G16.4427 Practical MRI 1 – 5 th March 2015 G16.4427 Practical MRI 1 Advanced pulse sequences.
-
Upload
darcy-allen -
Category
Documents
-
view
220 -
download
0
Transcript of G16.4427 Practical MRI 1 – 5 th March 2015 G16.4427 Practical MRI 1 Advanced pulse sequences.
G16.4427 Practical MRI 1 – 5th March 2015
G16.4427 Practical MRI 1
Advanced pulse sequences
G16.4427 Practical MRI 1 – 5th March 2015
Signal Formula for SE
Mxy negligible (TR >> T2, or spoiler gradient)
= 90° = 180°
MzA
short pulse (no T1 relaxationbetween A and B, or C and D)
Bernstein et al. (2004) Handbook of MRI Pulse Sequences
G16.4427 Practical MRI 1 – 5th March 2015
Multi-Echo SE• The transverse magnetization can be
repeatedly refocused into subsequent SEs by playing additional RF refocusing pulse– The series of echoes is called an echo train– Each echo number fits its own independent k-space
• The length of the echo train is limited by T2 decay– In most cases we are interested in 2 echoes (an
early and a late one). Question: if TR is long, what contrast will have the 2 resulting images?
G16.4427 Practical MRI 1 – 5th March 2015
Multi-Echo SE• The transverse magnetization can be
repeatedly refocused into subsequent SEs by playing additional RF refocusing pulse– The series of echoes is called an echo train– Each echo number fits its own independent k-space
• The length of the echo train is limited by T2 decay– In most cases we are interested in 2 echoes (an
early and a late one). if TR is long, the two images will be PD- and T2-weighted, respectively
G16.4427 Practical MRI 1 – 5th March 2015
Example of Dual-Echo SE Acquisition
Proton density-weightedTE/TR = 17/2200 ms
T2-weightedTE/TR = 80/2200 ms
G16.4427 Practical MRI 1 – 5th March 2015
Dual-Echo SE
Bernstein et al. (2004) Handbook of MRI Pulse Sequences
G16.4427 Practical MRI 1 – 5th March 2015
T2-Mapping• It is a common application of acquiring longer echo trains
(otherwise more than two echoes per TR are rarely acquired in MRI)
• In theory we can acquire long echo train of SEs and fit the signal intensity at each pixel to calculate T2
• In practice there are systematic errors that make it difficult to fit a monoexponential decay curve– Variable flip angle across slice profile– Stimulated echoes can introduce unwanted T1-weighting
variations into the echo-train signals– If magnitude reconstruction is used, the noise floor has
nonzero mean leading to incorrectly larger T2 values
G16.4427 Practical MRI 1 – 5th March 2015
Paper Discussion
G16.4427 Practical MRI 1 – 5th March 2015
Advanced pulse sequences
G16.4427 Practical MRI 1 – 5th March 2015
Echo Planar Imaging (EPI)• EPI is one of the fastest MRI pulse sequences– 2D image in few tens of milliseconds
• Allowed developing challenging MR applications– Diffusion, perfusion, cardiac imaging, etc.
• EPI uses a gradient-echo train– Typical to produce ~100 gradient echoes to produce a
low-resolution image from a single RF excitation• More prone to a variety of artifacts– Ghosting along phase-encoded direction
G16.4427 Practical MRI 1 – 5th March 2015
Ghosting Artifacts
• Not caused by motion, but by eddy currents, imperfect gradients, field non-uniformities, or a mismatch between the timing of the even and odd echoes– Which results in mis-registration of alternating lines of k-space
Phase errors may resultfrom the multiple positiveand negative passes through k-spaceGhost artifacts in the phase direction
G16.4427 Practical MRI 1 – 5th March 2015
GRE vs. EPI• In a simple GRE pulse sequence the transverse
magnetization decays as Mxy(t) = Mxy(0)e-t/T2*
• Half lifetime of Mxy is T2*ln2– A very small fraction of the lifetime is actually used
for data acquisition in GRE• EPI maximally uses the transverse magnetization
without additional RF excitations– Bipolar oscillating readout gradient produces a series
of echoes, each individually phase-encoded
G16.4427 Practical MRI 1 – 5th March 2015
GRE and EPI
Netl = echo train length = number of echoes following RF excitation
tesp = echo spacing (typically echoes are evenly spaced)
A series of echoesis produced beforeMxy decays away dueto T2* relaxation
Bernstein et al. (2004) Handbook of MRI Pulse Sequences
G16.4427 Practical MRI 1 – 5th March 2015
EPI Readout Gradient
Starts with a prephasing gradient that position the k-space trajectory to kx,min followed by a series of readout gradient lobes with alternating polarity. Question: what is the area of the prephasing gradient lobe Gx,p?
Bernstein et al. (2004) Handbook of MRI Pulse Sequences
G16.4427 Practical MRI 1 – 5th March 2015
EPI Readout Gradient
Starts with a prephasing gradient that position the k-space trajectory to kx,min followed by a series of readout gradient lobes with alternating polarity. Answer: Half the area of the first readout gradient area.
The second half of each lobe serves as prephasing for thefollowing gradient(that is why the polaritymust alternate)
Bernstein et al. (2004) Handbook of MRI Pulse Sequences
G16.4427 Practical MRI 1 – 5th March 2015
EPI Phase-Encoding Gradient
a) Constant phase-encoding gradient throughout the entire readout echo train (ky varies linearly with time)
• Gridding is needed before reconstructionb) A series of blips with the same polarity and identical area,
each played before the acquisition of an echo
Bernstein et al. (2004) Handbook of MRI Pulse Sequences
G16.4427 Practical MRI 1 – 5th March 2015
Gradient-Echo EPI
Bernstein et al. (2004) Handbook of MRI Pulse Sequences
G16.4427 Practical MRI 1 – 5th March 2015
Spin-Echo EPI
What is different compared to GRE-EPI?
G16.4427 Practical MRI 1 – 5th March 2015
Phase Contrast (PC) Imaging• A method to image moving magnetization by
applying flow-encoding gradients– Image flow in blood vessels and CSF, track motion
• The flow-encoding gradients translates velocity into the phase of the image– Bipolar gradient, as it produces linear proportionality
• The axis of the gradient determines the direction of flow sensitivity– Normally applied to only one axis at a time
• Typically performed with GRE pulse sequences, adding phase-encoding gradients. Why?
G16.4427 Practical MRI 1 – 5th March 2015
Typical PC Pulse Sequence
Typically a bipolar gradient is added to only one of the three logical axes at a time
Toggling of the bipolar gradient (dotted lines) varies the gradient first moment and introduces flow sensitivity along that axisBernstein et al.
(2004) Handbook of MRI Pulse Sequences
G16.4427 Practical MRI 1 – 5th March 2015
PC Acquisition and Reconstruction• Two complete sets of image are acquired varying
only the 1st moment of the bipolar gradient– The amount of such operator-selected variation
determines the amount of velocity encoding• The phases of the two images are subtracted on
a pixel-by-pixel basis in image domain– Allows to quantify flow direction, flow velocity and
volume flow rate– Phase-difference or complex-difference
reconstruction methods are in common use
G16.4427 Practical MRI 1 – 5th March 2015
Diffusion Imaging• In the presence of a magnetic field gradient, diffusion
of water molecules causes a phase dispersion of the transverse magnetization– The degree of signal loss depends on tissue type,
structure, physical and physiological state• Diffusion imaging is a family of techniques– E.g. DWI, DTI, DKI
• All diffusion pulse sequences contain diffusion-weighting gradients
• DWI typically employs a single b-value, other quantitative mapping methods at least 2 b-values
G16.4427 Practical MRI 1 – 5th March 2015
Diffusion-Weighting Gradients• Typically consist of two lobes with equal area• Amplitude is the maximum allowed• Pulse width is larger than most imaging gradients• When used, water diffusion can cause an attenuation
in proton MRI signals– Degree of attenuation depends on the product between
the diffusion coefficient D and the b-value– b-value is analogous to TE in T2-weighted sequences– Increasing gradient amplitude, separation of its lobes, or
pulse width of each lobe results in a higher b-value. How does diffusion-weighting change consequently?
G16.4427 Practical MRI 1 – 5th March 2015
Diffusion Weighting in GRE and SE
Spin Echo Gradient Echo
Bernstein et al. (2004) Handbook of MRI Pulse Sequences
G16.4427 Practical MRI 1 – 5th March 2015
Single-Shot Spin-Echo EPI• The most prevalent sequence due to high
acquisition speed (e.g. < 100 ms per image)• A pair of identical gradient lobes on either side of
the refocusing pulse• Gradient direction can be controlled by varying
its vector components along the 3 axes• To minimize TE the max amplitude is used to
achieve the desired b-value• What is another way to minimize TE?
G16.4427 Practical MRI 1 – 5th March 2015
Single-Shot Spin-Echo EPI• The most prevalent sequence due to high
acquisition speed (e.g. < 100 ms per image)• A pair of identical gradient lobes on either side of
the refocusing pulse• Gradient direction can be controlled by varying
its vector components along the 3 axes• To minimize TE the max amplitude is used to
achieve the desired b-value• Maximizing SR also reduces TE, but can cause
peripheral nerve stimulation
G16.4427 Practical MRI 1 – 5th March 2015
Diffusion-Weighted Single-Shot SE
Bernstein et al. (2004) Handbook of MRI Pulse Sequences
G16.4427 Practical MRI 1 – 5th March 2015
Diffusion-Weighted (DW) Imaging• In the presence of a gradient, molecular diffusion
attenuates the MRI signal exponentially:
• Tissue with fast diffusion experiences more signal loss low intensity in the DW image
• To remove the patient-orientation dependence, 3 DW images can be obtained with a DW gradient applied along the three orthogonal directions– If same b-value then isotropic DW image
(S and S0 are the voxel signal intensity with and without diffusion)
G16.4427 Practical MRI 1 – 5th March 2015
Example: White Matter Infarct
G16.4427 Practical MRI 1 – 5th March 2015
Quantitative Apparent Diffusion Coefficient (ADC) Mapping
• A series of DW images are acquired with multiple b-values:
• A linear fit between ln(S0/Si) and bi is performed on a pixel-by-pixel basis to find D
• The contrast of the ADC map is inverted compared to a DW image
• To keep a constant TE in all DW images, b-values are typically changed by varying the diffusion-gradient amplitude instead of its duration– Contribution from imaging gradients should be included in b-
value calculation to avoid overestimating ADC
…
G16.4427 Practical MRI 1 – 5th March 2015
DW Image vs. ADC Map
G16.4427 Practical MRI 1 – 5th March 2015
Fat Suppression• Because of its short T1, the bright appearance of
fat is a problem for T1-weighted images with short TR and short TE
• Fat is a main contributor to chemical shift artifacts
• There are several methods for fat suppression– Spectrally selective RF pulses• What are the drawbacks?
G16.4427 Practical MRI 1 – 5th March 2015
Fat Suppression• Because of its short T1, the bright appearance of
fat is a problem for T1-weighted images with short TR and short TE
• Fat is a main contributor to chemical shift artifacts
• There are several methods for fat suppression– Spectrally selective RF pulses• B1 and B0 inhomogeneities, not good at low field strengths
– Short TI recovery (STIR)• B1 inhomogeneities, long scan, signal from other tissues
G16.4427 Practical MRI 1 – 5th March 2015
Two-Point Dixon Pulse SequenceIn conventional spin echo Δ = 0
Question: what happens if Δ ≠ 0
Bernstein et al. (2004) Handbook of MRI Pulse Sequences
G16.4427 Practical MRI 1 – 5th March 2015
Two-Point Dixon Pulse SequenceIn conventional spin echo Δ = 0
If Δ ≠ 0, spins with different chemical shifts will be out of phase at kx = 0 (unless phase shift happens to be a multiple of 2π)
If the 180° pulse is delayed or advanced by Δ/2, the RF spin echo is delayed or advanced by Δ relative to where kx = 0
Consider a voxel with a water and a fat spin having a frequency difference fcs and assume there are no B0 inhomogeneities
In the 2-point Dixon technique, we acquire two SE images:1. Δ = 0 and normal acquisition
• Fat and water in-phase at kx = 0 (in-phase image)2. Δ = 1/(2fcs) and RF pulse advanced or delayed by Δ/2 = 1/(4fcs)
• Fat and water 180° out-of-phase at kx = 0 (out-of-phase image)3.
ϕ = 2πfcsΔ at kx = 0
G16.4427 Practical MRI 1 – 5th March 2015
Two-Point Dixon TechniqueBecause image contrast is heavily determined by the peak signal amplitude that occurs at kx = 0, the resulting complex images are approximately given by:
I0 = W + F I1 = W – F
Separate images of the water and fat magnetization can be reconstructed from:
W = (1/2) (I0 + I1) F = (1/2) (I0 - I1)
The water image W can be used as a fat suppressed image, whereas W and F separately provide information about the relative water and fat content of tissues
G16.4427 Practical MRI 1 – 5th March 2015
Limits of Two-Point Dixon• The 2-point Dixon technique assumes perfect B0
homogeneity which is almost never true– Due to ΔB0 fat and water have accumulated an additional phase
shift Δϕ = γ(ΔB0)Δ at kx = 0 in I1
– Although fat and water spins in any given voxel are still anti-parallel in the opposed-phase image, they might no longer be parallel or anti-parallel to the fat and water spins in the in-phase image
(example of 90° phase shift caused by B0 inhomogeneities)
Question: what is wrong with the combined images?
G16.4427 Practical MRI 1 – 5th March 2015
Example: Healthy Liver
AJR April 2010, vol. 194(4), p. 964-971
G16.4427 Practical MRI 1 – 5th March 2015
Example: Fatty Liver
AJR April 2010, vol. 194(4), p. 964-971
G16.4427 Practical MRI 1 – 5th March 2015
Any questions?
G16.4427 Practical MRI 1 – 5th March 2015
See you next Thursday!