisequipment was put into circulation and theright to check the seriesfor compliance with the requirements was granted.

This statement is provided to comply with the requirements of the GermanSound Emission Directive dated Januaty 18, 1991.

• Sound Pressure Lp < 70 dB(A)• At Operator Position• Normal Operation• According to ISO m9 (IYpe 'Iest)

Heratellerbeschelnlgung

Hiermit wird 6esc:heingt, daB das GerAt/System HP 7673 in Ubereinstimmungmit den Bestimmungen von Postverfilnung 1046/84 funkentstOrt ist.

Der Deutschen Bundespost wureJe das Inverkehrbringen diesesOerAtes/Systems angezeigt und die Berechtigung zur Oberprilfung der Serleauf Einhaltung der BestbnmungeneingerAumt.

Diese Information steht im Zusammenhang mit den Anforderungen derMaschinenllirminformationsverordnung vom 18Januar 1991.

• Schalldruckpegel Lp < 70 dB(A)• Am Arbeitsplatz• Normaler Betrieb• Nach DIN 4563S T. 19(lWprllfuna)

Page 130: HP 7673B - S service manual

Chapter 1. Setting UpThe Modules of the Automatic Sampler. . . . . . . . . . . . . . 1-2Preparing the Site 14Installing the Injectors. . . . . . . . . . . . . 1-7Installing the nay .•............................... 1-13Installing the Controller. . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-17Using B Modules with A Modules................... 1-24Preventive Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-28

Chapter 2. Example of OperationDesigned for the beginner, this IS-minute "dry run"illustrates how to prepare the automatic sampler and runa three vial sequence using the following methods of control:

• HP 3396AIntegrator with the Instrument Network (INET).• HP 3365ChemStation (DOS Series).• HP S880A Gas Chromatograph.• Standalone control using the injector's control switches.

Chapter 3. OperationPreparing Sample Vials .....................•...... 3-2Using the Injector Fan 3-8Controlling Sample Vial Temperatures 3-10Preparing the Solvent and Waste Bottles.. . . . . . . . 3-16Selecting and Installing Syringes .....•.............. 3-28Preparing the Inlet .....••......................... 3-34Adapting for Cool On-column Injection . . . . . . . . . . . . . . 3·36Setting the Run Parameters. . . . . . . . . . . . . . . . . . . . . . • • . 3-37Checklist 3-46Running the Samples 3-47Using 1\vo Injectors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-50nay Control Commands 3-54

Page 131: HP 7673B - S service manual

5. Install the SyringeUse a 23 gauge syringe with packed and capil1aIy inlets or a 26/23 gaugetapered syringe with on-column inlets. However, you cancomplete the dry runwithout a syringe.a. Open the injector door.b. Slide up the gray loop of the plunger carrier and loosen the plunger screw

about five turns.c. With your fingernail, lift and tum the tab of the needle support jaws so

that the upper jaw of white plastic swings toward you.d. Une up the top of the syringe barrel with the guide at the top of the metal

clip and snap it into place.e. Uft and close the white jaw around the needle.f. Slide down the gray loop of the plunger carrier and tighten the plunger

screw around the plunger button of the syringe. Slide the loop up anddown to be sure that it is holding the plunger. Close the injector door.

Plualler butto.

PIIIII,r Olrrler

0&0

allde

Syrllll' 01111I

Syrall'

Needl' IUpport Jawl

For more information on selecting and installing a syringe, see page 3~28.

2-7

Page 132: HP 7673B - S service manual

Preparing the GC and Controlling Device

To minimize the amount of time between injections, enter the appropriate setpoints for oven temperature, equilibration time, initial time, and data analysis.

You can minimize the time for the GC to be ready by setting the oven to a lowtemperature, remove any temperature ramp program, and set the equilibrationtime to zero. You can minimize the amount of time the integrator or computercollects data by turning the detector or signal off and setting the run time toabout 10 seconds.

Turn to the section for the controlling method that you are using and completethe list of tasks.

HP 3396A Integrator

1. Set the run time to 0.1 minutes bypressing CJiMD c::J CO CW).

2. Begin the dialogue to prepare a sequence by pressing the keysfor[PREP][SEQ).

3. Answer yes to INET SAMPLER CONTROL by pressing (I) [ INT!!I I.

4. Turn to page 2~12, column two for instructions on setting the runparameters.

• TIU.1 STOP• PUP SEQ

ALS INFOIUlATIONINET SAMPLEI CONTIOL [Y./N]; Y.

717SA SAMPLER;LOOP ADDU!!; a

FRONT INJECTOllINJ /BOTTLE a ->

2-8

Page 133: HP 7673B - S service manual

'CAUTION IFallur. to us. the on-column ayrlnge when inJecting Into anon-column Inlet could damagethe InJ.ctor, ayrInge, and column.

Inspecting Syringes

Before installing the syringe.

1. Roll the syringe on the edle of a clean flat surface. If the tip of the needleseems to follow a circularmotion, bend it aliJhtly ncar where it connects tothe syringe barrel and check it apin.

ae.d lI.r. n••c....ry

Flpre 3·21. Syriap partI aDd aHdIt lnapectloD.

2. Check for a rough needle. The needle may contain closely spacedconcentric ridges that act_like a miniature me and abrade pieces of theseptum into the inlet. The ridges are easy to see under lOX magnification.

If there are ridges. polish the needle by pulling it through a folded piece offine emory paper between you rmger and thumb until the ridges are gone.Be careful not to modify the special blunt tip of the syringe.

~. Check for a stickyplunger. Slide the plunger of the syringe up and down afew times. It should move smoothly-without sticking or binding.

3-28

Page 134: HP 7673B - S service manual

Installing Syring••

After completing this task. check your work with the instructions on page 3-32.There are three functioning parts to keep in mind.

• The syringe is held onto a plastic syringe carrier by metal spring clips.• The needle is supported bya set of plastic jaws at the end of a spring

loaded rod.

• The plunger carrier moves the syringe plunger up and down. The carrierholds the plunger with a "thumb screw.

1. Open the injector door.

2. Slide the gray loop of the plunger carrier up as faras it will go.

3. Find the black and white plastic needle support jaws.

4. With your fingernail, lift and turn the tab of the upper, white plastic jaw 80

that it swings toward you.

Nudl. upport Jewa 3

Syrl.g. ollpaltt.clltdto ayrl.g. CirrI_g.

"

FIpre 3.22. Plunler carrier and needle IUPPOrt.

3-30

Page 135: HP 7673B - S service manual

S. Position the syringe in front of the metal syringe clips. The top of thesyringe barrel should fit into the guide.

6. Push the syringe against the clips until it snaps in place.

7. Move the tip of the needle into the notch of the black plastic needlesupport Jaw. Uft and close the white plastic jaw around the needle.

8. Loosen the plunger screw about S turns,Slide the gray loop of the plungercarrier down SOthat the plunger button fits into th~ hole of the plungercarrier. TIghten the plunger screw.

Plaager carrier

C!> Gild.

I )--,-...,....---,---,---1Syrlagl ell,.

Plaager button

Syrngl

7}------__..N_II lappon IIWI

F"ure 3·23. Snapplq the ayrfnae Into place.

3-31

Page 136: HP 7673B - S service manual

Checking Your Work

o Move the gray loop of the plunger carrier up and down. If the syringeplunger does not move along with the carrier, return to step 8. Be sure theplunger carrier screw is tight.

o Be sure the jaws of the needle support are closed around the needle. Pushup on the bottom of the needle support jaws (compressing the spring). Ifthe jaws of the needle support do not slide up and down the needle freely,return to step 4. Also check for flaws on the needle as described in"Inspecting Syringes" on page 3-29.

o Check the alignment of the syringe needle with the cone shaped lead onthe septum nut to ensure an average septum life of 200 injections. Followthese instructions.

1. Pull down the syringe carriage until the needle tip is near the top ofthe inlet's septum retainer nut. Use the plastic tab on the syringecarriage. Figure 3-24 shows a syringe needle over the septum retainernut of a packed inlet.

I~

1If.

.........''N

Plulglr .crlw

The needle should be centered exactly over the hole in the septumretainer nut. Use a small light to see that the needle will hit the sep­tum without rubbing on the nut.

3-32

Page 137: HP 7673B - S service manual

ti

2 bracket StWQrts (18597-00235)(Drawn fiJIscae )

No. 2 PoiltPozOiv

3/B-il. cresc8ft wrench

No.2 POOt Pozidiv

4 pal head saews (0515-0915)(Drawn fuI selie )

3 standoffs (18597-20710)(Drawn fLt seas )

0-

oo

[tZZZZ77ZZZZ711~

3 Attach theTray Bracket

2 Attach the Tray Str1cIoffs

1 Attach the Bracket &worts

1 tray bl'acket (18597-00135-1)

r I ,- I I ~~ 3 pan head screws (0515-0915).!

( Drawn M scale )

g.. 1 pal head screw (0515-0910)( Dmm MI scale)

>- 1 flat headscrew (0515-0913)( Drawn fijJ scae )

'(,I

o

1',~' :l~~I .'~1/ !

I

l

ExpIoded~

~.

"

r-::

i

Page 138: HP 7673B - S service manual
Page 139: HP 7673B - S service manual

...~:

II

f. Line up the hole on the near side of the bracket Oabeled 5890) with thethreaded hole of the front mounting support. Screwone flat head screwinto the mounting support with a No.2 Point Pozidriv.

g. Tighten the four screws that attach the mounting supports to the sideof the flow panel.

d. Push the mounting supports you installed in step 1 so they are flushagainst the left side panel

e. Line up the hole on the far side of the bracket Qabeled 5890) with thethreaded hole of the rear mounting support. Screw one pan head screwinto the mounting support with a No.2 Point Pozidriv.

i.\

c. Using a No.2 Point Pozidriv, screw in the three pan head screwsthrough the holes labeled A and into the standoffs.

b. H the top of the standoff's do not line up with the holes labeled with theletter A, check the location of the standoffs against figure 2 on page 5.

a. Place the tray bracket on the mounting supports and the standoffsinstalled in steps 1 and 2. The top of the standoff's should fit into theholes labeled with the letter A on the tray bracket.

f,r-

f'~.

r~~:::

~

t~

1:.-

~_.,

[,~'.

~~C',

~

I

-c 3. Attach the Tray Bracket

ir;.

~tr:~' Figure 3. Attaching the tray bracket to the standoffs.

6

Figure 4. Attaching the tray bracket to the bracket supports.

7

Page 140: HP 7673B - S service manual

3Checking Your Work

Double-check the following;

o The three screws that attach the tray bracket to the standoffs on theinjector bracket must go through the holes labeled A on the traybracket.

'.

2. Attach the Tray Standoffs

a. Screw the three standoffs (hexagonal posts) into the injector mountingbracket. Use the holes identified in figure 2. Hyou do not use thecorrect holes, the tray will not be able to deliver sample vials to theinjector.

b. Tighten the standoffs with a 318-in. crescent wrench. .'

i~

~;~

'."~'.'..~.'.'.¥»o-:.".

1;:$

.j.. ~.:.•.~•..~~..~

~-:.;

..~

j-~

-::.~

'~

.,~

]'.'!l

-:::-.---~--;,~

i.:.;""':1~~--;~

~-'i

, - I~I

o All screws and standoffs are tight.

To mount the tray on the bracket and connect the cable to the controller,see chapter 1 of the operating manual "Setting Up."

8

Figure 2. Mountinz: bracket holes for the tray standoffs.

5

Page 141: HP 7673B - S service manual

;:..-.~,

~

~~

;;:.

~

~~.,.~.

~~_.

f

t,

f~.·.•··'···.··.··.··:·······.-=-:..

-*...

r~

r~~

~ ...~~

r~

""~

,~

["~X.

t.r

i:

1. Attach the Bracket Supports

a. Using a No.2 Point Pozidriv, attach the two bracket supports to theflow panel on the left-hand side of the GC. Do not tighten the screwsuntil the end of step 3.

II II!!lui~~e:..: j{F'10I1"

Figure 1. Attaching the bracket supports.

4

Page 142: HP 7673B - S service manual

Installing the Tray Bracket

Pull out the yellow worksheet from the centerof this guide, Organize thehardware and tools for each step. Full-scale drawings of the screws helpidentify which ones you need for each step. If any of the hardware ismissing, contact your HP representative.

...~.~

-~

.0;-

I_>5

-,.",.}c>-i

···.···;s..~.f~:::.:v "

,

"~

I;~

~

.,

.~-<

··I·.···.·.-.··.··.@.

-,

~-:-~~~

• --c,.

.2.........

2

2 ...........~('IIIIa.")

__t~

---~_....~.--,

.................fII-. ...)

..........~s....~~.-..-

~ 1 ...._a.~

,... ea-:III1DiI)

Exploded view with aI parts

(fh~Wa-

~tJystep

Patsyou need for each step

Tools you need for each step

--

.~~

3

Page 143: HP 7673B - S service manual

p:t-~.-~~

~... Before You StartT

"

1

I-f

IiI

t~f~u~-

r~

ff~

~

f;

~..

~

2

Before you install the tray bracket, you must do the following:

• Install the split top cover for the HP 5890 SERIES II GasChromatograph (Kit No. 07673-61(65)(only if installing onto anHP 5890AGas Chromatograph).

• Install the mounting bracket Cor the HP 18593B Injector Module(Kit No. 18957-60655).

• Remove any injectors f'rom the iDject.or mounting bracket.

• Install the left side panel (Part No. 05890-61480) for the HP 5890SERIES II Gas Chromatograph (only ifnot already installed). You canalso install the bracket on the support rails.

After you have completed the installation, place this link manual in thebinder of your operating manual for the HP 7673 Automatic Sampler(Part No. 07673-60995) after the tab for appendix C "Link. Manuals."

© Copyright 1989 by Hewlett-Packard CompanyAll Rights Reserved

Page 144: HP 7673B - S service manual

HEWLETT-PAC K.AR 0-~.;J

J"~-~j--

Manual Part No.07673-90205

~HEWLETT

~PACKARD

Printed fn U.S.A.October 1989

( =-= 0

(=-= ..

LinkManual

HP 7673 Automatic SamplerInstalling the Tray Bracket on 5890GCs

-~~.''''',-~

:i<-~:~

~~

"~~

Page 145: HP 7673B - S service manual

4 P\ace the Bracket

t>-

3I8-i1. crescenl: wrench

No.2 poi1t Pozidfiy

3/8-il. crescent wrench

2 pal headscrews (0515-0915)(Dfawn~ scale )

3 standoffs (18597-2071.0)(Drawn fIJ scale )

2 stIldoffs (16597-20745)( Drawn fu1 scale )

1 tray bracket support (18597-00195)

.ctZ77777777ZZZ}tD

0-

7 Attach theTray Standoffs

8 Attach the Tray Bracket~ .

b)­

.~( (

((

"U'11

.-

,. /

I (

(f

4 sta Idoffs (18597-20520)( Drawn ftI1 scale )

2~ standoffs (18597-20755)( Inwn U scale )

4 sta Idoffs (18597-20745)( Drawn fu1 scale )

No. 2 poD PozJiIvi'Mtwreoch

4 pal head screws (0515-0915)( Drawn fu1 scale)

1 sheet metal (18597-00245)

No. 2poi1t Pozictiv

3/8-it crescert wrench

1/4-il. I1Jt ctiYer

1 qectorbracket (18957-oot75)

0'0oo

•o

Dt??1 II I 221l=­

t???? I? I II J:-Q-

3 Attach the Injector Bracket~

~~c

No.2 pom Pozidriv

1 flat head saew (0515-0913)( Drawn fuI scale)

1 traybracket (18597-60515)

3 pan head saews (0515-0915)( Drawn fu1 scakl )

1 pal headsaew (0515-0910)( Drawn fij staHl )

0­0­~

9 Attach theTray Bracket

~',J.

'I _

f ((.\-, (

IaIhead saewaiYer

1 dum1y septl.rnru (18597-20580)( Drawn fiJI scaM )

1 septlm IlJt (19243-60500) .( Drawn fiJI scale )

No. 2 poi1t Pozktiv

2 aIitpnent tools (18597-40030)

(I

5 AIg1 the Bracket

6 Attach the MolJ1DJ Posts m 8eptJ..rn Nuts

t ~2 i1jectoc rTKUIli1g posts (07613-61035)

( Drawn 101 seale )

ttt' ~ If~

~,,

Page 146: HP 7673B - S service manual

r:( :

( i'\, ';

..

"

~" 1lis wor1<sheet is for Ilsta1ilg the ITlOlIlti1g bracket for the HP 18593B In;ectorModLie and the HP 18596B Tray MCKUe toCI1 HP 5880A Gas ChrornatDgaJtl.

.~~

-~

"r;-

I.~

I-'~

..0<~.:

--;~~

~i~

~ .",

( ( I 'I'\ '\'. ',~

J.~-'"t.

Page 147: HP 7673B - S service manual

f .

kt"t,~,,-~-

(:

~

t'~"r.f'~.,

':

5Connecting the Cables

Referring to the figure below, connect the HP 5880A control cable to therear of the HP 5880A and the rear of the controller. This control cable isthe only controlling cable needed, whether one or two injectors areinstalled.

The type of data-handling device and communications configuration you usemay require additional cables. Refer to manuals for that device.

6Operation

This section contains operating information for the automatic sampler,including:

• Filling and placing the bottles.

• Setting the run parameter.

Several eeamplee illustrate setting the run parameters. Refer to theHP 5880A lnstnDnent Manual for operating instructions for the GC itself.Refer to the HP 7673 Automatic Sampler Operating Manual for detailedexamples and information on preparing the sampler.

I...I\l

T.,YinrwID aTray

Riling and Placing the Bottles

Always place two 4-ml waste bottles into waste positions A and B on theturret. Place one or two 4-ml solvent bottles into thesolvent positionsaccording to the run parameters.

The following illustration shows bottles placed in both waste and bothsolvent positions.

19321~111

•~r~

t;{

~ 2021

Page 148: HP 7673B - S service manual

,"

19

,<;

J,~

c~

~

C"~

'~.~

,~

.~

J+~

j- "'J

.-}. '.''Jo~

-~-.~-~

~':;~-',",

.-~~

'.

.i'J,,.~~

.~---';~

Be Sire III threadthrnp b01llIbecnaeetor aIIdtile uameplale

o@

"

o

o

RMU'ODd lUluIIwerI»M1I gIHts

o

Attach the lower board guide to the card cage with the pan-head screws.

d. Directly above the selected slot, install the plastic board guide (notchedend a'WRy from you) inthe roof of the card cage.

e. Install the lower board guide, J2 connector, and ribbon cable assembly.

Route the J2 connector and cable behind the other J2 connectors.Thread the connector through the hole in the back. panel labeled ALSIBCD. Attach it with the small brass nuts and washers,

r. Reinstall the cable shieldingplate.

g. Install the sampler board (part number 19321-60010).

h. Reinstall the boards you removed in step b.

i Attach the side panel

j. Replace the blank key caps from the terminal keyboard with the AUTOSEQ keycaps in this kit.

LIST AUTO SEQ

INJECTOR 1:o. MODE (O=NORMAL. I=ON COLUMN) = 01. PRE-INJECTION SAMPLE WASHES = 22. VISCOSITY = 03. SAMPLE PUMPS = 64. SAMPLE VOLUME = 15. POST-INJECTIOff SOLVENT A WASHES .. 26. POST-INJECTION SOLVENT B WASHES = 27. INJECTIONS PER BOTTLE I: 18. FIRST BOTTLE = 19. LAST BOTTLE = 1

'J I

~. w_ \'22

When using the HP 5880A, do not use sample inserts in the turret.

When the tray is installed, all sample vials are loaded into the tray, and thesample position in the injector turret is left empty. WIthout a tray you canonly ron one sample at a time by placing it in the turret's sample position.

Setting the Run Parameters

lqjector and sequence parameters are maintained in an "auto sequencetable" within the HP 5880A. There is a separate table for each injector.The HP 5880A executes instructions provided in the auto sequence tableand ignores front panel switches on the injector itself.

A complete auto sequence table requires ten parameters. A default value isused if no value is entered for a given parameter. The following illustrationlists these parameters and their default values for the front injector.

List the parameters stored in the auto sequence table for the front injectorfrom terminal 11 by pressing

EJ~rn.

Page 149: HP 7673B - S service manual

~"

,

~',

For example, to set the sample volume parameter for the front injector to 2,press these keys on terminalll

8 (wg) CD c::::J CD rn·

Change a parameter in the auto sequence table by pressing

:;....t'-"

~.

e,~·

t..~».

;-;...

~.'-'"

2. Install the Sampler Board

a. Select a slot in the card cage for the sampler board. You can use slots9,10,11, or 12. Ifan HP767ln2A board is installed, it must beremoved

b. Removethe boards between the selected slot and the back of the GC.As each board is removed, record its location so that it can be placed inthe samelocation when reinstalled.

c. Removethe cable shielding plate.

8(~l [~eter] G:J [nlue of ]

paramettr rn·

~-

;

...~

For more information on simultaneous injection see "SYNC Mode" onpage 28.

Any of the ten parameters can be entered or changed manually orautomatically using keystroke or BASIC programs. If an incorrect entry ismade for any parameter, the HP 5880A will refuse the entry and indicatethe cause of the error. The first and last bottle parameters may be BASICvariables. All others must be simple numbers.

Injector #1 is the front injector and injector 112 the rear injector.

Ifyou do not enter an injector number, the edit affects the auto sequencetable associated with the terminal you are using. Terminal #1 controls thefront injector (Il) and terminal 12 controls the rear injector (#2).

The [mj II may be 3; "3" indicates that you want simultaneous injections(i.e., front and rear injectors inject at the same time).

You can change a parameter in the auto sequence table for either injector

from the same terminal. Indicate the number of the injector, the number of

the parameter, and the value of the parameter by pressing

8 [iII#]1 wg I [~....l c::::J [.=rl rn·RHIne

@

o Q)

-: 11110

"'\I"~

~

•i..·· ..•.c•..•.Ft~-

~--i

r~t'

~.

[

18 23

Page 150: HP 7673B - S service manual

1. Upgrade the ROM Board

a. Remove the right side panel of the GC.

b. Remove the ROM board (part number 19300-60155) from slot 2 of theHP 5880Acard cage.

e. Replace the group 4 EPROMs Oabeled G4L and G4R) with the EPROMsin this kit according to the follOwing table and illustration. The tableshows which EPROMs are in each accessory and whether they go intothe left or right position.

d. Insert the ROM board into slot 2.

Description of Parameters

O. INJECTION MODE: Specifies whether the injector is in the normal oron-column injection mode. The on-column injection mode is used with coolon-column inlets. For the on-column injection mode, the injector changesthe injection speed to 500 milliseconds and lowers the tip of the syringeneedle an additional 19 mm (814 inch) to the column.

• Set to "0" for normal injection mode. Use this setting with packed andcapillary inlets.

• Set to "1" for on-eolumn injection mode. Use a 26123-gauge syringe.

1. PRE-INJECTION SAMPLE WASHES: Specifies the number .oftimes the syringe is rinsed with sample before the injection. The injectorlowers the needle into the sample vial, fills the syringe to eight-tenths itsfull volume, and empties it into one of the waste bottles.

The allowable values for this parameter are 0 to 15.

Accessory NumberHP 19321AHP 19321BHP 19321C

Left (G4W19300-8007019300-80090193DO-80110

Right (G4ID19321-8008019321-8010019321-80120

~~,-i..;.::

1l:'~

~-~1

~

.~

An excessive number of sample washes can create a vacuum in the samplebottle and affect the reproducibility of the injection volume.

2. VISCOSITY: Specifies the number of seconds the plunger pauses at thetop of the pump stroke and injection stroke. For viscous samples, the pauseallows the sample to flow into the vacuum created in the syringe.

The allowable values for this parameter are 0 to 7.

3. SAMPLE PUMPS: Specifies the number of times the syringe plunger ismoved up and down while the needle is in the sample to expel air bubblesand improve reproducibility.

The allowable values for this parameter are .() to 15.

24

Replace

17

~

..~

-:~

.

..:.t.-.:.•.,.~.-..~

..._.~·'i.so

L'-

Page 151: HP 7673B - S service manual

~-­

~-

~0"~;--"--

•~~(

~.

~ '.

'i­~

~.~R,

~Jl.;...•.•~... '...;1'

~-

l!

"~~"

~

r

I~.'

~.~<.'

J.:.~

f"

~.

t~--

~lk

4Installing the Electrical Components

This section explains how to upgrade the EPROMs on the ROM board andhow to install the sampler board to the GC.

WARf\;INGi

TURN THE POWER swrrca ON THE GC TO "OFP' ANDDISCONNECT THE POWER CORD BEFORE INSTALLING THEAUTOMATIC SAMPLER BOARD IN THE GC.

t CAUTION ITurning off the power immedIately erases data stored In theInstrument's memory. To save this data. list the set points andprogrammed events. Reenter the data when you havecompleted the Instanation.

ThIs procedure requires protection against electrostaticdischarge. Use a static control wrist strap (part number9300-0969 for large wrist strap or part number 9301>-0970 forsman) connected to a ground. "you do not use staticprotection. you may damage the electronics of the board orcontroller.

16

4. SAMPLE VOLUME: Specifies the injection volume. Each volume unitcorresponds to one-tenth the syringe's capacity. For example, each unit fora 5-~ syringe is 0.5 ~

The allowable values for this parameter are 1 to 5.

I CAUTION IDo not set the volume parameter for splltless Inlets to volumesmore than 2 Jl.1. Larger volumes cause sample to be lostthrough the purge vent and may contaminate carrier Inlet lines.

5. POST-INJECTION SOLVENT "A" WASHES: Specifies the numberof times the syringe is rinsed with solvent from the solvent A bottle. Theinjector lowers the needle into the solvent A bottle, fills the syringe toeight-tenths its full volume, and empties it into one of the waste bottles.

The allowable values for this parameter are 0 to 15.

6. POST-INJECTION SOLVENI' "B" WASHES: Specifies the numberof times the syringe is rinsed with solvent from the solvent B bottle. Theinjector lowers the needle into the solvent B bottle, fills the syringe toeight-tenths its full volume, and empties it into one of the waste bottles.

The allowable values for this parameter are 0 to 15.

7. INJECTIONS PER BOTI'LE: Specifies the number of times the ronis repeated for each sample vial. For example, if the value is two, each vialis analyzed twice.

The allowable values for this parameter are 0 to 5.

I CAUTION IAn excessfve number of Injections per vIal can create a vacuumIn the vial. This may affect the reproducIbility of the injectionvoIum$.

25

Page 152: HP 7673B - S service manual

8. FIRST BOTI'LE: Specifies the tray or injector turret position of thefirst sample vial. Each position on the tray and turret is numbered. Trayvalues range from 1 to 100. Injector turret value is L This parameter maybe a BASIC variable.

Without a tray, the HP5880A GC can only run one sample at a time.

9. LAST BOTTLE: Specifies the tray or injector turret position of the lastsample vial. Tray values range from 1 to 100. Injector turret value is l.This parameter may be a BASIC variable.

An empty bottle position between the FIRST BOTrLE and the LASTBOTTLE parameters will be interpreted as an error, and the sequence willabort.

Starting a Run or Sequence

With a tray, p..... (ST",") [~:,o) rn.Without a bW, press [ST",") (~)8rn.GC readiness is required before injection. Conditions required include ovenreadiness and the absence of conflicting operations, such as already runningan analysis, auto sequences, or programs.

You can override the parameters in the autosequence table for FIRSTBOTTLE and LAST BOTTLE. The following example starts a sequence forsample vials in tray positions 6 through 26 using the injector associatedwith the terminal you use.

(START)!~ JCD C:::J WCD rn .

26

3Checking Your Work

Double-check. the following:

o Be sure. the new septum nutts) and dummy septum nutfs) are installedproperly.

o Be sure the three screws that attach the tray bracket to the standoffson the injector bracket go through the holes labeled B on the traybracket.

o Be sure the flat head screw for attaching the tray bracket is in thecorrect position. See page 14-

o All screws and standoffs are tight.

For detailed information on mounting andconnecti.ng the injector and tray,see chapter 1, "Setting Up" in your Operating Manual.

An example cable diagram for your GC is in the next section.

After you have completed the installation, place this Link Manual in thebinder of your Operating Manual for the HP 7673 Automatic Sampler(part number 07678-60995) after the tab for appendix C "Link. Manuals."

15

·~.'~r.~.·~.'

.'.~'

;;--0;.:

.~~

,.'-'·4

i~",:.j.~

1i

.,.:.~..~

~

-1.c;

J.~

·.·•..1-';

c.,;~1

.•..j'-z~

:··.i···~·:.··~'~'-:..,

"

.~c~

~

Page 153: HP 7673B - S service manual

f '~r~:

f _

~.'

~:;.;.-

~':!""..

t.•.·..·.·rriI:-:". ­~

!J-:~~

f·f~-

9. Attach the Tray Bracket

a Place the tray bracket on the mounting supports and the standoffsinstalled in steps 1 and 2. The top of the standoffs should fit into theholes labeledwith the letter B on the tray bracket.

b. lfthe top of the standoff's do not line up with the holes labeledwith theletter B, check the location of the standoffs against the figure onpage 12.

c. Using a No.2 point Pozidriv, screw in the three pan-head screwsthrough the holes labeled B and into the standoffs.

d. Line up the hole on the rear of the bracket Oabeled 5880) with thethreaded hole of the tray bracket support. Screw one pan-head screwinto the mounting support with a ne. 2 point Pozidriv.

e. Line up the hole on the front of the bracket (labeled 5880) with thethreadedhole of the tray bracket support. Screw one flathead screwinto the mounting support with a no. 2 point Pozidriv.

FIaIbeatIICI'IW

14

Stopping a Sequence

To slop • ron, press ( STOP )( We? Irn.This command can be used at any time. The run in progress is completed,and no further injections are made. The command must be entered on thesame terminal or channel that you started the sequence.

Single Sample

With a tray installed, you can interrupt a sequence and run a single samplefrom the turret of an injector by the followingprocedure.

1. Note the sample vial number currently running and interrupt thesequence by pressing

(STOP I(We? Irn.2. Change any run parameters using the EDIT SEQ keys explained on

page x,

3. Place the sample in the sample position of the turret.

4. Start the single run by pressing

(STAAT)( We? )8rn.5. Reset the run parameters for the sequence you interrupted, including

the FIRST BO'ITLE, and restart the sequence.

27

Page 154: HP 7673B - S service manual

1."il-..0:;-

:i~

,;~~~

-.,~

.jj~

-~~

"

~~~

'1.~.-~

-~

~l.~

,:~

.~-~4

"1

8tCItI:I of..S88QA

13

Rear IIoIes

If the injector bracket is installed over the B and D inlets, use thefront set of holes. -

If the injector bracket is installed over the A and C inlets, use the rearset of holes.

c. Using a no. 2 point Pozidriv, attach the tray bracket support with twopan-head screws.

a Using a no. 2 point Pozidriv, remove the two screws that hold on thepanel on the left-hand side of the GC. Remove the side panel

b. Position the new side panel (part number 19300-01260) and attach itwith two short standoffs (part number 18597-20745). Tighten themwith a 318-in. crescent wrench.

8. Attach the Tray Bracket Support

Started "". IInIJnI ..1

DIIIctor

Detector

........

aaaanel2

Q-flDIl ~UYH

cu.el1(JraIw DO. ofnm)

For these types of applications. the SYNC feature must be turned "ON" bypressing

The synchronous mode (SYNC) is used to synchronize injectors andterminals. When the synchronous mode is turned on, both terminals startat the same time. H there are two injectors, the injectors inject at the sametime.

Start the synchronous sequences from the terminal associated with thegreater number of runs by pressing

(START) GO ('Wa' )rn .The following illustration shows using two injectors and an oventemperature program. The injectors and terminals must be synchronized sothat they start at the same time.

When using an oven temperature program, enter the parameters for eachinjector from its own terminal. Also set the SYNC feature to "ON" on bothterminals.

28

SYNC Mode

Page 155: HP 7673B - S service manual

DItecttr

Deteder

Onl

I CAUTION IIn the simultaneous Injection mode, an error will result If the twoauto sequences can for the same bottle for the same Injection.In this case the error message

••• ERROR ••• AUTO SEQ BOTTLE CONFUCT

will be printed, and the auto sequence aborted.

CU11De12

CU11e11

JCAUTION IIf you start the sequence from the termlnal associated with thesmaller number of runs, the excess runs will not be performed.

The following illustration shows splitting a sample and using two columnsand two detectors to analyze it. The signal from each detector goes to adifferent terminal The terminals must be synchronized so that they startat the same time.

~ ~•

f~,I

a. Screw the three standoffs (part number 18597-20710) into the injectormounting bracket. Use the holes identified in the illustration. lfyou donot use the correct holes, the tray will not be able to deliver samplevials to the injector.

b. Tighten the standoffs with a 31B-in. crescent wrench.

7. Attach the Tray Standoffs

t~

~.

irrE(~

~

~,

-.;;=..::::~-.

tit~

~~Cor.~

['~'

b-~

"l

r'~~.,t

For more information regarding the use of SYNC, consult the ttHP 5880AInstrument Manual."

~'-.

12 29

Page 156: HP 7673B - S service manual

Example 1: Single Injector with a Tray

This example illustrates how to do a sequence of runs using the tray and thefront injector.

The following terminal display shows editing five parameters of the autosequence table. These parameters tell the GC that:

• The ilQector is mounted over a packed or capillary inlet.

• The injection volume is 1/1Oth of the syringe volume.

• Each sample vial is analyzed three times.

• The six sample vials are in tray positions 20 through 25.

EDIT AUTO SEQ O~ 0

EDIT AUTO SEQ 4~ 1

EDIT AUTO SEQ 7. 3

EDIT AUTO SEQ 8~ 20

EDIT AUTO SEQ 9. 25

LIST AUTO SEQ

INJECTOR 1:O. MODE CD-NORMAL. 1=ON COLUMN} = 01. PRE-INJECTION SAMPLE WASHES = 22. VISCOSITY = 03. SAMPLE PUMPS = 44. SAMPLE VOLUME m 15. POST-INJECTION SOLVENT A WASHES = 26. POST-INJECTION SOLVENT B WASHES = 27. INJECTIONS PER BOTTLE = 38. FIRST BOTTLE = 20

1:- l~9. lAST BOTTLE = 25

30

6. Attach the Dummy Septum Nuts and the Mounting Posts

a Screw on dummy septum nuts (part number 18597·20580) over theinlets that are not being used with the automatic sampler.

b. Screw each mounting post clockwise into the threaded holes of thebracket. Tighten the posts with a flathead screwdriver.

Ifyou are installing the tray bracket, continue on the next page.

Ifyou are not installing the tray bracket, turn to page 15.

11

=.

·.-3

..~~-c".=;

~ 0:.-;

c~l!~o--~}

- .~

..';1

...~...1i•..

'";~. -. ~

<~

\~~::l

.~'".~

:%.~

~'-¥;

~~~

'~

3.--_.~

;,"~;j

..~

-.-~

.~~='~

~~

.~;~

Page 157: HP 7673B - S service manual

!~;

~.C.. ~.'.rIi'~-

~.•

f

~itit>•".

,."' .''':••••••••••••••••••

t~~~-

;•,...!~..

f

~t[~.":

f,.,t~-.

r/

5. Align the Bracket

a. Attach new septum nuts (part number 19243-605(0) on the packed andon-eolumn inlets that are being used with the automatic sampler. Donot place septum nuts on the inlets that are not being used.

H the GC has a eapi11sry inlet, route the gas lines on the inlet nutover the mounting bracket. Push the inlet nut down and turn it clock.­wise. Tighten it with the inlet wrench.

The gas lines must not interfere with the feet of the injector. Smallwhite circles on the bracket show where the injector feet touch thebrackeL H necessary, gently bend the gas lines so they do not cross thewhite circles.

b. Place the plastic alignment toolls), honeycomb side up, over the inlet(s)ofyour GC. Move the bracket until both alignment tools fit flush onthe bracket. Hyoo cannot make them fit flush by moving the bracket,you may haveto adjust the position of the inlets.

c. Tighten the four screws that hold down the brack.etwith a no. 2 pointPozidriv. Remove and save the alignment tools for future adjustments.

10

Load the six sample bottles into the tray positions 20 through 25. Theinstrument wiD make18 runs (three runs for each of the six vials).

Top YIn If Quadrant

r '\

Start the sequence by pressing (e,"",.,) (~ I00.

31

Page 158: HP 7673B - S service manual

~

A

-~

5i.~

-~

~]

---,:;j

}].,,"-.cj

£4~I-¥,j1

t-::~

.,""~.ili.

~~:~

1i

;1

.~;'-~-•.•...•••:.-..•.•-....•.....

.~ - ,r' .

.~

Hlies for inletposltlu BauD

a Place the bracket on top of the standoff's.

The injector mounting bracket enables you to mount two injectors overnonadjacent inlets-either A and C positions or B and D positions. De­cide which positions you need.

b. Line up the top of the standoffs with the slots on the bracket. Screw thepan-head screws into the threaded holes of the standoffs. Do nottighten the screws until step 5.

When installing the bracket over the A and C positions, use only threescrews.

4. Place the Bracket

Detector

DeItc10r

OterI

Chamel2

C~m111

Enter the necessary changes to the rear injector on the terminal forchannel 2. The sample vials for the rear injector are in tray positions 5through 15.

This requires the SYNC mode to be turned OFF on both terminals. Press-

rnli!IlMiil (~) a i on both terminals.Gl*l R

The sequence for the front injector requires the same ron parameters as inthe first example, except each vial is ron only once (one injection per vial).From the terminal controlling the front injector, edit INJECTIONS PERBOTl'LE to "1."

In this example the instrument is used as two isothermal GCs (bothchannels of a dual channel instrument are used to analyze samplesasynchronously; both channels have different start times and durations).This mode of operation is only practical if the oven is run isothermally.

The following illustration represents the asynchronous configuration.

Example 2: Two Injectors with a Tray

f~;~-~

32 9

-~#

- --~

.:~

.'1c~

Page 159: HP 7673B - S service manual

c. Screw two brass-colored standoffs (part number 18597-20755) onto thebottom of the mounting bracket to act as support legs. The holes havesilver-colored pems (small threaded washers) on the top of the bracket.Tighten the standoffs with a 1/4-in. nut driver.

EDIT AUTO SEQ 7. 1

EDIT AUTO SEQ 8. 5

EDIT AUTO SEQ 9. 15

LIST AUTO SEQ

INJECTOR 2:O. MODE (O=NORMAL. l=ON COLUMN) = 01. PRE-INJECTION SAMPLE WASHES'" 22. VISCOSITY = 03. SAMPlE PUMPS • 440 SAMPLE VOLUME = 15. POST-INJECTION SOLVENT A WASHES = 26. POST-INJECTION SOLVENT B WASHES • 270 INJECTIONS PER BOTTLE = 180 FI RST BOTTLE '" 59. LAST BOTTLE'" 15

EDIT AUTO SEQ 7. 1

LIST AUTO SEQ

INJECTOR 1:00 MODE (O=NORMAL. l=ON COLUMN) = 01. PRE-INJECTION SAMPLE WASHES = 22. VISCOSITY = 03. SAMPLE PUMPS = 44. SAMPLE VOLUME = 15. POST-INJECTION SOLVENT A WASHES • 26. POST-INJECTION SOLVENT B WASHES = 270 INJECTIONS PER BOTTLE = 180 FIRST BOTTLE = 209. lAST BOTTLE = 25

f~~IL..-- ----'

~

The following illustration shows the editing commands and the autosequence tables for both injectors.

I•

IItU

I

o~"'-

t"i'~.

~~'"

~

I

~~.

r.!E"~ .~ ­i'":"

~.If

ft~- ...fi1.. 0'~.-,

r

t

,l:'I;,;

833

Page 160: HP 7673B - S service manual

Load the seventeen sample bottles into the tray positions 5 through 15 and20 through 25. The instrument will make 17 runs (one run for each vial).

Top Vin .. Quadrant

I

b. Replace the screws with the appropriate size standoffs.

• For capillary inlets, use two steel standoffs (part number18597-20520) to support the injector bracket. Tighten them with aIf4-in. nut driver. .· .. B

~-

• For packed inlets, use two small standoffs (part number18597-20745) to support the injector bracket. Tighten them with a318-in. crescent wrench.

~~

.?

>f-

-;/

~~

.~

~1

~

;1tJ

,I"'011_-_ci,

"

t'i;~

- .~~~

~

~

• When there is no inlet in a position, remove the small piece ofsheet metal that covers the position and replace it with the piece ofsheet metal (part number 18597-00245) and two steel standoffs(part number 18597-20520).

(OT_I(~:n rn·

r-J"l

Start the sequence for the front injector from the terminal controlling thefront injector by pressing

Start the sequence for the front injector from the terminal controlling thefront injector by pressing the same keys.

34

Save any standoffs you do not use for future adjustments or differentconfigurations.

7~

-: .--",..1

~~

Page 161: HP 7673B - S service manual

I"'!"'.~.

~r-,['c 3. Attach the Injector Bracket Supports

7~>.

f~

rt~~

~.!~.'

~.

t~.J'- .~~-

t~k~'

~',

171'-t.

6

The injector bracket is supported by standoffs (small hexagonal posts).Where you attach these standoffs depends on the type and position of theinlet. The illustrations in this section show a capillazy inlet in the Bposition and a packed inlet in the D position.

a. Remove the screws that hold down the inlets with a no. 2 pointPozidriv.

Error Messages

Depending on circumstances, the HP 5880Amay print an error messageonits terminal to indicate a fault condition or certain occurrences incommunications. The following is a list of these messages and a briefdefinition of their meaning:

..* ERROR *.. CHANNEL 2 (or 1) NOT INSTALLED

• The START or STOP button on injector 2 (or 1) bas been pushed. butno channe12 (or 1) has been configured in the instrument (nointegrator/plotter 2 or 1).

*** ERROR ***Aura SEQ BO'I"1'LE CONFLICT

• The instroment is running two sequences in simultaneous injectionmode, and the two sequences have called for the same bottle.

*** ERROR ***TRAY NOT INSTALLED

• The STARTAUTO SEQ command cannot be used without the traymodule installed. H no tray is installed. single samples may be run byusingthe [START [injI] AUTO SEQ -J command.

..* ERROR "*INJECTOR 1 (or 2)NOl" INSTALLED

• A START 1 (or 3) AUTO SEQ command has been attempted either withno injector on channel 1, or with the power controller turned off.

*** ERROR *** TRAY MISSING BOTTLE (I)

• The auto sequence calls for a bottle (Jf) in the tray. but there is none inthe requested position.

35

Page 162: HP 7673B - S service manual

36

.** ERROR **. INVALID BO'ITLE IN INJECTOR 1

• The tray ann is attempting to deliver a bottle to an injector thatcurrently contains a bottle.

••• ERROR ••• TRAY FAILURE

• Turn to chapter 5, "Troubleshooting', in the operating manual totroubleshoot the problem.

**. ERROR **. INVALID BOTI'LE IN TRAY POSITION (I)

• The tray arm is attempting to move a bottle from the injector back to atray position that currently contains a bottle.

••• ERROR ••• INJECTOR 1 (or 2) FAILURE

• Turn to chapter5, "Troubleshooting", in the operating manual totroubleshoot the problem.

~

.-~~

2. Install the New Top Cover

a Install the right cover (part number 19324-00(70) with two pan-headscrews.

b. It the GC hasa capillary inlet lightfs), remove the light and attach it tothe new lamp holder (part number 19321..()()()B(). Disconnect wires androute them through the tie wraps where necessary,

c. Attach the lamp holder and the left cover (part number 19321-0(010)with two pan-head screws.

'II I I~

I ,'"-~

~ 11I1111111111111I11I11111I11I111I11111111111I 1111111111a. 11I11

5

-(19':

,~.~

':t!_:i!l

J'~

J~,

]"_-.i

·i·..•

···'•.·····•.~--<:-.-"

,:;.-. -:.-~~

. ~

.:~

.1.

Page 163: HP 7673B - S service manual

I

~.'

"..

~

~il.=­~

I=j-c'

~~~.~

J~:.

7<.-_

~.!<c.

If~:

~

~t!'r

WARNING

TURN THE POWER SWITCH ON THE GC TO "OFF" ANDDISCONNECT THE POWER CORD BEFORE INSTAlUNG THEAUTOMATIC SAMPLER BOARD IN THE GC. ALLOW THE INLETSAND OVEN TO COOl BEFORE INSTALUNG THE INJECTORMOUNTING BRACKET. THE METAL SURFACES OF THE GC AREVERY HOT AND COULD BURN YOUR SKIN.

I CAUTION ITurning off the power immediately eraSes data stored In theInstrument's memory. To save this data. list the set points andprogrammed events by pressing (UST] [ENTER]. Reenter thed8t8 when you have completed the Installation.

1. Remove the ExIsting Top Cover, Side Panel, and Septum Nuts.

a Using a no. 2 point Pozidriv, remove the top cover and left side panel oftheGC.

b. Remove the septum nuts on the packed and on-column inlets.

c. For capillary inlets, loosen the inlet nut with the inlet wrench. Turn thenut counterclockwise until it is off the threads and lift it out of the way.

JCAUTION ITum off the gas flows before you remove the Inlet nut to &voIddamaging the pecklng In the Inlet liner. Avoid contaminating theInlet while" I, uncovered. The braze joints that hold the ga9nnes to the Inlet are fragne.

4

Page 164: HP 7673B - S service manual

2Installing the Brackets

-~

'",'~

-Pull out the yellow.worksheet from the. center of this link manual. Use thisworksheet to organize the hardware and tools for each step. Full-scaledrawings of the screws help identify which ones you need for each step. Ifyou do not use the worksheet for this complex task, you may not be able tocomplete the instaIlation successfuIly.

The necessary hardware is packaged in several locations. The injectorbracket is packaged with the controller module, the tray bracket ispackaged with the tray module, and the top cover is packaged with theelectronic materials.

~.:.".~-"-~I':d.i'lJ

1~

":1;

;1

3

....­-... ;1,........

'...........

1........0= ......

1............-.-..

2 ...........c.r

ExpIDded view with aI pats

oi i r-w

Step by step

Toom you needforeach step

PCH1S yoo need for eaen step

<,}

~.

,',,", :1'. ; . .

I' '. '.',' ~ ~.!

I '.' I• -....I •

I J>d ,. ri

--

Ifany of the hardware is missing, contact your HP representative.

Page 165: HP 7673B - S service manual

5;c~c

tt~ 1

~

This link manual contains the following information:

• Instructions for installing the mounting brackets for theHP 18593B Injector Module and HPI8596B Tray Module on anHP 5880A GC. Turn to page 3.

• Instructions for installing the necessary electrical components (printedcircuit board and EPROMs) on an HP 5880A GC. Turn to page 16.

Before you begin check to see that you ordered the correct option. Thefollowing table shows what features each option supports and theassociated part numbers.

This "link" manual contains information specific to using the automaticsampler with an HP 5880A Gas Chromatograph (00. Refer to theHP 7673 Automatic Sampler Operating Manual (part number 07673-6(995)for general operating information, including: mounting an injector, selectingconsumables, or calculating the maximum number of samples you can run.

~

iif·· ."'(

~

~ Before You Start~-~.~

~~~

f:

lc... ··..Il:iio~.

i"·

fr

Optien IIIIlber 1932tA 193218 19321C

AlS, Loop, D/A. AlS, Loop, WA. ALS,loop, DlA.Support" featanl V" D1ver, EFC, NP. VtM Drh'er, EFC. V<rNe Driver, sec,

fP'O. Extsma II1XJt and 00 ald DC!(NO). andSCC

Old EPROMI1818-3119 1818-3121 19335-800101818-3120 1818-3122 19335-80015

NewEPROMs 19300-80070 19300-8000O 19300-8011019300-80080 19300-80100 19300-80120

• Sample diagrams for connecting the cables for the automatic sampler toan HP 5880A GC. Turn to page 20.

• General operating instructions for using the automatic sampler with theHP 5880A GC. Turn to page 21.

~ Common error messages. Turn to page 35.

~,

@ Copyright 1989, 1990 by Hewlett-Packard CompanyAll Rights Reserved

2

Page 166: HP 7673B - S service manual

~~-_~r-._~?~i~~~~it~!:#:&0l1i$&~*) -'J. HEWLETT-PACKARD .~

Manual Part No.07673-90235

~HEWLETT

~PACKARD

PrInted In U.S.A.April 1990

r.tI

I

t\

{ =- D.·(ll ,..

UnkManual

HP 7673 Automatic SamplerInstalling and Operating the Injector

and Tray on HP 5880A GCs

-.-~

:-~

\·4"~-I-11

.,,0;

~j

~."'¥

~-~~

.-:'--~

-:-~~

--~

.~

Page 167: HP 7673B - S service manual

7~....",~..-.

2 seff-twilg screwS (0624-os4s)( Ofawn fuIsea )

3 standoffs (18597-20520)( lXawn ftJ scale )

1/4-i'1. ru ckiver

2 ilieCtDf bracI<et bust*1gs (18597-20735)( Drawn flj SClY )

1. i1}ector bracket (18957-00175)

o 0 0 0o

g~

til~""l2ii}e

oo00

2 Detach the !Ret Nut i1etwrench-3 Piace the Bracket

~ 2 pal headscrews (0515-0~n5)

( Ofawn UIscale )

4 Attach the IRet Nut i1Ietwrench-5 Aligl the Bracket

~No.2 Pm Pozkiriv

2 aIigi. nent tools (18597-40030)

1 Attach the Bracket~

6 Attach the MOlilting Posts flathead saewctiver

£ ~rl1OlI1Ii1g posts (07673-20865)

• ( Drawn ftJ scale )

D

~H.·I "'- '

I'~"I

II

II

II

I,II

II

~.~

~,.

~.

f:c'I

c: Exploded diagcrn

~:

l!F,c,~.

~

o~...-~-

t~

"

Page 168: HP 7673B - S service manual
Page 169: HP 7673B - S service manual

f~/

The gas lines must not interfere with the feet of the injector. Smallwhite circles on the bracket show where the injector feet touch thebracket. H necessary, gently bend the gas lines so they do not cross thewhite circles.

t1Figure 7. Route the gas lines over the bracket and reattach the inlet

nut.

b. Push the inlet nut down and turn it clockwise. Tighten it with the inletwrench.

4. Attach the Inlet Nut (Read only if the GC has a spliUsplitiess Injection port.If not, skip to step 5.)

a. Route the gas lines on the inlet nut over the mounting bracket. Formost configurations, the gas lines should be routed over the front orrear of the bracket.

b. Line up the top of the standoffs with the hole and slot on the bracket.Screw the pan head SCI'eVr'S into the threaded holes of the standoffs. Donot tighten the screws until step 5.

Figure S. Placing the mounting bracket on the GC.

a. Hold the bracket with the bend on the left side and slide it onto the cyl­inders that you attached to the side wall in step 1. Allow the bracket torest on top of the standoffs.

3. Place the Bracket

~.

t.~ -~rif:f:

F)

~f

t

~.~..;.i.f

t

t<

Figure 6. Start the pan head screws into the standoffs.

8 9

Page 170: HP 7673B - S service manual

';. -.~

,.-:<~. :~-~-

._~

5. Align the Bracket

a. Place the plastic alignment toolfs), honeycombed side up, over the in­letts) ofyour GC. Move the bracket until both alignment tools fit flushon the bracket. Hyou cannot make them fit flush by moving thebracket, you may have to adjust the position of the inlets.

b. Tighten the two screws that hold down the bracket with a No.2 pointPozidriv. Remove and save the bracket alignment tools for futureadjustments.

~-~

;-~'~:_;1

..~e

..'.'#11'.;"/_~

. ~~y--:,,~

:~~

:::~

.~.:,";" ', ...

' ..•...........•......

.~t~,-::~

Figure 4. Detach the inlet nut and lift it out of the way.

7

I CAUTION ITurn off the gas flows before you remove the inlet nut or youmay damage the packing In the inlet liner. Avoid contaminatingthe Inlet while It Is uncovered.

a. Loosen the inlet nut with the inlet wrench. Turn the nut counterclock­wise until it is off the threads and lift it out of the way. The braze jointsthat hold the gas lines to the inlet are fragile.

2. Detach the Inlet Nut (Read only if the GC has a split/splitless injection port.If not, skip to step 3.)

For a splitlsplitless inlet, route the gas lines over the bracket. Ifyou donot, you cannot replace the inlet liner without removing the bracket.

'~

Neede 9gLidetop 05eptunnut ~

base assembly~~m-~

Figure 8. Align the bracket and tighten the screws.

Ifyou have an on-column inlet on an HP 5890A with a cooling tower(Part No. 19245-20540), replace the cooling tower with a septum nut baseassembly (Part No. 19245-80520). Save the cooling tower to realign thebracket. H you have an lIP 5890 SERIES IT cool on-eolumn inlet,remove the needle guide top (Part No. 19245-20670). Save the needle guidetop to realign the bracket.

10

Page 171: HP 7673B - S service manual

~-'

.......[f'

~r~.

~.<

~:.-

f"'.~_.:

Ii;

t?'

i> •.lrIt~:el.1;'1_.

~p.

~..,...."....t"I',t~"r

f

.;.,.

;,.1,

t.~ 6

e. Screw one standoff onto the bottom of the mounting bracket to act as asupport leg. The hole has a silver-colored pem (small threaded washer)on the top, right-hand side of the bracket. Tighten the standoff with a1/4-in. nut driver.

~Figure 3. Attach the standoff to the bottom of the mounting bracket.

Ifa purged packed inlet is installed on the GC, this standoff couldinterfere with the heat sink (metal fins) around the inlet. Ifnecessary, useonly the left-hand side of the heat sink.

IfHP 5890 SERIES n cool on-column inlets are installed in both thefront and rear positions of the GC, do not install this standoff. Without thisstandoff', the stability of an injector in the rear position may be affected.

6. Attach the Mounting Posts

a. Screw each mounting post clockwise into the threaded holes of thebracket. Tighten the posts with a flathead screwdriver.

Figure 9. Attach the mounting posts.

11

Page 172: HP 7673B - S service manual

3Checking Your Work

Double-check. the following:

o If the GC has a splitJsplitless injection port, check the inlet nut.

o If the GC hasan on-eolumn injection port, be sure that you havereplaced the cooling tower with the septum nut base assembly.

o Ifgas lines are routed over the bracket, be sure the lines are not nearthe white circles.

o Check that the following are tight:

o Self-tapping screws.

o Pan head screws.

o Injector mounting posts.

Ifyou are installing a tray, install the tray bracket before mounting theinjector.

For detailed information on mounting and connecting the injector, seechapter 1 ofyour operating manual "Setting Up."

d. Screw two of the standoffs (small hexagonal posts) into the threadedholes located near the inlets. Tighten them with a 1I4-in. nut driver.

DII

I

Figure 2. Attach the standoffs near the inlets.

If the rear inlet is an HP 5890 SERIES n cool on-column inlet (AccessoryKit 19245B or 19245C) on an lIP 5890A GC, use a short standoff' in the rearposition (Part No. 19245-20760). This standoff is packaged with theaccessory kit.

..~

---:::f.. ,.,-~~

t.'~

.:;,~

~

~-p

~

/~:'1

t.·•.:.~l:1

J

".._~12 5

Page 173: HP 7673B - S service manual

~~.~:::-..

~~ 1. Attach the Injector Bracket Supports."_Ie'

4

~

Rea-

~ ~ ~ ~ ~,~

~~~ ~~

Remote connector

~~

r.:::-:l~

Ir.:::-:lL:.:::J

~IDfOO

!NET and REMOTE cable connections are located under the right top panelof the HP 5890. Other connections, such as the RS-232-C connector, arefound on the back.

The type of data-handling device and communications configuration thatyou use determines how you connect the automatic sampler to theHP 5890 GCs. This section contains some general information and twoexamples of cabling for the instrument network (lNET).

For cabling the automatic sampler to data-handling devices, see the link.manual for the specific device or system.

For detailed information on cabling to the HP 5890 SERIES II GasChromatograph, see the Site PreplInstallation Manual(Part No. 05890-90280).

Signal Cable Connections

Connecting to HP 5890 GCs

Figure 1. Attach the cylinders to the side of the GC.

b. Lift up the right-hand side of the top cover.

c. Using a 1/4-in. nut driver and the two self-tapping screws, attach thetwo small cylinders onto the left side wall of the chromatograph.

a. Remove the port pan by inserting a finger into the U-shaped cut on thepanel and pulling up.

tt-rt

~,I~.:

f

;.,..... -:--

t=~

rts,~)

f

~ ..~'.-;~-

INET comectors

FilUre 10. Signal cable connections.

413

Page 174: HP 7673B - S service manual

Instrument Network (INET)

The following diagrams showtwo !NET configurations for connecting theautomatic sampler to an HP 5890.

2Installing the Injector Bracket

..j~

.~

j)_~_1'

. -~-1:\~-F:

.:~

Pull out the yellow worksheet from the center of this link manual. Organizethe hardware and tools for each step. Full-scale drawings of the screws helpidentify which ones you need for each step. ITany of the hardware ismissing, contact your HP representative.

",.~.•.•...'...~"··1- :~

-.;;:;

'~

':;!

."~

:~,:::!

]

,-;;

,

~

1

.~'1

...-~51

,:,,~

.~"""3.1

3

--

.. 3'I'w .....

......-

2 ~~, :111 ....)

z........ <WW......

''''_~_

:2' :E.,.:~ra-- )._-­( .........)

,- _.-( .....

Exploded view with aI pats

2 _ .......

5 ....._~

~

4 ......... _

3 _ .. - D-

8 .-:II ......,.. .............

f a -­( .........)

Step by step

Pcr1s you need foreach step

Tools you need foreach step

--Conroller

IN

Catrtter&~

OUT II,. ,~.

.~=~------

OUTI IN

t - - - • - - - -I

II

I

I

I

I

- .!

5890sma I

HP 5890 SfRIES I Gas~

o

o

ASimple INET Loop,- . - - - - -- - - - - - - - - - - . - - - - - - - - - - . -.

Figure 11. Cable connections for a system using INEf.

I

I S<fI1lIerI "I

I

II

I

II

C jI

14

Page 175: HP 7673B - S service manual

t C

f;ctE1,:_".--'.

t

ic 1""

CorVoIer&~

secm:IlrtePm

IIRenDellStatIl BCO

I II

~;~~~ o~IIN_~...~.---

••

-Aiibj-""CMIlJ I

I

~.....'--=----

•••••I

-~

5890SOlD.

liP589lI SERES I Gas~

ASimple INET Loop

·.---.-----.. ----.--..-. -- -..

o

6

. ...

••••I

•••c'

I 5

.,

I,

WARNING

Allow the inlets and oven to cool before InstaJllng the injectormounting bracket. The metal surfaces of the GC are very hotand could bum your skin. If you have a split/splitlessInjection port, you could damage the column.

After you have completed the installation, place this link manual in thebinder ofyour operating manual for the HP 7673 Automatic Sampler(Part No. 07673-60995) after the tab for appendix C "Link Manuals."

• Instructions for installing the mounting bracket for thelIP 18593B Injector Module.

• Sample diagrams for connecting the cables for the automatic sampler tothe lIP 5890 Gas Chromatographs (GCs).

Before installing the injector mounting bracket on an. HP 5890A, you mustfirst install the split top cover for the HP 5890 SERIES ITGasChromatograph.

tBefore You Start'1::,'-.

k

: This manual contains the following information:~

~..•

~.,.i~

~-.

~,~<

~

t~ _

~

~Figure 12. Cable connections for a dual injector system using INET.

e Copyright 1989 by Hewlett-Packard CompanyAll Rights Reserved

,2 15

Page 176: HP 7673B - S service manual

~

~

;..~

.j-;;

'?.t:'

"'"-~.l..;

-,':;~

•.'··.··I···.~·.;.~-,

~

;l~-;~

..-~

Cc~,.,".~.- ,

,~

,~-~

....~

( =-= 0='

{ . =-= »

. LinkManual

HP 7673 Automatic SamplerInstalling the Injector Bracket and

cables to 5890 GCs

HEWLETT-PACKARD

PrInted In U.S.A.October 1989

~HEWLETT

~PACKARD

Manual Part No.07673-90195

Page 177: HP 7673B - S service manual

00 o::::J 0 CWOiI II

~ r-::-lp"'!!IU L.....IL.J .-----

.. ~ O~ • JoEE ~;' 8 .,r-'II- - ..

q

c2

1

Contents for Accessory Kits HP 18599R and 185998

1

1

lI;E:!:J 10

(>

;; I E ! IF E ~ I If I·IJ I I,COa:::JELHIJELlLD F I fF I I

> i ~[TI~~ITJ~ I IF I I~ I ... ~ = • I rr1

lZ

1

~

~Itontents for Accessory Kits HP 18599H.'~---------------------------'I

~u '~I~ I <0

~ I ~~ • I~~if> rJ O))~.•.• ao-...,. • ~ II I..l. liU Ill'

• J>'-----.--'lo~~~~p..D~F (tz

.,~o(). ;O~iii~R ~[lp'~o~ _.. -Ii[!] IU I ~

3

~-";-~..

•..

f

fItem Description Part No. QTY

1 ControllerlHP-IB Board 18594-60110 1

HP-IB Cable 8120-4654 1

~ M3 Screws 0515-0912 6 ..~,

T#6 Washers 2190-0065 6

Bezel, Blank 18594-00115 2

Bezel. Blank 18594-00116 2

Item Description Part No. QTY

1 RS-232-e Board 18594-60080 12 Controller EPROM 18594-8NNNN 1*

3 Controller Board 18594-60060 1**

RS-232-e Manual 18594-90300 1

* Included in HP 18599R kit only. NNNN indicates the last four digits of thepart number for most recent EPROM.

.. Included in HP 185995 kit only.

Page 178: HP 7673B - S service manual

"R~------4!

Contents for Accessory Kits 18599L and 18599M

'0 ,~

f1~:;.

.:~

-1i

i

,.

._-~1

,- .~-""

-}~

'%~---i<~

;-;;

',1-~

'-t

~-2

':':1--:i~

_"-;rt~

·..··.l'~I;1-~.:

···I~~·'L~

~~rro r-'ii' o! II; II; iI; II; .. I~ =t?1 i ~-~~.- ,I%J L....:!...JOi'i'tt¥l'l~ ~ - c::::::::s ~ ~ •• ;1!I C? l,IMWL.MtJr'rt.JMJW c:::=3 IoN'"

~ fmillcl:51 -I :"&~ • II ~((((ICD);.; ..•~-¥ .1 . •••((fNf9~ ~ b - Sf~ .g~.mm ~O

); ill EEl: QiP- -00+· ((0)~~o~ !l ~J 0;' ~. ··b

1

Contents for Accessory Kit HP 18596A

{

. ~:.iiiis rn

m.-=""'; "

m II 8,,§ _ ffiIrn 8o:J • CD

= !ijrnl rn8cq~ ~EI;!~ CD i jlO

3

1

Item Description Part No. QTY Item Description Part No.

1 Loop Board 18594-60090 1 • 1 Tray Logic Board 18594-600752 Controller EPROM 18594-8NNNN 1"3 Loop .y" Cable 18594-60650 14 Controller Board 18594-60060 1"" r

" Included in 18599L kit only. NNNN indicates the last four digits of thepart number for most recent EPROM.

• " Included in 18599M kit only.

QTY

1

Page 179: HP 7673B - S service manual

~

fk

ooסס00000180010001111010000101001100011101000010011101001011011000110101110011111eooo10001100101001110100101011011010111110001100111010110111110011101

•12345,789

10111213141516171.1.2021222324252&272829

• Select an address from 1 to 29

• Must be different Ihan aD otherInstrument HP-IB addresses.

Address SwItches

~

CD

N

­CD

-e::t

Set the HP-IB address switches on the rear panel as shown in the followingillustration. The address should be unique for each instnunent.

In the illustration the example switches on the right are set to address 8.This is the usual address for the first automatic sampler connected to aChemStation. For an additional automatic sampler, choose any unusedaddress.

For Use With an HP ChemStatlon (HP-UX Series)

The controllerlHP-IB board contains a control section and acommunications section. It is used to control the injector(s), the tray, andother devices attached to the controller.

The board communicates with lIP ChemStations via an lIP-IB cable. Itcommunicates with lIP 3350Lab Antomation Systems (A series only) via anRS-232-C Interface Board in the COMM slot and an RS-232-C cable to anlIP 35900B Dual Channel Interf'ace.

This section includes information for configuring the board with either aChemStation or Lab Automation System, for connecting the cables, and forinterpreting the turn-on behavior of the controller when the HB-IB boardis installed.

Iddms IWIcIIn

The following illustration shows the location of the address switches on thecontrollerlHP-IB board. Configure the board for either a ChemStation or aLab Automation System with the fonowing instructions. Be sure toproperly ground yourself to eliminate any static discharge problems.

~.

f·-~

.J~.

, "

~-

4~~;Hp·IB Board

~'

t Configuring the Board

t.~-

~,~~; .

"I~..

Turn back. to page 6 and install the HP-IB board.

r-i·

• Remove the plastic cap from the connector on the back of theControllerlHP-IB board. Save it for later use.

~-

• Install the controllerlHP-IB board in the CONTROLLER slot.

12 13

Page 180: HP 7673B - S service manual

.~

l-~~

-,~

·~-A

"····.·····.'::'···.1·······'·~•.····.·..·.·.·.··'··

~.;

-;

~~

z:

.~

,

.~;!~~

.~-,~

~~~t~·

'1._~~:t

E -'

1. Plug the tray cable into the connector of the tray board in the left-handslot on the back panel of the controller.

2. Secure the plug to the connector with a small flathead screwdriver.

Ifany of the following are true, the controller board is not compatible withthe tray board. Contact you HP representative to order the necessary part.

• The CONTROLLER slot hasa board in it with no connectors on thebezel (ie., a flat metal plate). The part number for this board is18594-60020. Look on the lower right-hand corner of the board.

• There is an EPROM on the controller board and its part number is18594-80010.

Connecting Cables

o

1

Baud rate setto9600

Jmper

4 2 1

2. Set the HP-IB address switches on the rear of the HP-ill board toaddress 8 as shown in the following illustration.

When configuring for use with a Lab Automation System, you mustconfigure the HP-ill and RS-232-C boards.

This configuration requires an lIP 35900B Dual Channel Interface. Themodel designation is on the Serial Number plate on the rear panel.

1. Locate the WI BAUD RATE jumper on theRS-232-C board (partnumber 18594-60080). Set the jumper to 9.6K as shown in the followingillustration.

For Use With an HP 3350 Lab Automation System (A series only)

Turn back to page 6 and install both boards in the controller.

• Install the RS-232-C Interface board in the COMM slot and the HP-ffiboard. in the CONTROLLER slot. Leave the plastic cap on the HP-fficonnector.

14 11

Page 181: HP 7673B - S service manual

~~.~.

t~.

::-.~

[;:Jr 3 Connecting Cables

With HP 3350 Lab Automation Systems

Connect an HP 17255D (or equivalent) RS-232-C cable from the connectoron the RS-232-C Interface Board to an instrument port on the rear panel ofan HP 35900A Dual Channel Interface.

With HP ChemStations (HP-UX Series)

1. Connect an HP 8120-4654 (or equivalent) HP-m cable from theconnector on the back of the board to either an instrument HP-IB porton the ChemStation or an HP-m connector on a cable that eventuallyreaches a ChemStation instrument port.

DO NOT connect the cable to an HP-m ChemStation disk or printerport.

tTray Boardi;.. The following instructions include checking the configuration of the~' controller board and connecting the tray cable.

....•.. For information on mounting and operating the tray, see the automatick sampler operating manual. For A modules, see part number" 07673-60990.t. For B modules, see part number 07673-60995.

~" ,;:-cheCklng the Configuration

:" The tray board you received with this accessory must be able to•.• communicate with the controller board. There are a few cases where the

~.'~' controller board must be modified.~

tl - Locate the board in the slot labeledC?~~R. Ifany of the followingf~ are true, turn back to page 6 and continue installing the tray board.

l....·.··.. . · There is a U6 EPROM on the controller board and it has a part number~"., whose last four digits are greater than 010 (i.e., a part number greater

. than 18594-80010).

• A controllerlHP-m board is in the CONTROLLER slot. A picture ofthe board is on page 8.

• A 5880Acommunication board is in the CONTROLLER slot. A pictureof the board is on page 8.

~:

HP7673ACIIdrtIer

pesslWylIIbtr

iJstrumelltSCllenStatitn

RS-232Pert

RS-232-CCOIIJICtIr

~.

~

HP 7613AtF359008 .

Ctmreler DualCbameIlItet1act

EPROM

-c::=J- ct -c::J-a-c::=J- .. ~ • -c::=J- a

<) ••

.M nn~r U. U U-O-~::'Q? rJ ~rtL-~ ~ U ,.

~

lr~a;

..

1015

Page 182: HP 7673B - S service manual

Turn-on Behavior

When you turn on an automatic sampler containing a co:n.trollerlHP-mboard, the fault lights (i.e., LEOs) on the front panel remain on untilinitialization is complete. It may may take from one to ten seconds,depending on what other boards are installed.

Red I I Green

-==-

-During initialization, the controller runs a series of intem.al diagnostics. Ifa problem is found, initialization stops and the read fault light flashes adiagnostic code. Possible codes are:

• 2 flashes-defective RAM on the controllerlHP-m board

• 3 flashes-defective ROM on the controllerlHP-m board

• 4 flashes-defective RAM on a bar code reader board

• 5 flashes-defective ROM on a bar code reader board

• 6 flashes-bar code reader cable not attached

• 7 flashes-bar code reader configuration switch set to TEST

The red fault light also flashes to indicate problems with the tray board. Ifthe tray board is installed but the~ is not attached, the red light turnson and remains on. This prevents seeing any of the diagnostic flashes fromthe controllerffiP-m board.

16

4. Attach the Top Plate and Top Cover

a. Slip on the metal top plate.

For an A module, complete this step.

b. Screwon the five screws with a No.2 Point Posidriv.

For a B module, complete these steps.

b. Screw on the four corners screws with a No.2 Point Posidriv.

c. Tighten the screws of the bezels with a No.1 Point Posidriv.

For a both modules, complete these steps.

d. Slip on the controller top cover.

e. Screwon the two screws with a No.1 Point Posidriv.

5. Connect Any Cables

For instructions, turn to the page for the board you are adding.

For the tray board, turn to page 11.

For the controDerlHP-m board, turn to page 15.

For the Loop board, turn to page 21.

9

..,-~

~

-j'::'_'i'-

~j-~~~: ;-'~

... ~- 'I:;

'}j-~:~

~~~

._~~..~,:

.-;~

-_1

.~

11~

~::-:-'

.~

.~",4. ~ j

j~~._~

~

1--;-~

:,j-:~

Page 183: HP 7673B - S service manual

~

j;.'

p-BorflI Name I PertNLJTiler Connectors ControDer Slot 6

~,"~;~-"

~~-

Tray Board18594-60075

cO ! TRAY RS-232-C Board

1- COnroIerIHP-IB Board18594-60110

CONTROlLER This section includes information for configuring the board. Be sure toproperly ground yourself to eliminate any static discharge problems.

LoopBoa-d ~~ - .OJ COMM18594-60090 • •

588CA ComTuicatioos ro eI Fi:[-. I~~ CONTROLLERBoard 18594-60050 L..:_....J.II......-'.IL-- ~

Bald rate settI !I6GO

For initial operation information, see section 2 of the "HP 7673 AutomaticSampler with RS-232-C Serial Interface Manual" (part number18594-903(0).

1. Locate the WI BAUD RATE jumper on the RS-232-C board (partnumber 18594-60080). Set the jumper to a baud rate that is supportedby your computer (i.e., 300, 600, 1?00, 2400, 4800, 9600 or 19200 baud).

Configuring the Board

r'

AU>:

COMM

co ••.-, . I~J COMM

~~~oo LoBpI-ad !~._O~

IHET Board18594-60030

RS-232-e Board18594-60080

~"t

".tfl'i'

l_

i""

mBoard18594-60040 cO .... IIIII~J AUX

In general, higher baud rates will result in faster communication andshorter delay times, but sometimes running at very high baud ratescould make it difficult for your computer to keep up with transmissionfrom the HP 7673. Some experimentation may be necessary.

For more information on band rate selection, see section 3 of the "HP7673 Automatic Sampler with RS-232-C Serial Interface Manual" (partnumber 18594-90300).

8 17

Page 184: HP 7673B - S service manual

.;-i~

''i.

'.J~

,,'1

"04~., :'

",'-~1

,~~.~

."q

7

Blaltbml

A

SlIt

SlIt

c•••m..

B

~ _b_1

<,

a. Note the orientation of the EPROM currently installed. Removethe installed EPROM.

b. Install the new EPROM (l8594-8NNNN) provided in the kit.

e. Re-instaD the board into the CONTROLLER slot.

d. Turn back to page 6, remove any board or blank bezel from theCOMM slot, and install the RS-232-C board in the COMM slot.

IF NNN is less than 120, remove the board in the CONTROLLER slotand lay it on a flat surface.

IF NNN is greater than or equal to 120, turn back to page 6, removeany board or blank. bezel from the COMM slot, and install the RS-232-Cboard in the COMM slot.

Tum back to page 6, remove any board or blank. bezel from the COMMslot, and install the RS-232-C board in the COMM slot.

• Ifyou are using kit number 18599R, go to step two.

• Ifyou are using kit number 185998, go to step three.

2. Locate the board in the slot labeled CONTROLLER There is anEPROM on this board with the part number written on it. Withoutremoving the board find the part number on the EPROM. The partnumber should look like 18594-80NNN.

<> • • • u8l~ ~ ~~:~.~ G· EPROM

<=J-" -e::J- .. '---.. <=J- a <=J-.. <=J- OJ

3. Remove the board in the CONTROLLER slot. Install the controllerboard (18594-60060) provided in the kit in the CONTROLLER slot.

18

Page 185: HP 7673B - S service manual

~CcL~

~.:

~~

F·3. Configure the Board(s) 7

r:'~.

~~

~

~ -.

f~

lr"e.:~

~ -~-

~10-_"~:

tt~ ~.

;j';-,0:;-

~

v

~-

J~

6

For instructions, turn to the page for the option board you are adding.

For the tray board, turn to page 10.

For the controDerlHP-m board, turn to page 12.

For the RS-232-C board. turn to page 17.

For the loop board, tum to page 19.

Install the Board(s)

a. Determine in which slot the option board belongs from the table onpage 8. If there is a board or blank bezel (i.e., metal plate or cover) inthe slot, remove it.

b. Line up the tab and plug(s) on the bottom edge of the board with theslot and eonnectorfs) on the base or the controller. Press the boardgently into place.

c. Be sure the board is seated correctly at two points:

• The tab of the board is in the slot.

• The plag(s) is firmly seated in the conneetorfs).

For an A module, also complete this step.

d. Using a No. 1 Point Pozidriv, attach the boards bezelto the back of thecontroller with two screws.

For a B module, also complete this step.

d. To be sure the front bezeltab is seated, wiggle the bezelback. and forthwith the tip of your fingers while pressing down on the board.

Loop Board

This section includes information for configuring the board. connectingcables, and setting run parameters. Be sure to properly ground yourself toeliminate any static discharge problems.

Configuring the Board

• Hyou are using kit number 18599L, go to step one.

• Hyou are using kit number 18599M, go to step two.

1. Locate the board in the slot labeled CONTROLLER. There is anEPROM on this board with the part number written on it. Withoutremoving the board find the part number on the EPROM. The partnumber should look like 18594-80NNN.

<> • • • ~ •

8l~ ~ ~~~.~ G· EPROM

-c:::J- r4 ~ a

'"-c:::J-a ~. ~ a

IF NNN is greater than or equal to 120, turn back to page 6 to completethe following:

a. Remove any board or blank bezelfrom the COMM slot.

b. Install the loop board in the COMM slot.

19

Page 186: HP 7673B - S service manual

~~']

..:-~~

IF NNN is less than 120, remove the board in the CONTROLLER slotand lay it on a flat surface.

a. Note the orientation of the EPROM currently installed. Removethe installed EPROM.

b. Install the new EPROM (18594-8NNNN) provided in the kit.

e. Re-instaD the board into the CONTROLLER slot.

d. Turn back. to page 6 to complete the following:

• Remove any board or blank bezelfrom the COMM slot.

• Install the loop board in the COMM slot.

2. Remove the board in the CONTROLLER slot. Insta1lthe controllerboard (18594-60060) provided in the kit in the CONTROLLER slot.

Turn back to page 6 and install the board in the controller to completethe following:

a. Remove any board or blank bezelfrom the COMM slot.

b. Install the loop board in the COMM slot.

Screws

-~...•.s:':.~

J::1--.iL;.~

';~

~

·····~·;······':··.·:.1..··;...5.~

;~'-~

-_J~

..~

~~,*

- -;~

-~

~~~",.

Front .qe

j~~

'l

,~1-:~

5

SCI'8W$B

HP 185948

Screws

20

Page 187: HP 7673B - S service manual

fbi":~

~: 1. Remove the Top Cover Connecting Cables~.

~.

t~,:.-"

This task is completed the same way for either an A module or a B module.See the illustrations on the page 5.

a. Remove the controller top coverby removing two screws located on therear edge of the plastic top cover with a No. 1 Point Pozidriv.

A single "Y" cable (18594-60650) is used to connect the HP 7673 AutomaticSampler to the 18653B SlECM.

1. Connect the 15-pin connector to the loop board in the controller andsecure it with the screws.

l2. Remove the Top Plate

~ For an A module, complete these steps.

it

~C,_.!".".~

a. Remove the five screws of the top plate with a No.2 Point Pozidriv.

,:

2. Plug the two plugs of the Y end of the cable into J4 and J5 on the HP18653B SlECM box. These connectors are keyed to define their properlocations.

J3 on the 18653B is used to connect remote start signals between thegas chromatograph and the 18653B.

GC

111&53­6012.5

~~ .SI!..C~ _ • ...J

1t5S14 60650

~I~ r'-'-'-'~'\.o~J3 I I

115M-lllXl8O

.-n«face Board

I

I

I

I. .U'P.7!!,3 . !:!os~_ • .J

r"-"-"--_·'

For a B module, complete these steps.

a. Remove the four screws at the corners of the top plate with a No.' 2Point Pozidriv.

b. Loosen the four screws that hold the boards and bezels to the metal topplate with a No.1 Point Pozidriv.

c. Remove the top plate by lifting the front edge and sliding it toward thefront of the controller.

b. Lift off the top plate.

4 21

Page 188: HP 7673B - S service manual

2

Pull out the yellow worksheet from the center of this link manual Find thecontent list for the accessory you ordered. Check the hardware youreceived against the list of hardware on the yellow worksheet. Ifany of thehardware is missing, contact your lIP representative.

j:1

N-:tel

l'"i'l.~

"

';':'

.._-:f

•">i

'··.·1··.'.·.

i-.;:.

1

-i.

-~

.~

"ill~J.~

~~

.."-.'!.'..]~

'~

~~

.'"

.~-

,·;1}4I,1.~

-~~

3

~

1I·11-

.....UI5M-6IIIlSII­1__

~

~;::q-- / __!---.in':'{';;~~ '. ~ ~. ,,:.,,:;~'L.:....t'~ '.Yto;-:'O,;.1'.";­~~~r·:~,,~'. 1 • : I . r-:. ..- ';.

~1il;I:.01--"- .J •!li~~t-~,- ....~;r~I-.-'"T'I' - I ...--nn;:n····,i~·

'0

- --.­I ~_

Z _RX*l1 ~"T'c-.• e-._

e-te.ts for Accessory Kits 19599L -.:I11l5ll9M

• __ it l.II5ML '*..."

- ~ .. 1J5IfIII '*..

You need a No. 1 Point Pozidriv and a No. 2 Point Posidriv to completethese instructions.

Installing a Board

Once the datasystem detennines that a run has finished and the next oneshould be started, it sends the next bottle number to the automatic samplerindicating that a new run is being initiated.

The Loop Board provides communication between the Lab AutomationSystem (HP 3350A and lIP 3357A) and the HP 7673 Automatic Sampler.

Communication is limited to passing the next bottle number to theautomatic sampler. All other parameters need to be entered via the controlswitches on the front panel of the HP 185938 Injector Module. For moreinformation on the control switches, see chapter 4, "Standalone Control" inthe operating manual (part number 07673-60995).

Valid bottle numbers with the tray installed are 01-00 (where 00 isinterpreted as 100). Valid bottle numbers when the tray is NOT installedare 01-03. Once a valid bottle number is sent to the automatic sampler, thesampler will begin its pre-run sequence. When an injection occurs, a startsignal is sent from the automatic sampler to the SlECM which starts theAID and the gas chromatograph.

22

Operation

Page 189: HP 7673B - S service manual

~

f"r~'

, Before You Startr-­~-

1

~..

~~c,.

~".

2

This link. manual describes how to install and configure the printed-circuitboard associated with the option you ordered into either an HP 18594A orHP 18594B Controller Module.

Before you start. do the following:

• Turn off the power of the controller and disconnect the power cord.

• Check the warranty identification label on the backof the controller todetermine whether you havean A module (i.e., Modell HP 18594A) ora B module CLe., Model# HP 18594B).

WARNING

HAZARDOUS VOLTAGES ARE PRESENT IN THIS INSTRUMENTWHEN THE POWER CORD IS CONNECTED. AVOID ANYDANGEROUS SHOCK HAZARD BY DISCONNECTING THE CORDBEFORE REMOVING THE CONTROLLER UD.

I CAUTION IThis procedure requires protection against electrostaticdischarge. Use a static control wrist strap (I.e., part number9300-0969 for large wrist strap or part number 9300-0970 forsmall) connected to a ground. If you do not use static

.protection, you may damage the electronics of the board orcontroller.

After you have completed the installation, place this link manual in thebinder ofyour operating manual for the HP 7673 Automatic Sampler(part number 07673-60995) after the tab for appendix C "Link Manuals."

© Copyright 1989 by Hewlett-Packard CompanyAll Rights Reserved

r,

Page 190: HP 7673B - S service manual

~~" -~

.1~i

~

1:i

>"~.r;,

....~

~l.~:.•.~.·.·'~

..........•~..~..•~'~'iJ

•~'i~;a,r,-.~.·.·;-.·.·••.··..•..~

Link ..Manual

HP 7673 Automatic SamplerInstalling Option Boards

HEWLETT-PACKARD

i1')

t

fI

PrInted In U.S.A.November 1989

~HEWLETT

~PACKARD

Manual Part No.18599-90155

Page 191: HP 7673B - S service manual

r I

1/4-.1. rM ctiver

00. 2 point Pozktiv saewttiver

2 pa1 head saews (0515-0915)( Drawn flj scale )

iietwrench

net wrench

2 aqrnerttools (18597-40030)

2 seff-tappilg saews (0624-0546)( [)(awn fuI scaie )

3 standoffs (18597-20520)( Drawn flj scale )

2 qector bracket bust*lgs (18597-20735)( Drawn fuIscale )

1 i1jectDf bfacket (18597-00205)

o 0o

~

o

g~t........ ZZZ1ZZJ-.

2 Detach the Inlet Nut

5 AigrIthe Bracket

4 Attact1 the Inlet Nut

3 Place the Bracket

1 Attach the Bracket~

(

i.

t r~~

'\

D

~_o.I "'~ ,

1",,-,I

II

II

II

II

II

I

r--toded ciqamfl uo·· I

~i

~Jr. -.

~~

~--­

~

~~

~/;.

~~r'k~..~

ff-:. ~

~.~

,

~'tr:I

-,

..~

~~

~

r

Page 192: HP 7673B - S service manual

~,l,O"",:,,1

\.

....

Page 193: HP 7673B - S service manual

I~.

Figure 8. Align the bracket and tighten the screws.

Turn to page 16 and check your work.

11

a. Place the plastic alignment toolfs), honeycombed side up, over theinletfs) ofyour GC. Move the bracket until both alignment tools fitflush on the bracket. Ifyou cannot make them fit flush by moving thebracket, you may have to adjust the position of the inlets.

b. Tighten the two screws that hold down the bracket with a no. 2 pointPozidriv screwdriver. Remove and save the bracket alignment tools forfuture adjustments.

6. Check Your Work

5. Align the Bracket(Read only if the GC has a split/splltless Inlet.If not, skip to step 5.)

~Figure'7. Route the z:as lines over the bracket and reattach the inlet

nut.

The gas lines must not interfere with the mounting of the injector. Ifnecessary, gently bend the gas lines.

b. Push the inlet nut down and turn it clockwise. Tighten it with the inletwrench.

a. Route the gas lines on the inlet nut over the mounting bracket. Formost configurations, the gas lines should be routed over the front orriht-rear corner of the bracket.

10

~.•

I:~.•...:;.;r'~: .

t 4. Attach the Inlet Nut

L.·.."Ixf<

~,~~-

Page 194: HP 7673B - S service manual

4Installing on HP 5880A GCs

This section reviews the steps for installing the upgrade injector bracketand shows an exploded View. For detailed instructions, refer to the link.manual "Installing and Operating the Injector and Tray on HP 5880A GCs"(part number 07673-90235). Some of the parts are packaged with theupgrade bracket and the rest with the HP 5880A injector/tray mountinghardware (part number 18597-6(675).

The exploded view on the next page shows the bracket being installed overpositions B and D.

Complete the following steps to install the upgrade injector bracket.

1. Remove the existing injector bracket and bracket supports.

2. Replace the pan head screws that hold the inlets to the GC top coverwith the appropriate standoffs (small hexagonal posts}.

• For a packed inlet, use the small standoffs.

• For a capillary inlet, use the large standoffs.

• For a position without an inlet, attach the piece ofsheet metal andtwo large standoffs.

3. Attach two large standoffs to the bottom of the bracket.

4. Place the bracket.

5. Align the bracket.

6. The next step is to install the tray bracket.

12

3. Place the Bracket

a Hold the bracket with the bend on the left side and slide it onto thecylinders that yon attached to the side wall in step 1. Allow the bracketto rest on top of the standoffs.

Figure 5. Placing the mounting bracket on the GC.

b. Line up the top of the standoffs with the hole and slot on the bracket.Screwthe pan head screws into the threaded holes of the standoff's. Donot tighten the screws until step 5.

Figure 6. Start the pan head screws into the standoffs.

9

.~

i',~

j~~

~i\:l.~~--~~

.

''''".,'........•"\~

...:.r.

,---1

,;~

~

1.~'"':~.~

~~

. ,;'iJ;j

.1'~

- .;,-

Page 195: HP 7673B - S service manual

~

t·:

~

,.~.e

~~..

~

~•.•;"!

""­'.

2. Detach the Inlet Nut (Read only if the GC has a spliUsplitless inlet.If not, sklp to step 3.)

For a split/splitless inlet, route the gas lines over the bracket. H you donot, you cannot replace the inlet liner without removing the bracket.

I CAUTION ITum off the gas flows before you remove the inlet riut or youmay damage the packing in the Inlet nner. Avoid contaminatingthe inlet while it Is uncovered.

a, Loosen the inlet nut with the inlet wrench. Turn the nutcounterclockwise until it is off the threads and lift it out of the way.The braze joints that hold the gas lines to the inlet are fragile.

Figure 4. Detach the inlet nut and lift it out or the way.

8

,.

I

.~e

Figure 9. Installing the upgrade bracket on an HP 5880A GC.

13

Page 196: HP 7673B - S service manual

I

4Installing on HP 5710/30/90 GCs

This section reviews the steps for installing the upgrade injector bracketand shows an exploded view. For detailed instructions, refer to the linkmanual "In.sta.lling and Connecting the Injector and Tray on HP 5710/30190GCs" (part number 07673-9(225).

You can mount only one injector on an HP 5710130190. To mount theinjector over the front inlet (i.e., in the A position), install the bracket overthe rear set of threaded holes. To mount the injector over the rear inlet(i,e., in the B position), install the bracket over the front set of threadedholes.

Complete the following steps to install the upgrade injector bracket.

1. Remove the existing bracket and replace the GC top cover.

2. Attach the plugs and septum nut.

3. Attach the appropriate standoff's (small hexagonal posts).

• For a packed inlet, use the medium-sized standoffs to support thetray, but do not use any standoffs to support the injector bracket.The injector can only be used with a packed port in the B position.

• For a capillary inlet, use the largest standoffs to support the trayand the smallest standoffs to support the injector bracket.

4. Attach the injector bracket.

5. Align the bracket.

6. The next step is to install the tray bracket.

14

j-

e. Screwone standoff onto the bottom of the mounting bracket to act as asupport leg. Thehole has a silver-colored pem (small threaded washer)on the top, right-hand side of the bracket. Tighten the standoff with a'114-in. nut driver.

oFigure 3. Attach the standort to the bottom of the mounting bracket.

Ifa purged packed inlet is installed on the GC, this standoff couldinterfere with the heat sink (metal fins) around the inlet. Ifnecessary, useonly the left-hand side of the heat sink.

IfHP 5800 SERIES n cool on-column inlets are installed in both thefront and rear positions of the GC, do not install this standoff. WIthout thisstandoff, the stability of an injector in the rear position may be affected.

7

'~

£;:.~

'\~','J

1'~

.,'1

,~" ".'."'.,~:.',',,'~1~~.". _~

-~.::i

. .'~

"

;~

~~,j

j»;

~,',;,~,;'~

:t,il.,~

;

~~

',,':,~,'oil--~:1

Page 197: HP 7673B - S service manual

~~

~

if-"'

~7.~'-.

t,

ft.'.'•..·.'...• ·f~.."t..'.'...·.~~- " ..k

<

?f

I~.,

~,f,~1:,.,.

6

d. Screw two of the standoffs (small hexagonal posts) into the threadedholes located near the inlets. Tighten them with a 1I4-in. nut driver.

oII

I

Figure 2. Attach the standoffs near the inlets.

If the rear inlet is an HP 5890 SERIESIT cool on-column inlet (Accessory/

Kit 19245B or 19245C) on an HP 5890A GC, use a short standoff in the rearposition (part number 19245-20760). This standoff is packaged withtheaccessory kit.

{i·

Ali011IneIlI tool

Figure 10. Installing the upgrade bracket on an HP 5710/30/90 GC.

15

Page 198: HP 7673B - S service manual

J .~

-t

:~

,':'-:

,.~

~

~~--~

~

]j~-..;.

t~.~

~jh

-~f~"o-r.

~

Fipre 1. Attach the cylinders to the side of the GC.

5

If installing on an lIP 5890A GC, you must first install the HP 5890SERIES II split top cover .. You.must also remove the existing injectorbracket and supports.

a Remove the port pan of the split top cover by inserting a finger into thaU-shaped cut on the panel and pulling up.

b. Lift. up the right-hand side of the top cover.

e. Using a 1/4-in. nut driver and the two self-tapping screws. attach thetwo small cylinders onto the left side wall of the chromatograph.

1. Attach the Injector Bracket Supports

i

The next step is to install the tray bracket. Refer to "Installing the TrayBracket on 5890 GCs" (part number 07673-90205).

3

Double-check the following:

o If the GC hasa splitJsplitless inlet, check the inlet nut.

o If the GC has an on-column inlet, be sure you havethe correct inlettop-either the cooling tower or the septum nut base assembly with theneedle guide top.

o Ifgas lines are routed over the bracket, be sure the lines are placedaccording to figure 7 on page 10.

o Check that the following are tight:

o Self-tapping screws.

o Pan head screws.

16

Checking Your Work

Page 199: HP 7673B - S service manual

rj,;:_­

s- 2

--

--

.._-J......-=-s-.­_-I2_-__

'or=::'"_---2 ~~

_-'0)1_-~"-

Exploded view with aI pais

5 .....-

0

3 _ .. - e-,

"'_..-

Step by step

Pa1syou need fofeachstep

Tools you need foreach step

~--:--------------t-I 1 _ .._-

--

Pull out the yellow worksheet from the center of this link manual. Organizethe hardware and tools for each step. Full-scale drawings of the screws helpidentify which ones you need for each step. Ifany of the hardware ismissing, contact your HP representative.

~-

i;'

f

~'Installing on HP 5890 GCs~

~.~•.~~---:

~.~.rI-f

tt'-e

~.,

t

t~.

~.

4

Page 200: HP 7673B - S service manual

3. Install the upgrade injector bracket. Instructions are in this linkmanual.

• For an HP 5890A or SERIES IT Get turn to page 4.

• For an HP 5880A Ge, turn to page 12.

• For an HP 5710/30190GC, turn to page 14.

4. Install the tray bracket. Instructions are packaged with the traybracket.

I

3

<'!l---?

~-~

"1.",£

..

-•..'•...,7.•·.'1.'1'..'··'.i-...:-~:~

-~- -y

--t~:.';)

'I., ..~

~·1

-;·'1"1

.~ ::;

..;'~~';

L-t

Page 201: HP 7673B - S service manual

~

"

1,.

..

WARNING

ALLOW THE INLETS AND OVEN TO COOL BEFOREINSTALUNG THE UPGRADE INJECTOR BRACKET. THE METALSURFACES OF THE GC ARE VERY HOT AND COULD BURNYOUR SKIN. IF YOU HAVE A SPUT/SPUTlESS lNLET, YOUCOULD DAMAGE THE COLUMN.

In order to install the tray, you must:

1. Remove the existing injector bracket.

2. Install' any necessary upgrades to your gas chromatograph (GC).

• Before installing the upgrade injector bracket on an HP 5890SERIES n GC with a serial number less than 2938Axxxx, you mustfirst install the new left side panel (part number 18597-60705).

• Before installing the upgrade injector bracket' on an HP 5890A, youmust first install the split top cover (part number 07673-61065) andthe left side panel (part number 18597-607(5).

• Before installing the upgrade injector bracket on an HP 5710/30190GC, you must replace the GC's top cover with the new top cover.

The upgrade injector bracket enables you to install an HP 1859GB TrayModule to deliver sample bottles to one or two HP 18593A InjectorModules .

~", ..•,..•.•.... ,

.',~r

~

l,.••.I.f

rli,

~.' Before You Start~Y;

-;'~-

'f~

e Copyright 1989 by Hewtett-Packard CompanyAll Rights Reserved

2

Page 202: HP 7673B - S service manual

;~-~j

Manual Part No.18597-90150

~HEWLETT

~PACKARO

PrInted In U.S.A.December 1989

1r

HEWLETT- PACKARD

( :II Oa'

{=a= &

LinkManual

HP 7673 Automatic SamplerInstaiUng the Upgrade Injector Bracket

.<f

~,t1,~';,- ,

"1'j.:~1J.~._--~~

-;";,.:;:;f

Ac~

":F:

~

--~

";;<!~

]--1

. --~

~-l

.L~--~~I:'T:ll

. -:~

Page 203: HP 7673B - S service manual

i~.

1/4-in. rut ciiver

4 stalcIoffs (18597-20660)( drawn fuI scale)

2standoffs (18597-20650)( li'awnf1iJ scae ~

2 Strldoffs (18597-20640) with packed ir*t( Drawn 0/1 scale)

2 plugs (6960-0085)( chwn fuI scale )

1 seph.m rut (18597-20670)( Drawn fIJI scae )

1 top cover (18597-00185)

No.2paR Pozidriv

5 PMhead screws (2510-0099)( chwn fuI scale )

No.2poi1t Pozidriv

4 pal headscrews (2510-0099)( drawn fiJI scale)

1 injector bracket (18957-0(175)

g.0'0o

00

o

Io

C': :: h ".::::.::::::::'}_

tos. tl, % tl, II " \, os t$ <I \~

4 Attach the Injector Bracket'

2 Attach the PkJgs and 5eptlm

3 Attach the Bracket SJ.worts

1 R~e the GC Top Gover

..-

ExpIoded~

If

e.

tl

g.- -

~.-~ . I

0

~' t• I

v,

I-_~» .I

00

~"

~ ~1:'

00 DOo~

I 1\ I• f

1000

"

00

I: , I

0000

0 0

i~.

w, I I

Page 204: HP 7673B - S service manual

;.-:j

f ~

5 Aign the Bracket

6 Attach theMourJ!i1g Posts

NO.2 point Pozictiv

2 arvrnent tools (18597-40030).

Flat head saewaiYef

2 mountilg posts (07673-61035)

,~

i:~

j.,]

YJ'.<.'~.:~:i

..~

~

f~~.-,---'"

:~._~

7 Attach the Tray StMdoffs

~~'7~?7~7Zl',Jmo3I8-il. Crescent wrench

3 standoffs (18597-20710)(drawn fljj sc.) Tlisworksheet isfor instaIi1g the rnot..rFfilg bfackets forthe HP 185938 Ir1ector

Mo<lie cr1d the HP 185968 Tray Me>roe toan HP 5710/3CV90 Gas Chromatogaph.

~

"I"i~

. .:h.~-_..~

NO.2 poilt Pozkiiv

1traybracket (18597-60515)

o •

I0 a- 3 pan head screws (0515-0915)( dIawn fij sca)e )

• .- 00 Og.

QIII 1PCfl headscrew (2510-0099)( drawn fIJI scale )

~ 1 flat headscrew (251 0-0121)( <hwn fUlscale)

8 Attach the Tray Bracket

Page 205: HP 7673B - S service manual

·.;.

~it.~r 5. Align the Bracket 6. Attach the Mounting Posts

e

a. Screw each mounting post clockwise into the threaded holes of thebracket. Tighten the posts with a flathead screwdriver.

Ifyou are not installing the tray bracket, turn to page 12.

Ifyou are installing the tray bracket, continue on the next page.

e

~I

a. Place the plastic alignment tool, honeycombed side up, over the inlet ofyour GC. Move the bracket until the alignment tool fits flush on thebracket. Ifyou cannot make it fit flush by moving the bracket, yon mayhave to adjust the position of the inlet.

b. Tighten the four screws that hold down the bracket with a no. 2 pointPozidriv. Remove and save the bracket alignment tool for futureadjustments.

~:

[,t

f~.~t:•~"

1,-

f.•·.·I.l':

~:~~

t,"'-',

8 9

Page 206: HP 7673B - S service manual

'~"':"~;

·:.·.1'.'"1

.~.~

:]

4...J..-.:1

..

1::~

~

;-1

·cJ•....~~,

,.~

'I

·1,~~j~

.~~

7

1e I

! ~e

8~/.

"A

6)11

dll

a. Hold the bracket with the bend on the left side. Allow the bracket torest on top of the threaded washers or standoffs.

b. Line up the top of the threaded washers or standoffs with the holes onthe bracket. Screwthe pan head screws (English thread) into thethreaded holes of the standoffs.

4. Attach the Injector Bracket

~-

e

~ ~, II 'Io

II

I

a. Screw the three standoffs (hexagonal posts) into the injector mountingbracket. Use the holes identified in the illustration. Ifyou do not usethe correct holes, the tray will not be able to deliver sample vials to theinjector.

b. Tighten the standoffs with a 318-in. crescent wrench.

10

7. Attach the Tray Standoffs

Page 207: HP 7673B - S service manual

b. Using a no. 2 point Pozidriv, screw in the three pan head screws (metricthread) through the holes labeled A and into the standoffs..

c. Line up the hole on the far side of the tray bracket Oabe1ed 5700) withthe threaded hole of the rear standoff'. Screw one pan head screw(English thread) into the mounting support with a no. 2 point Pozidriv.

d. Line up the hole on the near side of the tray bracket Oabeled 57(0) withthe threaded hole of the front standoff'. Screw one flat head screw(English thread) into the mounting support with a no. 2 point Pozidriv.

11

a. Place the tray bracket on the five standoffs installed in steps 3 and 7.

If the top of the standoffs do not line up with the holes labeled with theletter A, check the location of the standoffs against the figure onpage 10.

8. Attach the Tray Bracket

/

Tlnadd wathen tor:

a. Attach the necessary standoffs to the threaded washers on the top coverthat you installed in step l.

b. Tighten the standoffs with a V4-in. nut driver.

Save any standoffs you do not use for future adjustments or for differentconfigurations.

6

3. Attach the Bracket Supports

The injector and tray brackets are supported by standoffs (small hexagonalposts) that screw into threaded washersattached to the top cover. Whichthreaded washers you usedepends on the type and position of the inlet.

• For a packed inlet use the two medium-sized standoffs (part number18597-20640) to support the tray, but do not use any standoffs tosupport the injector bracket. The injector can be used with a packedport in the B position, but not with one in the A position.

• For a capillary inlet use the two largest standoffs (part number18597-20650) to support the tray and the two smallest standoffs (partnumber 18597-20660) to support the injector bracket.

• For an inlet in the A position attach any standoffs to the rear set of .threaded washers.

• For an inlet in the B position attach any standoffs to the front set ofthreaded washers.

~~.

~~

~Ii'

i

~••..

~,~,- -

~I'....1',?

{~r

I~.

~.

-c,

;;.:

~

Page 208: HP 7673B - S service manual

.,.:0:.1

'-,'

3Checking Your Work

Double-check the following:

D The new septum nut is properly installed.

D The three screws that attach the tray bracket to the standoffs on theinjector bracket must go through the holes labeled A on the traybracket.

2. Attach the Plugs and Septum Nut

With the HP 5710/30190 GC, the automatic sampler is limited to deliveringsample to one inlet and cannot be usedwith an on-column inlet.

a. There are two holes on the right-hand side ofthe top cover for anon-column capillary inlet. Cover one or both holes with the plasticplugs.

b. Replace the existing septum nut with the one supplied in this kit.

~ ~-~

,

1~,i,~

",'

D The flat head screw for attaching the tray bracket is in the correctposition. See page 11. e

e

."~'-);

~

l,,~

"~

5

e

PllIgs

e

8eptumnut

"

For detailed information on mounting and connecting the injector and tray,see chapter 1, "Setting Up", in your operating manual

D All screws and standoffs are tight.

After you have completed the installation, place this link manual in thebinder of your operating manual for the HP 7673 Automatic Sampler(part number 07673-60995) after the tab for appendix C "Link Manuals."

12

Page 209: HP 7673B - S service manual

v..~-r'if:~. 1. Replace the GC Top Cover 4,u,~~.

~.~

ft.Fe~:f

[~~,

~

To replace the top cover, you may have to remove hardware that is attachedto the existing top cover (e.g., the inlet control module).

a. Using a no. 2 point Pozidriv, remove the 5 screws that hold on the topcover of the GC.

b. Lift off the existing top cover.

c. Position the new top cover.

d. Using a no. 2 point Pozidriv, attach the top cover with 5 screws (Englishthread).

.'

Connecting the Cables

The type of data-handling device and communications configuration thatyou use determines how you connect the automatic sampler to theHP 5710130190 GCs. This section contains diagrams for controlling withthe instrument network (INET) or with stand-alone control.

Instrument Network (INET)

t\'.r,~.;

~~

tr~0:t.

t

E:

~.4

I

I

I

II

II

I

~

Connect the !NET cables between the !NET board in the controller and theintegrator. Connect the remote startlstop cable between the integrator andthe GC.

~

03392 ·60800 •

• Fir HP 3396 Iltegrator USI ,art BUlUlIf 03396-60629

13

Page 210: HP 7673B - S service manual

~

::';;-:'

.~

~-1':z

~"1...~.'.~

_ ..~

.~

. '~

.s-:.,.ii~?~

B;~

.4

'·.~.I.".iI

' ..\~-ci

#'it-'-."-

""~~4

....I'~

I IIIIII-=-t-~[ )

2=::)..............

1,,0IW~

......­( .........)

-­;---(--III..)

J ......~<---..."

.. 2.........................,........t ...... -.Iio)

._-~-

e-00-c:::r.

1:::)1

= ..

~n...~ ....

Exploded view with aI pats

3 _OR__

,- ..--

step bystep

Parts you need foreach step

2

Tools you needforeach step

--

Your special hardware kit contains all of the hardware you need to installthe new top cover, the injector bracket, and the tray bracket for theautomatic sampler, except for the brackets. Find the injector bracket in thecontroller box and the tray bracket in the tray box.

Pull out the yellow worksheet from the center of this link manual. Organizethe hardware and tools for each step. Full-scale drawings of the screws helpidentify which ones you need for each step. If any of the hardware ismissing, contact your HP representative.

Installing the Brackets

1894-50&00

RMT

~

HP5700

RMT

18594-80530

~

Connect the remote startJstop cable between a chart recorder or integratorand the GC with a"T cable.

WIthout a chart recorder, connect a cable between the controller and theGC.

Stand-alone Control with a Remote Start/Stop Signal

14 3'-I'

Page 211: HP 7673B - S service manual

I:J

&F"'

!1 With an AID converter, connect the three instruments with the appropriate

cable or cables. The following illustrations show two example.

~

35900-M620

1a594 -60820

35900-60800...,.

.14AEIIlOTE

START

RIfT

J4ROIOT£

START

\

r

..

ALLOW THE INLETS AND OVEN TO COOL BEFORE INSTALUNGTHE NEW TOP COVER AND INJECTOR MOUNTING BRACKET.THE METAL SURFACES OF THE GC ARE·VERY HOT AND COULDBURN YOUR SKIN.

WARNING

• Instructions for installing the mounting brackets for theHP 18593B Injector Module and HPl8596B Tray Module on anlIP 5710/30190 GC.

You can mount only one injector on an HP 5710/30190. To mount theinjector over the front inlet fi.e., in the A position), install the bracket overthe rear set of threaded holes. To mount the injector over the rear inlet(i.e., in the B position), install the bracket over the front set of threadedholes.

To change the position of the injector, you must change the position of themounting bracket,

• Sample diagrams for connecting the cables for the automatic sampler toan HP 5710/30190 GC.

• Instructions for installing the new top cover on an HP 5710/30/90 GasChromatograph (GC).

This manual contains the following information:

~_:

~',

E

f':<

" .,~~

t~

~.

~f

fr'"r.t1!'7~"\-

"

,~;~

~\Before You Startfi.-.c-.•.·.'!J,

1\

..© Copyright 1989 by Hewlett-Packard Company

All Rights Reserved

2 15

Page 212: HP 7673B - S service manual

Manual Part No.07673-90225

~HEWLETT

~~PACKARO

Printed In U.S.A.December 1989

I.i

I•t•,

HEWLETT- PACKARD

( ;]I 0.'

(~ ..Link

Manual

HP 7673 Automatic SamplerInstalHng and Connecting the Injector

and Tray on HP 5710/30/90 GCs

c~

~~l,:1-..•....'.~•....•........•....i..:•...•..~~'.

~-.-:.~-

0-

'~

;\~~.

--:;'-~

.1c~

~1

Page 213: HP 7673B - S service manual

1/4-it fllt diYer1 Attach the Bracket~ExpkxIed dagrrfn

i&'

~ I

2 qector bracket busmgs (18597-20735)( Drawn f1j scaie)

2 seI-twilg saews (0624-0546)( Inwn f1j seale )

3 saIdoffs (18597-20520)( DfIwn U scale )

1 i1)eckr tncket (18597-00205)

00o

g~tzzzzzzzzl-

o 0 L-..Jo 000 CJ<?'~~n 0

6 Attach t1e Mou1i1g Post Iattleadscrewdiver

f jnfu1li1g post(07673-20865)

( Inwn f1j seale )

t·;..

,:

"!

~J

0

I n; i

! 2 ~b~~D

i1etwnn:h

..3 Place tie Bracket

~ 2 pal head screws (O515-0915)( Drawn f1j scaie)

4 Attach the IlietNut net 'M'eI'dI-5 AIr;J1 the Bracket

~No. 2 Poirt Pazkiiv

2 aig IlllllU toots (1 8597-400(0)

~,8 •I ~~ 'I .A"~

1""I '

I•

II

II

I,II

I

fio·

r"-

~

.,~

~ .

E

~-:.::-

l-'-_t,;.";,

Page 214: HP 7673B - S service manual

-'

Page 215: HP 7673B - S service manual

,c

ti~:~..•.~

,

•. -.c...•.• '.~.•.•.••.•.•..

'i _

>'

,

~. -

t~:

=-~8Septlrn rJ.t ~base assen1lIy~~-~

a. Screwthe mounting post clockwise into the threaded hole of thebracket. Tighten the post with a flathead screwdriver.

To mount over an lIP 5890 SERIES n cool on-eolumn inlet, removethe needle guide top (part number 19245-20670). Save the needle guide topto realign the bracket.

To mount over an HP 5890A on-eolumn inlet, replace the cooling tower(part number 19245-20540) with a septum nut base assembly (part number19245-8(520), Save the cooling tower to realign the bracket.

The B module needs the following:

6. Attach the Mounting Post

The A module needs the cooling tower or needle guide top to be inplace.

Fi!UU 8. Align the bracket and tighten the screws.

The A and B injector modules need different parts when mounting over acoolon-column inlet.

a. Place the plastic alignment toolfs), honeycombed side up, over the in­letfs) ofyour~. Move the bracket until both alignment tools fit flushon thebrackeL Ifyou cannot make them fit flush by moving thebracket, you may have to adjust the position of the inlets.

b. Tighten the two screws that hold down the bracket with a No.2 pointPozidriv. Remove and save the bracket alignment tools for futureadjustments.

5. Align the Bracket

~.

{

~.

,~

~~':>c.r~-

}-

~

:;:.-

Figure 9. Attach the mounting post.

10 11}-~:

Page 216: HP 7673B - S service manual

3

Installing on HP 5880A GCs

This section reviews the steps for installing the injector bracket and showsan exploded view. For detailed instructions, refer to the link. manual"Installing and Operating the Injector and Tray on HP 5880A GCs" (partnumber 07673-90235). The parts are packaged with the HP 5880Ainjector/tray mounting hardware (part number 18597-60675)

You can mount two injectors on an HP 5880A GC in positions A and C or Band D. To mount the injector over positions A and C, align the bracket overthese positions. The exploded view on the next page shows the bracketbeing installed over positions B and D.

Complete the following steps to install the hybrid injector bracket.

1. Remove the existing injector bracket and supports.

2. Attach the appropriate standoffs (small hexagonal posts) next to theinlets.

• For a packed inlet, replace the pan head screws with the smallstandoffs.

• For a capillary inlet, replace the pan head screws with the largestandoffs.

• For a position without an inlet, attach the piece of sheet metal andtwo large standoffs.

3. Place the bracket.

4. Align the bracket.

5. Attach the mounting post.

12

~

4. Attach the Inlet Nut (Read only if the GC has 8 split/splitless Inlet.If not, skip to step 5.)

a. Route the gas lines on the inlet nut over the mounting bracket. Formost configurations, the gas lines should be routed over the front orrear of the bracket.

The gas lines must not interfere with the feet of the injector. Smallwhite prcles on the bracket showwhere the injector feet touch thebracket. Ifnecessary, gently bend the gas lines so they do not cross thewhite circles.

b. Push the inlet nut down and turn it clockwise. Tighten it with the inletwrench.

~Figure 7. Route the gas lines over the bracket and reattach the inlet

nut.

9

.yoc...,:.•

.~i..:,---';.-j

.:--::-;

~cf:

J~'.:",.'""~

~~.0J"F.·•..",·.·.1.:..f'._'"-__--o-:f-.".

c~~-

j":· ..-1_..~.

--~

,I,;.::.,;

-.·...-1.....~

~-l~--,'-~

Page 217: HP 7673B - S service manual

I·~, c.,.,.,. " ,,;C,.

,"... :,

~L

~. 3. Place the Bracket

r~:.

~

:::

~•.

~"

~.l -~.

It'f<

~.•.•.. ,.'.f~',.:f~:

[

&~

~,

t~,F,E

~8

r

a. Hold the bracket with the bend on the left side and slide it onto the cyl­inders that you attached to the side wall in step 1. Allow the bracket torest on top of the standoff's.

Figure 5. Placing the mountlng bracket 00 the GC.

b. Line up the top of the standoff's with the hole and slot on the bracket.Screw the pan head screws into the threaded holes of the standoff's. Donot tighten the screws until step 5.

Figure 6. Start the pan head screws Into the standoffs.

~~~

Figure 10. Installing the hybrid bracket 00 an HP 5880A GC.

13

Page 218: HP 7673B - S service manual

4Installing on HP 5710/30/90 GCs

This section reviews the steps for installing the injector bracket and showsan exploded view. For detailed instructions, refer to the link manual"Installing and Connecting the Injector and Tray on lIP 5710130190 GCs"(part number 07678-90225).

You can mount only one injector on an lIP 5710130190. To mount theinjector over the front inlet a.e., in the A position), install the bracket overthe rear set of threaded holes. To mount the injector over the rear inlet(i.e., in the B position), install the bracket over the front set of threadedholes.

Complete the following steps.to install the hybrid injector bracket.

1. Remove the existing injector bracket.

2. Replace the GC top cover and attach the plugs and septum nut.

3. Attach the appropriate standoffs (small hexagonal posts).

• For a packed~, use the medium-sized standoffs to support thetray, but do not use any standoffs to support the injector bracket.The i.nject.or can only be used with a packed-port in the B position.

• For a eapillaITinIet, use the largest standoffs to support the trayand the smallest standoffs to support the injector bracket.

4. Attach the injector bracket.

5. Align the bracket.

6. Attach the mounting post. There is only one mounting post.

14

2. Detach the Inlet Nut (Read only if the GC has a spllt/splitless Inlet.If not, skip to step 3.)

For a split/splitless inlet, route the gas lines over the bracket. lfyou donot, you cannot replace the inlet liner without removing the bracket.

I. CAUTION ITum off the gas flows before you remove the Inlet nut or youmay damage the packing In the inlet liner. Avoid contaminatingthe Inlet while it Is uncovered.

a Loosen the inlet nut with the inlet wrench. Turn the nut counterclock­wise until it is off the threads and lift it out of the 'Nay. The braze jointsthat hold the gas lines to the inlet are fragile.

Figure 4. Detach the inlet nut and lift it out of the way.

7

.~

:~

.....~>c.·.'1"~·lH'.~

'~J

1:'1

'~i,d

:~?

'~.~~

1!'--~

,~

:1ti~

.•...••••c••.~••..•.•,I,·.::"~I

1.!'~~~

.~

Page 219: HP 7673B - S service manual

,..."..

}f

AiIIllMlJt tool

MeDDtiII post

Figure 11. Installing the hybrid bracket on an HP 5710/30/90 GC.

Figure 3. Attach the standoff to the bottom or the mounting bracket.

oIfa purged packed inlet is installed 00 the GC, this standoff couldinterfere with the heat sink (metal 'fins) around the inlet. Ifnecessary, useonly the left-hand side of the heat sink.

If lIP 5890 SERIES n cool on-column inlets are installed in both thefront and rear positions of the GC, do not install this standoff. Without thisstandoff, the stability of an injector in the rear position may be affected.

e. Screwone standoff onto the bottom of the mounting bracket to act as asupport leg. The hole has a silver-eolored pem (small threaded washer)on the top, right-hand side of the bracket. Tighten the standoffwith a1/4-in. nut driver.

f~~,

~.

~-

~.ii;,itt

~

••

~,.

~~~

1":

~

~

6 15

Page 220: HP 7673B - S service manual

5Connecting the Adapter Cable

The following diagram illustrates how to attach the adapter cable requiredto connnect an HP 18593B Injector Module to an HP 18594A ControllerModule.

AcontroIItr

AIIaptw ea'il

8 IniedIr cabI.

E JFigure 12. Connecting the adaptor cable.

16

d. Screw two of the standoffs (small hexagonal posts) into the threadedholes located near the inlets. Tighten them with a 1/4-in. nut driver.

UII

I

Figure 2. Attach the standoffs near the inlets.

If the rear inlet is an HP 5890 SERIES II cool on-column inlet (AccessoryKit 19245B or 19245C) on an lIP 5890A GC, use a short standoff in the rearposition (part number 19245-20760). This standoff is packaged with theaccessory kit.

5

'-~'-::-~

',~~

-~~

W~

.~

1-t~-~

2~;,.t~:

- :;:;

...~

:~

i.;~

I

Page 221: HP 7673B - S service manual

~~

~~.

t 1. Attach the Injector Bracket Supports 6

-iI

~x,

f",r~'

~eic~'

r-lf'

I;

~.

r

r.. ·~.•.·.·...r

4

If installing on an HP 5890AGC, you must first install the necessaryupgrades listed on page 2.

a Remove the port pan by inserting a finger into the U-shaped cut on thepanel and pulling up.

b. Lift up the right-hand side of the top cover.

c. Using a 1/4-in. nut driver and the two self-tapping screws, attach thetwo small cylinders onto the left side waD of the chromatograph.

Figure 1. Attach the cylinders to the side of the GC.

~

Checking Your Work

Double-checkthe followingt

o If the GC bas a splitJsplitless inlet, check the inlet nut.

o If the GC bas an on-column inlet, be sure that you have the correct inlettop: either the cooling tower or the septwn nut base assembly, with orwithout the needle guide top.

o Check. that the following are tight:

o Self-tapping screws.

o Pan head screws.

o Injector mounting posts.

Ifyou are installing a tray, install the tray bracket before mounting theinjector. Instructions for installing the tray bracket are packaged with thetray.

17

Page 222: HP 7673B - S service manual

2Installing on HP 5890 GCs

Pull out the yellow worksheet from the center of this link manual. Organizethe hardware and tools for each step. Full-scale drawings of the screws helpidentify which ones you need for each step. Ifany of the hardware ismissing, contact your HP representative.

;!I.~.'.:.::.':~.',-":"5';-

'.'l'r

Expoded view wirh aI parts

e _ .._.. __

F l--- ­(... • ....1

. ~.~

-~~J

·1··•.··:·.':.·.·.·.:·'.·-.,

.._-._--6 ....._

A~

---r-'-a --Step bystep ----~1IlI_)

Parts you need foreach step 2 ......~~t;"'IIIal)

J .....~..,--Tools you need foreach step

b"~1 ,------

I....~ I 3 _ .. _ QID2 ......--.:~,-..0-.. _..- -

"

3

Page 223: HP 7673B - S service manual

poL~?~

¥. 1~

~.. Before You Starti--x

"--------------I~:>,.

~I's-

i:,l -"~_:

r.~-

trk

I'~,

r~­

kr·,,

2

The hybrid injector bracket enables you to mount an HP 18593A InjectorModule over the front inlet and/or an HP 18593B Injector Module over therear inlet of your gas chromatograph.

To install the hybrid bracket, remove the existing bracket and bracketsupports.

• For an lIP 5890A or SERIES II Ge, you can mount both injectors atthe same time. Turn to page 3.

Before installing the hybrid injector bracket on an HP 589OA, you mustfirst install the split top cover for the HP 5890 SERIES n GC.

Hyour automatic sampler includes a tray. you must install the new traybracket onto the existing tray bracket supports (often called side rails).Directions for installation are packaged with the new tray bracket.

• For an HP 5880A Ge. you can mount both injectors over inlets in posi­tions A and C or B and D. Turn to page 12.

• For an HP 5710/30/90 Ge. you can mount only one injector at a time.Turn to page 14.

• Toattach the adapter cable required to connnect an HP 18593B In­jector Module to an HP 18594A Controller Module, turn to page 16.

WARNING

ALLOW THE INLETS AND OVEN TO COOL BEFOREINSTALLING THE HYBRID INJECTOR MOUNTING BRACKET.THE METAL SURFACES OF THE GC ARE VERY HOT ANDCOULD BURN YOUR SKIN. IF YOU HAVE A SPUT/SPUTLESSINLET. YOU COULD DAMAGE THE COLUMN.

e Copyright 1989 by Hewlett-Packard CompanyAll Rights Reserved

~

~

Page 224: HP 7673B - S service manual

ManuaJ Part No.18597·90140

fi~ HEWLETT~~ PACKARD

Printed In U.S.A.December 1989

1,I}

"..

i,J

( :II 0='{:-= ..

LinkManual

HP 7673 Automatic SamplerInstalling the Hybrid Injector Bracket

;..";"~ -'.-!>

..-,~'---:-ji

.-~7;

j--..:;~,.'

'-1

..J--1

iI;

"':..(;-~

"D,

~o.]

-a

,

oj':S..•..~.-.'.:-,

:.~;;

Page 225: HP 7673B - S service manual

'~:"-- ..

~~"'ff'"

~.

~I==- -~

I

NO.2 PoD Pozidriv

1/4-i1. rwt driver

1port pa1 (5890-40072)

No.2Point Pozidriv

1/4-it ru cliver

3 pal head saews (0515-0910)( Drawn flil scaIe)

1seff--twflg screw (0624-0665)( Drawn fuIscale )

1 font pa18I bezel (05890-40242)

1 pal head saew (0515-0904)( Drawn fuIscale )

1 flat washer (3050-1178)(rnwn fuI scale )

1 rigt side oftop cover (05890-00402)

1 id shaft (05890-2<1780)

) 1 flatwasher (3050-1178)o

~

(

-g-

( 0

1000-

oo

~ <

c-

ro

i ]

5 SOC(l on the New Port Perl

4 Attach the New Fronl Pm Bezel

.3 Attach the New Top Cover

1 Remove the 5890A Top Cover

2 Assemble the New Top Cover

I g 0

(

~i'

,.\i

'\'

r-•~-,-

II

~

Exploded clagran

lo,ttrg,,'I

r

0>--,

~.

~-­1':'

,

~.

~

Page 226: HP 7673B - S service manual

c •.

................

Page 227: HP 7673B - S service manual

IF1-

r

2. Assemble the New Top Cover3. Attach the New Cover

Figure 5. The left-band side of the binge.

/

c. Using a 1/4-in. nut driver, a flat washer, and a self-tapping screw, securethe top cover to the right-hand side of the GC. H the GC hasagrounding wire, attach it to the threaded post with the lock washer andnut that you removed in step 1.

b. Push the hole on the right-hand side of the top cover over the knob onthe right-handside of the GC.

a. Place the cupped end of the lid shaft over the knob on the left-hand sideof the GC.

Figure 4. Assembling the new top cover.

a. Using a No.2 Point Pozidriv, a flat washer, and the largest pan headscrew, attach the cylindrical lid shaft to the right-hand section of thenew top cover.

t!

""

t~ '

',.~~- ;.

{

f"I~,.P'i

~~

Figure- 6. Attaching the right-hand side of the hinge.

\.6

7

Page 228: HP 7673B - S service manual

.~

Figure 7. Attaching the front grill.

cj

.'-"-S-

j"",~-

'~~

.,:,:",,

;,j..;,~

.,~

.~

.ii~

-~--:~

-~

-~

:~~

"- ~'~

.0'1<)',1

. ..~

:....:.,.....•.--~:'4

..0'-."'-.-

Figure 2. Dismantling the hinge of the top cover.

b. Using a 1J4-in. nut driver, remove the screws and washers that make upthe hinge of the top cover. If the GC hasa grounding wire, disconnectit by removing the nut and lock washer with a 9/32-in. nut driver.

Figure 3. Dismantling the hinge of the top cover.

5

c. Remove the top cover by pushing one of the metal tabs over the knobwith your thumb.

I

a. Open the oven door and place the front panel bezel (or grill) on the topfront edge of the GC.

b. Start each of the three pan head screws from below with your fingers.Tighten each screw with a No.2 Point Pozidriv,

4. Attach the New Front Panel Bezel

8

Page 229: HP 7673B - S service manual

(-­

I•

~•..

~.

~/l-:

[~'

f~\

..·t.-

t,l

t!' ~

[

t"

~_.

~-

1. Remove the 5890A Top Cover

a Remove the injection port pan from the top cover by pushing in thesides and lifting up. The next two paragraphs describe one method.

Using the thumbs ofboth hands, push to the left and lift the left-handside oftha port pan. The left-hand side should pop out of the top cover.

Next, pull up on the right-hand side with your fingers. The right-handside should pop out of the top cover. Lift the port pan off the top cover.

Figure 1. Pop out the injection port pan.

4

5. Snap on the New Port Pan (Only If you are not Installing an Injectorat this time.)

a Close the oven door and lower the right-hand side of the top rover.

b. Place the front edge of the port pan on the front bezel. Push the backedge with the U-shaped notch down until the port pan snaps into place"

Figure 8. Attaching the new port pan.

9

Page 230: HP 7673B - S service manual

IT

Installing the Split Top Cover

r

.~

--,-,"

~-­""1."..,~

>~

~.----,~

•.0;."

,..,~>~

j..-~

.....~..•.:

~.::,::

,~. "1........~

-:~

~.:.J.. .

_--'1'

~h-~

~

--

...~~

...,-

--

1..... e---::a

1 IIa-II~( .-)

._--

2_...._

Exploded view will aI pats

1_.. __

4. .-.:-.............

0-5 ..............

fl lOa] I

z::c::::::: If' I Ii 1..... ,.........

Pm you needfO( each step

Step by step

2

Tools you need foreach step

--

Pull out the yellow worksheet from the center of this link manual. Organizethe hardware and tools for each step. Full-scale drawings of the screws helpidentify which ones you need for each step. Ifany of the hardware is miss­ing, contact your HP representative.

Double-check the following:

o Be sure that all the screws are tight.

3

o Be sure the edge of the top cover is flush with the front grill. If it is notflush, go back to step 4.

Next, install the left side panel of the HP 5890 SERIES II GasChromatograph. The instructions are packaged with the left side panel.

Ifyou have already installed the left side panel. install the injector mountingbracket. The instructions are packaged with the controller.

Checking Your Work

103

Page 231: HP 7673B - S service manual

F"~.-

~.~-

f;

1~

~

~-, Before You Startt-";<.-

~,

~~:~.~-

~-,~.~-

'--~

~.~

~... -l.ti't",;

t~.II,~c

f

2

You must upgrade the top cover of an lIP 5890A Gas Chromatograph (GC)with the split top cover of the HP 5890 SERIES II GC before installing themounting bracket for the HP 18593B Injector Module. The SERIES IIdesign enables you to open and close the top cover while the injectormounting bracket is installed.

WAR.NING

Before Installation, tum off the oven and allow to cool. Youmust open the oven door to complete step 4.

After you have completed the installation, place this manual in the binder ofyour operating manual for the HP 7673 Automatic Sampler(Part No. 07673-60995) after the tab for appendix C "Link Manuals."

@ Copyright 1989 by Hewlelt·Packard CompanyAll Rights Reserved

~,

( '~2

Page 232: HP 7673B - S service manual

Manual Part No.07673-90215

r.c. HEWLETT~PACKARD

Printed In U.S.A.October 1989

t,\

1,,;'.

HEWLETT-PACKARD

( =- 0..(~ e

LinkManual

HP 7673 Automatic SamplerInstalling the HP 5890 SERIES II Split Top Cover

·~;x

~--

Ai";";~

o

..~_.~~

j.,--.~

,I.;~

~~~

"

i~

i-~

~

l~

..~

I..

-'~..'•....'"

-,,~~

;

.~

Page 233: HP 7673B - S service manual

I

~-: 2. Attach the New Left Side Panel~.

t~.

f

t~

a. Place the new left side panel on the GC and slide it forward. Be surethe front edge is seated. Be sure the gas lines are not in the way.

b. Using a No.2 Point Pozidriv, screw the two pan head screws into theside of the GC.

;

3Checking Your Work

After you have completed the installation, place this link manual in thebinder ofyour operating manual for the HP 7673 Automatic Sampler(Part No. 07673-6(995) after the tab for appendix C "Link Manuals."

Double-check. the following:

o Be sure the front edge of the left side panel is seated into the GC.

o Be sure aD screwsare tight.

Next, install the split top cover for the HP 5890 SERIES II GasChromatograph. The instructions are packaged with the split top cover.

If the split top cover is already installed, install the injector mountingbracket. The instructions are with the mounting hardware that is packagedwith the controller.

J

L

Figure 3. Attaching the new left side panel.

IIIIIIIIIIiIII1

II 111~1I11I1i1"1-: <, 11111111111111111

• I-:f'e

tf

~~p~~. -

~>~.

tie,1l\'".

!i~

45

Page 234: HP 7673B - S service manual

2Installing the Left Side Panel.

1. Remove the Old Left Side Panel

a Using a No.2 Point Pozidriv, remove the two pan head screws that holdthe existing left side panel onto the left side of the GC.

b. Lift off the old left side panel.

1111111111

Row access door

.Figure 2. Removing the old left side panel.

3

,'~.<t).'-~

~-t

':;(

,

...·'.·•.•..1'.·.·.···

1'.-;1~,

~-j

Page 235: HP 7673B - S service manual

I~L

1Before You Start

You must upgrade the left side panel of the HP 5890A Gas Chromatograph(00 before installing the mounting bracket for the HP 1859GB TrayModule.

Figure 1 lists the hardware and tools that you need to complete the installa­tion. Ifany of the hardware is missing, contact your HP representative.

NO.2 Poit Pozkiiy

~ --

2:-

~

:-..;.

o : •. .111111111111111

Il 0

IIIIII~ 11111111. 111111 ~1111 ~1I1111

Dan-

1 leftside palBI (05890-61480)

2 pal head screws (0515-0982)(Drawn fiJI scale)

Figure 1. Parts list.

© Copyright 1989 by Hewlett-Packard CompanyAll Rights Reserved

2

Page 236: HP 7673B - S service manual

Manual Part No.185i7-90170

~HEWLETT

~PACKARD

PrInted In U.S.A.October 1989

~.

HEWLETT-PACKARD

( =-= 0.·(~ ..

LinkManual

HP 7673 Automatic SamplerInstalling the HP 5890 SERIES II Left Side Panel

..

Page 237: HP 7673B - S service manual

sample Vials and Caps Sample Vials and caps with Injector Only

The following sample vials and caps should not be used with a tray.

2-m1 screw cap. glass sampia vials

100 cIecI' 9301-1128

Screw caps wiholtS8I:ia

100 9301-1131

PaVocange S8JD 10r snap caps ()( screw caps

100 9301-1130

11-mm idcrimp top, 1OQ-pJ~5ned~sanple rnicroviaIs

100 clear 9301-09n

11-nvn kI crinp top, 1OQ-pJ~ (al plastic)scmpie ITlicnMaJs

1000 clear 9301-0978

Payethy\ene snap caps, inckIdes recVcnnge S8JD

5181-1259100~

•11-mm criTCJ 1Dp, 1QO.p.J glass Slmpie micro¥iaIs.

12 cIBar 5080-8779

11-mm crin1t top, 2-m1 glass sam;lIe vials.

100 c&r'witKUcaps 5180-4197500 ctEa' '«til caps &beD: 5061-3350500 amber wIhcU caps 5061-3369500 lInber wQh caps &bat 5061-3348

A B 11-mm wkje oprilg crin1> top, 2-m1 sample vials. Clil

~be used wIh glass microYiaI ilserts

100 _ with ilserts 5181-1269100 _ withot.t ilserts 5181-3375100 ~ withot.t ilsefts 5181-3376100 nicn:lIIiaI inserts (A) 5181-1270500 IirnIBd YOI. ilserts(B) 5181-33n

~-

~_I

~.~

t~

~ -

i..'.'.'...:.f'f.:~~-

11-nvn crI:n1J caps with Te6on-coatBd. 5.58-fmI.raQ'ofange septa

L~

~ 100 akrT*un metaJ100 blue metal100 IJ'88I1 metal100 red metal

5181-12105181-12155181-12165181-1217

A-12A-13

Page 238: HP 7673B - S service manual

4-mI soh'ent orwash tQIJe wihscrew cap

Solvent and Waste Bottles

•T 144 c&I'

Diffusk)n caps

12

$epa~ 4-ml1xdes

144

9301-0723

07673-40180

9301-1031

3HP Recommended Supplies

This sectioecontainsa list of supplies available from Hewlett-Pactard'sAnalytical Direct.

Selecting and ordering the correct consumables like sample vials, syringes, andsepta are an important part of operating the lIP 7673 Automatic Sampler. Byusing Hewlett-Packard rerommended supplies, you can help maintain thereliability of the automatic sampler. For more information, see the currentissue of the Chroma/Qgrophy Usen Couz/og.

The following example illustrates the organization of the information.

s-aJ lIescriptioI 11-1T1Tl crln1l top, 2-m1 glass sanpIB vials.~

_~...._ ~q:wlIhcaps~.itnetlt.......,Or dIM 500 anmer willcaps & box

500 anDlrwithott caps

PIrt .......

':

j~

A-14 A-11

Page 239: HP 7673B - S service manual

Inlet Uners

You must use the correct inlet liner or insert with the automatic sampler. Ifyou do not, you may introduce peak area discrimination prOOlems into the

analysis.

SyrInges

23gauge syriIgefornets ether than on-<:cUnn

23n6 tapEIfed IlBEde syrilge for gennllUJlOS8 a1don-caum iEts

HlpI.n.B........

ZV26 tapered repiacement neecIes fa syri1ges. partramer 5181-3321 Q" 5181-1265

5181-3319

f~

~

WIth split or spIitI.ess inlets, you may try theseliners:

• 4-mm, borosilicate glass liner with glass wool plug (part number19251-60540)

• 4-mm sitantred, borosilicate glass liner with single taper and S1lan ized, glasswool plug (part number 5062-3587) .

WIth splitless inlets, you mayalso try these liners:

• 4-mm sitanired, borosilicate glass liner with a double taper' (part number5181-3315)

• 4-mm S1lantm1, borosilicate glass liner with single taper (part number5181-3316)

The single taper liners with or without the glass wool plug are possille choicesfor many environmental analyses. The single taper liner without the glass woolplug is usoaDy better when the anaIytes tend to degrade in the inlet.

The flow inverter liner (part number 18740-60840)and the quartz splitless liner(part number 18740-80220)are not reoomm.ended for use with the automaticsampler.

Practically every sample rontains nomoIati1e material that accumulates andpartially pryolyzes in the liner. Period.ally. the liner must be replaced orcleaned and deactivated.

A-10

161

16111

3

1o-~ fbced neecHe1~ fixed neEde5-~ fixed neecIe

1D-pJ, fixed neecIe1D-pJ, fixed neecIe5-pJ, fbced needle1~ removable neede5-pJ, remo¥abIe neecIe

9301-07139301-07259301-0892

5181-12675181-33605181-12735181-33215181-1265

A-15

Page 240: HP 7673B - S service manual

Inlet septa Inlet septa

11-mm low bleed septa tor aI packed 1RI spfitr'sPftlessi;lction. will 1fJ5890 SERIfS I, I11iIldnuntBlIlpenmn 350°C

11-fT1Tl OwbiBecI S8JD fa aD packed andspIIspIIessqectioo.wIh HP 589OA, 588OA, or571~maxInun~ 350°C

24 gray, sold 5080-8896

~lIIII

- LiIer

25 red, sokI 5181-1263

Inlet septa should last at least 200 injections before there is any sign ofleakage. The following factors impact the life expecta.ncy of septa.

• SyringeseIection, installation, and condition.

• Septum size, type, and installation.

• Injector mounting bracket alignment

• InIettemperature.

Change septa at the end of the day's work. It takes severalhours for the gasflow to sweepout any contaminants introduced to the inlet and for the flowrate to equi1iJrate.

Select the proper septa diameter for your GC from the followingdiagram.

'i

11-fT1Tl~ 101' use wih G('JMass SpectrometeI,Cool~ cI lIIeI ITmIffiI.IIl temperatl.r8 350°C

JCAUTION IThe ,.5-...red lewbleed septa are DOt l'eCO'MH'aded fer use witIltile HP7673Autoautic S8D1p1er. TIte smaller septa teIKI to ... IMI'e rreq.wreplacement

5-rrm <Isle sepia 1a cool. oo-coIumn iIleCtbl withHP 5890 SERES I, max!nun t81Tlpelahre 350°C

25 red, with 00Ie 5181-126025 red, sold 5181-1261

• Red: Recommended general purpose septa good foe inlettemperatures ~ ~oC.

• Gray: General purpose septa good for inlet temperatures~ 350°C.

5-mm

~rnn

5-mm

9.5-rrm

Type ci lnIet

11-mm

Packed {Sp!i/ SpfiUess

11-nvn

HP 5700 series

HP 5880A

Gas ctvornatogra.gt

HP5890

Septa are made from materials that give off volatiles especiaJly in heated inlets.H not swept away in a septum purge design, these volatiles may interfere withthe analysis. The following list of inlet septa may help you select the correcttype.

9301-1263red, said25e~

Ie-DbtI se,tI.

IR

~A-16 A-9

Page 241: HP 7673B - S service manual

f

The following diagram illustrates the shapes of the two fixed needles.Inlet Uners

4-nrn IdborosIIcaE glass liner (C) with one tapered end101' use willspiVspIIess i1Iets.

1 W/siaIized glass wool pkJg 5062-35871 wihoU: glass wool plug 5181-3316

4-nJnkI siaIized, borosicatB glass Iinef (A)withdo.tie taper toruse will spIIess inlets. The oominaIIinBr volume Is800ul,

1 5181-3315

4-nrn id borosic*l glass liner (B) with sftarized glasswool plug foruse wIh spitIspIIess inlets. ThenomilaI ..YOUne Is1000)ll.

~....'

~0Tk';,.

f,i€-:

~.~,

~ -~~.

ec.

• • •

~

k&F]

...

23.......

A ]I C

D

1 19251-60540

Uners fa use wittan liP 5880A pacted cokJrm JnIets

Inserts 10r coo oo-coIufm nets usilg HP 5890 SEAlES Y530-p. coUnns

Glass liner (0) foruse with til HP 5890A Of1-1' 571 tV'2W3G'9OA packed coIurm inlets, wIh OfwiIXUsepurn p.rgeF;;

t

f~r

A-a

Injection volumes varywith the sample volume parameter set on thecontrolling device, the size of the syringe, and whether the injector is set up fornanoIiter injection (lIP 18599N Nanoliter Adapter).

The following table mts the aWllwmate injection volumes available with theautomatic sampler.

8an1*l voome Standard kpetioo Halder IlIiectk!1Isettilg

SniloeSize 5u.J 10u.J 5pJ 10u.J1 0.5 1.0 0.1 0.22 1.0 2.0 as 1.03 1.5 3.0 1.0 2.0

4 2.0 4.0 1.5 3.0

5 2.5 5.0 2.0 4.0

Cool o.c0k • kill

s~e

I

IR--bIsert

25

1 stalldlld1 _ to1/4-il. metal1 adap: to1/B-itmetal25 ~ ilIet inIlr

1 tor stalldald COOJm

5080-8732

19359-2055018745-8001018709-800305080-8732

19245-20580

A-17

Page 242: HP 7673B - S service manual

FIXedand removable needle syringes are available to use with the automaticsampler.

• 23 gauge needle used with packed,porged-pack:ed, and spIitIsplitless inlets.

• 23/26 gauge tapered needle used with on-column mlets, TIm syringe canbe used asa general JKlIpose syringe with all inlets.

The syringe needle must havea flat tip. Do not use sharp-tipped needles (e.g.,point style 2). When used with the automatic sampler, these needles tear theinlet septum causing leaks. Also, sharp-tipped needles wipe off on the septumas they exit resulting in a large solvent tail on the chromatogram..

Syringes

5958-94505958-94415958-94425958-94435958-94445958-94455958-94465958-94475958-9448

18587-60560label appicatoc

18587ABlr'Cooe Reader labels

NlJrDlred1 to1,(XXl1,001 to 2,0002,001 kJ3,0003,001 kJ4,0004,001 to 5,0005,001 to6,0006,001 to7,0007,00110 8,0008,001 to 9,000

C...--:J

Bar Code Supplies

Label klcator tor

cawed vialsIIf"NlIImf1 vials

TestviaJ

18587-0012018587-D0130

18587-60540

sw,tip

RIIU,

The following diagram illustrates some of the aitical syringe dimensions. .•.~These dimensions do not mate up a complete set of specifications. Someof the ~dimensions are too difficult to measure without special instruments.

A-18

I 'U 1U------JH, n , ,: n ¥ n ~1U±D.1 rr--=-

AI..easioa II ..11....1'1

A-7

Page 243: HP 7673B - S service manual

~ ....

~

~

~:.".

~~

Solvent and Waste Bottles

4-ml bottles are placed in the injector turret to bold solvents used for washingthe syringe between samples and to hold the solvent and sample rinses. Duringa solvent wash. the syringe needle cannot reach the last 2.5 ml in the bottle.For more information, see "Filling and Placing the Bottles" on page 3-17.

You can use diffusion caps or septa on these bottles to reduce evaporation anddiffusion of your solvents and waste. HP reoommends diffusion caps oversepta

fOf' two reasons.

Miscellaneous

Vial racks

5 plastic, hokIS50 vials 9301-07225 foam. ho6ds 50 viats 9301..Q967

01673-20570

5181-1213

9301-0990

871D-0979

18596-40015

11-mm deCapping tea

11-1Mlcr'~ lQ tOO

11-1TIll~ I*Jg

1000

,....,sa.,/ screw torhokmg syrtnge pIooger

·.U

12£

~ set oftray quadrcns

A!~ 4 ~.~~-~....~

4-al1IInII1I'waslllIatUt

DiInIoICI'

ser.cap.----==~~

• For many common solvents, the rate of diffusion out of the bottle is lesswith a diffusion cap than with a septum that has been punctured with astandard syringe needle.

• The diffusion cap allows multiple entrances into a bottle withoutcontaminating the liquid inside the bottle with small pieces of septum

material

~.~'.-

~;r~

~­~..

..c-,..

E _.~--;

R~;-

&C.~..

~:

f

~

~?

~\~:

A-6A-19

Page 244: HP 7673B - S service manual

"

e......GIllie TIf

~..*-I A......,e-I Dbt: se,b.

I-SpragR-Iaert

Parts forHP 5890 SffUES IIcool on-coIurm inlet

NeecIe gtidetop 19245-206708eJmn oot baseassentit 19245-80520Spri1g 19245-60760

Injection portru101" '* 5880A1 1~~

Once the vial septum is punctured by the~ needle, the sample mayinteract with the septum material and affect the a.naI)'sis. There are three typesof septum used in vial caps. Each has a different resistance to interaction withsolvents.

• Redlorange: Made from natural robber and coated with Thfloo- on thesample side.A general purpose rubber formulation .intended primarily foraqueous liquid fill products with a pH range 4-0-7.5.Least resistant tosolvents after puncture.

• Red: Made from silicone rubber coated with Thfloo- on both sides. A highquality, low extractable silicone rubber intended fOr' applications requiringresistance to high extremes of heat. Somewhat resistant to solvents afterpuncture. Slightly poorer resealing characteristicswhen application callsfor multiple injections from the same vial

,11.~

MolI'Iting post1 07673-61035

• Black: Made from Vrtone . Most resistant to solvents.

Pcrtirlg or SlWCrt rod foruse wit!HP 5890 SERIES I GCs wih SIN greater than3033A321001 ~1525

I CAUTION IBlack septa made fI'MI Vitae are BOtrecoII1DleDded Cor use with the UP18593Aud B IDjec:ter Modules. UsiDcVitoB. may cause • syringeerror.

The following diagram illustrates the recommended and minimum diameter forvial septa.

~

Basic~ KI 07673-60840

6 1Q-p.l2:Vl61aA*ed syrilges144 4-mI soMIlfIu;h baties wIh caps144 dffuslon caps1000 2-m sam;:E vials wih cri11l caps5 viaf racks144 11-rrm 5eJD

A-20

NalOOlef adaPer1 HP 18599N

i.5I (0.221 iL) rtCOII.eHed

U5 (0.113"') .il'."

AI fiIIeaIots II 1Il1n.cm

H you are loading vials directly into the injector turret. you can use 2-mlsample vials with screw caps. Do not use screw cap vials with the tray.

Use amber glass vials for light-sensitive samples.

A-5

~';

Page 245: HP 7673B - S service manual

sample Vials and caps

~~

'Ib maintain the tray gripper's reliability, you must use glass vialswhich meetlIP specifications. Service callsand repairs found to be due to vials that do not

meet these specifications are not oovered under either warranty or serviceoootraet.

The following diagram illustrates most of the aitical dimensions for the samplevials and microria1 inserts. These dimensions do not makeup a complete set ofspecifications- Some of the dimensions are too difficnh to measure withoutspecial instruments.

11.7 :i:

( .

AD dl...... I.1IiIIl1Itters

II [---r

U-=1~---LI

g.I •• fIj ......1'111

2U

IIIa..WIaert

1 1--5 hI •I ~ i' , ..

f[."-

.,;:-

f

~

~..

f\t·

~

'"

,

~,to--r

A-4

Page 246: HP 7673B - S service manual

I

2Selecting Consumables

Thissection oontains infonnation to help you select eonsumables to use withthe lIP 7673 Automatic Sampler. Some aitical dimensions are provided forsamplevials, syringes, and septa. Ifyou are using consumabIes that do not havethesedimensions, the reliability of the automatic sampler maybe affected.

A-3

~

')i-~---~

";i5

'?

)i

'---~

1

Page 247: HP 7673B - S service manual

1Introduction

r~·.··.~.·.·.· ·>"--

~~-- -

~~.e­'.I

This is an integral part cl the lIP 7673 Automatic Sampler Operating Manual(part number 07673-60995). It is designed to be updated on a regnIar basis andkq)t in the manual's binder after the chapter tab for appendix A,"Cons:mnahles...

'Ibis booklet contains basic information on bow to select and orderconsnmables foc the automatic sampler. It is dirided into two parts.

• Detailed infoonation to heIp)'OU select the appropriate eonsamables fOl'the lIP 7673. 'Ibis section is divided into five parts.

• Sample vials and caps.

• Solvent and waste bottles.

• Syringes.• Inlet septa.

• Inlet liners,

• A list cl Hewlett-PactaId supplies recommended for the lIP 7673 andavailable from Analytical Direct.

e Copyrtght 1989, 1990. 1991 by HewIeIl-Packard companyAI RIghts Reserved

A-2

Page 248: HP 7673B - S service manual

""·":1\: ;!I'~/' - . " -';:-:',,\ I- ·"-;'f ~ ~I ~~I~;I ~',::' .·;1

a~ ..-c Q)-c.~ E tn

CD-U ca .aen ca

E

~U ::J.- tn

I:D..,

cas 0E 0I Ii « 0f--4 ••

[0/ c:(I ..,:I .~f--4 « "tJ

c~ M CD

'" C.

~ CD c.

'"c(

~ a.::E:

~

::c