Gas Chromatography: Detectors
![Page 2: Gas Chromatography: Detectors. The Ideal Detector n Adequate sensitivity - range 10 ^- 18 to 10^-15 g analyte/s n good stability and reproducibility n.](https://reader035.fdocuments.in/reader035/viewer/2022062517/56649f155503460f94c2aac3/html5/thumbnails/2.jpg)
The Ideal Detector
Adequate sensitivity - range 10^-18 to 10^-15 g analyte/s
good stability and reproducibility a linear response to analyte that
extends over several orders of magnitude
a temperature range from room temperature to at least 400 degrees C
![Page 3: Gas Chromatography: Detectors. The Ideal Detector n Adequate sensitivity - range 10 ^- 18 to 10^-15 g analyte/s n good stability and reproducibility n.](https://reader035.fdocuments.in/reader035/viewer/2022062517/56649f155503460f94c2aac3/html5/thumbnails/3.jpg)
The Ideal Detector
A short response time that is independent of flow rate
high reliability and ease of use similarity in response toward all
analytes of alternatively a higher predictable and selective response toward + classes of ananlytes
nondestructive of sample
![Page 4: Gas Chromatography: Detectors. The Ideal Detector n Adequate sensitivity - range 10 ^- 18 to 10^-15 g analyte/s n good stability and reproducibility n.](https://reader035.fdocuments.in/reader035/viewer/2022062517/56649f155503460f94c2aac3/html5/thumbnails/4.jpg)
Types of Detectors
Thermal Conductivity Detector Thermionic Detectors Atomic Emission Detector Flame Ionization Detector Electron Capture Detector
![Page 5: Gas Chromatography: Detectors. The Ideal Detector n Adequate sensitivity - range 10 ^- 18 to 10^-15 g analyte/s n good stability and reproducibility n.](https://reader035.fdocuments.in/reader035/viewer/2022062517/56649f155503460f94c2aac3/html5/thumbnails/5.jpg)
Thermal Conductivity Detector (TCD) Q. How does this work? Answer
Electronically measures the changes in the thermal conductivity of the gas.
The thermal conductivity changes due to the presence of the analyte
![Page 6: Gas Chromatography: Detectors. The Ideal Detector n Adequate sensitivity - range 10 ^- 18 to 10^-15 g analyte/s n good stability and reproducibility n.](https://reader035.fdocuments.in/reader035/viewer/2022062517/56649f155503460f94c2aac3/html5/thumbnails/6.jpg)
The Modulate Single Filament (TCD) A method to get rid of some noise ! How?
Both the analytical and reference gases are passed altrenately over a tiny filament
This is done at a set frequency, that sets the frequency of the output of the electrical signal
The amplitude fo the signal is proportional to the difference in the thermal conductivity of both gases
![Page 7: Gas Chromatography: Detectors. The Ideal Detector n Adequate sensitivity - range 10 ^- 18 to 10^-15 g analyte/s n good stability and reproducibility n.](https://reader035.fdocuments.in/reader035/viewer/2022062517/56649f155503460f94c2aac3/html5/thumbnails/7.jpg)
TCD
What is the best type of carrier gas for the TCD?
He and H have high thermal conductivity therefore they show the presence of other components of the gas because of the large drop in thermal conductivity
Other gases are not used in TCD
![Page 8: Gas Chromatography: Detectors. The Ideal Detector n Adequate sensitivity - range 10 ^- 18 to 10^-15 g analyte/s n good stability and reproducibility n.](https://reader035.fdocuments.in/reader035/viewer/2022062517/56649f155503460f94c2aac3/html5/thumbnails/8.jpg)
Advantages
simplicity large linear dynamic range ( about
10^5) general response to organic and
inorganic samples non-destructive character
![Page 9: Gas Chromatography: Detectors. The Ideal Detector n Adequate sensitivity - range 10 ^- 18 to 10^-15 g analyte/s n good stability and reproducibility n.](https://reader035.fdocuments.in/reader035/viewer/2022062517/56649f155503460f94c2aac3/html5/thumbnails/9.jpg)
Limitations
Low sensitivity (10^-8 solute/mL carrier gas)
Others exceed this by factor of 10^4 to 10^7
![Page 10: Gas Chromatography: Detectors. The Ideal Detector n Adequate sensitivity - range 10 ^- 18 to 10^-15 g analyte/s n good stability and reproducibility n.](https://reader035.fdocuments.in/reader035/viewer/2022062517/56649f155503460f94c2aac3/html5/thumbnails/10.jpg)
Thermionic Detectors (TID) How does this work?
The effluent is mixed with H Passes through a flame tip assembly
to ignite Hot gas flows around an electrically
heated rubidium silicate bead Heated bead forms a plasma What occurs in the plasma to produce
large amounts of ions for D and N containing molecules is not understood
![Page 11: Gas Chromatography: Detectors. The Ideal Detector n Adequate sensitivity - range 10 ^- 18 to 10^-15 g analyte/s n good stability and reproducibility n.](https://reader035.fdocuments.in/reader035/viewer/2022062517/56649f155503460f94c2aac3/html5/thumbnails/11.jpg)
TID
There is a large ion current that results
![Page 12: Gas Chromatography: Detectors. The Ideal Detector n Adequate sensitivity - range 10 ^- 18 to 10^-15 g analyte/s n good stability and reproducibility n.](https://reader035.fdocuments.in/reader035/viewer/2022062517/56649f155503460f94c2aac3/html5/thumbnails/12.jpg)
Advantages
Selective for organic compounds with P and N
This is useful for detecting pesticides that contain P
![Page 13: Gas Chromatography: Detectors. The Ideal Detector n Adequate sensitivity - range 10 ^- 18 to 10^-15 g analyte/s n good stability and reproducibility n.](https://reader035.fdocuments.in/reader035/viewer/2022062517/56649f155503460f94c2aac3/html5/thumbnails/13.jpg)
Atomic Emission Detection (AED) Eluent introduced into a microwave
- energized helium plasma coupled to a diode-array optical spectrometer
Plasma sufficiently energetic to atomize all of the elements in the sample and excite their characteristic atomic emission spectra
![Page 14: Gas Chromatography: Detectors. The Ideal Detector n Adequate sensitivity - range 10 ^- 18 to 10^-15 g analyte/s n good stability and reproducibility n.](https://reader035.fdocuments.in/reader035/viewer/2022062517/56649f155503460f94c2aac3/html5/thumbnails/14.jpg)
AED
Spectra observed using a spectrometer that employs a movable, flat diode array is capable of detecting omitted radiation between 170 to 1780 nm
The positional diode array is capable of measuring 2 to 4 elements at any given setting
Present software allows measurement of the concentration of 15 elements
![Page 15: Gas Chromatography: Detectors. The Ideal Detector n Adequate sensitivity - range 10 ^- 18 to 10^-15 g analyte/s n good stability and reproducibility n.](https://reader035.fdocuments.in/reader035/viewer/2022062517/56649f155503460f94c2aac3/html5/thumbnails/15.jpg)
Flame Ionization Detector Consists of a stainless steel burner
assembly installed in the detector compartment and a electrometer system in a separate unit adjacent to the gas chromatograph
Often it is installed in the tandem with the thermal conductivity cell
Effluent form the column enters burner base through millipore filters which remove contaminates
![Page 16: Gas Chromatography: Detectors. The Ideal Detector n Adequate sensitivity - range 10 ^- 18 to 10^-15 g analyte/s n good stability and reproducibility n.](https://reader035.fdocuments.in/reader035/viewer/2022062517/56649f155503460f94c2aac3/html5/thumbnails/16.jpg)
Flame Ionization Detector hydrogen mixed with gas stream at
bottom of jet and air or oxygen is supplied axially around the jet
hydrogen flame burns at the tip, which also functions as the cathode and is insulated form the body by a ceramic seal
Collector electrode is above the burner tip and is made of platinum
![Page 17: Gas Chromatography: Detectors. The Ideal Detector n Adequate sensitivity - range 10 ^- 18 to 10^-15 g analyte/s n good stability and reproducibility n.](https://reader035.fdocuments.in/reader035/viewer/2022062517/56649f155503460f94c2aac3/html5/thumbnails/17.jpg)
Flame Ionization Detector In series with flame gases is a
selection of resistors 10^7 to 10^10 ohms
vibrating reed electrometer used to provide sensitivities up to 5 x 10^13 Amps
Carbon counting device that produces a current proportional to number of ions or electrons formed in the flamed gases
![Page 18: Gas Chromatography: Detectors. The Ideal Detector n Adequate sensitivity - range 10 ^- 18 to 10^-15 g analyte/s n good stability and reproducibility n.](https://reader035.fdocuments.in/reader035/viewer/2022062517/56649f155503460f94c2aac3/html5/thumbnails/18.jpg)
Flame Ionization Detector Responds to all organic compounds
except for formic acid Response greatest with
hydrocarbons and decreases with substitution
Except for vapor of elements in Groups I and II, does not respond to inorganic compounds
Sensitivity high due to low noise level
![Page 19: Gas Chromatography: Detectors. The Ideal Detector n Adequate sensitivity - range 10 ^- 18 to 10^-15 g analyte/s n good stability and reproducibility n.](https://reader035.fdocuments.in/reader035/viewer/2022062517/56649f155503460f94c2aac3/html5/thumbnails/19.jpg)
Flame Ionization Detector Insensitivity to water, the
permanent gases, and inorganic compounds simplifies the resolution of components in analysis of aqueous extracts and in air pollution studies
![Page 20: Gas Chromatography: Detectors. The Ideal Detector n Adequate sensitivity - range 10 ^- 18 to 10^-15 g analyte/s n good stability and reproducibility n.](https://reader035.fdocuments.in/reader035/viewer/2022062517/56649f155503460f94c2aac3/html5/thumbnails/20.jpg)
Electron Capture Detector operates similar
to proportional counter for measurement of X-radiation
effluent form the column passes over a beta-emitter - I.e.) nickel-63 or tritium
![Page 21: Gas Chromatography: Detectors. The Ideal Detector n Adequate sensitivity - range 10 ^- 18 to 10^-15 g analyte/s n good stability and reproducibility n.](https://reader035.fdocuments.in/reader035/viewer/2022062517/56649f155503460f94c2aac3/html5/thumbnails/21.jpg)
ECD
electron from emitter causes ionization of carrier gas (N)
produces a burst of electrons standing current between
electrodes decreases in presence of organic species that capture the electrons
![Page 22: Gas Chromatography: Detectors. The Ideal Detector n Adequate sensitivity - range 10 ^- 18 to 10^-15 g analyte/s n good stability and reproducibility n.](https://reader035.fdocuments.in/reader035/viewer/2022062517/56649f155503460f94c2aac3/html5/thumbnails/22.jpg)
ECD
selective in its response and highly sensitive
Hewlett Packard makes one with a detection limit of less than 8 fg/sec for lindane
sensitive toward molecules with electronegative functional groups (halogens, peroxides, quinones, nitro groups)
insensitive towards amines, alcohols and hydrocarbons
![Page 23: Gas Chromatography: Detectors. The Ideal Detector n Adequate sensitivity - range 10 ^- 18 to 10^-15 g analyte/s n good stability and reproducibility n.](https://reader035.fdocuments.in/reader035/viewer/2022062517/56649f155503460f94c2aac3/html5/thumbnails/23.jpg)
ECD
important application: detection and determination of chlorinated insecticides
advantages does not alter the sample
significantly (in contrast to flame detection)
quick and easy relatively cheap
![Page 24: Gas Chromatography: Detectors. The Ideal Detector n Adequate sensitivity - range 10 ^- 18 to 10^-15 g analyte/s n good stability and reproducibility n.](https://reader035.fdocuments.in/reader035/viewer/2022062517/56649f155503460f94c2aac3/html5/thumbnails/24.jpg)
ECD
disadvantage : linear response range is usually limited to around 2 orders of magnitude