GaAs detectors for medical imaging G.I.Ayzenshtat a*, S.M. Gushin a, O.B. Koretskaya b, O.P....

15
GaAs detectors for medical imaging G.I.Ayzenshtat a* , S.M. Gushin a , O.B. Koretskaya b , O.P. Tolbanov b , A.V. Tyazhev b , E.A. Babichev c , V.R. Groshev c , G.A. Savinov c , A.P. Vorobiev d a Research Inst. of Semicond. Dev., Tomsk, Russia b Siberian Physical Technical Institute, Tomsk, Russia c Budker Institute of Nuclear Physics, Novosibirsk, Russia d Institute of High Energy Physics, Protvino, Russia [email protected]

Transcript of GaAs detectors for medical imaging G.I.Ayzenshtat a*, S.M. Gushin a, O.B. Koretskaya b, O.P....

Page 1: GaAs detectors for medical imaging G.I.Ayzenshtat a*, S.M. Gushin a, O.B. Koretskaya b, O.P. Tolbanov b, A.V. Tyazhev b, E.A. Babichev c, V.R. Groshev.

GaAs detectors for medical imaging

G.I.Ayzenshtata*, S.M. Gushina, O.B. Koretskayab, O.P. Tolbanovb, A.V. Tyazhevb, E.A. Babichevc, V.R.

Groshevc, G.A. Savinovc , A.P. Vorobievd

a Research Inst. of Semicond. Dev., Tomsk, Russia

b Siberian Physical Technical Institute, Tomsk, Russiac Budker Institute of Nuclear Physics, Novosibirsk, Russia

d Institute of High Energy Physics, Protvino, Russia

* [email protected]

Page 2: GaAs detectors for medical imaging G.I.Ayzenshtat a*, S.M. Gushin a, O.B. Koretskaya b, O.P. Tolbanov b, A.V. Tyazhev b, E.A. Babichev c, V.R. Groshev.

Low dose digital radiographic device

with GaAs - detectors • The aim of this work

was to test GaAs detectors in a real

radiographic device• Gas detector was

replaced by a GaAs one in a Low Dose Digital Radiographic Device “LDRD SIBERIA-N”

Detector

+

Electronics

Low dose digital

radiographic device

Page 3: GaAs detectors for medical imaging G.I.Ayzenshtat a*, S.M. Gushin a, O.B. Koretskaya b, O.P. Tolbanov b, A.V. Tyazhev b, E.A. Babichev c, V.R. Groshev.

The main properties of compensated SI-GaAs(Cr) detectors

• 1. SI-GaAs(Cr) detectors have no Schottky contacts, they have only ohmic ones. I-V characteristics are linear.

• 2.The electric field (E) occupies the whole volume of detector. E(kV/cm) const

• 3. Charge collection in GaAs-detectors is determined by electrons.

• 4. The detector material resistivity is extremely high (ρ>109 Ω∙cm).

-100 0 100

-5

5

0

I, nA

U, V

Cur

rent

Voltage

20 40 600

50

Energy

Cou

nts

keV

E const

Page 4: GaAs detectors for medical imaging G.I.Ayzenshtat a*, S.M. Gushin a, O.B. Koretskaya b, O.P. Tolbanov b, A.V. Tyazhev b, E.A. Babichev c, V.R. Groshev.

Attachment of GaAs-detectors to electronics

Iz = 0I = +I0

I = -I0

+U1 - U1

zX-Rays

The problem: the dark current of detector is about 10 nA. It is necessary to reduce the current by a factor of 10 for the electronics to operate. Solution of the problem: we offer to attach the detectors to electronics as shown in the figure.

Page 5: GaAs detectors for medical imaging G.I.Ayzenshtat a*, S.M. Gushin a, O.B. Koretskaya b, O.P. Tolbanov b, A.V. Tyazhev b, E.A. Babichev c, V.R. Groshev.

Symmetrical detector structures with the compensation of dark currents

The hypothetical detector structure

The real structure of the mammo- graphy detector (100 mkm pitch)

+ U - U

I = 0

Page 6: GaAs detectors for medical imaging G.I.Ayzenshtat a*, S.M. Gushin a, O.B. Koretskaya b, O.P. Tolbanov b, A.V. Tyazhev b, E.A. Babichev c, V.R. Groshev.

The results of current compensation measurements in the symmetrical device.

• The output current of the detector decreases by a factor 30

X - ray s

A S IC

U 1 U 2

T h e sc reen

0 50 100 150 200

0

200

400

600

I = 20 pA

+ U2

U1 = -200 V

I, pA

X- rays

screen

+U0 -U0

Page 7: GaAs detectors for medical imaging G.I.Ayzenshtat a*, S.M. Gushin a, O.B. Koretskaya b, O.P. Tolbanov b, A.V. Tyazhev b, E.A. Babichev c, V.R. Groshev.

D

etec

tor

1 D

etec

tor

2

Plane view of detector (h = m)

A fragment of hybrid circuit of

electronics

A fragment of detectors mounting

64-channel chip

Page 8: GaAs detectors for medical imaging G.I.Ayzenshtat a*, S.M. Gushin a, O.B. Koretskaya b, O.P. Tolbanov b, A.V. Tyazhev b, E.A. Babichev c, V.R. Groshev.

0 10 20 30 40 50 60 700

2000

4000

6000

8000

10000

I, mA

80 keV

Cou

nt (

Arb

. un

its)

0 2000 4000 60000

2000

4000

6000

8000

10000

mR/min

60 keV 80 keV 120 keV

Cou

nt (

Arb

. uni

ts)

The magnitude of the output signal

as a function of the Roentgen tube current

The magnitude of the output signal as a function of dosage rate for various

quantum energy

Page 9: GaAs detectors for medical imaging G.I.Ayzenshtat a*, S.M. Gushin a, O.B. Koretskaya b, O.P. Tolbanov b, A.V. Tyazhev b, E.A. Babichev c, V.R. Groshev.

0 1 2 3 4

500

1000

1500

2000

2500

With Pb screen

Without Pb screen

Sig

nal m

agni

tude

t, sec

Waveshape of the signal in vertical scanning of

the detector SNR2 as a function of

dosage rate

Page 10: GaAs detectors for medical imaging G.I.Ayzenshtat a*, S.M. Gushin a, O.B. Koretskaya b, O.P. Tolbanov b, A.V. Tyazhev b, E.A. Babichev c, V.R. Groshev.

«Channel form» obtained by vertical scanning

0,9 1,0 1,1 1,2 1,30

100

200

300

400

500

21

Downx, mm

1 (U = - 200 V) 2 (U = + 250 V)

Sig

nal m

agni

tude

UDown

To electronics

X-rays

Page 11: GaAs detectors for medical imaging G.I.Ayzenshtat a*, S.M. Gushin a, O.B. Koretskaya b, O.P. Tolbanov b, A.V. Tyazhev b, E.A. Babichev c, V.R. Groshev.

«Channel form» obtained by horizontal scanning

Detector

200 400 600 800 1000 12000,0

0,2

0,4

0,6

0,8

1,0Channel numbersU

x, mkm

30 31 32 33 34 35

The collimator moves along the horizontal axis

Collimator

The normalized amounts of signals for various channels during horizontal scanningTo electronics

X-rays

Page 12: GaAs detectors for medical imaging G.I.Ayzenshtat a*, S.M. Gushin a, O.B. Koretskaya b, O.P. Tolbanov b, A.V. Tyazhev b, E.A. Babichev c, V.R. Groshev.

Image of the test pattern obtained by the GaAs detector with a 400 mkm pitch(The limiting image resolution is 1.4 pairs

of lines per mm)

Page 13: GaAs detectors for medical imaging G.I.Ayzenshtat a*, S.M. Gushin a, O.B. Koretskaya b, O.P. Tolbanov b, A.V. Tyazhev b, E.A. Babichev c, V.R. Groshev.

-150 -100 -50 0 50 100 150-10

-5

0

5

10

15

SI-GaAs(Cr)

Epitaxy p-i-n GaAs(Cr)I, nA

U,V

Plane view of the detector (h = 200m) fabricated

from GaAs liquid-phase epitaxy

Comparision of I-V characteristics for two detectors.

1. The GaAs liquid-phase epitaxy detector 2.

The SI-GaAs(Cr) detector

Page 14: GaAs detectors for medical imaging G.I.Ayzenshtat a*, S.M. Gushin a, O.B. Koretskaya b, O.P. Tolbanov b, A.V. Tyazhev b, E.A. Babichev c, V.R. Groshev.

IT The image of the test pattern obtained by GaAs detector with a 200 m pitch

(The limiting image resolution is 2.8 pairs of lines per mm)

Page 15: GaAs detectors for medical imaging G.I.Ayzenshtat a*, S.M. Gushin a, O.B. Koretskaya b, O.P. Tolbanov b, A.V. Tyazhev b, E.A. Babichev c, V.R. Groshev.

Conclusion

1. It was shown experimentally that the GaAs material compensated with Cr can be used to fabricate efficient X-ray detectors.

2.The object image with the resolution 2.8 pairs of lines per mm has been obtained with GaAs detectors.

3.We consider the GaAs(Cr) liquid-phase epitaxy material to be a more promising material for obtaining the resolution more than 2.8 pairs of lines per mm.