 1 GAS CHROMATOGRAPHY (GC)   SAIMA ALEEM. M.PHIL  . Course no CHM  -754. COURSE SUPERVISOR: DR RAHAT  

description

gas

Transcript of gaschromatography-140516135208-phpapp02

Page 1: gaschromatography-140516135208-phpapp02

7/17/2019 gaschromatography-140516135208-phpapp02

http://slidepdf.com/reader/full/gaschromatography-140516135208-phpapp02 1/21

 

1

GAS

CHROMATOGRAPHY

(GC)  

 SAIMA ALEEM.

M.PHIL .

Course no CHM  -754. 

COURSE SUPERVISOR: DR RAHAT  

Page 2: gaschromatography-140516135208-phpapp02

7/17/2019 gaschromatography-140516135208-phpapp02

http://slidepdf.com/reader/full/gaschromatography-140516135208-phpapp02 2/21

 

2

GAS CHROMATOGRAPHY

INTRODUCTION

•  Gas chromatography is one of the most versatile and ubiquitous

analytical techniques in the laboratory. It is widely used for thedetermination of organic compounds.

•  Complex mixtures can be separate by this technique.

•  A gas chromatography is a chemical analysis instrument for separating

chemicals in a complex sample. 

  A gas chromatograph uses a flow-through narrow tube known as

the column, through which different chemical constituents of a sample

pass in a gas stream (carrier gas, mobile phase) at different rates.

  It depending on their various chemical and physical properties and their

interaction with a specific column filling called the stationary phase.   As the chemicals exit the end of the column, they are detected and

identified electronically. The function of the stationary phase in the

column is to separate different components, causing each one to exit the

column at a different time (retention time).

  Other parameters that can be used to alter the order or time of

retention are the carrier gas flow rate, column length and the

temperature.

Page 3: gaschromatography-140516135208-phpapp02

7/17/2019 gaschromatography-140516135208-phpapp02

http://slidepdf.com/reader/full/gaschromatography-140516135208-phpapp02 3/21

 

3

Gas chromatography

  In a GC analysis, a known volume of gaseous or liquid analyte is injected

into the "entrance" (head) of the column, usually using a

microsyringe (or, solid phase microextraction fibers, or a gas source

switching system).

 

As the carrier gas sweeps the analyte molecules through the column,

this motion is inhibited by the adsorption of the

analyte molecules either onto the column walls or onto packing

materials in the column.

  The rate at which the molecules progress along the column depends on

the strength of  adsorption, which in turn depends on the type of

molecule and on the stationary phase materials. 

 

Since each type of molecule has a different rate of progression, thevarious components of the analyte mixture are separated as they

progress along the column and reach the end of the column at different

times (retention time). A detector is used to monitor the outlet stream from the

column.

  The time at which each component reaches the outlet and the amount of that

component can be determined. 

AUTO SAMPLER 

• 

The auto sampler provides the means to introduce a sample automatically into theinlets.

•  Manual insertion of the sample is possible but is no longer common. Automatic

insertion provides better reproducibility and time-optimization.

•  Different kinds of auto samplers exist. Auto samplers can be classified in

relation to sample capacity (auto-injectors vs. auto samplers, where

auto-injectors can work a small number of samples), to robotic

technologies (XYZ robot vs. rotating robot – the most common), or to

analysis:

  Liquid.

  Static head-space by syringe technology.

  Dynamic head-space by transfer-line technology.

  Solid phase micro extraction (SPME).

CARRIER GAS

Page 4: gaschromatography-140516135208-phpapp02

7/17/2019 gaschromatography-140516135208-phpapp02

http://slidepdf.com/reader/full/gaschromatography-140516135208-phpapp02 4/21

 

4

•  In gas chromatography Helium, nitrogen, Hydrogen or argon gases are

used as a carrier gas.

•  The choice of gas is depend upon availability, purity required,

consumption and the type of detector employed.

• 

Helium is preferred, when thermal conductivity detectors are used.Because of it high thermal conductivity relative to that of the vapours of

most organic compounds.

SAFETY

•  Free standing gas cylinder must always be supported by means of

clamps or chains.

•  Waste gases, especially hydrogen, must be vented through an extraction

hood.

INJECTORS

Packed Column Injectors  In general, the sample injected onto a packed GC column ranges in

volume from 0.5 micro liters to 5 micro liter and usually contains the

materials of interest at concentrations ranging from 5% v/v to 10%

w/v.

Page 5: gaschromatography-140516135208-phpapp02

7/17/2019 gaschromatography-140516135208-phpapp02

http://slidepdf.com/reader/full/gaschromatography-140516135208-phpapp02 5/21

 

5

A Pack column injector

 

The sample is injected by a hypodermic syringe, through a siliconerubber septum directly into the column packing or into a flash heater.

Although the latter tends to produce broader peaks it also disperses the

sample radially across the column.

  Direct injection into the packing constrains the sample into a small

volume, but can cool the front of the packing.

  The silicone septum is compressed between metal surfaces in such a

manner that a hypodermic needle can pierce it, but when it is

withdrawn the hole is closed as a result of the septum compression and

there is no gas leak.  The glass liner prevents the sample coming in contact with the heated

metal wall and thus, reduces the chance of thermal decomposition. The

glass liner can be fitted with a separate heater and the volatalization

temperature can, thus, be controlled. This "flash heater" system is

available in most chromatographs.

Page 6: gaschromatography-140516135208-phpapp02

7/17/2019 gaschromatography-140516135208-phpapp02

http://slidepdf.com/reader/full/gaschromatography-140516135208-phpapp02 6/21

 

6

  By using a syringe with a long needle, the tip can be made to penetrate

past the liner and discharge its contents directly into the column

packing.

  This procedure is called 'on-column injection' and, as it reduces peak

dispersion on injection and thus, provides higher column efficiencies, isoften the preferred procedure.

Open Tubular Column Injection Systems

  Due to the very small sample size that must be placed on narrow bore

capillary columns; a split injection system is necessary.

Page 7: gaschromatography-140516135208-phpapp02

7/17/2019 gaschromatography-140516135208-phpapp02

http://slidepdf.com/reader/full/gaschromatography-140516135208-phpapp02 7/21

 

7

  As the sample passes the column opening, a small fraction is split off

and flows directly into the capillary column are called a split injector.

  Regulating the portion of the carrier gas that passes to waste changes

the split ratio. This achieved by an adjustable flow resistance situated in

the waste flow line. This device is only used for small diameter capillarycolumns where the charge size is critical.

  The device has certain disadvantages due to component differentiation

and the sample placed on the column may not be truly representative.

The solutes with the higher diffusivities (low molecular weight) are lost

preferentially to those with lower diffusivities (higher molecular

weights). Consequently, quantitative analyses carried out using the high

efficiency small diameter capillary columns may have limited accuracy

and precision, depending on the nature of the sample.

The Split Injection System  This problem was partially solved by using larger diameter

columns that would permit on-column injection.

Page 8: gaschromatography-140516135208-phpapp02

7/17/2019 gaschromatography-140516135208-phpapp02

http://slidepdf.com/reader/full/gaschromatography-140516135208-phpapp02 8/21

 

8

  There are also difficulties associated with this type of

injector. On injection, the sample breaks up into separate

portions, and bubbles form at the beginning of the column

causing the sample to be deposited at different positions

along the open tube as the solvent evaporates.  On starting to develop the separation, each local

concentration of sample acts as a separate injection.

Column

•  The actual components is effected in the Column

•  Where the nature of the solid support, type and amount of liquid phase,

method of packing length and temperature are important factor in

obtaining the desired resolution.

•  The Column is enclosed in a thermostatically controlled oven so that its

temperature is held constant to within 0.5 C. Thus ensuring

reproducible conditions.

•  The operating temperature may range from ambient to over 400 C and

for isothermal operation is kept constant during the separation process.

Type of Column

1. Packed columns

•  The internal diameter of Column is 2-6 mm.

The outer diameter of Column is 3-10 mm and it is coil for compactness

Page 9: gaschromatography-140516135208-phpapp02

7/17/2019 gaschromatography-140516135208-phpapp02

http://slidepdf.com/reader/full/gaschromatography-140516135208-phpapp02 9/21

 

9

Packed columns

Qualities of the packed column

•  Glass column must be used if any of the sample components are

decompose by compact with metal.

•  The material chosen as a inaet support should be uniform granular size

and capable of being packed in to a uniform bed in a column.

•  The surface area of the material should be large so as to promote

distribution of the liquid phase as a film and ensure the rapidattainment of equilibrium between the stationary phase and mobile

phase.

•  The most commonly used supports(e.g celite) are made from

diatomaceous materials which can hold liquid phases in amounts

Page 10: gaschromatography-140516135208-phpapp02

7/17/2019 gaschromatography-140516135208-phpapp02

http://slidepdf.com/reader/full/gaschromatography-140516135208-phpapp02 10/21

 

10

exceeding 20% without becoming too sticky to flow freely and can be

easily packed.

•  To be a good resolution the height is equivalent of the theoretical plate

is proportion to the average particle diameter.

•  Rapidly increase the pressure due to decrease the particle size.

•  It is necessary to increase the pressure for achieving the flow rate

through the column

Open columns

•  These capillary columns (i.d < 1mm) are increasingly used in GLC. 

• 

It is used for complex mixtures.

•  This result from a highly theoretical plate numbers which can be attain

with long column of this type for the relatively smaller pressure drop.

•  In these capillary columns the stationary phase s coated on the inner

wall of the tube,

Page 11: gaschromatography-140516135208-phpapp02

7/17/2019 gaschromatography-140516135208-phpapp02

http://slidepdf.com/reader/full/gaschromatography-140516135208-phpapp02 11/21

 

11

Here 2 types of capillaries are available.

1. Wall coated open tubular

•  Here the stationary phase is directly coated on the inner wall of tubing.

2. Supported coated open tubular

Page 12: gaschromatography-140516135208-phpapp02

7/17/2019 gaschromatography-140516135208-phpapp02

http://slidepdf.com/reader/full/gaschromatography-140516135208-phpapp02 12/21

 

12

•  It have finely divided layer of solid support material deposited on the

inner wall of the stationary phase is then coated

•  Supported coated open tubular is less efficient then wall coated open

tubular.

•  Capillary column are made up of thin wall stainless steel.

•  The dimension of the column, which is coiled, is 25- 200 m long and 0.2-

0.5mm i.d.

•  The drawback of this column is that they have low sample capacity than

pack column.

DETECTOR

The choice of detector will depend the concentration level of the

sample.

The important properties of a detector are as follows

1. SENSITIVITY

•  This is usually defined as the detector response (mV) per unit

concentration of analyte (mg/ml).

•  It is closely related to the detection limit.

Since high sensitivity often give low limit of detection.

•  The limit of detection will be rises if the detector produced noise.

• 

The sensitivity also determines the slop of calibration graph and

therefore influences the precision of the analysis.

2 . Linearity

Page 13: gaschromatography-140516135208-phpapp02

7/17/2019 gaschromatography-140516135208-phpapp02

http://slidepdf.com/reader/full/gaschromatography-140516135208-phpapp02 13/21

 

13

•  The linear range of the detector refer to the concentration range over

which the signal is directly proportional to the concentration of analyte.

•  Linearity in detector response will give linearity of the calibration graph

and allow later to be drawn with more certainty.

•  With the convex calibration curved, the precision is reduced at higher

concentrations where the slope of the curved is much lesser.

•  The high linearity range has high advantage but detector small range

will be used because of their other qualities, although they will need to

be recalibrated over a number of difference concentration ranges.

3. Stability

•  The important characteristic of a detector is the extent to which the

signal output remains constant with time, assuming there is a constant

input.

•  Lack of the stability can be exhibited in to two ways.

1. Baseline noise.

2. Drift.

1 .Baseline noise.

•  Baseline noise is caused by a rapid random variation in detector output

difficult to measure small peaks against the fluctuation background.

2 .Drift

•  Drift is often due to factors external to the detector and impose a more

fundamental limit on its performance, such as temperature change or

column bleed and so is controllable, where as noise is usually due to

poor contacts within detector and impose a more fundamental limit on

its performance.

•  It must be reproducible

TYPE OF DETECTOR

Page 14: gaschromatography-140516135208-phpapp02

7/17/2019 gaschromatography-140516135208-phpapp02

http://slidepdf.com/reader/full/gaschromatography-140516135208-phpapp02 14/21

 

14

Thermal conductivity detector

•  The most important of the bulk physical property detector is a thermal

conductivity detector which is universal, nondestructive, concentration

– sensitive detector. It is still widely used in gas chromatography.

•  Helium and hydrogen are the best carrier gases to use in conjunction

with this type of detector. Since their thermal conductivity are much

higher than the other gases.

Page 15: gaschromatography-140516135208-phpapp02

7/17/2019 gaschromatography-140516135208-phpapp02

http://slidepdf.com/reader/full/gaschromatography-140516135208-phpapp02 15/21

 

15

Thermal Conductivity Detector (TCD)

•  In the detector 2 pair of match filaments are arranged in a Wheatstone

bridge circuit, 2 filaments is opposite arm of the bridge is surrounded by

the carrier gas only.

•  The other 2 are surrounded by the effluent from the chromatographic

column.

•  There are 2 types of gas channels through the cell.

1. Sample channels.

2. Reference channels.

• 

When the pure carrier gas is pass over both the Sample and reference

filaments the bridge is balanced but when the vapours emerges from

the column, the rate of cooling of the sample filament become changed

and the bridge become unbalanced.

•  It is used for the detection limit of permanent gases light hydrocarbon

and compound which response poorly from the flame ionization

detector.

• 

It is used for the gas chromatography studies of metal chelates, e.g

mixture of beryllium, aluminium, gallium and indium

trifluoroacethyacetonates. For many general applications it is replaced

by Flame ionization detector.

Ionisation detector

•  An important characteristic of the common carrier gases is that they

behave as perfect insulators at normal temperature and pressures .The

increase conductivity due to the presence of a few charge molecules in

the effluent from the column thus provide the high sensitivity which is a

feather of the ionisation based detector.

•  Ionisation detector in current used include the flame ionisation

detector, thermionic ionisation detector, photo ionisation detector and

Page 16: gaschromatography-140516135208-phpapp02

7/17/2019 gaschromatography-140516135208-phpapp02

http://slidepdf.com/reader/full/gaschromatography-140516135208-phpapp02 16/21

 

16

election capture detector  each of course employing a different method

to generate an ion current.

•  The basis of this detector is that the effluent from the column is mixed

with hydrogen and burned in air to produce a flame which has sufficientenergy to ionize solute molecules having low Ionisation potential.

Flame Ionisation detector

•  The ion produced are collect at electrodes and a result ion current

measured.

•  The burner jet is the negative electrode whilst the anode is usually wire

or grid extending into the tip of the flame. Because the sample is destroy

in the flame.

Page 17: gaschromatography-140516135208-phpapp02

7/17/2019 gaschromatography-140516135208-phpapp02

http://slidepdf.com/reader/full/gaschromatography-140516135208-phpapp02 17/21

 

17

•  A stream-splitting device is inserted between the column and detector

and allows the bulk of the sample to by-pass the detector.

•  It is widely applicable for gas chromatography of organic compounds.

• 

It has high sensitivity, stability, fast response and linearly response with

analytes.

Electron capture detector

•  Most ionisation detectors are based on measurement of the increase in

current which occur when a more readily ionised molecule appears in

the gas stream.

•  the Electron capture detector is differ from the other ionisation

detectors .it exploits the recombination phenomenon, based on Electron

capture by compound having affinity for free electron .it only measurethe decrease in current.

Page 18: gaschromatography-140516135208-phpapp02

7/17/2019 gaschromatography-140516135208-phpapp02

http://slidepdf.com/reader/full/gaschromatography-140516135208-phpapp02 18/21

 

18

Electron capture detector

•  A beta rays is used to generate slow electron by ionisation of the carrier

gas (nitrogen preferred) flowing through the detector.

•  The slowly electron migrate to the anode under a fixed potential and

give rise to a steady baseline current•  When the electron capturing gas is emerges from the column. It will

react with electron by a negative ion of much higher mass with a

corresponding reduction in current flow.

•  It response due to electron affinity of the eluate molecules. Being

sensitive to compound containing halogens, sulphur, anhydrides,

nitrites, conjugated carbonyls, nitrates and organo metallic compounds.

•  It has high sensitivity.

Element sensitive detector

•  Element sensitive detector in current used includes “Thermionic

Ionisation detector, Flame photometric detector and atomicabsorption detector”. 

Thermionic Ionisation detector

•  It only response to nitrogen containing compound and phosphorus.

•  It contains an electrically heated rubidium silicate bead situated a few

millimeters above the detector jet tip and below the collector electrode.

•  The temperature of beat is maintained at 600-800 degree centigrade.

Flame photometric detector

•  It only response to sulphur and phosphorus.

•  Hydrogen – rich flame results in the formation of luminescent species

that emit light character of the heteroatom introduced in to the flame.

• 

When nitrogen is carrier gas, it can be used.

•  Hydrogen is introduced at the burner to initiate combustion.

Atomic absorption spectroscopy

•  It is used for the separation of organo metallic compound.

Page 19: gaschromatography-140516135208-phpapp02

7/17/2019 gaschromatography-140516135208-phpapp02

http://slidepdf.com/reader/full/gaschromatography-140516135208-phpapp02 19/21

 

19

TYPES OF CHROMATOGRAPHY

•  Gas solid (adsorption) chromatography.

•  Gas liquid (partition) chromatography.

The most important one of the two is gas liquid chromatography, used

in the foam of capillary column.

GAS-LIQUID CHROMATOGRAPHY  

•  In gas-liquid chromatography, the mobile phase is a gas such as helium

and the stationary phase is a high boiling point liquid absorbed onto a

solid.

 A flow scheme for gas-liquid chromatography

Injection of the sample

•  Very small quantities of the sample that you are trying to analyses

are injected into the machine using a small syringe. The syringe

needle passes through a thick rubber disc (known as a septum)

which reveals itself again when the syringe is pulled out.

•  The injector is contained in an oven whose temperature can be

controlled. It is hot enough so that all the sample boils and is

carried into the column as a gas by the helium (or other carrier

gas).

Page 20: gaschromatography-140516135208-phpapp02

7/17/2019 gaschromatography-140516135208-phpapp02

http://slidepdf.com/reader/full/gaschromatography-140516135208-phpapp02 20/21

 

20

Column: packed Column.

Retention time

•  The time taken for a particular compound to travel through the

column to the detector is known as its retention time. This time is

measured from the time at which the sample is injected to the

point at which the display shows a maximum peak height for that

compound.

•  Different compounds have different retention times. For a

particular compound, the retention time will vary depending on

the boiling point of the compound.

• 

 A compound which boils at a temperature higher than the column

temperature is going to spend nearly all of its time condensed as a

liquid at the beginning of the column. So high boiling point means a

long retention time.

•  The solubility in the liquid phase. The more soluble a compound is

in the liquid phase, the less time it will spend being carried. along

by the gas. High solubility in the liquid phase means a high

retention time.

•  The temperature of the column. A higher temperature will tend to

excite molecules into the gas phase - either because they evaporate

more readily, or because they are so energetic that the attractions

of the liquid no longer hold them.

•   A high column temperature shortens retention times for

everything in the column.

• 

The lower the temperature of the column, the better theseparation.

Detector

•  There are several different types of detector is use.

Page 21: gaschromatography-140516135208-phpapp02

7/17/2019 gaschromatography-140516135208-phpapp02

http://slidepdf.com/reader/full/gaschromatography-140516135208-phpapp02 21/21

21

•  The flame ionisation detector described below is commonly used

and is easier to describe and explain than the alternatives.

GAS SOLID CHOROMATOGRAPHY

 

Comprises all gas chromatographic methods in which the

stationary is an active solid (e.g. charcoal, molecular

sieves).Separation is achieved by adsorption of the components of

a sample. 

  The retention time is long, due to which it is less used then gas

liquid chromatography.