HD1 STRUCT GE Incubator 2019 09 0… · 6.air entrainment:astm c260 and astm c494, entrain 6%...
Transcript of HD1 STRUCT GE Incubator 2019 09 0… · 6.air entrainment:astm c260 and astm c494, entrain 6%...
A
B
1 2
C
D
3 4 5 COPYRIGHT DESIGN WEST ARCHITECTSc
ISSUED:
CHECKED BY:
DRAWN BY:
PROJECT #:
MARK
:DA
TE:
DESC
RIPT
ION:
795 N
ORTH
400 W
EST
SAL
T LA
KE C
ITY
UT 84
103
255 S
OUTH
300
WES
T L
OGAN
UT
8432
1
GE H
EALT
HCAR
E - I
NCUB
ATOR
925 W
EST
1800
SOU
THLO
GAN,
UTA
H 84
321
SEP. 2019
CLR
419591
DLB
S
STE
TA
UTRS CT
R P
UR
F OE
G
OF UT HA
ERE
NI
NISO
L
EN
A LA
No. 295553RASMUSSENCRAIG L.
9/6/2019
COLD FORM STEEL PRODUCT INFORMATION (SSMA DESIGNATION)
THE FIRST SET OF NUMBERS INDICATES THE MEMBER DEPTH IN 1/100 INCHES.
THE NEXT LETTER INDICATES THE PRIMARY FUNCTION:
S = STUD OR JOIST SECTION.
T = TRACK SECTION
U = CHANNEL SECTIONS.
F = FURRING CHANNEL SECTIONS.
THE SECOND SET OF NUMBERS INDICATES THE FLANGE WIDTH IN 1/100 INCHES.
THE LAST TWO NUMBERS INDICATE THE GAUGE OF STEEL:
33 MIL - 20 GAUGE
43 MIL - 18 GAUGE
54 MIL - 16 GAUGE
68 MIL - 14 GAUGE
97 MIL - 12 GAUGE
118 MIL - 10 GAUGE
STUDS
USE THREE (3) STUDS AT THE CORNER OF ALL EXTERIOR WALLS.
USE THREE (3) STUDS AT THE INTERSECTION OF ALL LOAD BEARING WALLS (EXTERIOR
AND/OR INTERIOR)
ENDS OF STUDS SHOULD SEAT FIRMLY IN RUNNER TRACK WITCH MUST HAVE FULL BEARING
ON STRUCTURE.
ATTACH EACH RUNNER TRACK LEG TO EACH STUD FLANGE WITH ONE (1) #10-16 SCREW, 5/8"
LONG.
NO NOTCHING OR COPING OF STUDS IS ALLOWED.
ALL BRACING SHALL BE INSTALLED AT THE TIME THE WALL IS ERECTED. FAILURE TO INSTALL
BRACING AT THIS TIME MAY COMPROMISE THE STRUCTURAL INTEGRITY OF THE BUILDING.
LATERAL BRACING SHALL BE PROVIDED AT 4'-0" O.C. (MAX.) FOR ALL LOAD BEARING WALLS.
BRACING SHALL BE PROVIDED FOR JOISTS AT 8'-0" O.C. (MAX.).
ATTACHMENTS
USE #10-16, 5/8" LONG, LOW PROFILE HEAD SCREWS FOR ALL STEEL TO STEEL CONNECTIONS
EXCEPT AS NOTED ON PLANS AND DETAILS.
A ¾" (MINIMUM) CLEARANCE MUST BE MAINTAINED FROM ALL EDGES OF THE STEEL MEMBERS
IN LOCATING SCREWS. A ¾" (MINIMUM) ON CENTER SPACING MUST BE MAINTAINED BETWEEN
ADJACENT SCREWS.
ALL SCREW RECOMMENDATIONS ARE BASED ON CCFSS TECHNICAL BULLETIN VOL. 2, NO. 1
WHICH OUTLINES THE PROPOSED AISI SPECIFICATION PROVISIONS FOR SCREW
CONNECTIONS, AND ALLOWABLE SCREW SHEAR STRENGTH VALUES PUBLISHED BY BUILDEX,
COMPASS AND GRABBER.
FOR ATTACHMENT OF SINGLE - LAYER ½" OR 5/8" GYPSUM BOARD OR SHEATHING TO STEEL
STUDS AND JOISTS, USE #10 - 16 SCREWS, 1" LONG BUGLE HEAD SCREWS.
WELDING PROCEDURE TO BE IN ACCORDANCE WITH AWS D1.3 - 81. USE 1/8" E60XX
ELECTRODE MINIMUM. MINIMUM WELD THROAT THICKNESS (T) MUST MATCH OR EXCEED THE
THICKNESS OF THE THINNEST CONNECTED SHEET.
IN WELDING, THE ZINC COATING ON STEEL FRAMING WILL BE BURNED AWAY; THEREFORE, A
ZINC RICH PAINT MUST BE APPLIED TO THE WELD AREA TO PROVIDE CORROSION
RESISTANCE.
MISCELLANEOUS
ALL CONNECTIONS SHALL BE COMPLETE AS PER THE PLANS AND SPECIFICATIONS AT THE
TIME OF INSTALLATION. FAILURE TO PROMPTLY COMPLETE CONNECTIONS MAY
COMPROMISE THE STRUCTURAL INTEGRITY OF THE BUILDING.
PRECAUTIONS MUST BE TAKEN TO AVOID CONSTRUCTION LOADS EXCEEDING DESIGN LIVE
LOADS. CONSTRUCTION LOADS HAVE NOT BEEN CONSIDERED IN THESE RECOMMENDATIONS.
SPLICING OF RUNNER TRACK TO BE ACCOMPLISHED BY SCREW ATTACHMENT OF A STUD
SECTION BETWEEN VERTICAL STUDS TO THE RUNNER TRACK USING #10 - 16 SCREWS, 5/8"
LONG LOW PROFILE SCREWS.
1. ALL DESIGN, CONSTRUCTION, AND INSPECTION SHALL BE IN CONFORMANCE WITH THE 2018
INTERNATIONAL BUILDING CODE (IBC) AND REFERENCED STANDARDS.
2. THE CONTRACTOR SHALL VERIFY ALL DIMENSIONS AND CONDITIONS AT THE SITE.
3. ALL OMISSIONS OR CONFLICTS BETWEEN THE VARIOUS ELEMENTS OF THE WORKING
DRAWINGS AND/OR SPECIFICATIONS SHALL BE BROUGHT TO THE ATTENTION OF THE
ENGINEER BEFORE PROCEEDING WITH ANY WORK INVOLVED.
4. DRAWINGS INDICATE THE FINISHED PRODUCT. THEY DO NOT INDICATE A METHOD OF
CONSTRUCTION. CONTRACTOR SHALL TAKE ALL PRECAUTIONS NECESSARY TO PROTECT
THE STRUCTURE DURING CONSTRUCTION. SUCH PRECAUTIONS SHALL INCLUDE, BUT NOT BE
LIMITED TO, BRACING, SHORING FOR CONSTRUCTION EQUIPMENT, ETC..
5. THE CONTRACTOR SHALL BE RESPONSIBLE FOR COMPENSATING THE OWNER FOR ANY
CHANGES MADE AS A RESULT OF A DEVIATION FROM THE CONTRACT DOCUMENTS,
DEVIATION FROM THE SPECIFICATIONS, FAULTY MATERIALS, OR FAULTY WORKMANSHIP.
6. OPTIONS ARE FOR THE CONTRACTOR'S CONVENIENCE. THE CONTRACTOR SHALL BE
RESPONSIBLE FOR COORDINATING ALL REQUIRED DESIGN CHANGES. COST ASSOCIATED
WITH ANY DESIGN WORK INITIATED BY THE OPTION SHALL BE BORN BY THE CONTRACTOR.
7. CONTRACTOR SHALL BE RESPONSIBLE FOR SAFETY AND PROTECTION WITHIN AND
ADJACENT TO THE JOB SITE.
8. TEMPORARY SHORING AND BRACING SHALL BE PROVIDED WHEREVER NECESSARY TO TAKE
CARE OF ALL LOADS TO WHICH THE STRUCTURE MAY BE SUBJECTED INCLUDING WIND. SUCH
BRACING SHALL BE LEFT IN PLACE AS LONG AS MAY BE REQUIRED FOR SAFETY OR UNTIL
ALL THE STRUCTURAL ELEMENTS ARE COMPLETE.
9. DURING AND AFTER CONSTRUCTION THE CONTRACTOR AND/OR OWNER SHALL KEEP LOADS
ON THE STRUCTURE WITHIN THE LIMITS OF THE DESIGN LOADS.
10. THE GENERAL CONTRACTOR SHALL HAVE SHOP DRAWINGS REVIEWED BY THE ENGINEER
PRIOR TO THE FABRICATION OR ERECTION FOR THE FOLLOWING ITEMS: REINFORCING
STEEL, STRUCTURAL STEEL, AND MISCELLANEOUS METALS.
11. ALL DETAILS, SECTIONS, AND NOTES ARE INTENDED TO BE TYPICAL AND SHALL APPLY TO
SIMILAR SITUATIONS UNLESS NOTED OR SHOWN OTHERWISE.
12. OBSERVATION VISITS TO THE JOB SITE BY THE OWNER, ENGINEER OR FIELD
REPRESENTATIVES OF THE ENGINEER SHALL NEITHER BE CONSTRUED AS INSPECTION NOR
APPROVAL OF CONSTRUCTION.
13. SIZES, LOCATIONS, AND ANCHORAGE'S OF EQUIPMENT SHALL BE VERIFIED IN THE FIELD WITH
EQUIPMENT MANUFACTURERS (SUPPLIERS) PRIOR TO PLACING CONCRETE OR FABRICATING
STEEL.
1. FOOTING ELEVATIONS OR FROST COVER SHOWN ON PLAN ARE TOP OF FOOTINGS AND ARE
MINIMUM DEPTH. DIFFERENT OR UNUSUAL CONDITIONS SHALL BE REPORTED TO THE
ENGINEER.
2. EXTERIOR WALL FOOTINGS SHALL BEAR AT A MINIMUM DEPTH OF 30" BELOW FINISHED
EXTERIOR GRADE.
3. NO FOOTINGS SHALL BE PLACED IN WATER OR ON FROZEN GROUND.
4. ANY SOIL CONDITION ENCOUNTERED DURING EXCAVATION THAT IS CONTRARY TO THE
CONDITIONS USED FOR DESIGN OF FOOTINGS, OR ON THE DRAWINGS SHALL BE BROUGHT
TO THE ATTENTION OF THE ENGINEER BEFORE PROCEEDING.
5. WHERE 6" DIAMETER OR LARGER PIPE PASSES THROUGH AN INTERIOR OR EXTERIOR
FOUNDATION WALL, STEP THE FOOTING DOWN TO PASS BELOW PIPE AND THEN STEP BACK
UP TO INDICATED ELEVATION. PROVIDE PIPE SLEEVE THROUGH FOUNDATION WALL. PLACE
BOTTOM OF PIPE SLEEVE 12" ABOVE TOP OF FOOTING.
7. SUBGRADE PREPARATION AS PER RECOMMENDATIONS OF THE GEOTECHNICAL REPORT
PREPARED BY GHS GEOTECHNICAL, INC. DATED APRIL 30, 2019
1. ALL CONCRETE SHALL MEET THE REQUIREMENTS OF ACI-301, “SPECIFICATION FOR
STRUCTURAL CONCRETE FOR BUILDINGS.” PROPORTIONING OF INGREDIENTS FOR EACH
CONCRETE MIX SHALL BE BY METHOD 2 OR THE ALTERNATE PROCEDURE GIVEN IN ACI-301.
PLACE CONCRETE PER ACI-304 AND CONFORM TO ACI-604 (306) FOR COLD WEATHER
PLACEMENT AND ACI-605 (305) FOR HOT WEATHER PLACEMENT, USE INTERIOR MECHANICAL
VIBRATORS WITH 7,000 RPM MINIMUM FREQUENCY. DO NOT OVER-VIBRATE. CONCRETE SHALL
BE PLACED MONOLITHICALLY BETWEEN CONSTRUCTION AND CONTROL JOINTS. PROTECT ALL
CONCRETE FROM PREMATURE DRYING, EXCESSIVE HOT OR COLD TEMPERATURE FOR SEVEN
DAYS AFTER PLACING.
2. STRENGTH
TWENTY-EIGHT DAY COMPRESSIVE STRENGTH SHALL BE:
4500 PSI, 5 ½ SACK
SLUMP: 4” + 1 INCH.
MAX. WATER/CEMENT RATIO: 0.45
3. STRUCTURAL CONCRETE EXPOSURE CLASS: F2
4. MATERIALS
CEMENT: ASTM 150, TYPE I.
COARSE AND FINE AGGREGATE: ASTM C33.
WATER SHALL BE CLEAN AND POTABLE.
5. ADMIXTURES
WATER REDUCING ADMIXTURE: ASTM C494, ADMIXTURES SHALL BE USED IN EXACT
ACCORDANCE WITH MANUFACTURER'S INSTRUCTIONS.
SYNERGIZED PERFORMANCE SYSTEMS: CONCRETE USING ADMIXTURES TO
PRODUCE FLOWABLE CONCRETE MAY BE USED SUBJECT TO ENGINEER'S APPROVAL.
6. AIR ENTRAINMENT:ASTM C260 AND ASTM C494, ENTRAIN 6% PLUS/MINUS 1 1/2% BY VOLUME IN
ALL EXPOSED CONCRETE.
7. NO OTHER ADMIXTURE PERMITTED UNLESS APPROVED BY THE ENGINEER OF RECORD.
8. THE FLATNESS AND LEVELNESS OF CONCRETE FLOORS SHALL BE MEASURED PER ASTM E
1155:STANDARD TEST METHOD FOR DETERMINING FLOOR FLATNESS AND LEVELNESS USING
THE F-NUMBER SYSTEM.
SPECIFIED OVERALL F-NUMBER FF 25/FL 20
MINIMUM LOCAL F-NUMBER FF 15/FL 12
9. A STATEMENT OF MIX DESIGN FOR ALL CONCRETE SHALL BE SUBMITTED TO AND REVIEWED BY
THE ENGINEER PRIOR TO COMMENCING WORK.
10. ALL CONCRETE WORK SHALL BE PLACED, CURED, STRIPPED, AND PROTECTED AS DIRECTED
BY THE SPECIFICATIONS AND ACI STANDARDS AND PRACTICES.
11. DOWEL VERTICAL BARS THE DEVELOPMENT LENGTH INTO STRUCTURE ABOVE AND FOOTINGS
BELOW. PROVIDE 90 DEGREE HOOK WHERE DEVELOPMENT LENGTH IS NOT POSSIBLE. LAP
ALL REINF. IN CONCRETE THE LISTED LENGTH U.N.O.
12. BEFORE CONCRETE IS POURED CHECK WITH ALL TRADES TO ENSURE PROPER PLACEMENT OF
ALL OPENINGS, SLEEVES, CURBS, CONDUITS, BOLTS, INSERTS, ETC. RELATIVE TO WORK.
13. ADD 2-#5 BARS MINIMUM AROUND ALL OPENINGS (UNLESS OTHERWISE NOTED) AND EXTEND
24" BEYOND CORNER OF OPENING.
14. ALL SLABS ON GRADE SHALL BE PLACED IN ALTERNATE PANELS WITH A MAXIMUM WIDTH OF
20' BETWEEN CONTROL OR CONSTRUCTION JOINTS; REFER TO TYPICAL DETAILS ON
DRAWINGS. UNLESS OTHERWISE NOTED, SLABS ON GRADE SHALL BE 4" THICK AND SHALL BE
REINFORCED WITH #3 BARS AT 18" ON CENTER EACH WAY.
15. PROVIDE A #4 NOSING BAR IN ALL CONCRETE STAIR TREADS.
16. REFER TO DRAWINGS FOR TYPICAL CONSTRUCTION JOINT DETAILS. UNLESS NOTED IN
DRAWINGS, ALL REINFORCEMENT SHALL BE CONTINUOUS THROUGH JOINTS.
17. WHERE EXTERIOR SLABS ON GRADE ABUT FOUNDATIONS OR COLUMNS PROVIDE 3/8"
PREFORMED EXPANSION JOINT WITH SEALANT.
1. ALL REINFORCEMENT SHALL BE DETAILED AND PLACED IN ACCORDANCE WITH BP-66(04): ACI
DETAILING MANUAL - 2004 AND ACI 350-06.
2. REINFORCING STEEL SHALL BE ASTM A615 GRADE 60. WELDED WIRE FABRIC SHALL
CONFORM TO ASTM A1064.
3. ALL REINFORCEMENT SHALL BE SECURELY TIED AND HELD IN PLACE.
4. REINFORCING BARS THAT ARE TO BE WELDED, INCLUDING DEFORMED BAR ANCHORS (D.B.A.)
SHALL COMPLY WITH ASTM A706 OR ANOTHER APPROVED WELDABLE GRADE AND SHALL BE
WELDED IN ACCORDANCE WITH THE A.W.S. RECOMMENDATIONS.
5. ALL CONTINUOUS REINFORCEMENT SHALL TERMINATE WITH A 90 DEG. TURN OR A SEPARATE
CORNER BAR. ALL SPLICES IN CONCRETE SHALL LAP THE LISTED LAP LENGTH.
6. THE FOLLOWING CONCRETE COVER SHALL BE PROVIDED FOR REINFORCEMENT:
A. CONCRETE CAST AGAINST AND PERMANENTLY EXPOSED TO EARTH: 3"
B. ALL OTHER CONCRETE:2"
7. PRIOR TO FABRICATION AND PLACEMENT, SHOP DRAWINGS FOR ALL REINFORCING STEEL
SHALL BE REVIEWED BY THE ENGINEER.
8. REFER TO WALL CORNER AND WALL INTERSECTION REINFORCING DETAIL. IN GENERAL, THE
WALL CORNER REINFORCING SIZES AND SPACING SHALL BE CALLED OUT ON THE PLANS
AND REFERENCED TO THESE DETAILS AND THE TYPICAL HORIZONTAL WALL REINFORCING
SHALL LAP WITH THE HORIZONTAL REINFORCING.
9. ALL BENDS, UNLESS OTHERWISE SHOWN, SHALL BE A 90 DEGREE STANDARD HOOK. REFER
TO STANDARD CONCRETE HOOK DETAILS.
10. ALL WALL CORNER AND WALL INTERSECTION REINFORCEMENT BARS SHALL BE CONTINUOUS
AROUND CORNERS AND THROUGH COLUMNS OR PILASTERS. REINFORCEMENT SHALL BE
EXTENDED INTO CONNECTING WALLS AND LAPPED ON THE OPPOSITE FACE OF THE
CONNECTING WALLS.
11. UNLESS INDICATED OTHERWISE, CONTRACTOR MAY SPLICE CONTINUOUS SLAB OR
LONGITUDINAL BEAM BARS AT LOCATION OF THEIR CHOOSING, EXCEPT THAT TOP BAR
SPLICES SHALL BE LOCATED AT MIDSPAN AND BOTTOM BAR SPLICES SHALL BE LOCATED AT
SUPPORTS. STAGGER SPLICES IN HORIZONTAL WALL BARS SO THAT NO TWO ADJACENT
BARS IN THE SAME OR OPPOSITE CURTAIN ARE SPLICED AT THE SAME LOCATION. ALL
REINFORCEMENT BENDS AND LAPS, UNLESS OTHERWISE NOTED, SHALL SATISFY THE
REQUIREMENTS OF THE STD. CONCRETE HOOK SCHEDULE AND THE CONCRETE
REINFORCEMENT LAP AND DEVELOPMENT SCHEDULES.
THE FOLLOWING STRUCTURAL DESIGN LOADS APPLY U.N.O.:
ROOF LIVE LOAD . . . . . . . . . . . . . . . . . . . . . . . . . L
r
= 20 PSF
ROOF DEAD LOAD . . . . . . . . . . . . . . . . . . . . . . . . .D
r
= 25 PSF
CEILING DEAD LOAD . . . . . . . . . . . . . . . . . . . . . . .D = 10 PSF
CEILING LIVE LOAD . . . . . . . . . . . . . . . . . . . . . . . .D = 10 PSF
STRUCTURE RISK CATEGORY . . . . . . . . . . . . . . II
SNOW LOAD:
GROUND SNOW LOAD. . . . . . . . . . . . . . . . P
g
= 42 PSF
SNOW IMPORTANCE FACTOR. . . . . . . . . I
s
= 1.00
SNOW EXPOSURE FACTOR. . . . . . . . . . . C
e
= 1.00
SNOW THERMAL FACTOR . . . . . . . . . . C
t
= 1.00
FLAT ROOF SNOW LOAD. . . . . . . . . . . . . P
f
= 30 PSF
SLOPED ROOF SNOW LOAD . . . . . . . . . . P
s
= 30 PSF
WIND:
BASIC WIND SPEED . . . . . . . . . . . . . . . . . . V = 105 MPH
WIND IMPORTANCE FACTOR. . . . . . . . . . . Iw = 1.00
WIND EXPOSURE. . . . . . . . . . . . . . . . . . . . C
SEISMIC:
SEISMIC IMPORTANCE FACTOR. . . . . . . . . . . . I
E
= 1.00
SPECTRAL RESPONSE ACCELERATION. . . . . S
S
= 1.093
SPECTRAL RESPONSE ACCELERATION. . . . . S
1
= 0.368
SEISMIC SOIL SITE CLASS . . . . . . . . . . . . . . . . D
SPECTRAL RESPONSE COEFFICIENT . . . . . . S
DS
= 0.776
SPECTRAL RESPONSE COEFFICIENT . . . . . . S
D1
= 0.613
SEISMIC DESIGN CATEGORY . . . . . . . . . . . . . D
ANALYSIS PROCEDURE USED: EQUIVALENT LATERAL FORCE
BASIC SEISMIC FORCE RESISTING SYSTEM:
LIGHT-FRAME (COLD-FORMED STEEL) WALLS w/ WOOD STRUCTURAL PANELS
RESPONSE MODIFICATION COEFFICIENT. . . . . R = 6.5
OVERSTRENGTH FACTOR . . . . . . . . . . . . . . . . . . Ω
0
= 2.5
DESIGN LOAD-BEARING VALUE OF SOILS = 2000 PSF
GEOTECHNICAL REPORT PREPARED BY GHS GEOTECHNICAL, INC. DATED APRIL 30, 2019
GENERAL CONCRETE
REINFORCING STEEL
FOOTINGS
DESIGN CRITERIA
PLYWOOD DIAPHRAGM FASTENERS
1. SUBSTITUTION OF FASTENERS OTHER THAN THOSE SPECIFIED IN THE STRUCTURAL
NOTES OR THE PLYWOOD DIAPHRAGM SCHEDULE IS NOT ALLOWED.
2. AT ROOF AND FLOOR DIAPHRAGMS, APPLY FASTENERS AT PANEL EDGE SPACING TO
ALL DRAG STRUTS, TENSION CHORDS, BLOCKING OVER BEARING AND SHEAR WALLS,
AND ANY OTHER SPECIAL DIAPHRAGM MEMBERS NOTED ON PLANS.
3. AT PLYWOOD SHEAR WALLS, APPLY FASTENERS AT PANEL EDGE SPACING TO TOP
AND BOTTOM TRACKS, END CHORDS, ALL VERTICAL ELEMENTS AT HOLD DOWN
ANCHORS, AND HORIZONTAL BLOCKING.
4. ALL PANEL EDGES OF PLYWOOD SHEAR WALLS MUST BE BLOCKED.
5. FASTENERS TO COMPLY WITH ASTM C1513.
1. FOLLOW RECOMMENDED PRACTICE FOR CONCRETE FORMWORK (ACI-347-14).
2. ALL SHORING SHALL BE THE RESPONSIBILITY OF THE CONTRACTOR. FORMWORK SUPPORTS
AND SHORING SHALL BE DESIGNED TO PROVIDE FINISHED CONCRETE SURFACES OF ALL
FACES LEVEL, PLUMB, AND TRUE TO THE DIMENSIONS AND ELEVATIONS SHOWN.
TOLERANCES AND VARIATIONS SHALL BE AS SPECIFIED.
FORM WORK
1. EXCEPT WHERE INDICATED ON THE DRAWINGS OR AS OTHERWISE APPROVED,
POST-INSTALLED ANCHORS SHALL CONSIST OF THE FOLLOWING ANCHOR TYPES AS
PROVIDED BY HILTI, INC. CONTACT HILTI AT (800) 879-8000 FOR PRODUCT RELATED
QUESTIONS.
a) ANCHORAGE TO CONCRETE
i) ADHESIVE ANCHORS FOR CRACKED AND UNCRACKED CONCRETE USE:
(1) HILTI HIT-HY 200 SAFE SET SYSTEM WITH HILTI HIT-Z ROD PER ICC ESR-3187.
(2) HILTI HIT-HY 200 SAFE SET SYSTEM WITH HILTI HOLLOW DRILL BIT SYSTEM WITH
HAS-E THREADED ROD PER ICC ESR-3187.
(3) HILTI HIT-RE 500-SD SAFE SET EPOXY ADHESIVE ANCHORING SYSTEM WITH
HAS-E THREADED ROD PER ICC ESR-2322 FOR SLOW CURE APPLICATIONS
POST-INSTALLED ANCHORS
COLD FORM STEEL
STRUCTURAL
NOTES
S-000
TYPICAL SINGLE MAT CORNER AND
INTERSECTION REINFORCING
"D"
NOTE 3NOTE 3
NO
TE
3
"D
"
NO
TE
3
"D"
STD ACI-318 90° HOOK
UNLESS OTHERWISE
NOTED
HORIZ WALL REINFORCING BETWEEN CORNERS AND
INTERSECTIONS AS SHOWN ON WALL SECTIONS. LAP
WITH CORNER AND INTERSECTION REINFORCING
LAP, SEE NOTE 5,
TYP
A
B
1 2
C
D
3 4 5 COPYRIGHT DESIGN WEST ARCHITECTSc
ISSUED:
CHECKED BY:
DRAWN BY:
PROJECT #:
MARK
:DA
TE:
DESC
RIPT
ION:
795 N
ORTH
400 W
EST
SAL
T LA
KE C
ITY
UT 84
103
255 S
OUTH
300
WES
T L
OGAN
UT
8432
1
GE H
EALT
HCAR
E - I
NCUB
ATOR
925 W
EST
1800
SOU
THLO
GAN,
UTA
H 84
321
SEP. 2019
CLR
419591
DLB
S
STE
TA
UTRS CT
R P
UR
F OE
G
OF UT HA
ERE
NI
NISO
L
EN
A LA
No. 295553RASMUSSENCRAIG L.
9/6/2019
DIAPHRAGM SCHEDULE
ALL GRADES OF STEEL (MIN. YIELD STRENGTHS)
D = FINISHED INSIDE BEND DIAMETER
d = BAR DIAMETER
D = 6d FOR #3 THROUGH #8
D = 8d FOR #9, #10 AND #11
D = 10d FOR #14 AND #18
d
B
4d or
2
1
2
" MIN.
D
HOOK A
DETAILING
DIMENSION
180° HOOK
d
12d
A
D
DETAILING
DIMENSION
90° HOOK
RECOMMENDED END HOOKS
BAR
SIZE
D
180° HOOKS
A B
90° HOOKS
A
#3 2
1
4
" 5" 3" 6"
#4 3" 6" 4" 8"
NOTES:
1. ALL REINFORCEMENT LAPS, UNLESS OTHERWISE NOTED, SHALL SATISFY THE MINIMUM
REQUIREMENTS SHOWN IN THE TABLE "REINFORCEMENT LAP LENGTHS."
2. TOP BARS SHALL BE DEFINED AS ANY HORIZONTAL BARS PLACED SUCH THAT MORE THAN 12"
OF FRESH CONCRETE IS CAST IN THE MEMBER BELOW THE BAR, IN ANY SINGLE POUR.
HORIZONTAL WALL BARS ARE CONSIDERED TOP BARS.
CONCRETE REINFORCEMENT LAP & DEVELOPMENT SCHEDULES
NTS
DIAPHRAGM LAYOUT
TYPICAL PLYWOOD ROOF
NOTES:
JOISTS AND TRACKS ARE TO BE ASTMA1003 STRUCTURAL GRADE 33 (GRADE 230) TYPE H STEEL FOR
MEMBERS WITH A DESIGNATION THICKNESS OF 33 AND 43 MILS, AND ASTM A1003 STRUCTURAL GRADE
50 (GRADE 340) TYPE H STEEL FOR MEMBERS WITH A DESIGNATION THICKNESS EQUAL TO OR GREATER
THAN 54 MILS.
SCREWS USED TO ATTACH WOOD STRUCTURAL PANELS ARE MINIMUM NO. 8 WHERE STRUCTURAL
MEMBERS HAVE A DESIGNATION THICKNESS OF 54 MILS OR LESS AND NO. 10 WHERE STRUCTURAL
MEMBERS HAVE A DESIGNATION THICKNESS GREATER THAN 54 MILS. SCREWS SHALL COMPLY WITH
ASTM C1513
TRACK MEMBERS TO HAVE A MINIMUM FLANGE WIDTH OF 1 1/4"
PANELS LESS THAN 12" WIDE ARE NOT PERMITTED.
FASTENERS ALONG THE EDGES IN SHEAR PANELS ARE PLACE FROM PANEL EDGES NOT LESS THAN 3/8".
MINIMUM
NOMINAL
PANEL
THICKNESS
PANEL
BOUNDARY
EDGE
OTHER
PANEL
EDGES
BLOCKING
DIAPHRAGM
HOOK EMBED
DEPTH
7"
9"
#5 3
3
4
" 7" 5" 10"
#6 4
1
2
" 8" 6" 1'-0"
1'-7"11
3
4
"1'-3"9
1
2
"#9
1'-4"8"11"6"#8
1'-2"7"10"5
1
4
"#7
1'-10"1'-1
1
4
"1'-5"10
3
4
'#10
#11 12" 1'-7" 1'-2
3
4
" 2'-0"
#14 18
1
4
" 2'-3" 1'-9
3
4
" 2'-7"
#18 24" 3'-0" 2'-4
1
2
" 3'-5"
(IN ACCORDANCE WITH ACI 318-14)
NTS
STD. CONCRETE HOOK SCHEDULE
12"
14"
23"
21"
18"
26"
31"
41"
16"
NOTE: EMBED BARS WITH HOOKS TO THE DEPTH
SHOWN.
SPAN
RATING
PANEL
FIELD
#10 AT 6"
NOTES FOR CORNER AND INTERSECTION REINFORCING DETAILS
1. TYPICAL HORIZONTAL WALL CORNER AND INTERSECTION REINFORCING
LAYOUT IS SHOWN TO AVOID CONGESTION AND PERMIT PROPER
PLACEMENT, FOR SIZE AND SPACING SEE PLANS. ALL HORIZONTAL
REINFORCING AT CORNERS AND INTERSECTIONS SHALL BE
FABRICATED AND INSTALLED WITH SPLICES LOCATED WHERE SHOWN
REGARDLESS OF BAR SIZE AND SPACING.
2. WHERE THE CORNER OR INTERSECTION REINFORCING SIZE AND
SPACING IS NOT SHOWN, NOTED OR TABULATED ON THE PLANS, THE
SIZE AND SPACING SHALL BE THE SAME AS THE WALL HORIZONTAL
REINFORCING SHOWN ON THE WALL SECTIONS OR AS NOTED FOR THE
REINFORCING BETWEEN THE CORNERS OR INTERSECTIONS.
3. EXCEPT WHERE OTHERWISE SHOWN ON THE DRAWINGS, THE LENGTH
INDICATED AS "NOTE 3" SHALL BE THE LESSER OF D/4, 10 FEET, OR 1.0
TIMES THE HEIGHT OF THE WALL, EXCEPT THAT IN NO CASE SHALL IT
BE LESS THAN THE SPLICE LENGTH.
4. D = LENGTH OF WALL PARALLEL TO THE BAR LENGTH IN QUESTION.
5. EXCEPT WHERE OTHERWISE SHOWN ON THE DRAWINGS, THE LENGTH
INDICATED AS "NOTE 5" SHALL BE EQUAL TO ONE "LAP LENGTH" AS
REQUIRED BY THE GENERAL STRUCTURAL NOTES. USE THE LAP
LENGTH AS REQUIRED FOR THE SMALLER OF THE TWO REINFORCING
BARS BEING SPLICED.
6. HORIZONTAL BARS IN WALLS SHALL HAVE SPLICES STAGGERED ONE
LAP LENGTH AS DEFINED IN NOTE 5.
7. ALTERNATE SPLICE LOCATIONS OF HORIZONTAL BARS IN WALLS
BETWEEN THE TWO CURTAINS SO THAT NO SPLICES AT THE SAME
ELEVATION ARE LOCATED TOGETHER.
5'-7"5'-1"4'-6"4'-0"3'-6"2'-5"2'-0"1'-7"1'-3"
7'-3"6'-7"5'-10"5'-2"4'-6"3'-1"2'-7"2'-1"1'-7"
#11#10#9#8#7#6#5#4#3
OTHER BAR
TOP BAR
BAR SIZE
REINFORCEMENT LAP LENGTHS
CONCRETE DESIGN STRENGTH - 4000 PSI
SHEATHING FASTENERS
WD-1 19/32" 40/20NO
MARK
NOTES
DIAPHRAGM
BOUNDARIES (SHEAR
WALLS, TENSION
CHORDS)
CONT. PANEL EDGE
PANEL EDGE INTERMEDIATE SUPPORTS
FRAMING, SEE PLAN
BLOCKING IF REQ'D
GENERAL SHEAR WALL NOTES
WALL STUDS AND TRACK ARE TO BE ASTMA1003 STRUCTURAL GRADE 33 (GRADE 230) TYPE H STEEL FOR MEMBERS
WITH A DESIGNATION THICKNESS OF 33 AND 43 MILS, AND ASTM A1003 STRUCTURAL GRADE 50 (GRADE 340) TYPE H
STEEL FOR MEMBERS WITH A DESIGNATION THICKNESS EQUAL TO OR GREATER THAN 54 MILS.
SCREWS ARE A MINIMUM NO. 8 AND COMPLY WITH ASTM C1513.
TRACK MEMBERS TO HAVE A MINIMUM FLANGE WIDTH OF 1 1/4"
PANELS LESS THAN 12" WIDE ARE NOT PERMITTED.
FASTENERS ALONG THE EDGES IN SHEAR PANELS ARE PLACE FROM PANEL EDGES NOT LESS THAN 3/8".
PROVIDE DOUBLED (BACK TO BACK) STUDS AT SHEAR WALL ENDS.
SHEAR WALL SCHEDULE
1. MAXIMUM SHEAR = 330 P.L.F.
USE 7/16" APA RATED SHEATHING ON ONE SIDE. ATTACH SHEATHING TO STUDS WITH NO. 8 (MINIMUM) SCREWS
AT 6” O.C. ALL EDGES AND 12" FIELD. TRACK SIZE TO MATCH WIDTH AND THICKNESS OF SPECIFIED WALL
STUDS. SEE PLANS FOR LOCATIONS AND SIZES OF HOLDOWNS.
#10 AT 6" #10 AT 12"
STRUCTURAL
NOTES
S-001
1. SPECIAL INSPECTION SHALL BE PROVIDED ACCORDING TO IBC
CHAPTER 17 FOR THE ITEMS IDENTIFIED IN THIS SECTION AND
ON THE CONTRACT DOCUMENTS.
2. THE NAMES AND CREDENTIALS OF SPECIAL INSPECTORS TO BE
USED SHALL BE SUBMITTED TO THE BUILDING DEPARTMENT
WHEN APPLYING FOR A BUILDING PERMIT.
3. SPECIAL INSPECTION REPORTS SHALL BE DELIVERED TO THE
OWNER BI-WEEKLY OR MORE FREQUENTLY AS REQUIRED BY
THE INSPECTOR OR BUILDING OFFICIAL.
4. OFF-SITE FABRICATION: WHERE FABRICATION OF STRUCTURAL
LOAD-BEARING MEMBERS AND ASSEMBLIES IS BEING
PERFORMED ON THE PREMISES OF A FABRICATORS SHOP,
SPECIAL INSPECTION OF THE FABRICATED ITEMS SHALL BE IN
ACCORDANCE WITH IBC SECTION 1704.2.5 UNLESS THE
FABRICATOR IS APPROVED ACCORDING TO IBC SECTION
1704.2.5.1.
5. CONCRETE CONSTRUCTION: SPECIAL INSPECTIONS AND
VERIFICATIONS SHALL BE PROVIDED IN ACCORDANCE WITH
TABLE 1705.3.
6. MASONRY CONSTRUCTION: LEVEL 2 SPECIAL INSPECTION
SHALL BE PROVIDED FOR MASONRY CONSTRUCTION IN
ACCORDANCE WITH SECTION ACI 530-13 TABLE 3.1.2.
7. THE FOLLOWING ELEMENTS OF THE SEISMIC-FORCE-RESISTING
SYSTEM IN ACCORDANCE WITH IBC SECTION 1705.12:
A. STRUCTURAL WOOD (IBC 1705.12.2)
8. SPECIAL INSPECTIONS OF WOOD TRUSSES WITH OVERALL
HEIGHTS OF 60 INCHES OR GREATER SHALL BE PERFORMED TO
VERIFY THAT THE INSTALLATION OF THE PERMANENT
INDIVIDUAL TRUSS MEMBER RESTRAINT/BRACING HAS BEEN
INSTALLED IN ACCORDANCE WITH THE APPROVED TRUSS
SUBMITTAL PACKAGE. FOR WOOD TRUSSES WITH A CLEAR
SPAN OF 60 FEET OR GREATER, THE SPECIAL INSPECTOR
SHALL VERIFY DURING CONSTRUCTION THAT THE TEMPORARY
INSTALLATION RESTRAINT/BRACING IS INSTALLED IN
ACCORDANCE WITH THE APPROVED TRUSS SUBMITTAL
PACKAGE.
9. SOILS: SPECIAL INSPECTION SHALL BE PROVIDED FOR
PLACEMENT OF FILL 12 INCHES OR MORE DEEP IN
ACCORDANCE WITH SECTION 1705.6.
10. EPOXY ANCHORS: PRIOR TO AND DURING EPOXY INJECTION TO
INSURE PROPER INSTALLATION AS PER MANUFACTURERS
REQUIREMENTS. CONTRACTOR SHALL SUBMIT PROPOSED
EPOXY MANUFACTURERS I.C.B.O. REPORT TO ENGINEER PRIOR
TO INSTALLATION.
EACH CONTRACTOR RESPONSIBLE FOR THE CONSTRUCTION OF A
SEISMIC-FORCE-RESISTING SYSTEM, DESIGNATED SEISMIC SYSTEM,
OR COMPONENT LISTED IN THE QUALITY ASSURANCE PLAN SHALL
SUBMIT A WRITTEN CONTRACTOR'S STATEMENT OF RESPONSIBILITY
TO THE BUILDING OFFICIAL AND TO THE OWNER PRIOR TO THE
COMMENCEMENT OF WORK ON THE SYSTEM OR COMPONENT IN
ACCORDANCE WITH IBC SECTION 1704.4. THE CONTRACTOR'S
STATEMENT OF RESPONSIBILITY SHALL CONTAIN THE FOLLOWING:
A. ACKNOWLEDGMENT OF AWARENESS OF THE SPECIAL
REQUIREMENTS CONTAINED IN THE QUALITY ASSURANCE
PLAN.
B. ACKNOWLEDGMENT THAT CONTROL WILL BE EXERCISED TO
OBTAIN CONFORMANCE WITH THE CONSTRUCTION
DOCUMENTS APPROVED BY THE BUILDING OFFICIAL.
C. PROCEDURES FOR EXERCISING CONTROL WITHIN THE
CONTRACTOR'S ORGANIZATION, THE METHOD AND
FREQUENCY OF REPORTING, AND THE DISTRIBUTION OF
REPORTS.
D. IDENTIFICATION AND QUALIFICATIONS OF THE PERSON(S)
EXERCISING SUCH CONTROL AND THE POSITION(S) IN THE
ORGANIZATION.
QUALITY ASSURANCE
CONTRACTOR RESPONSIBILITY
QUALITY ASSURANCE PLAN
VERIFICATION AND INSPECTION
CONTINUOUS DURING
TASK LISTED
PERIODICALLY DURING
TASK LISTED
1. VERIFY MATERIALS BELOW SHALLOW FOUNDATIONS ARE ADEQUATE TO ACHIEVE THE
DESIGN BEARING CAPACITY.
2. VERIFY EXCAVATIONS ARE EXTENDED TO PROPER DEPTH AND HAVE REACHED PROPER
MATERIAL.
3. PERFORM CLASSIFICATION AND TESTING OF COMPACTED FILL MATERIALS.
-- X
4. VERIFY USE OF PROPER MATERIALS, DENSITIES AND LIFT THICKNESSES DURING
PLACEMENT AND COMPACTION OF COMPACTED FILL.
5. PRIOR TO PLACEMENT OF COMPACTED FILL, OBSERVE SUBGRADE AND VERIFY THAT SITE
HAS BEEN PREPARED PROPERLY.
-- X
-- X
X --
-- X
TABLE 1705.6 REQUIRED SPECIAL INSPECTIONS AND TESTS OF SOILS
VERIFICATION AND INSPECTION CONTINUOUS PERIODIC
REFERENCED
STANDARD
IBC
REFERENCE
1. INSPECTION OF REINFORCING STEEL, INCLUDING PRESTRESSING TENDONS, AND
VERIFY PLACEMENT.
3. INSPECT ANCHORS POST-INSTALLED IN HARDENED CONCRETE MEMBERS.
a. ADHESIVE ANCHORS INSTALLED IN HORIZONTALLY OR UPWARDLY INCLINED
ORIENTATIONS TO RESIST SUSTAINED TENSION LOADS.
b. MECHANICAL ANCHORS AND ADHESIVE ANCHORS NOT DEFINED IN 3.a.
4. VERIFYING USE OF REQUIRED DESIGN MIX.
-- X ACI 318 CH. 20,
25.2, 25.3,
26.6.1-26.6.3
1908.4
--X
ACI 318:17.8.2.4
--
-- X
ACI 318: CH. 19,
26.4.3, 26.4.4
1904.1,
1904.2,
1908.2,
1908.3
5. PRIOR TO CONCRETE PLACEMENT, FABRICATE SPECIMENS FOR STRENGTH
TESTS, PERFORM SLUMP AND AIR CONTENT TESTS, AND DETERMINE THE
TEMPERATURE OF THE CONCRETE.
X --
ASTM C172,
ASTM C31,
ACI 318: 26.4, 26.12
1908.10
6. INSPECT CONCRETE AND SHOTCRETE PLACEMENT FOR PROPER APPLICATION
TECHNIQUES.
X --ACI 318: 26.5
1908.6,
1908.7,
1908.8
7. VERIFY MAINTENANCE OF SPECIFIED CURING TEMPERATURE AND TECHNIQUES.
-- X
ACI 318:
26.5.3-26.5.5
1908.9
8. INSPECT FORMWORK FOR SHAPE, LOCATION AND DIMENSIONS OF THE
CONCRETE MEMBER BEING FORMED. -- X
ACI 318:
26.11.1.2(b)
--
TABLE 1705.3 REQUIRED SPECIAL INSPECTIONS AND TESTS OF CONCRETE CONSTRUCTION
2. INSPECT ANCHORS CAST IN CONCRETE.
-- X
ACI 318:17.8.2
--
-- X
ACI 318: 17.8.2
--
VERIFICATION AND INSPECTION
CONTINUOUS DURING
TASK LISTED
PERIODICALLY DURING
TASK LISTED
1. VERIFY APPROPRIATE SHEATHING, FRAMING MEMBERS AT PANEL EDGES AND
FASTENERS ARE USED
2. VERIFY PROPER NAILING, BOLTING, ANCHORING AND OTHER FASTENING OF ELEMENTS
OF LATERAL FORCE RESISTING SYSTEM INCLUDING WOOD SHEAR WALLS, WOOD
DIAPHRAGMS, DRAG STRUTS, BRACES AND HOLD-DOWNS.
-- X
-- X
IBC SECTION 1705.11.1 REQUIRED SPECIAL INSPECTIONS FOR STRUCTURAL WOOD FRAMING
3. VERIFY THAT PROPER BOLTING AND SPACING ARE USED ON TIMBER TRUSSES
-- X
F1
F3
F1F1
F1
4" SLAB AT #3 AT 18"
O.C. EACH WAY AT
CENTER OF SLAB
H
D
1
H
D
1
HD1
HD1 HD1
F2
HD
1
HD
1
FOOTING SCHEDULE
MARK SIZE REINF.
F1 20"x10" PER DETAIL
16"x10" (2) #4 CONT.F2
NOTES
THICKENED SLAB FOOTING
F3 12"x10" (2) #4 CONT.
HOLDOWN SCHEDULE
MARK
HD1
FASTENERSTYPE
S/HDU4 (6) #14 SCREWS
NOTES: INSTALL PER MANUFACTURER'S RECOMMENDATIONS
A.B DIAMETER
5/8"
NOTES: ALL FOOTINGS ARE TO BE F1 U.N.O.
THICKENED SLAB FOOTING
DETAIL 1 SHEET S500
2
1
S500
3
S500
1
S500
1
S500
2
S500
STRUCTURAL KEY NOTES
1. RECESS FOUNDATION WALL AT DOOR OPENING AND POUR SLAB OVER WALL PER
DETAIL 5 SHEET S500.
2. CONCRETE CONTROL JOINT. SEE DETAIL 4 SHEET S-500.
GENERAL SHEET NOTES
1. REFER TO ARCHITECTURAL PLANS FOR ADDITIONAL DIMENSIONS AND ELEVATIONS.
1
A
B
1 2
C
D
3 4 5 COPYRIGHT DESIGN WEST ARCHITECTSc
ISSUED:
CHECKED BY:
DRAWN BY:
PROJECT #:
MARK
:DA
TE:
DESC
RIPT
ION:
795 N
ORTH
400 W
EST
SAL
T LA
KE C
ITY
UT 84
103
255 S
OUTH
300
WES
T L
OGAN
UT
8432
1
GE H
EALT
HCAR
E - I
NCUB
ATOR
925 W
EST
1800
SOU
THLO
GAN,
UTA
H 84
321
SEP. 2019
CLR
419591
DLB
S
STE
TA
UTRS CT
R P
UR
F OE
G
OF UT HA
ERE
NI
NISO
L
EN
A LA
No. 295553RASMUSSENCRAIG L.
9/6/2019
S-100
FOOTING
PLAN
H2
H2
H2
H1
H1
H
D
1
H
D
1
HD1
HD1
HD
1
HD
1
HD1
BW
2B
W2
BW3 BW3
BW
4
BW
4
BW
4
BW
4
B
W
2
BW
4
BW
4
J
1
STRUCTURAL KEY NOTES
1. MECHANICAL UNIT. SEE MECHANICAL PLANS.
2. PROVIDE CEILING JOIST, J2, AT 16" O.C. SEE ARCHITECTURAL PLANS FOR JOIST
ELEVATION.
3. PROVIDE DOUBLE CEILING JOIST, J2, BACK-TO-BACK INTERCONNECTED AT 12" O.C.
PER DETAIL 11 SHEET S-501.
4. PROVIDE SIMPSON MSTC28 w/ (30) #10 SCREWS AT ROOF JOIST TRACK CENTERED
OVER CORNER.
5. PROVIDE WEB STIFFENER FOR ROOF JOIST OVER HEADER.
MEMBERMARK
NOTES:
SEE DETAIL 10 SHEET S501 FOR HEADER ATTACHMENT
COLD FORM HEADER & BEAM SCHEDULE
H2
H1 (2) 1000S200-54
(2) 600S137-33
TYPE
BOXED
BOXED
ADDITIONAL COMMENTS
FY = 33 KSI
FY = 33KSI
MEMBERMARK
NOTES:
INTERCONNECTED AT 12" O.C.
BRACE AT MID-POINT
COLD FORM COLUMN SCHEDULE
C1 (2) 600S162-33
TYPE ADDITIONAL COMMENTS
BACK-TO-BACK FY = 33 KSI
C2 (2) 362S137-43 BACK-TO-BACK FY = 33 KSI
MEMBERMARK
J1
LIGHT GAUGE STEEL JOIST SCHEDULE
1200S125-68 AT 16" O.C.
ADDITIONAL COMMENTS
FY = 50 KSI
NOTES:
J2 600S162-43 AT 16" O.C.FY = 33 KSI
GENERAL SHEET NOTES
1. REFER TO ARCHITECTURAL PLANS FOR ADDITIONAL DIMENSIONS AND ELEVATIONS.
2. ALL EXTERIOR WALLS TO BE WALL TYPE BW1.
3. ROOF DIAPHRAGM TO BE TYPE WD-1. SEE DIAPHRAGM SCHEDULE ON SHEET S-002.
4. STUDS TO ALIGN DIRECTLY BELOW ROOF JOISTS, TYP ALL BEARING WALLS.
1
6
S-500
8
S-500
7
S-500
6
S-500
MEMBERMARK
NOTES:
COLD FORM WALL SCHEDULE
BW1 600S162-33 AT 16" O.C.
ADDITIONAL COMMENTS
FY = 33 KSI
BW2 362S137-43 AT 16" O.C.FY = 33 KSI
BRACE AT MIDPOINT
BRACE AT MIDPOINT
2
3
A
B
1 2
C
D
3 4 5 COPYRIGHT DESIGN WEST ARCHITECTSc
ISSUED:
CHECKED BY:
DRAWN BY:
PROJECT #:
MARK
:DA
TE:
DESC
RIPT
ION:
795 N
ORTH
400 W
EST
SAL
T LA
KE C
ITY
UT 84
103
255 S
OUTH
300
WES
T L
OGAN
UT
8432
1
GE H
EALT
HCAR
E - I
NCUB
ATOR
925 W
EST
1800
SOU
THLO
GAN,
UTA
H 84
321
SEP. 2019
CLR
419591
DLB
S
STE
TA
UTRS CT
R P
UR
F OE
G
OF UT HA
ERE
NI
NISO
L
EN
A LA
No. 295553RASMUSSENCRAIG L.
9/6/2019
BW3 350S162-33 AT 24" O.C.FY = 33 KSI
BRACE AT MIDPOINT
BW4 250S162-33 AT 16" O.C.FY = 33 KSI
BRACE AT MIDPOINT
C3 (2) 350S162-33 BACK-TO-BACK FY = 33 KSI
9
S-500
ROOF
FRAMING
PLAN
S-101
4
5
ROOF JOIST PER PLAN
TYP. WALL STUD AND
SPACING FOR PARAPET WALL
CONTINUOUS SHEATHING
PROVIDE FLAT STRAP w/ BLOCKING AT BREAKS
IN SHEATING. SEE DETAIL 6 SHEET S501
INSULATION PER
ARCHITECTURAL PLANS
SHEATHING PER PLAN
CONTINUOUS TRACK MEMBER. MATCH
SIZE AND THICKNESS OF JOIST
DIAPHRAGM BOUNDARY
FASTENER PER PLAN
CONTINUOUS TRACK AT WALL FOR
JOIST LEDGER. MATCH SIZE AND
THICKNESS OF CEILING JOIST
ARCHITECTURAL EIFS
(2) #10 x 3/4" SCREWS
AT EACH WALL STUD
SIMPSON L30 AT
EACH JOIST
CEILING JOIST PER PLAN
WALL SHEATHING PER PLAN
WALL STUD PER PLAN
ALIGN WALL STUD DIRECTLY
BENEATH ROOF JOIST
SEE DETAIL 6 FOR
WALL ANCHORAGE
ARCHITECTURAL EIFS
CONCRETE SLAB
PER PLAN
FOOTING REINF.
PER PLAN
MIN
F
RO
ST
D
EP
TH
6"
MIN
20"
#4 AT 12" O.C. VERTICAL
DOWEL w/ STD. HOOK
8"
#4 AT 12" O.C.
HORIZ. REINF.
#4 AT 8" O.C.
TRANSVERSE REINF.
3"
CLR
12"
10"
EXISTING STRUCTURE
SEE DETAIL 6 FOR
WALL ANCHORAGE
CONCRETE SLAB
PER PLAN
THICKENED
SLAB FOOTING
REINF. PER PLAN
CEILING JOIST PER PLAN
ROOF JOIST PER PLAN
TYP. WALL STUD AND
SPACING FOR PARAPET WALL
CONTINUOUS SHEATHING
PROVIDE FLAT STRAP w/ BLOCKING AT BREAKS
IN SHEATING. SEE DETAIL 6 SHEET S501
INSULATION PER
ARCHITECTURAL PLANS
SHEATHING PER PLAN
CONTINUOUS TRACK MEMBER. MATCH
SIZE AND THICKNESS OF JOIST
DIAPHRAGM BOUNDARY
FASTENER PER PLAN
WALL SHEATHING PER PLAN
ARCHITECTURAL EIFS
WALL STUD PER PLAN
ROOF JOIST BLOCKING
AT 24" O.C. FOR MIN. (2)
SPACES FROM WALL
CONTINUOUS TRACK AT WALL FOR
JOIST LEDGER. MATCH SIZE AND
THICKNESS OF CEILING JOIST
(2) #10 x 3/4" SCREWS
AT EACH WALL STUD
SIMPSON L30 AT
EACH JOIST
CEILING JOIST PER PLAN
ROOF JOIST BLOCKING
EACH SPACE
ROOF JOIST PER PLAN
SHEATHING PER PLAN
ALIGN WALL STUD DIRECTLY
BELOW ROOF JOIST
1'-4"
CONCRETE SLAB
PER PLAN
THICKENED
SLAB FOOTING
REINF. PER PLAN
1/8" WIDE x 1" DEEP SAWCUT FILL
WITH SEALANT SAW CUT SHALL
BE MADE WITHIN 4 TO 8 HOURS
AFTER CONCRETE POUR
CONCRETE SLAB AND
REINF. PER PLAN
FOOTING REINF.
PER PLAN
20"
10"
#4 AT 12" O.C.
HORIZ. REINF.
#4 AT 8" O.C.
TRANSVERSE REINF.
3"
CLR
8"
4"
THICKENED SLAB AT
DOOR OPENING
#4 AT 12" O.C. VERTICAL DOWEL
w/ STD. HOOK EACH END
CONTINUOUS TRACK AT WALL FOR
JOIST LEDGER. MATCH SIZE AND
THICKNESS OF CEILING JOIST
(2) #10 x 3/4" SCREWS
AT EACH WALL STUD
SIMPSON L30 AT
EACH JOIST
CEILING JOIST PER PLAN
ROOF JOIST PER PLAN
SHEATHING PER PLAN
ALIGN WALL STUD DIRECTLY
BELOW ROOF JOIST
ROOF JOIST BLOCKING
EACH SPACE
1
DETAIL
SCALE: 3/4" = 1'-0"
2
DETAIL
SCALE: 3/4" = 1'-0"
6
DETAIL
SCALE: 3/4" = 1'-0"
7
DETAIL
SCALE: 3/4" = 1'-0"
8
DETAIL
SCALE: 3/4" = 1'-0"
3
DETAIL
SCALE: 3/4" = 1'-0"
4
DETAIL
SCALE: 3/4" = 1'-0"
5
DETAIL
SCALE: 3/4" = 1'-0"
A
B
1 2
C
D
3 4 5 COPYRIGHT DESIGN WEST ARCHITECTSc
ISSUED:
CHECKED BY:
DRAWN BY:
PROJECT #:
MARK
:DA
TE:
DESC
RIPT
ION:
795 N
ORTH
400 W
EST
SAL
T LA
KE C
ITY
UT 84
103
255 S
OUTH
300
WES
T L
OGAN
UT
8432
1
GE H
EALT
HCAR
E - I
NCUB
ATOR
925 W
EST
1800
SOU
THLO
GAN,
UTA
H 84
321
SEP. 2019
CLR
419591
DLB
S
STE
TA
UTRS CT
R P
UR
F OE
G
OF UT HA
ERE
NI
NISO
L
EN
A LA
No. 295553RASMUSSENCRAIG L.
9/6/2019
9
DETAIL
SCALE: 3/4" = 1'-0"
STRUCTURAL
DETAILS
S-500
1
TYP
LATERAL BRIDGING
SCALE: N.T.S.
2" X 2" X 16 GA CLIP ANGLE
1
4
"
LESS THAN STUD WIDTH. ATTACH
USING (4) #10 SCREWS, CLIP
ANGLE AT EACH STUD.
1-1/2" COLD ROLLED
CHANNEL
TYPICAL STUD
NOTE:
STRAP MAY BE ELIMINATED WHERE
SHEATHING EXISTS. SHEATHING MUST
BE ATTACHED TO BLOCKING AS SHOWN
FOR STRAPS.
2
TYP
LATERAL BRACING
SCALE: N.T.S.
TYPICAL STUD
1-1/2" X 20 GA STRAP.
SCREW ATTACH TO
EVERY STUD FLANGE,
BOTH SIDES (STRAPPING
TO START & END ON
SOLID BLOCKING)
SCREW ATTACH BOTH
PARTS OF CUT FLANGE
TO STUD ON EACH SIDE
WITH #10 SCREWS.
MINIMUM 18 GA RUNNER TRACK
SOLID BLOCKING AT 10'-0" O.C., MAKE
RUNNER TRACK 8" LONGER THAN
OPENING WIDTH. CLIP FLANGES OF
TRACK 4" FROM EACH END. BEND
TRACK AT CLIPPED FLANGES.
ATTACH TRACK TO STUDS.
4
"
M
I
N
.
12" M
AX
.
1/2" M
AX
.
3
TYP SLIP TRACK
SCALE: N.T.S.
(2) #10 SCREWS TO METAL DECK OR (2)
HILTI 0.145"Ø P.A.F.(s) TO STEEL BEAM OR
SLAB ABOVE AT EACH STUD LOCATION
(2) #10 SCREWS TO METAL DECK OR (4)
HILTI 0.145"Ø P.A.F.(s) TO STEEL BEAM OR
SLAB ABOVE AT EACH JAMB LOCATION.
DEEP LEG "SLIP" TRACK W/
GAUGE TO MATCH STUDS
(600T250-43 MIN.)
TYPICAL STUD
ROW OF LATERAL BRACING
WITHIN 12" (MAX.) OF TOP
OF STUD. REF. TYPICAL
LATERAL BRIDGING DETAIL.
CORNER FRAMING
INTERSECTION FRAMING
4
INTERSECTING
WALL FRAMING
SCALE: N.T.S.
SCREWS AS REQ'D
SCREWS AS REQ'D STUD OR CLIP ANGLES
AS REQ'D FOR STUD
CONNECTION
O
P
E
N
I
N
G
2
"
M
I
N
.
3
"
M
I
N
.
S
P
A
C
I
N
G
F
R
O
M
E
D
G
E
3" M
IN
.
SP
CA
IN
G
3" MIN.
SPACING
TOP VIEW
2" M
IN
.
FR
OM
ED
GE
OF
S
LA
B
5
TYP DOOR JAMB
SCALE: N.T.S.
ATTACH TRACK TO SLAB
w/ (2) 0.145"Ø HILTI X-DNI
P.D.F.'s w/ 1" MIN. EMBED.
LOCATED AS SHOWN
ABOVE.
RUNNER TRACK
JAMB STUDS ATTACHED w/
TRACK. ATTACH TRACK TO JAMB
w/ (1) #10-16 SCREW AT 16" O.C.
ON EACH LEG, & (2) #10-16
SCREWS INTO WEBS AT 16" O.C.
(1) #10-16 SCREW AT
EACH FLANGE
6
TYP SHEAR WALL
SCALE: N.T.S.
SEE TRACK SPLICE DETAIL FOR
TOP TRACK SPLICE IN SHEAR
WALL
MIN DBL. STUD AT POST
EACH END OF SHEAR WALL STUDS
CONNECTED w/ #10-16 SCREWS
PER PLAN
PROVIDE 3" X 20 GA FLAT STRAP
W/ BLOCKING AT EDGE OF
SHEATHING.
7/16" APA RATED SHEATHING TO
RUN PERPENDICULAR TO FRAMING
SEE SHEAR WALL SCHEDULE FOR
SCREW REQUIREMENTS
3/8" SIMPSON TITEN HD w/ 2-1/2"
EMBEDMENT AT 32" O.C. w/
3"x3"x1/4" PLATE WASHER
HOLDOWN SEE PLAN FOR TYPE
AND LOCATION AND DETAIL 7
SHEET S-501
3
"
F
R
O
M
E
D
G
E
O
F
S
L
A
B
8
SHEAR WALL
HOLDOWN
SCALE: N.T.S.
TYPICAL TRACK
HOLDOWN PER PLAN
SHEAR WALL CHORD
PER PLAN
ANCHOR PER PLAN
9
RUNNER TRACK
SCALE: N.T.S.
STUDS TO FIT TIGHT
IN TRACK AT
BEARING WALLS
ATTACH RUNNER
TRACK TO STUD w/
(1) #10 SCREWS AT
EACH STUD FLANGE
7
TYP .
TRACK SPLICE
SCALE: N.T.S.
TYP. RUNNER TRACK
SPLICE USING SHORT
SECTION OF STUD
SCREW ATTACH BOTH ENDS OF
RUNNER TRACKS TO SHORT
SECTION OF STUD w/ (3) #10
SCREWS BOTH SIDES AS SHOWN
HEADER ATTACHMENT
1"
1"
1"
1"
10
TYP HEADER
SCALE: N.T.S.
CRIPPLE STUD
BOX HEADER
SEE HEADER SCHEDULE
FOR MEMBER SIZES.
ATTACH RUNNER TRACK TO
HEADER MEMBERS W/ (2)
#10 SCREWS AT 16" O.C.
4-1/2" LONG RUNNER TRACK SIZE &
GAUGE TO MATCH STUDS. SCREW
ATTACH TO CAP STUD W/ (6) #10
SCREWS & TO HEADER W/ (3) #10
SCREWS EACH SIDE
JAMB POST SEE HEADER
SCHEDULE FOR NUMBER
FULL HEIGHT JAMB SEE HEADER
SCHEDULE FOR NUMBER
HEADER ATTACHMENT SEE
DETAIL ABOVE
3" X 3" X 16 GA CLIP ANGLE
1
2
" LESS THAN
STUD WIDTH. ANGLES TOP AND BOTTOM
OF HEADER. ATTACH TO JAMB AND
RUNNER TRACK WITH (4) #10 SCREWS
EACH LEG AS SHOWN ABOVE
ATTACH JAMB
MEMBERS w/ (2)
#10 SCREWS AT
16" O.C.
ATTACH DOUBLED
(BACK TO BACK)
MEMBERS w/ (2) #10
SCREWS AT 12" O.C.
11
BACK TO BACK
MEMBERS
SCALE: N.T.S.
A
B
1 2
C
D
3 4 5 COPYRIGHT DESIGN WEST ARCHITECTSc
ISSUED:
CHECKED BY:
DRAWN BY:
PROJECT #:
MARK
:DA
TE:
DESC
RIPT
ION:
795 N
ORTH
400 W
EST
SAL
T LA
KE C
ITY
UT 84
103
255 S
OUTH
300
WES
T L
OGAN
UT
8432
1
GE H
EALT
HCAR
E - I
NCUB
ATOR
925 W
EST
1800
SOU
THLO
GAN,
UTA
H 84
321
SEP. 2019
CLR
419591
DLB
S
STE
TA
UTRS CT
R P
UR
F OE
G
OF UT HA
ERE
NI
NISO
L
EN
A LA
No. 295553RASMUSSENCRAIG L.
9/6/2019
STRUCTURAL
DETAILS
S-501