HA Collogen Composite AC

download HA Collogen Composite AC

of 12

Transcript of HA Collogen Composite AC

  • 8/13/2019 HA Collogen Composite AC

    1/12

    Wear 262 (2007) 13871398

    Wear characteristic and biocompatibility of somehydroxyapatitecollagen composite acetabular cups

    S.K. Roy Chowdhury a,, A.C. Kulkarni a, A. Basakb, S.K. Roy c

    a Department of Mechanical Engineering, Indian Institute of Technology, Kharagpur, IndiabDepartment of Chemistry, Indian Institute of Technology, Kharagpur, India

    cDepartment of Metallurgical and Materials Engineering, Indian Institute of Technology, Kharagpur, India

    Received 28 March 2006; received in revised form 6 December 2006; accepted 9 January 2007

    Available online 12 February 2007

    Abstract

    Both HDPE and UHMWPE have long been used successfully as socket materials in hip-joint replacements. Recently, however there are

    concerns over the adverse biological responses due to the wear debris of these polymers. Although a good deal of work to improve the performance

    of these polymers has been carried out a need still exists for an implant material with improved biocompatibility and mechanical properties.

    Hydroxyapatitecollagen composites have been prepared by precipitation of calcium phosphate on collagen in the past but very few of these

    attempts considered the mechanical strength of the composites that suits their realistic uses as implant material. Present work is an attempt to

    develop hybrid composites of hydroxyapatitecollagenhyaluronic acid or gelatin with sufficient adherence to both hard and soft tissues and also

    with good cohesive strength leading to improved mechanical and biological properties. It was possible to prepare acetabular cups of the newly

    developed composites by compression moulding for tests on a hip-joint simulator. Pin specimens for tests on a pin-on-disc apparatus were also

    moulded with these composites. Tests with the acetabular cups and pin specimens indicate that some of the newly developed materials offer wear

    resistance comparable to those of the presently used socket materials. Biocompatibility tests with these materials show that their haemolysis counts

    are well below the acceptable range. Hydroxyapatitecollagen composites with 10% hyaluronic acid offer suitable mechanical strengths, good

    friction and wear characteristics and acceptable level of haemolysis and therefore the composite may be considered to be a potential socket material

    of future generation. 2007 Elsevier B.V. All rights reserved.

    Keywords: Hydroxyapatitecollagen composite; Hip-joint; Acetabular cups; Wear; Biocompatibility

    1. Introduction

    Since Sir John Charnley proposed metal to plastic pairing for

    total hip replacements in early 1960s there has been a contin-

    uous search for more suitable materials for the prosthetic pairs

    [1,2].The initial choice of PTFE as a socket material was soon

    abandoned due to its high wear rate despite its extremely low

    frictional resistance. Subsequently, both HDPE and UHMWPE

    have been used successfully as socket materials over the last

    few decades [3,4]. Recently, however, there are concerns regard-

    ing the adverse biological responses due to UHMWPE wear

    debris. A good deal of work to improve the performance and

    biocompatibility of these polymers hasbeen carried out in recent

    Corresponding author. Tel.: +91 3222 282972; fax: +91 3222 282278.

    E-mail address:[email protected](S.K. Roy Chowdhury).

    years [5,6]. This also regenerated interestsin metals and ceramic

    prosthetic pairs. Here too, it was soon realized that long-term

    systematic problems might arise due to release of metal ions

    and increase in cobalt and chromium concentration in blood

    and urine[7].In view of this and also the need for bone aug-

    mentations in several other applications, an acute need for an

    artificial bone substitute with appropriate biocompatibility and

    suitable mechanical properties was felt. Two basic approaches

    were made to obtain such materials. One group of researchers

    considered polymer composites with hydroxyapatite fillers or

    thermally sprayed hydroxyapatite coatings on the metallic pros-

    theses. These attempts were successful in improving the wear

    resistance of metallic implants but the hydroxyapatitepolymer

    composites could not yield the desired compatibility. Wear

    debris at the contact between the prosthetic pairs of these com-

    posites were often found to contain sizeable amountof polymers.

    Another group of researchers considered that a composite of the

    0043-1648/$ see front matter 2007 Elsevier B.V. All rights reserved.

    doi:10.1016/j.wear.2007.01.023

    mailto:[email protected]://localhost/var/www/apps/conversion/tmp/scratch_4/dx.doi.org/10.1016/j.wear.2007.01.023http://localhost/var/www/apps/conversion/tmp/scratch_4/dx.doi.org/10.1016/j.wear.2007.01.023mailto:[email protected]
  • 8/13/2019 HA Collogen Composite AC

    2/12

    1388 S.K. Roy Chowdhury et al. / Wear 262 (2007) 13871398

    two major solid phases of bone, namely the collagen and the cal-

    cium phosphate would be a better choice to obtain near-natural

    bone substitutes[8]. An added incentive for the development

    of such collagencalcium phosphate composite was that the

    collagen has been said to promote bone healing and its regener-

    ation. This in turn promotes new bone growth and an eventual

    replacement by natural bone[9].

    Many attempts were made to develop such composites

    mainly by precipitation of calcium phosphate on collagen

    [8,10]. One of the earliest attempts to develop collagencalcium

    phosphate composite was made by Mittlmier and Nizzard[11]

    who mixed calcium phosphate granules with collagen web.

    Many other researchers attempted to produce these composites

    by mixing the preformed calcium phosphate granules in colla-

    gen suspension[1014]. However, very few of these attempts

    considered the mechanical strength of the composites that suits

    their realistic uses. The main interest there was to reproduce the

    physical properties of bone by nucleation and growth of calcium

    phosphate crystals from solution on collagen fibrils [1518].

    Lawson and Czernuszka [8] produced a collagencalcium phos-phate composite by precipitation method and concluded that the

    mechanical and biological properties of their composite were

    encouraging. However, such composites, in general, lack cohe-

    sive strength and they are not entirely suitable for manufacturing

    load-bearing implants. Attempts have been made to develop

    hybrid composites of hydroxyapatitecollagenhyaluronic

    acid with sufficient adherence to both hard and soft tissues

    and also with good cohesive strength, improved mechani-

    cal and biological properties [18]. The present work is an

    attempt to develop such hybrid composites with improved

    mechanical strengths to the extent that acetabular cups may

    be manufactured for use in total hip-joint replacements. Themechanical, tribological and biological characteristics of the

    cups made of the newly developed composites have also been

    studied.

    2. Materials and methods of preparation

    2.1. Materials

    In view of the present need for a near-natural prosthetic mate-

    rial it was felt that the development of the following groups

    of composites based on hydroxyapatite and collagen would be

    useful.

    2.1.1. Polymerhydroxyapatite composites with varying

    percentages of hydroxyapatite in polymer matrix

    Since HDPE and UHMWPE are widely used as socket

    materials in hip-joint prostheses hydroxyapatite reinforced com-

    posites of these polymers only were considered to be useful.

    The fundamental argument for developing this class of com-

    posites is that hydroxyapaptite being a natural body material

    the above composites would necessarily be more biocompatible

    and if wear debris contain more of filler material less harm to the

    body system is expected. There is some evidence [19] to suggest

    that the wear debris of these composites may contain low level

    of polymer.

    Table 1

    Mechanical properties of hydroxyapatite[8]

    Compressive strength (MPa) 310510

    Tensile strength (MPa) 40

    Vickers hardness (MPa) 4500

    Elastic modulus (GPa) 4090

    Fracture toughness K1c (MPa m1/2)

  • 8/13/2019 HA Collogen Composite AC

    3/12

    S.K. Roy Chowdhury et al. / Wear 262 (2007) 13871398 1389

    Table 3

    Properties of hydroxyapatite-filled polymer composites and the parent polymers

    Specimen materials Tensile strength at break0 (MPa) Elongation at break0(%) HardnessH(MPa) Product parameter00H

    HARHDPE (5%) 23.47 61.82 617.8 896,376

    HARHDPE (10%) 20.7 55.3 657.1 752,189

    HARUHMWPE (5%) 32.6 140 490.5 2,238,642

    HARUHMWPE (10%) 26.57 101.75 578.8 1,564,784

    HDPE 30 450 37 499,500UHMWPE 37 350 50.5 653,975

    Table 4

    Properties of composites using micro-mechanics approach

    Properties HARHDPE (5%) HARHDPE (10%) HARUHMWPE (5%) HARUHMWPE (10%)

    wf 0.05 0.1 0.05 0.1

    wm 0.95 0.9 0.95 0.9

    vf 0.135 0.249 0.135 0.248

    vm 0.864 0.752 0.8644 0.751

    pc (gm/cm3) 0.8556 0.7863 0.8578 0.7865

    Ec(GPa) 12.546 22.097 2.547 22.011

    c 0.3861 0.3755 0.4640 0.4424

    reinforced with 5 and 10% hydroxyapatite by weight designated

    as HARHDPE (5%), HARHDPE (10%), HARUHMWPE (5%)

    and HARUHMWPE (10%), respectively. The tensile strength,

    elongation at break and hardness were measured in an appro-

    priate tensile testing machine and a standard Vickers hardness

    tester respectively with the specimens made of all the four com-

    posites and the parent polymer following the ASTM standard

    [20,21].The results of these tests are given inTable 3.

    The composites developed may be considered to be isotropic

    and it is convenient to obtain some of the other relevant param-

    eters by micromechanical approach using the standard relations[22].The properties so obtained are given in Table 4.

    Here, suffixes m, c and f represent matrix, composite and

    fiber, respectively, and w is the weight fraction, v the volume

    fraction,p the density,Ethe elastic modulus and is the Pois-

    sons ratio. General observation in Tables 3 and 4 is that the

    elongation at break of thecomposites decrease whereas thehard-

    ness rises significantly compared to the parent polymers. The

    changes in tensile strength, Poisons ratio and density are not so

    significant but the elastic constant increases many folds.

    For tribological tests two types of specimens were necessary,

    one for the tests under simulated conditions using a pin-on-disc

    apparatus and the other for a hip-joint simulator under rela-

    tively more realistic in vitro test conditions. The specimens for

    the pin-on-disc tests are simply pins of 8 mm diameter for our

    apparatus and those for the tests on a hip-joint simulator are

    22 mm acetabular cups of standard dimensions. These speci-

    mens were prepared by compression moulding using an existing

    die described elsewhere [23]. The die allowsmoulding of 22 mm

    standard acetabular cups along with a cylindricalextension at the

    bottom part (Figs. 1a and 4).The objective of such a die design

    is to produce specimens for both the hip-joint simulator and the

    pin-on-disc apparatus within a single mould. Hydroxyapatite

    and the polymer matrix materials were mixed in different weightpercentages to makea total mass of 25 gm. This value was calcu-

    lated based on the volume of cup, pin and risers. The premixing

    was done in a Brabender Plasticorder mixing machine. The pres-

    sure and temperature were maintained at 0.6 MPa and 180 C,

    respectively. Thereafter the premix was chopped and blended in

    a mixer grinder. The prepared compound was then poured into

    the mould cavity of the die. A silicon spray was used as a mould-

    releasing agent. Pressure was then exerted on the die placed

    between the heating plates of a compression-moulding machine

    of 200 kN capacity. The load and the temperature were in the

    range of 120140 kN and 160185 C, respectively, depending

    on the composite being moulded. The die was then cooled in

    Fig. 1. (a) A typical moulded acetabular cup of 10% hydroxyapatite-filled UHMWPE composite along with two pins cut out of the main mould. (b) A schematic

    diagram of a typical acetabular cup.

  • 8/13/2019 HA Collogen Composite AC

    4/12

    1390 S.K. Roy Chowdhury et al. / Wear 262 (2007) 13871398

    Table 5

    Dimensions of the moulded UHMWPE cup and a commercially available HDPE cup

    Parent polymer D1(mm) D2(mm) d1(mm) Mass (gm)

    UHMWPE (moulded) 49.42 43.35 22.18 18.9484

    HDPE (commercially available) 49.37 43.33 22.17 19.2764

    Table 6

    Composition and mechanical properties of hydroxyapatitecollagenhyaluronic acid composites

    Serial number Material composition Mouldability Hardness

    (HV/20)

    Elastic modulus

    (MPa)

    Tensile

    strength (MPa)

    1 Hydroxyapatite 90%, collagen 10% Poor: specimens could not be

    prepared

    2 Hydroxyapatite 90%, collagen 9.2%,

    hyaluronic acid 0.8%

    Poor: specimens could not be

    prepared

    3 Hydroxyapatite 80%, collagen 15%,

    hyaluronic acid 5%

    Acetabular cup and pins prepared 90 100 96

    4 Hydroxyapatite 75%, collagen 15%,

    hyaluronic acid 10%

    Acetabular cup and pins prepared 117 350 25

    5 Hydroxyapatite 90%, collagen 9.2%,

    hyaluronic acid 0.8%, hydroxyapatitecollagenhyaluronic acid composite

    60%, gelatin 40%

    Acetabular cup and pins prepared 100 290 19

    Symbol () in the table indicates that measurement could not be made due to lack of mouldability.

    air for 3 h and finally the cup was ejected out of the mould

    cavity. The pin specimens were then cut from the cup ends. A

    typical moulded acetabular cup of a hydroxyapatite reinforced

    UHMWPE composite along with two pins cut out of the main

    mould is shown in Fig. 1(a). A schematic diagram of a typi-

    cal acetabular cup is shown in Fig. 1(b) and the dimensions of

    a moulded UHMWPE specimen are compared with those of a

    commercially available HDPE specimen in Table 5.It can beseen that the dimensions matches reasonably well and the head

    of a standard femoral implant fitted well into the hemispherical

    cavity of a cup.

    The preparation procedure for the specimens of

    HARUHMWPE composites was almost the same as that

    for HARHDPE except that UHMWPE was available in pow-

    dered form rather than in granular form and the premixing

    was not necessary. The pressure applied during compression

    molding was maintained again at 0.6 MPa, but slightly higher

    temperature of 200 C was necessary.

    2.2.2. Preparation of hydroxyapatitecollagen composites

    Hydroxyapatitecollagen composites have been prepared bymany workers in different forms [8,10,1315]. In our attempt

    hydroxyapatite particles were gradually added to deionized

    water and intensively mixed. Separately, the dispersion of very

    fine collagen fibers in deionized water was also prepared. Nine

    parts by weight of hydroxyapatite-in-dispersion and one part

    by weight of collagen-in-dispersion were mixed intensively in

    a magnetic stirrer for 24 h to form the complex precipitate. The

    precipitate was filtered and dried at a temperature of 37 C for

    72 h in a Petri dish. This forms the hydroxyapatitecollagen

    composite. However, attempt thereafter to produce acetabular

    cup and pin specimen by compression molding was not suc-

    cessful. This was because the required degree of cross-linking

    of collagen was not achieved [15]. The hydroxyapatitecollagen

    composite was synthesized further using hyaluronic acid which

    is known to provide binding capabilities. The samples con-

    taining hyaluronic acid was prepared by adding nine parts

    by weight of inorganic component, viz. hydroxyapatite-in-

    dispersion and one part by weight of organic component that

    contains 92% by weight of bovine collagen type-1 and 8% by

    weight of hyaluronic acid. Samples with increasing percentageof hyaluronic acid were prepared by intensive mixing of the two

    phases in a mechanical shaker. The samples with gelatin as a

    binderwere also prepared by adding 40%by weightgelatin pow-

    der and 60% by weight of the above composite with hyaluronic

    acid. The detail composition of all the composites prepared and

    their mechanical properties are shown in Table 6. The tensile

    strengths and elastic moduli of the samples were determined

    in a suitable tensile testing machine and the Vickers hardness

    number was obtained in a standard hardness tester using 2 kgf

    load.

    The scanning electron micrographs of some of the compos-

    ites are shown inFigs. 2 and 3and they are useful in studying

    the morphology of the composites. The micrographs show thatthe particles of hydroxyapatite are anchored in the complex of

    biopolymer matrix and a compact block structure is formed. At

    higher magnifications the microphotographs show very intimate

    contact of hydroxyapatite granules with the collagenhyaluronic

    acid complex. The particles are seen to be completely covered

    with the film of biopolymer conjugate and the structure of the

    material is rather dense.

    The micrographs of the other developed composites, such as

    hydroxyapatitecollagen composites with 5 and 10% hyaluronic

    acid show certain cross-linking between hydroxyapatite and

    collagen (Fig. 3).Acetabular cups were prepared by compres-

    sion moulding following the procedure described in Section

  • 8/13/2019 HA Collogen Composite AC

    5/12

    S.K. Roy Chowdhury et al. / Wear 262 (2007) 13871398 1391

    Fig. 2. Scanning electron micrographs of hydroxyapatitecollagenhyaluronic acid composite at serial number 2 in Table 6with magnifications: (a) 1000 and (b)

    5000.

    Fig. 3. Scanning electron micrographs of: (a) hydroxyapatitecollagenhyaluronic acid composite at serial number 3 in Table 6and (b) hydroxyapatitecollagen

    hyaluronic acid composite at serial number 4 in Table 6.

    Table 7

    Processing parameters for moulding acetabuylar cups of hydroxyapatitecollagenhyaluronic acid composites shown in Table 6

    Material composition Temperature (C) Load (kN) Processing time (min)

    Hydroxyapatite 90%, collagen 10% 185 110 15

    Hydroxyapatite 80%, collagen 15%, hyaluronic acid 5% 200 115 20

    Hydroxyapatite 75%, collagen 15%, hyaluronic acid 10% 210 115 25

    2.2.1. The processing parameters for moulding are giveninTable 7.

    Two such moulded specimens are shown inFig. 4.Attempts

    to prepare the moulded specimen with the composite at serial

    number l inTable 6(hydroxyapatite, 90%; collagen, 10%) were

    not very successful. The mould was of brittle nature and could

    not sustain sufficient compressive load for the purpose it was

    prepared. However, the specimens with hydroxyapatitecollgen

    composites with hyaluronic acid were suitable for the tests under

    the simulated conditions.

    2.2.3. Preparation of polymerhydroxyapatitecollagencomposites

    Attempts were made to develop composites with varying per-

    centage of hydroxyapatite and collagen in UHMWPE matrix. It

    was observed that mouldability andnecessary strengths could be

    obtained only with fairly large percentage of polymer content.

    Two such polymers were prepared and the specimenswere made

    of them. The processing parameters for moulding the compos-

    ites and their mechanical properties are given in Table 8and the

    dimensions of two typical pin specimens are given inTable 9.

    Fig. 4. The integral moulds for the acetabular cup and pin specimens for tests on a hip-joint simulator and a pin-on-disc apparatus. The mould materials are

    hydroxyapatitecollagen composites with (a) 5% hyaluronic acid and (b) 10% hyaluronic acid.

  • 8/13/2019 HA Collogen Composite AC

    6/12

    1392 S.K. Roy Chowdhury et al. / Wear 262 (2007) 13871398

    Table 8

    Processing parameters for moulding the polymerhydroxyapatitecolagen composites and their mechanical properties

    Material composition UHMWPE (50%)hydroxyapatite

    (40%)collagen (10%)

    UHMWPE (70%)hydroxyapatite

    (23%)collagen (7%)

    UHMWPE

    Mould processing temperature (C) 185 165 185

    Mould processing load (kN) 110 115 105

    Mould processing time (min) 1520 1520 1520

    Tensile strength (MPa) 11 17 35Youngs modulus (MPa) 110 170 500

    Hardness (MPa) 17 20 40

    Table 9

    Dimensions and weights of two typical pin specimens of the composites in Table 8

    Material composition Dimensions (mm) Weight (gm)

    DiameterD LengthL

    50% UHMWPEhydroxyapatitecollagen 9.2 23.2 1.74514

    70% UHMWPEhydroxyapatitecollagen 8.35 22.7 1.256

    3. Friction and wear tests

    As discussed earlier friction and wear tests of the new mate-

    rials were carried out using a pin-on-disc apparatus and also

    a hip-joint simulator. The tests are described in the following

    sections.

    3.1. Friction and wear tests on a pin-on-disc apparatus

    A commercially available standard pin-on-disc apparatus

    (Ducom, India TR-20) with a surgical grade stainless steel disc

    of 230 mm diameter was used. The roughness of the disc surface

    was maintained within a range of 0.030.05m cla by a com-bination of fine grinding and lapping. The pin specimens were

    cut out from the main mould as demonstrated inFig. 1(a). The

    pin ends were flattened for conformity and then cleaned in an

    ultrasonic bath. The lubricant used was a solution of 1.5 gm of

    carboxymethyl cellulose powder in 150 ml of distilled water.

    Although bovine serum would be a better substitute for the

    synovial fluid the carboxymethyle solution was used for the sim-

    ilarity of its rheological properties with those of synovial fluid

    [24]. The pin specimens were soaked in the carboxymethyle

    solution for 48 h and then dipped in 4% formaldehyde solution

    for 24 h for sterilisation. The disc was also cleaned thoroughly

    before carrying out the experiments. Some experiments with

    bovine serum as a lubricant were also carried out and the resultsare compared in Section3.1.2.

    Tests for considerably long time were carried out under con-

    stant loads with wear measurements made at regular intervals.

    It is generally assumed that the load is distributed over the artic-

    ular contact between the cup and the femoral implant as shown

    inFig. 5[25].In order to estimate the appropriate load for the

    pin-on-disc apparatus we refer to the load on the femoral head.

    In our experiments on a hip-joint simulator muscle forces were

    applied by equivalent springs and the resultant loadPwas of the

    order of 2287 N over the contact region.

    The maximum Hertzian contact pressure here is given by

    Pmax= 3P/(2a2

    ) where a is the Hertzian contact radius and

    P is the load. The contact radius a can be obtained using thevalues of elastic constants and the radii of the cups prepared

    (Tables 4, 6 and 8). Matching the calculated maximum contact

    pressurePmaxto the pin-on-disc configuration the pin load may

    be obtained. Based on this pin-on-disc tests were conducted over

    the load range between 10 and 70 N. The tests were carried out

    at an exaggerated sliding velocity of 0.24 m/s in order to com-

    pare the friction and wear characteristics of the newly developed

    composites under rigorous conditions[23].

    3.1.1. Friction and wear tests on a pin-on-disc apparatus

    with pins made of hydroxyapatite-filled polymer composites

    Two sets of tests with pins made of hydroxyapatite-filledpolymer composites (Table 3)were carried out in the presence

    of carboxymethyl cellulose solution. In each set two pin load-

    ings of 20 and 30 N were used. In the first set of experiments

    the wear tests for the HARUHMWPE with 5 and 10% rein-

    forcements were carried out over 100 min at a constant sliding

    velocity of 0.24 m/s. Wear was measured by weight loss method

    using a microbalance after each 20 min interval. Wear volumes

    in mm3, calculated using the densities in Table 4 are plotted

    against sliding distance inFig. 6for ease of comparison with

    other results.

    Fig. 5. Load distribution on a femoral implant head during articulation.

  • 8/13/2019 HA Collogen Composite AC

    7/12

    S.K. Roy Chowdhury et al. / Wear 262 (2007) 13871398 1393

    Fig. 6. Plots of wear volume in mm3 against sliding distance for hydroxyapatite

    reinforced UHMWPE composites at a constant sliding velocity of 0.24 m/s and

    under two different loading conditions on a pin-on-disc apparatus.

    Fig. 7. Plotsof wear volume inmm3 against loadat a constantslidingvelocity of

    0.24 m/s for hydroxyapatite reinforced UHMWPE composite on a pin-on-disc

    apparatus.

    The second set of experiments were carried out with

    HARUHMWPE of 5, 10 and 30% hydroxyapatite reinforce-

    ments and also with an unfilled UHMWPE under pin loadings of

    10, 30, 40, 50 and 70N at a constant sliding velocity of 0.24m/s

    for 20 min. Wear and the steady value of frictional force during

    sliding are plotted against load inFigs. 7 and 8,respectively.

    Fig. 8. Variation of coefficient of friction with normal load for hydroxyapatite

    reinforced UHMWPE composites on a pin-on-disc apparatus.

    3.1.2. Friction and wear tests on a pin-on-disc apparatus

    with pins made of hydroxyapatitecollagen-hyaluronic acid

    composites

    Pin specimens of only three composites in this class could be

    prepared as shown inTable 6and the friction and wear tests of

    these pins with both carboxymethyl solution and bovine serumas lubricants were carried out on the pin-on-disc apparatus at a

    constant sliding velocity of 0.24 m/s and under a load of 20 N.

    Variationsof wear volume andcoefficientof friction with sliding

    distance are shown inFigs. 9 and 10,respectively.

    A 6% by weight bovine serum solution in distilled water

    was prepared and a 2% by weight of sodium azide in distilled

    water was added as an antibacterial agent. Tests with this bovine

    serum solution as a lubricant were conducted under conditions

    identical to those with carboxymethyl solution as a lubricant and

    the results are included in Figs. 9 and 10. In addition results with

    UHMWPE pins are also shown for comparison.

    In order to see the effect of change in sliding velocity on

    the coefficient of friction some tests at sliding velocity rangingbetween 0.1 and 1 m/s were also carried out. Plots of coef-

    ficient of friction against sliding velocity for the developed

    hydroxyapatitecollagenhyaluronic acid composites under a

    constant load of 20 N with carboxymethyl solution as a lubricant

    is shown inFig. 11.

    3.1.3. Wear tests on a pin-on-disc apparatus with pins

    made of UHMWPEhydroxyapatitecollagen composites

    Plots of wear volume against sliding distance for the two

    UHMWPEhydroxyapatitecollagen composites (Table 8)are

    Fig. 9. Plots of wear volume in mm3 against sliding distance of the hydroxyapatitecollagenhyaluronicacid composites lubricated withboth carboxymethyl solution

    and bovine serum solution on a pin-on-disc apparatus.

  • 8/13/2019 HA Collogen Composite AC

    8/12

    1394 S.K. Roy Chowdhury et al. / Wear 262 (2007) 13871398

    Fig. 10. Plots of coefficient of friction against sliding distance of the hydroxyapatitecollagenhyaluronic acid composites lubricated with carboxymethyl solution

    and bovine serum solution on a pin-on-disc apparatus.

    Fig. 11. Plots of coefficient of friction against sliding velocity for the

    hydroxyapatitecollagenhyaluronic acid composites lubricated with car-

    boxymethyl solution as a lubricant on a pin-on-disc apparatus.

    shown inFig. 12.In these composites the UHMWPE contents

    were reduced to only 50 and 70% compared to the earlier com-

    posites (Table 3) with UHMWPE contents of 90% or more.These tests were carried out essentially to see the effect of the

    presence of both hydroxyapatite and collagen in the polymer

    matrix. These plots are also compared with UHMWPE results

    inFig. 12.

    3.2. Wear tests on a hip-joint simulator

    Some tests with the acetabular cups made of the newly

    developed composites were also carried out on an existing hip-

    Fig. 12. Plots of wear volume against sliding distance for the polymerhydro-

    xyapatitecollagen composites lubricatedwith carboxymethyl solutionon a pin-

    on-disc apparatus.

    joint-simulator described in details elsewhere [23]. Attempts

    were made to reproduce the real situation in the simulator by

    allowing two femurs with standard stainless steel implants to

    articulate simultaneously within two standard 22 mm diameteracetabular cups fitted within the cavities on either side of a

    pelvis. The pelvis is fixed in a structure but it is allowed to

    rotate about superiorinferior axis and the appropriate motion

    of the femurs and loading at joint was provided by a suitable

    arrangement using pneumatic cylinders. In order to test the

    newly developed cups two new cups were fitted within the

    pelvic sockets after soaking and cleaning in the stipulated

    manner described in Section 3.1. Tests with each pair were

    conducted for up to 12,000 cycles at a rate of 25 cycles per

    minute. The wear was measured, as before, by weight loss

    method using a precision balance at an interval of 1000 cycles.

    The load on the joints was maintained by a pressure regulator

    connected to the pneumatic cylinder at the loading platform.

    Efforts were made to determine the relative merits of the

    developed composites in terms of their wear resistance.

    Wear tests with only a few specimens could be carried

    out on the simulators due to certain restrictions of the sim-

    ulator and specimens. The results of hydroxyapatite-filled

    UHMWPE (Table 3) are shown in Fig. 13 and those with

    Fig.13. Variationof wearvolume against numberof walkingcyclesin a hip-joint

    simulator for the hydroxyapatite-filled UHMWPE composite.

  • 8/13/2019 HA Collogen Composite AC

    9/12

    S.K. Roy Chowdhury et al. / Wear 262 (2007) 13871398 1395

    Fig. 14. Plots of wear volume against number of walking cycles in a hip-joint

    simulator for UHMWPEhydroxyapatitecollagen composites.

    UHMWPEhydroxyapatitecollagen composites (Table 8)are

    given inFig. 14.

    4. Biocompatibility tests

    Biocompatibility refers essentially to the compatibility of

    materials with the biological systems. Since it is rarely pos-sible to find a fully biocompatible material, it is necessary to

    identify the materials, which are physiologically tolerable. In

    the current trend of biocompatibility studies, aspects of both

    biosafety and biofunctionality are considered [29]. Biosafety

    tests such as test on cytotoxicity and mutagenesis or car-

    cinogenesis are aimed at excluding the severe harmful effects

    of biomaterials on organisms. Among many the haemolysis

    count is an important parameter in testing the biocompatibil-

    ity of materials. In the present work, in vitro biocompatibility

    tests were carried out broadly in line with the ASTM stan-

    dards and the extent of haemolysis of the newly developed

    implant materials were determined. Haemolysis indicates pre-

    mature destruction of red blood cells when they come in contactwith water or other foreign elements. Haemolysis percentage is

    given by

    haemolysis(%) =OD (test) OD (negative)

    OD (positive) OD (negative) 100

    where the optical density OD is basically the light absorbency

    and this is given by log (Io/I),IoandIbeing the intensities of the

    original and transmitted lights respectively. The optical density

    was measured in a standard UV/vis spectrometer. The haemoly-

    sis counts of the developed composites were obtained following

    Fig. 15. Haemolysis counts of the developed composites.

    the standard procedure[23]using goats blood. The results are

    shown inFig. 15.

    5. Discussions

    The three groups of composites prepared and tested are: (a)

    polymerhydroxyapatite composites with varying percentage ofhydroxyapatite in polymer matrix, (b) hydroxyapatitecollagen

    composites with varying percentage of hyaluronic acid or

    gelatin and (c) polymerhydroxyapatitecollagen composites.

    The method of preparation of the composites and specimens are

    detailed in Section2 and the friction and wear test results are

    given in Section3. The biocompatibihty results are described

    in Section 4. All the results are compared with those of our

    moulded UHMWPE specimens. We now discuss the results in

    sequence.

    Figs. 68show the friction and wear tests results of the com-

    posites in group (a) above and it can be seen in Fig. 7 that

    the wear resistance of the hydroxyapatite reinforced UHMWPE

    composites arenoticeably higher than that of theparent polymer.Furthermore, the wear resistance of the composites increases

    to an extent with the increase in hydroxyapatite content. The

    increase in wearresistancecan be explainedusing Ratners equa-

    tion which is widely used for predicting polymer wear with some

    degree of success. The equation is given by:

    V =wx

    00H

    whereVis the wear volume, wthe normal load andxis the slid-

    ing distance. Table 3 shows that the product parameter00Hof

    the composites are much higher than that of the parent polymer

    and also the parameter increases with the increase in hydrox-yapatite content. Since the product parameter appears in the

    denominator in the equation the experimental observation that

    thewear resistance rises with theincreasein hydroxyapatite con-

    tent is justified. FromFigs. 7 and 8,it can be seen that although

    HARUHMWPE (30%) is most wear resistant its friction level

    is higher than the other composites. It therefore seems that

    HARUHMWPE (10%) offers an optimum tribo-performance

    within this group of composites. However, it will be shown later

    that the suitability of a new composite as a possible implant

    material cannot be judged by its friction and wear characteristics

    alone.

    Friction and wear test results of the composites in group (b)

    above with both carboxylmethyl solution and bovine serum aslubricants are shown inFigs. 911.For all, the new materials

    wear rises initially with sliding distance and then it levels off.

    This is a typical behaviour of polymer wear. Hydroxyapatite

    layer over the rough counterface is subjected to a small-scale

    fracture process, which again is a common mechanism of wear

    of brittle polycrystalline materials[26].Wear process here may

    be considered to be a sequence of steps involving deformation

    and fracture of hydroxyapatite layer at the asperity contact

    regions followed by a fresh formation of debris layer which

    gradually builds up to cause a decelerating wear pattern. The

    results also indicate that with a higher percentage of hyaluronic

    acid (10%) the wear is least among the present groups of

  • 8/13/2019 HA Collogen Composite AC

    10/12

    1396 S.K. Roy Chowdhury et al. / Wear 262 (2007) 13871398

    composites. This may be because addition of hyaluronic

    acid increases cohesion within a composite structure and

    consequently the bonding strength rises. Gelatin also offers

    bonding strength but fairly large proportion of gelatin is needed

    to increase the wear resistance of a composite. It can be seen in

    Fig. 9that the composite (serial number 5 inTable 6)with 40%

    gelatin is more wear resistant than the composite with a lower

    percentage (5%) of hyaluronic acid but the composite with 10%

    hyaluronic acid (serial number 4, Table 6) is more wear resistant

    than the composites with 5% hyaluronic acid, 40% gelatin and

    also the unfilled UHMWPE. It can be seen in Fig. 10that the

    coefficient of friction of hydroxyapatitecollagenhyaluronic

    acid composites falls sharply initially and then reaches a steady

    value with the further increase in sliding distance. This again

    is a typical behaviour of polymer friction where transfer layer

    builds up as sliding proceeds.Fig. 11shows that the coefficient

    of friction of all the composites falls initially with increasing

    sliding velocity and then reaches a steady value. The figure also

    shows that the friction level of the hydroxyapatitecollagen

    composite with 10% hyaluronic acid is the lowest among thecomposites tested. Considering that the composite is most wear

    resistant among the present sets of new materials developed

    and that its biocompatibility is likely to be of very high order

    due to the presence of both hydroxyapatite and collagen it is

    certainly a strong competitor as a future implant material The

    results in Fig. 9 also show clearly that the wear levels of the

    composites are higher in the presence of bovine serum solution

    than carboxymethyl solution as a lubricant. It was shown by

    several authors that its protein content plays a key role in

    wear results. Scholes and Unsworth[27]observed that varying

    degree of protein might be adsorbed on the surfaces of ceramic

    and polymer specimens. They concluded that the adsorbedfilm of protein influenced the lubricating condition and wear

    of the artificial joints. Another observation in Fig. 10 is that

    the coefficient of friction is lower in the presence of bovine

    serum than carboxymethyl solution as a lubricant. This is again

    because the adsorbed protein layer on the metallic counterface

    reduces the transfer layer formation and the counterface

    roughness is retained to some extent leading to high wear level

    but low friction. McKellop et al. [28] using a reciprocating

    pin-on-plate rig investigated the effects of different lubricants on

    the frictional resistance between UHMWPE and 316L stainless

    steel. In the presence of bovine blood serum, the coefficient of

    friction between the two surfaces (= 0.12) was lower than that

    in the presence of either distilled water (= 0.18) or Ringerssolution (= 0.27). Heavy transfer film on the metal surface

    was reported with Ringers solution and distilled water.

    Fig. 12 shows the variation of wear volume against

    sliding distance for the composites in group (c) above

    (polymerhydroxyapatitecollagen composites) lubricated with

    carboxymethyl solution on a pin-on-disc apparatus. The

    wear levels of both the composites are seen to be higher

    than the parent polymer. These two composites were devel-

    oped primarily to compare their performance with that of

    hydroxyapatite-filled UHMWPE composites without any col-

    lagen (Figs. 6 and 7) because it was felt that if the wear

    performance of polymerhydroxyapatitecollagen composites

    compared well with those of hydroxyapatite-filled UHMWPE

    then the former would score high as implant material because

    of its improved biocompatibility. However, the results inFig. 12

    did not support the hypothesis.

    The wear test results for the composites in group (a)

    on the hip-joint-simulator shown in Fig. 13 indicate that

    HARUHMWPE-30% (Tables 3 and 4) is most wear resis-

    tant and this is in agreement with the pin-on-disc results.

    However, since the pin-on-disc results also indicate that the

    coefficient of friction of HARUHMWPE-30% is high these

    results confirm the previous conclusion that HARUHMWPE-

    10% offers the best tribo-performance among the composites

    in group (a). Wear test results for the composites in group (c)

    (Fig. 14) show that none of the composites is suitable from

    the wear point of view although they may have enhanced bio-

    compatibility. Again this observation is in agreement with the

    pin-on-disc results inFig. 12.The results inFig. 15are encour-

    aging for all the hydroxyapatitecollagen based composites

    because the accepted norm is that if the haemolysis percentage

    is less than 10 the test material is taken as haemocompati-ble and if it is less than 5 the material is considered to be

    highly haemocompatible. It can be seen in Fig. 15that all the

    hydroxyapatitecollagen based composites are haemocompat-

    ible. However, the haemocompatibility of the composites fell

    with the increasing percentage of hyaluronic acid. The haemo-

    compatibility also fell in the presence of gelatin and it seems

    that gelatin affects the haemocompatibility more adversely than

    hyaluronic acid. This probably indicates that haemocompati-

    bility arises essentially due to the presence of the two major

    solid phases of bone, namely the collagen and the calcium phos-

    phate and the presence of any other binding element affects the

    haemocompatibility adversely.Wear factors of some of the newly developed composites as

    tested on a pin-on-disc apparatus and on a hip-joint simulator

    along with their haemolysis counts are given inTable 10.

    First observation to be made here is that the wear factors

    of the in-house moulded UHMWPE specimens tested both

    on a pin-on-disc and a hip-joint simulator are higher than

    the range of values (0.8E06 to 1.1E06mm3/Nm) reported

    in literatures [3032]. It must be emphasised here that the

    UHMWPE specimens were moulded with ultra-high molecular

    weight polyethylene powders (Aldrich Chemical Company,

    Inc., India; density: 0.84) in the same manner as the specimens

    for the new composites. This was done for consistency in the

    method of manufacturing of all the specimens. However, someinaccuracy in our moulding procedure is expected and therefore

    the friction and wear characteristics of these specimens are not

    likely to match those of the commercially available UHMWPE

    specimens. Nevertheless, the present results are believed to

    be quite satisfactory for a comparative assessment of the new

    composites with reference to the parent polymer. The second

    important observation here is that the wear factors of the new

    composites lie within the range of 4.15E05 to 1.80E03 for

    the pin-on-disc tests and 2.67E5 to 1.70E3 for the hip-joint

    simulator. Within this range the hydroxyapatite reinforced

    UHMWPE composites (group (a), Tables 3 and 4) seem to

    be relatively more wear resistant. However, these composites

  • 8/13/2019 HA Collogen Composite AC

    11/12

    S.K. Roy Chowdhury et al. / Wear 262 (2007) 13871398 1397

    Table 10

    Wear factors and haemolysis counts of some of the composites developed

    Composites Wear factor (mm3/Nm) Haemolysis count

    Pin-on-disca tests Hip-jointa simulator tests

    HARUHMWPE (5%) 2.02E04 4.68E05

    HARUHMWPE (10%) 6.58E05 2.67E05

    HARUHMWPE (30%) 9.42E05 1.43E05 Hydroxyapatitecollagen composite with no

    binding agent

    Poor mouldability; specimens

    could not be prepared

    2.2

    Hydroxyapatitecollagen composite with

    0.8% hyaluronic acid.

    Poor mouldability; specimens

    could not be prepared

    3.8

    Hydroxyapatitecollagen composite with

    5% hyaluronic acid.

    2.06E04 5.0

    Hydroxyapatitecollagen composite with

    10% hyaluronic acid.

    4.15E05 6.5

    Hydroxyapatitecollagen composite with

    gelatin.

    1.41E04 9.0

    50%UHMWPEhydroxyapatitecollagen

    composite

    1.80E03 1.70E03

    70%UHMWPEhydroxyapatitecollagen

    composite

    1.09E04 1.03E03

    UHMWPE 4.86E04b 5.63E05c 7.0a Tests on both the pin-on-disc and hip-joint simulator tests were earned out with carboxymethyl solution as a lubricant.b Average value fromFigs. 7, 9 and 12.c Average value fromFigs. 13 and 14.

    are not likely to have biocompatibility of high order due

    to the large polymer content. The hydroxyapatitecollagen

    composites with varying percentage of hyaluronic acid and

    gelatin (group (b)) probably come next in order of their wear

    resistance. Composites in group (c) do not offer any meaningful

    wear resistance. If we consider both the wear resistance and

    biocompatibility hydroxyapatitecollagen composite with

    10% hyaluronic acid probably offers the best combination.

    Although the hydroxyapatitecollagen composites with no

    binding agent are extremely biocompatible they do not offer

    sufficient mechanical strength to be used as a socket material.

    Finally, considering both friction and wear results along with

    the haemolysis count it seems that for higher biocompatibil-

    ity some compromise on the wear performance of materials is

    necessary.

    6. Conclusions

    A number of hydroxyapatitecollagen based compositeshave

    been developed using a chemical synthesis route and their

    mechanical properties, such as hardness, tensile strength andmouldability could be improved when they are synthesized fur-

    ther using hyaluronic acid or gelatin. It was possible to prepare

    acetabular cups of some of the new composites by compres-

    sion moulding for tests on a hip-joint simulator. Pin specimens

    for pin-on-disc tests were also prepared. Tests with both the

    acetabular cups and pin specimens indicate that some of the

    newly developed materials offer wear resistance comparable to

    moulded UHMWPE specimens. Following standard test meth-

    ods the biocompatibilities of the composites were determined

    and the haemolysis counts for all the materials were well below

    the acceptable range of 10. Considering both the mechanical and

    tribological characteristics hydroxyapatitecollagencomposites

    with 10% hyaluronic acid seems to be a potential candidate for

    implant materials of future generation with higher biocompati-

    bility. Finally, it must be mentioned that this is the first attempt

    to establish the possible future use of the new composites as

    socket materials and certainly more work is needed to bring it

    to the stage of commercial acceptance.

    References

    [1] J. Charnley, Anchorage of femoral head prosthesis to shaft of femur, J.

    Bone Joint Surg. B 42 (1960) 28.

    [2] J. Charnley, Arthroplasty of the hip: a new operation, Lancet 1 (1) (1961)

    11291132.

    [3] K.J. Brown, J.R. Atkinson, D. Dowson, V. Wright, The wear of ultra high

    molecular weight polyethylene and a preliminary study of its relation to

    the in vivo behaviour of replacement hip joints, Wear 4 (1976) 255264.

    [4] H. Marrs, D.C. Barton, R.A. Jones, I.M. Ward, J. Fisher, Comparative wear

    under four different tribological conditions of acetylene enhanced cross-

    linked ultra-high molecularweightpolyethylene, J. Mater.Sci. Mater.Med.

    10 (1999) 333342.

    [5] J. Fisher, Quantitative analysis of wear debris from UHMWPE that has

    and has not been sterilised by gamma irradiation, J. Bone Joint Surg. B 80(1998) 340344.

    [6] S.M. Kurtz, O.K. Muratoglu, M. Evans, A.A. Edidin, Advances in the

    processing, sterilization and crosslinking of ultra-high molecular weight

    polyethylene for total hip joint arthroplasty, Biomaterials 20 (1999)

    16591688.

    [7] A.H. Schaffer, A. Pilger, H.W. Ruediger, Increased blood cobalt and

    chromium after total hip joint replacement, J. Toxicol. Clin. Toxicol. 37

    (7) (1999) 839.

    [8] A.C. Lawson, J.T. Czernuszka, Collagen-calcium phosphate composites,

    Proc. Inst. Mech. Eng., Part H 212 (1998) 378390.

    [9] B.D. Katthagen, Bone regeneration with bone substitutes, an animal study,

    Ph.D. Thesis, Flinders University, 1987.

    [10] J.E. Zerwekh, S. Kourosh, R. Scheinberg, T. Kitano, M.L. Edwards, D.

    Shin, D.K. Selby, Fibrillar collagen-biphasic calcium phosphate composite

    as a bone material in a spinal fusion, J. Orthop. Res. 10 (1992) 565572.

  • 8/13/2019 HA Collogen Composite AC

    12/12

    1398 S.K. Roy Chowdhury et al. / Wear 262 (2007) 13871398

    [11] H. Mittlmier, M. Nizzard, Knochenregeneration mit industriell gefertigtem

    collagen-apatite implant. In Osteogenese and knochenwachstum, (Eds.)

    M.H. Hackenbroch, H.J. Reflor, M. Jagaer, 1982 (Theime).

    [12] S.W. Young, W.A. Andrews, H. Muller, B. Constantz, Induction of frac-

    ture healing using fibrous calcium phosphate compositespherulites, Invest.

    Radiol. 2 (5) (1991) 470473.

    [13] A. Roivra, J.R. Amedee, M. Radaud, Colonization of a calcium phosphate/

    elastin-olubilized peptide-collagen composite material by human oesto-

    blast, Biomaterials 17 (1996) 15351540.[14] M. Kazim, J.A. Katowitz, M. Fallon, K.I. Piest, Evaluation of colla-

    gen/hydroxyapatite implant for orbital reconstructive surgery, Ophthal.

    Plast. Reconstr. Surg. 8 (2) (1992) 94108.

    [15] K.S. Ten Huisen, R.I. Martin, M. Klimkiewicz, P.W. Brown, Formation

    and properties of synthetic bone composites; hydroxyapatite-collagen, J.

    Biomed. Mater. Res. 29 (1995) 803810.

    [16] N.J. Mathers, J.T. Czernuszka, Growth of hydroxyapatite on type one col-

    lagen, J. Mater. Sci. Lett. 10 (17) (1991) 992993.

    [17] M. Iijima, Y. Moriwaki, Y. Kuboki, In vitro crystal growth of octacal-

    cium phosphate on type one-collagen fibers, J. Cryst. Growth 137 (1994)

    553560.

    [18] D. Bakos, M. Soldan, I. Hernandez-Fuentes, Hydroxyapatite-

    collagen-hyaluronic acid composite, Biomaterials 20 (1999) 191

    195.

    [19] D. Saha, S.K. Roy Chowdhury, A. Banthia, Characterisation of particulatedebris obtained from hip-joint prosthesis, in: Proceedings of the National

    Conference on Biomaterials, IIT, Kharagpur, India, 2000.

    [20] Annual Book of ASTM Standards, F-732, Practice for reciprocating pin-

    on-flat Evaluation of friction and wear properties of polymeric materials

    for use in total joint prosthesis. Standard guide for measuring and report-

    ing friction coefficient, Designation G115-93, vol. 03.02, May 1993,

    486495.

    [21] Annual Book of ASTM Standards, Standard Test Method for Wear Testing

    with a Pin-on-Disk Apparatus, Designation G99-90, vol. 03-02, January

    1990, pp. 399403.

    [22] C. Zweben, in: C. Zweben, et al. (Eds.), Mechanics of Composite Materi-

    als, Delaware Composites Encyclopedia, Technomic publishing company,

    Lancaster, 1981, pp. 3748.

    [23] S.K. RoyChowdhury, A. Mishra, B. Pradhan, D. Saha, Wear characteristics

    andbiocompatibilityof somepolymer compositeacetabularcups,Wear256

    (2004) 10261036.[24] R.M. Hall, A. Unsworth, B.M. Wroblewski, I.C. Burgess, Frictional char-

    acteristicsof explanted Charnleyhip prosthesis, Wear 175 (1994) 159166.

    [25] D.R. Carter, R. Vasu, W.H. Harris, Stress distribution in the acetabular

    region-IIthe effect of cement thickness and metal backing of total hip

    acetabular component, J. Biomech. 15 (3) (1982) 165170.

    [26] M. Kalin, S. Jahanmir, Lewis K. Ives, Effect of counterface roughness on

    abrasive wear of hydroxyapatite, Wear 52 (2002) 679685.

    [27] S. Scholes, T. Unsworth, Tribology of hard bearing surfaces for use in hip

    prostheses, Wear 152 (1999) 125135.

    [28] H. Mckellop, I. Clark, K. Markolf,H. Amstutz,Frictionand wearproperties

    of polymer, metal and ceramic prosthetic joint materials evaluated on a

    multichannel screeningdevice, J. Biomed. Mater. Res. 15 (1981) 616653.

    [29] C.J. Kirkpatrick, F. Bittinger, M. Wagner, H. Kohler, T.G. Van Kooten,

    C.L. Klein, M. Otto, Current trends in biocompatibility testing, Proc. Inst.

    Mech. Eng., Part H 212 (1998) 7584.[30] V. Saikko, A multidirectional motion pin-on-disc wear test method for

    prosthetic joint materials, J. Biomed. Mater. Res. 41 (1998) 5864.

    [31] T.J. Joyce, A. Unsworth, A multidirectional wear screening device and

    preliminary results of UHMWPE articulating against stainless steel, Bio-

    Med. Mater. Eng. 10 (2000) 241249.

    [32] V. Saikko, A twelve station anatomic hip joint simulator, Proc. Inst. Mech.

    Eng., Part H 219 (2005) 437448.