Getting started sound and vibration measurements
-
Upload
felix-ramirez-ramirez-moroyoqui -
Category
Engineering
-
view
109 -
download
4
Transcript of Getting started sound and vibration measurements
NI Sound and Vibration Assistant
Getting Started with the NI Sound and Vibration Assistant
Getting Started with the NI Sound and Vibration Assistant
May 2013371938D
Support
Worldwide Technical Support and Product Informationni.com
Worldwide OfficesVisit ni.com/niglobal to access the branch office Web sites, which provide up-to-date contact information, support phone numbers, email addresses, and current events.
National Instruments Corporate Headquarters11500 North Mopac Expressway Austin, Texas 78759-3504 USA Tel: 512 683 0100
For further support information, refer to the Technical Support and Professional Services appendix. To comment on National Instruments documentation, refer to the National Instruments Web site at ni.com/info and enter the Info Code feedback.
© 2007â2013 National Instruments. All rights reserved.
Important Information
WarrantyThe media on which you receive National Instruments software are warranted not to fail to execute programming instructions, due to defects in materials and workmanship, for a period of 90 days from date of shipment, as evidenced by receipts or other documentation. National Instruments will, at its option, repair or replace software media that do not execute programming instructions if National Instruments receives notice of such defects during the warranty period. National Instruments does not warrant that the operation of the software shall be uninterrupted or error free.
A Return Material Authorization (RMA) number must be obtained from the factory and clearly marked on the outside of the package before any equipment will be accepted for warranty work. National Instruments will pay the shipping costs of returning to the owner parts which are covered by warranty.
National Instruments believes that the information in this document is accurate. The document has been carefully reviewed for technical accuracy. In the event that technical or typographical errors exist, National Instruments reserves the right to make changes to subsequent editions of this document without prior notice to holders of this edition. The reader should consult National Instruments if errors are suspected. In no event shall National Instruments be liable for any damages arising out of or related to this document or the information contained in it.
EXCEPT AS SPECIFIED HEREIN, NATIONAL INSTRUMENTS MAKES NO WARRANTIES, EXPRESS OR IMPLIED, AND SPECIFICALLY DISCLAIMS ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. CUSTOMERâS RIGHT TO RECOVER DAMAGES CAUSED BY FAULT OR NEGLIGENCE ON THE PART OF NATIONAL INSTRUMENTS SHALL BE LIMITED TO THE AMOUNT THERETOFORE PAID BY THE CUSTOMER. NATIONAL INSTRUMENTS WILL NOT BE LIABLE FOR DAMAGES RESULTING FROM LOSS OF DATA, PROFITS, USE OF PRODUCTS, OR INCIDENTAL OR CONSEQUENTIAL DAMAGES, EVEN IF ADVISED OF THE POSSIBILITY THEREOF. This limitation of the liability of National Instruments will apply regardless of the form of action, whether in contract or tort, including negligence. Any action against National Instruments must be brought within one year after the cause of action accrues. National Instruments shall not be liable for any delay in performance due to causes beyond its reasonable control. The warranty provided herein does not cover damages, defects, malfunctions, or service failures caused by ownerâs failure to follow the National Instruments installation, operation, or maintenance instructions; ownerâs modification of the product; ownerâs abuse, misuse, or negligent acts; and power failure or surges, fire, flood, accident, actions of third parties, or other events outside reasonable control.
CopyrightUnder the copyright laws, this publication may not be reproduced or transmitted in any form, electronic or mechanical, including photocopying, recording, storing in an information retrieval system, or translating, in whole or in part, without the prior written consent of National Instruments Corporation.
National Instruments respects the intellectual property of others, and we ask our users to do the same. NI software is protected by copyright and other intellectual property laws. Where NI software may be used to reproduce software or other materials belonging to others, you may use NI software only to reproduce materials that you may reproduce in accordance with the terms of any applicable license or other legal restriction.
End-User License Agreements and Third-Party Legal NoticesYou can find end-user license agreements (EULAs) and third-party legal notices in the following locations:
⢠Notices are located in the <National Instruments>\_Legal Information and <National Instruments> directories.
⢠EULAs are located in the <National Instruments>\Shared\MDF\Legal\license directory.
⢠Review <National Instruments>\_Legal Information.txt for more information on including legal information in installers built with NI products.
TrademarksRefer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for more information on National Instruments trademarks.
ARM, Keil, and µVision are trademarks or registered of ARM Ltd or its subsidiaries.
LEGO, the LEGO logo, WEDO, and MINDSTORMS are trademarks of the LEGO Group. ©2013 The LEGO Group.
TETRIX by Pitsco is a trademark of Pitsco, Inc.©2013
FIELDBUS FOUNDATION⢠and FOUNDATION⢠are trademarks of the Fieldbus Foundation.
EtherCAT® is a registered trademark of and licensed by Beckhoff Automation GmbH.
CANopen® is a registered Community Trademark of CAN in Automation e.V.
DeviceNet⢠and EtherNet/IP⢠are trademarks of ODVA.
Go!, SensorDAQ, and Vernier are registered trademarks of Vernier Software & Technology. Vernier Software & Technology and vernier.com are trademarks or trade dress.
Xilinx is the registered trademark of Xilinx, Inc.
Taptite and Trilobular are registered trademarks of Research Engineering & Manufacturing Inc.
FireWire® is the registered trademark of Apple Inc.
Linux® is the registered trademark of Linus Torvalds in the U.S. and other countries.
Handle Graphics®, MATLAB®, Real-Time Workshop®, Simulink®, Stateflow®, and xPC TargetBox® are registered trademarks, and TargetBox⢠and Target Language Compiler⢠are trademarks of The MathWorks, Inc.
Tektronix®, Tek, and Tektronix, Enabling Technology are registered trademarks of Tektronix, Inc.
The Bluetooth® word mark is a registered trademark owned by the Bluetooth SIG, Inc.
The ExpressCard⢠word mark and logos are owned by PCMCIA and any use of such marks by National Instruments is under license.
The mark LabWindows is used under a license from Microsoft Corporation. Windows is a registered trademark of Microsoft Corporation in the United States and other countries.
Other product and company names mentioned herein are trademarks or trade names of their respective companies.
PatentsFor patents covering National Instruments products/technology, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your media, or the National Instruments Patent Notice at ni.com/patents.
Export Compliance InformationRefer to the Export Compliance Information at ni.com/legal/export-compliance for the National Instruments global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and other import/export data.
WARNING REGARDING USE OF NATIONAL INSTRUMENTS PRODUCTS(1) NATIONAL INSTRUMENTS PRODUCTS ARE NOT DESIGNED WITH COMPONENTS AND TESTING FOR A LEVEL OF RELIABILITY SUITABLE FOR USE IN OR IN CONNECTION WITH SURGICAL IMPLANTS OR AS CRITICAL COMPONENTS IN ANY LIFE SUPPORT SYSTEMS WHOSE FAILURE TO PERFORM CAN REASONABLY BE EXPECTED TO CAUSE SIGNIFICANT INJURY TO A HUMAN.
(2) IN ANY APPLICATION, INCLUDING THE ABOVE, RELIABILITY OF OPERATION OF THE SOFTWARE PRODUCTS CAN BE IMPAIRED BY ADVERSE FACTORS, INCLUDING BUT NOT LIMITED TO FLUCTUATIONS IN ELECTRICAL POWER SUPPLY, COMPUTER HARDWARE MALFUNCTIONS, COMPUTER OPERATING SYSTEM SOFTWARE FITNESS, FITNESS OF COMPILERS AND DEVELOPMENT SOFTWARE USED TO DEVELOP AN APPLICATION, INSTALLATION ERRORS, SOFTWARE AND HARDWARE COMPATIBILITY PROBLEMS, MALFUNCTIONS OR FAILURES OF ELECTRONIC MONITORING OR CONTROL DEVICES, TRANSIENT FAILURES OF ELECTRONIC SYSTEMS (HARDWARE AND/OR SOFTWARE), UNANTICIPATED USES OR MISUSES, OR ERRORS ON THE PART OF THE USER OR APPLICATIONS DESIGNER (ADVERSE FACTORS SUCH AS THESE ARE HEREAFTER COLLECTIVELY TERMED âSYSTEM FAILURESâ). ANY APPLICATION WHERE A SYSTEM FAILURE WOULD CREATE A RISK OF HARM TO PROPERTY OR PERSONS (INCLUDING THE RISK OF BODILY INJURY AND DEATH) SHOULD NOT BE RELIANT SOLELY UPON ONE FORM OF ELECTRONIC SYSTEM DUE TO THE RISK OF SYSTEM FAILURE. TO AVOID DAMAGE, INJURY, OR DEATH, THE USER OR APPLICATION DESIGNER MUST TAKE REASONABLY PRUDENT STEPS TO PROTECT AGAINST SYSTEM FAILURES, INCLUDING BUT NOT LIMITED TO BACK-UP OR SHUT DOWN MECHANISMS. BECAUSE EACH END-USER SYSTEM IS CUSTOMIZED AND DIFFERS FROM NATIONAL INSTRUMENTS' TESTING PLATFORMS AND BECAUSE A USER OR APPLICATION DESIGNER MAY USE NATIONAL INSTRUMENTS PRODUCTS IN COMBINATION WITH OTHER PRODUCTS IN A MANNER NOT EVALUATED OR CONTEMPLATED BY NATIONAL INSTRUMENTS, THE USER OR APPLICATION DESIGNER IS ULTIMATELY RESPONSIBLE FOR VERIFYING AND VALIDATING THE SUITABILITY OF NATIONAL INSTRUMENTS PRODUCTS WHENEVER NATIONAL INSTRUMENTS PRODUCTS ARE INCORPORATED IN A SYSTEM OR APPLICATION, INCLUDING, WITHOUT LIMITATION, THE APPROPRIATE DESIGN, PROCESS AND SAFETY LEVEL OF SUCH SYSTEM OR APPLICATION.
© National Instruments v Getting Started with the NI Sound and Vibration Assistant
Contents
About This ManualRequired Software .........................................................................................................viiRelated Documentation..................................................................................................vii
Chapter 1Introduction to the NI Sound and Vibration Assistant
Chapter 2Working with Projects
Opening a Project...........................................................................................................2-1Running a Project and Displaying Signals ....................................................................2-3Configuring a Step .........................................................................................................2-5Moving Steps and Handling Errors ...............................................................................2-6
Chapter 3Logging Data
Recording a Signal.........................................................................................................3-1Viewing a Logged Signal ..............................................................................................3-3Logging Signals with Predefined Start and Stop Conditions ........................................3-4
Chapter 4Performing Sound and Vibration Measurements
Opening Previously Recorded Data...............................................................................4-1Playing Back Data .........................................................................................................4-2Analyzing Data ..............................................................................................................4-5Advanced Playback........................................................................................................4-7Signal Types in the Sound and Vibration Assistant ......................................................4-8Saving, Exporting, and Printing Signals ........................................................................4-8
Saving Signals to File......................................................................................4-8Exporting Signals to Microsoft Excel .............................................................4-9Printing Signals ...............................................................................................4-9Creating Reports in the Sound and Vibration Assistant..................................4-9Exporting Documentation to HTML...............................................................4-9
Contents
Getting Started with the NI Sound and Vibration Assistant vi ni.com
Chapter 5Performing Sweep Measurements
Defining Sweep Ranges and Outputs ............................................................................ 5-1Creating a Project to Sweep ............................................................................ 5-1Defining the Sweep......................................................................................... 5-2
Plotting Sweep Results .................................................................................................. 5-5
Chapter 6Performing Order Analysis
Performing Order Analysis on Logged Data................................................................. 6-1Loading the UFF58b Data File ....................................................................... 6-2Measuring the Rotational Speed of the Engine............................................... 6-2Identifying Dominant Orders .......................................................................... 6-3Tracking the Magnitude of Specific Orders.................................................... 6-6
Performing Order Analysis During Data Acquisition ................................................... 6-7Importing the Logged Data File...................................................................... 6-7Measuring the Rotational Speed of the PC Fan .............................................. 6-8Identifying Dominant Orders .......................................................................... 6-8Tracking the Magnitude of Specific Orders.................................................... 6-10
Chapter 7Extending Sound and Vibration Assistant Projects with LabVIEW
Importing LabVIEW VIs into the Sound and Vibration Assistant as Steps.................. 7-1Converting Sound and Vibration Assistant Projects to LabVIEW Block Diagrams .... 7-5
Chapter 8Where to Go from Here
Sound and Vibration Assistant Example Projects ......................................................... 8-1Opening a Sound and Vibration Example Solution ...................................................... 8-1Using Hardware with the Sound and Vibration Assistant............................................. 8-1Web Resources .............................................................................................................. 8-2
Appendix ATechnical Support and Professional Services
© National Instruments vii Getting Started with the NI Sound and Vibration Assistant
About This Manual
Use this manual to familiarize yourself with the National Instruments Sound and Vibration Assistant. You can perform interactive measurements and use basic Sound and Vibration Assistant features to acquire and analyze signals.
This manual contains exercises that help you begin working with the Sound and Vibration Assistant. These exercises teach you how to run projects, configure steps, work with signals, perform sweep measurements, log data, and extend the Sound and Vibration Assistant with NI LabVIEW Graphical Programming.
The Sound and Vibration Assistant extends existing NI SignalExpress technology. Throughout this manual, you will notice references or examples in SignalExpress. These references and examples apply to both the Sound and Vibration Assistant and SignalExpress.
Required SoftwareThe Sound and Vibration Assistant requires you to install the following software:
⢠If you use the Sound and Vibration Assistant with National Instruments data acquisition (DAQ) hardware, you must install NI-DAQmx 8.9.5 or later for NI-DAQmx support.
⢠If you want to convert Sound and Vibration Assistant projects to LabVIEW block diagrams, you must install LabVIEW 8.2 or later before you install the Sound and Vibration Assistant.
Related DocumentationIn addition to the exercises in this manual, you can learn about the Sound and Vibration Assistant with the animated Sound and Vibration Assistant Tutorial. Complete the following steps to access this tutorial:
1. Select Start»All Programs»National Instruments»Sound and Vibration»NI Sound and Vibration Assistant to launch the Soundand Vibration Assistant.
2. Select Help»Show Welcome Dialog.
3. Scroll down to the Sound and Vibration Assistant heading.
About This Manual
Getting Started with the NI Sound and Vibration Assistant viii ni.com
4. Expand this heading.
5. Click the View a five minute interactive tutorial link.
You also can refer to the Sound and Vibration Assistant Help, available by selecting Help»Sound and Vibration Assistant Help and navigating on the Contents tab to Step Reference»Sound and Vibration Assistant Steps, for more information as you read this manual.
© National Instruments 1-1 Getting Started with the NI Sound and Vibration Assistant
1Introduction to the NI Sound and Vibration Assistant
The NI Sound and Vibration Assistant optimizes virtual instrumentation for sound and vibration engineers by offering instant interactive measurements that require no programming. You can use the Sound and Vibration Assistant interactively to acquire, generate, analyze, compare, import, log, and play back sound and vibration signals.
The Sound and Vibration Assistant extends NI SignalExpress technology to provide specific sound and vibration measurements and analysis. You can use the Sound and Vibration Assistant for audio testing, acoustic measurements, environmental noise testing, machine condition monitoring, vibration analysis, noise and vibration control, and noise, vibration, and harshness (NVH) measurements.
You also can extend the functionality of the Sound and Vibration Assistant by importing a custom virtual instrument (VI) created in NI LabVIEW or by converting a Sound and Vibration Assistant project to a LabVIEW block diagram so that you can continue development in LabVIEW.
© National Instruments 2-1 Getting Started with the NI Sound and Vibration Assistant
2Working with Projects
You can use the NI Sound and Vibration Assistant to perform measurements by adding and configuring steps in an interactive measurement environment. A step is a configurable function that acquires, generates, analyzes, loads, or saves signals. Most steps process input signals and produce output signals. A saved sequence of configured steps is a Sound and Vibration Assistant project.
This chapter teaches you how to load and run existing projects and how to configure steps in these projects.
Opening a ProjectComplete the following steps to load an example project in the Sound and Vibration Assistant:
1. Select Start»All Programs»National Instruments»Sound and Vibration»NI Sound and Vibration Assistant to launch the Soundand Vibration Assistant.
Notice that the Sound and Vibration Assistant is split into views that display various types of information. The primary view appears in the center of the application window and contains tabs. If the Sound and Vibration Assistant opens in the default configuration, the Data View tab, the Recording Options tab, and the Project Documentation tab appear in the primary view.
The primary view is surrounded by supplementary views. In the default configuration, the Project View appears on the left, and the context help appears on the right.
2. If the Sound and Vibration Assistant does not open in the default configuration, select View»Reset Layout to reset the application to the default configuration. You can use the View menu to display tabs and views or reset the layout at any time.
3. Select Help»Open Example»Sound and Vibration Assistant, double-click the Tutorial folder, and double-click the SVA_Analyze Signal From UFF File (Tutorial).seproj Sound and Vibration Assistant project.
Chapter 2 Working with Projects
Getting Started with the NI Sound and Vibration Assistant 2-2 ni.com
4. Examine the window that appears, as shown in Figure 2-1, to learn about different components of the Sound and Vibration Assistant.
Figure 2-1. SVA_Analyze Signal from UFF File (Tutorial).seproj
The left pane displays the Project View, which presents the order of operations, or steps, for the project.
The center pane displays the Data View tab, which displays the signals that the project generates and analyzes.
The right pane displays the context help and updates when you hover the cursor over a new step, tab, parameter, or display. The upper half of the context help displays information about the steps and tabs, and the lower half of the context help displays information about the parameters.
1 Run button2 Work area pull-down menu3 Project status4 Step
5 Input6 Outputs7 Project View8 Logged data window
9 Primary view with Data View tab visible10 Record button11 View tabs12 Context help
9
7
8
6 12
111 10
4
5
2
3
Chapter 2 Working with Projects
© National Instruments 2-3 Getting Started with the NI Sound and Vibration Assistant
Running a Project and Displaying SignalsThe Sound and Vibration Assistant has two execution modesâRun Once and Run Continuously. When you click the down arrow on the Run button, shown at left, and select the Run Once option, the Sound and Vibration Assistant executes all steps in the project once. When you click the down arrow on the Run button and select the Run Continuously option, the Sound and Vibration Assistant executes all steps in the project continuously until you click the Stop button, shown at left.
The Stop button appears in place of the Run button as the project runs. While the steps in the project execute, the Data View tab updates continuously. You can change the measurement configurations while the project runs and view the results immediately.
You also can configure the run mode of a project. When you click the down arrow on the Run button and select Configure Run, you can set how long to run the project by setting the iterations, seconds, or setting the project to run continuously. If you click the Run button without selecting options from the pull-down menu, the project runs continuously until you click the Stop button.
Complete the following steps to run the example project and display signals:
1. Click the Run button to execute all steps in the project continously.
When you run a project, steps analyze input signals and generate new output signals as a result of the analysis. In this project, the Load from UFF58 step loads a triangle wave from a UFF58 file, and the Power Spectrum step and the Peak Search step analyze the triangle wave and return new outputs. In the Project View, the Sound and Vibration Assistant designates inputs with red arrows and outputs with blue arrows.
2. Drag the number of peaks output signal of the Peak Search step from the Project View to the Data View tab to display the signal.
Chapter 2 Working with Projects
Getting Started with the NI Sound and Vibration Assistant 2-4 ni.com
The Sound and Vibration Assistant creates a new graph on the Data View tab. The Sound and Vibration Assistant does not display the number of peaks signal in an existing table or graph because the number of peaks signal is a different type of data than the data in the existing graphs. The Sound and Vibration Assistant automatically recognizes different types of signals and renders them in the appropriate displays. The project appears as shown in Figure 2-2.
Figure 2-2. Outputs of SVA_Analyze Signal From UFF File (Tutorial).seproj
3. Select Help»Sound and Vibration Assistant Help and click the Search tab. Enter âsignal typesâ and double-click on Signal Types in SignalExpress to find more information about signal types.
The help provides information about using the Sound and Vibration Assistant and SignalExpress functionality such as projects, steps, and signals.
Chapter 2 Working with Projects
© National Instruments 2-5 Getting Started with the NI Sound and Vibration Assistant
Configuring a StepA step is a configurable function that acquires, generates, analyzes, loads, or saves signals. Most steps process input signals and produce output signals. You can configure the operation of a step by specifying values in the Step Setup tab. While a project runs, you can modify the configuration of steps and view immediate changes on the Data View tab, and you can adjust the measurements until you achieve the results you need.
Complete the following steps to configure the Power Spectrum step and the Peak Search step:
1. Select File»Save Project As and save the project as My First Project.seproj in a convenient location.
2. Double-click the Power Spectrum step in the Project View. The Step Setup tab for the Power Spectrum step appears as shown in Figure 2-3. The graph on the Step Setup tab displays a preview of the input signal.
Figure 2-3. Power Spectrum Step Setup Tab
Chapter 2 Working with Projects
Getting Started with the NI Sound and Vibration Assistant 2-6 ni.com
3. Click the Input tab in the Step Setup tab. Notice that the Input signal pull-down menu specifies a time waveform. The Load from UFF58 step produces the time waveform signal for the Power Spectrum step.
4. Click the Peak Search step in the Project View. The Step Setup tab changes from displaying the configuration of the Power Spectrum step to displaying the configuration of the Peak Search step.
5. Click the Input tab in the Step Setup tab. Notice that the Input Signal pull-down menu specifies a spectrum. The Power Spectrum step produces the spectrum signal for the Peak Search step.
6. Click the Configuration tab in the Step Setup tab. The settings indicate that the Peak Search step performs a peak search on the spectrum to locate a single maximum peak above a threshold of -40. The step generates two measurements as outputsâthe frequency and amplitude of the peak and the number of peaks.
7. On the Configuration tab in the Step Setup tab, select Multiple Peaks from the Search type pull-down menu to measure every peak above the threshold of the input spectrum. The signals in the Peaks table on the Step Setup tab update to reflect the change you made.
8. Click the Stop button to stop the project.
When you click the Stop button, the project stops running after completing the current iteration. You can click the down arrow on the Stop button and select the Abort option to completely stop the project without finishing the current iteration.
9. Select File»Save Project to save the project.
10. Select File»Close Project to close the project.
Moving Steps and Handling ErrorsMost steps in Sound and Vibration Assistant projects depend on input data, which means steps can operate only on signals exported from previous steps in the Project View. The Input signal pull-down menu on the Input tab in the Step Setup tab of a step displays only compatible signals exported from previous steps. When the output of a step becomes the input of another step, the steps become dependent on each other, and the two steps execute sequentially at the same rate. The first step generates an output signal that the second step must receive as an input before executing.
You can move a step within a project by dragging it up or down in the Project View. You can delete a step by right-clicking the step name in the Project View and selecting Delete from the shortcut menu. However, when you move or delete a step, the status of signals in the project changes.
Chapter 2 Working with Projects
© National Instruments 2-7 Getting Started with the NI Sound and Vibration Assistant
For example, if you delete a step that generates output signals, the operation of the project breaks if any of the deleted output signals are input signals for other steps. This causes an error indicator to appear.
You also can cut, copy, and paste steps within a project by pressing the <Ctrl-X>, <Ctrl-C>, and <Ctrl-V> keys, respectively, or by right-clicking a step in the Project View and selecting Cut, Copy, Paste Before Selected Step, or Paste After Selected Step from the shortcut menu.
Complete the following steps to move steps and handle an error:
1. Select File»New Project to open a new project.
2. Save the project as My Error.seproj in a convenient location.
3. Click the Add Step button, shown at left, and select Processing»Standardized Filter to add the Standardized Filter step.
Tip You can add steps to a project from the Add Step menu, the Add Step palette, or the shortcut menu that appears when you right-click in the Project View.
4. Click the Cancel button when the Missing Inputs dialog box appears, as shown in Figure 2-4.
Figure 2-4. Missing Inputs Dialog Box
The Missing Inputs dialog box prompts you to select a step to produce the necessary input signal if an input signal does not exist.
Chapter 2 Working with Projects
Getting Started with the NI Sound and Vibration Assistant 2-8 ni.com
5. Notice an error indicator, shown at left, appears at the bottom of the Step Setup tab and on the step in the Project View.
Note The Sound and Vibration Assistant logs all errors and warnings on the Event Log tab while a project runs. To display the Event Log, select View»Event Log. Refer to the Sound and Vibration Assistant Help, available by selecting Help»Sound and Vibration Assistant Help, for more information about errors and warnings. Click the Search tab and enter errors.
6. Click the Details button on the Step Setup tab to see the full description of the error in the Details dialog box. You also can click the error indicator in the Project View to display the full description. To handle this error, you can add a step to provide the missing input.
7. Close the Details dialog box.
8. Click the Add Step button and select Create Signals»Create Analog Signal. The Missing Input error still exists because the Create Analog Signal step does not provide the direct input signal to the Standardized Filter step.
9. Drag the Create Analog Signal step above the Standardized Filter step in the Project View. The Standardized Filter step uses the sine wave output of the Create Analog Signal step as an input signal, removing the error indicator.
10. Save and close the project.
© National Instruments 3-1 Getting Started with the NI Sound and Vibration Assistant
3Logging Data
You can use the Sound and Vibration Assistant to record measurements, including any time-domain, numeric, or Boolean step output.
This chapter teaches you how to record data using the integrated data logging features in the Sound and Vibration Assistant. You learn how to record a specified signal, play back that signal, and analyze that signal using analysis steps. You also learn how to use the Recording Options tab to log signals based on specified start or stop conditions.
Note References or examples of SignalExpress also apply to the Sound and Vibration Assistant.
Recording a SignalComplete the following steps to record a signal:
1. Select Help»Open Example»SignalExpress, double-click the Tutorial folder, and double-click Logging.seproj.
This project uses the Create Analog Signal step to generate a signal based on a formula.
2. Save the project as My Logged Signal.seproj in a convenient location.
3. Click the Record button, shown at left, to open the Logging Signals Selection dialog box.
The Logging Signals Selection dialog box displays the signals in the project available for recording. You can select one signal or multiple signals to record. You also can specify a name and description forthe log.
4. Place a checkmark in the signal checkbox to record the signal generated in the Create Analog Signal step.
5. Click the OK button to close the Logging Signals Selection dialog box and begin recording the signal. The logging operation continues until you click the Stop button.
Chapter 3 Logging Data
Getting Started with the NI Sound and Vibration Assistant 3-2 ni.com
6. Click the Stop button to stop logging the signal. If you have not logged a signal before, the First Log Complete dialog box appears. Click the OK button to close the dialog box.
The Logged Data window, below the Project View, displays a list of all logged data in the current project, as shown in Figure 3-1.
Figure 3-1. Logged Data Window
By default, the Sound and Vibration Assistant names the logged data according to the date and time you recorded the data. The Sound and Vibration Assistant saves logged data in the .tdms file format.
Note Visit ni.com/info and enter tdms for information about the .tdms file format.
7. Select Tools»Options and select the Logging option to specify the directory for the Sound and Vibration Assistant to save the logged data and to customize various preferences for the logged data.
8. Click the OK button to close the Options dialog box.
9. Save the project.
Chapter 3 Logging Data
© National Instruments 3-3 Getting Started with the NI Sound and Vibration Assistant
Viewing a Logged SignalComplete the following steps to view the logged data:
1. Drag the data log from the Logged Data window to the Data View tab. The Data View tab displays the logged data in a graph and a preview graph beneath, as shown in Figure 3-2.
Figure 3-2. Logging.seproj
The preview graph provides a method for zooming in, zooming out, and panning through data in the Data View tab. The preview graph appears by default when you view logged data. When viewing live or non-logged data, you can right-click the graph on the Data View tab and select Visible Items»Preview to display the preview graph.
Chapter 3 Logging Data
Getting Started with the NI Sound and Vibration Assistant 3-4 ni.com
2. Click the Zoom In button, shown at left, next to the preview graph to zoom in on the logged signal. The cursors on the preview graph show the subset of the data currently displayed on the Data View tab. Click and drag the cursors to increase or decrease the subset of data you are viewing.
3. Use the scroll bar beneath the preview graph to scroll through the data.
4. Right-click the preview graph and select Show All from the shortcut menu to change the graph to display the entire signal.
Logging Signals with Predefined Start and Stop Conditions
You can configure start and stop conditions that signals must meet before the Sound and Vibration Assistant starts recording or stops recording the signals. Complete the following steps to log data based on start and stop conditions:
1. Click the Recording Options tab or select View»Recording Options.
2. Select Signal Selection from the Category list on the Recording Options tab.
Chapter 3 Logging Data
© National Instruments 3-5 Getting Started with the NI Sound and Vibration Assistant
3. Place a checkmark next to the signal in the Record column, as shown in Figure 3-3.
Figure 3-3. Signal Selection
The Record button changes to the Record While Running button, shown at left, in the enabled position with the green checkmark. When the Record While Running button is enabled, the Sound and Vibration Assistant records the selected signal when you click the Run button.
4. Select Start Conditions from the Category list on the Recording Options tab.
5. Click the Add button under the Start condition list to customize a start condition for the logging task.
a. In the Condition type pull-down menu, verify that Signal trigger is selected to specify that the Sound and Vibration Assistant begins recording when the input signal meets the specified condition.
b. In the Signal pull-down menu, verify that signal is selected.
Chapter 3 Logging Data
Getting Started with the NI Sound and Vibration Assistant 3-6 ni.com
c. In the Trigger type pull-down menu, verify that Rising slope is selected to begin recording the signal based on the value of the edge of the signal on a positive slope.
d. Enter 1 in the Trigger value field to begin recording when the signal crosses 1 on a rising slope.
The Recording Options tab appears as shown in Figure 3-4.
Figure 3-4. Recording Options Start Conditions
6. Select Stop Conditions from the Category list on the Recording Options tab.
Chapter 3 Logging Data
© National Instruments 3-7 Getting Started with the NI Sound and Vibration Assistant
7. Click the Add button under the Stop condition list to customize a stop condition for the logging task.
a. In the Condition type pull-down menu, verify that Duration is selected to specify that the Sound and Vibration Assistant stops recording after a specified amount of time passes.
b. In the Duration control, verify that 5 appears to specify to record the signal for 5 seconds after the signal meets the start condition.
8. Click the Run button. The Sound and Vibration Assistant begins recording the signal when the signal crosses level 1 on a rising slope and continues recording the signal for 5 seconds.
The following indicators in the Recording status section of the Recording Options tab update while the project runs:
⢠Recording displays on when the signal meets the start condition and logging is in progress.
⢠Disk information displays the available hard disk space on the computer for the log.
⢠Current estimated log size displays the size of the log file on disk.
⢠Recording time available displays the amount of time you can continue recording the log before running out of disk space.
⢠Current log started on displays the start time of the current log.
The Start Conditions, Stop Conditions, Alarms, and Events pages of the Recording Options tab also include indicators that display the status of start and stop conditions, alarms, and events that you can configure.
9. Save and close the project.
© National Instruments 4-1 Getting Started with the NI Sound and Vibration Assistant
4Performing Sound and Vibration Measurements
You can use the Sound and Vibration Assistant to perform sound and vibration measurements interactively. You can analyze and process logged data by playing the logged data back through analysis steps. You also can analyze and process live data by using National Instruments data acquisition (DAQ) hardware, such as the NI dynamic signal acquisition (DSA) devices.
This chapter teaches you how to work with sound and vibration data in the Sound and Vibration Assistant, including how to analyze previously recorded data, plot measurements on graphs, and save signals to a file.
Opening Previously Recorded DataYou can analyze previously recorded data with the Sound and Vibration Assistant. You also can analyze data acquired using DAQ hardware.
Complete the following steps to open previously recorded soundpressure data:
1. Open a new project.
2. Select File»Import»Logged Signals from SignalExpress TDMS Files, navigate to the National Instruments\Sound and
Vibration Assistant\Examples\Log File - Sound
Pressure\ directory, and click the Current Folder button.
(Windows XP/2000) The National Instruments folder is located in the C:\Documents and Settings\All Users\
Shared Documents\ directory.
(Windows Vista) The National Instruments folder is located in the C:\Users\Public\Public Documents\ directory.
3. Notice that the logged data appears under Logs in the Logged Data window.
Chapter 4 Performing Sound and Vibration Measurements
Getting Started with the NI Sound and Vibration Assistant 4-2 ni.com
4. Select Playback»Sound Pressure from the work area pull-down menu at the top of the Project View, as shown in Figure 4-1.
Figure 4-1. Work Area Pull-down Menu
Use work areas to perform multiple Sound and Vibration Assistant operations from within the same project. You can acquire data, process signals, record data, and perform measurements on logged data without opening a new project. When you save a project, the Sound and Vibration Assistant saves every work area within the project in the same project file.
The default work area, Monitor/Record, allows you to take measurements, analyze live data, and log data. The Playback work area allows you to use logged data that you record in the Monitor/Record work area as an input for an analysis step.
You now can analyze the sound pressure data in the Sound Pressure.tdms file.
Note You must have DAQ hardware and NI-DAQmx 8.9.5 or later installed to run projects with live sound pressure data. If you want to use live sound pressure data in the following exercises, create a new project, click the Add Step button, and select Acquire Signals»DAQmx Acquire»Analog Input»Sound Pressure. In the Add Channels To Task dialog box that appears, you can select the analog input channel(s) you want to analyze.
Playing Back DataComplete the following steps to analyze the data in the Sound Pressure.tdms file:
1. Save the project as My Analysis.seproj in a convenient location.
2. Click the Add Step button and select Analysis»Time-Domain Measurements»Sound Level.
3. Click the Input tab in the Step Setup tab. Notice that the Input signal pull-down menu already contains the sound pressure signal from the .tdms file. The Sound and Vibration Assistant automatically selects the logged data as the input to the Sound Level step.
Chapter 4 Performing Sound and Vibration Measurements
© National Instruments 4-3 Getting Started with the NI Sound and Vibration Assistant
4. Click the Weighting tab in the Step Setup tab and select A Weighting from the Weighting pull-down menu to simulate the loudness of the low-level tones.
5. Click the Averaging tab in the Step Setup tab and remove the checkmarks from the Leq and Peak checkboxes to compute only the running equivalent continuous sound level and the exponential averaged sound level of the signal.
6. Drag the Sound Pressure signal from the Logged Data window to the Data View tab.
Note Viewing logged signals in Playback mode displays only one iteration of the data at a time. To view the complete logged data, select Monitor/Record from the work area pull-down menu to switch to the Monitor/Record work area. You also can view the complete logged data on the Playback Options tab.
7. Run the project.
Notice that the recorded sound pressure data appears in the Data View tab with a timeline slider at the top of the view. The project runs until it reaches the end of the logged data file.
Tip When the project is not running, you can click and drag the slider in the timeline to view different portions of the previously recorded data.
Chapter 4 Performing Sound and Vibration Measurements
Getting Started with the NI Sound and Vibration Assistant 4-4 ni.com
8. Right-click the graph and select Visible Items»Cursors fromthe shortcut menu to display two interactive cursors, as shown in Figure 4-2.
Figure 4-2. Displaying Cursors on the Graph
9. Click and drag the cursors to increase or decrease the subset of data you are viewing. As you drag the cursors, the Sound and Vibration Assistant displays the x and y values of the cursors in the cursor table below the graph in the Data View tab.
10. Save the project.
Chapter 4 Performing Sound and Vibration Measurements
© National Instruments 4-5 Getting Started with the NI Sound and Vibration Assistant
Analyzing DataComplete the following steps to perform more measurements on the sound pressure data:
1. Drag the running Leq output of the Sound Level step to the Data View tab. The Sound and Vibration Assistant creates a new chart in the Data View tab.
2. Drag the exponential output of the Sound Level step to the existing chart on the Data View tab.
3. Run the project.
Notice on the Data View tab that running Leq and exponential update in the chart as the playback data progresses.
4. Click the Add Step button and select Analysis»Frequency-Domain Measurements»Octave Analysis.
5. Notice that the warning Frequency range coerced appears at the bottom of the Step Setup tab. The Octave Analysis step automatically coerces the frequency range based on the sample rate of the input signal. Select 1k from the High frequency (Hz) pull-down menu to change the frequency range and eliminate the warning.
6. Click the Input tab in the Step Setup tab and notice that the Input Signal is the same sound pressure signal you used with the Sound Level step.
7. Click the Configuration tab in the Step Setup tab and select A Weighting from the Weighting pull-down menu.
Chapter 4 Performing Sound and Vibration Measurements
Getting Started with the NI Sound and Vibration Assistant 4-6 ni.com
8. Drag the octave output of the Octave Analysis step to the Data View tab.
The Sound and Vibration Assistant displays the sound pressure, running Leq, exponential, and octave charts, as shown in Figure 4-3.
Figure 4-3. My Analysis.seproj with Octave Analysis
9. Drag the total band power output of the Octave Analysis step to the chart with running Leq and exponential on the Data View tab.
10. Right-click the chart and select View As»Table Display from the shortcut menu to view the numerical results as a table instead of a graph.
11. Run the project.
Notice that the octave graph updates as the playback progresses. You have now performed sound level and octave analysis on the sound pressure data.
12. Save and close the project.
Chapter 4 Performing Sound and Vibration Measurements
© National Instruments 4-7 Getting Started with the NI Sound and Vibration Assistant
Advanced PlaybackYou can configure advanced data playback options by using the Playback Options tab. The Playback Options tab displays a preview of the logged data and allows you to select a subset of that data to play back or run through analysis steps.
Complete the following steps to configure advanced playback options:
1. Select File»Open Project, navigate to My Logged Signal.seproj, and click the Open button.
2. Select Playback from the work area pull-down menu.
3. Click the Playback Options tab or select View»Playback Options. The Playback Options tab appears as shown in Figure 4-4.
Figure 4-4. Playback Options Tab
4. Select signal from the second log in the Logged signal listbox.
5. Enter 1 in the Start time (s) field and 4 in the Stop time (s) field to play back or analyze a subset of the logged signal beginning one second and ending four seconds after the start of the log.
6. Click the Add Step button and select Processing»Filter.
7. Drag the filtered data output of the Filter step to the Data View tab.
Chapter 4 Performing Sound and Vibration Measurements
Getting Started with the NI Sound and Vibration Assistant 4-8 ni.com
8. Click the Run button. The Sound and Vibration Assistant filters the subset of the signal you specified on the Playback Options tab and displays the resulting filtered signal on the Data View tab. Notice that the signal starts at 1 second and stops at 4 seconds.
9. Save the project.
Refer to the Sound and Vibration Assistant Help, available by selecting Help»Sound and Vibration Assistant Help, for more information about logging data, such as specifying alarm conditions and advanced playback options.
Signal Types in the Sound and Vibration AssistantSome steps, such as the Arithmetic step, can operate on multiple signal types. For example, you can use the Arithmetic step to operate on time-domain or frequency-domain signals. The Arithmetic step changes behavior based on the type of input signals you select for the step. For example, if you add two time-domain signals, the Sound and Vibration Assistant adds only their amplitudes. However, if you add two frequency-domain phase signals, the Sound and Vibration Assistant adds the appropriate phase shift.
Saving, Exporting, and Printing SignalsYou can use the Sound and Vibration Assistant to document signals or continue analysis in another software application. This section teaches you how to save signals to an ASCII file, export data to Microsoft Excel, print signals, and use the built-in project documentation feature to document the Sound and Vibration Assistant project.
Saving Signals to FileComplete the following steps to save a signal from My Logged Signal.seproj to a file:
1. Click the Add Step button and select Load/Save Signals»Save to ASCII/LVM. You can use a Load/Save Signals step to save data to a file every time the project runs.
2. Click the File Settings tab in the Step Setup tab and click the Browse button, shown at left, next to the Export file path field.
3. Navigate to the National Instruments\Sound and Vibration
Assistant\Examples\Tutorial\ directory and enter octave.txt as the filename.
Chapter 4 Performing Sound and Vibration Measurements
© National Instruments 4-9 Getting Started with the NI Sound and Vibration Assistant
4. Select Overwrite from the If file already exists pull-down menu.
5. Run the project to save the resulting signal to the specified ASCII file.
6. Save the project.
7. Navigate to the folder containing octave.txt and open the file in a text editor to view the saved data.
Exporting Signals to Microsoft ExcelTo export signal data to Microsoft Excel, launch Excel and drag the output signal of a step in the Sound and Vibration Assistant to an Excel spreadsheet. The data is tab delimited.
You also can right-click a display on the Data View tab and select Export To»Microsoft Excel from the shortcut menu to export data from the selected graph on the Data View tab. You also can open saved data, such as octave.txt, in Microsoft Excel.
Printing SignalsTo print an image of a graph, display the Data View tab and select File»Print»Print Data View.
Creating Reports in the Sound and Vibration AssistantSelect View»Project Documentation to display the Project Documentation tab. You can describe the project using text and images such as graphs. You can drag a step output from the Project View into the Project Documentation tab to display a graph of the output signal. If the project is running, the graph on the Project Documentation tab automatically updates to match the current value of the step output.
To print the documentation, open the Project Documentation tab and select File»Print»Print Documentation, or click the Print Documentation button.
Exporting Documentation to HTMLTo export the documentation to HTML, display the Project Documentation tab and select File»Export»Export Documentation to HTML. You also can right-click on the Project Documentation tab and select Export Documentation to HTML from the shortcut menu to export the documentation.
© National Instruments 5-1 Getting Started with the NI Sound and Vibration Assistant
5Performing Sweep Measurements
You can create sweep operations in the Sound and Vibration Assistant to automate measurements in order to characterize and validate designs. You can use the sweep measurements to gather data from devices under test over a range of conditions in order to document the performance of the devices. For example, you can use sweep operations to vary the frequency of a stimulus signal or vary the level of a supply voltage while taking measurements.
This chapter teaches you how to set up sweep operations using the Sweep step in the Sound and Vibration Assistant. You learn how to characterize the performance of a filter by sweeping through a range of frequency values and measuring the output of the filter. You also learn how to display sweep results and perform multidimensional sweeps for more complex measurements.
You can sweep physical signals generated from National Instruments arbitrary waveform generators, function generators, dynamic signal acquisition (DSA) devices, or multifunction I/O (MIO) devices.
Note You must have DAQ hardware with analog inputs and outputs and NI-DAQmx 8.9.5 or later installed to complete the following exercises with physical signals.
Defining Sweep Ranges and OutputsYou can use the Sweep step in the Sound and Vibration Assistant to define automated measurements for complex, repeatable sweep operations.
Creating a Project to SweepComplete the following steps to create a project to sweep:
1. Open a new project.
2. Save the project as My Sweep.seproj in a convenient location.
3. Add the Create Analog Signal step.
Chapter 5 Performing Sweep Measurements
Getting Started with the NI Sound and Vibration Assistant 5-2 ni.com
4. On the Configuration tab in the Step Setup tab, enter 10000 in the Block size (samples) field.
5. Add the Filter step.
6. On the Configuration tab in the Step Setup tab, select Bandpass from the Type pull-down menu. Notice that the Filter Magnitude Response (dB) graph updates as you change options in the Filter Specifications section.
7. Enter 2000 in the Low cutoff (Hz) field and 5000 in the High cutoff (Hz) field.
8. Click the Add Step button and select Analysis»Frequency-Domain Measurements»Tone Measurements.
9. Click the Single-Tone Measurements tab in the Step Setup tab and place checkmarks in the THD+N and THD checkboxes and remove checkmarks from the SNR, SINAD, and SFDR checkboxes in the Distortion and Channel Evaluation sections.
10. Save the project.
Defining the SweepComplete the following steps to define a frequency range to sweep through a filter:
1. Click the Add Step button and select Execution Control»Sweep.
2. Click the Add button in the Sweep Configuration tab in the Step Setup tab to display the list of sweepable parameters from each step in the project, as shown in Figure 5-1.
Figure 5-1. Sweep Parameter Selection Dialog Box
Chapter 5 Performing Sweep Measurements
© National Instruments 5-3 Getting Started with the NI Sound and Vibration Assistant
3. Select the Frequency (Hz) parameter under Create Analog Signal: sine wave and click the OK button.
Notice in the Project View, as shown in Figure 5-2, that the Sweep step encloses the Create Analog Signal step, which provides the signal to sweep.
Figure 5-2. Sweep Step Enclosing Create Analog Signal Step
4. On the Sweep Configuration tab, select Exponential from the Type pull-down menu.
5. Enter 2000 in the Start: Frequency (Hz) field and 5000 in the Stop: Frequency (Hz) field.
Chapter 5 Performing Sweep Measurements
Getting Started with the NI Sound and Vibration Assistant 5-4 ni.com
6. Enter 20 in the Number of points field.
The Sweep Configuration tab appears as shown in Figure 5-3.
Figure 5-3. Sweep Configuration Tab
You configured the Sweep step to specify a range of values to iterate through the Frequency (Hz) parameter of the Create Analog Signal step. The Create Analog Signal step uses the defined frequency range to generate a sine wave at each of these frequencies. You can use the Sweep step to iterate through any sweepable parameter value of any sweepable step in a project.
7. Click the Sweep Output tab in the Step Setup tab.
8. Click the Add button to display the list of sweepable outputs from each step in the project.
Chapter 5 Performing Sweep Measurements
© National Instruments 5-5 Getting Started with the NI Sound and Vibration Assistant
9. Select the fundamental amplitude output under Tone Measurements and click the OK button to plot this measurement against the swept Frequency (Hz) parameter.
Notice that the Sweep step creates a loop around all the steps in the Project View, as shown in Figure 5-4, to include all the steps in the sweep operation.
Figure 5-4. Sweep Step Including All Steps in the Sweep Operation
10. Save the project.
Plotting Sweep ResultsComplete the following steps to run the sweep measurement:
1. Drag the filtered data output of the Filter step to the Data View tab.
2. Drag the sine wave output of the Create Analog Signal step to the Data View tab.
3. Select the Run Once option to execute the sweep measurement.
Chapter 5 Performing Sweep Measurements
Getting Started with the NI Sound and Vibration Assistant 5-6 ni.com
The project generates a sine wave stimulus signal using the Create Analog Signal step, passes it through a bandpass filter using the Filter step, and measures the THD+N and THD levels of the filter output using the Tone Measurements step. The Filter step acts as a simulated unit under test, so the project uses no hardware.
The sine wave output on the graph iterates through the specified range of frequencies.
4. Drag the fundamental amplitude vs. Frequency (Hz) signal from the Sweep loop to the Data View tab to display the output of the sweep.
The Sound and Vibration Assistant creates a new graph. The data from the sweep operation is an XY array that requires a separate graph, as shown in Figure 5-5.
Figure 5-5. Sweep Project
5. Select the Run Once option to execute the sweep measurement.
The frequency response of the Filter step plots on the new graph while the project runs. The graph displays the transfer function of the filter, which is the fundamental amplitude output versus the frequency.
Chapter 5 Performing Sweep Measurements
© National Instruments 5-7 Getting Started with the NI Sound and Vibration Assistant
Note By default, the Sound and Vibration Assistant does not clear displays on the Data View tab between iterations of a sweep. Because the frequency response of the Filter step is the same for each iteration, the graph that displays the signal does not appear to update when you run the project. You can use the Data page of the Options dialog box to specify whether the Sound and Vibration Assistant clears displays between iterations of a sweep. Select Tools»Options to display the Options dialog box.
6. Save and close the project.
You can use the Sweep step to sweep multiple parameters simultaneously by adding additional parameters on the Sweep Configuration tab of the Step Setup tab. Sweeping two or more parameters simultaneously is called a parallel sweep. For example, if you want to vary the amplitude of a stimulus signal, you can run a parallel sweep to maximize the precision of the acquisition by varying the input range of a digitizer or DSA device as you vary the signal level. As the signal level increases, you can increase the input range of the measurement device to ensure you use the entire resolution for the measurement.
© National Instruments 6-1 Getting Started with the NI Sound and Vibration Assistant
6Performing Order Analysis
If you have the Sound and Vibration Measurement Suite, you can use the Sound and Vibration Assistant to perform order-domain measurements on a device under test (DUT). You can perform these measurements, also known as order analysis, either during the test or after the test. This chapter provides examples that demonstrate both situations.
In the first example, you perform order analysis on a universal file format binary (UFF58b) file that contains data acquired from a run-up test on a four-cylinder engine. This example demonstrates how you can use acquired test data to perform order analysis.
In the second example, you use the Playback feature of the Sound and Vibration Assistant to simulate a run-up test on a PC fan. During this simulated test, you perform order analysis on the acquired data. This example demonstrates how you can acquire and analyze data simultaneously.
Both examples involve run-up tests. Because run-up tests change the rotational speed of the DUT over the entire operating range, run-up tests are useful for identifying orders that cause significant noise or vibration in the DUT.
Performing Order Analysis on Logged DataThe example in this section uses logged data from a run-up test on a four-cylinder engine. You perform order analysis on this data to identify the orders that cause significant engine noise. This data includes the following two signals from the test:
⢠A tachometer pulse signal that measures the rotational speed of the engine. The tachometer generates one pulse per revolution of the engine.
⢠A microphone signal that measures the sound the engine emits. In this situation, the measured sound represents unwanted noise.
Chapter 6 Performing Order Analysis
Getting Started with the NI Sound and Vibration Assistant 6-2 ni.com
In the following sections, you complete these steps:
1. Load the acquired data from a UFF58b file.
2. Measure the rotational speed of the engine during the course of the test.
3. Identify the orders that cause significant engine noise.
4. Analyze the magnitude of these specific orders.
Loading the UFF58b Data FileComplete the following steps to open the example project:
1. Select Help»Open Example»Sound and Vibration Assistant, double-click the Tutorial folder, and double-click SVA_Order Analysis From UFF File (Tutorial).seproj.
2. Save the project as My Engine.seproj in a convenient location.
The Data View tab shows two signals: tachometer and sound pressure. The tachometer signal in the upper-left corner shows how the tachometer voltage changes during the run-up test. The sound pressure signal in the upper-right corner shows how the sound pressure, or noise, changes during the same test.
Measuring the Rotational Speed of the EngineBecause order analysis measures the rotational speed of the DUT, the first step in any order analysis procedure is measuring the speed profile of the DUT. A speed profile illustrates how the rotational speed of the DUT changes over time. You measure rotational speed in revolutions per minute (RPM).
You can use the Analog Tacho Processing step to acquire a speed profile of the DUT. This step converts a tachometer pulse signal into RPM. Complete the following steps to measure the speed profile of the engine:
1. Click the Add Step button and select Analysis»Order-Domain Measurements»Analog Tacho Processing.
2. On the Input tab in the Step Setup tab, notice the Tachometer signal is the source of analysis, or Input signal. This signal appears in the Tachometer signal graph on the upper portion of the Step Setup tab.
The middle portion of the Step Setup tab displays the Speed profile graph. This graph represents an RPM measurement this step makes on the Tachometer signal. This measurement shows how the engine RPM changes during the course of the run-up test.
Chapter 6 Performing Order Analysis
© National Instruments 6-3 Getting Started with the NI Sound and Vibration Assistant
Figure 6-1 shows the Speed profile graph in the Step Setup tab.
Figure 6-1. Speed Profile of the Engine
In Figure 6-1, notice how the engine RPM increases from about 1300 RPM to 5200 RPM over 10.84 seconds. You can use this speed profile as the basis for order analysis.
3. Drag the speed profile output of the Analog Tacho Processing step to the empty graph of the Data View tab.
The Data View tab now displays three graphs as a function of time: the tachometer graph shows the tachometer voltage amplitude, the sound pressure graph shows the sound pressure level, and the speed profile graph shows the engine RPM.
4. Save the project.
Identifying Dominant OrdersNow that you have used the tachometer signal to acquire a speed profile, you can use the microphone signal to perform order analysis. A common tool for this analysis is a spectral map in the form of a colormap. A colormap shows you an overview of how the amplitude of a signal relates to either frequency or order. A colormap also shows how the amplitude of a signal relates to either time or speed.
You also can configure a colormap to display order as a function of RPM. An RPM-order colormap displays how the sound pressure levels at different orders change with engine RPM. You can use a colormap to identify the dominant orders, which are orders that cause high sound pressure levels.
Chapter 6 Performing Order Analysis
Getting Started with the NI Sound and Vibration Assistant 6-4 ni.com
You use the Spectral Map step to display a colormap. Complete the following steps to display a colormap of the recorded sound pressure levels:
1. Click the Add Step button and select Analysis»Order-Domain Measurements»Spectral Map.
2. On the Input/Output tab in the Step Setup tab, select time waveforms»Sound Pressure from the Input signal pull-down menu.
The colormap now displays the noise frequency, in Hz, as a function of time. The colormap also displays the amplitude of the pressure level as a range of colors. Low sound pressure levels are in green and high sound pressure levels are in red.
Notice that several high sound pressure levels occur between 600 to 750 Hz, as shown in Figure 6-2. This frequency range probably includes a resonant frequency of some parts of the engine.
Figure 6-2. Frequency-Time Colormap Display
3. Click the Configuration tab in the Step Setup tab and select RPM-Order from the Plot type pull-down menu.
The colormap still shows the amplitude of the sound pressure level in color. However, the colormap now shows how order changes with RPM instead of how frequency changes with time.
Chapter 6 Performing Order Analysis
© National Instruments 6-5 Getting Started with the NI Sound and Vibration Assistant
4. Double-click the maximum value on the x-axis and enter 20 as the new value. This adjusts the scale of the colormap to focus on orders 0 to 20.
The Colormap display appears as shown in Figure 6-3.
Figure 6-3. RPM-Order Colormap Display
In Figure 6-3, notice that the 2nd order contains red areas from 2500 to 3500 RPM and from 4100 to 5100 RPM. These red areas indicate high sound pressure levels that occur in these RPM ranges. This behavior at the 2nd order is expected because combustion, which causes noise, occurs twice per revolution in a four-cylinder engine.
Also notice that high sound pressure levels exist at the 8th, 10th, 12th, and 14th orders at different RPM ranges. These orders are harmonics, or multiples, of the 2nd order. Therefore, you also can expect these orders to cause significant noise. All of these orders are dominant orders.
5. Drag the colormap output of the Spectral Map step to the Data View tab. This tab now displays four graphs.
6. Repeat step 4 for the colormap display on the Data View tab by clicking the maximum value and replacing the current value with 20.
7. Save the project.
Chapter 6 Performing Order Analysis
Getting Started with the NI Sound and Vibration Assistant 6-6 ni.com
Tracking the Magnitude of Specific OrdersThe last step of the order analysis procedure is tracking the magnitude of two dominant orders. The colormap displays this data already. However, a colormap shows a broad overview of this data. A magnitude graph provides a simpler way of tracking the numeric data for specific orders.
You can use the Order Tracking step to show the magnitude of one or more orders as a function of RPM. Complete the following steps to display the magnitude of the 10th and 12th orders over the entire RPM range:
1. Click the Add Step button and select Analysis»Order-Domain Measurements»Order Tracking.
2. On the Input/Output tab in the Step Setup tab, select time waveforms»Sound Pressure from the Input signal pull-down menu.
3. Click the Configuration tab in the Step Setup tab and enter 10 and 12 in the Orders to track list.
4. Select Speed from the X-axis selection pull-down menu to change the x-axis to RPM.
Tip You can configure the Order Tracking step in the Step Setup tab to display the order magnitude as a function of time, speed, or number of revolutions.
5. Drag the magnitudes output of the Order Tracking step to the Data View tab.
6. Right-click the new magnitudes graph and select Visible Items»Legend.
The magnitudes graph appears as shown in Figure 6-4.
Figure 6-4. Magnitude Plot of 10th and 12th Orders
In Figure 6-4, notice the high magnitudes for the 12th order exist around 3400 RPM. This peak corresponds to the red area for the 12th order on the colormap. The magnitudes graph provides this information in a format that is simpler to read than a colormap.
Chapter 6 Performing Order Analysis
© National Instruments 6-7 Getting Started with the NI Sound and Vibration Assistant
The peak for the 10th order exists around 4200 RPM. Again, the colormap confirms this analysis.
7. Save and close the project.
Performing Order Analysis During Data AcquisitionYou can use the Sound and Vibration Assistant to perform order analysis while acquiring data from NI DAQ hardware, such as NI DSA devices.
In this section, you play back a file that contains logged data. You can perform order analysis on this data while the data plays back. In this way, the example simulates acquiring test data while performing order analysis on this data simultaneously. Refer to the Playing Back Data section of Chapter 4, Performing Sound and Vibration Measurements, for information about playing back logged data.
In the example in this section, the DUT is a PC fan with seven blades, mounted on a motor with four coils. The DSA device measures the following two signals from this fan:
⢠A tachometer pulse signal that measures the speed of the fan. The tachometer generates two pulses per revolution of the fan.
⢠An accelerometer signal that measures the vibration of the fan.
You perform order analysis on these signals to identify the orders that cause significant fan vibration.
In the following sections, you complete these steps:
1. Import and play back the acquired data from a Sound and Vibration Assistant log file to simulate real-time data acquisition.
2. Measure the rotational speed of the fan during the course of the test.
3. Identify the orders that cause significant fan vibration.
4. Analyze the magnitude of these specific orders.
Importing the Logged Data FileComplete the following steps to open the data from the PC fan run-up test:
1. Open a new project.
2. Select File»Import»Logged Signals from SignalExpress TDMS Files, navigate to the National Instruments\Sound
and Vibration Assistant\Examples\Log File - PC Fan
Run-up\ directory, and click the Current Folder button.
Chapter 6 Performing Order Analysis
Getting Started with the NI Sound and Vibration Assistant 6-8 ni.com
3. Save the project as My PC Fan.seproj in a convenient location.
4. Notice that PC Fan Run-up appears under Logs in the Logged Data window.
5. Select Playback»PC Fan Run-up from the work area pull-down menu. The Data View tab now displays a timeline slide at the top of the tab.
Measuring the Rotational Speed of the PC FanAs in the Performing Order Analysis on Logged Data section of this chapter, you must measure the speed profile of the DUT before you can perform any order analysis. Complete the following steps to measure the speed profile of the PC fan:
1. Add the Analog Tacho Processing step.
2. On the Input tab in the Step Setup tab, notice that the default Input signal is the Analog Tachometer Signal.
3. Click the Configuration tab in the Step Setup tab and enter 2 in the Pulse/Revolution field because for this PC fan, two pulses represent one revolution. Notice that the RPM values on the y-axis of the Speed profile graph reduce by half.
4. Drag the speed profile output of the Analog Tacho Processing step to the Data View tab.
5. Run the project to play back the data.
Notice that the speed profile graph updates as the timeline slide moves. This movement shows how the Sound and Vibration Assistant calculates the fan speed as you acquire a tachometer signal from the DUT.
Tip While in the Playback work area, you can select Operate»Set Playback Rate to change the speed of the data playback.
6. Save the project.
Identifying Dominant OrdersAs in the Performing Order Analysis on Logged Data section of this chapter, the next step is identifying the dominant orders. However, instead of using a colormap, the following steps use an order spectrum graph to identify these orders. Order spectrum graphs are useful for displaying numeric data, whereas colormaps are useful for obtaining a broad overview of this data.
Chapter 6 Performing Order Analysis
© National Instruments 6-9 Getting Started with the NI Sound and Vibration Assistant
You can use the Order Spectrum step to display an order spectrum graph. Complete the following steps to measure the order spectrum of the vibration signal:
1. Click the Add Step button and select Analysis»Order-Domain Measurements»Order Spectrum.
2. On the Input tab in the Step Setup tab, select Active Log»Acquired Signals»Vibration Signal from the Input signal pull-down menu. The Order spectra graph updates to display vibration amplitude as a function of order during a single iteration of the run-up test.
3. Click the Spectrum Settings tab in the Step Setup tab and select Linear from the Magnitude scale pull-down menu to provide a linear scale that is easier to read than decibels.
On the Order spectra graph, notice that significant vibrations do not occur after the 10th order. You can change the scale of the graph to focus on the orders that do cause vibrations.
4. Enter 10 as the new value for the Maximum order field.
The Order spectra graph appears as shown in Figure 6-5.
Figure 6-5. Order Spectra Graph of Vibration Signal
5. Drag the order spectrum output of the Order Spectrum step to the Data View tab. This tab now displays the speed profile graph and the order spectrum graph.
6. Run the project. Notice that the order spectrum graph updates as the project runs.
Chapter 6 Performing Order Analysis
Getting Started with the NI Sound and Vibration Assistant 6-10 ni.com
7. Right-click the order spectrum graph and select View As»Waterfall Graph. A waterfall graph is useful for showing a history of data throughout the test.
8. Run the project. Notice that the waterfall graph displays a history of the order spectra throughout the run-up test. The z-axis is the test iteration number.
Figure 6-6 shows the waterfall graph.
Figure 6-6. Waterfall Graph of the Order Spectra
The waterfall graph shows peaks at the 4th and 7th orders. These orders are the dominant orders. This behavior is expected because the four coils in the PC fan motor cause vibrations at the 4th order and the seven blades cause vibrations at the 7th order.
9. Save the project.
Tracking the Magnitude of Specific OrdersThe last step is tracking the magnitude of the vibration at these orders throughout the test. Complete the following steps to display the magnitude of the vibration at the 4th and 7th orders:
1. Add the Order Tracking step.
2. On the Input/Output tab in the Step Setup tab, select Active Log»Acquired Signals»Vibration Signal from the Input signal pull-down menu.
Chapter 6 Performing Order Analysis
© National Instruments 6-11 Getting Started with the NI Sound and Vibration Assistant
3. Click the Configuration tab in the Step Setup tab and enter 4 and 7 in the Orders to track list to track the orders which correspond with the vibrations of the four coils in the PC fan motor and the seven fan blades.
4. Enter 0.5 in the Bandwidth (order) field and enter 500m in the Time interval (s) field.
5. Drag the magnitudes output of the Order Tracking step to the Data View tab. The Data View tab now displays three graphs: the speed profile graph, the waterfall graph of the order spectra, and the magnitude graph of vibrations at the 4th and 7th orders.
6. Right-click the magnitudes graph and select Visible Items»Legend.
7. Run the project. The magnitudes graph displays the magnitude of vibrations at the 4th and 7th orders during the run-up test, as shown in Figure 6-7.
Figure 6-7. Magnitude Plot of 4th and 7th Orders
These peaks are the same as the peaks on the waterfall graph. The highest peak for the vibrations at the 4th order occurred at 13 seconds and the highest peak for the vibrations at the 7th order occurred at 8 seconds.
8. Save and close the project.
© National Instruments 7-1 Getting Started with the NI Sound and Vibration Assistant
7Extending Sound and Vibration Assistant Projects with LabVIEW
You can use the Sound and Vibration Assistant to define automated measurements by using built-in steps for acquiring, generating, analyzing, or logging signals. You can extend the functionality of the Sound and Vibration Assistant with LabVIEW Graphical Programming in the following ways:
⢠You can build VIs in LabVIEW and import them into the Sound and Vibration Assistant to expand the built-in measurement options and provide custom step functionality.
⢠You can convert Sound and Vibration Assistant projects to LabVIEW block diagrams to continue developing the application in LabVIEW. By doing so, you can use the interactive tools in the Sound and Vibration Assistant to define measurement procedures, and then convert the sequence into a LabVIEW block diagram to build automated test applications.
Note References or examples of SignalExpress also apply to the Sound and Vibration Assistant.
Importing LabVIEW VIs into the Sound and Vibration Assistant as Steps
Use the Run LabVIEW VI steps in the Sound and Vibration Assistant to call custom LabVIEW VIs. You can call a LabVIEW VI from the Sound and Vibration Assistant to do the following:
⢠Read or write data to more file formats
⢠Use the VI to further analyze data
⢠Define a measurement algorithm
Chapter 7 Extending Sound and Vibration Assistant Projects with LabVIEW
Getting Started with the NI Sound and Vibration Assistant 7-2 ni.com
Complete the following steps to import a VI from LabVIEW with a Run LabVIEW VI step:
Note You must have LabVIEW 8.6 or later to complete the exercises in this section.
1. Select Help»Open Example»SignalExpress, double-click the Tutorial folder, and double-click User Step.seproj.
This project uses the Create Analog Signal step to generate a signal.
2. Save the project as My Imported VI.seproj in a convenient location.
3. Click the Add Step button and select Run LabVIEW VI»Run LabVIEW 8.6 VI.
The VI you run in this exercise was saved in LabVIEW 8.6. You must use the version of the Run LabVIEW VI step that matches the version of LabVIEW you saved the VI in.
4. Click the Browse button in the Select VI section on the Step Setup tab and select Limiter-LV86.vi in the SignalExpress\Examples\Tutorial directory.
Note If you receive an error stating Unknown LabVIEW Version, open the VI in LabVIEW 8.6 and save the VI. This ensures that the VI is saved in the correct version of LabVIEW and executes with the Run LabVIEW 8.6 VI step in the Sound and Vibration Assistant without error. Then click the Refresh VI button on the Step Setup tab.
Chapter 7 Extending Sound and Vibration Assistant Projects with LabVIEW
© National Instruments 7-3 Getting Started with the NI Sound and Vibration Assistant
Verify that the Step Setup tab appears as shown in Figure 7-1.
Figure 7-1. Limiter VI Step Setup Tab
The Limiter-LV86 VI accepts a time-domain waveform as an input, clips the signal above and below values that you specify in the Step Setup tab, and returns the clipped waveform as an output signal.
When you import a LabVIEW VI, the Sound and Vibration Assistant maps the inputs of the VI as input signals or parameters and the outputs of the VI as output signals.
An input signal appears in the Project View as an input to a step, which means that you can pass signals as inputs to a VI. A parameter is a value you can configure in the Step Setup tab of a step. You also can sweep parameters dynamically using the Sweep step. In this project, the VI has an input signal, Time waveform in, and scalar parameters, Upper limit and Lower limit.
5. Click the Configure VI tab in the Step Setup tab to display the VI.
6. Enter 100 in the Upper limit field and 20 in the Lower limit field to specify the limits.
Chapter 7 Extending Sound and Vibration Assistant Projects with LabVIEW
Getting Started with the NI Sound and Vibration Assistant 7-4 ni.com
7. Drag the Clipped waveform out output of the Run LabVIEW 8.6 VI step to the Data View tab.
8. Drag the sine wave output of the Create Analog Signal step to the Data View tab. The graph on the Data View tab displays the two signals: the complete sine wave and the clipped waveform, as shown in Figure 7-2.
Figure 7-2. Sine Waveform with Clipped Waveform
9. Save and close the project.
Refer to the Sound and Vibration Assistant Help, available by selecting Help»Sound and Vibration Assistant Help, for more information about
Chapter 7 Extending Sound and Vibration Assistant Projects with LabVIEW
© National Instruments 7-5 Getting Started with the NI Sound and Vibration Assistant
using LabVIEW VIs in the Sound and Vibration Assistant and developing VIs that work well in the Sound and Vibration Assistant.
Converting Sound and Vibration Assistant Projects to LabVIEW Block Diagrams
The Sound and Vibration Assistant can convert Sound and Vibration Assistant projects into LabVIEW block diagrams.
Note To convert a Sound and Vibration Assistant project to a LabVIEW VI, you must have the LabVIEW 8.2 Full Development System or later installed. You also must have the LabVIEW Support for Sound and Vibration Analysis component installed to the version of LabVIEW to which you want to generate code.
Complete the following steps to convert a Sound and Vibration Assistant project to a LabVIEW block diagram:
1. Select Help»Open Example»Sound and Vibration Assistant, double-click the Tutorial folder, and double-click SVA_Analyze Signal From UFF File (Tutorial).seproj.
2. Select Tools»Generate Code»LabVIEW Diagram.
Note If you have multiple versions of LabVIEW installed, this option generates code in the last version of LabVIEW you launched.
3. Save the new VI as My LabVIEW VI.vi and click the OK button. The Sound and Vibration Assistant generates the new VI and opens the VI in the last version of LabVIEW you opened.
The resulting LabVIEW block diagram reflects the execution of the project in the Sound and Vibration Assistant. The LabVIEW block diagram consists of LabVIEW Express VIs wired together. Each Express VI correlates to a step in the Sound and Vibration Assistant project. You can double-click an Express VI to display the same Step Setup as in the Sound and Vibration Assistant. You also can right-click an Express VI and select Open Front Panel from the shortcut menu to convert the Express VI into a LabVIEW subVI. When you convert an Express VI into a subVI, you cannot convert the subVI back into an Express VI. You can view the block diagram to see how the LabVIEW block diagram executes and modify the functionality of the VI.
The LabVIEW front panel displays the time waveforms input signal, the spectrum input signal, the outputs of averaging done and
Chapter 7 Extending Sound and Vibration Assistant Projects with LabVIEW
Getting Started with the NI Sound and Vibration Assistant 7-6 ni.com
averages completed from the Power Spectrum step, and the outputs of peaks and number of peaks from the Peak Search step.
4. Double-click the UFF58 Read File Express VI on the block diagram. Verify that the UFF58 file in the UFF58 filepath field is Triangle wave (1000 Hz).uff and click the OK button.
5. Click the Run button, shown at left, to run the VI. On the front panel in LabVIEW, the time waveforms graph and the spectrum graph update continuously, as shown in Figure 7-3, until you stop the VI.
Figure 7-3. LabVIEW Front Panel
6. Click the Stop button, shown at left, to stop the VI.
7. Save the VI and close LabVIEW.
Note When you convert a Sound and Vibration Assistant project with logging or hardware configuration, the Sound and Vibration Assistant generates a LabVIEW block diagram with only one Express VI containing the entire contents of the project. You cannot convert the generated Express VI into a subVI.
© National Instruments 8-1 Getting Started with the NI Sound and Vibration Assistant
8Where to Go from Here
Refer to the following resources for more information about the Sound and Vibration Assistant.
Sound and Vibration Assistant Example ProjectsThe Sound and Vibration Assistant provides example projects that demonstrate capabilities of the Sound and Vibration Assistant with National Instruments hardware. Access these projects by selecting Help»Open Example»Sound and Vibration Assistant. Review these examples to learn more about the features of the Sound and Vibration Assistant or to start with a project that closely resembles your needs.
Opening a Sound and Vibration Example SolutionYou can find completed solutions for each chapter of this manual by selecting Help»Open Example»Sound and Vibration Assistant. For example, open SVA_Analyze Signal from UFF File.seproj to view the completed solution for Chapter 2, Working with Projects.
Using Hardware with the Sound and Vibration AssistantThe Sound and Vibration Assistant supports a variety of National Instruments hardware for acquiring and generating signals. You can generate or acquire and log analog signals in the Sound and Vibration Assistant using National Instruments MIO devices, DSA devices, high-speed digitizers, or arbitrary waveform generator and function generator devices. You also can synchronize multiple devices in a system by sharing clocks and trigger signals between devices. You must have NI-DAQmx 8.9.5 or later installed to use the Sound and Vibration Assistant with NI hardware.
Refer to the Sound and Vibration Assistant Help, available by selecting Help»Sound and Vibration Assistant Help, for more information about using hardware with the Sound and Vibration Assistant.
Chapter 8 Where to Go from Here
Getting Started with the NI Sound and Vibration Assistant 8-2 ni.com
Web ResourcesRefer to the National Instruments Web site at ni.com/ soundandvibration for additional resources.
© National Instruments A-1 Getting Started with the NI Sound and Vibration Assistant
ATechnical Support and Professional Services
Log in to your National Instruments ni.com User Profile to get personalized access to your services. Visit the following sections of ni.com for technical support and professional services:
⢠SupportâTechnical support at ni.com/support includes the following resources:
â Self-Help Technical ResourcesâFor answers and solutions, visit ni.com/support for software drivers and updates, a searchable KnowledgeBase, product manuals, step-by-step troubleshooting wizards, thousands of example programs, tutorials, application notes, instrument drivers, and so on. Registered users also receive access to the NI Discussion Forums at ni.com/forums. NI Applications Engineers make sure every question submitted online receives an answer.
â Standard Service Program MembershipâThis program entitles members to direct access to NI Applications Engineers via phone and email for one-to-one technical support, as well as exclusive access to self-paced online training modules at ni.com/self-paced-training. All customers automatically receive a one-year membership in the Standard Service Program (SSP) with the purchase of most software products and bundles including NI Developer Suite. NI also offers flexible extended contract options that guarantee your SSP benefits are available without interruption for as long as you need them. Visit ni.com/ssp for more information.
For information about other technical support options in your area, visit ni.com/services, or contact your local office at ni.com/contact.
Appendix A Technical Support and Professional Services
Getting Started with the NI Sound and Vibration Assistant A-2 ni.com
⢠Training and CertificationâVisit ni.com/training for training and certification program information. You can also register for instructor-led, hands-on courses at locations around the world.
⢠System IntegrationâIf you have time constraints, limited in-house technical resources, or other project challenges, National Instruments Alliance Partner members can help. To learn more, call your local NI office or visit ni.com/alliance.
You also can visit the Worldwide Offices section of ni.com/niglobal to access the branch office Web sites, which provide up-to-date contact information, support phone numbers, email addresses, and current events.
NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã
NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã ã¹ã¿ãŒãã¢ããã¬ã€ã
NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã
2013 幎 5 æ
371938D
ãµããŒã
ãµããŒã
æè¡ãµããŒãã®ãæ¡å ni.com/support
äžçåå°ã®ãªãã£ã¹
ni.com/niglobalããããåãåããå ããµããŒãé»è©±çªå·ãé»åã¡ãŒã«ã¢ãã¬ã¹ãçŸåšå®æœäžã®ã€ãã³ãã«é¢ããææ°æ å ±ãæäŸããååœçŸå°ãªãã£ã¹ã®ãŠã§ãããŒãžã«ã¢ã¯ã»ã¹ã§ããŸãã
æ¥æ¬ãã·ã§ãã«ã€ã³ã¹ãã«ã¡ã³ãæ ªåŒäŒç€Ÿ
ã 105-0012 æ±äº¬éœæž¯åºè倧é 1-9-9 éæäžåç£è倧éãã« 8F/9F Tel: 0120-527196
National Instruments ç±³åœæ¬ç€Ÿ11500 North Mopac Expressway Austin, Texas 78759-3504 USA Tel: 512 683 0100
ãµããŒãæ å ±ã®è©³çŽ°ã«ã€ããŠã¯ããæè¡ãµããŒãããã³ãããã§ãã·ã§ãã«ãµãŒãã¹ããåç §ããŠãã ãããã
ã·ã§ãã«ã€ã³ã¹ãã«ã¡ã³ãã®ããã¥ã¡ã³ãã«é¢ããŠãæèŠããå¯ãããã ãå Žåã¯ããã·ã§ãã«ã€ã³ã¹ãã«ã¡ã³ã
ã®ãŠã§ããµã€ããni.com/jp/infoã«ãã Info Code ã«ãfeedbackããšãå ¥åãã ããã
© 2007â2013 National Instruments. All rights reserved.
å¿ ããèªã¿ãã ãã
ä¿èšŒNIã®ãœãããŠã§ã¢è£œåãèšé²ãããŠããåªäœã¯ãçŽ æããã³è£œé æè¡äžã®æ¬ é¥ã«ããããã°ã©ãã³ã°äžã®åé¡ã«å¯ŸããŠãåé æžãªã©ã®æžé¢ã«ãã£ãŠç€ºãããåºè·æ¥ãã 90æ¥éä¿èšŒèŽããŸããNIã¯ãä¿èšŒæéäžã«ãã®ãããªæ¬ é¥ã®éç¥ãåãåã£ãå ŽåãåŒç€Ÿã®è£éã«ãããããã°ã©ãã³ã°ã®æ瀺ã©ããã«å®è¡ã§ããªããœãããŠã§ã¢åªäœãä¿®çã亀æèŽããŸããNIã¯ããœãããŠã§ã¢ã®æäœãäžæãããªãããšãããã³æ¬ é¥ã®ãªãããšãä¿èšŒèŽããŸããã
ã客æ§ã¯ãä¿èšŒã®å¯Ÿè±¡ãšãªã補åãNIã«è¿åŽããåã«ãè¿åç¢ºèª (RMA: Return Material Authorization)çªå·ãNIããååŸããããã±ãŒãžå€ã«æèšããå¿ èŠããããŸããNIã¯ãä¿èšŒãåãã§ããéšåãã客æ§ã«è¿åŽãã茞éè²»ãè² æ ããããŸãã
æ¬æžã®å 容ã«ã€ããŠã¯äžå šãæããŠãããæè¡çå 容ã«é¢ãããã§ãã¯ãå ¥å¿µã«è¡ã£ãŠãããŸããæè¡çãªèª€ããŸãã¯èª€æ€ããã£ãå ŽåãNIã¯ãæ¬æžãææããã客æ§ãžã®äºåã®éåãªããæ¬æžã®æ¬¡ã®çãæ¹èšããæš©å©ãæããŸãã誀ããšæãããåæããããŸããããNIãžãé£çµ¡ãã ãããNIã¯ãæ¬æžããã³ãã®å 容ã«ããããŸãã¯ããã«é¢é£ããŠçºçããæ害ã«å¯ŸããŠãäžå責任ãè² ããŸããã
NIã¯ãããã«èšèŒããã以å€ãæ瀺ãŸãã¯é»ç€ºã®ä¿èšŒã¯èŽããŸãããç¹ã«ãååæ§ãŸãã¯ç¹å®çšéãžã®é©åæ§ã«é¢ããä¿èšŒã¯èŽããŸãããNIåŽã®é倱ãŸãã¯äžæ³šæã«ããçºçããæ害ã«å¯Ÿããã客æ§ã®è³ åè«æ±æš©ã¯ãã客æ§ã補åã«æ¯æãããéé¡ãäžéãšããŸããNIã¯ãããŒã¿ã®æ¶å€±ãå©çã®æ倱ã補åã®äœ¿çšã«ããæ倱ãä»éçãŸãã¯éæ¥çæ害ã«å¯ŸããŠããã®æ害ãçºçããå¯èœæ§ãéç¥ãããŠããå Žåã§ããäžåã®è²¬ä»»ãè² ããŸãããNIã®éå®ä¿èšŒã¯ã蚎èšæ¹åŒãå¥çŽäžã®è²¬ä»»ãŸãã¯äžæ³è¡çºã«å¯Ÿãã責任ãåãããé倱責任ãå«ããé©çšãããŸããNIã«å¯Ÿãã蚎èšã¯ã蚎èšåå ã®çºçãã1幎以å ã«æèµ·ããå¿ èŠããããŸããNIã¯ãNIã®åççã«ç®¡çå¯èœãªç¯å²ãè¶ ããåå ã«ããçºçããå±¥è¡é 延ã«é¢ããŠã¯äžåã®è²¬ä»»ãè² ããŸãããææè ãã€ã³ã¹ããŒã«ãæäœãä¿å®ã«é¢ããNIã®æ瀺æžã«åŸããªãã£ããããææè ã«ãã補åã®æ¹é ãä¹±çšã誀çšããŸãã¯äžæ³šæãªè¡åãããã«åé»ããµãŒãžãç«çœã措氎ãäºæ ã第äžè ã®è¡çºããã®ä»ã®åççã«ç®¡çå¯èœãªç¯å²ãè¶ ããäºè±¡ã«ããçºçããæ害ãæ¬ é¥ãåäœäžè¯ãŸãã¯ãµãŒãã¹ã®åé¡ã«ã€ããŠã¯ãæ¬æžã«å®ããä¿èšŒã®å¯Ÿè±¡ãšãªããŸããã
èäœæš©èäœæš©æ³ã«åºã¥ããNational Instruments CorporationïŒç±³åœãã·ã§ãã«ã€ã³ã¹ãã«ã¡ã³ã瀟ïŒã®æžé¢ã«ããäºåã®èš±å¯ãªããæ¬æžã®ãã¹ãŠãŸãã¯äžéšãåçè€åãèšé²ãæ å ±æ€çŽ¢ã·ã¹ãã ãžã®ä¿åãããã³ç¿»èš³ãå«ããé»åçãŸãã¯æ©æ¢°çãªãããªã圢åŒã«ãã£ãŠãè€è£œãŸãã¯è»¢èŒããããšãçŠæ¢ããŸãã
National Instrumentsã¯ä»è ã®ç¥ç財ç£ãå°éããŠãããã客æ§ãåæ§ã®æ¹éã«åŸãããŸããããé¡ãããããŸããNIãœãããŠã§ã¢ã¯èäœæš©æ³ãã®ä»ç¥ç財ç£æš©ã«é¢ããæ³åŸã«ããä¿è·ãããŠããŸããNIãœãããŠã§ã¢ãçšããŠä»è ã«åž°å±ãããœãããŠã§ã¢ãã®ä»ã®ãããªã¢ã«ãè€è£œããããšã¯ãé©çšããã©ã€ã»ã³ã¹ã®æ¡ä»¶ãã®ä»ã®æ³çèŠå¶ã«åŸã£ãŠãã®ãããªã¢ã«ãè€è£œã§ããå Žåã«éãå¯èœã§ãããã®ãšããŸãã
ãšã³ããŠãŒã¶äœ¿çšèš±è«Ÿå¥çŽããã³ä»ç€Ÿè£œåã®æ³ç泚æäºé ãšã³ããŠãŒã¶äœ¿çšèš±è«Ÿå¥çŽïŒEULAïŒããã³ä»ç€Ÿè£œåã®æ³ç泚æäºé ã¯ä»¥äžã®å Žæã«ãããŸãã
⢠泚æäºé ã¯ã<National Instruments>Â¥_Legal Informationããã³<National Instruments>ãã£ã¬ã¯ããªã«ãããŸãã
⢠EULAã¯ã<National Instruments>Â¥SharedÂ¥MDFÂ¥LegalÂ¥Licenseãã£ã¬ã¯ããªã«ãããŸãã
⢠NI補åãšãšãã«äœæããã€ã³ã¹ããŒã©ã«æ³åŸæ å ±ãçµã¿èŸŒãæ¹æ³ã«ã€ããŠã¯ã<National Instruments>Â¥_Legal Information.txtããèªã¿ãã ããã
åæšNational Instrumentsã®åæšã®è©³çŽ°ã«ã€ããŠã¯ãni.com/trademarksã«æ²èŒãããŠãããNI Trademarks and Logo Guidelinesã ãã芧äžããã
ARM, Keil, and ΌVision are trademarks or registered of ARM Ltd or its subsidiaries.LEGO, the LEGO logo, WEDO, and MINDSTORMS are trademarks of the LEGO Group. ©2013 The LEGO Group.TETRIX by Pitsco is a trademark of Pitsco, Inc.©2013FIELDBUS FOUNDATION⢠and FOUNDATION⢠are trademarks of the Fieldbus Foundation.EtherCAT® is a registered trademark of and licensed by Beckhoff Automation GmbH.CANopen® is a registered Community Trademark of CAN in Automation e.V.DeviceNet⢠and EtherNet/IP⢠are trademarks of ODVA.Go!, SensorDAQ, and Vernier are registered trademarks of Vernier Software & Technology. Vernier Software & Technology and vernier.com are trademarks or trade dress.Xilinx is the registered trademark of Xilinx, Inc.Taptite and Trilobular are registered trademarks of Research Engineering & Manufacturing Inc.FireWire® is the registered trademark of Apple Inc.Linux® is the registered trademark of Linus Torvalds in the U.S. and other countries.Handle Graphics®, MATLAB®, Real-Time Workshop®, Simulink®, Stateflow®, and xPC TargetBox® are registered trademarks, and TargetBox⢠and Target Language Compiler⢠are trademarks of The MathWorks, Inc.Tektronix®, Tek, and Tektronix, Enabling Technology are registered trademarks of Tektronix, Inc.
The Bluetooth® word mark is a registered trademark owned by the Bluetooth SIG, Inc.The ExpressCard⢠word mark and logos are owned by PCMCIA and any use of such marks by National Instruments is under license.The mark LabWindows is used under a license from Microsoft Corporation. Windows is a registered trademark of Microsoft Corporation in the United States and other countries.
æ¬ææžäžã«èšèŒããããã®ä»ã®è£œååããã³äŒæ¥åã¯ãããããã®äŒæ¥ã®åæšãŸãã¯åå·ã§ãã
National Instruments Alliance Partner Programã®ã¡ã³ããŒã¯National Instrumentsããç¬ç«ããŠããäºæ¥äœã§ãããNational Instrumentsãšäœã代çåºãããŒãããŒã·ãããŸãã¯ãžã§ã€ã³ãã»ãã³ãã£ãŒã®é¢ä¿ã«ãããŸããã
ç¹èš±National Instrumentsã®è£œå /æè¡ãä¿è·ããç¹èš±ã«ã€ããŠã¯ããœãããŠã§ã¢ã§åç §ã§ããç¹èš±æ å ± (ãã«ãâç¹èš±æ å ± )ãã¡ãã£ã¢ã«å«ãŸããŠããpatents.txtãã¡ã€ã«ããŸãã¯ãNational Instruments Patent NoticeãïŒni.com/patentsïŒã®ãã¡ã該åœãããªãœãŒã¹ããåç §ããŠãã ããã
茞åºé¢é£æ³èŠã®éµå®ã«é¢ããæ å ±ãã·ã§ãã«ã€ã³ã¹ãã«ã¡ã³ãã®èŒžåºé¢é£æ³èŠéµå®ã«å¯Ÿããæ¹éã«ã€ããŠããŸãå¿ èŠãªHTSã³ãŒããECCNããã®ä»ã®ã€ã³ããŒã /ãšã¯ã¹ããŒãããŒã¿ãååŸããæ¹æ³ã«ã€ããŠã¯ãã茞åºé¢é£æ³èŠã®éµå®ã«é¢ããæ å ±ã
ïŒni.com/legal/export-complianceïŒãåç §ããŠãã ããã
National Instruments Corporation補åã䜿çšããéã®èŠå(1) National Instruments CorporationïŒä»¥äžãNIããšããïŒã®è£œåã¯ãå€ç§ç§»æ€ãŸãã¯ããã«é¢é£ãã䜿çšã«é©ããæ©åšã®åãã£ã補åãšããŠããŸãã¯åäœäžè¯ã«ãã人äœã«æ·±å»ãªé害ãåãŒããããã®ããçåœç¶æè£ çœ®ã®éèŠãªæ©åšãšããŠèšèšãããŠãããããã®ä¿¡é Œæ§ããããã©ããã®è©Šéšãå®è¡ãããŠããŸããã
(2) äžèšãå«ãããŸããŸãªçšéã«ãããŠãäžé©åãªèŠå ã«ãã£ãŠãœãããŠã§ã¢è£œåã®æäœã®ä¿¡é Œæ§ãæãªããããããããããŸããããã«ã¯ãé»åäŸçµŠã®å€åãã³ã³ãã¥ãŒã¿ããŒããŠã§ã¢ã®èª€äœåãã³ã³ãã¥ãŒã¿ã®ãªãã¬ãŒãã£ã³ã°ã·ã¹ãã ãœãããŠã§ã¢ã®é©åæ§ãã¢ããªã±ãŒã·ã§ã³éçºã«äœ¿çšããã³ã³ãã€ã©ãéçºçšãœãããŠã§ã¢ã®é©åæ§ãã€ã³ã¹ããŒã«æã®ééãããœãããŠã§ã¢ãšããŒããŠã§ã¢ã®äºææ§ã®åé¡ãé»åç£èŠã»å¶åŸ¡æ©åšã®èª€äœåãŸãã¯æ éãã·ã¹ãã ïŒããŒããŠã§ã¢ããã³ /ãŸãã¯ãœãããŠã§ã¢ïŒã®äžæçãªé害ãäºæãã¬äœ¿çšãŸãã¯èª€çšããŠãŒã¶ãŸãã¯ã¢ããªã±ãŒã·ã§ã³èšèšè ã®åŽã®ãã¹ãªã©ããããŸãããããã«éå®ãããŸããïŒä»¥äžããã®ãããªäžé©åãªèŠå ãç·ç§°ããŠãã·ã¹ãã æ éããšããïŒãã·ã¹ãã æ éã財ç£ãŸãã¯äººäœã«å±å®³ãåãŒãå¯èœæ§ïŒèº«äœã®æå·ããã³æ»äº¡ã®å±éºãå«ãïŒã®ããçšéã®å Žåã¯ãã·ã¹ãã æ éã®å±éºãããããã1ã€ã®åœ¢åŒã®ã·ã¹ãã ã«ã®ã¿äŸåãã¹ãã§ã¯ãããŸãããæ害ãæå·ãŸãã¯æ»äº¡ãšãã£ãäºæ ãé¿ããããããŠãŒã¶ãŸãã¯ã¢ããªã±ãŒã·ã§ã³èšèšè ã¯ãé©æ£ã§æ éãªã·ã¹ãã æ éé²æ¢çãåãå¿ èŠããããŸããããã«ã¯ãã·ã¹ãã ã®ããã¯ã¢ãããŸãã¯åæ¢ãå«ãŸããŸãããããã«éå®ãããŸãããåãšã³ããŠãŒã¶ã®ã·ã¹ãã ã¯ã«ã¹ã¿ãã€ãºãããNIã®ãã¹ãçšãã©ãããã©ãŒã ãšã¯ç°ãªãããããããŠãŠãŒã¶ãŸãã¯ã¢ããªã±ãŒã·ã§ã³èšèšè ããNIã®è©äŸ¡ããããšã®ãªãããŸãã¯äºæããŠããªãæ¹æ³ã§ãNI補åãä»ã®è£œåãšçµã¿åãããŠäœ¿çšããå¯èœæ§ããããããNI補åãã·ã¹ãã ãŸãã¯ã¢ããªã±ãŒã·ã§ã³ã«çµ±åããå Žåã¯ããŠãŒã¶ãŸãã¯ã¢ããªã±ãŒã·ã§ã³èšèšè ããNI補åã®é©åæ§ãæ€èšŒã確èªãã責任ãè² ããã®ãšããŸããããã«ã¯ããã®ãããªã·ã¹ãã ãŸãã¯ã¢ããªã±ãŒã·ã§ã³ã®é©åãªèšèšãããã»ã¹ãå®å šã¬ãã«ãå«ãŸããŸãããããã«éå®ãããŸããã
© National Instruments v NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã
ç®æ¬¡
ãã®ããã¥ã¢ã«ã«ã€ããŠå¿ èŠãªãœãããŠã§ã¢...............................................................................................................................viié¢é£ããã¥ã¡ã³ã ...................................................................................................................................vii
第 1 ç« NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã®æŠèŠ
第 2 ç« ãããžã§ã¯ãã®æäœ
ãããžã§ã¯ããéã...............................................................................................................................2-1ãããžã§ã¯ãã®å®è¡ãšä¿¡å·ã®è¡šç€º ....................................................................................................2-3ã¹ããããæ§æãã...............................................................................................................................2-4ã¹ãããã®ç§»åããã³ãšã©ãŒåŠç ....................................................................................................2-6
第 3 ç« ããŒã¿ããã°ãã
ä¿¡å·ãèšé²ãã .......................................................................................................................................3-1ãã°ãããä¿¡å·ã衚瀺ãã .................................................................................................................3-2äºåå®çŸ©ããéå§ / åæ¢æ¡ä»¶ã§ä¿¡å·ããã°ãã ..........................................................................3-4
第 4 ç« é³é¿ãæ¯åã®æž¬å®ãè¡ã
以åã«éé²æžã¿ã®ããŒã¿ãéã.........................................................................................................4-1ããŒã¿ãåçãã ...................................................................................................................................4-2ããŒã¿ã解æãã ...................................................................................................................................4-4äžçŽåç.....................................................................................................................................................4-6é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã«ãããä¿¡å·ã¿ã€ã .....................................................................4-7ä¿¡å·ãä¿åããšã¯ã¹ããŒãããã³å°å·ãã ..................................................................................4-7
ä¿¡å·ããã¡ã€ã«ã«ä¿åãã................................................................................................4-7Microsoft Excel ãžä¿¡å·ããšã¯ã¹ããŒããã..............................................................4-8ä¿¡å·ãå°å·ãã......................................................................................................................4-8é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã§ã¬ããŒããäœæãã ...............................................4-8ããã¥ã¡ã³ãã HTML ã«ãšã¯ã¹ããŒããã ................................................................4-8
第 5 ç« ã¹ã€ãŒã枬å®ãè¡ã
ã¹ã€ãŒãã®ç¯å²ãšåºåãå®çŸ©ãã ....................................................................................................5-1ã¹ã€ãŒããããããžã§ã¯ããäœæãã..........................................................................5-1ã¹ã€ãŒããå®çŸ©ãã .............................................................................................................5-2
ã¹ã€ãŒãã®çµæããããããã.........................................................................................................5-5
ç®æ¬¡
NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã vi ni.com
第 6 ç« æ¬¡æ°è§£æãè¡ã
ãã°ããŒã¿ã«å¯ŸããŠæ¬¡æ°è§£æãè¡ã ............................................................................................... 6-1UFF58b ããŒã¿ãã¡ã€ã«ãããŒããã .......................................................................... 6-2ãšã³ãžã³ã®å転é床ã枬å®ãã ...................................................................................... 6-2䞻次æ°ãç¹å®ãã................................................................................................................. 6-3ç¹å®ã®æ¬¡æ°ã®æ¯å¹ ããã©ããã³ã°ãã ......................................................................... 6-5
ããŒã¿éé²äžã«æ¬¡æ°è§£æãè¡ã ........................................................................................................ 6-7ãã°ããŒã¿ãã€ã³ããŒããã........................................................................................... 6-7PC ãã¡ã³ã®å転é床ã枬å®ãã ................................................................................... 6-8䞻次æ°ãç¹å®ãã................................................................................................................. 6-8ç¹å®ã®æ¬¡æ°ã®æ¯å¹ ããã©ããã³ã°ãã ......................................................................... 6-10
第 7 ç« é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã®ãããžã§ã¯ãã LabVIEW ã§äœ¿çšãã
LabVIEW VI ãã¹ããããšããŠé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã«ã€ã³ããŒããã ............ 7-1é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã®ãããžã§ã¯ãã LabVIEW ã®
ãããã¯ãã€ã¢ã°ã©ã ã«å€æãã ................................................................................................ 7-5
第 8 ç« è©³çŽ°æ å ±
é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã®ãµã³ãã«ãããžã§ã¯ã ............................................................ 8-1é³é¿ / æ¯åãµã³ãã«ã®è§£çãéã.................................................................................................... 8-1é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ããããŒããŠã§ã¢ãšäœµçšãã........................................................ 8-1ãŠã§ããªãœãŒã¹ ....................................................................................................................................... 8-2
ä»é² Aæè¡ãµããŒãããã³ãããã§ãã·ã§ãã«ãµãŒãã¹
© National Instruments vii NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã
ãã®ããã¥ã¢ã«ã«ã€ããŠ
ãã®ããã¥ã¢ã«ã§ã¯ããã·ã§ãã«ã€ã³ã¹ãã«ã¡ã³ãã®é³é¿ / æ¯åèšæž¬ã¢ã·
ã¹ã¿ã³ãã«ã€ããŠèª¬æããŸããé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã®æ©èœã察話
åŒã«äœ¿çšããä¿¡å·ã®éé²ã解æãã§ããŸãã
ãŸãããã®ããã¥ã¢ã«ã«ã¯ãé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ããåããŠäœ¿çšã
ããŠãŒã¶ã®ããã®æŒç¿ãå«ãŸããŠããŸãããããã®æŒç¿ã§ã¯ããããžã§ã¯
ãã®å®è¡æ¹æ³ãã¹ãããã®æ§ææ¹æ³ãä¿¡å·ã®æäœæ¹æ³ãã¹ã€ãŒã枬å®ã®å®
è¡æ¹æ³ãããŒã¿ã®ä¿åæ¹æ³ãããã³ LabVIEW ã®ã°ã©ãã£ã«ã«ããã°ã©ã
ã³ã°ã«ããé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã®æ¡åŒµæ¹æ³ã«ã€ããŠèª¬æããŸãã
é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã¯æ¢åã® NI SignalExpress æè¡ãæ¡åŒµãã
ãœãããŠã§ã¢ã§ãããã®ããã¥ã¢ã«ã§ã¯ SignalExpress ã«èšåããŠãã
ãªãã¡ã¬ã³ã¹ããã®äŸã䜿çšãããŠããŸãããã®ãããªãªãã¡ã¬ã³ã¹ãäŸ
ã¯é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãããã³ SignalExpress ã®äž¡æ¹ã§é©çšãã
ãŸãã
å¿ èŠãªãœãããŠã§ã¢
é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã§ã¯ã以äžã®ãœãããŠã§ã¢ã®ã€ã³ã¹ããŒã«ã
å¿ èŠã§ãã
⢠é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãããã·ã§ãã«ã€ã³ã¹ãã«ã¡ã³ãã®ããŒã¿
éé²ïŒDAQïŒããŒããŠã§ã¢ãšäœµçšããå Žåãå¿ ã NI-DAQmx ããŒ
ãžã§ã³ 8.9.5 以éãã€ã³ã¹ããŒã«ããŠãã ããã
⢠é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã®ãããžã§ã¯ãã LabVIEW ã®ãããã¯
ãã€ã¢ã°ã©ã ã«å€æããå Žåã¯ããŸã LabVIEW ããŒãžã§ã³ 8.2 以é
ãã€ã³ã¹ããŒã«ããŠããé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ããã€ã³ã¹ããŒã«
ããŠãã ããã
é¢é£ããã¥ã¡ã³ã
ãã®ããã¥ã¢ã«ã«ããæŒç¿ã®ä»ã«ãåç»ã䜿çšãããé³é¿ / æ¯åèšæž¬ã¢ã·
ã¹ã¿ã³ããã¥ãŒããªã¢ã«ãã«åŸã£ãŠé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã®æäœã
ç¿åŸã§ããŸãããã®ãã¥ãŒããªã¢ã«ãåç §ããã«ã¯ã以äžã®æé ã«åŸã£ãŠ
ãã ããã
1. ã¹ã¿ãŒãâãã¹ãŠã®ããã°ã©ã â National Instruments â Sound and Vibration â Sound and Vibration Assistant ãéžæããé³é¿ /æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ããèµ·åããŸãã
2. ãã«ãâãããããã€ã¢ãã°ã衚瀺ãéžæããŸãã
ãã®ããã¥ã¢ã«ã«ã€ããŠ
NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã viii ni.com
3. é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã®èŠåºãã衚瀺ããããŸã§äžã«ã¹ã¯ããŒ
ã«ããŸãã
4. ãã®èŠåºããå±éããŸãã
5. 5 åéã®å¯Ÿè©±åŒãã¥ãŒããªã¢ã«ãèŠãã®ãªã³ã¯ãã¯ãªãã¯ããŸãã
ãã®ããã¥ã¢ã«ãèªãã§ããéäžã§ãããé ç®ã«ã€ããŠã®è©³çŽ°æ å ±ãæ€çŽ¢
ããã«ã¯ããé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ããã«ãããåç §ããŠãã ããã
ãã«ãâé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ããã«ããéžæããã¹ããããªãã¡ã¬
ã³ã¹âé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã®ã¹ãããã«ç§»åããŠç®æ¬¡ã¿ãã§èª¿ã¹
ãããšãã§ããŸãã
© National Instruments 1-1 NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã
1NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã®æŠèŠ
NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã¯ããã°ã©ãã³ã°ãäžèŠãªå¯Ÿè©±åŒã®ä»®æ³
èšæž¬ç°å¢ã§ãé³é¿ãæ¯åã枬å®ãããšã³ãžãã¢ã«ãšã£ãŠæé©åãããŠããŸ
ããé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã§ã¯ãä¿¡å·ã®éé²ãçæã解æãæ¯èŒã
ã€ã³ããŒããä¿åãããã³é³é¿ãæ¯åä¿¡å·ã®åçã察話åŒã«è¡ãããšãã§
ããŸãã
ãŸããNI SignalExpress ã®æè¡ãæ¡åŒµããé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã
ã䜿çšããããšã«ãããåºæã®é³é¿ / æ¯åèšæž¬ã解æãè¡ããããã«ãªã
ãŸããé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã®çšéã«ã¯ãé³å£°ãã¹ããé³é¿æž¬å®ã
ç°å¢éšé³ãã¹ããæ©æ¢°ç¶æ ç£èŠãæ¯å解æãéšé³ / æ¯åå¶åŸ¡ãNVHïŒéš
é³ / æ¯å / ããŒã·ã¥ãã¹ïŒæž¬å®ãªã©ããããŸãã
ãŸããNI LabVIEW ã§äœæããã«ã¹ã¿ã ä»®æ³èšæž¬åšïŒVIïŒã®ã€ã³ããŒã
ããé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã®ãããžã§ã¯ãã LabVIEW ãããã¯ã
ã€ã¢ã°ã©ã ã«å€æããããšã«ãããLabVIEW ã§ã®éçºãå¯èœãšãªããé³
é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã®æ©èœãæ¡åŒµããããšãã§ããŸãã
© National Instruments 2-1 NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã
2ãããžã§ã¯ãã®æäœ
NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã䜿çšããŠã察話åŒèšæž¬ç°å¢ã§ã¹ããã
ãè¿œå ãæ§æããããšã«ãã枬å®ãè¡ããŸããã¹ããããšã¯ãä¿¡å·ã®éé²
ãçæã解æãããŒããããã³ä¿åãè¡ãæ§æå¯èœãªé¢æ°ã§ããã»ãšãã©
ã®ã¹ãããã¯ãå ¥åä¿¡å·ãåŠçããŠãåºåä¿¡å·ãçæããŸããæ§æããã
ã¹ãããã®ã·ãŒã±ã³ã¹ãä¿åãããã®ãé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã®ã
ããžã§ã¯ããšãªããŸãã
ãã®ç« ã§ã¯ãæ¢åã®ãããžã§ã¯ããããŒããŸãã¯å®è¡ããæ¹æ³ãšããã
ãžã§ã¯ãã®ã¹ããããæ§æããæ¹æ³ã«ã€ããŠèª¬æããŸãã
ãããžã§ã¯ããéã
é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã§ãµã³ãã«ãããžã§ã¯ããããŒãããã«ã¯ã
以äžã®æé ã«åŸã£ãŠãã ããã
1. ã¹ã¿ãŒãâãã¹ãŠã®ããã°ã©ã â National Instruments â Sound and Vibration â Sound and Vibration Assistant ãéžæããé³é¿ /æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ããèµ·åããŸãã
é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã¯ãããŸããŸãªçš®é¡ã®æ å ±ã衚瀺ããè€
æ°ã®ãã¥ãŒã«åããããŠããŸããã¢ããªã±ãŒã·ã§ã³ãŠã£ã³ããŠã®äžå€®
ã«è¡šç€ºãããã¿ãä»ãã®ãã¥ãŒãããã©ã€ããªãã¥ãŒãšåŒã³ãŸããé³
é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ããåããŠèµ·åãããšããã©ã€ããªãã¥ãŒã«
ããã©ã«ãã§ããŒã¿ãã¥ãŒã¿ããèšé²ãªãã·ã§ã³ã¿ãããããžã§ã¯ã
ããã¥ã¡ã³ãã¿ãã衚瀺ãããŸãã
ãã©ã€ããªãã¥ãŒã®åšãã«ã¯ãè£å©ãã¥ãŒã衚瀺ãããŸããããã©ã«
ãã®æ§æã§ã¯ãå·ŠåŽã«ãããžã§ã¯ããã¥ãŒãå³åŽã«è©³çŽ°ãã«ãã衚瀺
ãããŸãã
2. é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ããããã©ã«ãã®ã¬ã€ã¢ãŠãã§è¡šç€ºãããŠ
ããªãå Žåã¯ã衚瀺âã¬ã€ã¢ãŠãããªã»ãããéžæããŠããã©ã«ãã®
æ§æã«æ»ãããšãã§ããŸããã¿ãããã¥ãŒã¯ã衚瀺ã¡ãã¥ãŒãã衚瀺
ãšé衚瀺ãåãæ¿ããããããã©ã«ãã®ã¬ã€ã¢ãŠãã«ãªã»ããããã
ãšãã§ããŸãã
3. ãã«ãâãµã³ãã«ãéãâ Sound and Vibration Assistant ãéžæ
ããTutorialãã©ã«ããããã«ã¯ãªãã¯ããŠéããé³é¿ / æ¯åèšæž¬
ã¢ã·ã¹ã¿ã³ããããžã§ã¯ã SVA_Analyze Signal From UFF File
(Tutorial).seprojãããã«ã¯ãªãã¯ããŸãã
第 2 ç« ãããžã§ã¯ãã®æäœ
NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã 2-2 ni.com
4. å³ 2-1 ã®ãããªãŠã£ã³ããŠã衚瀺ãããããé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿
ã³ãã«æèŒãããŠããããŸããŸãªã³ã³ããŒãã³ãã確èªããŠãã ã
ãã
å³ 2-1 ãSVA_Analyze Signal from UFF File (Tutorial).seprojããããžã§ã¯ã
å·ŠåŽã®ããŒã³ããããžã§ã¯ããã¥ãŒã§ããããžã§ã¯ãã®æäœæé ãã¹
ãããã衚瀺ããŸãã
äžå€®ã®ããŒã³ã«ããŒã¿ãã¥ãŒã¿ãã衚瀺ãããŸããããã«ã¯ããã
ãžã§ã¯ããçæã解æããä¿¡å·ã衚瀺ããŸãã
å³åŽã®ããŒã³ã«è©³çŽ°ãã«ãã衚瀺ãããŸããæ°èŠã®ã¹ããããã¿ãã
ãã©ã¡ãŒã¿ããŸãã¯è¡šç€ºã«ã«ãŒãœã«ã移åãããšå 容ãæŽæ°ãããŸ
ãã詳现ãã«ãã®äžååã«ã¯ã¹ããããã¿ãã®æ å ±ããäžååã«ã¯ã
ã©ã¡ãŒã¿ã®æ å ±ã衚瀺ãããŸãã
1 å®è¡ãã¿ã³
2 äœæ¥ãšãªã¢ãã«ããŠã³ã¡ãã¥ãŒ
3 ãããžã§ã¯ãã¹ããŒã¿ã¹
4 ã¹ããã
5 å ¥å
6 åºå
7 ãããžã§ã¯ããã¥ãŒ
8 ãã°ããŒã¿ãŠã£ã³ããŠ
9 ãã©ã€ããªãã¥ãŒïŒããŒã¿ãã¥ãŒã¿ããéžæããç¶æ ïŒ
10 èšé²ãã¿ã³
11 衚瀺ã¿ã
12 詳现ãã«ã
9
7
8
6
12
111 10
4
5
2
3
第 2 ç« ãããžã§ã¯ãã®æäœ
© National Instruments 2-3 NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã
ãããžã§ã¯ãã®å®è¡ãšä¿¡å·ã®è¡šç€º
é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã«ã¯ã2 ã€ã®å®è¡ã¢ãŒãïŒ1 åå®è¡ãšé£ç¶å®
è¡ïŒããããŸããå·Šã«ç€ºãå®è¡ãã¿ã³ã®äžç¢å°ãã¯ãªãã¯ãã1 åå®è¡ãª
ãã·ã§ã³ãéžæãããšãé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã¯ãããžã§ã¯ãã®ã
ã¹ãŠã®ã¹ãããã 1 åå®è¡ããŸããå·Šã«ç€ºãå®è¡ãã¿ã³ã®äžç¢å°ãã¯
ãªãã¯ããé£ç¶å®è¡ãªãã·ã§ã³ãéžæãããšãé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³
ãã¯ãå·Šã«ç€ºãåæ¢ãã¿ã³ãã¯ãªãã¯ãããŸã§ããããžã§ã¯ãã®ãã¹ãŠã®
ã¹ããããé£ç¶çã«å®è¡ããŸãã
ãããžã§ã¯ãã®å®è¡äžã¯ãå®è¡ãã¿ã³ãåæ¢ãã¿ã³ã«å€åããŸãããã
ãžã§ã¯ãã®ã¹ãããå®è¡äžã¯ãããŒã¿ãã¥ãŒãç¶ç¶çã«æŽæ°ãããŸããã
ããžã§ã¯ãã®å®è¡äžã§ãã枬å®ã®æ§æãå€æŽããããå³æã«æž¬å®çµæãè¡š
瀺ããããšãã§ããŸãã
ãŸãããããžã§ã¯ãã®å®è¡ã¢ãŒããæ§æããããšãã§ããŸããå®è¡ãã¿ã³
ã®äžç¢å°ãã¯ãªãã¯ããå®è¡æ¹æ³ãæ§æãéžæãããšããããžã§ã¯ãã®å®
è¡æéãå埩åæ°ãç§æ°ã§èšå®ãããããããžã§ã¯ããé£ç¶å®è¡ãããã
ã«èšå®ã§ããŸãããã«ããŠã³ã¡ãã¥ãŒãããªãã·ã§ã³ãéžæããªãã§å®è¡
ãã¿ã³ãã¯ãªãã¯ãããšãåæ¢ãã¿ã³ãã¯ãªãã¯ãããŸã§ããããžã§ã¯ã
ãé£ç¶çã«å®è¡ããŸãã
ãµã³ãã«ãããžã§ã¯ããå®è¡ããŠä¿¡å·ã衚瀺ããã«ã¯ã以äžã®æé ã«åŸã£
ãŠãã ããã
1. å®è¡ãã¿ã³ãã¯ãªãã¯ãããããžã§ã¯ãã®ãã¹ãŠã®ã¹ããããé£ç¶ç
ã«å®è¡ããŸãã
ãããžã§ã¯ããå®è¡ãããšãåã¹ãããã¯å ¥åä¿¡å·ã解æãã解æçµ
æããæ°ããåºåä¿¡å·ãçæããŸãããã®ãããžã§ã¯ãã§ã¯ã
ãUFF58 ããããŒããã¹ãããã UFF58 ãã¡ã€ã«ããäžè§æ³¢ãããŒã
ããããã¯ãŒã¹ãã¯ãã«ãã¹ããããšãããŒã¯æ€åºãã¹ãããã¯ãã®
äžè§æ³¢ã解æããåŸãæ°èŠåºåãè¿ããŸãããããžã§ã¯ããã¥ãŒã§
ã¯ãå ¥åã¯èµ€ãç¢å°ãåºåã¯éãç¢å°ã§ç€ºãããŸãã
2. ãããžã§ã¯ããã¥ãŒã®ãããŒã¯æ€åºãã¹ãããã®ããŒã¯æ°åºåä¿¡å·ã
ããŒã¿ãã¥ãŒã¿ãã«ãã©ãã°ããŸãã
é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã®ããŒã¿ãã¥ãŒã§æ°èŠã°ã©ããäœæãã
ãŸããããŒã¯æ°ä¿¡å·ã¯æ¢åã®ã°ã©ããšããŒã¿ã¿ã€ããç°ãªããããæ¢
åã®è¡šãŸãã¯ã°ã©ãã«ã¯è¡šç€ºãããŸãããé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³
ãã§ã¯ç°ãªãä¿¡å·ã¿ã€ããèªåçã«èªèããé©åãªæ¹æ³ã§ãã®ä¿¡å·ã
衚瀺ããŸãããããžã§ã¯ãã¯å³ 2-2 ã®ããã«è¡šç€ºãããŸãã
第 2 ç« ãããžã§ã¯ãã®æäœ
NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã 2-4 ni.com
å³ 2-2 ãSVA_Analyze Signal from UFF File (Tutorial).seprojããããžã§ã¯ãã®åºå
3. ãã«ãâé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ããã«ããéžæããæ€çŽ¢ã¿ããã¯
ãªãã¯ããŸãããä¿¡å·ã¿ã€ãããšå ¥åããSignalExpress ã®ä¿¡å·ã¿ã€
ããããã«ã¯ãªãã¯ããŠãä¿¡å·ã¿ã€ãã®è©³çŽ°ã調ã¹ãŸãã
ãã®ãã«ãã«ã¯ãSignalExpress ã®ãããžã§ã¯ããã¹ããããä¿¡å·ãª
ã©ã®æ©èœã®è©³çŽ°ãªèª¬æãèšèŒãããŠããŸãã
ã¹ããããæ§æãã
ã¹ããããšã¯ãä¿¡å·ã®éé²ãçæã解æãããŒããããã³ä¿åãè¡ãæ§æ
å¯èœãªé¢æ°ã§ããã»ãšãã©ã®ã¹ãããã¯ãå ¥åä¿¡å·ãåŠçããŠãåºåä¿¡å·
ãçæããŸããã¹ãããèšå®ã¿ãã§å€ãæå®ãããšãã¹ãããã®æäœãæ§
æã§ããŸãããããžã§ã¯ãã®å®è¡äžã§ããã¹ãããã®æ§æãå€æŽããŠãã®
çµæãããŒã¿ãã¥ãŒã¿ãã«è¡šç€ºããå¿ èŠãªæž¬å®çµæãåŸããããŸã§æž¬å®ã
調æŽããããšãã§ããŸãã
第 2 ç« ãããžã§ã¯ãã®æäœ
© National Instruments 2-5 NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã
ããã¯ãŒã¹ãã¯ãã«ãã¹ãããããã³ãããŒã¯æ€åºãã¹ããããæ§æãã
ã«ã¯ã以äžã®æé ã«åŸã£ãŠãã ããã
1. ãã¡ã€ã«âãããžã§ã¯ããå¥åã§ä¿åãéžæãããããžã§ã¯ãã My First Project.seprojãšããååã§ä»»æã®ãã£ã¬ã¯ããªã«ä¿åã
ãŸãã
2. ãããžã§ã¯ããã¥ãŒã®ããã¯ãŒã¹ãã¯ãã«ãã¹ããããããã«ã¯ãªã
ã¯ããŸããå³ 2-3 ã®ãããªããã¯ãŒã¹ãã¯ãã«ãã¹ãããã®ã¹ããã
èšå®ã¿ãã衚瀺ãããŸããã¹ãããèšå®ã¿ãã®ã°ã©ãã«å ¥åä¿¡å·ã®ã
ã¬ãã¥ãŒã衚瀺ãããŸãã
å³ 2-3 ããã¯ãŒã¹ãã¯ãã«ãã®ã¹ãããèšå®ã¿ã
3. ã¹ãããèšå®ã¿ãã®å ¥åã¿ããã¯ãªãã¯ããŸããå ¥åä¿¡å·ãã«ããŠã³
ã¡ãã¥ãŒãæé波圢ãæå®ããŠããããšã«æ³šæããŠãã ããã
ãUFF58 ããããŒããã¹ããããããã¯ãŒã¹ãã¯ãã«ãã¹ãããã®æ
é波圢信å·ãçæããŸãã
4. ãããžã§ã¯ããã¥ãŒã®ãããŒã¯æ€åºãã¹ããããã¯ãªãã¯ããŸãã
ã¹ãããèšå®ã¿ãã¯ããããŒã¯æ€åºãã¹ãããã®æ§æ衚瀺ããããã
ã¯ãŒã¹ãã¯ãã«ãã¹ãããã®æ§æ衚瀺ã«åãæ¿ãããŸãã
5. ã¹ãããèšå®ã¿ãã®å ¥åã¿ããã¯ãªãã¯ããŸããå ¥åä¿¡å·ãã«ããŠã³
ã¡ãã¥ãŒãã¹ãã¯ãã«ãæå®ããŠããããšã«æ³šæããŠãã ããããã
ã¯ãŒã¹ãã¯ãã«ãã¹ãããããããŒã¯æ€åºãã¹ãããã®ã¹ãã¯ãã«ä¿¡
å·ãçæããŸãã
第 2 ç« ãããžã§ã¯ãã®æäœ
NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã 2-6 ni.com
6. ã¹ãããèšå®ã¿ãã®å ¥åã¿ããã¯ãªãã¯ããŸããæ§æã¿ãã§ã¯ã
ãããŒã¯æ€åºãã¹ããããã¹ãã¯ãã«ã«å¯ŸããŠããŒã¯æ€åºãè¡ããã
ããå€ -40 ãè¶ ããæ倧ããŒã¯ã 1 ã€æ€åºããããã«èšå®ãããŠããŸ
ãã ãã®ã¹ãããã§ã¯åºåãšã㊠2 ã€ã®æž¬å®çµæïŒããŒã¯ã®åšæ³¢æ°ãš
æ¯å¹ ãããã³ããŒã¯æ°ïŒãçæãããŸãã
7. ã¹ãããèšå®ã¿ãã®æ§æã¿ãã§ãæ€åºã¿ã€ããã«ããŠã³ã¡ãã¥ãŒãã
è€æ°ããŒã¯ãéžæãããšãå ¥åã¹ãã¯ãã«ã®ãããå€ãè¶ ãããã¹ãŠ
ã®ããŒã¯ã枬å®ããŸããã¹ãããèšå®ã¿ãã®ããŒã¯è¡šã«ããä¿¡å·ãæŽ
æ°ãããè¡ã£ãå€æŽãåæ ãããŸãã
8. åæ¢ãã¿ã³ãã¯ãªãã¯ããŠãããžã§ã¯ããåæ¢ããŸãã
åæ¢ãã¿ã³ãã¯ãªãã¯ãããšãçŸåšã®å埩ãå®äºããåŸã§ãããžã§ã¯
ããåæ¢ããŸããåæ¢ãã¿ã³ã®äžç¢å°ãã¯ãªãã¯ããäžæãªãã·ã§ã³
ãéžæãããšããããžã§ã¯ãã¯ç·æ¥åæ¢ããæ®ãã®ã¹ãããã¯å®è¡ã
ããŸããã
9. ãã¡ã€ã«âãããžã§ã¯ããä¿åãéžæããŠãããžã§ã¯ããä¿åããŸ
ãã
10. ãã¡ã€ã«âãããžã§ã¯ããéãããéžæãããããžã§ã¯ããéããŸ
ãã
ã¹ãããã®ç§»åããã³ãšã©ãŒåŠç
é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã®ãããžã§ã¯ãã®ã¹ãããã¯ãå ¥åããŒã¿ã«
ãã£ãŠç°ãªããŸããã€ãŸããåã¹ãããã§æäœã§ããã®ã¯ããããžã§ã¯ã
ãã¥ãŒã«ãã以åã®ã¹ããããããšã¯ã¹ããŒããããä¿¡å·ã®ã¿ãšãªããŸ
ããã¹ãããã®ã¹ãããèšå®ã¿ãã®å ¥åã¿ãã§ãå ¥åä¿¡å·ãã«ããŠã³ã¡
ãã¥ãŒã§ã¯ã以åã®ã¹ããããããšã¯ã¹ããŒããããäºææ§ã®ããä¿¡å·ã®
ã¿ã衚瀺ããŸããã¹ãããã®åºåãå¥ã®ã¹ãããã®å ¥åã«ãªããšããã® 2ã€ã®ã¹ãããã¯çžäºäŸåãšãªããé çªã«åãã¬ãŒãã§å®è¡ãããŸããæå
ã®ã¹ããããåºåä¿¡å·ãçæãããã®åºåä¿¡å·ã 2 ã€ç®ã®ã¹ãããã®å ¥
åãšããŠåãåãããå®è¡ãéå§ãããŸãã
ãããžã§ã¯ãå ã®ã¹ãããã¯ããããžã§ã¯ããã¥ãŒã§äžäžã«ãã©ãã°ããŠ
移åã§ããŸãããŸãããããžã§ã¯ããã¥ãŒã§ã¹ããããå³ã¯ãªãã¯ãã
ã·ã§ãŒãã«ããã¡ãã¥ãŒããåé€ãéžæãããšãã¹ããããåé€ã§ããŸ
ãããã ããã¹ãããã®ç§»åãåé€ãè¡ã£ãå Žåã«ã¯ããããžã§ã¯ãã®ä¿¡
å·ã¹ããŒã¿ã¹ãå€æŽãããŸããããšãã°ãåºåä¿¡å·ãçæããã¹ãããã
åé€ãããšãåé€ãããã¹ãããããã®åºåä¿¡å·ãä»ã®ã¹ãããã®å ¥åä¿¡
å·ã«ãªã£ãŠããå Žåã¯ãããžã§ã¯ãã®æäœãäžæãããŸããæäœãäžæã
ãããšãšã©ãŒã€ã³ãžã±ãŒã¿ã衚瀺ãããŸãã
第 2 ç« ãããžã§ã¯ãã®æäœ
© National Instruments 2-7 NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã
ãããžã§ã¯ãå ã®ã¹ãããã®åãåã / ã³ã㌠/ 貌ãä»ããããã«ã¯ãã
ããã <Ctrl-X>ã<Ctrl-C>ã<Ctrl-V> ããŒãæŒããããããžã§ã¯ããã¥ãŒ
ã®ã¹ããããå³ã¯ãªãã¯ããŠè©³çŽ°ã¡ãã¥ãŒããåãåããã³ããŒãéžæã
ãã¹ãããã®åã«è²Œãä»ãããŸãã¯éžæããã¹ãããã®åŸã«è²Œãä»ããéž
æããŸãã
ã¹ãããã移åããŠãšã©ãŒãåŠçããã«ã¯ã以äžã®æé ã«åŸã£ãŠãã ã
ãã
1. ãã¡ã€ã«âæ°èŠãããžã§ã¯ããéžæããŠããããžã§ã¯ããäœæããŸ
ãã
2. My Error.seprojãšããååã§ä»»æã®ãã£ã¬ã¯ãã«ä¿åããŸãã
3. å·Šã«ç€ºãã¹ããããè¿œå ãã¿ã³ã¯ãªãã¯ããåŠçâæšæºãã£ã«ã¿ãéž
æããŠãæšæºãã£ã«ã¿ãã¹ããããè¿œå ããŸãã
ãã³ã ãããžã§ã¯ããžã®ã¹ãããã®è¿œå ã¯ãã¹ããããè¿œå ã¡ãã¥ãŒãã¹ããããè¿œ
å ãã¬ããããŸãã¯ãããžã§ã¯ããã¥ãŒãå³ã¯ãªãã¯ããŠè¡šç€ºããã詳现ã¡
ãã¥ãŒããè¡ãããšãã§ããŸãã
4. å³ 2-4 ã«ç€ºãããã«ãæªæ§æã®å ¥åãã€ã¢ãã°ããã¯ã¹ã衚瀺ããã
ããã£ã³ã»ã«ãã¿ã³ãã¯ãªãã¯ããŸãã
å³ 2-4 æªæ§æã®å ¥åãã€ã¢ãã°ããã¯ã¹
å ¥åä¿¡å·ãååšããªãå Žåã¯ãå¿ èŠãªå ¥åä¿¡å·ãçæããã¹ãããã
éžæããããã®æªæ§æã®å ¥åãã€ã¢ãã°ããã¯ã¹ãéããŸãã
5. å·Šã«ç€ºããšã©ãŒã€ã³ãžã±ãŒã¿ãã¹ãããèšå®ã¿ãã®äžéšãšãããžã§ã¯
ããã¥ãŒã®ã¹ãããå ã«è¡šç€ºãããããšã«æ³šæããŠãã ããã
第 2 ç« ãããžã§ã¯ãã®æäœ
NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã 2-8 ni.com
ã¡ã¢ ãããžã§ã¯ãã®å®è¡äžã«çºçãããã¹ãŠã®ãšã©ãŒããã³èŠåã¯ãã€ãã³ããã°
ã¿ãã«èšé²ãããŸããã€ãã³ããã°ã衚瀺ããã«ã¯ãã€ãã³ããã°ã衚瀺ãã
ã«ã¯ã衚瀺âã€ãã³ããã°ãéžæããŸãããšã©ãŒããã³èŠåã®è©³çŽ°ã«ã€ããŠã¯ã
ãã«ãâé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ããã«ããéžæããŠè¡šç€ºã§ãããé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ããã«ãããåç §ããŠãã ãããæ€çŽ¢ã¿ããã¯ãªãã¯ãããš
ã©ãŒãšå ¥åããŸãã
6. ã¹ãããèšå®ã¿ãã®è©³çŽ°ãã¿ã³ãã¯ãªãã¯ãããšã詳现ãã€ã¢ãã°
ããã¯ã¹ã«ãšã©ãŒã®èª¬æã衚瀺ãããŸãããŸãããããžã§ã¯ããã¥ãŒ
ã®ãšã©ãŒã€ã³ãžã±ãŒã¿ãã¯ãªãã¯ããŠãã説æã衚瀺ããããšãã§ã
ãŸãããã®ãšã©ãŒãåŠçããã«ã¯ãã¹ããããè¿œå ããŠæ°èŠå ¥åãäœ
æããŸãã
7. 詳现ãã€ã¢ãã°ããã¯ã¹ãéããŸãã
8. ã¹ããããè¿œå ãã¿ã³ãã¯ãªãã¯ããä¿¡å·ãäœæâã¢ããã°ä¿¡å·ãäœ
æãéžæããŸãããã¢ããã°ä¿¡å·ãäœæãã¹ãããã¯ãæšæºãã£ã«ã¿ã
ã¹ããããžã®çŽæ¥å ¥åä¿¡å·ãæž¡ããªããæªæ§æã®å ¥åãšã©ãŒãæ®ããŸ
ãã
9. ãããžã§ã¯ããã¥ãŒã§ããã¢ããã°ä¿¡å·ãäœæãã¹ãããããæšæº
ãã£ã«ã¿ãã¹ãããã®äžã«ãã©ãã°ããŸãã移åããããšã«ããã
ãæšæºãã£ã«ã¿ãã¹ãããããã¢ããã°ä¿¡å·ãäœæãã¹ãããã®æ£åŒŠ
æ³¢åºåãå ¥åä¿¡å·ãšããŠäœ¿çšããããããšã©ãŒã€ã³ãžã±ãŒã¿ã¯è¡šç€ºã
ããªããªããŸãã
10. ãããžã§ã¯ããä¿åããŠéããŸãã
© National Instruments 3-1 NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã
3ããŒã¿ããã°ãã
é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã䜿çšãããšãä»»æã®æéé åãæ°å€ããŸã
ã¯ããŒã«ã®ã¹ãããåºåãå«ã枬å®ãèšé²ã§ããŸãã
ãã®ç« ã§ã¯ãé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã§çµ±åããŒã¿ãã®ã³ã°æ©èœã䜿
çšããããŒã¿ã®èšé²æ¹æ³ã«ã€ããŠèª¬æããŸããããšãã°ãæå®ããä¿¡å·ã®
èšé²æ¹æ³ãä¿¡å·ã®åçæ¹æ³ã解æã®ã¹ãããã䜿çšããä¿¡å·ã®è§£ææ¹æ³ã«
ã€ããŠèª¬æããŸãããŸããèšé²ãªãã·ã§ã³ã¿ãã§æå®ããéå§æ¡ä»¶ãŸãã¯
åæ¢æ¡ä»¶ã«åºã¥ããŠä¿¡å·ããã°ããæ¹æ³ã«ã€ããŠã説æããŸãã
ã¡ã¢ SignalExpress ã®ãªãã¡ã¬ã³ã¹ãäŸã¯é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã«ãåæ§ã«é©
çšãããŸãã
ä¿¡å·ãèšé²ãã
ä¿¡å·ãèšé²ããã«ã¯ã以äžã®æé ã«åŸã£ãŠãã ããã
1. ãã«ãâãµã³ãã«ãéãâ SignalExpress ãéžæããTutorial
ãã©ã«ããããã«ã¯ãªãã¯ããŠéããLogging.seprojãããã«ã¯
ãªãã¯ããŸãã
ãã®ãããžã§ã¯ãã§ã¯ããã¢ããã°ä¿¡å·ãäœæãã¹ãããã䜿çšãã
åŒã«åºã¥ããŠä¿¡å·ãçæããŸãã
2. ãããžã§ã¯ãã My Logged Signal.seprojãšããååã§ä»»æã®
ãã£ã¬ã¯ããªã«ä¿åããŸãã
3. å·Šã«ç€ºãèšé²ãã¿ã³ãã¯ãªãã¯ãããã°ã«èšé²ããä¿¡å·ãéžæãã€ã¢
ãã°ããã¯ã¹ãéããŸãã
ãã°ã«èšé²ããä¿¡å·ãéžæãã€ã¢ãã°ããã¯ã¹ã¯ããããžã§ã¯ãã®èš
é²å¯èœãªä¿¡å·ã衚瀺ããŸãã1 ã€ãŸãã¯è€æ°ã®ä¿¡å·ãèšé²ããããéž
æã§ããŸãããã°ã®ååãšèª¬æãæå®ããããšãã§ããŸãã
4. signal ãã§ãã¯ããã¯ã¹ããªã³ã«ããŠãã¢ããã°ä¿¡å·ãäœæã¹ãã
ãã§çæãããä¿¡å·ãèšé²ããŸãã
5. OK ãã¿ã³ãã¯ãªãã¯ãããã°ã«èšé²ããä¿¡å·ãéžæãã€ã¢ãã°ãã
ã¯ã¹ãéããŠãä¿¡å·ã®èšé²ãéå§ããŸããåæ¢ãã¿ã³ãã¯ãªãã¯ãã
ãŸã§ããã°æäœãç¶ç¶ãããŸãã
6. åæ¢ãã¿ã³ãã¯ãªãã¯ããä¿¡å·ã®ãã®ã³ã°ãåæ¢ããŸããåããŠä¿¡å·
ããã°ããå Žåã¯ãæåã®ãã°äœæãå®äºãã€ã¢ãã°ããã¯ã¹ã衚瀺
ãããŸããOK ãã¿ã³ãã¯ãªãã¯ãããã€ã¢ãã°ããã¯ã¹ãéããŸãã
第 3 ç« ããŒã¿ããã°ãã
NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã 3-2 ni.com
å³ 3-1 ã«ç€ºãããã«ããã°ããŒã¿ãŠã£ã³ããŠããããžã§ã¯ããã¥ãŒã®
äžã«è¡šç€ºãããŸãããã®ãŠã£ã³ããŠã«ã¯ãçŸåšã®ãããžã§ã¯ãã§ãã
ãŸã§ã«èšé²ããããã¹ãŠã®ããŒã¿ã®ãªã¹ãã衚瀺ãããŸãã
å³ 3-1 ãã°ããŒã¿ãŠã£ã³ããŠ
ããã©ã«ãã§ã¯ãããŒã¿ãèšé²ããæ¥ä»ãšæéã«å¿ããŠèªåçã«ãã°
ããŒã¿ã«ååãä»ããããŸãããã°ããŒã¿ã¯ã.tdmsãã¡ã€ã«åœ¢åŒ
ã§ä¿åãããŸãã
ã¡ã¢ .tdmsãã¡ã€ã«åœ¢åŒã®è©³çŽ°ã«ã€ããŠã¯ãni.com/jp/infoã«ã¢ã¯ã»ã¹ãã
ãtdmsããšå ¥åããŠãã ããã
7. ãã°ããŒã¿ãä¿åãããã£ã¬ã¯ããªãå€æŽãããããã°ããŒã¿ã®èšå®
ãã«ã¹ã¿ãã€ãºããã«ã¯ãããŒã«âãªãã·ã§ã³ãéžæãããã°ãªã
ã·ã§ã³ãéžæããŸãã
8. OK ãã¿ã³ãã¯ãªãã¯ããŠãªãã·ã§ã³ãã€ã¢ãã°ããã¯ã¹ãéããŸ
ãã
9. ãããžã§ã¯ããä¿åããŸãã
ãã°ãããä¿¡å·ã衚瀺ãã
ãã°ããŒã¿ã衚瀺ããã«ã¯ã以äžã®æé ã«åŸã£ãŠãã ããã
1. ãã°ããŒã¿ãŠã£ã³ããŠã®ãã°ãããŒã¿ãã¥ãŒã¿ãã«ãã©ãã°ããŸãã
å³ 3-2ã®ããã«ããŒã¿ãã¥ãŒã¿ãã«ãã°ããŒã¿ãšã°ã©ããã¬ãã¥ãŒã
衚瀺ãããŸãã
第 3 ç« ããŒã¿ããã°ãã
© National Instruments 3-3 NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã
å³ 3-2 Logging.seproj
ã°ã©ããã¬ãã¥ãŒã§ã¯ãããŒã¿ãã¥ãŒã¿ãã®ããŒã¿ã®ãºãŒã ã€ã³ / ã¢ãŠãããã³ããã³ã°ãè¡ããŸããããã©ã«ãã§ã¯ãã°ã©ããã¬ãã¥ãŒ
ã¯ãã°ããŒã¿ã衚瀺ãããšè¡šç€ºãããŸããã©ã€ãããŒã¿ãŸãã¯ãã°ã
ããŠããªãããŒã¿ã確èªããå Žåã«ã°ã©ããã¬ãã¥ãŒã衚瀺ããã«
ã¯ãããŒã¿ãã¥ãŒã¿ãã®ã°ã©ããå³ã¯ãªãã¯ããŠè¡šç€ºé ç®âãã¬
ãã¥ãŒãéžæããŸãã
2. ã°ã©ããã¬ãã¥ãŒã®é£ã«ãããºãŒã ã€ã³ãã¿ã³ãã¯ãªãã¯ãããã°ã
ããä¿¡å·ã§ãºãŒã ã€ã³æäœãè¡ããŸããã°ã©ããã¬ãã¥ãŒã®ã«ãŒãœã«
ã¯ãããŒã¿ãã¥ãŒã¿ãã«çŸåšè¡šç€ºãããŠããããŒã¿ã®ãµãã»ããã瀺
ããŸãã衚瀺ããããŒã¿ã®ãµãã»ãããå¢æžããã«ã¯ãã«ãŒãœã«ãã¯
ãªãã¯ããã©ãã°ããŸãã
第 3 ç« ããŒã¿ããã°ãã
NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã 3-4 ni.com
3. ã°ã©ããã¬ãã¥ãŒã®äžã«ããã¹ã¯ããŒã«ããŒã§ãããŒã¿ãã¹ã¯ããŒã«
ããŸãã
4. ã°ã©ãäžã§ä¿¡å·å šäœã衚瀺ãããã«ã¯ãã°ã©ããã¬ãã¥ãŒãå³ã¯ãªã
ã¯ããŠãã¹ãŠè¡šç€ºãéžæããŸãã
äºåå®çŸ©ããéå§ / åæ¢æ¡ä»¶ã§ä¿¡å·ããã°ãã
é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã§ä¿¡å·ã®èšé²ãèšé²åæ¢ãè¡ãåã«ãéå§ã
ãã³åæ¢æ¡ä»¶ã«äžèŽããããã«ä¿¡å·ãæ§æã§ããŸããéå§ããã³åæ¢æ¡ä»¶
ã«åºã¥ããŠããŒã¿ããã°ããã«ã¯ã以äžã®æé ã«åŸã£ãŠãã ããã
1. èšé²ãªãã·ã§ã³ã¿ããã¯ãªãã¯ãããã衚瀺âèšé²ãªãã·ã§ã³ãéžæ
ããŸãã
2. èšé²ãªãã·ã§ã³ã¿ãã®ã«ããŽãªãªã¹ãã§ä¿¡å·éžæãéžæããŸãã
3. å³ 3-3 ã®ããã«ãä¿¡å·ã®èšé²åã«ãã§ãã¯ããŒã¯ãä»ããŸãã
å³ 3-3 ä¿¡å·éžæ
èšé²ãã¿ã³ã¯ãå·Šã«ç€ºãå®è¡äžã«èšé²ãã¿ã³ã«å€ãããŸãããã®ãã¿
ã³ãæå¹ã«ãªã£ãŠããéã¯ãå®è¡ãã¿ã³ãã¯ãªãã¯ããããšãã«éžæ
ããä¿¡å·ãèšé²ãããŸãã
4. èšé²ãªãã·ã§ã³ã¿ãã®ã«ããŽãªãªã¹ãã§éå§æ¡ä»¶ãéžæããŸãã
第 3 ç« ããŒã¿ããã°ãã
© National Instruments 3-5 NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã
5. ãã°ã¿ã¹ã¯ã®éå§æ¡ä»¶ãå€æŽããã«ã¯ãéå§æ¡ä»¶ãªã¹ãããŒãžã§è¿œå
ãã¿ã³ãã¯ãªãã¯ããŸãã
a. æ¡ä»¶ã¿ã€ããã«ããŠã³ã¡ãã¥ãŒã§ãä¿¡å·ããªã¬ãéžæãããŠãã
ããšã確èªããŸãããã®èšå®ã§ã¯ãå ¥åä¿¡å·ãæå®ãããæ¡ä»¶ã
æºãããšèšé²ãéå§ãããŸãã
b. ä¿¡å·ãã«ããŠã³ã¡ãã¥ãŒã§ãsignal ãéžæãããŠããããšã確
èªããŸãã
c. ããªã¬ã¿ã€ããã«ããŠã³ã¡ãã¥ãŒã§ç«ã¡äžããã¹ããŒããéžæã
ããŠããããšã確èªããŸãããã®èšå®ã§ã¯ãæ£ã®ã¹ããŒãã§ã®ä¿¡
å·ã®ãšããžã®å€ã«åºã¥ããŠä¿¡å·ã®èšé²ãéå§ãããŸãã
d. ããªã¬å€ã« 1ãšå ¥åããŠãç«ã¡äžããã¹ããŒãã§ä¿¡å·ãã¬ãã« 1ãè¶ ãããšãã«èšé²ãéå§ãããããæå®ããŸãã
å³ 3-4 ã®ãããªèšé²ãªãã·ã§ã³ã¿ãã衚瀺ãããŸãã
å³ 3-4 èšé²ãªãã·ã§ã³ã®éå§æ¡ä»¶
第 3 ç« ããŒã¿ããã°ãã
NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã 3-6 ni.com
6. èšé²ãªãã·ã§ã³ã¿ãã®ã«ããŽãªãªã¹ãã§åæ¢æ¡ä»¶ãéžæããŸãã
7. ãã°ã¿ã¹ã¯ã®åæ¢æ¡ä»¶ãå€æŽããã«ã¯ãåæ¢æ¡ä»¶ãªã¹ãããŒãžã§è¿œå
ãã¿ã³ãã¯ãªãã¯ããŸãã
a. æ¡ä»¶ã¿ã€ããã«ããŠã³ã¡ãã¥ãŒã§ãæç¶æéãéžæãããŠããã
ãšã確èªããŸãããã®èšå®ã§ã¯ãæå®ãããæéãçµéãããšèš
é²ãåæ¢ãããŸãã
b. æç¶æéãã£ãŒã«ãã« 5ãšå ¥åããŠãä¿¡å·ãéå§æ¡ä»¶ã«äžèŽã
ãŠãã 5 ç§éèšé²ããããæå®ããŸãã
8. å®è¡ãã¿ã³ãã¯ãªãã¯ããŸããç«ã¡äžããã¹ããŒãã§ä¿¡å·ã®å€ã 1ãè¶ ãããšä¿¡å·ã®èšé²ãéå§ãã5 ç§éèšé²ãããŸãã
ãããžã§ã¯ãã®å®è¡äžã¯ãèšé²ãªãã·ã§ã³ã¿ãã®çŸåšã®ãã°ã®ç¶æ ã»
ã¯ã·ã§ã³ã®ä»¥äžã®è¡šç€ºãæŽæ°ãããŸãã
⢠信å·ãéå§æ¡ä»¶ãæºãããŠãããã°ãå®è¡ãããŠããéãèšé²äž
ã ON ã«ãªããŸãã
⢠ãã£ã¹ã¯æ å ±ã¯ããã°ã«äœ¿çšã§ããã³ã³ãã¥ãŒã¿ã®æå¹ãªããŒã
ãã£ã¹ã¯ã¹ããŒã¹ã衚瀺ããŸãã
⢠çŸåšã®æšå®ãã°ãµã€ãºã¯ããã£ã¹ã¯äžã®ãã°ãã¡ã€ã«ã®ãµã€ãºã
瀺ããŸãã
⢠æ®ãèšé²æéã¯ã䜿çšå¯èœãªãã£ã¹ã¯ã¹ããŒã¹ã«å¯ŸããŠãã°ãç¶
ç¶ã§ããæéã瀺ããŸãã
⢠çŸåšã®ãã°éå§æéã¯çŸåšã®ãã°ãéå§ãããæéã瀺ããŸãã
èšé²ãªãã·ã§ã³ã¿ãã®éå§æ¡ä»¶ãåæ¢æ¡ä»¶ãã¢ã©ãŒã ãã€ãã³ãããŒ
ãžã«ã¯ãèšå®ãããéå§ / åæ¢æ¡ä»¶ãã¢ã©ãŒã ãã€ãã³ãã®ç¶æ ã瀺
ãã€ã³ãžã±ãŒã¿ã衚瀺ãããŸãã
9. ãããžã§ã¯ããä¿åããŠéããŸãã
© National Instruments 4-1 NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã
4é³é¿ãæ¯åã®æž¬å®ãè¡ã
é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã䜿çšããŠãé³é¿ãæ¯åã®æž¬å®ã察話åŒã«è¡
ãããšãã§ããŸãã解æã¹ãããã§ãã°ããŒã¿ãåçããŠã解æãåŠçã
ãããšãã§ããŸãããŸããNI ãã€ãããã¯ä¿¡å·éé²ïŒDSAïŒããã€ã¹ãª
ã©ã®ãã·ã§ãã«ã€ã³ã¹ãã«ã¡ã³ãã®ããŒã¿éé²ïŒDAQïŒããŒããŠã§ã¢ã
䜿çšããŠéé²ããã©ã€ãããŒã¿ã®è§£æãåŠçãå¯èœã§ãã
ãã®ç« ã§ã¯ã以åã«éé²æžã¿ã®ããŒã¿ã®è§£æã枬å®çµæã®ã°ã©ã衚瀺ãä¿¡
å·ã®ãã¡ã€ã«ãžã®ä¿åãå«ããé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã§ã®é³é¿ããŒ
ã¿ããã³æ¯åããŒã¿ã®åŠçæ¹æ³ã«ã€ããŠèª¬æããŸãã
以åã«éé²æžã¿ã®ããŒã¿ãéã
é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã䜿çšããŠã以åã«éé²æžã¿ã®ããŒã¿ã解æ
ããããšãã§ããŸãããŸããDAQ ããŒããŠã§ã¢ã䜿çšããŠéé²ããããŒ
ã¿ã解æããããšãã§ããŸãã
以åã«èšé²ãããé³å§ããŒã¿ãéãã«ã¯ã以äžã®æé ã«åŸã£ãŠãã ããã
1. æ°èŠãããžã§ã¯ããéããŸãã
2. ãã¡ã€ã«âã€ã³ããŒãâ LabVIEW SignalExpress TDMS ãã¡ã€ã«ã®
ãã°ä¿¡å·ãéžæããNational InstrumentsÂ¥Sound and Vibration AssistantÂ¥ExamplesÂ¥Log File - Sound PressureÂ¥
ãã£ã¬ã¯ããªã«ç§»åããŠãçŸåšã®ãã©ã«ããã¯ãªãã¯ããŸãã
ïŒWindows XP/2000ïŒNational Instrumentsãã©ã«ãã¯
C:Â¥Documents and SettingsÂ¥All UsersÂ¥å ±æããã¥ã¡ã³ã Â¥
ãã£ã¬ã¯ããªã«ãããŸãã
ïŒWindows VistaïŒNational Instrumentsãã©ã«ã㯠C:Â¥UsersÂ¥
PublicÂ¥Public DocumentsÂ¥ãã£ã¬ã¯ããªã«ãããŸãã
3. ãã°ããŒã¿ããããã°ããŒã¿ããŠã£ã³ããŠã®ãã°ã«è¡šç€ºãããŠããã
ãšã«æ³šç®ããŠãã ããã
4. å³ 4-1 ã«ãããšããããããžã§ã¯ããã¥ãŒäžæ¹ã®äœæ¥ãšãªã¢ãã«ããŠ
ã³ã¡ãã¥ãŒããåçã¢ãŒãâ Sound Pressure ãéžæããŸãã
å³ 4-1 äœæ¥ãšãªã¢ãã«ããŠã³ã¡ãã¥ãŒ
第 4 ç« é³é¿ãæ¯åã®æž¬å®ãè¡ã
NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã 4-2 ni.com
åããããžã§ã¯ãã§è€æ°ã®é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã®æäœãå®è¡
ããå Žåã«äœæ¥ãšãªã¢ã䜿çšããŸããæ°èŠãããžã§ã¯ããéããã«ã
ããŒã¿éé²ãä¿¡å·åŠçãããŒã¿ã®èšé²ããã°ããŒã¿ã§ã®æž¬å®å®è¡ãå¯
èœã§ãããããžã§ã¯ããä¿åãããšããããžã§ã¯ãã®ãã¹ãŠã®äœæ¥ãš
ãªã¢ããããžã§ã¯ããã¡ã€ã«ã«ä¿åãããŸãã
ããã©ã«ãã®ç£èŠ / èšé²äœæ¥ãšãªã¢ã§ã¯ã枬å®ã®å®è¡ãã©ã€ãããŒã¿
ã®è§£æãããŒã¿ã®ãã°ãè¡ãããšãã§ããŸããåçäœæ¥ãšãªã¢ã§ã¯ã
ç£èŠ / èšé²äœæ¥ãšãªã¢ã§ãã°ããããŒã¿ãå ¥åãšããŠè§£æã¹ãããã
å®è¡ããããšãã§ããŸãã
ããã§ããã¡ã€ã« Sound Pressure.tdmsã«å«ãŸããé³å§ããŒã¿ã®è§£æ
ãã§ããããã«ãªããŸãã
ã¡ã¢ é³å§ã®éé²ããŒã¿ããããžã§ã¯ãã§å®è¡ããã«ã¯ãNI-DAQmx 8.9.5 以éãã€ã³
ã¹ããŒã«ããDAQ ããŒããŠã§ã¢ã䜿çšããå¿ èŠããããŸãã以éã®æŒç¿ã§é³å§
ã®çããŒã¿ã䜿çšãããå Žåã¯ãæ°èŠãããžã§ã¯ããäœæãã¹ããããè¿œå ã
ã¿ã³ãã¯ãªãã¯ããŠãä¿¡å·ãéé²â DAQmx éé²âã¢ããã°å ¥åâé³å§ãéžæ
ããŸããããã§è¡šç€ºãããã¿ã¹ã¯ã«ãã£ã³ãã«ãè¿œå ãã€ã¢ãã°ããã¯ã¹ã§ã
解æãããã¢ããã°å ¥åãã£ã³ãã«ãéžæã§ããŸãã
ããŒã¿ãåçãã
ãã¡ã€ã« Sound Pressure.tdmsã«ããããŒã¿ã解æããã«ã¯ã以äžã®
æé ã«åŸã£ãŠãã ããã
1. ãããžã§ã¯ãã My Analysis.seprojãšããååã§ä»»æã®ãã£ã¬ã¯
ããªã«ä¿åããŸãã
2. ã¹ããããè¿œå ãã¿ã³ãã¯ãªãã¯ãã解æâæéé å枬å®âé³å§ã¬ã
ã«ãéžæããŸãã
3. ã¹ãããèšå®ãã€ã¢ãã°ããã¯ã¹ã®å ¥åã¿ããã¯ãªãã¯ããŸããå ¥å
ä¿¡å·ãã«ããŠã³ã¡ãã¥ãŒã« .tdmsãã¡ã€ã«ããã®é³å§ä¿¡å·ããã§ã«
å«ãŸããŠããããšã«æ³šç®ããŠãã ãããé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã
ã¯ãã°ããŒã¿ããé³å§ã¬ãã«ãã¹ããããžã®å ¥åãšããŠèªåçã«éžæ
ããŸãã
4. ã¹ãããèšå®ã¿ãã®éã¿ä»ãã¿ããã¯ãªãã¯ããéã¿ä»ããã«ããŠã³
ã¡ãã¥ãŒãã A ç¹æ§ãéžæããŠãäœã¬ãã«ããŒã³ã®ã©ãŠããã¹ãã·
ãã¥ã¬ãŒã·ã§ã³ããŸãã
5. ã¹ãããèšå®ã¿ãã®å¹³ååã¿ããã¯ãªãã¯ããLeq ããã³ããŒã¯ã®
ãã§ãã¯ããã¯ã¹ããªãã«ããŠãä¿¡å·ã®é£ç¶ Leq ããã³ææ°å¹³åé³
å§ã¬ãã«ã®ã¿ãèšç®ããŸãã
6. é³å§ä¿¡å·ãããã°ããŒã¿ããŠã£ã³ããŠããŒã¿ãã¥ãŒã¿ããžãã©ãã°ã
ãŸãã
第 4 ç« é³é¿ãæ¯åã®æž¬å®ãè¡ã
© National Instruments 4-3 NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã
ã¡ã¢ åçã¢ãŒãã§ãã°ãããä¿¡å·ã衚瀺ããå Žåã1 床㫠1 ã€ã®å埩ããŒã¿ã®ã¿ã
衚瀺ããŸãããã¹ãŠã®ãã°ããŒã¿ã衚瀺ããã«ã¯ãäœæ¥ãšãªã¢ãã«ããŠã³ã¡
ãã¥ãŒããç£èŠ / èšé²ãéžæããŠãç£èŠ / èšé²ãäœæ¥ãšãªã¢ã«åãæ¿ããŸãã
ãŸããåçãªãã·ã§ã³ã¿ãã§ãããã¹ãŠã®ãã°ããŒã¿ã衚瀺ã§ããŸãã
7. ãããžã§ã¯ããå®è¡ããŸãã
èšé²æžã¿ã®é³å§ããŒã¿ãããŒã¿ãã¥ãŒã¿ãã«è¡šç€ºããããã¥ãŒã®äžæ¹
ã«æé軞ã¹ã©ã€ããããããšã«æ³šç®ããŠãã ããããããžã§ã¯ãã¯ã
ãã°ãããããŒã¿ãã¡ã€ã«ã®æåŸã«å°éãããŸã§å®è¡ãããŸãã
ãã³ã ãããžã§ã¯ããå®è¡äžã§ãªããšãã¯ãæé軞ã¹ã©ã€ããã¯ãªãã¯ããŠãã©ãã°
ãããšã以åã«èšé²æžã¿ã®ããŒã¿ã®äžéšåã衚瀺ã§ããŸãã
8. ã°ã©ããå³ã¯ãªãã¯ããã·ã§ãŒãã«ããã¡ãã¥ãŒãã衚瀺é ç®âã«ãŒ
ãœã«ãéžæããŠãå³ 4-2 ã«ç€ºã 2 ã€ã®å¯Ÿè©±åŒã«ãŒãœã«ã衚瀺ããŸãã
å³ 4-2 ã°ã©ãäžã«ã«ãŒãœã«ã衚瀺ãã
第 4 ç« é³é¿ãæ¯åã®æž¬å®ãè¡ã
NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã 4-4 ni.com
9. 衚瀺ããããŒã¿ã®ãµãã»ãããå¢æžããã«ã¯ãã«ãŒãœã«ãã¯ãªãã¯ã
ãã©ãã°ããŸããã«ãŒãœã«ã移åãããã³ã«ãã«ãŒãœã«ã® X å€ãš Yå€ãæŽæ°ãããããŒã¿ãã¥ãŒã¿ãã®ã°ã©ãã®äžã«ããã«ãŒãœã«ã®è¡šã«
衚瀺ãããŸãã
10. ãããžã§ã¯ããä¿åããŸãã
ããŒã¿ã解æãã
é³å§ããŒã¿ã«ããã«æž¬å®ãè¿œå ããã«ã¯ã以äžã®æé ã«åŸã£ãŠãã ããã
1. ãé³å§ã¬ãã«ãã¹ãããã®é£ç¶ Leq åºåãããŒã¿ãã¥ãŒã¿ãã«ãã©ã
ã°ããŸããé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã®ããŒã¿ãã¥ãŒã§æ°èŠãã£ãŒ
ããäœæãããŸãã
2. ãé³å§ã¬ãã«ãã¹ãããã®ææ°åºåãããŒã¿ãã¥ãŒã¿ãã®æ¢åã®
ãã£ãŒãã«ãã©ãã°ããŸãã
3. ãããžã§ã¯ããå®è¡ããŸãã
ããŒã¿ãã¥ãŒã§ã¯ãåçããŒã¿ã®é²è¡ã«åŸã£ãŠé£ç¶ Leq ããã³ææ°
ããã£ãŒãå ã§æŽæ°ãããããšã«æ³šç®ããŠãã ããã
4. ã¹ããããè¿œå ãã¿ã³ãã¯ãªãã¯ãã解æâåšæ³¢æ°é å枬å®âãªã¯
ã¿ãŒã解æãéžæããŸãã
5. èŠååšæ³¢æ°ç¯å²ã匷å¶ãããŸãããã¹ãããèšå®ã¿ãã®äžæ¹ã«è¡šç€ºã
ããããšã«æ³šç®ããŠãã ãããããªã¯ã¿ãŒã解æãã¹ãããã¯ãå ¥å
ä¿¡å·ã®ãµã³ãã«ã¬ãŒãã«åºã¥ããŠèªåçã«åšæ³¢æ°ç¯å²ã匷å¶ããŸãã
åšæ³¢æ°ç¯å²ãå€æŽããŠèŠåã解æ¶ããã«ã¯ãé«åšæ³¢æ°ïŒHzïŒãã«ã
ãŠã³ã¡ãã¥ãŒãã 1k ãéžæããŸãã
6. ã¹ãããèšå®ã¿ãã®å ¥åã¿ããã¯ãªãã¯ããå ¥åä¿¡å·ããé³å§ã¬ã
ã«ãã¹ãããã§äœ¿çšããé³å§ä¿¡å·ãšåãã§ããããšã確èªããŠãã ã
ãã
7. ã¹ãããèšå®ã¿ãã®æ§æã¿ããã¯ãªãã¯ããéã¿ä»ããã«ããŠã³ã¡
ãã¥ãŒãã A ç¹æ§ãéžæããŸãã
8. ããªã¯ã¿ãŒã解æãã¹ãããã®ãªã¯ã¿ãŒãåºåãããŒã¿ãã¥ãŒã¿ãã«
ãã©ãã°ããŸãã
第 4 ç« é³é¿ãæ¯åã®æž¬å®ãè¡ã
© National Instruments 4-5 NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã
å³ 4-3 ã«ç€ºãããã«ãé³å§ãé£ç¶ Leqãææ°ãããã³ãªã¯ã¿ãŒãã®
åãã£ãŒãã衚瀺ãããŸãã
å³ 4-3 ãªã¯ã¿ãŒã解æãè¡ã£ã My Analysis.seproj
9. ããªã¯ã¿ãŒã解æãã¹ãããã®åž¯åå ãã¯ãŒåèšåºåããããŒã¿
ãã¥ãŒã¿ãã®é£ç¶ Leq ããã³ææ°ã衚瀺ãããŠãããã£ãŒããžã
ã©ãã°ããŸãã
10. æ°å€çµæãã°ã©ãã§ã¯ãªãè¡šã§è¡šç€ºãããå Žåã¯ããã£ãŒããå³ã¯
ãªãã¯ã衚瀺ãªãã·ã§ã³âè¡šãéžæããŸãã
11. ãããžã§ã¯ããå®è¡ããŸãã
ãªã¯ã¿ãŒãã®ã°ã©ããåçã®é²è¡ã«åŸã£ãŠæŽæ°ãããããšã«æ³šç®ããŠ
ãã ãããããã§ãé³å§ããŒã¿ã«ãããé³å§ã¬ãã«ãšãªã¯ã¿ãŒã解æ
ã¯å®äºã§ãã
12. ãããžã§ã¯ããä¿åããŠéããŸãã
第 4 ç« é³é¿ãæ¯åã®æž¬å®ãè¡ã
NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã 4-6 ni.com
äžçŽåç
åçãªãã·ã§ã³ãã¥ãŒã§ãäžçŽããŒã¿åçãªãã·ã§ã³ãæ§æããŸããåç
ãªãã·ã§ã³ã¿ãã§ã¯ããã°ããŒã¿ã®ãã¬ãã¥ãŒã衚瀺ããåçããããŒã¿
ã®ãµãã»ãããéžæãããã解æã¹ãããã§ããŒã¿ãåŠçããããšãã§ã
ãŸãã
äžçŽåçãªãã·ã§ã³ãæ§æããã«ã¯ã以äžã®æé ã«åŸã£ãŠãã ããã
1. ãã¡ã€ã«âãããžã§ã¯ããéããéžæããMy Logged
Signal.seprojãåç §ããŠéããã¿ã³ãã¯ãªãã¯ããŸãã
2. äœæ¥ãšãªã¢ãã«ããŠã³ã¡ãã¥ãŒããåçãéžæããŸãã
3. åçãªãã·ã§ã³ã¿ããã¯ãªãã¯ãããã衚瀺âåçãªãã·ã§ã³ãéžæ
ããŸããå³ 4-4 ã®ãããªåçãªãã·ã§ã³ã¿ãã衚瀺ãããŸãã
å³ 4-4 åçãªãã·ã§ã³ã¿ã
4. ãã°ä¿¡å·ãªã¹ãããã¯ã¹ã® 2 çªç®ã®ãã°ãã signal ãéžæããŸãã
5. ãã°ãããä¿¡å·ã®ãã¡ãã°éå§ãã 1 ç§åŸã«å§ãŸã 4 ç§åŸã«çµãã 3ç§éã®ãµãã»ãããåçãŸãã¯è§£æããã«ã¯ãéå§æéïŒsïŒãã£ãŒ
ã«ãã« 1ããåæ¢æéïŒsïŒãã£ãŒã«ãã« 4ãå ¥åããŸãã
6. ã¹ããããè¿œå ãã¿ã³ãã¯ãªãã¯ããåŠçâãã£ã«ã¿ãéžæããŸãã
7. ããã£ã«ã¿ãã¹ãããã®ãã£ã«ã¿åŸã®ããŒã¿åºåãããŒã¿ãã¥ãŒã¿ã
ã«ãã©ãã°ããŸãã
第 4 ç« é³é¿ãæ¯åã®æž¬å®ãè¡ã
© National Instruments 4-7 NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã
8. å®è¡ãã¿ã³ãã¯ãªãã¯ããŸããåçãªãã·ã§ã³ã¿ãã§æå®ããä¿¡å·ã®
ãµãã»ããããã£ã«ã¿åŠçããããŒã¿ãã¥ãŒã¿ãã«çµæãšããŠåŸãã
ãããã£ã«ã¿åŠçæžã¿ä¿¡å·ã衚瀺ããŸããä¿¡å·ããã°éå§ãã 1 ç§
åŸã«å§ãŸãå 4 ç§åŸã«çµããããšã確èªããŠãã ããã
9. ãããžã§ã¯ããä¿åããŸãã
ã¢ã©ãŒã æ¡ä»¶ãäžçŽåçãªãã·ã§ã³ã®æå®ãªã©ããŒã¿ãã®ã³ã°ã®è©³çŽ°ã«ã€
ããŠã¯ããã«ãâé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ããã«ããéžæããŠè¡šç€ºãã
ããé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ããã«ãããåç §ããŠãã ããã
é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã«ãããä¿¡å·ã¿ã€ã
ã¹ãããã®äžã«ã¯ããæŒç®ãã¹ããããªã©ãè€æ°ã®ä¿¡å·ã¿ã€ããåŠçã§ãã
ãã®ããããŸããããšãã°ããæŒç®ãã¹ãããã§ã¯ãæéé åãŸãã¯åšæ³¢æ°
é åã®ä¿¡å·ãåŠçããããšãã§ããŸãããã®ãæŒç®ãã¹ãããã¯ãéžæã
ãå ¥åä¿¡å·ã®ã¿ã€ãã«å¿ããŠç°ãªãåäœãããŸããããšãã°ã2 ã€ã®æé
é åä¿¡å·ãå ç®ãããšããã®æ¯å¹ ã®ã¿ãå ç®ãããŸããããã«å¯Ÿãã2 ã€
ã®åšæ³¢æ°é åäœçžä¿¡å·ãå ç®ãããšãé©åãªäœçžã·ãããè¿œå ãããŸãã
ä¿¡å·ãä¿åããšã¯ã¹ããŒãããã³å°å·ãã
é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã§ä¿¡å·ãããã¥ã¡ã³ãåããããå¥ã®ãœãã
ãŠã§ã¢ã¢ããªã±ãŒã·ã§ã³ã§è§£æãè¡ãããšãã§ããŸãããã®ã»ã¯ã·ã§ã³ã§
ã¯ãä¿¡å·ã® ASCII ãã¡ã€ã«ãžã®ä¿åãMicrosoft Excel ãžã®ããŒã¿ã®ãšã¯ã¹
ããŒããä¿¡å·ã®å°å·ãæšæºã®ããã¥ã¡ã³ãæ©èœã䜿çšããé³é¿ / æ¯åèšæž¬
ã¢ã·ã¹ã¿ã³ããããžã§ã¯ãã®ããã¥ã¡ã³ãäœæã®æ¹æ³ã«ã€ããŠèª¬æããŸãã
ä¿¡å·ããã¡ã€ã«ã«ä¿åããMy Signals.seprojã®ä¿¡å·ããã¡ã€ã«ãžä¿åããã«ã¯ã以äžã®æé ã«
åŸã£ãŠãã ããã
1. ã¹ããããè¿œå ãã¿ã³ãã¯ãªãã¯ããä¿¡å·ãããŒã / ä¿åâ
ASCII/LVM ã«ä¿åãéžæããŸãããããžã§ã¯ããå®è¡ããããã³
ã«ããä¿¡å·ãããŒã / ä¿åãã¹ãããã䜿çšããããŒã¿ããã¡ã€ã«ãž
ä¿åããŸãã
2. ã¹ãããèšå®ã¿ãã®ãã¡ã€ã«èšå®ã¿ããã¯ãªãã¯ããå·Šã«ç€ºãããšã¯
ã¹ããŒããã¡ã€ã«ã®ãã¹ãã£ãŒã«ãã®æšªã«ããåç §ãã¿ã³ãã¯ãªãã¯
ããŸãã
3. National Instruments¥Sound and Vibration Assistant¥
ExamplesÂ¥TutorialÂ¥ãã£ã¬ã¯ããªãåç §ãããã¡ã€ã«åãšããŠ
octave.txtãšå ¥åããŸãã
4. ãã¡ã€ã«ãæ¢åã®å Žåãã«ããŠã³ã¡ãã¥ãŒããäžæžããéžæããŸãã
第 4 ç« é³é¿ãæ¯åã®æž¬å®ãè¡ã
NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã 4-8 ni.com
5. ãããžã§ã¯ããå®è¡ããçµæã®å€ãç¹å®ã® ASCII ãã¡ã€ã«ã«ä¿åã
ãŸãã
6. ãããžã§ã¯ããä¿åããŸãã
7. octave.txtãå«ããã©ã«ããåç §ããããã¹ããšãã£ã¿ã§ãã¡ã€ã«
ãéããŠä¿åãããŠããããŒã¿ã衚瀺ããŸãã
Microsoft Excel ãžä¿¡å·ããšã¯ã¹ããŒãããMicrosoft Excel ãžä¿¡å·ããŒã¿ããšã¯ã¹ããŒãããã«ã¯ãExcel ãèµ·åãã
é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã«ããã¹ãããã®åºåä¿¡å·ã Excel ã®ã¹ã
ã¬ããã·ãŒããžãã©ãã°ããŸããããŒã¿ã¯ã¿ãåºåããšãªããŸãã
ãŸããããŒã¿ãã¥ãŒã¿ãã®è¡šç€ºãå³ã¯ãªãã¯ããã·ã§ãŒãã«ããã¡ãã¥ãŒ
ãããšã¯ã¹ããŒãâ Microsoft Excel ãéžæããŠãããŒã¿ãã¥ãŒã¿ãã®
éžæããã°ã©ãããã®ããŒã¿ããšã¯ã¹ããŒãããŸãããŸããoctave.txt
ãªã©ä¿åæžã¿ã®ããŒã¿ã Microsoft Excel ã§éãããšãã§ããŸãã
ä¿¡å·ãå°å·ããã°ã©ãã®ç»åãå°å·ããã«ã¯ãããŒã¿ãã¥ãŒã¿ãã衚瀺ãããã¡ã€ã«âå°
å·âããŒã¿ãã¥ãŒãå°å·ãéžæããŸãã
é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã§ã¬ããŒããäœæãããããžã§ã¯ãããã¥ã¡ã³ãã¿ãã衚瀺ããã«ã¯ã衚瀺âãããžã§ã¯ãã
ãã¥ã¡ã³ããéžæããŸããããã¹ããã°ã©ããªã©ã®ç»åã䜿çšããŠãã
ãžã§ã¯ãã®èª¬æãèšèŒã§ããŸãããããžã§ã¯ããã¥ãŒã®ã¹ãããã®åºåã
ãããžã§ã¯ãããã¥ã¡ã³ãã¿ãã«ãã©ãã°ãããšãåºåä¿¡å·ã®ã°ã©ããè¡š
瀺ãããŸãããããžã§ã¯ãå®è¡äžã®å Žåã¯ããããžã§ã¯ãããã¥ã¡ã³ãã¿
ãã®ã°ã©ããã¹ãããã®åºåã®ææ°å€ãåæ ããããã«èªåçã«æŽæ°ãã
ãŸãã
ããã¥ã¡ã³ããå°å·ããã«ã¯ããããžã§ã¯ãããã¥ã¡ã³ãã¿ããéãã
ãã¡ã€ã«âå°å·âããã¥ã¡ã³ããå°å·ãéžæããããããã¥ã¡ã³ããå°å·
ãã¿ã³ãã¯ãªãã¯ããŸãã
ããã¥ã¡ã³ãã HTML ã«ãšã¯ã¹ããŒãããããã¥ã¡ã³ãã HTML ãžãšã¯ã¹ããŒãããã«ã¯ããããžã§ã¯ãããã¥ã¡
ã³ãã¿ããéãããã¡ã€ã«âãšã¯ã¹ããŒãâããã¥ã¡ã³ãã HTML ã«ãš
ã¯ã¹ããŒããéžæããŸãããŸãããããžã§ã¯ãããã¥ã¡ã³ãã¿ããå³ã¯
ãªãã¯ããã·ã§ãŒãã«ããã¡ãã¥ãŒããããã¥ã¡ã³ãã HTML ã«ãšã¯ã¹
ããŒããéžæããŠãããã¥ã¡ã³ãã®ãšã¯ã¹ããŒããã§ããŸãã
© National Instruments 5-1 NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã
5ã¹ã€ãŒã枬å®ãè¡ã
é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã§ã¹ã€ãŒãæäœãäœæããŠã枬å®ãèªåå
ããç¹æ§è©äŸ¡ãèšèšæ€èšŒãè¡ãããšãã§ããŸããåºç¯ãªæ¡ä»¶ã§ã¹ã€ãŒã枬
å®ã䜿çšããŠæ€æ»å¯Ÿè±¡ããã€ã¹ããããŒã¿ãéé²ãããã®æ§èœãããã¥ã¡
ã³ãåããããšãã§ããŸããããšãã°ã枬å®äžã«ãã¹ã€ãŒãæäœã«ãããŠ
åºæ¿ä¿¡å·ã®åšæ³¢æ°ãé»æºã®é»å§ã¬ãã«ãå€æŽããããšãã§ããŸãã
ãã®ç« ã§ã¯ãé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã§ãã¹ã€ãŒããã¹ãããã䜿çš
ããæäœã®èšå®æ¹æ³ã«ã€ããŠèª¬æããŸããç¹å®ã®ç¯å²ã§åšæ³¢æ°å€ãã¹ã€ãŒ
ãããŠãã£ã«ã¿åºåã®æž¬å®ãè¡ãããšã§ããã£ã«ã¿ã®æ§èœãè©äŸ¡ããæ¹
æ³ãã¹ã€ãŒãçµæã®è¡šç€ºæ¹æ³ãããè€éãªæž¬å®ã®ããã®å€æ¬¡å ã¹ã€ãŒãã®
å®è¡æ¹æ³ã«ã€ããŠã説æããŸãã
ãŸãããã·ã§ãã«ã€ã³ã¹ãã«ã¡ã³ãã®ä»»æ波圢çºçåšãé¢æ°çºçåšããã€
ãããã¯ä¿¡å·éé²ïŒDSAïŒããã€ã¹ããŸãã¯ãã«ããã¡ã³ã¯ã·ã§ã³ I/OïŒMIOïŒããã€ã¹ããçæãããç©çä¿¡å·ãã¹ã€ãŒãããããšãã§ããŸ
ãã
ã¡ã¢ 以äžã®æŒç¿ãç©çä¿¡å·ã䜿çšããŠè¡ãã«ã¯ãNI-DAQmx 8.9.5 以éãã€ã³ã¹ããŒ
ã«ããã¢ããã°å ¥åºåãæèŒãã DAQ ããŒããŠã§ã¢ã䜿çšããå¿ èŠããããŸ
ãã
ã¹ã€ãŒãã®ç¯å²ãšåºåãå®çŸ©ãã
ãã¹ã€ãŒããã¹ãããã§ã¯ãè€éã§å埩çãªã¹ã€ãŒãæäœã§ãããããå®
矩ããããšã«ãã枬å®ãèªååããããšãã§ããŸãã
ã¹ã€ãŒããããããžã§ã¯ããäœæãããããžã§ã¯ããäœæããã«ã¯ã以äžã®æé ãå®è¡ããŸãã
1. æ°èŠãããžã§ã¯ããéããŸãã
2. ãããžã§ã¯ãã My Sweep.seprojãšããååã§ä»»æã®ãã£ã¬ã¯ããª
ã«ä¿åããŸãã
3. ãã¢ããã°ä¿¡å·ãçæãã¹ããããè¿œå ããŸãã
4. ã¹ãããèšå®ã¿ãã®æ§æã¿ãã§ããããã¯ãµã€ãºïŒSïŒãã£ãŒã«ãã«
10000ãšå ¥åããŸãã
5. ãã£ã«ã¿ã¹ããããè¿œå ããŸãã
第 5 ç« ã¹ã€ãŒã枬å®ãè¡ã
NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã 5-2 ni.com
6. ã¹ãããèšå®ã¿ãã®æ§æã¿ãã§ãã¿ã€ããã«ããŠã³ã¡ãã¥ãŒãããã³
ããã¹ãéžæããŸãããã£ã«ã¿ä»æ§ã»ã¯ã·ã§ã³ã§ãªãã·ã§ã³ãå€æŽã
ããšãã£ã«ã¿æ¯å¹ å¿çïŒdBïŒã°ã©ããæŽæ°ãããããšã«æ³šç®ããŠã
ã ããã
7. äœåã«ãããªãïŒHzïŒãã£ãŒã«ãã« 2000ãšå ¥åããé«åã«ãããªã
ïŒHzïŒãã£ãŒã«ãã« 5000ãšå ¥åããŸãã
8. ã¹ããããè¿œå ãã¿ã³ãã¯ãªãã¯ãã解æâåšæ³¢æ°é å枬å®âãªã¯
ã¿ãŒã解æãéžæããŸãã
9. ã¹ãããèšå®ã¿ãã®ã·ã³ã°ã«ããŒã³æž¬å®ã¿ããã¯ãªãã¯ããæªã¿ãã
ã³ãã£ã³ãã«è©äŸ¡ã»ã¯ã·ã§ã³ã«ãã THD+N ããã³ THD ã®ãã§ãã¯
ããã¯ã¹ããªã³ã«ããSN æ¯ãSINAD ããã³ SFDR ã®ãã§ãã¯ããã¯
ã¹ããªãã«ããŸãã
10. ãããžã§ã¯ããä¿åããŸãã
ã¹ã€ãŒããå®çŸ©ããåšæ³¢æ°ç¯å²ãå®çŸ©ãããã£ã«ã¿ãä»ããŠã¹ã€ãŒãæäœãè¡ãã«ã¯ã以äžã®
æé ã«åŸã£ãŠãã ããã
1. ã¹ããããè¿œå ãã¿ã³ãã¯ãªãã¯ããå®è¡å¶åŸ¡âã¹ã€ãŒããéžæããŸ
ãã
2. å³ 5-1 ã®ããã«ãã¹ã€ãŒãæ§æã¿ãã®è¿œå ãã¿ã³ãã¯ãªãã¯ãããã
ãžã§ã¯ãã®åã¹ãããã®ã¹ã€ãŒããã©ã¡ãŒã¿ã®ãªã¹ãã衚瀺ããŸãã
å³ 5-1 ã¹ã€ãŒããã©ã¡ãŒã¿éžæãã€ã¢ãã°ããã¯ã¹
第 5 ç« ã¹ã€ãŒã枬å®ãè¡ã
© National Instruments 5-3 NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã
3. ã¢ããã°ä¿¡å·ãäœæ : æ£åŒŠæ³¢ããåšæ³¢æ° (Hz) ãã©ã¡ãŒã¿ãéžæãã
OK ãã¿ã³ãã¯ãªãã¯ããŸãã
å³ 5-2 ã«ç€ºãããã«ããããžã§ã¯ããã¥ãŒå ã§ããã¹ã€ãŒããã¹ãã
ããã¹ã€ãŒãããããã®ä¿¡å·ãšãªããã¢ããã°ä¿¡å·ãäœæãã¹ããã
ãå²ãŸããŠããããšã«æ³šç®ããŠãã ããã
å³ 5-2 ãã¹ã€ãŒããã¹ãããã«å²ãŸããŠãããã¢ããã°ä¿¡å·ãäœæãã¹ããã
4. ã¹ã€ãŒãæ§æã¿ãã®ã¿ã€ããã«ããŠã³ã¡ãã¥ãŒããææ°ãéžæããŸ
ãã
5. éå§ : åšæ³¢æ°ïŒHzïŒãã£ãŒã«ãã« 2000ãšå ¥åããçµäº : åšæ³¢æ°
ïŒHzïŒãã£ãŒã«ãã« 5000ãšå ¥åããŸãã
第 5 ç« ã¹ã€ãŒã枬å®ãè¡ã
NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã 5-4 ni.com
6. ãã€ã³ãæ°ãã£ãŒã«ãã« 20ãšå ¥åããŸãã
ã¹ã€ãŒãæ§æã¿ãã¯å³ 5-3 ã®ããã«ãªããŸãã
å³ 5-3 ãã¹ã€ãŒãæ§æãã¿ã
ãã¹ã€ãŒããã¹ããããæ§æãããã¢ããã°ä¿¡å·ãäœæãã¹ãããã®åš
æ³¢æ°ïŒHzïŒãã©ã¡ãŒã¿ã®å埩ããåå€ã®ç¯å²ãæå®ããŸãããã¢ãã
ã°ä¿¡å·ãäœæãã¹ãããã§ã¯ãå®çŸ©ãããåšæ³¢æ°ç¯å²ãããšã«ãååš
æ³¢æ°ã§æ£åŒŠæ³¢ãçæããŸãããã¹ã€ãŒããã¹ãããã§ã¯ããããžã§ã¯
ãã®ãã¹ãŠã®ã¹ã€ãŒãå¯èœãªã¹ãããã®ã¹ã€ãŒããã©ã¡ãŒã¿ã®åå€ã«
察ããŠå埩ããããæå®ã§ããŸãã
7. ã¹ãããèšå®ã¿ãã®ã¹ã€ãŒãåºåã¿ããã¯ãªãã¯ããŸãã
8. è¿œå ãã¿ã³ãã¯ãªãã¯ãããããžã§ã¯ãã®åã¹ãããã®ã¹ã€ãŒãåºå
ãªã¹ãã衚瀺ããŸãã
9. ããŒã³æž¬å®ããåºæ¬æ¯å¹ åºåãéžæããOK ãã¿ã³ãã¯ãªãã¯ããŠã
ã¹ã€ãŒããããåšæ³¢æ°ïŒHzïŒãã©ã¡ãŒã¿ã«å¯ŸããŠãã®æž¬å®ã衚瀺ã
ãŸãã
第 5 ç« ã¹ã€ãŒã枬å®ãè¡ã
© National Instruments 5-5 NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã
ã¹ã€ãŒãæäœã§äœ¿çšããã¹ãããããã¹ãŠå«ãããããå³ 5-4 ã«ç€ºã
ããã«ããã¹ã€ãŒããã¹ããããã«ãŒããäœæããŠãããžã§ã¯ã
ãã¥ãŒã®å šã¹ããããå²ãã§ããããšã«æ³šç®ããŠãã ããã
å³ 5-4 ã¹ã€ãŒãæäœã§å¿ èŠãªãã¹ãŠã®ã¹ããããå²ãã§ãããã¹ã€ãŒããã¹ããã
10. ãããžã§ã¯ããä¿åããŸãã
ã¹ã€ãŒãã®çµæããããããã
ã¹ã€ãŒã枬å®ãå®è¡ããã«ã¯ã以äžã®æé ã«åŸã£ãŠãã ããã
1. ããã£ã«ã¿ãã¹ãããã®ãã£ã«ã¿åŸã®ããŒã¿åºåãããŒã¿ãã¥ãŒã¿ã
ã«ãã©ãã°ããŸãã
2. ãã¢ããã°ä¿¡å·ãäœæãã¹ãããã®æ£åŒŠæ³¢åºåãããŒã¿ãã¥ãŒã¿ãã«
ãã©ãã°ããŸãã
3. 1 åå®è¡ãªãã·ã§ã³ãå床éžæããŠã¹ã€ãŒããå®è¡ããŸãã
ãã®ãããžã§ã¯ãã§ã¯ããã¢ããã°ä¿¡å·ãäœæãã¹ãããã«ããæ£åŒŠ
æ³¢ã®åºæ¿ä¿¡å·ãçæãããŸãããã®ä¿¡å·ã¯ãããã£ã«ã¿ãã¹ãããã«
ãããã³ããã¹ãã£ã«ã¿ãä»ããŠæž¡ããããããŒã³æž¬å®ãã¹ãããã«
ãããã£ã«ã¿åºåã® THD+N ããã³ THD ã®ã¬ãã«ã枬å®ãããŸãã
第 5 ç« ã¹ã€ãŒã枬å®ãè¡ã
NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã 5-6 ni.com
ããã£ã«ã¿ãã¹ãããã¯ã·ãã¥ã¬ãŒããããæ€æ»å¯Ÿè±¡ãšããŠåäœãã
ãããããŒããŠã§ã¢ã¯äœ¿çšããŸããã
ã°ã©ãäžã®æ£åŒŠæ³¢åºåã¯ãæå®ãããåšæ³¢æ°ç¯å²ã§ç¹°ãè¿ãåºåãã
ãŸãã
4. ã¹ã€ãŒãã«ãŒãã®äžã®åºæ¬æ¯å¹ vs. åšæ³¢æ°ïŒHzïŒä¿¡å·ãããŒã¿ãã¥ãŒ
ã¿ãã«ãã©ãã°ããã¹ã€ãŒãåºåã衚瀺ããŸãã
æ°èŠã§ã°ã©ããäœæãããŸããå³ 5-5 ã®ããã«ãã¹ã€ãŒãæäœã®ããŒ
ã¿ã¯ãç°ãªãã°ã©ããå¿ èŠãšãã XY é åã§ãã
å³ 5-5 ã¹ã€ãŒããããžã§ã¯ã
5. 1 åå®è¡ãªãã·ã§ã³ãéžæããŠã¹ã€ãŒã枬å®ãå®è¡ããŸãã
ãããžã§ã¯ãã®å®è¡äžã«ãæ°èŠã°ã©ãã§ããã£ã«ã¿ãã¹ãããã®åšæ³¢
æ°å¿çïŒæ¯å¹ ïŒdBïŒå¯Ÿåšæ³¢æ°ïŒããããããããŸããã°ã©ãã¯ãåº
æ¬æ¯å¹ åºå察åšæ³¢æ°ã®ãã£ã«ã¿ã®äŒéé¢æ°ã衚瀺ããŸãã
ã¡ã¢ ããã©ã«ãã§ã¯ãããŒã¿ãã¥ãŒã¿ãã®è¡šç€ºã¯ã¹ã€ãŒãã®å埩ããšã«æ¶å»ãããŸ
ãããããã£ã«ã¿ãã¹ãããã®åšæ³¢æ°å¿çã¯ã©ã®å埩ã§ãåãã§ããããããã
ãžã§ã¯ãã®å®è¡äžã«ä¿¡å·ã衚瀺ããã°ã©ããæŽæ°ãããŠããªãããã«èŠããŸãã
ãªãã·ã§ã³ãã€ã¢ãã°ããã¯ã¹ã®ããŒã¿ããŒãžã§ãã¹ã€ãŒãã®å埩ããšã«
第 5 ç« ã¹ã€ãŒã枬å®ãè¡ã
© National Instruments 5-7 NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã
ãã¥ãŒãæ¶å»ãããã©ããéžæã§ããŸããããŒã«âãªãã·ã§ã³ãéžæããŠãªã
ã·ã§ã³ãã€ã¢ãã°ããã¯ã¹ã衚瀺ããŸãã
6. ãããžã§ã¯ããä¿åããŠéããŸãã
ã¹ãããèšå®ã¿ãã®ã¹ã€ãŒãæ§æã¿ãã«å¥ã®ãã©ã¡ãŒã¿ãè¿œå ããããš
ã§ããã¹ã€ãŒããã¹ãããã䜿çšããè€æ°ã®ãã©ã¡ãŒã¿ãåæã«ã¹ã€ãŒã
ããããšãã§ããŸããè€æ°ã®ãã©ã¡ãŒã¿ãåæã«ã¹ã€ãŒãããããšããå¹³
è¡ã¹ã€ãŒããšåŒã³ãŸããããšãã°ãåºæ¿ä¿¡å·ã®æ¯å¹ ãå€æŽããã«ã¯ãå¹³è¡
ã¹ã€ãŒããå®è¡ããŠããžã¿ã€ã¶ã DSA ããã€ã¹ã®ä¿¡å·ã¬ãã«ã®å€æŽã«å¿
ããŠå ¥åç¯å²ã調æŽããããšã§ãéé²ã®ç²ŸåºŠãæ倧éã«é«ããããšãã§ã
ãŸããä¿¡å·ã¬ãã«ãäžæãããã枬å®ããã€ã¹ã®å ¥åç¯å²ãåºããã°ãã
ã®æž¬å®ã«èšæž¬åšãæã€å解èœãæ倧éã«æŽ»ããããšãã§ããŸãã
© National Instruments 6-1 NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã
6次æ°è§£æãè¡ã
NI é³é¿ / æ¯åãœãããŠã§ã¢ããã±ãŒãžããè³Œå ¥ããŠããã ããšãé³é¿ /æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã䜿çšããŠæ€æ»å¯Ÿè±¡ããã€ã¹ïŒDUTïŒã«å¯Ÿãã次æ°
é å枬å®ãè¡ãããšãã§ããŸãããããã®æž¬å®ã¯æ¬¡æ°è§£æãšããŠç¥ãããŠ
ããããã¹ãäžã§ããã¹ãåŸã§ãå®è¡ã§ããŸãããã®ç« ã§ã¯ãäž¡æ¹ã®ç¶æ³
ã«ã€ããŠèª¬æãããµã³ãã«ããããŸãã
æåã®ãã¥ãŒããªã¢ã«ã§ã¯ããŠãããŒãµã«ãã¡ã€ã«ãã©ãŒããããã€ããª
ïŒUFF58bïŒãã¡ã€ã«ã«å¯ŸããŠæ¬¡æ°è§£æãå®è¡ããŸãããã®ãã¡ã€ã«ã«ã¯ 4æ°çãšã³ãžã³ã®ã©ã³ããããã¹ãããéé²ãããããŒã¿ãå«ãŸããŸããã
ã®ãµã³ãã«ã¯éé²ããããã¹ãããŒã¿ã«å¯ŸããŠæ¬¡æ°è§£æãè¡ãããã®æ¹æ³
ã説æããŸãã
2 çªç®ã®ãµã³ãã«ã§ã¯ãé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã®åçæ©èœã䜿çšã
㊠PC ãã¡ã³ã®ã©ã³ããããã¹ããã·ãã¥ã¬ãŒã·ã§ã³ããŸãããã®ã·ãã¥
ã¬ãŒã·ã§ã³ãã¹ãäžã«ãéé²ãããããŒã¿ã«å¯ŸããŠæ¬¡æ°è§£æãå®è¡ããŸ
ãããã®ãµã³ãã«ã¯ããŒã¿ã®éé²ãšè§£æãåæã«å®è¡ããæ¹æ³ã説æããŸ
ãã
ãããã®ãµã³ãã«ã¯äž¡æ¹ãšãã©ã³ããããã¹ãã«é¢é£ããŠããŸããã©ã³
ããããã¹ãã§ã¯ DUT ã®å転é床ãåäœç¯å²å šäœã«ããã£ãŠå€åããã
ããDUT ã«ãããé¡èãªéšé³ãæ¯åã®åå ãšãªã次æ°ãç¹å®ããã®ã«åœ¹
ç«ã¡ãŸãã
ãã°ããŒã¿ã«å¯ŸããŠæ¬¡æ°è§£æãè¡ã
ãã®ã»ã¯ã·ã§ã³ã®ãµã³ãã«ã§ã¯ã4 æ°çãšã³ãžã³ã®ã©ã³ããããã¹ãã«ã
ããã°ããŒã¿ã䜿çšããŸãããã®ããŒã¿ã«æ¬¡æ°è§£æãå®è¡ããŠãé¡èãªãš
ã³ãžã³éšé³ã®åå ãšãªã次æ°ãç¹å®ããŸãããã®ããŒã¿ã«ã¯ãã¹ãã«ãã
以äžã® 2 ã€ã®ä¿¡å·ãå«ãŸããŠããŸãã
⢠ãšã³ãžã³ã®å転é床ã枬å®ããã¿ã³ã¡ãŒã¿ãã«ã¹ä¿¡å·ãã¿ã³ã¡ãŒã¿ã¯
ãšã³ãžã³ã 1 å転ããããšã«ãã«ã¹ã 1 ã€çæããŸãã
⢠ãšã³ãžã³ããçããé³ã枬å®ãããã€ã¯ããã³ä¿¡å·ããã®ç¶æ³ã§ã¯ã
枬å®ãããé³ã¯äžèŠãªéšé³ã瀺ããŸãã
以äžã®ã»ã¯ã·ã§ã³ã§ã¯ã次ã®æé ã«åŸã£ãŠãã ããã
1. éé²ãããããŒã¿ã UFF58b ãã¡ã€ã«ããããŒãããŸãã
2. ãã¹ãå®æœäžããšã³ãžã³ã®å転é床ã枬å®ããŸãã
第 6 ç« æ¬¡æ°è§£æãè¡ã
NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã 6-2 ni.com
3. é¡èãªãšã³ãžã³éšé³ã®åå ãšãªã次æ°ãç¹å®ããŸãã
4. ç¹å®ãããããã®æ¬¡æ°ã®æ¯å¹ ã解æããŸãã
UFF58b ããŒã¿ãã¡ã€ã«ãããŒããããµã³ãã«ãããžã§ã¯ããéãã«ã¯ã以äžã®æé ã«åŸã£ãŠãã ããã
1. ãã«ãâãµã³ãã«ãéãâé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ããéžæãã
Tutorialãã©ã«ããããã«ã¯ãªãã¯ããããã« SVA_Order
Analysis From UFF File (Tutorial).seprojãããã«ã¯ãªãã¯
ããŸãã
2. ãããžã§ã¯ãã My Engine.seprojãšããååã§ä»»æã®ãã£ã¬ã¯ã
ãªã«ä¿åããŸãã
ããŒã¿ãã¥ãŒã¿ãã«ã¿ã³ã¡ãŒã¿ãšé³å§ã® 2 ã€ã®ä¿¡å·ã衚瀺ãããŸããå·Š
äžé ã®ã¿ã³ã¡ãŒã¿ä¿¡å·ã¯ã©ã³ããããã¹ãäžã«ã¿ã³ã¡ãŒã¿é»å§ãã©ã®ãã
ã«å€åãããã瀺ããŸããå³äžé ã®é³å§ä¿¡å·ã¯åãã¹ãäžã«é³å§ïŒã€ãŸã
éšé³ïŒãã©ã®ããã«å€åãããã瀺ããŸãã
ãšã³ãžã³ã®å転é床ã枬å®ãã次æ°è§£æã§ã¯ DUT ã®å転é床ã枬å®ããããã次æ°è§£æãå®è¡ããã«ã
ãã£ãŠã®æåã®æé ã¯åžžã« DUT ã®é床ãããã¡ã€ã«ã®æž¬å®ãšãªããŸãã
é床ãããã¡ã€ã«ã¯ DUT ã®å転é床ãæéãšãšãã«å€åããæ§åã瀺ã
ãŸããå転é床ã¯å転 / åïŒRPMïŒåäœã§æž¬å®ããŸãã
DUT ã®é床ãããã¡ã€ã«ãéé²ããã«ã¯ãã¢ããã°ã¿ã³ã¡ãŒã¿åŠçãã¹
ãããã䜿çšããŸãããã®ã¹ãããã§ã¯ã¿ã³ã¡ãŒã¿ãã«ã¹ä¿¡å·ã RPM ã«
å€æããŸãããšã³ãžã³ã®é床ãããã¡ã€ã«ã枬å®ããã«ã¯ã以äžã®æé ã«
åŸã£ãŠãã ããã
1. ã¹ããããè¿œå ãã¿ã³ãã¯ãªãã¯ãã解æâ次æ°é å枬å®âã¢ããã°
ã¿ã³ã¡ãŒã¿åŠçãéžæããŸãã
2. ã¹ãããèšå®ã¿ãã®å ¥åã¿ãå ã§ãã¿ã³ã¡ãŒã¿ä¿¡å·ã解æãœãŒã¹ïŒã€
ãŸãå ¥åä¿¡å·ïŒã«ãªã£ãŠããããšã«æ³šç®ããŠãã ããããã®ä¿¡å·ã¯ã¹
ãããèšå®ã¿ãã®äžéšã«ããã¿ã³ã¡ãŒã¿ä¿¡å·ã°ã©ãã«è¡šç€ºãããŸãã
ã¹ãããèšå®ã¿ãã®äžééšã«ã¯é床ãããã¡ã€ã«ã°ã©ãã衚瀺ãããŸ
ãããã®ã°ã©ãã¯ã¿ã³ã¡ãŒã¿ä¿¡å·ã§ãã®ã¹ããããçæãã RPM 枬
å®å€ãè¡šããŸãããã®æž¬å®å€ã¯ã©ã³ããããã¹ãäžã«ãšã³ãžã³ RPMãå€åããæ§åã瀺ããŠããŸãã
第 6 ç« æ¬¡æ°è§£æãè¡ã
© National Instruments 6-3 NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã
å³ 6-1 ã«ãã¹ãããèšå®ã¿ãã®é床ãããã¡ã€ã«ã°ã©ãã瀺ããŸãã
å³ 6-1 ãšã³ãžã³ã®é床ãããã¡ã€ã«
å³ 6-1 ã§ããšã³ãžã³ RPM ã 10.84 ç§éã« 1300 RPM ãã 5200 RPMã«å¢å€§ããŠããæ§åã«æ³šç®ããŠãã ããããã®é床ãããã¡ã€ã«ã次
æ°è§£æã®åºãšããŠäœ¿çšããŸãã
3. ãã¢ããã°ã¿ã³ã¡ãŒã¿åŠçãã¹ãããã®é床ãããã¡ã€ã«åºåãããŒ
ã¿ãã¥ãŒã¿ãã®ã°ã©ãã«ãã©ãã°ããŸãã
ãããšãããŒã¿ãã¥ãŒã¿ãã«ãæéã®é¢æ°ãšã㊠3 ã€ã®ã°ã©ããè¡š
瀺ãããŸããã¿ã³ã¡ãŒã¿ã°ã©ãã¯ã¿ã³ã¡ãŒã¿é»å§ã®æ¯å¹ ããé³å§ã°ã©
ãã¯é³å§ã¬ãã«ãããŸãé床ãããã¡ã€ã«ã°ã©ãã¯ãšã³ãžã³ RPM ã
衚瀺ããŸãã
4. ãããžã§ã¯ããä¿åããŸãã
䞻次æ°ãç¹å®ããé床ãããã¡ã€ã«ã®éé²ã«ã¿ã³ã¡ãŒã¿ä¿¡å·ã䜿çšããã®ã§ã次ã«ãã€ã¯ã
ãã³ä¿¡å·ã䜿çšããŠæ¬¡æ°è§£æãå®è¡ããŠã¿ãŸãããããã®è§£æã«å¯Ÿããäž
è¬çãªããŒã«ã¯ã«ã©ãŒããã圢åŒã®ã¹ãã¯ãã«ãããã§ããã«ã©ãŒããã
ã¯ä¿¡å·ã®æ¯å¹ ãåšæ³¢æ°ãŸãã¯æ¬¡æ°ã«é¢é£ããŠããæ§åãæŠç¥çã«ç€ºããŸ
ãããŸããã«ã©ãŒãããã¯ä¿¡å·ã®æ¯å¹ ãæéãŸãã¯é床ã«é¢é£ããŠããæ§
åã瀺ããŸãã
ãŸããã«ã©ãŒããããæ§æã㊠RPM ã®é¢æ°ãšããŠæ¬¡æ°ã衚瀺ããããšã
ã§ããŸããRPM- 次æ°ã«ã©ãŒãããã¯ããšã³ãžã³ RPM ã®å€åã«å¿ããŠç°
ãªã次æ°ã®é³å§ã¬ãã«ãå€åããæ§åã衚瀺ããŸããé«é³å§ã¬ãã«ã®åå
ãšãªã次æ°ã§ãã䞻次æ°ã®ç¹å®ã«ã«ã©ãŒãããã䜿çšã§ããŸãã
ã«ã©ãŒãããã衚瀺ããã«ã¯ãã¹ãã¯ãã«ããããã¹ãããã䜿çšããŸ
ããèšé²æžã¿ã®é³å§ã¬ãã«ã®ã«ã©ãŒãããã衚瀺ããã«ã¯ã以äžã®æé ã«
åŸã£ãŠãã ããã
1. ã¹ããããè¿œå ãã¿ã³ãã¯ãªãã¯ãã解æâ次æ°é å枬å®âã¢ããã°
ã¿ã³ã¡ãŒã¿åŠçãéžæããŸãã
2. ã¹ãããèšå®ã¿ãã®å ¥å / åºåã¿ãã§ãå ¥åä¿¡å·ãã«ããŠã³ã¡ãã¥ãŒ
ããæé波圢âé³å§ãéžæããŸãã
第 6 ç« æ¬¡æ°è§£æãè¡ã
NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã 6-4 ni.com
ã«ã©ãŒãããã¯ãæéã®é¢æ°ã§ããéšé³åšæ³¢æ°ã Hz åäœã§è¡šç€ºããŸ
ãããŸãããã®ã«ã©ãŒãããã§ã¯é³å§ã¬ãã«ã®æ¯å¹ ãè²ã§è¡šç€ºããŠã
ãŸããé³å§ãäœãå Žåã¯ç·ãé«ãå Žåã¯èµ€ã§ç€ºãããŸãã
å³ 6-2 ã«ç€ºãããã«ã600 ïœ 750 Hz ã®éã®æ°ãæã«é«ãé³å§ã¬ãã«ã
çŸãããŠããããšã«æ³šç®ããŠãã ããããã®åšæ³¢æ°ç¯å²ã¯ããããã
ãªãããã®ãšã³ãžã³éšåã®å ±æ¯åšæ³¢æ°ãå«ãã§ãããšèããããŸãã
å³ 6-2 åšæ³¢æ° - æéã«ã©ãŒããã衚瀺
3. ã¹ãããèšå®ã¿ãã®æ§æã¿ããã¯ãªãã¯ããããããã¿ã€ããã«ããŠ
ã³ã¡ãã¥ãŒãã RPM- 次æ°ãéžæããŸãã
æåã®ãã¡ã¯ãã«ã©ãŒãããã«ã¯é³å§ã¬ãã«ã®æ¯å¹ ãè²ã§è¡šç€ºãããŸ
ããããããæéã®å€åã«å¿ããåšæ³¢æ°ã®å€åã®ä»£ããã« RPM ã®å€
åã«å¿ãã次æ°ã®å€åãã«ã©ãŒãããã«è¡šç€ºãããããã«ãªããŸãã
4. X 軞ã®æ倧å€ãããã«ã¯ãªãã¯ããæ°èŠã®å€ãšã㊠20ãšå ¥åããŸ
ãããã®æäœã«ããã0 ïœ 20 ã®æ¬¡æ°ç¯å²ã®ã¿ã衚瀺ãããããã«ã«
ã©ãŒãããã®ã¹ã±ãŒã«ã調æŽãããŸãã
第 6 ç« æ¬¡æ°è§£æãè¡ã
© National Instruments 6-5 NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã
ã«ã©ãŒããã衚瀺ã¯å³ 6-3 ã®ããã«è¡šç€ºãããŸãã
å³ 6-3 RPM- 次æ°ã«ã©ãŒããã衚瀺
å³ 6-3 ã§ã2 次㯠2500 ïœ 3500 RPM ããã³ 4100 ïœ 5100 RPM ã«èµ€
ãé åãå«ãã§ããããšã«æ³šç®ããŠãã ããããããã®èµ€ãé åã¯ã
ãã® RPM ç¯å²ã§é«ãé³å§ã¬ãã«ãçºçããŠããããšã瀺ããŸãã
4 æ°çãšã³ãžã³ã§ã¯ 1 å転ã«ã€ã 2 åççŒãããã®åºŠã«éšé³ãçºçã
ãã®ã§ã2 次ã«ããããã®ãããªæåã¯äºæããããã®ã§ãã
ãŸããç°ãªã RPM ç¯å²ã§ã8 次ã10 次ã12 次ãããã³ 14 次ã«ãã
ãŠé«ãé³å§ã¬ãã«ãååšããŠããããšã«æ³šç®ããŠãã ããããããã®
次æ°ã¯ 2 次ã®åé³ïŒã€ãŸãåæ°ïŒã§ãããããã£ãŠããããã®æ¬¡æ°
ãé¡èãªéšé³ã®åå ãšãªããšèããããŸãããããã®æ¬¡æ°ãã¹ãŠãäž»
次æ°ãšãªããŸãã
5. ãã¹ãã¯ãã«ããããã¹ãããã®ã«ã©ãŒãããåºåãããŒã¿ãã¥ãŒã¿
ãã«ãã©ãã°ããŸãããã®ã¿ãã« 4 ã€ã®ã°ã©ãã衚瀺ãããŸãã
6. æ倧å€ãã¯ãªãã¯ããŠçŸåšã®å€ã 20ã«çœ®ãæããããŒã¿ãã¥ãŒã¿ã
ã®ã«ã©ãŒããã衚瀺ã«å¯ŸããŠæé 4 ãç¹°ãè¿ããŸãã
7. ãããžã§ã¯ããä¿åããŸãã
ç¹å®ã®æ¬¡æ°ã®æ¯å¹ ããã©ããã³ã°ãã次æ°è§£æãå®è¡ããã«ããã£ãŠã®æåŸã®æé 㯠2 ã€ã®äž»æ¬¡æ°ã®æ¯å¹ ã®ã
ã©ããã³ã°ã§ãããã®ããŒã¿ã¯ã«ã©ãŒãããã«æ¢ã«è¡šç€ºãããŠããŸããã
ã ããã«ã©ãŒãããã§ã¯ãã®ããŒã¿ã®å€§ãŸããªæŠèŠã衚瀺ãããŠããã ã
ã§ããæ¯å¹ ã°ã©ãã«ãããç¹å®ã®æ¬¡æ°ã«å¯Ÿããæ°å€ããŒã¿ã®ãã©ããã³ã°
ãç°¡åã«è¡ããŸãã
第 6 ç« æ¬¡æ°è§£æãè¡ã
NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã 6-6 ni.com
1 ã€ãŸãã¯è€æ°ã®æ¬¡æ°ã RPM ã®é¢æ°ãšããŠè¡šç€ºãããã«ã¯ãã次æ°ã
ã©ããã³ã°ãã¹ãããã䜿çšããŸããRPM ç¯å²å šäœã«ããã 10 次ãã
ã³ 12 次ã«ãããæ¯å¹ ã衚瀺ããã«ã¯ã以äžã®æé ã«åŸã£ãŠãã ããã
1. ã¹ããããè¿œå ãã¿ã³ãã¯ãªãã¯ãã解æâ次æ°é å枬å®â次æ°ã
ã©ããã³ã°ãéžæããŸãã
2. ã¹ãããèšå®ã¿ãã®å ¥å / åºåã¿ãã§ãå ¥åä¿¡å·ãã«ããŠã³ã¡ãã¥ãŒ
ããæé波圢âé³å§ãéžæããŸãã
3. ã¹ãããèšå®ã¿ãã®æ§æã¿ããã¯ãªãã¯ãããã©ããã³ã°ãã次æ°ãª
ã¹ãã« 10ããã³ 12ãšå ¥åããŸãã
4. X 軞ã¹ã±ãŒã«ãã«ããŠã³ã¡ãã¥ãŒããé床ãéžæããX 軞ã RPM ã«
å€æŽããŸãã
ãã³ã ã次æ°ãã©ããã³ã°ãã¹ãããã®ã¹ãããèšå®ã¿ãã§æ§æãè¡ããšã次æ°æ¯å¹ ã
æéãé床ããŸãã¯å転æ°ã®é¢æ°ãšããŠè¡šç€ºã§ããŸãã
5. ã次æ°ãã©ããã³ã°ãã¹ãããã®æ¯å¹ åºåãããŒã¿ãã¥ãŒã¿ãã«ã
ã©ãã°ããŸãã
6. æ°èŠã®æ¯å¹ ã°ã©ããå³ã¯ãªãã¯ãã衚瀺é ç®âå¡äŸãéžæããŸãã
æ¯å¹ ã°ã©ãã¯å³ 6-4 ã®ããã«è¡šç€ºãããŸãã
å³ 6-4 10 次ããã³ 12 次ã®æ¯å¹ ãããã
å³ 6-4 ã§ã12 次ã«å¯ŸããŠé«ãæ¯å¹ ã 3400RPM è¿èŸºã«çŸãããŠãã
ããšã«æ³šç®ããŠãã ããããã®ããŒã¯ã¯ã«ã©ãŒãããäžã§ 12 次ã«å¯Ÿ
ããèµ€ãé åã«çžåœããŸããæ¯å¹ ã°ã©ãã«ãããã«ã©ãŒãããããç°¡
åã«èªã¿åãã圢åŒã§ãã®æ å ±ã衚瀺ãããŸãã10 次ã«å¯ŸããããŒ
ã¯ã¯ 4200 RPM è¿èŸºã«çŸãããŠããŸããããã§ããã«ã©ãŒãããã§
ãã®è§£æã確èªã§ããŸãã
7. ãããžã§ã¯ããä¿åããŠéããŸãã
第 6 ç« æ¬¡æ°è§£æãè¡ã
© National Instruments 6-7 NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã
ããŒã¿éé²äžã«æ¬¡æ°è§£æãè¡ã
é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã䜿çšãããšãNI DSA ããã€ã¹ãªã©ã® NI DAQ ããŒããŠã§ã¢ããããŒã¿éé²äžã«æ¬¡æ°è§£æãè¡ããŸãã
ãã®ã»ã¯ã·ã§ã³ã§ã¯ããã°ããŒã¿ãå«ããã¡ã€ã«ãåçããŸããããŒã¿ã®
åçäžã«ãã®ããŒã¿ã«å¯ŸããŠæ¬¡æ°è§£æãè¡ãããšãã§ããŸãããã®ãã
ã«ã以äžã®ãµã³ãã«ã§ã¯ãã¹ãããŒã¿ã®æ¬¡æ°è§£æãšåæã«å®è¡ããããŒã¿
éé²ã®ã·ãã¥ã¬ãŒã·ã§ã³ãè¡ããŸãããã°ããŒã¿åçã®è©³çŽ°ã«ã€ããŠã¯ã
第 4 ç« ãé³é¿ãæ¯åã®æž¬å®ãè¡ããã®ãããŒã¿ãåçãããã®ã»ã¯ã·ã§
ã³ãåç §ããŠãã ããã
ãã®ã»ã¯ã·ã§ã³ã®ãµã³ãã«ã§ã¯ãDUT 㯠7 æçŸœæ ¹ã® PC ãã¡ã³ã§ããã®
ãã¡ã³ã¯ 4 ã³ã€ã«ã¢ãŒã¿ã«åãä»ããããŠããŸããDSA ããã€ã¹ã¯ãã®
ãã¡ã³ãã以äžã® 2 ã€ã®ä¿¡å·ã枬å®ããŸãã
⢠ãã¡ã³ã®å転é床ã枬å®ããã¿ã³ã¡ãŒã¿ãã«ã¹ä¿¡å·ãã¿ã³ã¡ãŒã¿ã¯
ãã¡ã³ã 1 å転ããããšã«ãã«ã¹ã 2 ã€çæããŸãã
⢠ãã¡ã³ã®æ¯åã枬å®ããå é床èšä¿¡å·ã
ãããã®ä¿¡å·ã«å¯ŸããŠæ¬¡æ°è§£æãè¡ãããã¡ã³ã«ããé¡èãªæ¯åã®åå ãš
ãªã次æ°ãç¹å®ããŸãã
以äžã®ã»ã¯ã·ã§ã³ã§ã¯ã次ã®æé ã«åŸã£ãŠãã ããã
1. éé²ãããããŒã¿ãé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã®ãã°ãã¡ã€ã«ãã
ã€ã³ããŒããåçããŠããªã¢ã«ã¿ã€ã ããŒã¿éé²ã®ã·ãã¥ã¬ãŒã·ã§ã³
ãè¡ããŸãã
2. ãã¹ãå®æœäžããã¡ã³ã®å転é床ã枬å®ããŸãã
3. é¡èãªãã¡ã³æ¯åã®åå ãšãªã次æ°ãç¹å®ããŸãã
4. ç¹å®ãããããã®æ¬¡æ°ã®æ¯å¹ ã解æããŸãã
ãã°ããŒã¿ãã€ã³ããŒãããPC ãã¡ã³ã®ã©ã³ããããã¹ãããããŒã¿ãéãã«ã¯ã以äžã®æé ã«åŸã£
ãŠãã ããã
1. æ°èŠãããžã§ã¯ããéããŸãã
2. ãã¡ã€ã«âã€ã³ããŒãâ LabVIEW SignalExpress TDMS ã®ãã°ä¿¡å·
ãéžæããNational InstrumentsÂ¥Sound and Vibration
AssistantÂ¥ExamplesÂ¥Log File - PC Fan Run-upÂ¥ãã£ã¬ã¯ããª
ãåç §ããŠãçŸåšã®ãã©ã«ããã¿ã³ãã¯ãªãã¯ããŸãã
3. ãããžã§ã¯ãã My PC Fan.seprojãšããååã§ä»»æã®ãã£ã¬ã¯ã
ãªã«ä¿åããŸãã
第 6 ç« æ¬¡æ°è§£æãè¡ã
NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã 6-8 ni.com
4. ãã°ããŒã¿ãŠã£ã³ããŠã®ãã°ã®äžã« PC Fan Run-up ã衚瀺ãããŸ
ãã
5. äœæ¥ãšãªã¢ãã«ããŠã³ã¡ãã¥ãŒããåçã¢ãŒãâ PC Fan Run-up ã
éžæããŸããããŒã¿ãã¥ãŒã¿ãã§ãã¿ãã®äžéšã«æé軞ã¹ã©ã€ããè¡š
瀺ãããŸãã
PC ãã¡ã³ã®å転é床ã枬å®ãããã®ç« ã®ããã°ããŒã¿ã«å¯ŸããŠæ¬¡æ°è§£æãè¡ããã®ã»ã¯ã·ã§ã³ã§èª¬æãã
ããã«ã次æ°è§£æãè¡ãåã«ã¯ DUT ã®é床ãããã¡ã€ã«ã®æž¬å®ãåžžã«å¿
èŠãšãªããŸããPC ãã¡ã³ã®é床ãããã¡ã€ã«ã枬å®ããã«ã¯ã以äžã®æ
é ã«åŸã£ãŠãã ããã
1. ãã¢ããã°ã¿ã³ã¡ãŒã¿åŠçãã¹ããããè¿œå ããŸãã
2. ã¹ãããèšå®ã¿ãã®å ¥åã¿ãå ã§ãããã©ã«ãã®å ¥åä¿¡å·ãã¢ããã°
ã¿ã³ã¡ãŒã¿ä¿¡å·ã«ãªã£ãŠããããšã«æ³šç®ããŠãã ããã
3. ãã®ãã¡ã³ã§ã¯ 2 ã€ã®ãã«ã¹ã§ 1 å転ãè¡šãã®ã§ãã¹ãããèšå®ã¿
ãã®æ§æã¿ããã¯ãªãã¯ãããã«ã¹æ°ãã£ãŒã«ãã« 2ãšå ¥åããŸãã
ãããšãé床ãããã¡ã€ã«ã°ã©ãã® Y 軞äžã® RPM å€ãåæžããããš
ã«æ³šç®ããŠãã ããã
4. ãã¢ããã°ã¿ã³ã¡ãŒã¿åŠçãã¹ãããã®é床ãããã¡ã€ã«åºåãããŒ
ã¿ãã¥ãŒã¿ãã«ãã©ãã°ããŸãã
5. ãããžã§ã¯ããå®è¡ããŠããŒã¿ãåçããŸãã
é床ãããã¡ã€ã«ã°ã©ããæé軞ã¹ã©ã€ãã®åãã«åãããŠæŽæ°ãã
ãããšã«æ³šç®ããŠãã ããããã®åãã¯ãDUT ããã¿ã³ã¡ãŒã¿ä¿¡å·
ãéé²ããªããé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãããã¡ã³ã®é床ãèšç®ã
ãŠããæ§åã瀺ããŸãã
ãã³ã åçäœæ¥ãšãªã¢ã衚瀺ãããŠããéã¯ãæäœâåçã¬ãŒããèšå®ãéžæããŠ
ããŒã¿ã®é床ãå€æŽã§ããŸãã
6. ãããžã§ã¯ããä¿åããŸãã
䞻次æ°ãç¹å®ãããã®ç« ã®ããã°ããŒã¿ã«å¯ŸããŠæ¬¡æ°è§£æãè¡ããã®ã»ã¯ã·ã§ã³ã§èª¬æãã
ããã«ã次ã®æé ã§ã¯äž»æ¬¡æ°ãç¹å®ããŸãããã ããã«ã©ãŒãããã䜿çš
ãã代ããã«ã以äžã®æé ã§ã¯æ¬¡æ°ã¹ãã¯ãã«ã䜿çšããŠãããã®æ¬¡æ°ã
ç¹å®ããŸããã«ã©ãŒããããæ°å€ããŒã¿ã®å€§ãŸããªæŠèŠãææ¡ããã®ã«é©
ããŠããã®ã«å¯Ÿãã次æ°ã¹ãã¯ãã«ã°ã©ãã¯æ°å€ããŒã¿ã衚瀺ããã®ã«é©
ããŠããŸãã
第 6 ç« æ¬¡æ°è§£æãè¡ã
© National Instruments 6-9 NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã
ã次æ°ã¹ãã¯ãã«ãã¹ãããã䜿çšããŠã次æ°ã¹ãã¯ãã«ã°ã©ãã衚瀺ã
ãŸããæ¯åä¿¡å·ã®æ¬¡æ°ã¹ãã¯ãã«ã枬å®ããã«ã¯ã以äžã®æé ã«åŸã£ãŠã
ã ããã
1. ã¹ããããè¿œå ãã¿ã³ãã¯ãªãã¯ãã解æâ次æ°é å枬å®â次æ°ã¹ã
ã¯ãã«ãéžæããŸãã
2. ã¹ãããèšå®ã¿ãã®å ¥åã¿ãã§ãå ¥åä¿¡å·ãã«ããŠã³ã¡ãã¥ãŒããã¢
ã¯ãã£ããã°â Acquired Signals â Vibration Signal ãéžæããŸ
ãã次æ°ã¹ãã¯ãã«ã°ã©ããæŽæ°ãããã©ã³ããããã¹ã 1 ååã®
å®è¡äžã®æ¯åæ¯å¹ ã次æ°ã®é¢æ°ãšããŠè¡šç€ºããŸãã
3. ã¹ãããèšå®ã¿ãã®ã¹ãã¯ãã«èšå®ã¿ããã¯ãªãã¯ããæ¯å¹ ã¹ã±ãŒã«
ãã«ããŠã³ã¡ãã¥ãŒããç·åœ¢ãéžæããŠããã·ãã«åäœããèªã¿åã
ãããç·åœ¢ã¹ã±ãŒã«ã«èšå®ããŸãã
次æ°ã¹ãã¯ãã«ã°ã©ãã§ãé¡èãªæ¯åã 10 次以éã«çŸãããŠããªã
ããšã«æ³šç®ããŠãã ãããã°ã©ãã®ã¹ã±ãŒã«ãå€æŽããŠãæ¯åã®åå
ãšãªã次æ°ã衚瀺ã§ããŸãã
4. æ倧次æ°ãã£ãŒã«ãã®æ°èŠã®å€ãšã㊠16ãšå ¥åããŸãã
次æ°ã¹ãã¯ãã«ã°ã©ãã¯å³ 6-5 ã®ããã«è¡šç€ºãããŸãã
å³ 6-5 æ¯åä¿¡å·ã®æ¬¡æ°ã¹ãã¯ãã«ã°ã©ã
5. ã次æ°ã¹ãã¯ãã«ãã¹ãããã®æ¬¡æ°ã¹ãã¯ãã«åºåãããŒã¿ãã¥ãŒã¿
ãã«ãã©ãã°ããŸãããããšããã®ã¿ãã«ã¯é床ãããã¡ã€ã«ã°ã©ã
ãšæ¬¡æ°ã¹ãã¯ãã«ã°ã©ãã衚瀺ãããŸãã
6. ãããžã§ã¯ããå®è¡ããŸãã次æ°ã¹ãã¯ãã«ã°ã©ãããããžã§ã¯ãã
å®è¡ãããã«ã€ããŠæŽæ°ãããããšã«æ³šç®ããŠãã ããã
7. 次æ°ã¹ãã¯ãã«ã°ã©ããå³ã¯ãªãã¯ããŠã衚瀺ãªãã·ã§ã³âæ»åã°ã©
ããéžæããŸããæ»åã°ã©ãã¯ãã¹ããéããŠã®ããŒã¿ã®å±¥æŽã衚瀺
ããã®ã«é©ããŠããŸãã
第 6 ç« æ¬¡æ°è§£æãè¡ã
NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã 6-10 ni.com
8. ãããžã§ã¯ããå®è¡ããŸããæ»åã°ã©ããã©ã³ããããã¹ãäžã«æ¬¡æ°
ã¹ãã¯ãã«ã®å±¥æŽã衚瀺ããŠããããšã«æ³šç®ããŠãã ãããZ 軞ã¯ã
ã¹ãåæ°ãè¡šããŠããŸãã
å³ 6-6 ã«æ»åã°ã©ãã瀺ããŸãã
å³ 6-6 次æ°ã¹ãã¯ãã«ã®æ»åã°ã©ã
æ»åã°ã©ã㯠4 次ããã³ 7 次ã«ãããããŒã¯ã瀺ããŠããŸãããã
ãã®æ¬¡æ°ã䞻次æ°ã§ããPC ãã¡ã³ã¢ãŒã¿ã® 4 ã€ã®ã³ã€ã«ã 4 次ã®æ¯
åãããŸã 7 æã®çŸœæ ¹ã 7 次ã®æ¯åãçºçããŠããã®ã§ããã®ãã
ãªæåã¯äºæããããã®ã§ãã
9. ãããžã§ã¯ããä¿åããŸãã
ç¹å®ã®æ¬¡æ°ã®æ¯å¹ ããã©ããã³ã°ããæåŸã®æé ã§ã¯ããããã®æ¬¡æ°ã§ã®æ¯åã®æ¯å¹ ããã©ããã³ã°ããŸãã4次ããã³ 7 次ã§ã®æ¯åæ¯å¹ ã衚瀺ããã«ã¯ã以äžã®æé ã«åŸã£ãŠãã ã
ãã
1. 次æ°ãã©ããã³ã°ã¹ããããè¿œå ããŸãã
2. ã¹ãããèšå®ã¿ãã®å ¥å / åºåã¿ãã§ãå ¥åä¿¡å·ãã«ããŠã³ã¡ãã¥ãŒ
ããã¢ã¯ãã£ããã°â Acquired Signals â Vibration Signal ãéž
æããŸãã
3. ã¹ãããèšå®ã¿ãã®æ§æã¿ããã¯ãªãã¯ãããã©ããã³ã°ãã次æ°ãª
ã¹ãã« 4ããã³ 7ãšå ¥åããŠãPC ãã¡ã³ã® 4 ã€ã®ã³ã€ã«ãš 7 æã®çŸœ
æ ¹ã®æ¯åã«çžåœãã次æ°ããã©ããã³ã°ããŸãã
4. 垯åå¹ ïŒæ¬¡æ°ïŒãã£ãŒã«ãã« 0.5ãšããŸãæéééïŒsïŒãã£ãŒã«ã
ã« 500mãšå ¥åããŸãã
5. ã次æ°ãã©ããã³ã°ãã¹ãããã®æ¯å¹ åºåãããŒã¿ãã¥ãŒã¿ãã«ã
ã©ãã°ããŸãããããšãããŒã¿ãã¥ãŒã¿ãã«ãé床ãããã¡ã€ã«ã次
æ°ã¹ãã¯ãã«ã®æ»åã°ã©ãã4 次㚠7 次ã§ã®æ¯åã®æ¯å¹ ã°ã©ãã® 3 ã€
ã®ã°ã©ãã衚瀺ãããŸãã
第 6 ç« æ¬¡æ°è§£æãè¡ã
© National Instruments 6-11 NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã
6. æ¯å¹ ã°ã©ããå³ã¯ãªãã¯ãã衚瀺é ç®âå¡äŸãéžæããŸãã
7. ãããžã§ã¯ããå®è¡ããŸããå³ 6-7 ã«ç€ºãããã«ãæ¯å¹ ã°ã©ãã¯ã©ã³
ããããã¹ãäžã® 4 次ããã³ 7 次ã§ã®æ¯åã®æ¯å¹ ã衚瀺ããŸãã
å³ 6-7 4 次ããã³ 7 次ã®æ¬¡æ°æ¯å¹ ãããã
ãããã®ããŒã¯ã¯æ»åã°ã©ãã§ã®ããŒã¯ãšåãã§ãã4 次ã§ã®æ¯åã®
æ倧ããŒã¯ã¯ 13 ç§ã®æç¹ã«ããŸã 7 次ã§ã®æ¯åã®æ倧ããŒã¯ã¯ 8 ç§
ã®æç¹ã«çŸããŠããŸãã
8. ãããžã§ã¯ããä¿åããŠéããŸãã
© National Instruments 7-1 NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã
7é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã®ãããžã§ã¯ãã LabVIEW ã§äœ¿çšãã
é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã§ã¯ãä¿¡å·ã®éé²ãçæã解æããŸãã¯ãã°
ãªã©ã®æšæºã¹ãããã䜿çšããèªåèšæž¬ãå®çŸ©ããããšãã§ããŸãã以äž
ã®æ¹æ³ã«ãã LabVIEW ã°ã©ãã£ã«ã«ããã°ã©ãã³ã°ã§é³é¿ / æ¯åèšæž¬ã¢
ã·ã¹ã¿ã³ãã®æ©èœãæ¡åŒµã§ããŸãã
⢠LabVIEW 㧠VI ãäœæãããã® VI ãé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã«
ã€ã³ããŒãããŠæž¬å®ãªãã·ã§ã³ãæ¡åŒµããã«ã¹ã¿ã ã¹ãããé¢æ°ãäœ
æããŸãã
⢠é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã®ãããžã§ã¯ãã LabVIEW ãããã¯ã
ã€ã¢ã°ã©ã ã«å€æããLabVIEW ã§åŒãç¶ãã¢ããªã±ãŒã·ã§ã³éçºã
è¡ãããšãã§ããŸãããã®ãããªæ¹æ³ã§ãé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³
ãã®å¯Ÿè©±åŒããŒã«ã«ãã枬å®ãè¡ãé çªãå®çŸ©ãããã®äžé£ã®é çªã
LabVIEW ãããã¯ãã€ã¢ã°ã©ã ã«å€æããŠèªåãã¹ãã¢ããªã±ãŒ
ã·ã§ã³ãæ§ç¯ããŸãã
ã¡ã¢ SignalExpress ã®ãªãã¡ã¬ã³ã¹ãäŸã¯é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã«ãåæ§ã«é©
çšãããŸãã
LabVIEW VI ãã¹ããããšããŠé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã«ã€ã³ããŒããã
ã«ã¹ã¿ã LabVIEW VI ãåŒã³åºãã«ã¯ãé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã®
LabVIEW VI ãå®è¡ã¹ãããã䜿çšããŸããããšãã°ãé³é¿ / æ¯åèšæž¬ã¢
ã·ã¹ã¿ã³ããã LabVIEW VI ãåŒã³åºããšã以äžã®ããšãè¡ãããšãã§
ããŸãã
⢠å€æ§ãªãã¡ã€ã«åœ¢åŒã®ããŒã¿ãèªã¿æžããã
⢠VI ã䜿çšããŠããŒã¿ãããã«è§£æãã
⢠枬å®ã¢ã«ãŽãªãºã ãå®çŸ©ãã
LabVIEW VI ãå®è¡ã¹ããã㧠LabVIEW ãã VI ãã€ã³ããŒãããã«ã¯ã
以äžã®æé ã«åŸã£ãŠãã ããã
第 7 ç« é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã®ãããžã§ã¯ãã LabVIEW ã§äœ¿çšãã
NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã 7-2 ni.com
ã¡ã¢ ãã®ã»ã¯ã·ã§ã³ã®æŒç¿åé¡ã«ã¯ãLabVIEW 8.6 以éã®ããŒãžã§ã³ãå¿ èŠã§ãã
1. ãã«ãâãµã³ãã«ãéãâ SignalExpress ãéžæããTutorial
ãã©ã«ããããã«ã¯ãªãã¯ããŠéããUser Step.seprojãããã«
ã¯ãªãã¯ããŸãã
ãã®ãããžã§ã¯ãã§ã¯ãã¢ããã°ä¿¡å·ãäœæã¹ãããã§ä¿¡å·ãçæã
ãŸãã
2. ãããžã§ã¯ãã My Imported VI.seprojãšããååã§ä»»æã®ãã£
ã¬ã¯ããªã«ä¿åããŸãã
3. ã¹ããããè¿œå ãã¿ã³ãã¯ãªãã¯ããŠãLabVIEW VI ãå®è¡â
LabVIEW 8.6 VI ãå®è¡ãéžæããŸãã
ãã®æŒç¿ã§å®è¡ãã VI 㯠LabVIEW 8.6 ã§ä¿åãããŠããŸããVI ãä¿åãã LabVIEW ã®ããŒãžã§ã³ã«äžèŽãã LabVIEW VI ãå®è¡ã¹
ãããã®ããŒãžã§ã³ã䜿çšããå¿ èŠããããŸãã
4. ã¹ãããèšå®ã¿ã㧠VI ãéžæã»ã¯ã·ã§ã³ã®åç §ãã¿ã³ãã¯ãªãã¯ãã
SignalExpressÂ¥ExamplesÂ¥Tutorialãã£ã¬ã¯ããªã«ãã
Limiter-LV86.viãéžæããŸãã
ã¡ã¢ ãäžæ㪠LabVIEW ããŒãžã§ã³ã§ããã®ãšã©ãŒã衚瀺ãããå Žåã¯ã
LabVIEW 8.6 ã§ãã® VI ãéããŠä¿åããŠãã ãããããã«ãããVI ãLabVIEW ã®æ£ããããŒãžã§ã³ã§ä¿åãããé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã®
LabVIEW VI ãå®è¡ã¹ãããã§å®è¡ãããŸããã¹ãããèšå®ã¿ãã® VI ãæŽæ°ã
ã¿ã³ãã¯ãªãã¯ããŸãã
第 7 ç« é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã®ãããžã§ã¯ãã LabVIEW ã§äœ¿çšãã
© National Instruments 7-3 NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã
å³ 7-1ã®ãããªã¹ãããèšå®ã¿ãã衚瀺ãããããšã確èªããŠãã ã
ãã
å³ 7-1 Limiter VI ã®ã¹ãããèšå®ã¿ã
Limiter-LV86 VI ã¯æéé å波圢ãå ¥åãšããŠåãå ¥ããã¹ãããèš
å®ã¿ãã§æå®ããä¿¡å·ç¯å²ã®äžéãè¶ ããå€ãšäžéãäžåã£ãå€ãã¯
ãªããããã¯ãªãããã波圢ãåºåä¿¡å·ãšããŠè¿ããŸãã
LabVIEW VI ãã€ã³ããŒãããéãé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã§
ã¯ãVI ã®å ¥åã¯å ¥åä¿¡å·ãŸãã¯ãã©ã¡ãŒã¿ãšããŠããã㊠VI ã®åºå
ã¯åºåä¿¡å·ãšããŠå²ãåœãŠãããŸãã
å ¥åä¿¡å·ã¯ããããžã§ã¯ããã¥ãŒã§ã¹ãããã®å ¥åãšããŠè¡šç€ºããã
ãããä¿¡å·ãå ¥åãšã㊠VI ã«æž¡ãããšãã§ããŸãããã©ã¡ãŒã¿ã¯ã
ã¹ãããã®ã¹ãããèšå®ã¿ãã§æ§æã§ããå€ã§ãããã¹ã€ãŒããã¹
ãããã䜿çšãããšãåçã«ãã©ã¡ãŒã¿ãã¹ã€ãŒãããããšãã§ããŸ
ãããã®ãããžã§ã¯ãã§ã¯ãVI ã«å ¥åä¿¡å·ã® Time waveform inãããã³ã¹ã«ã©ãã©ã¡ãŒã¿ã® Upper limit ãš Lower limit ããããŸãã
5. ã¹ãããèšå®ã¿ãã® VI ãæ§æã¿ããã¯ãªãã¯ã㊠VI ã衚瀺ããŸãã
6. Upper limit ãã£ãŒã«ãã« 0.5ããLower limit ãã£ãŒã«ãã« -0.9
ãå ¥åããŠå¶éå€ãæå®ããŸãã
7. LabVIEW VI ãå®è¡ã¹ãããã® Clipped waveform out ãããŒã¿
ãã¥ãŒã¿ãã«ãã©ãã°ããŸãã
第 7 ç« é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã®ãããžã§ã¯ãã LabVIEW ã§äœ¿çšãã
NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã 7-4 ni.com
8. ãã¢ããã°ä¿¡å·ãäœæãã¹ãããã®æ£åŒŠæ³¢åºåãããŒã¿ãã¥ãŒã¿ãã«
ãã©ãã°ããŸããå³ 7-2 ã«ç€ºãããã«ãããŒã¿ãã¥ãŒã¿ãã®ã°ã©ã
ã¯ãå®å šãªæ£åŒŠæ³¢ãšã¯ãªããããã波圢㮠2 ã€ã®ä¿¡å·ã衚瀺ããŸãã
å³ 7-2 ã¯ãªããããã波圢ãšæ£åŒŠæ³¢
9. ãããžã§ã¯ããä¿åããŠéããŸãã
é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã®ããŒããŠã§ã¢ãšã®äœµçšãããã³é³é¿ / æ¯å
èšæž¬ã¢ã·ã¹ã¿ã³ã察å¿ã® VI ã®éçºã®è©³çŽ°ã«ã€ããŠã¯ããã«ãâé³é¿ / æ¯
åèšæž¬ã¢ã·ã¹ã¿ã³ããã«ããéžæããŠãé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ããã«
ãããåç §ããŠãã ããã
第 7 ç« é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã®ãããžã§ã¯ãã LabVIEW ã§äœ¿çšãã
© National Instruments 7-5 NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã
é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã®ãããžã§ã¯ããLabVIEW ã®ãããã¯ãã€ã¢ã°ã©ã ã«å€æãã
é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã§ã¯ããããžã§ã¯ãã LabVIEW ã®ãããã¯
ãã€ã¢ã°ã©ã ã«å€æã§ããŸãã
ã¡ã¢ é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã®ãããžã§ã¯ãã LabVIEW VI ã«å€æããã«ã¯ã
LabVIEW éçºã·ã¹ãã ã®ããŒãžã§ã³ 8.2 以éãå¿ èŠã§ãããŸããã³ãŒããçæ
ãã LabVIEW ã®ããŒãžã§ã³ã« LabVIEW Support for Sound and Vibration Analysis ã³ã³ããŒãã³ããã€ã³ã¹ããŒã«ãããŠããå¿ èŠããããŸãã
é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã®ãããžã§ã¯ãã LabVIEW ã®ãããã¯ãã€
ã¢ã°ã©ã ã«å€æããã«ã¯ã以äžã®æé ã«åŸã£ãŠãã ããã
1. ãã«ãâãµã³ãã«ãéãâé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ããéžæãã
Tutorialãã©ã«ããããã«ã¯ãªãã¯ããããã« SVA_Analyze
Signal From UFF File (Tutorial).seprojãããã«ã¯ãªãã¯ã
ãŸãã
2. ããŒã«âã³ãŒãçæâ LabVIEW ãã€ã¢ã°ã©ã ãéžæããŸãã
ã¡ã¢ è€æ°ã® LabVIEW ã®ããŒãžã§ã³ãã€ã³ã¹ããŒã«ãããŠããå ŽåãæåŸã«èµ·åã
ã LabVIEW ã®ããŒãžã§ã³ã§ã³ãŒãçæãè¡ãããŸãã
3. æ°èŠã® VI ã My LabVIEW VI.viãšããååã§ä¿åããOK ãã¿ã³ã
ã¯ãªãã¯ããŸããé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã¯æ°èŠ VI ãçæãã
ååéããããŒãžã§ã³ã® LabVIEW ã§ãã® VI ãéããŸãã
LabVIEW ã®ãããã¯ãã€ã¢ã°ã©ã ã®çµæã¯ãé³é¿ / æ¯åèšæž¬ã¢ã·ã¹
ã¿ã³ãã®ãããžã§ã¯ãã®å®è¡ã«åæ ãããŸããLabVIEW ã®ãããã¯
ãã€ã¢ã°ã©ã ã¯ãé ç·ããã LabVIEW Express VI ããæ§æãããŸ
ããå Express VI ã¯ãé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã®ãããžã§ã¯ã
ã®åã¹ãããã«çžåœããŸããExpress VI ãããã«ã¯ãªãã¯ãããšã
é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã®ã¹ãããèšå®ã衚瀺ãããŸãããŸãã
Express VI ãå³ã¯ãªãã¯ããã·ã§ãŒãã«ããã¡ãã¥ãŒããããã³ã
ããã«ãéããéžæãããšãExpress VI ã LabVIEW ã®ãµã VI ã«å€
æã§ããŸããExpress VI ããµã VI ã«å€æããå Žåããã®ãµã VI ãExpress VI ã«æ»ãããšã¯ã§ããŸããããããã¯ãã€ã¢ã°ã©ã ã衚瀺
ããLabVIEW ã®ãããã¯ãã€ã¢ã°ã©ã ã§ã®å®è¡ç¶æ³ã®ç¢ºèªã VI ã®æ©èœå€æŽãè¡ãããšãã§ããŸãã
LabVIEW ã®ããã³ãããã«ã«ã¯ãæéæ³¢åœ¢å ¥åä¿¡å·ãã¹ãã¯ãã«å ¥
åä¿¡å·ãããã¯ãŒã¹ãã¯ãã«ãã¹ãããããã®å¹³ååå®äºããã³å¹³å
å®äºæ°ã®åºåããããŒã¯æ€çŽ¢ãã¹ãããããã®ããŒã¯ããã³ããŒã¯æ°
ã衚瀺ãããŸãã
第 7 ç« é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã®ãããžã§ã¯ãã LabVIEW ã§äœ¿çšãã
NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã 7-6 ni.com
4. ãããã¯ãã€ã¢ã°ã©ã ã§ãUFF58 Read FileãExpress VI ãããã«ã¯
ãªãã¯ããŸããUFF58 ãã¡ã€ã«ãã¹ãã£ãŒã«ãã® UFF58 ãã¡ã€ã«ã
Triangle wave (1000 Hz).uffã§ããããšã確èªããOK ãã¿ã³ã
ã¯ãªãã¯ããŸãã
5. å·Šã«ç€ºãå®è¡ãã¿ã³ãã¯ãªãã¯ããVI ãå®è¡ããŸããå³ 7-3 ã«ç€ºã
ããã«ãLabVIEW ã®ããã³ãããã«ã§ã¯ãVI ãåæ¢ãããŸã§æéæ³¢
圢ã°ã©ããšã¹ãã¯ãã«ã°ã©ããç¶ç¶çã«æŽæ°ãããŸãã
å³ 7-3 LabVIEW ã®ããã³ãããã«
6. å·Šã«ç€ºãåæ¢ãã¿ã³ãã¯ãªãã¯ã㊠VI ãåæ¢ããŸãã
7. VI ãä¿åããLabVIEW ãéããŸãã
ã¡ã¢ é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã®ãããžã§ã¯ãã«ãã°ãããã³ããŒããŠã§ã¢æ§æ
ãå«ãŸããŠããç¶æ ã§å€æããå Žåãé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã¯ãããžã§
ã¯ãã«å«ãŸãããã¹ãŠã®ã³ã³ãã³ãã 1 ã€ã® Express VI ãšããŠçæããŸããç
æããã Express VI ããµã VI ã«å€æããããšã¯ã§ããŸããã
© National Instruments 8-1 NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã
8詳现æ å ±
é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã«ã€ããŠã®è©³çŽ°ã¯ã以äžã®ãªãœãŒã¹ãåç §ã
ãŠãã ããã
é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã®ãµã³ãã«ãããžã§ã¯ã
é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã«ã¯ããã·ã§ãã«ã€ã³ã¹ãã«ã¡ã³ãã®ããŒã
ãŠã§ã¢ã䜵çšããå Žåã«äœ¿çšã§ãããµã³ãã«ãããžã§ã¯ãããããŸãããã
ãã®ãããžã§ã¯ãã«ã¢ã¯ã»ã¹ããã«ã¯ããã«ãâãµã³ãã«ãéãâé³é¿ /æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ããéžæããŸãããããã®ãµã³ãã«ãåç §ããããšã«
ãããããã«å€ãã®é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã®æ©èœã®ç¿åŸãããã³å¿ èŠ
ãšããã¢ããªã±ãŒã·ã§ã³ã«è¿ããããžã§ã¯ããéžæããäœæ¥ãéå§ããããš
ãã§ããŸãã
é³é¿ / æ¯åãµã³ãã«ã®è§£çãéã
ãã«ãâãµã³ãã«ãéãâé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ããéžæããŠããã®
ããã¥ã¢ã«ã®åç« ã«ããæŒç¿ã®è§£çãèŠãããšãã§ããŸããããšãã°ã
第 2 ç« ããããžã§ã¯ãã®æäœãã®è§£çã衚瀺ããã«ã¯ãSVA_Analyze
Signal from UFF File.seprojãéããŸãã
é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ããããŒããŠã§ã¢ãšäœµçšãã
é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã§ã¯ãããŸããŸãªãã·ã§ãã«ã€ã³ã¹ãã«ã¡ã³
ã補ã®ä¿¡å·éé² / çæããŒããŠã§ã¢ããµããŒããããŠããŸãããã·ã§ãã«
ã€ã³ã¹ãã«ã¡ã³ãã® MIO ããã€ã¹ããã€ãããã¯ä¿¡å·éé²ïŒDSAïŒãã
ã€ã¹ãé«éããžã¿ã€ã¶ãä»»æ波圢çºçåšãé¢æ°çºçåšããã€ã¹ã䜿çšã
ãŠãé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã§ã¢ããã°ä¿¡å·ã®éé²ããã®ã³ã°ãè¡ã
ããšãã§ãããŸããããã€ã¹éã§ã¯ããã¯ãå ±æãä¿¡å·ãããªã¬ããããš
ã§ãã·ã¹ãã ã§è€æ°ããã€ã¹ãåæããããšãã§ããŸããé³é¿ / æ¯åèšæž¬
ã¢ã·ã¹ã¿ã³ãã NI ããŒããŠã§ã¢ãšäœµçšããã«ã¯ãNI-DAQmx ããŒãžã§ã³
8.9.5 以éãã€ã³ã¹ããŒã«ãããŠããå¿ èŠããããŸãã
é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ãã®ããŒããŠã§ã¢ãšã®äœµçšã®è©³çŽ°ã«ã€ããŠã¯ã
ãã«ãâé³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ããã«ããéžæããŠãé³é¿ / æ¯åèšæž¬
ã¢ã·ã¹ã¿ã³ããã«ãããåç §ããŠãã ããã
第 8 ç« è©³çŽ°æ å ±
NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã 8-2 ni.com
ãŠã§ããªãœãŒã¹
ãã®ä»ã®ãªãœãŒã¹ã®è©³çŽ°ã«ã€ããŠã¯ãni.com/soundandvibrationã«
ã¢ã¯ã»ã¹ããŠããã·ã§ãã«ã€ã³ã¹ãã«ã¡ã³ãã®ãŠã§ããµã€ããåç §ããŠã
ã ããã
© National Instruments A-1 NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã
Aæè¡ãµããŒãããã³ãããã§ãã·ã§ãã«ãµãŒãã¹
ãã·ã§ãã«ã€ã³ã¹ãã«ã¡ã³ãã®ãŠã§ããµã€ãïŒni.comïŒã®ãŠãŒã¶ãã
ãã¡ã€ã«ã«ãã°ã€ã³ããŠãã客æ§åããµãŒãã¹çšã«ã«ã¹ã¿ãã€ãºãããã¢
ã¯ã»ã¹ããŒãžã衚瀺ããŸããæè¡ãµããŒãããã³ãã®ä»ã®ãµãŒãã¹ã«ã€ã
ãŠã¯ãåŒç€Ÿã®ãŠã§ããµã€ãïŒni.comïŒã®äžèšã®ã»ã¯ã·ã§ã³ãåç §ããŠã
ã ããã
⢠ãµããŒã â æè¡ãµããŒãïŒni.com/supportïŒã«ã¯ä»¥äžã®ãªãœãŒã¹
ããããŸãã
â ã»ã«ããã«ããªãœãŒã¹ â 質åã«å¯Ÿããåçããœãªã¥ãŒã·ã§ã³ãå¿ èŠãªå Žåã¯ããã·ã§ãã«ã€ã³ã¹ãã«ã¡ã³ãã®ãŠã§ããµã€ãïŒni.com/supportïŒã§ãœãããŠã§ã¢ãã©ã€ããšã¢ããããŒããæ€çŽ¢å¯èœãªæè¡ãµããŒãããŒã¿ããŒã¹ã補åããã¥ã¢ã«ããã©ãã«ã·ã¥ãŒãã£ã³ã°ãŠã£ã¶ãŒããçš®é¡è±å¯ãªãµã³ãã«ããã°ã©ã ããã¥ãŒããªã¢ã«ãã¢ããªã±ãŒã·ã§ã³ããŒããèšæž¬åšãã©ã€ããªã©ããå©çšããã ããŸãããŠãŒã¶ç»é²ãããã客æ§ã¯ãNI ãã£ã¹ã«ãã·ã§ã³ãã©ãŒã©ã ïŒni.com/jp/dforumïŒã«ã¢ã¯ã»ã¹ããããšãã§ããŸãã
â æšæºãµããŒãã»ä¿å®ããã°ã©ã ïŒSSPïŒâNI ã®ã¢ããªã±ãŒã·ã§ã³ãšã³ãžãã¢ã«ããé»è©±ãŸã㯠E ã¡ãŒã«ã§ã®åå¥ãµããŒããni.com/elearningã§ã® e ã©ãŒãã³ã°ãã¬ãŒãã³ã°ã¢ãžã¥ãŒã«ãžã®ã¢ã¯ã»ã¹ãå¯èœãšãªãããã°ã©ã ã§ãããœãããŠã§ã¢è£œåïŒäžéšãé€ãïŒãŸã㯠NI Developer Suite ãå«ããã³ãã«ãœãããŠã§ã¢ããè³Œå ¥ã®ã客æ§ã«ã¯ããæšæºãµããŒãã»ä¿å®ããã°ã©ã (SSP)ãã® 1 幎éã®ã¡ã³ããŒã·ãããèªåçã«ãæäŸããããŸãããŸããã客æ§ã®ããŒãºã«åãã㊠SSP å¥çŽã延é·ããããšã«ããããã®ç¹å žãç¶ç¶ããŠç¢ºå®ã«ãå©çšããã ãããªãã·ã§ã³ããçšæããŠãããŸãã詳现ã«ã€ããŸããŠã¯ãni.com/sspãã芧ãã ããã
ãã®ä»ã®æè¡ãµããŒããªãã·ã§ã³ã«ã€ããŠã¯ãni.com/servicesãã芧ããã ãããni.com/contactãããåãåãããã ããã
⢠ãã¬ãŒãã³ã°ãšèªå® â ãã¬ãŒãã³ã°ããã³èªå®ããã°ã©ã ã«ã€ããŠ
ã¯ãni.com/trainingãåç §ããŠãã ããããŸããäžçåå°ã§ç»é²
å¯èœãªãªã³ãµã€ããã¬ãŒãã³ã°ãæäŸããŠãããŸãã
⢠ã·ã¹ãã ã€ã³ãã°ã¬ãŒã·ã§ã³ â æéã®å¶çŽãããå Žåã瀟å ã®æè¡
ãªãœãŒã¹ãäžè¶³ããŠããå ŽåããŸãã¯ãããžã§ã¯ãã§ç°¡åã«è§£æ¶ããª
ãåé¡ãããå Žåãªã©ã¯ããã·ã§ãã«ã€ã³ã¹ãã«ã¡ã³ãã®ã¢ã©ã€ã¢ã³
ä»é² A æè¡ãµããŒãããã³ãããã§ãã·ã§ãã«ãµãŒãã¹
NI é³é¿ / æ¯åèšæž¬ã¢ã·ã¹ã¿ã³ã A-2 ni.com
ã¹ããŒãããŒã«ãããµãŒãã¹ããå©çšããã ããŸãã詳ããã¯ãæå¯
ãã® NI å¶æ¥æã«ãé»è©±ããã ãããni.com/jp/allianceãã芧
ãã ããã
⢠é©å宣èšïŒDoCïŒâ é©å宣èšãšã¯ãé©å宣èšæžã«ããããŸããŸãªæ¬§
å·é£åçäºäŒæ什ãžã®é©å宣èšã§ãããã®å¶åºŠã«ãããé»ç£äž¡ç«æ§
ïŒEMC) ã«å¯ŸãããŠãŒã¶ä¿è·ã補åã®å®å šæ§ã«é¢ããæ å ±ãæäŸãã
ãŸããã䜿çšã®è£œåã®é©å宣èšã¯ãni.com/certificationïŒè±èªïŒ
ããå ¥æã§ããŸãã
⢠Calibration Certificateâ ã䜿çšã®è£œåã§ãã£ãªãã¬ãŒã·ã§ã³ããµ
ããŒããããŠããå Žåãni.com/calibrationãã Calibration Certificate ïŒè±èªïŒãååŸã§ããŸãã
åŒç€ŸãŠã§ããµã€ãã® Worldwide Offices ã»ã¯ã·ã§ã³ïŒni.com/
niglobalïŒè±èªïŒïŒããã¯ããåãåããå ããµããŒãé»è©±çªå·ãé»å
ã¡ãŒã«ã¢ãã¬ã¹ãçŸè¡ã®ã€ãã³ãçã«é¢ããææ°æ å ±ãæäŸããååœæ¯ç€Ÿ
ã®ãŠã§ããµã€ãã«ã¢ã¯ã»ã¹ã§ããŸãã