WorkHorse Horizontal ADCP Quick Start Guide P/N 957-6210-00 (December 2008)

description

MANUAL DE USO DEL HADCP

Transcript of HADCPQK

  • WorkHorse Horizontal ADCP Quick Start Guide

    P/N 957-6210-00 (December 2008)

  • Table of Contents Horizontal ADCP Quick Start Guide .......................................................................................................... 1

    Introduction ............................................................................................................................................................... 1 How to Contact Teledyne RD Instruments ............................................................................................................... 1 What This Guide Covers .......................................................................................................................................... 1

    Section 1.0: Overview ................................................................................................................................. 2 Packing List .............................................................................................................................................................. 3 Hardware Overview .................................................................................................................................................. 4 Power Overview ....................................................................................................................................................... 5 Serial Communication Overview .............................................................................................................................. 6 RS-232 to RS-422 Converter Overview ................................................................................................................... 6 Software Overview ................................................................................................................................................... 7 Installing the Software .............................................................................................................................................. 9

    Section 2.0: Bench Testing ...................................................................................................................... 10 Visual Inspection .................................................................................................................................................... 11 Setup the H-ADCP ................................................................................................................................................. 12 Changing the Baud Rate in the H-ADCP ................................................................................................................ 13 Testing the H-ADCP ............................................................................................................................................... 14 Compass Alignment ............................................................................................................................................... 16

    Preparing for Calibration ................................................................................................................................... 16 Compass Calibration Verification ...................................................................................................................... 17 Compass Calibration Procedure ....................................................................................................................... 17

    Section 3.0: Installation and Setup ......................................................................................................... 19 H-ADCP Preparation .............................................................................................................................................. 19 Pressure Sensor Preparation ................................................................................................................................. 20 Site Preparation ...................................................................................................................................................... 20 H-ADCP Installation ............................................................................................................................................... 21 Testing the Installed H-ADCP ................................................................................................................................. 22

    Section 4.0: Collecting Current Profile Data with SurfaceView............................................................ 24

    Section 5.0: Data Playback with SurfaceView ........................................................................................ 29

    List of Figures Figure 1. H-ADCP 300 kHz Transducer .......................................................................................... 2 Figure 2. H-ADCP 300 kHz Transducer (Long Range) ................................................................... 2 Figure 3. H-ADCP 600 kHz Transducer .......................................................................................... 3 Figure 4. I/O Cable Overview ......................................................................................................... 5 Figure 5. 25-pin to 9-pin RS-232 to RS-422 Converter ................................................................... 6 Figure 6. H-ADCP Connections .................................................................................................... 12 Figure 7. Compass Calibration ..................................................................................................... 18

  • UNOTES

  • Horizontal ADCP Quick Start Guide

    P/N 957-6210-00 (December 2008) page 1

    Horizontal ADCP Quick Start Guide

    Introduction Thank you for purchasing the Teledyne RD Instruments (TRDI) WorkHorse Horizontal ADCP (H-ADCP). This Quick Start Guide will lead you through the steps required for a successful deployment. Please read the entire guide, and then follow the instructions in the order they are presented.

    How to Contact Teledyne RD Instruments If you have technical issues or questions involving a specific application or deployment with your instrument, contact our Field Service group: Teledyne RD Instruments Teledyne RD Instruments Europe

    14020 Stowe Drive Poway, California 92064

    2A Les Nertieres 5 Avenue Hector Pintus 06610 La Gaude, France

    Phone +1 (858) 842-2600 Phone +33(0) 492-110-930

    FAX +1 (858) 842-2822 FAX +33(0) 492-110-931

    Sales [email protected] U Sales [email protected] U

    Field Service [email protected] U Field Service [email protected] U

    Customer Service Administration [email protected] U Web: Uhttp://www.rdinstruments.com U

    24 Hour Emergency Support +1 (858) 842-2700

    What This Guide Covers The Quick-Start Guide covers the basics of H-ADCP testing and deployments using the Surfa-ceView software. It will guide you through the inspection of components and bench testing of the H-ADCP. It will also tell you how to install and collect data with the H-ADCP.

    Additional information can be found in the H-ADCP Operation Manual that is supplied on CD-ROM and the SurfaceView Users Guide that is supplied on the SurfaceView software CD. If you ordered the directional wave measurement upgrade you will also be sent the WavesMon software package. The Users Guide is supplied on the WavesMon software CD.

    NOTE. To purchase a printed copy of the system documentation, contact our Customer Service department at [email protected] or call (858) 842-2600 and order the H-ADCP Manual kit.

  • Horizontal ADCP Quick Start Guide

    page 2 Teledyne RD Instruments

    Section 1.0: Overview The first step is to familiarize yourself with the components. Check to make sure that every-thing has been shipped to you, and then read the short descriptions of the hardware and soft-ware that comes with the H-ADCP.

    This Section Covers:

    Packing List

    Hardware Overview

    Software Overview

    Installing the Software

    Figure 1. H-ADCP 300 kHz Transducer

    Figure 2. H-ADCP 300 kHz Transducer (Long Range)

  • Horizontal ADCP Quick Start Guide

    P/N 957-6210-00 (December 2008) page 3

    Figure 3. H-ADCP 600 kHz Transducer

    Packing List You should have the following items.

    Inventory Horizontal ADCP I/O cable Tools and Spare Parts kit Shipping crate (please save all foam

    for reshipping use) H-ADCP Quick Start Guide H-ADCP Documentation CD SurfaceView software CD TRDI Tools software CD The Directional Wave Measurement upgrade includes the following: WavesMon software CD

    NOTE. If you ordered RS-422 communications, a RS-232 to RS-422 adapter will be added.

    5 meter I/O CABLE

    SHIPPING CRATE

    SOFTWARE WAVES UPGRADE

    HORIZONTAL ADCPDUMMY PLUG

    (INSTALLED ON ADCP)

    TOOLS AND SPARE PARTS KIT

    SurfaceView WavesMonWaveViewRDI Tools

    25 meter EXTENSION CABLE

    POWER SUPPLY

    20-60V!

    DOCUMENTATION CD

  • Horizontal ADCP Quick Start Guide

    page 4 Teledyne RD Instruments

    Hardware Overview The H-ADCP Horizontal ADCP provides real time current profiles from a fixed position. Take a moment to familiarize yourself with the features of the H-ADCP, paying special attention to the orientation of the X, and Y-axis. Understanding the orientation of the H-ADCP is crucial for proper installation.

    UH-ADCP Components UI/O Cable Connector Input/Output (I/O) cable connects the Workhorse H-ADCP to the computer. Urethane Face The urethane face covers the transducer ceramics. Never set the transducer on a hard surface. The urethane face may be damaged. Housing The standard housing allows deployment depths to 200M Thermistor The Thermistor measures the water temperature. Pressure Sensor (depth)- The optional pressure sensor measures water pres-sure. Transducer Head The electronics and transducer ceramics are mounted to the transducer head

    Mounting Orientation

    UX Axis - Pitch is measured around the X-axis and should be set to zero. UY- Axis - Roll is measured around the Y- axis and should be set to zero. Pressure Sensor The pressure sen-sor must be facing up towards the water surface

    1 3 2 URETHANE FACE

    ANODE

    ANODES

    I/O CABLE

    PRESSURE SENSOR(OPTIONAL)

    TRANSDUCER HEAD

    HOUSING

    I/O CABLE CONNECTOR

    +X

    -X

    +Y

    1

    2

    3

    TOP

  • Horizontal ADCP Quick Start Guide

    P/N 957-6210-00 (December 2008) page 5

    Power Overview H-ADCPs require +20 to 50 VDC to operate. The included AC Adapter runs on any standard AC power and supplies +48 VDC to run the H-ADCP for testing and setup. Substitute your own power supply with a voltage of 45 to 50 VDC for deployments.

    Figure 4. I/O Cable Overview

    REDYEL

    BLK/WHTBLK

    41

    32985

    P2

    12564

    37

    POWER +POWER -

    COMMUNICATION RETURN

    RS-232 IN / CH A RS-485A / RS-422 OUT ARS-232 OUT / CH A RS-485B/ RS-422 OUT B

    CH B RS-485A / RS-422 IN ACH B RS-485B / RS-422 IN B

    P1BLKWHTBLUBRNGRN

    P3

    76

    5

    4

    12

    3

    12 3

    4

    PIN 1PIN 6

  • Horizontal ADCP Quick Start Guide

    page 6 Teledyne RD Instruments

    Serial Communication Overview The standard communications settings for a H-ADCP is RS-422, 9600-baud, no parity, 8 data bits and 1 stop bit.

    You can set the H-ADCP for baud rates other than 9600 baud using the CB command (see Changing the Baud Rate in the H-ADCP U). If you make the new baud rates permanent with the CK command, BBTalk will search for the correct settings if you use the Auto Detect function, which searches for the H-ADCPs current serial port and baud rate (see Testing the H-ADCP U). If you tell the programs these settings, you will save time required for searching. Both WinSC and BBTalk will use the last communication setting for future use. Change Communication Setting. Changing a switch setting on the PIO board can change the communication setting between RS-232 and RS-422. The switch is in plain view on the top circuit board, near the cable connectors. Its settings are plainly marked on the board. If the serial protocol is set for RS-422 and your computer expects RS-232, you will need an RS-232 to RS-422 adapter between the H-ADCP cable and your computer.

    RS-232 to RS-422 Converter Overview The adapter converts RS-232-to-RS-422. This adapter (includes cable) is supplied with the H-ADCP.

    Figure 5. 25-pin to 9-pin RS-232 to RS-422 Converter

    NOTE. The converter supplied with the H-ADCP is powered by the computers COM port. It may be necessary to use a different converter when very long communication lines are used (maximum 1300 meters).

    3162147

    RS422 RDA

    DATA COMMON

    RS422 RDBRS422 TDARS422 TDB

    P2ADAPTER SIDE

    SOCKET

    P1ADCP SIDE

    PLUGREDORNYEL

    GRNBLU

    32985 DATA COMMON

    RS422 TDARS422 TDBRS422 RDARS422 RDB

    P2

    25-pin to9-pin adapter(use as needed)

    P1

  • Horizontal ADCP Quick Start Guide

    P/N 957-6210-00 (December 2008) page 7

    Software Overview You will need several software programs to test and collect current profile data with the H-ADCP.

    SurfaceView is TRDIs real-time data collection program for horizontal H-ADCPs. This guide does not go into detail about the operation of SurfaceView. For more information see the Sur-faceView Users Guide. BBTalk is a terminal emulator that allows your computer to communicate with the H-ADCP. This guide does not go into detail about the operation of BBTalk. For more information see the TRDI Tools Users Guide.

    If you ordered the directional wave measurement upgrade you will also need WavesMon and WaveView. WavesMon is the directional wave measurement package for the H-ADCP. Use WaveView to view wave data. It is an enormously useful tool for turning waves data into waves information.

    SurfaceView Main Screen SurfaceView is TRDIs real-time data collection program for H-ADCPs.

    BBTalk Main Screen BBTalk is a terminal emulator program that lets your computer communicate with the H-ADCP.

  • Horizontal ADCP Quick Start Guide

    page 8 Teledyne RD Instruments

    WavesMon Main Screen WavesMon is the directional wave mea-surement package used for the H-ADCP. This program is included if you ordered the directional wave mea-surement upgrade. For more information on how to use this software package, see the WavesMon Users Guide and WavesMon Quick Start Guide.

    WaveView Main Screen WaveView is geared toward analysis and presentation of processed wave data created by the software program WavesMon. WaveView provides the big picture view of the wave climate as well as details of individual height spec-tra and directional spectra. Time-series of significant wave height, peak period, peak direction, water level, and height spectra make it easy to find significant events. This program is included if you ordered the directional wave measurement up-grade. For more information on how to use this software package, see the WaveView Users Guide.

  • Horizontal ADCP Quick Start Guide

    P/N 957-6210-00 (December 2008) page 9

    Installing the Software You will be installing several software packages. These will be required for testing, installa-tion, and data collection.

    TRDI software requires a Windows compatible computer with the following specifications:

    Windows XP or Windows 2000

    Pentium III 600 MHz class PC (higher recommended)

    64 megabytes of RAM (128 MB RAM recommended)

    50 MB Free Disk Space plus space for data files (A large, fast hard disk is recom-mended)

    One Serial Port (two or more High Speed UART Serial Port recommended)

    Minimum display resolution of 1024 x 768, 256 color (higher recommended)

    CD-ROM Drive

    Mouse or other pointing device

    Software Installation a. Insert the compact disc into your CD-

    ROM drive and then follow the browser instructions on your screen. If the browser does not appear, com-plete Steps "b" through "d."

    b. On the Windows task bar, click the Start button, and then click Run.

    c. Type :launch. For example, if your CD-ROM drive is drive D, type d:launch.

    d. Follow the browser instructions on your screen

  • Horizontal ADCP Quick Start Guide

    page 10 Teledyne RD Instruments

    Section 2.0: Bench Testing The bench-testing process ensures that the H-ADCP is working properly before you put it in the water. The bench-test procedure will involve a visual inspection for obvious damage, and powered tests that will verify that the H-ADCPs electronics and transducers are functioning.

    This Section Covers:

    Visual Inspection

    Setup the H-ADCP

    Testing the H-ADCP

    Compass Alignment

  • Horizontal ADCP Quick Start Guide

    P/N 957-6210-00 (December 2008) page 11

    Visual Inspection Before connecting the H-ADCP, make a quick visual inspection of the components to make sure nothing is damaged.

    UI/O Cable and Dummy Plug Remove the strain relief. Roll the rubber retaining strap up so that it does not hinder plug removal. Remove the dummy plug by pulling with a straight-out motion. Do not wiggle the connector up; this can damage the con-nector pins. A slight side-to-side wiggle as you are pulling on the cable is OK.

    Visual Inspection Inspect the pins for damage. Make sure that they are straight and that there is no corrosion on the metal surfaces. Inspect the H-ADCP for damage. There should be no cracks or peeling surfaces. Check the following: Urethane faces Pressure housing Anode condition

  • Horizontal ADCP Quick Start Guide

    page 12 Teledyne RD Instruments

    Setup the H-ADCP You will now connect the H-ADCP to a power supply and a computer with the TRDI software installed.

    Figure 6. H-ADCP Connections

    Remove the dummy plug and connect the I/O cable to the H-ADCP. Do so by pushing straight in against the connec-tor. Remember, don't wiggle the con-nector up, push it straight in. Connect the I/O cable to your comput-er's communication port. Connect the I/O cable power connector to a 20 to 60 volt DC power supply. Note: Place a light amount of DC-111 lubricant on the connector pins (rubber portion only). This will make it easier to connect or remove the I/O cable and dummy plug.

    NOTE. H-ADCP units are pre-configured to use RS-232 or RS-422 communications protocols. Your computers communication port must match the H-ADCP, or use an adapter. To change the communication protocol, refer to the H-ADCP Operation Manual. If you ordered the RS-422 configuration, the H-ADCP is supplied with a RS-232 to RS-422 converter. The RS-232 to RS-422 converter must be disconnected from the H-ADCP when the cable has been disconnected from the computer. If left connected, the internal battery used to maintain ROM inside the H-ADCP can be drained or false Breaks may be sent.

    COMPUTER

    H-ADCP

    32

    1

    ANODE

    POWER SUPPLY

    48 VDC

    20-60V!

    100-240 VAC50/60 Hz

    I/O CABLETO SERIAL PORT(COM1 or COM2)

    BBTalk

    RS-232 to RS-422

    CONVERTOR

    Beam 1

    Beam 3

    Beam 2

  • Horizontal ADCP Quick Start Guide

    P/N 957-6210-00 (December 2008) page 13

    Changing the Baud Rate in the H-ADCP The H-ADCP can be set to communicate at baud rates from 300 to 115200. The factory default baud rate is always 9600 baud. The baud rate is controlled via the CB-command. The follow-ing procedure explains how to set the baud rate and save it in the H-ADCP. This procedure assumes that you will be using the program BBTalk that is supplied by TRDI.

    [BREAK Wakeup A] WorkHorse Broadband H-ADCP Version 16.xx Teledyne RD Instruments (c) 1996-2005 All Rights reserved. >? >cr1 [Parameters set to FACTORY defaults]

    Connect the H-ADCP to the computer and apply power. Start the BBTalk program and establish communications with the H-ADCP. Wa-keup the H-ADCP by sending a break signal with the End key. At the ">" prompt in the communication window, type CR1 then press the Enter key. This will set the H-ADCP to the factory default settings.

    BAUD RATE CB-command

    300 CB011 1200 CB111 2400 CB211 4800 CB311 9600 CB411 (Default) 19200 CB511 38400 CB611 57600 CB711 115200 CB811

    Send the CB-command that selects the baud rate you wish. The table on the left shows the CB-command settings for different baud rates with no parity and 1 stop bit. For example, to change the baud rate to 115200, at the ">" prompt in the com-munication window, type cb811 then press the Enter key.

    The CB? command will identify the communication setting.

    >cb? CB = 411 ----------------- Serial Port Control (Baud [4=9600]; Par; Stop) >cb811 >CK [Parameters saved as USER defaults] >cb? CB = 811 ----------------- Serial Port Control (Baud [8=115200]; Par; Stop) >

    BBTalk will send the command CK to save the new baud rate setting. Exit BBTalk. The H-ADCP is now set for the new baud rate. The baud rate will stay at this setting until you change it back with the CB-command.

    Exit BBTalk so the communication port is available for use with other pro-grams.

  • Horizontal ADCP Quick Start Guide

    page 14 Teledyne RD Instruments

    Testing the H-ADCP Before deploying the H-ADCP, it is a good idea to make sure that it is working properly. This simple test checks that the H-ADCP is able to communicate with the computer.

    Start BBTalk Start the BBTalk program (for help on using BBTalk, see the TRDI Tools Us-er's Guide). On the Connect To screen, select WorkHorse. Select the COM port the H-ADCP cable is connected to. Click Next.

    Enter the Baud Rate, Parity, Stop Bits, and Flow Control. If you are unsure of the settings, leave them at the default settings. Click Next, than Finish.

    Communications Settings On the File menu, click Break (you can also press the End key to send a break or use the Toolbar and press the blue B button). You should see the wakeup message appear on the log file window. If the wakeup message is not readable, on the File menu, click Properties. Click the Auto Detect ADCP button. Click OK when the H-ADCP is detected. Try to wakeup the H-ADCP again.

  • Horizontal ADCP Quick Start Guide

    P/N 957-6210-00 (December 2008) page 15

    >cr1 [Parameters set to FACTORY defaults] >ck [Parameters saved as USER defaults] >ps0 Instrument S/N: 3192 Frequency: 307200 HZ Configuration: 3 BEAM, HORIZONTAL Beam Angle: 25 DEGREES Beam Pattern: CONVEX Orientation: UP Sensor(s): HEADING TILT 1 TILT 2 DEPTH TEMPERATURE PRESSURE Pressure Sens Coefficients: (c3,c2,c1,offset) 0.00,0.00,0.01,-1.15 Temp Sens Offset: -0.22 degrees C CPU Firmware: 11.07 [0] Boot Code Ver: Required: 1.13 Actual: 1.13 DEMOD #1 Ver: ad48, Type: 1f DEMOD #2 Ver: ad48, Type: 1f PWRTIMG Ver: 85d3, Type: 6 Board Serial Number Data: D1 00 00 00 E6 FD 28 09 PIO727-3013-04X B4 00 00 00 E7 09 74 09 REC727-1015-04X 34 00 00 00 E6 F8 29 09 CPU727-2000-00H A9 00 00 02 C9 37 69 09 DSP727-2001-04F >

    Testing the H-ADCP At the ">" prompt in the communication window, enter the direct command CR1 then press the Enter key. This will set the H-ADCP to the factory default set-tings. At the ">" prompt in the communication window, enter the direct command CK then press the Enter key. This will save the factory default setting as the default power up condition. At the ">" prompt in the communication window, enter the direct command PS0 then press the Enter key. This will dis-play the H-ADCP system configuration data.

    >pa PRE-DEPLOYMENT TESTS CPU TESTS: RTC......................................PASS RAM......................................PASS ROM......................................PASS RECORDER TESTS: PC Card #0...............................NOT DETECTED Card Detect............................PASS Communication..........................PASS DOS Structure..........................PASS Sector Test (short)....................PASS PC Card #1...............................NOT DETECTED DSP TESTS: Timing RAM...............................PASS Demod RAM...............................PASS Demod REG...............................PASS FIFOs....................................PASS SYSTEM TESTS: XILINX Interrupts... IRQ3 IRQ3 IRQ3 ...PASS Receive Loop-Back........................PASS Wide Bandwidth...........................PASS Narrow Bandwidth.........................PASS RSSI Filter..............................PASS Transmit.................................PASS SENSOR TESTS: H/W Operation............................PASS >

    At the ">" prompt in the communication window, enter the direct command PA then press the Enter key. This will run the H-ADCP Pre-deployment tests.

  • Horizontal ADCP Quick Start Guide

    page 16 Teledyne RD Instruments

    >Pt3 Correlation Magnitude: Narrow Bandwidth Lag Bm1 Bm2 Bm3 0 255 255 255 1 144 144 146 2 70 71 73 3 37 33 35 4 24 18 18 5 13 12 8 6 6 5 2 7 3 2 1 High Gain RSSI: 71 67 74 Low Gain RSSI: 16 16 18 SIN Duty Cycle: 49 52 49 COS Duty Cycle: 48 50 48 Receive Test Results = $000 ... PASS >

    At the ">" prompt in the communication window, enter the direct command PT3 then press the Enter key. This will run the H-ADCP Receive Path test.

    Compass Alignment The compass calibration algorithm corrects for the distortions caused by the local magnetic fields to give you an accurate measurement. You should be aware of the following items:

    We recommend against calibrating the H-ADCP while on a ship or platform. The ships motion and magnetic fields from the hull and engine will likely prevent suc-cessful calibration.

    If you think your mounting fixture or frame has some magnetic field or magnetic permeability, calibrate the H-ADCP inside the fixture. Depending on the strength and complexity of the fixtures field, the calibration procedure may be able to cor-rect it.

    Preparing for Calibration

    Connect the H-ADCP as shown. Place the H-ADCP on a piece of strong cardboard on top of a smooth wooden (non-magnetic) table. If a wooden table is not available, place the H-ADCP on the floor as far away from metal objects as possible. Use the cardboard to ro-tate the H-ADCP during calibrationthis way you will not scratch the H-ADCP. Place the H-ADCP so that the anode is facing down and the pressure sensor is facing up.

    CAUTION. If you calibrate the compass with the pressure sensor facing downward, the compass calibration will be invalid. Compass errors in excess of 5 degrees may occur.

    COMPUTER

    H-ADCP

    32

    1

    ANODE

    POWER SUPPLY

    48 VDC

    20-60V!

    100-240 VAC50/60 Hz

    I/O CABLETO SERIAL PORT(COM1 or COM2)

    BBTalk

    RS-232 to RS-422

    CONVERTOR

    Beam 1

    Beam 3

    Beam 2

  • Horizontal ADCP Quick Start Guide

    P/N 957-6210-00 (December 2008) page 17

    Compass Calibration Verification Compass calibration verification is an automated built-in test that measures how well the com-pass is calibrated. The procedure measures compass parameters at every 5 of rotation for a full 360 rotation. When it has collected data for all required directions, the H-ADCP com-putes and displays the results.

    WorkHorse H-ADCP Teledyne RD Instruments (c) 2003 All rights reserved. Firmware Version: xx.xx >

    Start BBTalk. See the TRDI Tools Us-ers Guide for assistance on using BBTalk. Using BBTalk, send a Break to wake up the H-ADCP.

    >AX ---------------------------------------------------------------------- TRDI Compass Error Estimating Algorithm Press any key to start taking data after the instrument is setup. Rotate the unit in a plane until all data samples are acquired... rotate less than 5/sec. Press Q to quit. N NE E SE S SW W NW N ^ ^ ^ ^ ^

    At the > prompt, type AX and press the Return key.

    HEADING ERROR ESTIMATE FOR THE CURRENT COMPASS CALIBRATION: OVERALL ERROR: Peak Double + Single Cycle Error (should be < 5(): ( 1.55( DETAILED ERROR SUMMARY: Single Cycle Error: ( 1.54( Double Cycle Error: ( 0.07( Largest Double plus Single Cycle Error: ( 1.61( RMS of 3rd Order and Higher + Random Error: ( 0.31(

    Rotate the H-ADCP slowly 360 degrees (approximately 5 degrees per second). Pay particular attention to the Overall Error. If the overall error is less than 2, the compass does not require alignment. You can align the compass to reduce the overall error even more (if desired).

    Compass Calibration Procedure The built-in automated compass calibration procedure is similar to the alignment verification. The H-ADCP uses the first rotation(s) to compute a new calibration matrix and the final rota-tion to verify the calibration. It will not accept the new matrix unless the calibration was car-ried out properly, and it asks you to verify that you want to use the new calibration if it is not as good as the previous calibration. There are two compass calibrations to choose from; one only corrects for hard iron while the other corrects for both hard and soft iron characteristics for materials rotating with the ADCP. Hard iron effects are related to residual magnetic fields and cause single cycle errors while soft iron effects are related to magnetic permeability that distorts the earths magnetic field and causes double cycle errors.

    For the H-ADCP that is normally in a fixed horizontal attitude we use option b of the AF command that corrects for both hard and soft iron characteristics in a single plane.

  • Horizontal ADCP Quick Start Guide

    page 18 Teledyne RD Instruments

    Figure 7. Compass Calibration

    WorkHorse Broadband H-ADCP Version 16.xx Teledyne RD Instruments (c) 1996-2005 All Rights reserved. >?

    Using BBTalk, send a Break to wake up the ADCP.

    >AR Do you really want to write over the active fluxgate calibration data [y or n]?Y Fluxgate Calibration Matrices Updated with Factory Original Values. >

    At the > prompt, type AR and press the Return key. Type Y to return the Flux-gate Calibration Matrices with the facto-ry original values.

    >AF ---------------------------------------------------------------------- Field Calibration Procedure Choose calibration method: a. Remove hard iron error (single cycle) only. b. Remove hard and soft iron error (single + double cycle). c. Help. d. Quit.

    At the > prompt, type AF and press the Return key. Choose option b to start the calibration procedure. When prompted, rotate the H-ADCP slowly 360 degrees (approx-imately 5 degrees per second).

    Successfully evaluated compass performance after the field calibration update. Press any key to continue... Making new calibration parameters permanent. Calibration parameters have been updated in RAM and Flash. ---------------------------------------------------------------------- Field Calibration Procedure Hard and Soft Iron Calibration Compass field calibration procedure is complete. >>> Total error before calibration: 2.3 > Total error after calibration: 0.2

  • Horizontal ADCP Quick Start Guide

    P/N 957-6210-00 (December 2008) page 19

    Section 3.0: Installation and Setup Proper H-ADCP and site preparation is critical for a successful deployment. In this section we will prepare the H-ADCP for deployment and ensure that the installation site meets the neces-sary requirements. Then, we will install the H-ADCP in the water and make sure that it is functioning.

    This Section Covers:

    H-ADCP Preparation

    Pressure Sensor Preparation

    Site Preparation

    H-ADCP Installation

    Testing the Installed H-ADCP

    H-ADCP Preparation Before you put the H-ADCP into the water, you must prepare it for deployment.

    Check the Housing Seal If the housing is not properly sealed prior to deployment, the H-ADCP will be destroyed. Visually inspect the mating surfaces of the housing to ensure that there are no obstructions. The faces should be parallel to one another and there should be no gaps. Check to en-sure that the hardware holding the hous-ing together are tight.

    CAUTION. If the H-ADCP was opened, read the H-ADCP Operation Manual for details on H-ADCP re-assembly.

  • Horizontal ADCP Quick Start Guide

    page 20 Teledyne RD Instruments

    Pressure Sensor Preparation Horizontal ADCP Teledyne RD Instruments (c) 2008 All rights reserved. Firmware Version: xx.xx > > >AZ >

    Zero the Pressure Sensor Use the AZ-command to zero out the pressure sensor at the deployment site, before putting the H-ADCP into the wa-ter. Connect and apply power to the system. Start BBTalk and wakeup the H-ADCP (press the END key). Type AZ and press the Return key. Exit BBTalk.

    Site Preparation The H-ADCP must be rigidly mounted to a structure, such as a large vessel, solid wall, pier, or pile. The mounting structure must be firm, stable, without settlement or displacement over time.

    See the H-ADCP Operation Manual for Outline Installation Drawings.

    Mounting Location In order to gain the maximum profiling range at a site, we recommend the H-ADCP be mounted near the middle of the water depth at the mounting struc-ture location. For a site with a signifi-cant seasonal change of water level, the H-ADCP may need to be mounted at different elevations according to sea-sonal water level change. In deep water we recommend the H-ADCP be mounted 5 to 20 meters below the sur-face. UH-ADCP Orientation UX-Axis - Pitch is measured around the X-axis and should be set to zero. UY-Axis - Roll is measured around the Y- axis and should be set to zero. Pressure Sensor The pressure sen-sor must be facing up towards the water surface

    ANODE

    PRESSURESENSOR

    +20 TO 60 VDC

    SoftwareSurfaceView

    2

  • Horizontal ADCP Quick Start Guide

    P/N 957-6210-00 (December 2008) page 21

    H-ADCP Installation The H-ADCP must be mounted with its transducer orientation in the horizontal direction. That is, the three transducers, which are for water velocity measurement, must be looking horizon-tally (also called side-looking). The pressure sensor, which is for water depth measurement, must be looking up towards the water surface.

    NOTE. For CRITICAL H-ADCP Waves Array installation information, refer to the H-ADCP Wave System Installation Checklist in the H-ADCP Operation Manual.

    Attach H-ADCP to Structure Attach the H-ADCP to a rigid structure, solid wall, pier, or pile. The mounting structure must be firm, stable, without settlement or displacement over time. Double-check all of the connections, seals, and bolts. Connect the I/O cable and make sure that the strain relief is attached to the cable. Bolt the H-ADCP to the structure. Make sure that the I/O cable is routed so that it is not subject to becoming entangled or snagged.

    NOTE. TRDI offers optional kits for attaching the H-ADCP to a structure. For additional information, please contact our field service department (see How to Contact Teledyne RD Instruments).

    I/O CABLE

    48 VDC

    H-ADCP

    SurfaceViewSurfaceWaves

  • Horizontal ADCP Quick Start Guide

    page 22 Teledyne RD Instruments

    Testing the Installed H-ADCP Once the H-ADCP is mounted in the water, its time to communicate with it and make adjust-ments. SurfaceView will be used to adjust the pitch and roll settings as well as to check for ab-normalities in the acoustic backscattering intensity profile.

    Start SurfaceView. On the Setup Wizard, click Real-Time. Click Next.

    Once communications is established, on the Communications Settings screen, click Next. Accept the default setting on the next several screens. Click Finish.

    Pitch and Roll Adjustment The deployment screen of the Surface-View software indicates the pitch and roll angles of the H-ADCP. Adjust the mount until the roll is zero. The H-ADCP rolls about the Y-axis. The roll must be zero. Adjust the mount until the pitch is zero. The H-ADCP pitches about the X-axis. Note: A slight upward or downward pitch may be required to avoid beam interference from the bottom or surface.

  • Horizontal ADCP Quick Start Guide

    P/N 957-6210-00 (December 2008) page 23

    Intensity Plot Start collecting data (see Section 4.0: Collecting Current Profile Data with Sur-faceView) and check the intensity profile for abnormal data. What you are looking for are unusually large spikes, especially ones that stay fixed along the Y-axis of the plot. Spikes that don't move are usually fixed objects in the water like rocks or pylons. Spikes that move along the Y-axis are usually moving objects like schools of fish or logs. It is important to mount the H-ADCP in a location free from fixed obstructions. The velocity data will be questionable if the intensity profile shows abnormalities.

  • Horizontal ADCP Quick Start Guide

    page 24 Teledyne RD Instruments

    Section 4.0: Collecting Current Profile Data with SurfaceView This section has simple instructions for collecting current profile data with a Horizontal ADCP.

    Start SurfaceView Start SurfaceView by double clicking the SurfaceView icon. On the Start Wizard, click Real-Time Data Acquisition. Click Next.

    Setup Communications A message box will display whenever the wizard is waiting for a response from the COM port device. If the connection did not get automati-cally made, on the Communications Settings dialog box, select the COM Port, Baudrate, Databits, and Stopbits that the H-ADCP is connected to. If you are unsure of the setting, use Auto Detect. Click Next.

  • Horizontal ADCP Quick Start Guide

    P/N 957-6210-00 (December 2008) page 25

    Output Raw Data File (Prefix) Surfa-ceView uses the Deployment Name (Prefix) to create the data file made dur-ing data collection. Output Directory Use the Output Directory field to select where the raw data file will be stored. Click the Pick Directory button to select the directory. Use the Date/Time in File Name selec-tion to add the Date and Time to the file name. Enter a Maximum File Size in the box. When the data file size reaches the maximum limit, the file will be closed and a new data file generated with an incremented file name. Click Next.

    Click the Browse button and select a path to save the Peak/Mean Data Log file (*.CSV). CSV files can not be over-written. If you see a warning message, click OK and select a different path or file name by clicking the Browse button. Click Next to continue.

  • Horizontal ADCP Quick Start Guide

    page 26 Teledyne RD Instruments

    Select Use Wizard and click Next.

    Enter the H-ADCP Setup Choices SurfaceView enters the default settings in the Parameter Setup screen based on the H-ADCP frequency. Leave the default settings for the Num-ber of Cells, Cell Size, and Blank Dis-tance for this Quick Start example.

  • Horizontal ADCP Quick Start Guide

    P/N 957-6210-00 (December 2008) page 27

    Salinity (ppt) Enter the salinity of the water. Coordinates for Velocity Select Beam, Instrument, Ship Coordinates or Earth. Magnetic Variation or Offset (deg) Use a chart to determine the local mag-netic variation in your area and enter the value in the box (only available if using Earth coordinates). ADCP Xducer Depth If the ADCP does not have a pressure sensor, check the Use Manual Depth box and set the depth of the ADCP transducer faces below the surface. This value is used by the ADCP to calculate the speed of sound. ADCP Xducer Heading If the ADCP does not have a compass, check the Use Manual Heading box and enter the heading of the ADCP relative to forward or relative to true north. Click Next.

    Review the commands that will be sent to the ADCP. See the WorkHorse Commands and Output Data Format Guide for details on each command.

  • Horizontal ADCP Quick Start Guide

    page 28 Teledyne RD Instruments

    Start Data Collection Click Finish to begin data collection.

    See the SurfaceView Users Guide for details on the displays.

  • Horizontal ADCP Quick Start Guide

    P/N 957-6210-00 (December 2008) page 29

    Section 5.0: Data Playback with SurfaceView This section has simple instructions for a data playback of Horizontal ADCP current profile data.

    Start SurfaceView by double clicking the SurfaceView icon. On the Start Wizard, click Playback / Processing. Click Next.

    Select the Data File On the Playback File Selection screen select the file to be played by clicking the Browse button and click the Open button. Click Next.

  • Horizontal ADCP Quick Start Guide

    page 30 Teledyne RD Instruments

    Select the Bin Selection by clicking the Change Bin Selection box. Enter the bins you want and click OK. Click Next.

    Click the Browse button and select a path to save the Peak/Mean Data Log-ging file (*.CSV). CSV files can not be overwritten. If you see a warning mes-sage, click OK and select a different path or file name. Click Next.

    Start Playback Click Finish to begin data playback.

  • Horizontal ADCP Quick Start Guide

    P/N 957-6210-00 (December 2008) page 31

    UNOTES

  • Horizontal ADCP Quick Start Guide

    page 32 Teledyne RD Instruments

    UNOTES

    Horizontal ADCP Quick Start GuideIntroductionHow to Contact Teledyne RD InstrumentsWhat This Guide Covers

    Section 1.0: OverviewPacking ListHardware OverviewPower OverviewSerial Communication OverviewRS-232 to RS-422 Converter OverviewSoftware OverviewInstalling the Software

    Section 2.0: Bench TestingVisual InspectionSetup the H-ADCPChanging the Baud Rate in the H-ADCPTesting the H-ADCPCompass AlignmentPreparing for CalibrationCompass Calibration VerificationCompass Calibration Procedure

    Section 3.0: Installation and SetupH-ADCP PreparationPressure Sensor PreparationSite PreparationH-ADCP InstallationTesting the Installed H-ADCP

    Section 4.0: Collecting Current Profile Data with SurfaceViewSection 5.0: Data Playback with SurfaceView