Gcms and Hplc Methods

download Gcms and Hplc Methods

of 8

Transcript of Gcms and Hplc Methods

  • 7/28/2019 Gcms and Hplc Methods

    1/8

    Journal of Chromatography A, 1141 (2007) 9097

    Gas chromatography/mass spectrometry versus liquidchromatography/fluorescence detection in the analysis

    of phenols in mainstream cigarette smoke

    Serban C. Moldoveanu , Melissa Kiser

    R.J. Reynolds Tobacco Co., 950 Reynolds Boulevard, Winston-Salem, NC 27105, USA

    Received 19 October 2006; received in revised form 28 November 2006; accepted 29 November 2006

    Available online 19 December 2006

    Abstract

    A newgas chromatographic/massspectrometric(GC/MS)technique for the analysis of hydroxybenzenes (phenols)in mainstream cigarettesmoke

    has been developed. The technique allows the measurement of 24 individual compounds, and the sum of a few other alkyl-dihydroxybenzenes.

    A critical evaluation is done for the new technique and for an established high-performance liquid chromatographic (HPLC) technique reported

    in the literature for the analysis of hydroxybenzenes in cigarette smoke, which uses fluorescence detection. Compared with the HPLC procedure,

    the new technique has similar accuracy, precision, and robustness. However, the GC/MS procedure allows for a larger number of phenols to be

    analyzed simultaneously, and eliminates any potential interference that may appear in the HPLC method. Using the GC/MS analysis, it was found

    that besides the main phenols typically measured in mainstream cigarette smoke such as phenol, catechol, hydroquinone, and cresols, many other

    phenols that are present at lower levels can be quantitated in mainstream cigarette smoke.

    2006 Elsevier B.V. All rights reserved.

    Keywords: Phenols; Hydroxybenzenes; HPLC; GC/MS; Mainstream cigarette smoke

    1. Introduction

    Hydroxybenzenes (phenols) are present in cigarette smoke

    and contribute to cigarette sensory properties [1]. Also, some

    phenols are considered toxic compounds [2] and their pres-

    ence in smoke has been related to environmental and health

    issues. Thus, the analysis of phenols in smoke has been the

    subject of several studies, some reported in peer reviewed liter-

    ature [310]. Outside the tobacco industry, numerous published

    reports describe phenol analysis, which has been commonly per-

    formed on samples such as water [1118], food [1921], plant

    materials (fruits, vegetables) [22], pharmaceuticals [23,24], etc.

    Phenols can be analyzed by both high-performance liquid chro-

    matographic (HPLC) and gas chromatography (GC). The HPLC

    analysis can be done with UV detection [25,26] or fluores-

    cence detection [3,23,27]. Several advantages of fluorescence

    versus UV detection are described in the literature, the sensi-

    Corresponding author. Tel.: +1 336 741 7948; fax: +1 336 728 9112.

    E-mail address: [email protected] (S.C. Moldoveanu).

    tivity and selectivity of fluorescence detection being in general

    better than that of UV detection [23]. The analysis of phenols by

    GC or gas chromatography/mass spectrometry (GC/MS) is usu-

    ally done after derivatization [7,10,24,28,29]. Other analytical

    techniques for phenols determination are also reported such as

    micellar electrokinetic chromatography [30]. The phenols typi-

    cally analyzed in cigarette smoke by the HPLC method include

    1,4-dihydroxybenzene (hydroquinone), 1,2-dihydroxybenzene

    (catechol), 1,3-dihydroxybenzene (resorcinol), hydroxyben-

    zene (phenol), 4-methyl + 3-methyl-hydroxybenzene (p- + m-

    cresol), and 2-methylhydroxybenzene (o-cresol). Although

    2-methoxyphenol (guaiacol) is not usually analyzed by HPLC

    this compound can be easily measured by this technique.

    Besides the six (seven, if 2-methoxyphenol is added) phenols

    typically analyzed by HPLC/fluorescence, a number of other

    phenols are present in cigarette smoke [31]. They are not listed

    as biologically active agents [2] and their levels are relatively

    low. Nevertheless, their presence in smoke may contribute to

    the properties of smoke, and a full chemical characterization of

    cigarette smoke should include them. However, the analysis of

    these additional phenols by HPLC/fluorescence raises a selec-

    0021-9673/$ see front matter 2006 Elsevier B.V. All rights reserved.

    doi:10.1016/j.chroma.2006.11.100

    mailto:[email protected]://localhost/var/www/apps/conversion/tmp/scratch_2/dx.doi.org/10.1016/j.chroma.2006.11.100http://localhost/var/www/apps/conversion/tmp/scratch_2/dx.doi.org/10.1016/j.chroma.2006.11.100mailto:[email protected]
  • 7/28/2019 Gcms and Hplc Methods

    2/8

    S.C. Moldoveanu, M. Kiser / J. Chromatogr. A 1141 (2007) 9097 91

    tivity problem, since the separation of some of these phenols

    may be difficult by HPLC. Also, their positive identification

    without a mass spectrum can be questionable. For this reason, a

    GC/MS technique for phenol analysis in cigarette mainstream

    smoke has been developed and is described in this study. The

    application of the HPLC/fluorescence analysis and of GC/MS

    analysis on the same samples allows a good comparison of

    advantages and disadvantages of both techniques.

    2. Experimental

    2.1. Smoke collection

    Smoke collection was done with a Borgwaldt RM20 CSR

    smoking machine, tuned for conditions similar to those recom-

    mended by US Federal Trade Commission (FTC) [32]. That

    consisted of 35 mL puff volume, each puff taken every 2 s, with

    60 s puff interval, the cigarettes being smoked to 3 mm distance

    to the filter overwrap. Other smoking regimes [3336] can also

    be utilized depending on the purpose of the analysis. Five differ-entcigarettes were used in thestudy. This included two Kentucky

    references cigarettes indicated as 1R5F and 2R4F (University of

    Kentucky, Kentucky Tobacco Research & Development Center,

    KY, USA) and three commercial cigarettes. The description of

    these cigarettes is given in Table 1.

    Smoke from five cigarettes was collected in each run on a

    92 mm Cambridge pad. The results previously reported in the

    literature [3] indicate that phenols are effectively collected from

    mainstream smoke by this procedure. A chromatographic stan-

    dard consisting of 100L solution containing 250g/mL of

    4-chlorophenol was added to the pad. The pad was extracted on

    a mechanical shaker for 30 min with 25 mL water containing 1%acetic acid and 0.1% ascorbic acid (both from Aldrich/Sigma,

    Milwaukee, WI, USA). The solution extract was used as is,

    after filtration through 0.45m pore size polyvinilydene flu-

    oride (PVDF) filters, for the HPLC analysis. The solutions can

    be kept in a freezer for at least 1 week without affecting the

    phenols content. For the GC/MS analysis a sample prepara-

    tion using solid-phase extraction (SPE) was necessary before

    the chromatographic step.

    2.2. HPLC/fluorescence detection analytical procedure

    The analysis by HPLC/fluorescence followed a procedure

    described in the literature [3]. One difference in the present work

    was the use of a fixed amount of extracting solution (25 mL)

    Fig. 1. HPLC/fluorescence chromatogram obtained for the main phenols in

    mainstream smoke of a 2R4F Kentucky reference cigarette.

    instead of adjusting this volume to obtain a solution contain-

    ing approximately 1 mg/mL of total particulate matter (TPM).

    The ascorbic acid in the present method was added to prevent

    any potential oxidation of the phenols. Also, a chromatographic

    standard (further described) was used in the present work. Theinstrumentation used for the HPLC analysis consisted of two

    Waters 510 pumps, a 717plus Waters autosampler and a 474

    Waters scanning fluorescence detector (Waters, Milford, MA,

    USA). The separation was achieved on a Beckman Ultrasphere

    ODS column 15 cm4.6mmI.D.,5m particle size (Beckman

    Coulter, Fullerton, CA, USA). The separation took place under

    gradient conditions using two solutions, one being water with

    1% acetic acid (sol. A) and the other acetonitrile with 1% acetic

    acid (sol. B). The acetic acid is necessary for improving the sta-

    bility of phenols fluorescence [3]. The initial solution contained

    4% sol. B (and it is not pure water) in order to avoid the outgas-

    ing commonly seen when mixing pure water and acetonitrile.The flow rate was 1.4 mL/min. The composition of the mobile

    phase was changed (linear) to reach 31% sol. B at 10.5 min and

    100% sol. B at 15.5 min. Sol. B was flown through the column

    for another 4.5 min and then the conditions were restored to the

    initial mobile phase composition. Multiple injections were done

    every 27 min. Thefluorescencewas measured initially at 304 nm

    ex and 338 nmem. The conditions were changed to 274 nmexand 298 nm em after 3.5 min, and to 232 nm ex and 310 nm

    em after 13.2 min. The data acquisition was performed for

    20 min. Theinjection volume was8L. A typical chromatogram

    obtained in these conditions for the mainstream smoke of the

    2R4F Kentucky reference cigarette is shown in Fig. 1.

    For the quantitation of phenols, calibration curves were

    generated using a series of four standards. The standards

    Table 1

    Description of Kentucky reference and commercial cigarettes

    Descriptor 1R5F Kentucky ref. 2R4F Kentucky ref. 16.2 mg tar 10.6 mg tar 5.0 mg tar

    FTC tar (mg/cig) 1.7 8.9 16.2 10.6 5.0

    Cigarette length (mm) 84 84 83 83 83

    Filter length (mm) 32 27 21 27 27

    Filter ventilation (%) 72 28 23 32 54

    Blend type American American American American American

    Nicotine (mg/cig) 0.16 0.75 1.31 0.92 0.5

    CO (mg/cig) 3 12.0 13.9 10.7 7.4

  • 7/28/2019 Gcms and Hplc Methods

    3/8

  • 7/28/2019 Gcms and Hplc Methods

    4/8

    S.C. Moldoveanu, M. Kiser / J. Chromatogr. A 1141 (2007) 9097 93

    Table 3

    List of standards of phenols, their retention times, and m/z values used for the

    detection/quantitation

    No. Compound Retention

    time (min)

    m/z

    1 Phenol 6.88 166

    2 o-Cresol 8.57 180

    3 m-Cresol 8.76 1804 p-Cresol 9.08 180

    5 2-Ethylphenol 10.28 194

    6 2,5-Dimethylphenol 10.70 194

    7 3,5-Dimethylphenol 11.07 194

    8 2,4-Dimethylphenol 11.20 194

    9 2-Methoxyphenol 11.28 196

    10 4-Ethylphenol 11.59 194

    11 (I.S.) 4-Chlorophenol 11.71 185

    12 2,6-Dimethylphenol 11.79 194

    13 2,3-Dimethylphenol 12.02 194

    14 3,4-Dimethylphenol 12.32 194

    15 3-Methoxyphenol 13.17 196

    16 4-Methoxyphenol 13.47 196

    17 Catechol (1,2-dihydroxybenzene) 13.88 254

    18 Resorcinol (1,3-dihydroxybenzene) 16.05 254

    19 4-Methylcatechol 16.27 268

    20 Hydroquinone (1,4-dihydroxybenzene) 16.73 254

    21 3-Methylcatechol 16.71 268

    22 3-Methylresorcinol 18.19 268

    23 2-Methylresorcinol+ methylhydroquinone 18.66 268

    24 4-Ethylresorcinol 19.90 282

    25 2,5-Dimethylresorcinol 20.18 282

    the ions (m/z) used for the measurement of TMS derivatives.

    The standards solution was processed through the SPE similar

    to a smoke extract (2 mL added on the cartridge). The individ-

    ual concentrations in the standards solution varied between 2.0and 2.5g/mL. The SIM chromatogram for an extract of smoke

    from a 2R4F cigarette generated the chromatogram shown in

    Fig. 3.

    In addition to the dimethyl- and/or ethyl-dihydroxybenzenes

    (C2-dihydroxybenzenes) for which standards were available,

    several other peaks in the chromatogram shown in Fig. 3, eluting

    between 20.5 and 22.5 min were identified based only on their

    spectrum as C2- or C3-dihydroxybenzenes (C3 indicating any

    alkyl with three carbon atoms). Because of the similarity of the

    spectra of these compounds, the exact position of the substitu-

    Fig. 3. SIM chromatogram of the phenols from the smoke of a 2R4F cigarette.

    Peak identification based on retention times given in Table 3.

    tion on the benzene ring was not possible (no standards were

    obtained for these compounds).

    The quantitation of phenols by the GC/MS technique was

    also done with calibration curves. The calibration curves for

    hydroquinone, catechol, resorcinol, phenol, p-cresol, m-cresol,

    o-cresol, and guaiacol were generated with the same series of

    four standards used for the calibration of HPLC method. Curves

    representing quantity versus peak areas normalized by the area

    of the internal standard (4-chlorophenol) were generated. The

    curves were linear and theR2 values for the dependence were all

    above 0.993. For the other phenols only three point calibrations

    were generated (phenols from Aldrich/Sigma, Milwaukee, WI,

    USA).

    3. Results and discussion

    Several aspects of the two analytical techniques for phenols

    analysis, the HPLC and the GC/MS, were investigated with the

    purpose of their comparison. These aspects included selectiv-

    ity, accuracy, precision, limit of detection and of quantitation,recovery, and robustness. A comparison of advantages and dis-

    advantages of the two techniques was possible following the

    description of these parameters.

    3.1. Methods selectivity

    The selectivity of the HPLC procedure was difficult to ver-

    ify, since no sample was available containing the same matrix

    but without phenols. However, potential interferences from the

    matrix are not likely for the phenols found in relatively high

    levels in smoke. The extraction with water of smoke conden-

    sate eliminates many compounds that are not water soluble, andthe fluorescence of phenols is selective. The separation of p-

    cresol from m-cresol was not achieved, and the results for the

    two compounds were given as their sum. Some changes in the

    elution gradient that modified the retention times of the analytes,

    as well as the use of a longer column (Ultrasphere ODS 5 m,

    4.6 mm25 cm) did not change peak areas for the analytes for

    samples generated from a 2R4F Kentucky reference cigarette

    as well as those from the analyzed commercial cigarettes. This

    basically indicated that the measured peaks are pure.

    Selectivity was enhanced for the GC/MS procedure com-

    pared to the HPLC. The SPE processing of the sample provided

    besides concentration and solvent change, a cleanup step. The

    availability of the mass spectra for identification and the use ofselected ions for quantitation also increased the method selec-

    tivity. This allowed the separation of cresols (m- and p-cresol

    were not separated using the HPLC procedure) and the peaks

    of other hydroxybenzenes that were not analyzed by HPLC

    were very well separated in the GC/MS method (see Fig. 2).

    Only 2-methylresorcinol and methylhydroquinone had identi-

    cal retention times and very similar mass spectra and were not

    separated in the conditions described in this study.

    A potential interference in the measurement of hydro-

    quinone for both GC/MS and HPLC procedures may occur if

    p-benzoquinone is present. This substance is reduced to hydro-

    quinone in the presence of ascorbic acid. In order to prove that

  • 7/28/2019 Gcms and Hplc Methods

    5/8

    94 S.C. Moldoveanu, M. Kiser / J. Chromatogr. A 1141 (2007) 9097

    Table 4

    Comparison for the results of phenols from 2R4F cigarette obtained by the HPLC procedure using solution for pad extraction with and without 0.1% of ascorbic acid

    Compound Average g/cig. HPLC

    with ascorbic acid

    RSD % Average g/cig. HPLC

    without ascorbic acid

    RSD % Difference %

    Phenol 9.21 2.3 9.19 3.5 0.22

    o-Cresol 2.34 6.1 2.14 2.8 8.93

    m +p-Cresol 6.95 4.6 6.77 4.7 2.62

    Catechol 35.60 1.9 35.81 3.8 0.59Hydroquinone 28.80 4.9 28.71 3.8 0.31

    Resorcinol 0.84 10.6 0.79 8.9 6.13

    Guaiacol 1.61 4.0 1.52 4.5 5.75

    Table 5

    Comparison with literature data for the results obtained using the HPLC and the GC/MS procedures for the phenols from 2R4F and from 1R5F cigarettes

    2R4F 1R5F

    Compound Average

    g/cig.

    HPLC

    RSD %

    HPLC

    Average

    g/cig.

    GC/MS

    RSD %

    GC/MS

    Ref. [38]

    g/cig.

    Average

    g/cig.

    HPLC

    RSD %

    HPLC

    Average

    g/cig.

    GC/MS

    RSD %

    GC/MS

    Ref. [10]

    g/cig.

    Phenol 9.21 2.3 7.49 4.5 7.32 1.18 8.6 0.76 3.5 0.9

    o-Cresol 2.34 6.1 2.45 2.6 1.89 0.47 13.1 0.35 0.8 0.2m +p-Cresol 6.95 4.6 6.22 2.7 5.84 0.74 4.6 0.62 2.4 0.3

    Catechol 35.60 1.9 35.88 1.8 37.90 6.42 8.2 6.90 3.8 7.4

    Hydroquinone 28.80 4.9 32.74 5.7 32.40 6.58 8.2 6.30 3.4 4.9

    Resorcinol 0.84 10.6 1.27 7.9 0.91 0.24 23.5 0.51 4.5 0.6

    Guaiacol 1.61 4.0 1.42 3.2 Not available 0.37 13.6 0.20 6.1 Not available

    the addition of 0.1% ascorbic acid in the extraction solution

    of phenols does not affect the results regarding hydroquinone

    (or other phenols), the analysis for the 2R4F cigarette was

    done by the same procedure as previously described without

    the addition of ascorbic acid in the Cambridge pad extract-

    ing solution [3]. The results are shown for triplicate samples

    with each chromatographic runs performed twice (total of six

    data points) in Table 4 for the HPLC method only. The agree-

    ment between the data is very good. Further comparison of the

    accuracy of the results obtained with the present procedure for

    2R4F and 1R5F Kentucky reference cigarettes with the litera-

    ture data also proved to be very good. This indicates that the

    equilibrium quinone/hydroquinone is strongly displaced toward

    hydroquinone in mainstream cigarette smoke.

    3.2. Methods accuracy

    The results obtained for 2R4F and 1R5F Kentucky refer-

    ence cigarettes were compared with the data from the literature[10,35] ([3] shows data only for 1R4F Kentucky reference

    cigarette). Only results for seven common phenols in cigarette

    smoke were available for comparison. The results are given in

    Table 5. For this comparison, triplicates of each sample were

    smoked and processed, followed by chromatographic runs per-

    formed twice for each sample (total of six data points). The

    results from Table 5 show good agreement between the HPLC

    results, the GC/MS results, and also the results published in lit-

    erature ([38] provides data from a collaborative study between

    six laboratories). Larger discrepancies were noticed only in the

    results for resorcinol, which is present in cigarette smoke at

    lower levels compared to the other phenols shown in Table 5.

    This indicates that both the HPLC and the GC/MS procedure

    have good accuracy.

    3.3. Precision, limit of detection and of quantitation

    The RSD values given in Table 5 show that both the HPLCandGC/MS procedures havegood precision. Forthe HPLC anal-

    ysis of the 1R5F cigarette the RSD values were higher than

    those obtained by the GC/MS procedure. The fluorescence sig-

    nal for the 1R5F samples was relatively low compared to the

    noise, which may explain the increased RSD values. However, a

    lower volume of solution used to extract the Cambridge pads can

    increase the concentration of phenols in the analyzed solution

    and, therefore can possibly improve the repeatability of the tech-

    nique. Calculated as 3SD (where SD is the standard deviation

    for a sample with low levels of phenols), the limit of detection

    (LOD) obtained from Table 5 is given in Table 6. The limits

    of detection given in Table 6 can be considerably improved if

    certain changes are performed in the sample processing. For the

    Table 6

    Limit of detection calculated as 3SD, where SD is the standard deviation for

    a low sample

    Compound LOD g/cig. HPLC LOD g/cig. GC/MS

    Phenol 0.30 0.08

    o-Cresol 0.18 0.01

    m +p-Cresol 0.10 0.04

    Catechol 1.58 0.79

    Hydroquinone 1.62 0.64

    Resorcinol 0.17 0.07

    Guaiacol 0.15 0.04

  • 7/28/2019 Gcms and Hplc Methods

    6/8

    S.C. Moldoveanu, M. Kiser / J. Chromatogr. A 1141 (2007) 9097 95

    Table 7

    Other phenols detected in the mainstream cigarette smoke of 2R4F and 1R5F cigarettes

    No. Compound Average 2R4F g/cig. RSD % 2R4F Average 1R5F g/cig. RSD % 1R5F

    1 m-Cresol 1.76 2.26 0.15 0.49

    2 p-Cresol 4.46 2.68 0.47 2.41

    3 2-Ethylphenol 0.57 2.40 0.17 1.10

    4 2,5-Dimethylphenol 0.57 1.44 0.14 4.15

    5 3,5-Dimethylphenol 0.62 2.28 0.16 11.906 2,4-Dimethylphenol 1.09 1.16 0.26 3.35

    7 4-Ethylphenol 1.43 0.98 0.28 3.56

    8 2,6-Dimethylphenol 0.58 11.35 0.23 14.72

    9 2,3-Dimethylphenol 1.33 1.16 0.14 3.19

    11 3,4-Dimethylphenol 0.50 11.96 0.15 7.10

    12 3-Methoxyphenol 0.28 9.44 0.13 4.90

    13 4-Methoxyphenol 0.46 12.23 0.22 2.52

    14 4-Methylcatechol 4.32 4.78 0.88 3.48

    15 3-Methylcatechol 4.52 3.42 0.93 4.04

    15 3-Methylresorcinol 0.83 11.06 0.30 2.00

    17 2-Methylresorcinol+ methylhydroquinone 3.91 2.11 0.65 5.96

    18 4-Ethylresorcinol 0.40 6.87 0.17 6.85

    19 2,5-Dimethylresorcinol 1.26 7.33 0.33 7.12

    20 Other C2-dihydroxybenzenes 3.97 8.10 0.49 9.10

    21 Sum of C3-dihydroxybenzenes 0.63 4.33 0.27 5.97

    GC/MS technique for example, the volume of the solution of

    particulate phase smoke extract added to the SPE cartridge can

    be increased from 2 mL to a higher volume, without resulting

    in any breakthrough of the analytes. However, for the analysis

    of phenols in mainstream cigarette smoke this increase was not

    considered necessary. The values for the limit of quantitation

    (LOQ) can be evaluated as 3LOD.

    3.4. Recovery

    The recovery of the HPLC procedure has been previ-

    ously reported [3]. The efficiency of phenols extraction from

    the cigarette smoke particulate matter (TPM) collected on a

    Cambridgepad using a 1% aceticacidaqueous solution was thor-

    oughly evaluated in this previously reported HPLC technique

    [3]. For this reason, the dissimilarity between the extraction of

    phenols from TPM or from the clean pad was not expected to

    generate considerable differences. The recovery for the GC/MS

    wasperformedbyaddingonaCambridgepad100L solution of

    a mixture of standards containing between 500 and 625 g/mL

    of individual phenols. The pad was further processed by the

    procedure previously described, and the results compared to theinitial amount added. Recoveries between 93 and 105% were

    obtained for all the phenols analyzed. This indicated that no

    recovery problems are encountered during pad extraction and

    the SPE step.

    3.5. Robustness

    Robustness refers to the quality of an analytical procedure

    to not be influenced by small experimental modifications dur-

    ing its performance. Both HPLC and the GC/MS techniques

    described in this study were simple and no difficult steps were

    encountered. Some attentionis required in the GC/MS procedure

    during the drying of the SPE cartridge. The presence of water

    in the DMF eluate, noticeable by strong heating of the sam-

    ple when the reagent BSTFA is added for derivatization, leads

    to lower peaks in the chromatogram and incorrect quantitation

    (slight heating of thesamples still occurseven with drysamples).

    However, after 1 h of drying the SPE cartridge with ambient air,

    the potential problem of hydrolyzing the TMS derivatives was

    eliminated.

    3.6. Comparison of HPLC and GC/MS procedures

    Both HPLC and GC/MS procedures give good results for the

    analysis of the major phenols from cigarette mainstream smoke.

    The HPLC procedure is simpler since after the pad extraction,

    the solutions are directly subject to the chromatographic pro-

    cess. A slightly lower precision noticed during the analysis of

    the 1R5F cigarette by the HPLC method is not critical, and can

    be improved by using smaller extraction volumes of the Cam-

    bridge pad. The advantage of the GC/MS technique consists

    mainly in the extension of the list of analytes and in the separa-

    tion of m- and p-cresols. Also, the detection based on the ions

    characteristic for each phenol further eliminates the chances forinterferences. Theresults for the level of other phenols present in

    smoke of the 2R4F and 1R5F Kentucky reference cigarettes as

    obtained by the GC/MS technique are given in Table 7. As seen

    from Table 7, only the C1-dihydroxybenzenes and to a lower

    extent C2-dihydroxybenzenes are present in smoke of the two

    Kentucky reference cigarettes at appreciable levels compared

    to those of the major phenols. Most other phenols in smoke

    besides phenol, catechol, hydroquinone, and cresols are present

    at lower levels, however their total sum accounts for about 25%

    of total phenols in smoke. The same conclusion was obtained

    from the analysis of smoke from other cigarettes. As an exam-

    ple, the analysis of phenols from the mainstream smoke of three

  • 7/28/2019 Gcms and Hplc Methods

    7/8

    96 S.C. Moldoveanu, M. Kiser / J. Chromatogr. A 1141 (2007) 9097

    Table 8

    Comparison of phenol levels in three common commercial cigarettes

    Cigarette

    No. Compound 5.0 mg tar 10.6 mg tar 16.2 mg tar

    Average g/cig. RSD % Average g/cig. RSD % Average g/cig. RSD %

    1 Phenol 4.86 2.85 9.29 5.30 16.07 4.26

    2 o-Cresol 1.23 0.03 2.62 4.14 3.70 4.96

    3 m-Cresol 0.93 1.91 1.97 2.25 2.95 4.44

    4 p-Cresol 2.12 2.69 5.02 3.25 6.97 4.69

    5 2-Ethylphenol 0.31 1.60 0.70 3.97 0.75 2.85

    6 2,5-Dimethylphenol 0.29 2.63 0.67 2.98 0.79 3.72

    7 3,5-Dimethylphenol 0.36 9.59 0.82 2.04 1.00 3.40

    8 2,4-Dimethylphenol 0.56 2.68 1.28 2.54 1.47 2.61

    9 2-Methoxyphenol 0.65 4.05 1.43 2.31 2.40 2.48

    10 4-Ethylphenol 0.77 3.32 1.67 1.78 2.10 3.67

    11 2,6-Dimethylphenol 0.30 7.55 0.73 9.33 0.68 2.43

    12 2,3-Dimethylphenol 0.54 6.29 0.84 11.34 1.62 1.32

    13 3,4-Dimethylphenol 0.29 3.10 0.60 5.95 0.73 4.24

    14 3-Methoxyphenol 0.15 6.42 0.36 9.53 0.33 11.61

    15 4-Methoxyphenol 0.25 11.88 0.63 5.36 0.53 3.36

    16 Catechol 20.00 4.61 45.20 2.48 62.35 1.58

    17 Resorcinol 0.72 9.86 1.81 8.48 1.80 1.01

    18 4-Methylcatechol 2.43 4.52 6.19 6.08 7.59 2.13

    19 Hydroquinone 19.78 6.36 43.00 6.41 56.44 1.81

    20 3-Methylcatechol 2.37 3.74 5.58 1.75 6.50 2.92

    21 3-Methylresorcinol 0.46 7.94 1.16 7.57 1.19 0.58

    22 2-Methylresoecinol+ methylhydroquinone 1.83 7.89 3.57 13.11 5.70 2.32

    23 4-Ethylresoecinol 0.21 5.23 0.53 7.51 0.47 2.22

    24 2,5-Dimethylresorcinol 0.63 7.09 1.49 9.84 1.73 3.28

    25 C2-Dihydroxyphenol 1.29 10.94 2.93 11.10 4.65 2.76

    26 C3-Dihydroxyphenol 0.34 12.96 0.57 4.61 0.81 1.99

    Total phenols 63.69 140.68 190.29

    common commercial cigarettes of different tar levels was per-formed. The cigarettes included a 5.0 mg FTC tar cigarette,

    (where tar is defined as the weight of total mainstream smoke

    condensate (or wet particulate matter) minus the weight of nico-

    tine and water) a 10.6 mg tar and a 16.2 mg tar cigarette, with

    the description given in Table 1. The results for the phenols anal-

    ysis are given in Table 8. As it can be calculated from Table 8

    the proportion of different phenols relative to the total amount of

    phenols in smoke is notvery differentfor differentcigarettes,and

    phenol, catechol, hydroquinone, and cresols account for about

    75% of total phenols in smoke.

    4. Conclusions

    This study presents a new GC/MS analytical technique for

    phenol analysis, which allows the measurement of 24 individ-

    ual compounds and of the sum of certain C2-dihydroxybenzenes

    and of C3-dihydroxybenzenes. The main characteristics of the

    GC/MS analytical procedure are compared to those of a HPLC

    technique with fluorescence detection. Both techniques pro-

    vide very reliable results for the analysis of phenols. Both

    methods are equally accurate for the analysis of 67 major

    phenolic compounds from mainstream cigarette smoke. Using

    the GC/MS procedure it was found that besides the phenols

    typically analyzed in cigarette smoke some other phenols are

    present, accounting together for about 25% of total phenols inmainstream smoke. However, the GC/MS method requires two

    additional steps compared to the HPLC method (the SPE and the

    derivatization). Also, the GC/MS equipment is typically more

    expensivethan theHPLC one. Forroutine laboratoriesthe HPLC

    technique is probably recommended, while the GC/MS is nec-

    essary only when a detailed analysis of phenols in mainstream

    cigarette smoke is desired.

    References

    [1] D.L. Davis, M.T. Nielsen, Tobacco, Production, Chemistry and Technol-ogy, Blackwell Science, Cambridge, UK, 1999, p. 281.

    [2] D. Hoffmann, I. Hoffmann, Beitr. Tabakforsch. Int. 18 (1998) 49.

    [3] C.H. Risner, S.L. Cash, J. Chromatogr. Sci. 28 (1990) 239.

    [4] Y. Hoshika, J. Chromatogr. 144 (1977) 181.

    [5] T.J. Clark, J.E. Bunch, J. Chromatogr. Sci. 34 (1996) 272.

    [6] E.T. Oakley, J.O. Millham, L. Weissbecker, Anal. Chim. Acta 31 (1964)

    272.

    [7] E.J. Nanni, M.E. Lovette, R.D. Hicks, K.W. Fowler, M.F. Borgerding, J.

    Chromatogr. 505 (1990) 365.

    [8] B.A. Tomkins, R.A. Jenkins, W.H. Griest, R.R. Reagan, S.K. Holladay, J.

    Assoc. Off. Anal. Chem. 67 (1984) 919.

    [9] R.E. Fresenius, J. Anal. Appl. Pyrol. 8 (1985) 561.

    [10] J.B. Forehand, G.L. Dooly, S.C. Moldoveanu,J. Chromatogr. A 898 (2000)

    111.

    [11] Z. Zhi, A. Rios, M. Valcarcel, Analyst 121 (1996) 1.

  • 7/28/2019 Gcms and Hplc Methods

    8/8

    S.C. Moldoveanu, M. Kiser / J. Chromatogr. A 1141 (2007) 9097 97

    [12] P. MuBmann, K. Levsen, W. Radeck, Fresenius J. Anal. Chem. 348 (1994)

    654.

    [13] B. Gawdzik, J. Gawdzik, U. Czerwinska-Bil, J. Chromatogr. 509 (1990)

    135.

    [14] M.T. Galceran, O. Jauregui, Anal. Chim. Acta 304 (1995) 75.

    [15] M. Saraji, M. Bakhshi, J. Chromatogr. A 1098 (2005) 30.

    [16] A. Vermeulen, K. Welvaert, J. Vercammen, J. Chromatogr. A 1071 (2005)

    41.

    [17] F.E. Suliman, S.S. Al-Kindi, S.M. Al-Kindy, H.A. Al-Lawati, J. Chro-matogr. A 1101 (2006) 179.

    [18] F. Zhou, X. Li, Z. Zeng, Anal. Chim. Acta 538 (2005) 63.

    [19] M. Naczk, F. Shahidi, J. Chromatogr. A 1054 (2004) 95.

    [20] L.-K. Ng, P. Lafontaine, J. Harnois, J. Chromatogr. A 873 (2000) 29.

    [21] R.C. Mejas, R.N. Marn, M.G. Moreno, C.G. Barroso, J. Chromatogr. A

    995 (2003) 11.

    [22] K. Robards, J. Chromatogr. A 1000 (2003) 657.

    [23] R. Gatti, M.G. Gioia, A.M. Di Pietra, V. Cavrini, Anal. Chim. Acta 447

    (2001) 89.

    [24] S.C. Moldoveanu, V. David, Sample Preparation in Chromatography, Else-

    vier, Amsterdam, 2002, p. 669.

    [25] K. Ogan, E. Katz, Anal. Chem. 52 (1981) 160.

    [26] L. Yang, Z. Wang, L. Xu, J. Chromatogr. A 1104 (2006) 230.

    [27] L. Zhang, L. Zhang, W. Zhang, Y. Zhang, Anal. Chim. Acta 543 (2005) 52.

    [28] T. Heberer, H.-J. Stan, Anal. Chim. Acta 341 (1997) 21.

    [29] M.F. Borgerding, J.A. Bodnar, H.L. Chung, P.P. Mangan, C.C. Morrison,

    C.H.Risner, J.C.Rogers,D.F.Simmons, M.S.Uhrig, F.N. Wendelboe, D.E.

    Wingate, L.S. Winkler, Food Chem. Toxicol. 36 (1997) 169.

    [30] M.V. Bruijnsvoort, S.K. Sanghi, H. Poppe, W.T. Kok, J. Chromatogr. A

    757 (1997) 203.

    [31] R.L. Stedman, Chem. Rev. 68 (1968) 153.

    [32] H.C. Pillsbury, C.C. Bright, K.J. OConnor, F.H. Irish, J. Assoc. Off. Anal.

    Chem. 52 (1969) 458.[33] Routine Analytical Cigarette Smoking Machine-Definitions and Standard

    Conditions. International Standard Organization, Geneva, 3308 (E): ISO

    1991, p. 1.

    [34] Tobacco and Tobacco ProductsAtmosphere for Conditioning and Test-

    ing. International Standard Organization, Geneva, 3402 (E): ISO 1991, p.

    1.

    [35] G.N. Connoly, H. Saxner, Memorandum, The Commonwealth of Mas-

    sachusetts, Executive Office of Health and Human Services, MA, 19

    August1997.

    [36] CigarettesDetermination of Totaland Nicotine FreeDry Particulate Mat-

    ter Using a Routine Analytical Smoking Machine. International Standard

    Organization, Geneva, 4387 (E): ISO 1991, p. 1.

    [37] I. Rodrguez, M.P. Llompart, R. Cela, J. Chromatogr. A 885 (2000) 291.

    [38] P.X. Chen, S.C. Moldoveanu, Beitr. Tabakforsch. Int. 20 (2003) 448.