  • Vivid S6

  • DisplaySize:1280x960pixelswith256shadesofgray and16.7millionsimultaneouscolorsavailable

    Scannersoftwaresupportsdisplayresolutionof 800x600pixel

    Screentiltsatangleof20backwardsto10forward

    Screenswivelatangle:22leftorright

    Screencanbetiltedforwardhorizontallyformobility andtransportation

    Wide-anglevisibility

    Automaticormanualdigitalbrightnessandcontrast adjustmentforoptimalviewingindifferentambient lightconditions

    Separateadjustmentforexternalmonitor brightness/contrast

    Display Formats Instant-reviewscreendisplays12simultaneous loops/imagesforaquickstudyreview

    Scanplanepositionindicatorandprobetemperature aredisplayedwithallmulti-planeTEEprobes

    Imageorientationmarker

    Selectabledisplayconfigurationofduplexandtriplexmodes:side-by-sideortop-bottomduringlive,digitalreplayandclipboardimagerecall

    Single,dualandquad-screenview

    Split-screenview

    Display Annotations MechanicalIndex(MI)

    Thermalindex:applicationdependent

    Patientname/IDandadditionalpatientinformation

    Hospitalname

    Time/date

    Trackball-drivenannotationarrows

    Scanningparameters

    Application

    Probename

    Stressprotocolparameters

    Activemodedisplay

    ParameterannotationfollowASEstandard

    Multi-languagesupportforuserinterface, keyboardoverlay,reportsanddocumentation

    Product DescriptionTheVivid S6isahigh-performanceultrasoundsystem forcardiovascularandsharedservicesapplications. Itoffersaninnovativeergonomicdesign,superbimage quality,advancedconnectivity,productivitytoolsand advancedtechnology.CompatibilitywiththeVivid productfamilyoffersflexibilityinlabconfigurationand upgradeopportunities.Enjoyincreaseddiagnostic confidenceandahighstandardofproductivity.

    System ArchitectureTheVivid S6isbasedonGEsTruScanarchitecture commontoallGEUltrasoundsystems,EchoPACPC workstation,EchoPACsoftwareandnetworksolutions. Itfeaturesasoftware-driven,PC-basedplatform,raw datastoragewithadvancedpost-processingcapabilities, seamlessDICOMStandardconnectivity,andcompatibility withtheGEfamilyofcardiovascularultrasoundsystems. Innovativetoolsofferadvancedconnectivity,remote monitoring,andconsultationtoimproveproductivityandqualityofcare.

    Coded HarmonicsProducesexcellentqualityimages evenfromdifficult-to-imagepatients.

    Data Acquisition Programmablesystemarchitecture

    Application-specificchannelarchitecture:theVividS6 employsaflexible,digitalbeam-formerarchitecture capableofusingupto2048channelsdependingon specificapplicationrequirements

    Application-specificdigitalbeamformingalgorithm foreachmode

    Supportsphasedarray,linearandcurvedarray, TEEandnon-imagingpenciltransducers

    Receivefocusing,aperture,apodizationandfrequency responseareallcontinuouslyvariableasafunction ofdepth

    Wide-aperturemodeforimprovedpenetrationon convexandlineararrayprobes

    Data Processing Echodataprocessingofphase,amplitudeandfrequency

    Easilyupgradeable

    Digitalrawdatareplayallowsforimagepostprocessinganduncompromisedofflinemeasurementandanalysis

    Display Screen High-resolution,widefield-of-view,flat17-inchTFT LCDscreen

    2

  • Tissue ImagingGeneral

    Variabletransmitfrequenciesfor resolution/penetrationoptimization

    HighResolution(HR)-zoomordisplay-zoomwith zoomareacontrol

    Variablecontourfilteringforedgeenhancement

    Variabledynamicrangeandtransmitpowersettings

    Depthrangeupto30cmprobespecific

    Selectablegrayscaleparameters:gain,reject,gray-maps,DDP,persistenceandcompressioncanbeadjustedinlive,digitalreplayandimageclipboardrecall

    AutomaticallycalculatedTGCcurvesrequireless operatorinteraction

    ContinuousTissueOptimization(CTO)ofthe2D-mode imagetooptimizetheuniformityandbrightnessofthe tissuecontinuouslyinreal-time

    SelectableAutomaticTissueOptimization(ATO) ofthe2D-modeimageindigitalreplayorimage clipboardrecall.ATOmaybecombinedwith CTOuserselectable

    ClearVessel(patentpending):aimstoreduce reverberationsandclutternoisewhilescanning thecarotid

    Smartdepth:automaticallyoptimizingtransmitpatternparametersaccordingtoscan-depthsetting(option)

    2D-mode

    Sectortiltandwidthcontrol

    CodedOctaveImaging(COI):second-generationharmonictissueimagingprovidingimprovedspatialandcontrastresolutionoverconventionalimaging*featuresreducednoiseandimprovedwalldefinitionwithoutsacrificingframeratemakingitthetissuewell-suitedinmostcases, forallpatientgroups

    * Comparative claims made throughout this document are comparing Vivid S6 imaging to conventional imaging techniques.

    Confocalimaging:allowsmultipletransmitfocalzonesoverrangeofviewandahigh-vectordensityprobesdependentanduseradjustable

    ExpandedcardiologyperformanceontheM4S-RS probe,includingsixlevelsofharmonicsandultra-highframerates

    Harmonictissueimagingonalllinearandconvexprobes

    AdaptiverejectforimprovedcardiacIQ

    UDClarityandUDSpecklereduceimaginganadvancedimageprocessingtechniquetoremovespeckleinrealtimeexaminingtherelativedifferencebetweenneighboring pixelvaluesanddeterminingwhetherthegrayscalevariationshaveasharpdifference,followatrend,orarerandominnature

    CodedPhaseInversion(CPI)forimprovedcontrastresolution

    Variableimagewidth:areductioneitherincreasesframerateorallowstoincreasethenumberoffocalzoneswhilemaintainingtheframerateapplicationdependentonlinear/convexarrayprobes

    Multiple-anglecompoundimaging:multipleco-planarimages(on4C-RSandonlinearprobes)fromdifferent anglescombinedintoasingleimageinrealtimeimprovingborderdefinition,contrastresolutionandreducingangulardependenceofborderoredge

    Virtualconvex:providesalargerfield-of-viewinthefarfieldonLinearprobes(option)

    Virtualapex:providesawiderfield-of-viewwithphasedarrayprobes,effectivewithcertainapplications

    Dualfocus(oncardiacapplications):offersadditionalfocalzoneforaddedspatialandcontrastresolutionfromheartbaseuptoapicalareas(cardiologyapplicationonly)

    Left/rightandup/downinvertinlive,digitalreplayor imageclipboardrecall

    Digitalreplayforretrospectiverevieworautomatic loopingofimagesallowingforadjustmentofparameterssuchasgain,compression,reject,anatomicalM-mode,persistenceandreplayspeed

    DataDependentProcessing(DDP)performstemporal processing,whichreducesrandomnoisebutleaves motionofsignificanttissuestructureslargelyunaffectedcanbeadjustedevenindigitalreplay

    Differentgrayandcolorized2Dmapsuserselectable inreal-time,digitalreplayorimageclipboardrecall

    M-mode Trackball-steerableM-modelineavailablewithall imagingprobesmaxsteeringangleisprobedependent

    Simultaneousreal-time2D-modeandM-mode

    M-modePRF1kHz:allimagedataacquiredare combinedtogivehigh-qualityrecordingregardless ofdisplayscrollspeed

    Digitalreplayforretrospectivereviewofspectraldata

    Severaltop-bottomformats,side-by-sideformat andtime-motion-onlyformatcanbeadjustedinlive ordigitalreplay

    Selectablehorizontalscrollspeed: 1,2,3,4,6,8,12,16secondsacrossdisplay

    3

  • Horizontalscrollcanbeadjustedinlive,digitalreplay orimageclipboardrecall

    Anatomical M-mode Vingmed-patented,anyplaneM-modedisplayderivedfrom2Dcineloop

    M-modecursorcanbeadjustedatanyplane

    Canbeactivatedfromreal-timescan,digitalreplayor imageclipboardrecall

    AnatomicalcolorM-modeavailableinreal-timescan, digitalreplayorimageclipboardrecall

    Measurementandanalysiscapability

    AnatomicaltissuevelocityM-mode(option)

    Color DopplerGeneral SteerablecolorDoppleravailablewithallimagingprobesmaxsteeringangleisprobedependent

    Trackball-controlledRegionofInterest(ROI)position/size

    Removalofcolormapfromthetissueduringdigitalreplayorimageclipboardrecall

    Digitalreplayforretrospectivereviewofcolororcolor M-modedataallowingforadjustmentofparameters,suchascolor/tissuepriorityandcolorgain,evenonstoreddata

    Powerfuldigitalsignalprocessingmaintaininghigh colorframeratesofmorethan300fpsapplication andtransducerdependent

    PRFsettings:userselectable

    Advancedregressionwallfiltergivesefficientsuppressionofwallclutter

    Foreachencodingprinciple,multiple-colormapscanbeselected,includingvariancemaps,inlivedigitalreplayorimageclipboardrecall

    Morethan65,000simultaneouscolorsprocessedprovidingsmoothdisplay,2Dcolormapscontainingamultitudeofcolorhues

    Simultaneousdisplayofgrayscale2Dand2Dwithcolorflowinliveordigitalreplayorimageclipboardrecall

    Colorinvert:userselectableinliveordigitalreplayorinimageclipboardrecall

    Variablecolorbaseline:userselectableinliveordigitalreplayorinimageclipboardrecall

    Multivariatecolorpriorityfunctiongivesreliable delineationofdisturbedflowsevenacrossbright areasofthe2D-modeimage

    ColorDopplerfrequencycanbechangedindependentlyfrom2Dforoptimalflow

    Color Doppler Imaging

    Digitalsignalprocessingpowermaintainshigh framerateswithlargeRegionofInterests(ROIs) evenforverylowPRFsettings

    VariableRegionofInterest(ROI)sizeinwidthanddepth

    User-selectableradialandlateralaveragingto reducestatisticaluncertaintyinthecolorvelocity andvarianceestimates

    DataDependentProcessing(DDP)performstemporal processinganddisplaysmoothingtoreducepossibilityforlossoftransienteventsofhemodynamicsignificance

    DigitalreplayforretrospectiverevieworautomaticloopingofcolorimagesallowingforadjustmentofparameterssuchasDDP,baselineshift,colormaps,color/tissuepriority andcolorgainevenonfrozen/recalleddata

    Application-dependentmultivariatemotiondiscriminatorreducesflashartifacts

    SmartDepth:automaticallyadjuststransmitpatternparametersaccordingtodepthofcolorROI(option)

    Multiple-anglecompoundimagingin2Dmodeis maintainedwhileinColorDopplermode

    Color Angio (Color Intensity Imaging)

    Angle-independentmodeforvisualizationofsmallvesselswithincreasedsensitivitycomparedtostandardcolorflow

    Color M-mode

    VariableRegionofInterest(ROI)lengthandposition userselectable

    User-selectableradialaveragingtoreducestatistical uncertaintyinthecolorvelocityandvarianceestimates

    Selectablehorizontalscrollspeed:1,2,3,4,6,8,12,16 secondsacrossdisplaycanbeadjustedduringlive, digitalreplayorimageclipboardrecall

    Real-time2DimagewhileincolorM-mode

    Samecontrolsandfunctionsavailableasinstandard 2DcolorDoppler

    Anatomical Color M-mode

    Vingmed-patented,anyplane,colorM-modedisplay derivedfromcolorDopplercineloop

    Availableinreal-timescan,digitalreplayorimage clipboardrecall

    AlsoapplicabletoTissueVelocityImaging(option)

    Measurementandanalysiscapability

    4

  • B-Flow (option) B-Flowisadigitalimagingtechniquethatprovidesreal-time visualizationofvascularhemodynamicsbydirectlyvisual-izingbloodreflectorsandpresentingthisinformationinagrayscaledisplay

    UseofGE-patentedtechniquestoboostbloodechoesand topreferentiallysuppressnon-movingtissuesignals

    B-Flowisavailableformostvascularand sharedserviceapplications

    Blood Flow Imaging (BFI) (option) CombinescolorDopplerwithgrayscalespeckleimaging

    Allowsbetterdelineationofbloodflowwithoutbleedingintotissueorvesselwall

    Spectral DopplerGeneral

    OperatesinPW,HPRFPWandCWmodes

    Trackball-steerableDoppleravailablewithallimagingprobesmaxsteeringangleisprobedependent

    SelectableDoppleroptimization

    Real-timeduplexortriplexoperationinPWDopplermodeforallvelocitysettings

    Frameratecontrolforoptimizeduseofacquisitionpowerbetweenspectrum,2DandcolorDopplermodesinduplexortriplexmodes

    SpectralanalysiswithanequivalentDFTrateof0.2ms

    AutomaticSpectrumOptimization(ASO)providesa singlepress,automatic,real-timeoptimizationofPW orCWspectrumscale,andbaselinedisplay

    Dynamicgaincompensationfordisplayofflowswith varyingsignalstrengthsoverthecardiaccycle

    Dynamicrejectgivesconsistentsuppressionof backgrounduserselectableinreal-time,digitalreplay orimageclipboardrecall

    Digitalreplayforretrospectivereviewofspectral Dopplerdata

    Severaltop-bottomformats,side-by-sideformatand time-motiononlyformatcanbeadjustedinliveor digitalreplayorinimageclipboardrecall

    Selectablehorizontalscrollspeed:1,2,3,4,6,8,12,16secondsacrossdisplaycanbeadjustedinliveor digitalreplayorinimageclipboardrecall

    AdjustablespectralDopplerdisplayparameters: gain,reject,compress,colormapscanbeadjusted inliveordigitalreplayorinimageclipboardrecall

    Adjustablevelocityscale

    Wallfilterswitharangeof10-3000Hz (velocityscaledependent)

    Anglecorrectionwithautomaticadjustmentofvelocityscaleinlive,digitalreplayandimageclipboardrecall

    Stereospeakersmountedinthefrontpanel

    Displayannotationsoffrequency,mode,scales, Nyquistlimit,wallfiltersetting,anglecorrection andacousticpowerindices

    PW/HPRF PW Doppler

    AutomaticHPRFDopplermaintainsitssensitivity evenforshallowdepthsandwiththehighestPRFs

    DigitalvelocitytrackingDoppleremploysprocessing inrangeandtimeforhigh-qualityspectraldisplays

    User-adjustablebaselineshiftinPWlive,digitalreplay andimageclipboardrecall

    Adjustablesamplevolumesizeof1-15mm (probedependent)

    Maximumsamplevolumedepth30cm

    Tissue Doppler Imaging

    MyocardialPWDopplerprovidesreal-timeDoppler spectralinformationforspecifiedmyocardialmotion allowingforinstantaneoustissuevelocitymeasurement

    CW Doppler

    HighlysensitivesteerableCWavailablewithall phasedarrayandpencilprobes

    Tissue Velocity Imaging and Tissue TrackingTissue Velocity Imaging TVI

    MyocardialDopplerimagingwithcoloroverlayon tissueimage

    TissueDopplerdatacanbeacquiredinbackground duringregular2Dimaging

    Segmentalwallmotionanalysiscanbeobtainedwith useofanatomicalM-modefromdigitalreplayorimageclipboardrecall

    Tissuecoloroverlaycanberemovedtoshowjustthe 2Dimage,stillretainingthetissuevelocityinformation

    Tissue Tracking

    Real-timedisplayofthetimeintegralofTVIfor quantitativedisplayofmyocardialsystolicdisplacement

    Myocardialdisplacementiscalculatedanddisplayedasacolor-codedoverlayonthegrayscaleandM-modeimagedifferentcolorsrepresentdifferentdisplacementranges

    5

  • Tissue Synchronization Imaging TSI (option) Parametricimagingwhichgivesinformationabout synchronicityofmyocardialmotion

    Delayedmyocardialsegmentsproduceredoverlaywhereassegmentsmovinginnormalrhythmaregreen

    WaveformtraceavailabletoobtainquantitativetimetopeakmeasurementfromTSIImage

    AvailableinlivescanningaswellasanofflinecalculationderivedfromTVIdataincludingvelocitytracevisualization

    Strain/Strain Rate Imaging (option) Tissuedeformation(Strain)andrateofdeformationarecalculatedanddisplayedasreal-time,color-codedoverlayonthe2DImage

    Automated Function Imaging (AFI) (option) Parametricimagingtoolwhichgivesquantitativedata forglobalandsegmentalwallmotion

    Adecisionsupporttoolforregionalassessmentofthe LVsystolicfunction

    AllowsassessingthecompleteLeftVentriclewithall segmentsataglancebycombiningthreelongitudinalviewsintoonecomprehensivebulls-eyeview

    IntegratedintoM&Apackagewithworksheetsummary

    2Dstrainbaseddatamovesintoclinicalpractice

    SimplifiedworkflowwithadaptiveROI,quicktipsand combineddisplayoftracesfromallsegments

    Automated Ejection-Fraction Calc (option) AutomatedEFmeasurementtoolbasedon2D-speckletrackingalgorithm

    IntegratedintoM&Apackagewithworksheetsummary

    Cine Memory Holdsupto180,000frames(withdefaultimagesettings:2,500to5,600)

    High-fidelityloopsandimagesmaybereviewedby scrollingorbyrunningcineloops

    TruScanarchitectureoffersbroadpost-processing capabilitiesofrecalledimagesandloopsallowing manipulationofparameterssuchasgain,baseline, colormaps,sweepspeedsandcinespeed

    ImageClipboardforthumbnailstorageand reviewofsavedimagesandloops

    Trackball-controlledcinereview

    Uninterruptible Power Supply Incaseofpowerfailureoraccidentalshutdown, whenpowerisrestoredwithinlessthan10minutes, thesystemautomaticallyturnsoninstantly, maintainingexactsystemstatepriortoshutdown

    Forlongerperiodsofpowerinterruptions,the systemautomaticallysavesdataandturnsoff intoStandbymode

    Physiological Traces IntegratedECGorexternalECGleadinput

    InternallygeneratedrespiratorytraceusingECGleads(option)

    Externalrespiratorinterface(option)

    High-resolutiondisplayoftheECGandrespiratorytrace

    UserAdjustabletracegain/positionand trace-invertcontrols

    Userpre-settabletracegain/positioncontrol

    AutomaticQRScomplexdetection

    Analysis Program Personalizedmeasurementprotocolsallowindividual setandorderofmeasurementandanalysisitems

    Measurementscanbelabeledseamlesslybyusing protocolsorpostassignments

    Bodymarkiconsforlocationandpositionofprobe

    Cardiaccalculationpackageincludingextensivemeasure-mentsanddisplayofmultiplerepeatedmeasurements

    Vascularmeasurementspackage

    Measurementsassignabletoprotocolcapability

    ParameterannotationfollowASEstandard

    Measurementsassignabletoreportgenerator

    Dopplerautotracefunctionwithautomaticcalculations inbothliveordigitalreplayorimageclipboardrecall

    Seamlessdatastorageandreportcreation

    Measurementsaresummarizedinworksheets individualresultscanbeeditedordeleted

    User-assignableparameters

    Reporttemplatescanbecustomizedonboard

    ASE-baseddefaulttextmodules(English)usercustomizable

    Imageviewduringreporting

    AbilitytoexportReportinPDForCHMformat

    GeneratereporttemplatesbytheReportDesigneror importfromEchoPACPC

    6

  • Smart Stress Echo (option) Stresspackagewithmemorybufferofferspharmaceutical,treadmillandbicyclestressexamprotocolswithuser-configurabletemplatesandshufflemode

    SmartStressfunctionwiththeabilitytosaveover 17imagingparametersfromeachimagingplane theseimagingparametersarerecalledateachstress level,therebyrequiringnosystemadjustments

    Referenceloopdisplayduringacquisitionforcomparingrestingimagesorprevious-levelimagestoeachstresslevel(dualscreen)

    Advancedandflexiblestress-echoexaminationcapabilities

    Imageacquisition,review,all-segmentscoring andreporting

    Stresstreadmill-exercisewithmorethan120secondsofrawdatacontinuouscapture

    Possibilityofextensivepost-processingofimages underreview

    Wallmotionscoring(Bulls-eyeandsegmental)

    TemplateEditortocustomizethenumberofstresslevels,numberofviews,numberofheartcycles,andsystolicorfull-cyclecapture

    OB/GYN Application Module (option) OBpackageforfetalgrowthanalysiscontaining morethan100biometrytables

    DedicatedOB/GYNreports

    Fetalgraphicalgrowthcharts

    Growthpercentiles

    Multi-gestationalcalculations(uptofour)

    ProgrammableOBtables

    Expandedworksheets

    User-selectablefetalgrowthparametersbasedon European,AmericanorAsianmethodscharts

    GYNpackageforovaryanduterusmeasurements andreporting

    Quantitative Analysis (Q Analysis) (option) Qanalysissoftwarepackage,designedforanalysisofTVIrelated(TissueTracking,Strain,Strainrate,TSI)rawdata

    QuantitativeprofilesforTVI,tissuetracking,strainandstrainratecanbederived.Uptoeighttracescanbe generatedfromselectedpointsinthemyocardium

    Sample-areapointsmaybedynamicallyanchoredtomovewiththetissuewhenrunningthecineloop

    ArbitraryCurvedanatomicalM-Mode

    Cinecompounddisplayscineloopsgeneratedfroma temporalaveragingofmultipleconsecutiveheartcycles

    Quantitativeprofiles(Q-Analysis)canbederivedondatatransferredtoEPPCworkstation

    Intima-Media Thickness (IMT) Measurement Program (option) Automaticmeasurements(patentpending)ofcarotid arteryIntima-MediaThickness(IMT)onanyacquiredframe

    On-boardIMTpackageprovidesnon-interrupted workflowfullyintegratedwithM&A,worksheet, archivingandreportingfunctions

    Robustalgorithmprovidesquick,reliablemeasurementswhichcanbestoredtotheon-boardarchiveforreviewandreporting

    IMTmeasurementcanbemadefromfrozenimages orimagesretrievedfromarchive

    IMTpackagesupportsmeasurementsofdifferentregions oftheintimainthecarotidvessel(e.g.,Lt./Rt./CCA/ICAetc.)

    FrameforIMTmeasurementcanbeselectedinrelation totheECGwaveform

    IMTprotocolallowstopre-definethesizeandlocation oftheIMTROIrelativetosomeverticalmarker(bulb), tosupportconsistentresults

    User Interface ErgonomicFlexFitdesignwithleft/rightswivel andup/downarm-mobilityofkeyboardand monitorpermittingbothphysiologicalsittingor standingoperation

    Easy-to-learnuserinterfacewithintelligentkeyboard

    Keyboardwithapplication-specificassignablerotaries andpushbuttonsforprimarycontrols

    Interactiveback-lightingofapplication-specific pushbuttons

    Full-size,alphanumerickeyboardwith adjustablebacklighting

    Application-specificsecondarycontrolsavailablethroughslidebarsoperatedbyafour-wayrocker

    SlidepotTGCcurvewithsixpots

    Dedicatedrotaryforoverallgainfor2D-mode

    DedicatedrotaryforM-mode,CFMorDopplercontrolledbyactivemode

    Digitalharvestingofimagesandloopsinto imageclipboard

    Patientbrowserscreenforregistrationofdemographicdataandquickreviewofimageclipboardcontents

    7

  • 8 Fullyprogrammableuserpresetsforprobe/applicationdefaultsettings

    Supportforinternationalkeyboardcharactersin 12languages

    Integratedspeakers

    Probeandgelholdersonbothsidesofkeyboard

    Threeuser-programmableFLEXkeysforeasyaccesstoanoften-usedfunctioninglobalandonapplicationlevel

    Wideband Probes Electronicselectionbetweenfoursolid-stateand onestandaloneDopplerprobeconnectors

    Threeprobe-socketsareRStypeonesocketisLOGIQ typetosupportTEEprobeswithLOGIQconnectorsonly

    PROBE FREQUENCY RANGE CATAlOG #

    Phased Array Sector Probes

    M4S-RS 1.53.6MHz H40452LH

    5S-RS 2.05.0MHz H4000PC

    6S-RS 2.78.0MHz H45021RP

    7S-RS 3.58.0MHz H4000PE

    10S-RS 4.511.5MHz H4000PF

    linear Array Probes

    8L-RS 4.013.0MHz H40402LT

    9L-RS 3.510.0MHz H40442LL

    12L-RS 6.013.0MHz H40402LY

    Convex Array (Curved) Probes

    4C-RS 1.86.0MHz H4000SR

    8C-RS 4.011.0MHz H40402LS

    Convex Array Transvaginal Probe

    e8C-RS 4.011.0MHz H40402LN

    Intra-Operative Probes

    i12L-RS 5.013.0MHz H40402LW

    Doppler Pencil Probes

    2D(P2D) 2.0MHz H4830JE

    6D(P6D) 6.0MHz H4830JG

    Multi-Plane Transesophageal Phased Array Probes

    6Tc-RS 2.98.0MHz H45551ZE

    9T-RS 4.010.0MHz H45531YM

    6Tc 2.98.0MHz H4551ZD

    9T 4.010.0MHz H45521DY

    Biopsy Bracket Support (option) On-screenbiopsyGuide-Line,Guide-Zoneand depthmeasureforCivcomulti-anglebiopsy bracket,supportingprobemodels

    M4S-RS

    4C-RS

    8L-RS

    9L-RS

    12L-RS

    e8C-RS

    Supported Applications (probe dependent) Cardiac(adults,pediatricsandfetalheart)

    Vascular

    Pediatric

    Neonatalcephalic

    Transcranial(adultcephalic)

    Abdominal

    Gynecological

    Obstetrical

    MusculoskeletalincludingSuperficial

    SmallParts

    Breast

    NerveImaging

    Coronary

    Advanced OptionsContrast Imaging*

    Allcontrastagentsshouldbeusedasdescribed onthelabelbythecontrastagentmanufacturers

    lVO Contrast (option)*

    LVContrast(on M4S-RS, 5S-RS 6Tc, and 6T-RS or 6Tc-RS probes)enhancesdelineationoftheLVborderin combinationwithultrasoundcontrastagents.ThenewimplementationofGEsCodedPhaseInversion(CPI) provideshigh-resolutiondetectionofcontrastinthe LVcavityandexcellentsuppressionofmyocardial tissuesignals

    * Harmonic imaging for supporting contrast agent imaging was developed by Schering.

  • 9lOGIQView (option) LOGIQView:providestheabilitytoconstructandview

    astatic2Dimagewithwiderfield-of-viewofalinear arraytransducerthisallowsviewingandmeasure- mentsofanatomythatislargerthanwhatwouldfit inasingleimage

    Image Management and Archiving Built-inpatientarchivewithimages/loops,patient information,examinationinformationandtexts, measurementsandreport

    Rawdataworkflowwithinstantdatamanagement

    Dataareeitherstoredinternallyortoaremotearchive(EchoPACorImageVaultserver)

    Rawdataallowschangestogain,baseline,colormaps,sweepspeeds,etc.forrecalledimagesandloops

    DICOM3.0ImageFormat:DICOMincorporatesrawimagedatainformationwithallitsdatamanagementflexibilityintotheimagecommunicationstandardDICOM

    ImagescanbedirectlystoredorexportedinDICOM formattoaDICOMserver(PACS) (optionseeDICOMNetworkConnectivity)

    Imageclipboardforstamp-sizedpreviewimagestoallowrecallingimagesorloopsofchoicedirectly

    2D,CFMandTVIdataatmaximumframeratemaybereviewedbyscrollingorbyrunningcineloops

    Internalarchivedatacanbeexportedtoremovableimage storagethroughDVD/CD-RW,USBflashcard(option),ExternalHard-disk(option),Magnet-OpticalDisk(option) inRaw-DataandDICOMformat

    Internalharddisk:forstoringprograms,application settings,ultrasoundimagesandpatientarchive

    Over120Gbytediskspaceforexamarchivestorage

    ConfigurableHTML-basedreportfunction

    Reporttemplatedesignerpackage

    Raw-Data,DICOM,AVI,MPEGandJPEGexport

    Built-inDVDwriter(supportsCD-RandDVD-R)

    Self-contained DICOM Viewer ExamscanbeexportedtoCD/DVDorUSBmediawith anintegratedEZDICOMCDviewerTM

    Self-containedEZDICOMCDviewerTM allowsreview ofexamsfrommediaonastandardPC

    Excel Export AllowsexportofallarchivedmeasurementandtextualpatientinformationinstandardMicrosoftExcelfiles

    EchoPAC Connectivity ConnectivityandimageanalysiscapabilityofVividS6 fromEchoPACPC

    EchoPACPCallowsinstantaccesstoultrasoundrawdataprovidedbythesystem

    Comprehensivereview,analysisandpost-processing capabilitiesonEchoPACPC

    Advancedquantitativeanalysisand post-processingcapabilities

    Q-AnalysisonrawdatafromVividS6onEchoPACPC

    Threeuserlevelshelporganizingdata securityrequirements

    DICOM Media Support DICOMmedia:read/writeimagesonDICOMformat

    DICOM Network Connectivity (option)ProvidescommunicationtoaDICOMserverand DICOMprinter.Includes:

    Ethernetnetworkconnection

    VerificationAE

    ImageexportAE(networkstorage)

    ModalityworklistAE

    Storagecommitment

    Performedprocedurestep

    StoragetoDICOMserver

    DICOMstructuredreportSCUforcardiacandvascular

    Verify:providesverificationofanactiveconnection betweenthescannerandanotherDICOMdevice

    DICOMprint

    SupportoftwopatientIDfieldsinDICOM

    DICOM Modality Worklist (option) Modalityworklist:givesaccesstoalistofpatients providedbyaworklistserver

    DICOM Print (option) AllowsprintingimagesviaaDICOMprinter

  • 10

    Database Importation from Vivid 3 and Vivid 4 Systems (option) Providesaone-timedatabaseimportcapabilityfrom aVivid3oraVivid4system

    Allowsusertoreviewpreviousexamsgeneratedon aVivid3oraVivid4systeminDICOMformat

    Allowstoaddexaminationstopatientsalreadyexisting intheimporteddatabase

    Virtual Printer (option) ProvidestheabilitytosendPrintcommandstoany oftwoprintersevenwhennotconnectedtoaprinteruponre-connectionofprinter,thesystemautomaticallyproduceshardcopiesfromprintimagessavedin chronologicalsuccessionondisk

    MPEGvue (option) UsingMPEGvue,examsmaybestoredontoremovablemediaoronremotenetworkedshareddrivetogetherwithintegratedMPEGvueplayerforviewingonstandardPC

    Smartemailfeatureallowstransparenttransmission ofimagesfromaseparatePCviaemailusingresidentOutlookemailclient

    Patientmanagementutilityonthereceivingstandard PCprovidesabilitytoorganizetheexamsondifferent sub-directoriesontheusersharddisk

    eVue (option) Allowsinteractiveviewingofimages,loopsorfull examsfromremotelocationonanyPC,usingLAN orwirelessLAN

    Insite Express Connection (ExC) Enables Remote Service and Training* Easy,FlexibleandSecureconnectivityconfiguration

    The"ContactGE"on-screenbuttondirectlygeneratesareal-timeservicerequesttotheGEonlineengineeringorapplicationspecialist.Ittakesasnapshotofthesystem atthetimeoftheservicerequesttoenableanalysisofproblembeforecustomercontact.

    VirtualConsoleObservation(VCO)enablesthecustomerto allowdesktopscreenstobeviewedandcontrolledremotely overtheencryptedtunneltoenablereal-timetraining,deviceconfigurationandclinicalapplicationsupport.

    * Operation of Insite Express Connection is dependent on the infrastructure being available. Check with your local GE service representative.

    FileTransferenablesthecustomer(BiomedorClinician) todirectlytransfersysteminformation(e.g.systemlogs,images,parametricdata)toGEproductengineeringteams

    Softwarereloadprovidesremoteapplication reconstructionandrecoverycapabilitiesinthe eventofsystemcorruption

    Peripherals (options) USBblackandwhitevideoprinterwithcontrolfrom systempanel

    USBcolorvideoprinterwithcontrolfromsystempanel

    USBinkjetprinter,supportingink-savingmode

    ExternalMOD5.25"drive

    USBflashdriveforexamexport

    USBflashmemorycard

    USBwirelessnetworkinterfaceaccessories

    ExternalUSBHardDiskUSB2.0RAID1mirrors thecontentsofoneharddiskontoanother effectivesize:1TB

    Accessories (options) InterfacecableforExternalECG

    ECGadapterforDIN-typepediatricselectrodeleads

    Three-pedalfootswitchwith programmablefunctionality

    Rearandfolderboxes

    Disposablehygeniccoversfortheoperatorpanel

    Console Probeandgelholder

    Handrestandhandles

    Fourswivelwheelsfrontwheelbreaks

    Rearwheelsdirectionlock

    Inputs and Outputs DVI/VGAvideooutput

    Audioout

    FourUSB-2connectorstwoatrearandtwoatfront tosupportvideoprinters,MOD,flashmemorycards

    LANEthernet

  • 10

    Dimensions and Weight Depth:70cm/27.6"(approx.)

    Width:55cm/21.7"

    Height:123cm/48.4"to143cm/56.3"

    Minimumheightwithfoldedscreen:95cm/37.4"

    Weight:

  • Healthcare Re-imaginedGEisdedicatedtohelpingyoutransformhealthcare deliverybydrivingcriticalbreakthroughsinbiology andtechnology.Ourexpertiseinmedicalimaging andinformationtechnologies,medicaldiagnostics, patientmonitoringsystems,drugdiscovery,and biopharmaceuticalmanufacturingtechnologiesis enablinghealthcareprofessionalsaroundtheworld todiscovernewwaystopredict,diagnoseandtreat diseaseearlier.WecallthismodelofcareEarlyHealth.Thegoal:tohelpcliniciansdetectdiseaseearlier, accessmoreinformationandinterveneearlierwithmoretargetedtreatments,sotheycanhelptheir patientslivetheirlivestothefullest .Re-think, Re-discover,Re-invent,Re-imagine.

    GEHealthcare9900InnovationDriveWauwatosa,WI53226U.S.A.

    www.gehealthcare.com

    2011GeneralElectricCompanyAllrightsreserved.

    June2011

    GeneralElectricCompanyreservestherighttomakechangesinspecificationsandfeaturesshownherein,ordiscontinuetheproductdescribedatanytimewithoutnoticeorobligation.ContactyourGErepresentativeforthemostcurrentinformation.

    GE,GEMonogram,Vivid,EchoPACandVingmed aretrademarksofGeneralElectricCompany.

    ExcelisaregisteredtrademarkofMicrosoftCorporation.

    ScheringisatrademarkofBayerHealthcareAG.

    GEMedicalSystemsUltrasound&PrimaryCare Diagnostics,LLC,asubsidiaryofGeneralElectric Company,doingbusinessasGEHealthcare.

    DOC0762992