General Specifications <<Contents>> <<Index>> Model AM012 Magnetic Flow Converter Yokogawa Electric Corporation 2-9-32 Nakacho, Musashino-shi, Tokyo, 180-8750 Japan Phone: 81-422-52-5690 Fax.: 81-422-52-2018 GS 1E6K2-E GS 1E6K2-E ©Copyright June 1989 (YK) 5th Edition June 2012 (KP) n OUTLINE AMO12 is a signal generator used to check the calibration and operation of AM11 MAGNETIC FLOW CONVERTERS,AM12 MAGNETIC FLOW CONVERTERS,AM100A/200A-INTEGRAL TYPE MAGNETIC FLOWMETER,YMA11 MAGNETIC FLOW CONVERTER, COMPACT YEWMAG MAGNETIC FLOWMETER (style A and style B)and ADMAG AE MAGNETIC FLOWMETER. As the excitation load is build-in, the calibration and operation of a converter can be checked without a detector. Span can be set within the range of flow rates 0 to 10 m/s.Output signal correponding to setting span can be set to 0,25, 50, 75, 100% with the setting dial. n STANDARD SPECIFICATIONS Case: Material:Polyester resin Size:110 (height) x235(width) x364(rear)mm. Color: Light grayish green. Connecting Cable: Power cord for AM012. Signal cable. Excitation cable. Setting Ranges: Span Setting Dial: 0 to 10 m/s (minimum span 0.01 m/s). Output Setting Dial: CHECK, 0, 25, 50, 75, 100%. Accuracy: 0.1 m/s or more: �0.04% of set value. �0.04% of set value. 0.04% of set value. 0.1 m/s or less: Flow rate of �0.04 mm/s. �0.04 mm/s. 0.04 mm/s. Power: 90 to 132 V AC or 180 to 264 V AC, 47 to 63 Hz Ambient Temperature: 0 to 40 Weight: 4.2kg Model Suffix code Description AM012 .................. Calibrator for Magneric Flowmeter Power code -1 .............. -2 .............. -3 .............. 100 V version 100 V version 200 V version Style code *A .......... Style A Optional specification /CYM /AE /SE With Cable for compact YEWMAG With Junctionbox for ADMAG AE and ADMAG CA With Junctionbox for ADMAG SE, and special cable Power Code Power plug is attached as described below. n OPTIONAL SPECIFICATION Cable for Compact YEWMAG For the calibration of Compact YEWMAG. With the connector for signal and excitation cable. Option code /CYM Junctionbox for ADMAG AE and ADMAG CA When using the AM012 calibrator with the ADMAG AE and ADMAG CA, remove only the amplifier from the converter and put it in the dedicated junctionbox for calibration. Use this item when only the junctionbox for AM012 is to be procured. Option code /AE Junctionbox for ADMAG SE When using the AM012 calibrator with the ADMAG SE, remove only the amplifier from the converter and put it in the dedicated junctionbox for calibration. Use this item when only the junctionbox for AM012 is to be procured. Option code /SE n ACCESSORIES Power code.............................................................. 1set Signal cable ............................................................ 1set Excitation cable ...................................................... 1set Junction box (/AE, /SE) ........................................... 1set Cable for ADMAG SE (/SE) ................................ ... 1set Cable for Compact YEWMAG (/CYM) .................... 1set

description

transmitter

Transcript of GS01E06K02-00E_005

  • GeneralSpecifications

    Model AM012Magnetic Flow Converter

    Yokogawa Electric Corporation2-9-32 Nakacho, Musashino-shi, Tokyo, 180-8750 JapanPhone: 81-422-52-5690 Fax.: 81-422-52-2018

    GS 1E6K2-E

    GS 1E6K2-ECopyright June 1989 (YK)5th Edition June 2012 (KP)

    n OUTLINE AMO12 is a signal generator used to check the calibration and operation of AM11 MAGNETIC FLOW CONVERTERS,AM12 MAGNETIC FLOW CONVERTERS,AM100A/200A-INTEGRAL TYPE MAGNETIC FLOWMETER,YMA11 MAGNETIC FLOW CONVERTER, COMPACT YEWMAG MAGNETIC FLOWMETER (style A and style B)and ADMAG AE MAGNETIC FLOWMETER. As the excitation load is build-in, the calibration and operation of a converter can be checked without a detector. Span can be set within the range of flow rates 0 to 10 m/s.Output signal correponding to setting span can be set to 0,25, 50, 75, 100% with the setting dial.

    n STANDARD SPECIFICATIONSCase:Material:Polyester resinSize:110 (height) x235(width) x364(rear)mm.Color: Light grayish green.Connecting Cable: Power cord for AM012.

    Signal cable. Excitation cable.

    Setting Ranges:Span Setting Dial: 0 to 10 m/s (minimum span 0.01 m/s).Output Setting Dial: CHECK, 0, 25, 50, 75, 100%.Accuracy:

    0.1 m/s or more: 0.04% of set value.0.04% of set value.0.04% of set value. 0.1 m/s or less: Flow rate of 0.04 mm/s.0.04 mm/s.0.04 mm/s.

    Power: 90 to 132 V AC or 180 to 264 V AC, 47 to 63 HzAmbient Temperature: 0 to 40Weight: 4.2kg

    Model Suffix code Description

    AM012 .................. Calibrator for Magneric Flowmeter

    Power code-1 ..............-2 ..............-3 ..............

    100 V version100 V version200 V version

    Style code *A .......... Style AA

    Optional specification

    /CYM/AE

    /SE

    With Cable for compact YEWMAGWith Junctionbox for ADMAG AE and ADMAG CAWith Junctionbox for ADMAG SE, and special cable

    Power CodePower plug is attached as described below.

    n OPTIONAL SPECIFICATIONCable for Compact YEWMAGFor the calibration of Compact YEWMAG.With the connector for signal and excitation cable. Option code /CYM

    Junctionbox for ADMAG AE and ADMAG CAWhen using the AM012 calibrator with the ADMAG AE and ADMAG CA, remove only the amplifier from the converter and put it in the dedicated junctionbox for calibration. Use this item when only the junctionbox for AM012 is to be procured. Option code /AE

    Junctionbox for ADMAG SE When using the AM012 calibrator with the ADMAG SE, remove only the amplifier from the converter and put it in the dedicated junctionbox for calibration. Use this item when only the junctionbox for AM012 is to be procured. Option code /SE

    n ACCESSORIESPower code.............................................................. 1setSignal cable ............................................................ 1setExcitation cable ...................................................... 1setJunction box (/AE, /SE) ........................................... 1setCable for ADMAG SE (/SE) ................................ ... 1setCable for Compact YEWMAG (/CYM) .................... 1set

  • 2All Rights Reserved. Copyright 1989, Yokogawa Electric Corporation

    5th Edition June 01, 2012-00 GS 1E6K2-EGS 1E6K2-E

    n EXTERNAL DIMENSIONS

    AM012

    Junction box (For AE100/200 and CA100/200) Junction box (For SE100/200)

    unit : mm

    Subject to change without notice.

    n CABLE CONNECTIONS(1) AE14, AE100/200, AM11, AM12, AM100A/B, AM200A/B, YMA11

    AM012AM11AM12AM100A/B AM200A/B YMA11

    Connector Cable

    Flow rateABC

    ABC

    Excitation currentEx1Ex2G

    Ex1Ex2G

    Power supply Power plug Power outlet

    (2) SE100M/200M

    AM012SE100M/200M

    Connector Cable

    Flow-rate Connector Connector on amplifier

    Excitation current Connector Connector on amplifier

    Power supply Power plug Power outlet

    (3) Compact YEWMAG style A

    AM012Compact YEWMAG

    Connector Cable

    Flow rate

    Connector

    Connector on right side ofamplifier

    Excitation current

    Power supply Power plug Power outlet

    (4) Compact YEWMAG style B

    AM012Compact YEWMAG

    Connector Cable

    Flow-rate ConnectorConnector on left side of amplifier

    Excitation current ConnectorConnector on right side ofamplifier

    Power supply Power plug Power outlet

    = = = = = = ORDERING INSTRUCTIONS = = = = = =

    Model and Suffix codes

    = = = = = = RELATED INSTRUMENTS = = = = = =

    AM11 .................................................. GS 1E6C1-EAE100M/200M .................................... GS 1E7B1-02EAE14 .................................................. GS 1E7A0-01ECA100/200 ......................................... GS 1E8B1-ESE100M/200M ................................... GS 1E10B1-01E