Guide to Cleaning and Disinfecting PEOPL Respironics Masks ... Files... · Respironics Europe:...
Transcript of Guide to Cleaning and Disinfecting PEOPL Respironics Masks ... Files... · Respironics Europe:...
PEOPLE. PRODUCTS. PROGRAMS.
www.respironics.com
Guide to Cleaning and DisinfectingRespironics Masks
ComfortClassic™
ComfortSelect™
Simplicity™
Total™ Face Mask Profile™ Lite
For multiple patient, institutional use.
RESPIRONICS MASK TEMPERATURE EXPOSURE TIME RESULT NO. CYCLES TESTED NOTES
CLEANING
Clean the mask prior to using any of the disinfection or sterilization methods listed below. Use the same cleaning method for all masks.
• Remove the headgear and spacer, if appropriate.
• Soak the parts in a commercially available enzymatic cleaner (e.g., Klenzyme™).
• Clean the mask with a soft bristle brush in a solution of cool tap water and a commercially available anionic detergent(e.g., Manu Klenz™). Do not use cleaning products that contain conditioners or moisturizers because they will leave a residue.
• Rinse thoroughly under cool running tap water. Air dry.
• If adhering substances have not been adequately removed, replace the mask.
• See options below for disinfecting or sterilizing specific masks.
DISINFECTION OR STERILIZATIONRespironics recommends 4 processes to disinfect or sterilize its masks.
Please follow this grid to determine which process is appropriate for each mask.
DISINFECTION/STERILIZATION PROCESS
Profile™ Lite
Simplicity™
ComfortClassic™
ComfortSelect™
Total™
70˚ C
70˚ C
70˚ C
70˚ C
70˚ C
30 minutes
30 minutes
30 minutes
30 minutes
30 minutes
High level disinfection
High level disinfection
High level disinfection
High level disinfection
High level disinfection
30
30
30
30
10
FOR ALL MASKS:• Inspect the mask after
processing. If any components are damaged, replace the mask.
• If the swivel does not move freely after disinfection, replace the split washer (part #1002720 – pkg. of 5).
• If the swivel does not move freely after disinfection, replace the split washer (part #1002720 – pkg. of 5).
• If the swivel does not move freely after disinfection, replace the sealing ring (part #452027 – pkg. of 5).
• If the swivel does not move freely after disinfection, replace the sealing ring (part #452027 – pkg. of 5).
• Verify function of the flapper valve after processing. If the flapper does not move freely,replace the mask.
RESPIRONICS MASK PASTEURIZATION CIDEX® CIDEX O.P.A. STERRAD® SYSTEM 100
Profile™ Lite Nasal Gel Mask ■ ■ DO NOT USE ■
Simplicity™ Nasal Mask ■ ■ ■ ■
ComfortClassic™ Nasal Mask ■ ■ ■ ■
ComfortSelect™ Nasal Mask ■ ■ ■ ■
Total™ Face Mask ■ ■ ■ ■
Each recommended disinfection or sterilization process is listed here. Please refer to the following grids for processing information on specific masks.
PASTEURIZATION
CIDEX
CIDEX O.P.A.
Profile™ Lite
Simplicity™
ComfortClassic™
ComfortSelect™
Total™
45 minutes
45 minutes
45 minutes
45 minutes
45 minutes
Per manufacturer’sinstruction
Per manufacturer’sinstruction
Per manufacturer’sinstruction
Per manufacturer’sinstruction
Per manufacturer’sinstruction
High level disinfection
High level disinfection
High level disinfection
High level disinfection
High level disinfection
30
30
30
30
10
FOR ALL MASKS:• Inspect the mask after
processing. If any components are damaged, replace the mask.
• Slight discoloration of the maskcushion after processing is normal.
• If the swivel does not move freely after disinfection, replace the split washer (part #1002720 – pkg. of 5).
• If the swivel does not move freely after disinfection, replace the split washer (part #1002720 – pkg. of 5).
• If the swivel does not move freely after disinfection, replace the sealing ring (part #452027 – pkg. of 5).
• If the swivel does not move freely after disinfection, replace the sealing ring (part #452027 – pkg. of 5).
• Verify function of the flapper valve after processing. If the flapper does not move freely,replace the mask.
Simplicity™
ComfortClassic™
ComfortSelect™
Total™
12 minutes
12 minutes
12 minutes
12 minutes
Per manufacturer’sinstruction
Per manufacturer’sinstruction
Per manufacturer’sinstruction
Per manufacturer’sinstruction
High level disinfection
High level disinfection
High level disinfection
High level disinfection
30
30
30
10
FOR ALL MASKS:• Inspect the mask after
processing. If any components are damaged, replace the mask.
• Slight discoloration of the maskcushion after processing is normal.
• If the swivel does not move freely after disinfection, replace the split washer (part #1002720 – pkg. of 5). Thegel spacer is incompatible withCidex O.P.A. DO NOT DISIN-FECT/STERILIZE THE GELSPACER. IT SHOULD BECLEANED ONLY.(part #1002464).
• If the swivel does not move freely after disinfection, replace the split washer (part #1002720 – pkg. of 5).
• If the swivel does not move freely after disinfection, replace the sealing ring (part #452027 – pkg. of 5).
• Verify function of the flapper valve after processing. If the flapper does not move freely,replace the mask.
RESPIRONICS MASK EXPOSURE TIME RINSE RESULT NO. CYCLES TESTED NOTES
RESPIRONICS MASK EXPOSURE TIME RINSE RESULT NO. CYCLES TESTED NOTES
STERRAD SYSTEM 100
1013256 LK 12/19/02Reorder No. 1011138
www.respironics.com
Customer Service: 1-800-345-6443 • 724-387-4000Respironics Europe: +33-(0)1-55-60-19-80Respironics Asia Pacific: +852-234-342-18
©2002 Respironics, Inc. ComfortClassic, ComfortSelect, Profile, Simplicity and Total are trademarks of Respironics, Inc.CIDEX and STERRAD System are registered trademarks of Advanced Sterilization Products.Klenzyme and Manu Klenz are trademarks of Steris Corporation.
Profile Lite™
Simplicity™
ComfortClassic™
ComfortSelect™
Total™
Machine controlled
Machine controlled
Machine controlled
Machine controlled
Machine controlled
Sterilization
Sterilization
Sterilization
Sterilization
Sterilization
30
30
30
30
10
FOR ALL MASKS:• Inspect the mask after
processing. If any components are damaged, replace the mask.
• Slight discoloration of the maskcushion after processing is normal.
• If the swivel does not move freely after disinfection, replace the split washer (part #1002720 – pkg. of 5).
• If the swivel does not move freely after disinfection, replace the split washer (part #1002720 – pkg. of 5).
• If the swivel does not move freely after disinfection, replace the sealing ring (part #452027 – pkg. of 5).
• If the swivel does not move freely after disinfection, replace the sealing ring (part #452027 – pkg. of 5).
• Verify function of the flapper valve after processing. If the flapper does not move freely,replace the mask.
RESPIRONICS MASK EXPOSURE TIME RESULT NO. CYCLES TESTED NOTES
English
Cleaning and Disinfecting the ComfortGelTM Mask
CLEANING
Clean the mask prior to using any of the disinfection or sterilization methods listed below.
• Remove the headgear and spacer, if appropriate.
• Soak the parts in a commercially available enzymatic cleaner (e.g., Klenzyme™).
• Clean the mask with a soft bristle brush in a solution of cool tap water and a commerciallyavailable anionic detergent (e.g., Manu Klenz™). Do not use cleaning products that containconditioners or moisturizers because they will leave a residue.
• Rinse thoroughly under cool running tap water. Air dry.
• If adhering substances have not been adequately removed, replace the mask.
CAUTIONCidex OPA and STERRAD may also be used to disinfect the ComfortGel faceplate and forehead armonly. The Gel cushion and Comfort Flap are not compatible with these disinfection processes and mustbe replaced prior to reuse.
PASTEURIZATION
70o C 30 minutes High level disinfection 30 • Inspect the mask after processing. If any components are damaged, replace the mask.
• Tested for durability at 75o C.
CIDEX 2.4%
45 minutes Per manufacturer’sinstruction
High level disinfection 30 • Inspect the mask after processing. If any components are damaged, replace the mask.
• Slight discoloration of the mask cushion after processing is normal.
TEMPERATURE EXPOSURE TIME RESULT NO. CYCLES TESTED NOTES
RINSEEXPOSURE TIME RESULT NO. CYCLES TESTED NOTES
DISINFECTION/STERILIZATION PROCESS
www.respironics.com
1024727JW 05/17/05
Customer Service: 1-800-345-6443 • 724-387-4000
Respironics Europe: +33-(0)1-55-60-19-80
Respironics Asia Pacific: +852-234-342-18
©2005 Respironics, Inc.
Contact Customer Service to obtain mask disinfecting instructions for ComfortGel (#1026815).
Contact Customer Service to obtain mask disinfecting instructions for other Respironics masks (#1011138).
STERRAD®®
MachineControlled
N/A Sterilization 30 • Inspect the mask after processing. If any components are damaged, replace the mask.
• Slight discoloration of the mask cushion after pro-cessing is normal.
• Mask cushion must be replaced prior to reuse.
RINSEEXPOSURE TIME RESULT NO. CYCLES TESTED NOTES
CIDEX O.P.A.
12 minutes Per manufacturer’sinstruction
High level disinfection 30 • Inspect the mask after processing. If any components are damaged, replace the mask.
• Slight discoloration of the mask cushion after processing is normal.
• Mask cushion must be replaced prior to reuse.
RINSEEXPOSURE TIME RESULT NO. CYCLES TESTED NOTES
CAUTIONCidex OPA and STERRAD may also be used to disinfect the ComfortGel faceplate andforehead arm only. The Gel cushion and Comfort Flap are not compatible with these disinfection processes and must be replaced prior to reuse.
www.respironics.com
Cleaning and Disinfecting the ComfortLiteTM Mask
CLEANINGClean the mask prior to using any of the disinfection or sterilization methods listed below. • Remove the headgear and spacer, if appropriate. • Soak the parts in a commercially available enzymatic cleaner (e.g., Klenzyme™). • Clean the mask with a soft bristle brush in a solution of cool tap water and a commercially
available anionic detergent (e.g., Manu Klenz™). Do not use cleaning products that contain conditioners or moisturizers because they will leave a residue.
• Rinse thoroughly under cool running tap water. Air dry. • If adhering substances have not been adequately removed, replace the mask.
DISINFECTION/STERILIZATION PROCESS
CIDEX®
45 minutes
Per manufacturer’s instruction
High level disinfection
30 • Inspect the mask after processing. If any components are damaged, replace the mask.
• Slight discoloration of the mask cushion after processing is normal.
EXPOSURE TIME RINSE RESULT NO. CYCLES TESTED NOTES
CIDEX® O.P.A.
12 minutes
Per manufacturer’s instruction
High level disinfection
30 • Inspect the mask after processing. If any components are damaged, replace the mask.
• Slight discoloration of the mask cushion after processing is normal.
EXPOSURE TIME RINSE RESULT NO. CYCLES TESTED NOTES
PASTEURIZATION
70˚ C (performance tested
at 75˚ C)
High level disinfection
30 • Inspect the mask after processing. If any components are damaged, replace the mask.
• Slight discoloration of the mask cushion after processing is normal.
TEMPERATURE EXPOSURE TIME RESULT NO. CYCLES TESTED NOTES
30 minutes
1028073JW 11/01/05
Customer Service: 1-800-345-6443 • 724-387-4000Respironics Europe: +33-1-55-60-19-80Respironics Asia Pacific: +81-3-5280-9611
©2005 Respironics, Inc., and its affiliates. All rights reserved.
Contact Customer Service to obtain mask disinfecting instructions for the ComfortLite (#1026813).
Contact Customer Service to obtain mask disinfecting instructions for other Respironics masks (#1011138).
WAVICIDE®
45 minutes
Per manufacturer’s instruction
High level disinfection
30 • Inspect the mask after processing. If any components are damaged, replace the mask.
• Slight discoloration of the mask cushion after processing is normal.
EXPOSURE TIME RINSE RESULT NO. CYCLES TESTED NOTES
CONTROL 3®
10 minutes
Per manufacturer’s instruction
High level disinfection
30 • Inspect the mask after processing. If any components are damaged, replace the mask.
• Slight discoloration of the mask cushion after processing is normal.
EXPOSURE TIME RINSE RESULT NO. CYCLES TESTED NOTES
www.respironics.com
Cleaning and Disinfecting the ComfortLite™ 2 Mask
CLEANINGClean the mask prior to using any of the disinfection or sterilization methods listed below. • Remove the headgear and spacer, if appropriate. • Soak the parts in a commercially available enzymatic cleaner (e.g., Klenzyme™). • Clean the mask with a soft bristle brush in a solution of cool tap water and
a commercially available anionic detergent (e.g., Manu Klenz™). Do not use cleaning products that contain conditioners or moisturizers because they will leave a residue.
• Rinse thoroughly under cool running tap water. Air dry. • If adhering substances have not been adequately removed, replace the mask.
DISINFECTION/STERILIZATION PROCESS
CIDEX® 2.4%
45 minutes
Per manufacturer’s instruction
High level disinfection
30 • Inspect the mask after processing. If any components are damaged, replace the mask.
• Slight discoloration of the mask cushion after processing is normal.
EXPOSURE TIME RINSE RESULT NO. CYCLES TESTED NOTES
CIDEX® O.P.A.
12 minutes
Per manufacturer’s instruction
High level disinfection
30 • Inspect the mask after processing. If any components are damaged, replace the mask.
• Slight discoloration of the mask cushion after processing is normal.
EXPOSURE TIME RINSE RESULT NO. CYCLES TESTED NOTES
PASTEURIZATION
70˚ C (performance tested
at 75˚ C) High level disinfection 30• Inspect the mask after processing. If any components
are damaged, replace the mask.
• Slight discoloration of the mask cushion after processing is normal.
TEMPERATURE EXPOSURE TIME RESULT NO. CYCLES TESTED NOTES
30 minutes
90˚ C 1 minute
1028235JW 01/11/06
Contact Customer Service to obtain mask disinfecting instructions for the ComfortLite 2 (#1032071).
Contact Customer Service to obtain mask disinfecting instructions for other Respironics masks (#1011138).
WAVICIDE®
45 minutes
Per manufacturer’s instruction
High level disinfection
30 • Inspect the mask after processing. If any components are damaged, replace the mask.
• Slight discoloration of the mask cushion after processing is normal.
EXPOSURE TIME RINSE RESULT NO. CYCLES TESTED NOTES
CONTROL 3®
10 minutes
Per manufacturer’s instruction
High level disinfection
30 • Inspect the mask after processing. If any components are damaged, replace the mask.
• Slight discoloration of the mask cushion after processing is normal.
EXPOSURE TIME RINSE RESULT NO. CYCLES TESTED NOTES
Customer Service: 1-800-345-6443 • 724-387-4000Respironics Europe: +33 1 47 52 30 00Respironics Asia Pacific: +81-3-5280-9611
©2006 Respironics, Inc., and its affiliates. All rights reserved.
Cleaning and Disinfecting the ComfortFullTM 2 Mask
CLEANING
Clean the mask prior to using any of the disinfection or sterilization methods listed below.
• Remove the headgear and spacer, if appropriate.
• Soak the parts in a commercially available enzymatic cleaner (e.g., Klenzyme™).
• Clean the mask with a soft bristle brush in a solution of cool tap water and a com-mercially available anionic detergent (e.g., Manu Klenz™). Do not use cleaning prod-ucts that contain conditioners or moisturizers because they will leave a residue.
• Rinse thoroughly under cool running tap water. Air dry.
• If adhering substances have not been adequately removed, replace the mask.
DISINFECTION/STERILIZATION PROCESS
CIDEX 2.4%
45 minutes
Per manufacturer’s instruction
High level disinfection
30 • Inspect the mask after processing. If any components are damaged, replace the mask.
• Slight discoloration of the mask cushion after processing is normal.
RINSEEXPOSURE TIME RESULT NO. CYCLES TESTED NOTES
CIDEX O.P.A.
12 minutes
Per manufacturer’s instruction
High level disinfection
30 • Inspect the mask after processing. If any components are damaged, replace the mask.
• Slight discoloration of the mask cushion after processing is normal.
RINSEEXPOSURE TIME RESULT NO. CYCLES TESTED NOTES
www.respironics.com
PASTEURIZATION
70° C (performance tested
at 75° C) High level disinfection 30• Inspect the mask after processing. If any
components are damaged, replace the mask.
• Slight discoloration of the mask cushion after processing is normal.
TEMPERATURE EXPOSURE TIME RESULT NO. CYCLES TESTED NOTES
30 minutes
90° C 1 minute
1031389JW 12/29/05
WAVICIDE®
45 minutes
Per manufacturer’s instruction
High level disinfection
30 • Inspect the mask after processing. If any components are damaged, replace the mask.
• Slight discoloration of the mask cushion after processing is normal.
RINSEEXPOSURE TIME RESULT NO. CYCLES TESTED NOTES
Hydrogen Gas Plasma Sterilization (STERRAD®)
Machine Controlled
N/A
Sterilization
30 • Inspect the mask after processing. If any components are damaged, replace the mask.
• Slight discoloration of the mask cushion after processing is normal.
RINSEEXPOSURE TIME RESULT NO. CYCLES TESTED NOTES
Contact Customer Service to obtain the most current mask disinfecting instructions for the ComfortFull 2 (#1026814).
Contact Customer Service to obtain mask disinfecting instructions for other Respironics masks (#1011138).
Customer Service: 1-800-345-6443 • 724-387-4000Respironics Europe: +33 1 47 52 30 00Respironics Asia Pacific: +81-3-5280-9611
©2006 Respironics, Inc., and its affiliates. All rights reserved.