Page 1 OPERATING INSTRUCTIONS [GROWTH CHAMBER] Model : GC-300/1000 Manual No : 00HAA0002010 (Rev. 3)

Transcript of GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not...

Page 1: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 1

OPERATING INSTRUCTIONS

[GROWTH CHAMBER] Model : GC-300/1000

Manual No : 00HAA0002010 (Rev. 3)

Page 2: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 2

◆ NOTICE.

Thank you for purchasing Jeio Tech’ s product. This operation manual forms a definition of warning marks according to the level of importance and danger in order to use the product safely and correctly and prevent the users from accidents or injuries. Hence, please use the product in accordance with the instructions.

JEIOTECH CO., LTD

Page 3: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 3

■ Notice for the manual

1. Always keep this manual near to the instrument.

2. Check the manual in detail for correct operation

3. This manual is only for instruction of how to operate the instrument. Therefore,

anything out of the contents of the manual is not guaranteed.

4. It is prohibited to print and distribute the manual partly or entirely without

leave.

5. Please contact where you purchased the instrument if faults are found in the

manual.

■ Alert mark for safety operation

1. The alert marks are for safety operation and protect user and instrument from

damage.

2. Each mark is identified following to the degree of risks or importance.

User can damage seriously, or lose his or her life

if they operate the instrument wrongly.

User can damage or the instrument can be damaged if

they operate the instrument wrongly.

This is indicated as “ground connection"

The instrument should be set on a plat part.

This is for reference

3. Pay attention to the contents of alert marks.

4. Specially important “alert marks" are attached on the main body.(On the door)

5. Please change the alert marks if they are worn out

☞ Ask alert marks to where the instrument is purchased from if they are needed to

replace.

Page 4: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 4

■ Notice for safety and alteration

1. Read the manual carefully for safety before starting to operate the instrument.

2. Any damage that is happened due to carelessness is out of the manufacturer's

responsibility.

3. It's prohibited to alter the instrument.

4. Departing, fixing, and altering the instrument at your will can cause electric

shock, fire, and dysfunction.

5. Please contact where you purchased the instrument when some necessary parts should

be changed.

6. External shock to this instrument can cause damage and dysfunction.

■ Exemption

1. Any claim which is out the quality guaranteed by the manufacturer is our of the

manufacturer's responsibility.

2. Any damage that is from unexpected fault or damage of user by Acts of God is out

of Manufacturer's responsibility.

Page 5: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 5

◈ Unpacking your growth chamber

Please follow the orders as below.

1. Check the instrument.

1. Check the package condition before unpacking

2. Unpack carefully

3. Check the instrument carefully if damage is happened while transportation.

4. Check the accessories and related parts.

2. How to deal with damaged items

1. Please contact where you purchased the instrument if there is damage to the

instrument.

2. Keep the instrument as you received for the first time and check with the

transporter.

3. Issue claims as soon as you can to where you purchased the instrument.

3. Warranty for transportation

1. Responsibilities for damages happened while transportation is up to transporters.

2. The manufacturer will deal with any damage except for periods of transportation.

3. The instrument that is not transported by manufacturer's entrusted transporter

is out of manufacturer's responsibility.

Page 6: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 6

4. Composition of the unit 1) GC-300

<Image> <Description / Quantity>

Unit

1EA

Operating instructions

1EA

Shelves

3 EA

Shelf bracket

12EA

Industrial plug

/socket

20A / 1EA

Silicon hose

0.5M caliber 8mm / 2EA

1.5M caliber 10mm / 1EA

Page 7: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 7

Water tank

10 L / 1EA

Water tank prop

1EA

2) GC-1000

<Image> <Description / Quantity>

Unit

1EA

Operating instructions

1EA

Selves

6 EA

Shelf bracket

24EA

Page 8: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 8

Industrial plug/ socket

30A / 1EA

Silicon hose

0.5M caliber 10mm / 3EA

1.5M caliber 10mm / 1EA

Upper light

1 SET

Metal lamp

400W / 3EA

Recorder instructions

Optional / 1EA

Page 9: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 9

CONTENTS ◆ NOTICE …………………………………………………………………………………………… 3

◆ Transportation …………………………………………………………………………………… 5

◆ Composition of the unit ……………………………………………………………………… 6

OPERATING MANUAL

1 Alert and notice for Safety ………………………………………………………………… 10

2 Installation of the composition …………………………………………………………… 14

3.3-1 General purposes ……………………………………………………………………… 16

3-2 Features ……………………………………………………… ………………………… 17

4. Name of parts

4-1. GC-300 ……………………………………………………………………………………… 19

4-2. GC-1000 ……………………………………………………………………………………… 20

4-2. Lamp ………………………………………………………………………………………… 21

5. How to use Controller

5-1. Name & Operation ………………………………………………………………………… 22

5-2. How to operate Controller …………………………………………………………… 24

6. How to solve problems ……………………………………………………………………… 40

7. Daily Checking ………………………………………………………………………………… 41

8. How to Clean …………………………………………………………………………………… 42

9. Water tank usage …………………………………………………………………………… 44

10. Specification ………………………………………………………………………………… 46

11. Warranty ………………………………………………………………………………………… 47

12. How to waste ………………………………………………………………………………… 47

Page 10: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 10

Alert for safety

♣ Read the notice carefully

♣ This is for keeping user and their property from damage

Warning means that failure to follow this

safety statement could result in serious

personal injury, or death.

Check Voltage, Phase, and Capacity for

correct connection.

☞ Wrong connection can cause fire or

electric shock.

☞ Power cord should be connected to consent

with ground terminal

Don't insert a lot of plug into the

outlet at once!

☞ It causes fire and electric shock

Power with ground connection should be

used.

☞ Power without ground connection

causes serious damage to users or

equipment . Do not ground with gas or

water pipe.

Please check the power supply and

frequency on the ID plate before

installation.

☞ The unit requires a grounded and

polarized 220 VAC, single phase.

GC-300 : 16A, 2.0 SQ,

GC-1000 : 31A, 5.5SQ

Do not settle the equipment near to

places where gas can be leaked.

Do not use the equipment in places

where industrial oil, smoke, metal

dusts are existed

☞ That causes fire or electric shock

Off the power if the equipment has

strange sound and smell, smoke etc.

☞ Ask After Service.

It causes fire and electric shock.

v

◈ Setting

1

Page 11: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 11

Alert for safety

♣ Check the list as below during setting and operation.

♣ This is for keeping user and their property from damage.

Warning means that failure to follow this

safety statement could result in serious

personal injury, or death.

Do not put explosive or inflammable

materials into the chamber

☞ The equipment is not explosion-

proof structure. Sample can be

exploded in the chamber

Do not keep inflammable materials near

to the instrument.

☞ It can cause fire or explosion.

Do not depart, fix, and alter the

instrument by yourself.

☞ It can cause fire or electric shock

☞ It can cause dysfunction of the

instrument.

☞ Use only for the original purpose.

Do not put heavy things on the power

line. Do not put the equipment on the

line.

☞ It causes fire and electric shock.

Do not use the equipment near to places

where vibration occurs.

☞ It causes damage to equipment and is

possible that the equipment does not

works properly.

Do not let the product take any strong

shock or vibration.

☞ It causes abnormal operation or

trouble.

◈ Operating

1

Page 12: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 12

Notice for safety

♣ Check the list as below during setting and operation.

♣ This is for keeping user and their property from damage

Warning means that failure to follow this

safety statement could result in serious

personal injury, or death.

Do not clean the instrument with

solvent abstergent. Use smooth cloths.

☞ It can cause fire and deformity

Do not spread insecticide or

inflammable spray on the equipment.

☞ It causes damage to equipment ,

fire, electric shock as well as harms

to user.

Keep out of the direct sunlight.

☞ It may cause fire and abnormal

operation.

☞ When use in the direct sunlight,

please set up a proper interception

to make shade

Set the instrument apart from wet

materials.

☞ It can cause dysfunction of the

instrument and electric shock.

☞ Please ask After Service if the

instrument is wet.

Do not spread insecticide or

inflammable spray on the equipment.

☞ It causes damage to equipment ,

fire, electric shock as well as harms

to user.

Do not ask After Service to anyone who

is not from the manufacturer or its

entrusted party.

☞ It can cause dysfunction of the

instrument if the control is different

from Manufacturer's.

◈ Operating

1

Page 13: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 13

Notice for safety

♣ Check the list as below during setting and operation

♣ This is for keeping user and their property from damage

Warning means that failure to follow this

safety statement could result in serious

personal injury, or death.

▪ Setting place

Keep the equipment more than 1.5m from lighting and 20cm from the back. Make wide space for opening left and right side of the equipment. ☞ It can cause damage while changing fluorescent lamp.

Set the equipment on a plat place and do not heat or push down the equipment.

The equipment needs proper tool for

transportation.

■ Checking after settle down.

☞ Check whether the electric lines are at correct locations.

☞ Check whether the wire ground is correct

☞ Check whether there is spare place near the instrument. This is for setting

and after service of the instrument.

☞ Check the water level in the water tank and connect the tank to where it

should be located.

☞ Please operate the instrument after checking all the necessary part for

safety.

◈ Setting & operating

GC-300 : 250kg GC-1000 : 550kg

1

Page 14: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 14

Installation of the composition

2-1. Installing the GC-300 Water Tank

Following the bellow direction to install the water tank.

1. Setting the water tank prop on ②or ③ wherever you want with 6mm bolts.

2. Setting the water tank ① on where you want.

3. Connecting the water tank ① with the inlet ④ by using the silicon hose ⑤. (1.5M)

4. Pouring distilled water into the water tank ① and opening the valve.

(Using distilled water improves durability of the unit.)

5. Connecting two silicon hoses (0.5M) with part ⑥.

(On part ⑥, the right with a valve is a draining valve to drain water from the

inside of chamber. The left without a valve is a pipe for over flow.)

2

1

2

3

4

5

6

Page 15: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 15

2-2. Installing of the GC-1000 upper light

Following the below direction to install the upper light.

(Must take off the main plug in safety.)

1. Ready for the upper light first as the image <1>.

2. Put the upper light up on the upper side of unit as the image <2>, set the light

with bolts (4X10) that be set on the unit.

(Use a ladder with at least two people in safety.)

3. On the image <3>, open the upper cover (A) and put the lighting line in through the

hole (B).

4. As the image <4>, connecting the lighting line with (L1+, L1-, L2+, L2-, L3+, L3-,

FAN+, FAN-) from the left one in order. (blue= +, black= - )

5. Close the upper cover and open the lighting part <1> then set three metal lamps up.

<1>

<3> <4>

B A

C

<2>

Page 16: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 16

-1. General purposes

* Agriculture researches by using plant growth

* Plant tissue culturing

* Seeds germination researches

* Photosynthesis and light safety tests

* Growth tests for various animals and plants

* Preservation tests for food

* Genes and life science researches

* Increase, breeding, acclimation and photosynthesis tests of plants

* Marine plant, water microorganism, cultivation of insects and light safety tests

* Minute ocean currents cultivation

* Cultivation of sensitive microorganisms with heat and light

* Transmutation tests of plants

The equipment can make the optimal condition (temperature, humidity, intensity of

illumination) for the above various tests and researches like photosynthesis of plants

test.

3

Page 17: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 17

-2. Features

1. Accurate Temp. Control

▶ Perfect Temp. and heating control by PID Controlled Microprocessor Touch

Pad Type

2. Proper airflow for growing plants

▶ Air flowing from bottom to top by 2 fans (Top FAN(GC-300), bottom FAN(GC-

1000)

3. Controlling heat rising

▶ Minimizing heat rising by light and easy setting the condition of

illumination of day and night.

▶ Heat can be taken out through top part. (GC-1000)

4. Protect inner air from leaking out

▶ Outer part of chamber and door have been produced from Adiabatic material

and specially made silicon packing protects inner air from leaking out by

making close adherence between inner glass door and chamber.

5. Magnetic Packing

▶ Door with Magnetic Packing functions shock absorber and duplicate air

tightness.

6. Programming operation

▶ Ideal test condition by programming setting value by 10 grades for temp.

illumination, humid (GC-1000) . Repeatable test is possible till 999 Cycles.

7. Safety Circuit

▶ Over Temp. LED is on and power becomes off when the temp. is reached on Over

Temp. Limit.

3

Page 18: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 18

-2. Features

8. Corrosion-proof material

▶ Chamber material is made of Stainless Steel (SUS 304) and outer body is made

of Steel Plate with coating for protecting from corrosion.

9. Glass door for observation inside the chamber

▶ Tempered glass helps the door safe and durable. User can observe inside the

chamber without temp. change by opening only outer door.

10. Making perfect environmental condition

▶ Designed for maintaining stable condition inside the chamber

▶ Temp. illumination, humid. can be programmed by grades and illuminating

condition of day and night is controllable.

- GC-300 : Temp. illuminating program (Humid, CO2 OPTION)

- GC-1000 : Temp. illumination, humid program (CO2 OPTION)

3

Page 19: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 19

Name of each part

10 7

9 8

1 3

6

4

1. CONTROL PANEL

For inputting and checking setting

condition

2. Over temp. limit

For stopping heater when temp. is rising

over the set value unusually

3. Main Power Switch

On and off the main power switch.

4. Water Tank

For supplying water to control humidity.

5 Lamps

For supplying light into the chamber

6 Casters

Wheels For transporting and fixing the

equipment

7 Pair glass

Double glass for maintaining temp inside

the chamber

8 Filters

Separable filter can be easily cleaned.

(Clean filters once a week for maintaining

good condition of Compressor)

9 Shelf level adjustor

Shelves can be adjusted following to

sample size

10.Inner Door

User can observe inside the chamber

Tempered glass

11. Door

4-1. GC-300

4

11

2

5

Page 20: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 20

Name of each part

2

4

5

7 9

10

11

12

1. Recorder (OPTIONAL)

For recording Temp. and Humid. data

2. Control Panel

For inputting and checking setting condition

3. Over temp. limit

For stopping heater when temp. is rising over

the set value unusually.

4. Main Power Switch

On and off the main power switch

5. Lamps

For supplying light into the chamber

6. Casters

Wheels For transporting and fixing the

equipment

7. Pair glass

Double glasses for maintaining temp. inside

the chamber.

8. Door Limit Switch

To stop operating except for lamp when the

inner door is open.

9. Filters

Separable filter can be easily cleaned.

(Clean filters once a week for maintaining

good condition of Compressor)

10. Shelf level adjustor

Shelves can be adjusted following to sample

size.

11. Inner Door

User can observe inside the chamber Tempered

glass

12. Door

4-2. GC-1000

4

3

6

1

8

Page 21: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 21

Structure and feature of Lamp

1. Metal Halide Lamp

① Consuming power : 400W(per 1EA)

② Long life span, excellent light

efficiency.

③ Small heat loading capacity does

not affect samples. The instrument

is easy to be installed with small

maintenance cost.

④ Life span: 6000 hrs.

2. Fluorescent lamp

① High-grade light efficiency with

using economic-type electronic

safety device.

② Each lamp has it’ s own ON/Off

switch. - Illumination capacity can

be controlled step by step.

3. How to replace ramps.

① Open the top part by keeping the

handle.

② The lamp is very hot. Please wait

at least 10 min. after power-off.

③ When replacing fluorescent lamps,

turn it to clockwise rotation about

90°, and pull it to take off from

the socket.

④ Be careful when you replace the

lamps.

☞ High temp. “Burns" (Over 100℃)

Be careful when you replace the lamps. You can be

dead or have serious damage.

4-3. Lamp

(Showing Lamp part)

2

1

Page 22: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 22

Operating Controller

1. LCD display

For checking operating condition of Plant

Growth Chamber.

2. Condition display

1. LCD display

3. Controller buttons 4. Set value input

button

2. Condition display

5-1 Name and Operation

Showing temp. controlling heater operation.

Showing humid. Controlling heater operation

Showing CO2 supplying condition

5

Page 23: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 23

LAMP condition display

3. Controller operation button

4. Set value inputting button

Showing PROG/MANU operation mode

Showing Auto-tuning operation condition

Showing on/off condition of compressor

Showing whether or not Timer is used

Showing water level (Light is on and humi. Control

stops when water is lacked,)

START/STOP : START/STOP button

PROG/MANU : PROG/MANU MODE change button

MODE : Parameter setting button

ENTER/A-T: Set value changing button

☞ Auto-tuning starts when the button is put for 5 sec.

(In case of performing mode)

1~0 : Set value no. inputting button

◀ : Set value unit changing button

- : (-)mark inputting button

Page 24: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 24

Each set value is shown on SvT orderly.

SvT : Set temp. value

SvH : Set humid. value

SvC : Set CO2 density value

At First

You can see the message when

putting the instrument on.

Starting

Manual

Mode

Starting

Manual

Mode

.

Display shows the present condition of the

chamber when the instrument is on.

T : Present temp. in the chamber.

Co2 : CO2 Density

H : Present humid. In the chamber

Lux : Intensity of illumination in the

chamber

Tm : Remained time SvT : Set temp.

<Ref>

Manual Mode : 1 Step can be controlled

Program Mode : 10 Steps can be controlled

5-2 Controller operation

Page 25: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 25

1. Operation (start to control)

1. Put on the Main Power Switch on the right top part of the instrument.

2. Put on the Main Power Switch after putting on circuit breaker (This is for GC-1000)

1-1. Changing control mode

1-2. Performing screens for each of control mode.

Manual mode control screen

(Manual Mode) Screens

for

each

Control

Mode

Program mode control screen

The instrument works when you put the Start button

on in condition of Control stand-by

<Exchanging Program/manual mode>

PROG/MANU modes are exchanged by putting the button

for 5 sec.

ON

OFF

GC-300 GC-1000

Page 26: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 26

2. Operating manual mode

2-1. Putting set values in manual modes.

Exchanging

and

putting

Set values

Put the mode button on.

2-2. Setting control modes

<Ref>

T : Temp. input Co2 : CO2 input

H : Humid input

Tm : Time input F : Lamp input

(1 : ON, 0: OFF)

☞ Time, Temp. Humid. And CO2(GC-1000) values can be

set with and button or with Number

button.

☞ button is for moving of each unit.

☞ button is for moving of each item.

<Input set range>

Temp. : 0℃ ~ 50℃ Time : 00 : 00 ~ 99hr. 59min.

Humid. :50 ~ 90RH% CO2 density : 0 ~ 5000ppm(OPTION)

Values are returned to the present ones if they are not set within 1 min.

Page 27: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 27

2-3. Lamp setting

2-2. Set values are inputted following to Lamp diagram.

GC-300 LAMP diagram GC-1000 LAMP diagram

☞ Examples of Lamp setting(GC-300)

No. Lux(lx) Set Values Lamp Diagram

1 1,500 0000 010 2

2 3,500 0000 101 1,3

3 5,500 0000 111 1,2,3

4 6,000 0101 010 2,4,6

5 8,000 0101 101 1,3,4,6

6 10,000 1010 010 2,5,7

7 12,000 1010 101 1,3,5,7

8 14,000 1111 010 2,4,5,6,7

9 17,000 1111 101 1,3,4,5,6,7

10 20,000 1111 111 1,2,3,4,5,6,7

* Illuminating measurement- Values are measured from the middle of the chamber toward

topside.

<Lamp diagram>

1(No.8) 1(No.7) 1(No.6) 1(No.5) 1(No.4) 1(No.3) 1(No.2) 1(No.1)

☞ Location of ramps are shown orderly.

1(No.8) is not used with GC-300 and GC-1000

<Ref> 1=> Lamp On, 0 => Lamp Off.

Page 28: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 28

☞ Examples of Lamp setting(GC-1000)

No. Lux (lx) Set Values Lamp Diagram

1 5,000 0101 000 4, 6 or 5, 7

2 7,500 0000 010 2

3 9,500 1111 000 4, 5, 6, 7

4 12,500 0101 010 2/4, 6 or 5, 7

5 15,000 0000 101 1, 3

6 18,500 1111 010 2 / 4, 5, 6, 7

7 21,000 0101 101 1, 3 / 4, 6, or 5, 7

8 24,000 0000 111 1, 2, 3

9 28,000 1111 101 1, 3 / 4, 5, 6, 7

10 31,500 0101 111 1, 2, 3 / 4, 6 or 5,

7

11 35,000 1111 111 1, 2, 3 / 4, 5, 6, 7

* Illuminating measurement- Values are measured from the middle of the chamber toward

topside.

2-4. Operating Manual mode

2-3. Operate the instrument by putting “MODE"and“START"buttons after inputting set

values

1. Button “Mode” after inputting each control item

2. The instrument is operated by manual mode when “START"

key is buttoned.

Page 29: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 29

3. Example of performing manual mode

▶ The discription shows how to perform in Manual modes.

Please follow the orders as below

Main Power Switch should be ON.

1. Put MODE button

2. Input tmep, CO2, Humid. And time.

ex)T: 30.0℃, Co2: 600ppm, H: 60.5%, Tm : 01:00

3. F : Input No.1,3,5,7

ex) F: 1010101

(Using ◀, ENTER, and No. button )

4. Get out from the screen by putting MODE

button

Put START/STOP button.

.

The instrument works following to the inputted set

values.

T, Co2, H, and Lux on the display show Temp. Co2

density, Humid. and Illumination in the chamber.

Set values SvT, SvH, SvC are shown orderly.

● Put PROG/MANU button for 5 sec. if

you want to start with Program mode.

<Example of inputting set value>

This is how to operate the instrument by inputting fixed temp.

humid. Illumination, CO2 , and time as below.

ex) Temp.: 30.0℃, CO2: 600ppm, H: 60.5%, Time: 1hr

Illumination: 12000lx(Lamp No. 1,3,5,7)

Page 30: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 30

4. Operating Program Mode

4-1. Inputting set values in Program Mode

The mode is changed to Program Mode with beep sounds by putting

PROG/MANU button for 5 sec.

☞ Set values of program mode can be inputted while operation and

stop of the instrument.

(Referring to 4-4 parameter input)

Put MODE button

.

4-2. How to set program mode

<Ref>

T : Temp.input Co2 : CO2 input

H : Humid. input

Tm : Time input F : Lamp input

(1 : ON, 0: OFF)

Next STEP input screen is shown by putting button

Input values at Each STEP by the same way.

Inputting

Set value

☞ Time, Temp. Humid. And CO2(GC-1000) values can be set with

and button or with Number button.

☞ button is for moving of each unit.

☞ button is for moving of each item.

<Input set range>

Temp. : 0℃ ~ 50℃ Time : 00 : 00 ~ 99hr. 59min.

Humid. :50 ~ 90RH% CO2 density : 0 ~ 5000ppm(OPTION)

Values are returned to the present ones if they are not set within 1 min.

Lamp Input is the same as Manual mode (Please refer to 2-3 Lamp)

Page 31: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 31

4-3. Input and close of STEP

Maximum STEP nos. of the instrument is 10 ea.

If time is inputted as 00:00 at STEP 10, the steps before STEP 10 are inputted.

☞ Total 9 steps are inputted if time is inputted as 00:00 at STEP 10

Temp. set graph for each STEPs.

Page 32: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 32

4-4. Input parameter

Put MODE button after inputting “STEP"

The screen shows as below in case of inputting “STEP" value while operation

<Run Cycle No> Input Step repeating times.

Maximum: 999 times. Input “000" to repeat

infinitely.

<Lamp On Delay>

Lamp is on or off as indicated at the term

between steps. The unit is sec.

(**:This is for protecting plant growth from fast

change of illumination)

The screen shows as below when inputting STEP value after stopping operation.

<Pgm-On Delay>

This is shown when setting program on “Controller

Stop" condition.

The instrument works after the set time(Min.)

(Inputting“0000"doesn't works anything)

4-5. Performing Program Mode

1. In case of inputting “STEP" value while operation

1. Put “MODE" button after finishing each control item.

2. “Program mode" starts if you put “START" button.

Page 33: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 33

2. In case of inputting STEP value after stopping operation

4-6. Example of setting STEP input of Program Mode

If you want to repeat the step 60 times from 1 to 5

Example of setting to repeat 60 times

1. Input values with and or Nos. from step1 to Step 5.

2. Set values(Tm:) as 00:00 on Step 6.

3. “Run Cycle Number" mode is on if you put button.

Set “Run Cycle Number" as 60

4. Put button again

☞ Total Cycle(130) = STEP 5 ⅹRun Cycle Number(60)

4-7. Changing “Operation STEP" on Program Mode.

▶ This function is for changing present STEP while operating on program mode.

Select a STEP you want with putting ◀(SHIFT) button when you

want to change from program mode to present mode.

STEP 1 -> STEP 2 -> STEP 3 is selected repeatedly

when you put the SHIFT button

The instrument works automatically after the set time(min.) for

“Pgm-On Delay"if you put “MODE" button after inputting each

control item.

Step 1

Step 2

Step 3

Page 34: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 34

5. Example of operating Program Mode. 6.

▶ The discription shows how to perform in Program mode.

Please follow the orders as below

Put on Main Power Switch

1. Put “MODE" button

2. Input Temp. CO2, Humid. and Time.

ex) T: 60.0℃, Co2: 500ppm, H: 60.5%, Tm : 01:00

3. F : Input nos. 1,2,3,4,6

ex) F: 00101111

(Use ◀, ENTER, or No. button)

Next 5 step input screen is shown if you put

“MODE" button.

1. input Temp. CO2, Humid. and time.

ex) T: 30.0℃, Co2: 400ppm, H: 80.5%, Tm : 01:00

2. F : Input No.1,2,3

ex) F: 00000111

3. If second STEP input is done, input values of

next STEP by buttoning “MODE'

4. Input set values by 10 steps with the sam way.

● Put PROG/MANU button for 5 sec. if the

screen starts with manual mode

<Example of inputting set value>

This is how to operate the instrument by inputting fixed

temp. humid. Illumination, CO2 , and time as below.

<STEP 1>

Temp.: 30.0℃, CO2: 600ppm, H: 60.5%, Time: 1hr

Illumination: 14000lx(Lamp No. 1,2,3,4,6)

<STEP 2>

Temp: 30.0℃, CO2: 400ppm, H: 80.5%, Time: 1hr.

Illumination: 12000lx(Lamp No. 1,2,3)

Page 35: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 35

Parameter input screen is shown if STEP input is done

till 10 steps or input the time of STEP as 00:00

1. This is for setting repeating time of each

programmed STEP. Maximum 999 times can be

inputted. STEPs are repeated infinitely if you

input as “000"

ex) To set 10 STEP , input Run Cycle No as 005

10 STEP ⅹ 005 = 50 (Total Cycle)

2. Input Lamp-On delay time

ex) 60s : 60sec

3. Input Pgm-On delay time

ex) 0005 : 5min(The instrument works 5min. later)

4. Move to stand-by condition by putting

Button.

Put button.

This is operated automatically if you input Pgm-On

delay

T, Co2, H, Lux on the display shows present temp.

CO2 density, humid. and illumination of the

chamber.

The set values SvT, SvH, SvC are shown by turns.

The present STEP and cycle are shown by turns on

the display

Page 36: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 36

6. Operation and setting of timer. 6-1. Operating theory of timer

Timer works automatically by the set value when you put “START"

button.

1. Timer works when it satisfies some cases as below.

* If present temp. is under set value. (Set values(Tm:) as 01:00)

* If present temp is over set temp. (Set values(Tm:) as 01:00)

Page 37: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 37

6-2. Setting timer

<Setting range>

▶ 00:00 – 99:59(Hr:Min.)

▶ Timer does not work if “Tm:00:00" is shown.

“Tm:**:**"is indicated on the screen.

▶ Timer signal lamp is on when the time is over.

“Time-out"is shown on the screen.

▶ Every control output and relay is off when time is

over.

▶ Put “START/STOP" button to get out from “Time-

out"

● Timer does not work if a temp. is not reached to set temp.

In case of this, the Timer lamp is flickering

● Timer works when the present temp. is over the set temp.

In case of this, the Timer lamp is on and “:"part of time

signal is flickering.

Page 38: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 38

7. Operating “AUTO TUNING"

7-1. Start to “AUTO TUNING"

Operating mode Waiting mode

“Auto-tuning" selected

Screen. Auto-

tuning

LED

screen

Example of Temp. auto-tuning.

Please put “AT" button for 5 sec in START condition, then you

can see selectable screens for Auto-tuning

Auto-tuning is only done in Operating mode

1. Select a control item for “Auto-tuning" with Enter button.

2. “Auto-tuning" is operated by putting “MODE" button and

“Auto-tune" signal lamp is flickering.

Condition of AUTO-TUNE signal lamp

Page 39: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 39

7-2. Example of Temp. AUTO TUNING

7-3. Example of Humid. AUTO TUNING

If you move cursor to Humidity

by putting Enter button, the

screen shows Humidity (Y) and

others (N)

“Auto-tuning" is operated if

you put MODE button and AT-TUNE

lamp is on.

If you move cursor to TEMP by

putting Enter button, the screen

shows Temp(Y) and others (N).

“Auto-tuning" is operated if

you put MODE button and AT-TUNE

lamp is on.

Page 40: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 40

How to solve problems(check before ask)

Symptoms Confirmation Arrangement

Check the main power

Operate after checking

whether it's rated

voltage.

Check whether the power plug

is disconnected from an

outlet

put a power plug in an

outlet after turn off

the main power switch.

Check whether Run LED on

Display is blowing out.

Operate by pushing

START/STOP button.

If equipment does not

operate

Check weather it is power

failure.

Operate the equipment

after powering it on.

Please check if an obstacle

is in an intake ventilation

Remove obstacles from

the intake.

If temperature does not

go down

Please confirm whether there

is too much dirt in Filter.

Clean the Condenser Fin

And Filter.

The machine becomes off

even though turning on

the power switch.

Leakage

Stop operating and

contact with purchase

store

Check a power plug is

disconnected from an outlet

correctly

Turn off the power

switch and connect the

plug into the outlet. Display does not show

anything even though the

power switch gets on

Check whether the power

switch is ok.

Ask a purchase store

for After Service.

6

Page 41: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 41

Daily Checking

Section Daily

check

Weekly

check

Monthly

check

Yearly

check Note

Cleanliness

state of CASE ○ Condition Check

State of a

fluorescent

Lamp ○ Condition Check

State of an

illuminometer ○

CO2

Regulator ○

Air Filter

Cleaning ○

Connecting

condition ○

Noise and

Vibration ○

Controller

Check ○ Condition Check

① Selects a manager, and please check very often and

don't make other person to operate except for the

person in charge of operating.

② Don't repair product in the state that connected the

power. This makes the equipment to be broken earlier

7

Page 42: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 42

How to clean

◆ To clean appearance.

1. Turn off main power switch of the

equipment.

2. Pulls out a power plug of a

equipment and turn off the power.

3. Wipe the equipment with sponge or

the soft cloth that wetted a

neutral detergent.

4. Wipe the equipment with cloth in

order to remove a neutral

detergent.

5. Wipe with dry cloth.

◆ To clean air filter

1. If Filter is polluted, pneumatic

circulation decreases and obstructs

normal operation of the equipment.

2. Wash Air Filter according to a degree

of pollution with a vacuum cleaner or

water, and dry it at the shade where

ventilation is fluent in after washing

* Don’ t use insecticide or an

organic solvent when you clean the

equipment

GC-300

GC-1000

8

Page 43: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 43

How to clean

◆ To clean internal chamber

1. Turn off main power switch of the

equipment.

2. Take off inner Shelves and a

Shelves hanger of the equipment.

3. Wipe the inside with soft cloth

wetted with a neutral detergent.

4. Wipe with wet cloth in order to

remove a neutral detergent

5. Wipe with dry cloth.

◆ To replace lamp

1. Check the condition of a door

fluorescent lamp periodically.

2. Replace it after turn off the power

of the equipment to avoid electric

shock.

◆ To store equipment for a long

time without operation

1. Off the power by unplugging the

equipment.

2. Wipe with soft cloth cleanly.

3. Please keep it after packing.

8

Page 44: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 44

Water Tank Usage

1. How to use GC-300 Water Tank

Fill water into the water tank following to the

orders as below.

1. Take out the Water Tank from the supporter.

2. Open the cover and fill water into the

Water Tank.

3. Keep the water tank back to the supporter.

1. Water Tank : for controlling Humidity

by supplying water in the chamber.

(Capacity: 10L.)

2. Supporter: To keep the water tank on the body.

1

2

(Water Tanks are indicated by

orange colored line)

(Water Tanks can be set by 2 kinds shown)

User can select the position of water tank

as the pictures.

If the water tank lacks of water, lamp is

getting on and the humidity controlling is

stopped operating.

Use a ladder to move the water tank from

the supporter.

- Name and function of water tank

9

Page 45: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 45

2. GC-1000 Water Tank Usage

If the water tank lacks of water, lamp is getting on and the humidity controlling is stopped operating.

Please fill water in Water Tank according to the

following procedures.

1. Open water inlet of the bottom in front of the

equipment. (Pull the handle of Water inlet

forward)

2. Fill water by using funnel or kettle

3. Push water tank inlet back to the original

position.

1. Waterspout: Condensation steam is supplied to

Water Tank again.

2. Water Inlet : Please fill water in Water Tank

according to the following procedures..

(Total Capacity of Water Tank is 20L.)

3. Water level gauge : It shows water level in

the Water Tank.

2

(Water Tank part.)

1

- Water Tank names and function

3

If the water tank lacks of water, lamp is

getting on and the humidity controlling is

Page 46: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 46

Specification

Model GC-300TL GC-300TLH GC-1000TLH

Chamber Volume 300L 1000L

Permissible

Environmental Condition

Temperature 5℃ to 40℃

Maximum Relative Humidity 80%

Altitude up to 2,000m

Range +5℃ ~ 50℃(Lamp off), +10℃ ~ 50℃(Lamp on),

+20℃ ~ 50℃(with humidity)

Fluctuation ±0.1℃ at 25℃(without humidity) Temperature

Variation ±0.8℃ at 25℃(without humidity)

Range 0~20,000Lux 0~35,000Lux

Controls 10 Step 11 Step Illumination

Lamp FL 32W×18EA FL 32W×16EA

METAL 400W×3EA

40~80RH%(at20~35℃) 50 ~ 90RH% (at 20~35℃)Range -

70~90RH%(at36~50℃) 60 ~ 90RH% (at 36~50℃)Humidity

Fluctuation - ±3% at 60%RH

Chamber Stainless steel, 0.8t Stainless steel, 1.0t

Body Steel 1.2t, powder coating Steel 1.6t, powder

coating

Shelves Stainless steel wire, electro polished

Material

Inner door Tempered safety glass

Range 300 to 5,000ppm CO2(Optional)

Sensor NDIR CO2 sensor

Controller Microprocessor PID controller

Safety device Power Breaker, Level monitor, Door limit

Over temp. limit Hydraulic over temp. limit

Refrigerator 1/2HP 3/4HP

Temp. Heater Fin type 750W ⅹ 2EA Fin type 1500Wⅹ 2EA

Humid. Heater - 1300W 1500W

Internal(mm) 510ⅹ540ⅹ1100 1200ⅹ800ⅹ1080 Dimensions

(WⅹDⅹH) External(mm) 700ⅹ805ⅹ1900 1410ⅹ1070ⅹ2150

Electric Requirements AC 230V, 50/60Hz

Power Consumption 12.5A 16.0A 30.0A

Weight(net) 250Kg 550kg

* The above specification can be changed by quality of a product and a performance

elevation.

10

Page 47: GC-300.1000 영문 09.02.10~현재 -유효...materials into the chamber The equipment is not explosion- proof structure. Sample can be exploded in the chamber Do not keep inflammable

Page 47

Warranty

1. Service under Warranty period

If trouble occurs during product use, user can get free service for 2 year

limited warranty from the date of purchase.

2. Exceptional cases even during under warranty period

User can not be credited by warranty in case of as below.

① If trouble occurs by an act of God.

② If the equipment breaks down due to misusing of available voltage.

③ If damage occurs by dropping a product, or impact.

④ If damage occurs in an appearance by organic solvents such as thinner, benzene.

⑤ If damage occurs without following to notice in the manual

⑥ If damage occurs by fixing the equipment by any person who is not related

with Jeio tech.

⑦ If damage occurs by a mistake of a customer

How to waste products

Check the notice as below when you are going to waste the equipment

How to waste each model

Element Model Weight (kg) Dimension (mm) How to waste

Main Body GC-300TL 250 kg 700ⅹ805 19ⅹ 00

Main Body GC-300TLH 250 kg 700 805 19ⅹ ⅹ 00

Main Body GC-1000TLH 550 kg 1410 1ⅹ 070 2150ⅹ

Connect with a company that

is dealing with waste

matter.

11