~ ENERGY SOLUTIONS
EnergySolutions Services, Inc. 628 Gallaher Road Kingston, TN 37763 Phone: (877) 462-4873 Fax: (865) 376-8331
IS .ert1 1cate w1 Th ' c 'fj ·u b e accompame 'Y a 1 ration . db C l'b . Ch arts or ea mgs w ere app11ea R d' h r bl e
CUSTOMER INFORMATION INSTRUMENT INFORMATION
Customer Name: Duratek Instrument Services Manufacturer: Ludlum
Address:628 Gallaher Road, Kingston, TN 37763 Model: 2360 Serial Number: 193647
Contact Name: Tony Riggs Probe: N/A Serial Number: N/A
Customer Purchase Order I Work Order Calibration Method: Number: N/A Number: 2011-11583 Electronic
INSTRUMENT CALIBRATION INFORMATION
Ratemeter Response(::!: 10% of Calibration Time
Instrument Calibration Standard Values) Standard Value Base
Tolerances Range Standard Value (cpm)
CALIBRATION CERTIFICATE
Scaler Response
As Found As Left CPM (minutes) ::1:2% As Found As Left
XI IOO 100 100 1,000 CPM 0.1 90-110 98 98
X I 250 250 250 I,OOO CPM 0.5 450 - 550 486 486
XI 400 400 400 1,000 CPM 1 900-1,100 990 990
X 10 1,000 1,000 1,000 1,000CPM 2 1.8K-2.2K 1,989 1,989
X 10 2,500 2,500 2,500 1,000 CPM 5 4.5K-5.5K 4,956 4,956
X 10 4,000 4,000 4,000 1,000 CPM 10 9K-11K 9,913 9,913
X 100 10,000 10,000 10,000
X 100 25,000 25,000 25,000
X 100 40,000 40,000 40,000
X 1000 100,000 100,000 100,000 Calibrated in accordance with OEM Technical Manual
X 1000 250,000 250,000 250,000
X 1000 400,000 400,000 400,000
STATEMENT OF CERTIFICATION
We Certify that the instrument listed above was calibrated and inspected prior to shipment and that it met all the Manufacturers published operating specifications. We further certify that our Calibration Measurements are traceable to the National Institute of Standards and Technology. (We are not responsible for damage incurred during shipment or use of this instrument). Instrument
d4-Calibrated By: tv\.?~ . .' Reviewed By: Date: /?/;k Calibration Date: 07/12/2011
*Calibration Due (6mo): 01112/2012 , ,
*Calibration Due (12mo): 07/12/2012 Calibration due date IS de endant on users re ulato ep g re UJrements. ry q
~ ENERGY SOLUTIONS
Page 2 of2 Model: 2360 Serial Number: 193647
M&TE Environmental Conditions
Volt Meter ID# 94710023 Cal Due: 10/28/2011 Barometer ID# 3590 Cal Due: 09/21111
Pulser ID# 151067 Cal Due: 08/27/2011 Thermometer ID# 3590 Cal Due: 09/21/11
Humidity ID# 958670 Cal Due: 06/07/2012 Temp: 21.1·c Pressure: 741 mmHg Humidity: 43%
Special Test
BAT Check Sat ( .J) Unsat ( ) Geotropism Sat ( .J) Unsat ( )
LCD Display Check Sat ( .J) Unsat ( ) Audio Check Sat ( .J) Unsat ( )
Mechanical Zero Sat ( -.J) Unsat ( ) LowBATSet Sat ( -.J) Unsat ( )
Reset Sat ( -.J ) U nsat ( )
HV Analog Display Sat ( ..J) Unsat ( ) As Found As Left
High Voltage Calibration(± 10%) Alpha Sensitivity = 119mv Alpha Sensitivity = 120 mv
Voltage Tolerance As Found As Left Beta Sensitivity = 3.6mv Beta Sensitivity = 3.5 mv
500 450-550 556 556 Beta Window = 30.4 mv Beta Window = 30mv
1000 900-1100 1,056 1,056 Beta Setpoints-Pulser counts detected at 3.5mv ± lmv and shut off at 30mv for beta. For Alpha channel counts detected
1500 1350-1650 1,559 1,559 · at 120mv and above. "
H.V. Set With Detector Not Connected Overload to be set with detector to be used
COMMENTS
Calibrated in accordance with OEM Technical Manual
See detector certificate for High Voltage setting
**Calibrated with 5ft cable**
Instrument dL Calibrated By: M·~· Reviewed By: Date: %h/ *Calibration Due (6mo): 01/12/2012 / ,
Calibration Date: 07112/2011 *Calibration Due (12mo): 07/12/2012
call bran on aue a ate IS ae enaant on users regulatory reqmrements. p
8 Duratek··
Duratek Instrument Services 628 Gallaher Road Kingston, TN 37763 Phone: (865) 376-8337 Fax: (865)376-8331
CALIBRATION CERTIFICATE
IS ert1 1cate WI Th. c ·r. ·n b e accompame 'Y a 1 ration · db c rb · Ch arts or ea mgs w ere app11ca e R d. h r bl
CUSTOMER INFORMATION DETECTOR INFORMATION
Customer Name: EnergySolutions- Instrument Services Manufacturer: Ludlum
Address: 628 Gallaher Road Kingston, TN 37763 Detector Model: 43-93
Contact Name: Tony Riggs Serial Number: 238107
Customer Purchase: I Work Order Evaluation Method: Order Number: N/A Number: 2011-11538 Source
DETECTOR EVALUATION INFORMATION
Source Nuclide Serial Number Activity (dpm) 2 Pi Emissions Net Response (cpm) Efficiency(%)
Th-230 119738 18,600 8,640/ min 3,355 18.0% (4pi) 38.8% (2pi)
*Pu-239 019442 13,607 N/A 2,974 21.9% (4pi) *N/A (2pi)
Tc-99 099608 21,311 10,500/ min 3,413 16.0% (4pi) 32.5% (2pi)
SCALER INFORMATION DETECTOR INFORMATION
Model Serial Number Due Date Background Operating Voltage Threshold
2360 193647 07/12/2012 0.4 800V Alpha (120mV)
2360 193647 07/12/2012 209 800V Beta (3.5-30mV)
ATTACHMENTS
Voltage Plateau: ..JYES NO MDA/Cross Talk Evaluation: ..JYES NO
COMMENTS LINEARITY TEST
"'No 2pi emission rate Is listed on the source certificate. Count I (Toe) 3,671
Calibrated with 5ft cable Count 2 (Mid) 3,672
Linearity test performed with Tc-99#099608. Count 3 (Heel) 3,524
Calibrated in accordance with original equipment technical Average 3,622
manual. Pass/Fail PASS (+/-10% Tolerance)
STATEMENT OF CERTIFICATION
We Certify that the detector listed above was evaluated for proper operation prior to shipment and that it met all the Manufacturers published operating specifications. We further certifY that our Calibration Measurements are traceable to the National Institute of Standards and Technology. (We are not responsible for damage incurred during shipment or use of this detector). Detector
d~ Certified By: M·~· Reviewed By: Date: C/o2/t; I *Certification Due (6mo): 01/12/2012 ;!'
Certification Date: 07/12/2011 I *Certification Due (12mo): 07/12/2012 CalibratiOn due date 1s de enaant on users re p llato gu ry requtrements.
Page I of I
.Durotek~ Duratek Instrument Services
628 Gallaher Road
Kingston, TN 3776
Phone: (865) 376-8337
Fax: (865) 376-8331
Email : [email protected]
CALIBRATION CERTIFICATE
http :1/www. energysol utions .com/
IS erh 1cate WI Th" c "fi ·11 b e accompame >Y a 1 rauon . db C l "b . Ch arts or ea mgs w ere app11ca e R d. h r bl
CUSTOMER INFORMATION INSTRUMENT INFORMATION
Customer Name: Energy Solutions Manufacturer:
Address: 628 Gallaher Rd,Kingston TN 37763 Model : 2360 S.N. 193647
Contact Name: Tony Riggs Probe: 43-93 S.N. 238107
Customer Work Order PO No.: N/A Number: 2011-11247 Calibration Method: Source
Source Information
Isotope Source 10 Certifcation Date Activity ( dpm) a Source Pu-239 019442 6/1/92 13,607 ~Source Tc-99 099608 8/8/96 21,311
Ludlum Model 2929 High Voltage Plateau with crosstalk
Operating High Back~ round Alpha Source Beta Source Crosstalk Efficiency Voltage Voltage Alpha Beta Alpha Beta Alpha Beta a to J3 J3 to a a J3
750 1.0 118.0 2,855.0 263.0 0.0 2,643.0 5.30% 0.00 20.97% 11 .85% 775 2.0 167.0 2,872.0 471 .0 1.0 3,130.0 11.24% 0.00 21.09% 13.90% 800 2.0 206.0 3,125.0 790.0 3.0 3,797.0 20.01% 0.03% 22.95% 16.85% 825 1.0 232.0 3,080.0 1,104.0 1.0 4,259.0 30.62% 0.00% 22.63% 18.90% 850 0.0 240.0 3,061.0 2,252.0 2.0 4,490.0 71 .32% 0.05% 22.50% 19.94%
STATEMENT OF CERTIFICATION We Certify that the Instrument listed above was calibrated and inspected prior to shipment and that it met all the Manufacturers published operating
specifications. We further certify that our Calibration Measurements are traceable to the National Institute of Standards and Technology. (We are not responsible for damage incurred during shipment or use of this instrument).
Comments :
Calibrated By: ~u:~ ?. . J' Reviewed By: 4 Date: ::f/OZP/ ~ /
Calibration Date: 7/12/1 1 Calibration Due: 7/11/12
Page 1 of 1
.::
~ ENERGY SOLUTIONS
CALIBRATION CERTIFICATE
EnergySolutions Services, Inc. 628 Gallaher Road Kingston, TN 371·6) Phone: (87~)'462-4873 Fax: (865) 376-8331
'I This Certificate w1l be accompanied by a 1 ration Cl'b. Ch arts or. Rea m~~ W Cl:C 8PPIIC d' ' ti ' j· abl e .. -· - .. - .. """"'
·~· .. ~. -~· .... ' .
CUSTOMER INFORMATION .. ,
DETECTOR INFORMATION •. ' ''
Customer Name: EncrgySolutions Services, Inc. Manufacturer: Ludlum
Address :- 628 Gallaher Road Kingston, TN 37763 ' ~ ~ Detector Mode1: 43-93 ! .
Contact Name: Tony Riggs ·' • Serial Numl)er: ·1 : 195818 .. : ' ,Ht. t.... ,, j
Customer Purchase: I Work Order · ~ ..... ~- '::- -EYaluation 'Metliod:.. ••. -. ••• j ·,.J .• :. .. ,., ~ -· -... - ... .... _. ._ . . · ...
Order Number: N/A Number: 2011-11271 Source · ~. '. _:·;, :1("" ~ · ·.: . • , : ·.
DETECTOR EVALUATION INFORMATION '·
Source Nuclide Serial Number Activity (dpm) 2 PI Emissions Net Response (epm) Efficiency(%)
*Pu-239 019442 13,607 N/A 2,923 21.5% (4pl) *N/A (2pi)
Th-230 119738 18,600 8,640/ min 3,115 16.8"(6 (4pl) 36.1% (2pi)
Tc-99 099608 21,311 10,500/ min •.' .~ .• ?77.3 12.5efo{~pi) ~6.45% (2pl)
-SCALER INFORM~TION DETECTOR INFORMATION .... "-·· .. ' '"·· ••',/ ·- .............. -.. . '-~ ... .
Model Serial Number Due Date **Background Operating Voltage Threshold
2360 276935 02/0712012 0 650V Alpha (120mV) -2360 276935 02/07/2012 116 650V Beta (3.5-30mV)
ATTACHMENTS
Voltage Plateau: ..JYES NO MDA/Cross Talk Evaluation: YES ..JNO
COMMENTS LINEARITY TEST
* No 2pi emission rate is listed on the source certificate. Count 1 (Toe) 2,648
** 5 minute background performed. Count 2 (Mid) 2,753
Linearity test performed with Tc-99#099608. Count 3 (Heel) 2,931 Calibrated with 5ft cable
Calibrated in accordance with original equipment technical Average 2,777.3
manual. Pass/Fall PASS (+/-15% Tolerance per LMI)
STATEMENT OF CERTIFICATION
We Certify that the detector listed above was evaluated for proper operation prior to shipment and that it met all the Manufacturers published operating specifications. We further certif)• that our Calibration Measurements are traceable to the National Institute of Standards and Technology. (We are not responsible for damage incurred during shipment or use of this detector). Detector
$~ Certified By: Reviewed By: v#.J;6~~ Date: ~~~ ¥11/ , I *Certification Due (6ri1o)f 08/24/2011 Certification Date: 02/24/2011 I *certification Due (12mo): 02/24/2012 t..auoratJon aue aate IS aepenaant on users regulatory requirements.
Page I of I
Designer and Manufacturer of
Scientific ond Industrial lnstrvments
LUDLUM MEASUREMENTS, INC. POST OFFICE BOX 810 PH . 325-235-5494 501 OAK STREET FAX NO. 325-235-4672 SWEETWATER, TEXAS 79556, U.S.A.
Bench Test Data For Detector
Detector ~3-23 Serial No. lt.. z.f. i..IJ..Y Order#. 20169058
Customer ENERGY SOLUTIONS Alpha Input Sensitivity 12/l mV
Counter 2360 Serial No. Z.'&'l..lr' Beta Input Sensitivity ,l.f" mV
Count Time !Minute Beta Window .1a mV
Other
./
High Voltage
h/J~
~z..r
lro
67~
"700
Background
Alpha Beta
tJ .Jt:J
t) (,(
I IOV
() Jtfr-
t:) ZOR'
'
Isotope Ia 1J9 Size r.Jo9oa,/.A"'
Alpha Bela
'IJ'/'1 l(f.z
SSiJ t .U?
hO'fJ o/tJ/
fi11'J """ 7fJ7"7 ~,r
Distance Source to Detector
Isotope Size
Alpha
'{
(
_]
(
7
'
Bela
131"1
27f/J.
:· ¥1109
Got9
~~~-..,
'
0 Gas Proportional detector count rate decreased !> 10% after 15 hour static test using 39" cable.
~ utfzc.t:.,
Isotope S"t10t9{) Size 1fTPB'#,..
Alpha Beta
I') //)f/2.
tJ /I.7L
, ZJlo
t:J "VItl
I .'?'lb/
'
0 Gas proportional detector count rate decreased !> 10% after 5 hour static test using 39'' cable and alpha/beta counter.
Signature ------,?""~~ ..... llflf1Z'U'IMJ:..-.6i~~~...:::~'7=:.__----------
• Serving The Nuclear Industry Since 1962 • >RM C4B 03/l l /2010 Poge 2 ol 'l-----