GRABCAD CHALLENGE VA Innovation Creation Series Prosthetic & Assistive Technology Challenge By: Reinaldo Wiener Rocca

Transcript of Grabcad tremors

Page 1: Grabcad tremors

GRABCAD CHALLENGEVA Innovation Creation Series

Prosthetic & Assistive Technology Challenge

By: Reinaldo Wiener Rocca

Page 2: Grabcad tremors

PROBLEM STATEMENT

“Essential tremor is a common neurological disease and the most common movement disorder, which impacts hundreds of millions of people worldwide. Specifically, tremors are an "involuntary shaking movement that is repeated over and over” and when they occur in the hand, can be cumbersome and make daily tasks difficult, such as using a utensil, phone, or tool.” This is the description of the problem obtained from the Grabcad Challenge.

The objective is reduce the tremors effect to enable the patient to do day to day task without difficulty.

Page 3: Grabcad tremors

MODEL

Assumptions (risks): 1 dof model, hinge mechanism (wrist) The system is in resonance (ωnat = 7 Hz) Only steady state response considered Harmonic force and response

Technical requirements: Reduce a range of tremors (they range from a

frequency of 7 Hz with a low amplitude (0.1-0.015 cm) to a larger amplitude at 1 Hz) 

Fit under a shirt  Require no power source  Permit full thumb, finger, wrist and forearm

motion

Page 4: Grabcad tremors

MAIN IDEA OF THE SOLUTION

Reduce amplitudes of the tremors: To illustrate this part, harmonic response for a

forced-harmonic vibration is recalled:

In this sense, to reduce resonance effect the following can be done: Natural frequency of the system must be reduced (ωnat

<< ω), increasing the mass. But there is a limit since a lot of weight can be uncomfortable for the patient.

Increasing the stiffness will reduce the amplitude of the response since it is inversely proportional to it.

Overdamp the system: increasing the damping to a value above the critical. In this case even if natural frequency is equal to the frequency of the force (r=1), an amplification factor of 0.5 can be achieved.

Page 5: Grabcad tremors

PROPOSAL 1OVERDAMP THE SYSTEM

The overall concept is a glove that have fibers made of specific material and its good appealing.

Use of viscoelastic materials that give the desired behavior (like thermoplastic polyurethane). It is expected that if strain rate increases, the material absorb more vibration energy, same way as a damper do.

Use flexible dampers, that are within the glove (is hidden) and moves when wrist moves. It is important that the gloves wrap the stem of the damper to avoid buckling during compression.

Page 6: Grabcad tremors

PROPOSAL 1

This picture represent the core of the glove.

The red pieces are rigid and made of plastic, this helps the fixation of the dampers. It reduces relative movement between the hand and the dampers.

The elastic property of the glove fibers helps to attach the entire assembly to the hand.

The hand is able to move freely for smooth and slow movements.

Page 7: Grabcad tremors

PROPOSAL 1

The elastic fibers of the glove attach the stem to the hand and fingers, preventing the buckling during compression.

The green piston-cylinder set reduces wrist tremors and the blue ones reduce the finger tremors.

The idea is that this core design will be not visible for the patient since lies within the glove.

The distribution of the sets helps to reduce movements in a variety of directions.

Page 8: Grabcad tremors

PISTON-CYLINDER SET

Both stem and cylinder are made of flexible materials (polymers). In the case of the stem it should be able to handle compression and tension in an elastic range.

The energy of the movement of the wrist is dissipated through the viscous forces generated between the fluid within the cylinder and the piston (similarly as in a damper).

The geometry of the piston can be changed to obtain the desired damping and even a non-Newtonian fluid can be used to obtain a very high damping at regular load rates.

Page 9: Grabcad tremors

PROPOSAL 2STIFF THE SYSTEM

In this case it is desirable to stiff the hand when tremors occurs, but the hand is flexible when not.

It can be done using small tubes rounded by a wire and filled with magnetic fluid (represented in orange in the picture).

When a current flow pass through it, a magnetic flux is induced and the fluid “solidifies”.

The control system is able to decided what activates this function or even enable the user to decide.

Control

Page 10: Grabcad tremors

PROPOSAL 2STIFF THE SYSTEM

Despite for this idea a power source is needed I decided to share anyway. Solar cells, the cell phone energy or charger that take advantage of the hand movement, can be used to supply energy to this solution.

The main control can communicate with the smartphone through an app that tells how frequent the patient suffers tremors, check battery status and other useful functions.

Page 11: Grabcad tremors

CONCLUSIONS

Reducing the natural frequency to 1 Hz will imply a mass increase of 34kg, which is very high value.

After some research it was seen that what unable people to perform day-to-day activities (like writing) is actually the amplitude of the tremor, if this amplitude is reduced it will easier to handle daily tasks.

In general a solution that over damps the hand system will help to reduce tremors.

Another possibility is to create a device that suppress the electrical signals that induces the involuntary muscle movements. In this case the signal must be of high amplitude or a specific frequency to be easy to detect.