Highly sensitive cell concentration detection by resonant ...ap.polyu.edu.hk/apzhang/PDF_papers/2019...
Transcript of Highly sensitive cell concentration detection by resonant ...ap.polyu.edu.hk/apzhang/PDF_papers/2019...
-
0733-8724 (c) 2018 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission. See http://www.ieee.org/publications_standards/publications/rights/index.html for more information.
This article has been accepted for publication in a future issue of this journal, but has not been fully edited. Content may change prior to final publication. Citation information: DOI 10.1109/JLT.2019.2907786, Journal ofLightwave Technology
Abstract—This paper presents an original design for a volume
refractive index (RI) sensor for cell concentration detection based
on the unique resonant optical tunneling effect (ROTE). In this
design, the solution sample is introduced into the ROTE resonant
cavity, whose reflection spectrum presents a large shift of its
resonant dip in response to the tiny change of the solution RI.
Performance calibration using the polystyrene (PS) particle
solution indicates that the sensitivity is 19,000 nm/RIU and the
total Q factor is 620. Experiments with hepatoma cells have
obtained sensitivity values as high as 2.53 nm/(amol/ml) and a
detection limit of 1.2 × 105 cells/ml. Compared with previously
reported optical cell sensors, this original work implements the
ROTE mechanism for the first time and presents superior device
performance. Thus, the proposed design has a high potential for
use in biomedical applications, such as drug discovery, cell
research and medical diagnoses.
Index Terms—resonant optical tunneling effect, optical
biosensor, cell concentration, figure of merit.
I. INTRODUCTION
ELL concentration (number of cells/ml) is considered a
key indicator that reflects a series of cell processes, such as
viral infection [1-3], abnormal hematopoiesis [4], drug
reactions [5], and autoimmune disease [6]. Refractive index (RI)
measurements are widely used to quantitatively analyze
extremely small variations in the chemical components of a
solution. For example, 10-9 refractive index unit (RIU) is
equivalent to 1 femto-mol/L of salt in water. Recently, RI
measurements have been implemented and have real-time and
label-free methods of detecting cell concentrations at high
resolution [7].
Most of the current typical sensing mechanisms/structures of
RI sensors are based on near-field optics [8], such as optical
fiber Bragg grating (FBG) [9], long period fiber grating (LPG)
[10-11], photonic crystal/photonic crystal fiber (PCF) [12-13],
and whispering gallery mode (WGM) [14-18]. In addition, 2D
materials have been widely used in optical sensors because of
This study is financially supported by the National Natural Science
Foundation of China (No. 61501316, 61471255, 51622507, 51505324,
81602506 and 61377068), the Shanxi Provincial Foundation for Returned Scholars (2015-047), 863 project (2015AA042601), Excellent Talents
Technology Innovation Program of Shanxi Province (201605D211027),
Hundreds of Talents of Shanxi Province, and Research Grants Council of Hong Kong (N_PolyU505/13, 152184/15E and 152127/17E).
A. Q. Jian, L. Zou, G. Bai, Q. Q. Duan, Y. X. Zhang and S. B. Sang are with
Taiyuan University of Technology, MircoNano System Research Center,
College of Information and Computer Science, No. 79, Yingze Road (West),
their distinctive optical properties [19-21]. Y. J. Xiang et al.
designed a series of hybrid structure surface plasmon resonance
(SPR) sensors, in which 2D materials are employed for sensor
performance enhancement. For example, a graphene layer was
utilized to prevent the metal layer (Al) from oxidizing and to
increase both light and biomolecule adsorption to enhance the
sensitivity [22]. Moreover, the graphene layer acted as a
sensitive film to facilitate coupling between the two surface
plasmon polaritons (SPPs) modes, which are capable of high-
resolution RI sensing in the THz range [23]. Furthermore, MoS2
was found to be more absorptive of light energy and the
sensitivity can be further improved [24]. The heterostructures
of a limited number of black phosphorus (BP)-graphene/TMDC
(MoS2, WS2, MoSe2, WSe2) layers were also introduced to the
SPR sensors, and the performance of the various combinations
of 2D materials was compared [25].
Hyperbolic metamaterials, which are characterized by
hyperbolic dispersion, present a sensitive multilayer structure
and have drawn intensive research attention recently, and they
show considerable potential for RI sensing. Significant
sensitivity enhancement can be achieved for high-order modes
due to their spectral positions closer to the resonance of certain
permittivities [26] and stronger electromagnetic fields near the
interface [27]. Based on its unique characteristics, a compact
plasmonic biosensor platform has been developed with a
maximum of 30,000 nm/RIU and a record figure of merit (FOM,
which is defined as the ratio of the sensitivity to the full width
half maximum) of 590 [28].
For these detection methods above, evanescent waves near
the two-media interface are exploited. The strong confinement
of the evanescent waves near the interface favors the strong
light-analyte interaction. However, the intensity of evanescent
waves decays exponentially with distance away from the
interface, which inevitably limits the spatial interaction depth
to the subwavelength range. These sensors cannot be used to
effectively measure the entire biological sample with a size
larger than the wavelength (e.g., eukaryotic cells at 10–100 µm)
or the samples/particles naturally suspended in the solution.
Taiyuan, 030024, Shanxi, China.(e-mail: [email protected];
[email protected]; [email protected]; [email protected];
[email protected]; [email protected]; Q. W. Zhang is with Key Laboratory of Specialty Fiber Optics and Optical
Access Networks, Shanghai University, Shanghai 200072, China
([email protected]). X. M. Zhang is with The Hong Kong Polytechnic University, Department of
Applied Physics, Hung Hom, Kowloon, Hong Kong SAR, China (e-mail:
Highly sensitive cell concentration detection by
resonant optical tunneling effect
Aoqun Jian, Lu Zou, Gang Bai, Qianqian Duan, Yixia Zhang, Qianwu Zhang, Shengbo Sang, and
Xuming Zhang
C
mailto:[email protected]:[email protected]:[email protected]:[email protected]
-
0733-8724 (c) 2018 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission. See http://www.ieee.org/publications_standards/publications/rights/index.html for more information.
This article has been accepted for publication in a future issue of this journal, but has not been fully edited. Content may change prior to final publication. Citation information: DOI 10.1109/JLT.2019.2907786, Journal ofLightwave Technology
However, in the case of volume RI sensing, light waves
completely propagate through the solution containing the
targeted analyte and interact with every particle in the shaft,
which is particularly useful for detecting samples with ultralow
concentrations. Volume RI sensing can be realized by some
classical methods/schemes, e.g., Fabry-Pérot (FP) etalon.
However, the complex fabrication process of a highly reflective
mirror with tiny absorption prevents its comprehensive
commercial application [29-30].
Since the first demonstration by Hayashi’s group [31], the
resonant optical tunneling effect (ROTE) has shown its unique
ability to bridge wave optics and quantum physics [32-37] and
has considerable potential for use in both theoretical studies
[38-39] and practical applications as photonic devices [40-43].
This study will present the first experimental demonstration of
a RI sensor based on the ROTE. This sensor overcomes the
spatial limitation of the evanescent field and can achieve high
sensitivity detection of whole cells. With simple fabrication, the
sensor successfully detects hepatoma cell concentrations with a
resolution of 1.2 × 105 cells/ml, and its FOM value is
comparable to that of other cutting-edge optical methods [7, 44].
II. MATERIAL AND METHODS
A. Concept and sensor design
The ROTE originates from a relatively simple phenomenon,
the optical tunneling effect (a.k.a., frustrated total internal
reflection, FTIR). When the incident angle θ is larger than the
critical angle of the interface between high and low RI media,
some of the evanescent light can pass through the total
reflection interface to form tunneled light if the low-RI layer is
thin enough [45]. Briefly, the ROTE refers to the resonance
effect of such tunneled light in a well-designed resonator, and
its structure consists of five layers with a high-low-high-low-
high distribution of RI, which are named in turn as follows:
input layer, first tunneling gap, central slab, second tunneling
gap, and output layer along the light propagating direction as
shown in Fig. 1a.
By using the finite difference time domain (FDTD) method,
the electric field distributions of the ROTE structure (one-
dimensional model) at the resonant wavelength are illustrated
in Fig. 1b (S-polarized light) and Fig. 1c (P-polarized light).
Fig. 1. (a) Schematic of the multilayer structure for the resonant optical
tunneling effect (ROTE). The electric field distributions of light propagation in the ROTE structure at the resonant wavelengths of reflection for (b) S-polarized
light and (c) P-polarized light. (d) Design of the ROTE sensor for the volume
RI detection.
Except for the length of the central slab (15 μm), which is
shortened to save the simulation resource and time, the
parameters involved in the simulation are equivalent to the
values shown in TABLE I. For both polarization states, the
electric field tends to decrease in the first tunneling gap and then
form a locally enhanced mode in the central slab, similar to a
TABLE I
PARAMETERS OF THE RESONANT OPTICAL TUNNELING EFFECT (ROTE)
MULTILAYERED STRUCTURE
Parameter Symbol Value
Incidence angle α 61°
Width of tunneling gap d 3 μm
Width of central slab g 300 μm
Refractive index of input and
output layers nin, nout 1.59+9.84 × 10
-9 i
Refractive index of the
tunneling gaps n1 1.308+5 × 10
-6 i
Refractive index of central slab n2 To be determined
-
0733-8724 (c) 2018 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission. See http://www.ieee.org/publications_standards/publications/rights/index.html for more information.
This article has been accepted for publication in a future issue of this journal, but has not been fully edited. Content may change prior to final publication. Citation information: DOI 10.1109/JLT.2019.2907786, Journal ofLightwave Technology
standing wave, before it finally decays in the second tunneling
gap. These two graphs show that the incident wave propagates
across the whole ROTE structure, thereby enabling volume RI
sensing.
According to our previous study [40], the ROTE
transmission is more sensitive to RI changes in the central slab
than to that of the tunneling gaps. Therefore, the central slab is
chosen as the sensing element rather than the tunneling gaps. A
schematic diagram of the sensor based on ROTE is shown in
Fig. 1d. In this study, K9 prisms coated with low-RI polymer
layers are used to form the photonic barriers and the cell
solution injected between the two photonic barriers acts as the
central slab. The polymer layer (MY-131 series, RI = 1.308 RIU)
forms a low-RI dielectric layer between the prism (RI = 1.59
RIU) and the liquid sample to be tested (RI = 1.35–1.37 RIU).
In this way, the RI distribution has the order of high-low-high-
low-high, which meets the requirement of the ROTE. The
detailed parameters of the sensor device are listed in TABLE I.
Because the ROTE peak/dip of the S-polarized light is much
sharper than that of P-polarized light (by a factor of 1000 [40]),
the sensor is chosen to work in the S-polarized state in the
following experiments to achieve a better performance.
B. Preparation of the ROTE multilayered structure
Because the triangular prism is too heavy to be safely fixed
on the spin coating machine, the low-RI polymer film cannot
be directly coated on the triangular prism surface. Thus, it is
coated on a glass slide with the same material (K9) as the
triangular prism. Then, the glass slide is seamlessly adhered to
the triangular prism reflector by a UV curing adhesive. Because
the thickness of the low-RI medium strongly influences the
experimental results, it is carefully checked by ellipsometry
before the experiment.
C. Experimental setup
The schematic diagram of the experimental setup is shown in
Fig. 2a. Because of the low absorption of the cell in the near
infrared and infrared wavelengths, a tunable infrared fiber laser
(TIFL, New Focus TLB-6700) is employed as the light source
to reduce the influence of the laser thermal effect on cell
physiology. The polarization state of the emitted laser is tuned
to be S-polarized by a fiber polarization controller (FPC) and
verified by a polarization splitting prism (not shown in Fig. 2a).
Before the incident laser is collimated by the single mode fiber
(SMF) collimator, an optical fiber attenuator is utilized to adjust
the laser intensity. In the experiment, the incident angle (61°) is
greater than the total internal reflection angle by 1°. Finally, the
reflected laser is collected by a photodetector (New Focus
1811-FC-AC), recorded together with the synchronous signal
from the laser by a digital storage oscilloscope (DSO, GDS-
2302A), and further transferred as the reflection spectrum of the
ROTE structure.
In the experiments, the position of the reflected spot is
recorded and the incident angle can be calculated based on
geometric relationships. The thickness of the low RI film is
measured by an ellipsometer (AST SE200BA) in advance.
According to the theoretical study, the free spectral range (FSR)
of the ROTE reflection spectrum is mainly determined by the
incident angle, the width of the gap and the RI of the medium
in the gap. When the incident angle and RI are input as
constants, the FSR is only dependent on the separation of the
prisms. Therefore, the width of the analyte region can be
derived by fitting the experimental curve in the case of the gap
filled with the liquid using a definite RI. Before every series of
cell density measurements, the gaps are filled with the RI
matching fluid and the FSR is utilized as a reference for the
separation adjustment. Thus, the sensor is calibrated and the
uniformity of the analyte’s thickness is ensured in the
experiment. Different analyte solutions are prepared with
different concentrations, and the RIs of these solutions are
measured by a commercial digital RI detector (AR200 digital
hand-held refractometer). Then, one of the solutions is injected
into the gap (i.e., the central slab of the ROTE) between the two
triangular prisms with proper pressure to overcome the surface
tension. The thickness of the liquid analyte layer is equal to the
width of the gap. Because the irradiations out of resonance are
totally reflected at the high-low RI interface, their intensities are
considered to be the reference for output normalization. After
each measurement, the gap is washed with alcohol and fully
dried to restore the initial operation conditions. Finally, the
wavelength shift of the ROTE characteristic dip in the reflected
spectrum is recorded for further analysis.
Fig. 2. Experimental setup of the ROTE sensor. (a) Entire measurement system
that utilizes a tunable infrared laser (TIFL) to sweep the wavelength range to detect the reflection spectrum; and (b) close-up and (c) photograph of the ROTE
structure.
III. RESULTS AND DISCUSSION
A. Performance calibration using polystyrene particles
The sensor performance is first calibrated using polystyrene
(PS) particles, which have equivalent properties as the material
of interest, before the sensor is used to measure real cells. The
PS is an optically transparent material, and the dispersed PS
particles in aqueous solution are similar to the cells in culture
fluid. Therefore, PS particles are often used as a substitute for
cell experiment. The concentration of the PS particle solution
-
0733-8724 (c) 2018 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission. See http://www.ieee.org/publications_standards/publications/rights/index.html for more information.
This article has been accepted for publication in a future issue of this journal, but has not been fully edited. Content may change prior to final publication. Citation information: DOI 10.1109/JLT.2019.2907786, Journal ofLightwave Technology
can be evaluated with ultrahigh accuracy by weighing the dry
Fig. 3. (a) Shift of the resonant dip with the change of the RI of the analyte. (b)
Measured thermal drift as a function of the test time. (c) Enlarged view of the
resonant wavelength shift. The experimental reflection spectra are compared with the simulation results.
PS particles to be dissolved. These particles are utilized to
mimic the cell solution and to preliminarily characterize the
performance of the ROTE sensor. In the experiment, different
amounts of PS particles (5 µm diameter, standard deviation ≤
0.027 µm, Duke Scientific Corp.) are dissolved in anhydrous
ethanol as the target analyte. Because the RI of the PS particles
(1.590 RIU) is larger than that of anhydrous ethanol (1.361
RIU), a higher concentration of PS particles in the mixed
solution would result in a greater RI. The experimental results
indicate that the ROTE absorption dip in the reflection spectrum
experiences a redshift as the RI of the intermediate analyte layer
increases (Fig. 3a). The sensitivity obtained is 19,000 nm/RIU,
and the Q value of the ROTE absorption dip is 620.
Because long-term laser transmission will directly heat the
analyte in the central slab and change its RI during the
measurement, the variations in the ROTE resonant wavelength
over the test time under a particular laser power are recorded
(Fig. 3b). The resonant wavelength has a linear relationship
with the measured time, and the slope is ‒5 pm/min. To avoid
this problem, the thermal drift is removed in all measurements
in this study. A close-up of the ROTE reflection spectrum is
presented in Fig. 3c to clearly show the details of the
wavelength shift due to the RI variation. The experimental
results are consistent with the theoretical analysis obtained by
the transfer matrix method (TMM), and additional details on
this method can be found in the authors’ previous work [40].
The slight difference between the theoretical and experimental
results may be related to certain negligible factors, such as the
optical dispersion of the materials involved in the system (low
RI polymer and liquid analyte) and the parallelization of the
prisms.
B. Q factor analysis of the ROTE dip
The experimental ROTE reflection spectrum and the
simulation results are shown in Fig. 4. According to Gansch et
al. [46], the total Q factor of the resonator can be written as
follows:
radabsstrtotal QQQQ
1111 (1)
where Qrad is the Q factor due to the coupling loss of the
incident light, Qstr accounts for the loss of reflection and
refraction in the model structure, and Qabs originates from the
loss of material absorption.
The values of these Q factors can be obtained from the
simulation and the experiment. For instance, Qstr can be
determined from the simulation without considering the
material absorption (the red curve in Fig. 4). However, if the
complex value of RI is used when determining the material
absorption (the orange curve in Fig. 4), the ROTE characteristic
dip becomes significantly broader, and a value of Qabs = 650 is
obtained by comparing these two simulation curves. Based on
the experimentally measured spectrum (the green curve in Fig.
4), a value of Qtotal = 620 is directly obtained. Then, a
comparison of the differences between the simulated and the
experimental spectra generates a Qrad value of 1.3 × 104. To
achieve a high quality factor, we improve our device based on
multiple aspects, such as selecting low absorption materials and
optimizing the structural dimensions. Compared with the
simulations performed by our group, negative factors remain in
the experiment, such as the parallelization of the prisms, which
leads to a decline in the Q factor.
-
0733-8724 (c) 2018 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission. See http://www.ieee.org/publications_standards/publications/rights/index.html for more information.
This article has been accepted for publication in a future issue of this journal, but has not been fully edited. Content may change prior to final publication. Citation information: DOI 10.1109/JLT.2019.2907786, Journal ofLightwave Technology
Fig. 4. Comparison of the reflection spectra of the ROTE sensor with the FP
etalon. The ROTE spectra are obtained from the experiment and the simulations
under different conditions (with/without absorption).
Based on Eq. (1), the calculated value of Qstr is 1.28 × 106,
which is very high and demonstrates that the ROTE structure is
an excellent resonator. For further comparison, the reflection
spectrum of an ideal FP etalon with 99.6% reflectivity mirrors
is also plotted in Fig. 4. The cavity length of the FP etalon is
equivalent to the effective length of the ROTE structure. The
FP cavity has a Qstr value of 3 × 105, which is lower than that of
the ROTE structure by one order of magnitude. Since a higher
Q factor is more favorable for a higher resolution of volume
sensing, the ROTE structure is used in this work to detect the
cell concentration instead of the classical FP cavity.
The loss of energy is mainly caused by absorption by the
material in the system, which includes three parts (materials),
such as the prism (glass), the low-RI layer (polymer), and the
analyte solution (alcohol solution with PS particles). The main
source of absorption is from the liquid analyte due to its high
RI imaginary part and its large thickness (~300 μm). A dilemma
arises when we try to detect a low concentration solution based
on RI because higher sensitivity indicates a stronger light-
analyte interaction, which also induces stronger absorption of
the solvent (typically water or phosphate buffered solution, PBS)
and thus a lower Q factor and reduced sensor resolution.
Therefore, prerequisites for excellent sensor performance
include the reasonable optimization of parameters and a
consideration of the balance between sensitivity and the Q
factor.
C. Cell detection experiments
Hepatoma cells of different concentrations are used as the
analyte to be measured in this experiment. The initial cell
amount is counted on a hemocytometer. Pure cells are separated
from the cell culture fluid by centrifugation. Then, different
amounts of cell-containing solutions (2, 2.2, 2.4, 2.6, and 2.8 µl)
are extracted using a pipette before they are mixed with 500-µl
RI matching solution (Cargille AA, 1.4580 RIU), which can
maintain the cell morphology during the measurement process.
Five different concentrations of cells are prepared (1.19 × 105,
2.39 × 105, 3.59 × 105, 4.79 × 105, and 5.99 × 105 cells/ml). Fig.
5a shows the ROTE reflection spectra of the five sample
solutions. The RI variations can be retrieved by the shift of the
ROTE dip. Fig. 5b plots the dip wavelength as a function of the
cell concentration (the Q factor is approximately equal to 950).
The sensitivity reaches as high as 2.53 nm/(amol/ml), and the
cell concentration of 1.2 × 105 cells/ml can be easily
discriminated.
Fig. 5. (a) ROTE dip presents a redshift as the cell concentration of the analyte
increases. (b) Dependence of the ROTE resonant wavelength on the cell concentration.
TABLE II compares our ROTE sensors with some other
classical cell sensors in the literature, such as localized surface
plasmon resonance (LSPR) sensors [44], FP etalons [47], titled
TABLE II
COMPARISON OF OUR ROTE SENSOR WITH OTHER CLASSICAL SENSORS IN
THE LITERATURE
Working
principle
Sensitivity
(nm/RIU) FOM
Detection Limit
(cells/ml)
LSPR [44] 535 104~106
FP etalon [47] 1013
LSPR + FP etalon [49] 496 ~104
TFBG [7] 180 2500 ~105
PCF [48] 38000 4.6×10-7 RIU
ROTE 19000 7600 1.2 × 105
LSPR, localized surface plasmon resonance FP, Fabry–Pérot
TFBG, tilted fiber Bragg grating PCF, photonic crystal fiber
-
0733-8724 (c) 2018 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission. See http://www.ieee.org/publications_standards/publications/rights/index.html for more information.
This article has been accepted for publication in a future issue of this journal, but has not been fully edited. Content may change prior to final publication. Citation information: DOI 10.1109/JLT.2019.2907786, Journal ofLightwave Technology
fiber grating (TFBG) sensors [7] and PCF sensors [48]. From
TABLE II, the ROTE sensor is superior in sensitivity and
comparable in FOM. In addition, the detection limit of the
ROTE sensor is among the best. These findings demonstrate the
excellent performance of our ROTE cell concentration sensor.
IV. CONCLUSIONS
In this study, ultrahigh sensitive cell concentration detection
is experimentally demonstrated by exploiting a special physical
mechanism of the ROTE. By using PS particles as the model,
the ROTE sensor has been calibrated to have a sensitivity of
19,000 nm/RIU and a Q value of 620. The experiments with
hepatoma cells have shown that the detection limit of the
concentration resolution is as low as 1.2 × 105 cells/ml, which
is comparable to the best results obtained by other optical cell
sensors. The ROTE sensor will be particularly useful for drug
screening, environment protection and drinking water safety.
REFERENCES
[1] U. Jarocka, R. Sawicka, A. Sirko, J. Radecki and H. Radecka.
“Electrochemical Immunosensor for Detection of Antibodies against
Influenza a Virus H5n1 in Hen Serum,” Biosens. Bioelectron, vol. 55, no.
9, pp. 301, 2014.
[2] V. Peltola, J. Mertsola and O. Ruuskanen. “Comparison of Total White
Blood Cell Count and Serum C-Reactive Protein Levels in Confirmed
Bacterial and Viral Infections,” J. Pediatr., vol. 149, no. 5, pp. 721-724,
2006.
[3] H. Vaisocherová, K. Mrkvová, M. Piliarik, P. Jinoch, M. Šteinbachová
and J. Homola. “Surface Plasmon Resonance Biosensor for Direct
Detection of Antibody against Epstein-Barr Virus,” Biosens. Bioelectron,
vol. 22, no. 6, pp. 1020-1026, 2007.
[4] D. A. Jaitin, H. Keren-Shaul, N. Elefant and I. Amit. “Each Cell Counts:
Hematopoiesis and Immunity Research in the Era of Single Cell
Genomics,” Semi. Immunol., vol. 27, no. 1, pp. 67-71, 2015.
[5] C. V. Fletcher, K. Staskus, S. W. Wietgrefe, M. Rothenberger, C. Reilly,
J. G. Chipman, G. J. Beilman, A. Khoruts, A. Thorkelson and T. E.
Schmidt. “Persistent Hiv-1 Replication Is Associated with Lower
Antiretroviral Drug Concentrations in Lymphatic Tissues,” Proc. Natl.
Acad. Sci. USA, vol. 111, no. 6, pp. 2307-2312, 2014.
[6] Z. S. Wallace, H. Mattoo, M. Carruthers, V. S. Mahajan, E. D. Torre, L.
Hang, M. Kulikova, V. Deshpande, S. Pillai and J. H. Stone.
“Plasmablasts as a Biomarker for Igg4-Related Disease, Independent of
Serum Igg4 Concentrations,” Ann. Rheum. Dis., vol. 74, no. 1, pp. 190,
2015.
[7] T. Guo, F. Liu, Y. Liu, N. K. Chen, B. O. Guan and J. Albert. “In-Situ
Detection of Density Alteration in Non-Physiological Cells with
Polarimetric Tilted Fiber Grating Sensors,” Biosens Bioelectron, vol. 55,
no. 9, pp. 452-458, 2014.
[8] C. Li, G. Bai, Y. Zhang, M. Zhang and A. Jian. “Optofluidics
Refractometers,” Micromachines, vol. 9, no. 3, pp. 136, 2018.
[9] D. Sun, T. Guo and B. O. Guan. “Label-Free Detection of DNA
Hybridization Using a Reflective Microfiber Bragg Grating Biosensor
with Self-Assembly Technique,” J. Lightwave Technol., vol. PP, no. 99,
pp. 1-1, 2017.
[10] Z. He, F. Tian, Y. Zhu, N. Lavlinskaia and H. Du. “Long-Period Gratings
in Photonic Crystal Fiber as an Optofluidic Label-Free Biosensor,”
Biosens Bioelectron, vol. 26, no. 12, pp. 4774-4778, 2011.
[11] L. Rindorf and O. Bang. “Highly Sensitive Refractometer with a
Photonic-Crystal-Fiber Long-Period Grating,” Opt. Lett., vol. 33, no. 6,
pp. 563-565, 2008.
[12] Y. L. Hoo, W. Jin, L. Xiao, J. Ju and H. L. Ho. “Numerical Study of
Refractive Index Sensing Based on the Anti-Guide Property of a
Depressed-Index Core Photonic Crystal Fiber,” Sensor. Actuat. B-Chem.,
vol. 136, no. 1, pp. 26-31, 2009.
[13] D. K. Wu, B. T. Kuhlmey and B. J. Eggleton. “Ultrasensitive Photonic
Crystal Fiber Refractive Index Sensor,” Lasers and Electro-Optics, 2009
and 2009 Conference on Quantum electronics and Laser Science
Conference. CLEO/QELS 2009. Conference on, pp. 322-324, 2009.
[14] Y. Luo, X. Chen, M. Xu, Z. Chen and X. Fan. “Optofluidic Glucose
Detection by Capillary-Based Ring Resonators,” Opt. Laser Technol.,
vol. 56, no. 1, pp. 12-14, 2014.
[15] K. Scholten, X. Fan and E. T. Zellers. “A Microfabricated Optofluidic
Ring Resonator for Sensitive, High-Speed Detection of Volatile Organic
Compounds,” Lab Chip, vol. 14, no. 19, pp. 3873, 2014.
[16] F. Vollmer, D. Braun, A. Libchaber, M. Khoshsima, I. Teraoka and S.
Arnold. “Protein Detection by Optical Shift of a Resonant Microcavity,”
Appl. Phys. Lett., vol. 80, no. 21, pp. 4057-4059, 2002.
[17] S. X. Zhang, L. Wang, Z. Y. Li, Y. Li, Q. Gong and Y. F. Xiao. “Free-
Space Coupling Efficiency in a High-Q Deformed Optical Microcavity,”
Opt. Lett., vol. 41, no. 19, pp. 4437, 2016.
[18] Y. Zhi, X. C. Yu, Q. Gong, L. Yang and Y. F. Xiao. “Single Nanoparticle
Detection Using Optical Microcavities,” Adv. Mater., vol. 29, no. 12,
2017.
[19] R. Irshad, K. Tahir, B. Li, Z. Sher, J. Ali and S. Nazir. “A Revival of 2d
Materials, Phosphorene: Its Application as Sensors,” J. Ind. Chem. 2018.
[20] P. Kang, M. C. Wang and S. Nam. “Bioelectronics with Two-
Dimensional Materials,” Microelectron. Eng., vol. 161, pp. 18-35, 2016.
[21] C. Zhu, D. Du and Y. Lin. “Graphene and Graphene-Like 2d Materials
for Optical Biosensing and Bioimaging: A Review,” 2D Mater., vol. 2,
no. 3, pp. 032004, 2015.
[22] L. Wu, J. Guo, H. Xu, X. Dai and Y. Xiang. “Ultrasensitive Biosensors
Based on Long-Range Surface Plasmon Polariton and Dielectric
Waveguide Modes,” Photonics Res., vol. 4, no. 6, pp. 262-266, 2016.
[23] J. Zhu, B. Ruan, Q. You, J. Guo, X. Dai and Y. Xiang. “Terahertz
Imaging Sensor Based on the Strong Coupling of Surface Plasmon
Polaritons between Pvdf and Graphene,” Sensor. Actuat. B-Chem., vol.
264, pp. 398-403, 2018.
[24] L. Wu, Y. Jia, L. Jiang, J. Guo, X. Dai, Y. Xiang and D. Fan. “Sensitivity
Improved Spr Biosensor Based on the Mos 2/Graphene–Aluminum
Hybrid Structure,” J. Lightwave Technol., vol. 35, no. 1, pp. 82-87, 2017.
[25] L. Wu, J. Guo, Q. Wang, S. Lu, X. Dai, Y. Xiang and D. Fan. “Sensitivity
Enhancement by Using Few-Layer Black Phosphorus-Graphene/Tmdcs
Heterostructure in Surface Plasmon Resonance Biochemical Sensor,”
Sensor. Actuat. B-Chem., vol. 249, pp. 542-548, 2017.
[26] N. Vasilantonakis, G. Wurtz, V. Podolskiy and A. Zayats. “Refractive
Index Sensing with Hyperbolic Metamaterials: Strategies for Biosensing
and Nonlinearity Enhancement,” Opt. Express, vol. 23, no. 11, pp.
14329-14343, 2015.
[27] F. Abbas and M. Faryad. “A Highly Sensitive Multiplasmonic Sensor
Using Hyperbolic Chiral Sculptured Thin Films,” J. Appl. Phys., vol. 122,
no. 17, pp. 173104, 2017.
[28] K. V. Sreekanth, Y. Alapan, M. ElKabbash, E. Ilker, M. Hinczewski, U.
A. Gurkan, A. De Luca and G. Strangi. “Extreme Sensitivity Biosensing
Platform Based on Hyperbolic Metamaterials,” Nat. Mater., vol. 15, no.
6, pp. 621, 2016.
[29] G. D. Cole, W. Zhang, M. J. Martin, J. Ye and M. Aspelmeyer. “Tenfold
Reduction of Brownian Noise in High-Reflectivity Optical Coatings,”
Nat. Photonics, vol. 7, no. 8, pp. 644-650, 2013.
[30] G. D. Cole, W. Zhang, B. J. Bjork, D. Follman, P. Heu, C. Deutsch, L.
Sonderhouse, J. Robinson, C. Franz and A. Alexandrovski. “High-
Performance near- and Mid-Infrared Crystalline Coatings,” Optica, vol.
3, no. 6, 2016.
[31] S. Hayashi, H. Kurokawa and H. Oga. “Observation of Resonant Photon
Tunneling in Photonic Double Barrier Structures,” Opt. Rev., vol. 6, no.
3, pp. 204-210, 1999.
[32] G. M. Gehring, A. C. Liapis, S. G. Lukishova and R. W. Boyd. “Time-
Domain Measurements of Reflection Delay in Frustrated Total Internal
Reflection,” Phys. Rev. Lett., vol. 111, no. 3, pp. 030404, 2013.
[33] A. Q. Jian and X. M. Zhang. “Resonant Optical Tunneling Effect: Recent
Progress in Modeling and Applications,” IEEE J. Quantum Elect., vol.
19, no. 3, pp. 9000310-9000310, 2013.
[34] A. C. Liapis, G. M. Gehring, S. G. Lukishova and R. W. Boyd.
“Simulating Quantum-Mechanical Barrier Tunneling Phenomena with a
Nematic-Liquid-Crystal-Filled Double-Prism Structure,” Mol. Cryst. Liq.
Cryst., vol. 595, no. 1, pp. 136-143, 2014.
-
0733-8724 (c) 2018 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission. See http://www.ieee.org/publications_standards/publications/rights/index.html for more information.
This article has been accepted for publication in a future issue of this journal, but has not been fully edited. Content may change prior to final publication. Citation information: DOI 10.1109/JLT.2019.2907786, Journal ofLightwave Technology
[35] S. Longhi. “Quantum‐Optical Analogies Using Photonic Structures,”
Laser Photonics Rev., vol. 3, no. 3, pp. 243-261, 2009.
[36] A. C. Liapis, L. J. Bissell and R. W. Boyd. “Single-Photon Experiments
with Liquid Crystals for Quantum Science and Quantum Engineering
Applications,” SPIE OPTO, pp. 111-129, 2015.
[37] A. Jian, G. Bai, Y. Cui, C. Wei, X. Liu, Q. Zhang, S. Sang and X. Zhang.
“Optical and Quantum Models of Resonant Optical Tunneling Effect,”
Opt. Commun., vol. 428, pp. 191-199, 2018.
[38] Y. J. Chang. “On the Lasing-Like Transmission Via Radiation-Mode-
Enabled Te Resonant Optical Tunneling in Asymmetric, Passive,
Layered Media with Metal,” Opt. Express, vol. 22, no. 23, pp. 28941-
28953, 2014.
[39] W. Li. “Resonant Tunneling Condition and Transmission Periodic
Characteristics for a Metal Barrier in the Fabry-Perot Cavity,” Mater.
Res. Express, vol. 3, no. 12, pp. 126201, 2016.
[40] A. Q. Jian, X. M. Zhang, W. M. Zhu and M. Yu. “Optofluidic
Refractometer Using Resonant Optical Tunneling Effect,”
Biomicrofluidics, vol. 4, no. 4, pp. 43008, 2010.
[41] N. Yamamoto and N. Ohtani. “All-Optical Switching and Memorizing
Devices Using Resonant Photon Tunneling Effect in Multi-Layered
Gaas/Algaas Structures,” JPN. J. Appl. Phys., vol. 43, no. 4A, pp. 1393-
1397, 2004.
[42] W. M. Zhu, T. Zhong, A. Q. Liu, X. M. Zhang and M. Yu.
“Micromachined Optical Well Structure for Thermo-Optic Switching,”
Appl. Phys. Lett., vol. 91, no. 26, pp. 257, 2007.
[43] A. Jian, C. Wei, L. Guo, J. Hu, J. Tang, J. Liu, X. Zhang and S. Sang.
“Theoretical Analysis of an Optical Accelerometer Based on Resonant
Optical Tunneling Effect,” Sensors, vol. 17, no. 2, pp. 389, 2017.
[44] L. Fei, M. K. Wong, S. K. Chiu, L. Hao, J. C. Ho and S. W. Pang.
“Effects of Nanoparticle Size and Cell Type on High Sensitivity Cell
Detection Using a Localized Surface Plasmon Resonance Biosensor,”
Biosens. Bioelectron, vol. 55, no. 9, pp. 141-148, 2014.
[45] L. Brekhovskikh. “Waves in Layered Media,” vol. 16, Elsevier, 2012.
[46] R. Gansch, S. Kalchmair, P. Genevet, T. Zederbauer, H. Detz, A. M.
Andrews, W. Schrenk, F. Capasso, Lon, M. Ccaronar and G. Strasser.
“Measurement of Bound States in the Continuum by a Detector
Embedded in a Photonic Crystal,” Light-Sci. Appl., vol. 5, no. 9, pp.
e16147, 2016.
[47] J. M. Zhu, Y. Shi, X. Q. Zhu, Y. Yang, F. H. Jiang, C. J. Sun, W. H. Zhao
and X. T. Han. “Optofluidic Marine Phosphate Detection with Enhanced
Absorption Using a Fabry-Pérot Resonator,” Lab Chip, 2017.
[48] D. K. C. Wu, B. T. Kuhlmey and B. J. Eggleton. “Ultrasensitive Photonic
Crystal Fiber Refractive Index Sensor,” Opt. Lett., vol. 34, no. 3, pp. 322-
324, 2009.
[49] S. Zhu, H. Li, M. Yang and S. W. Pang. “High Sensitivity Plasmonic
Biosensor Based on Nanoimprinted Quasi 3d Nanosquares for Cell
Detection,” Nanotechnology, vol. 27, no. 29, pp. 295101, 2016.