Functional Brain Signal Processing: EEG & fMRI
Lesson 11
Kaushik Majumdar
Indian Statistical Institute Bangalore Center
M.Tech. (CS), Semester III, Course B50
![Page 2: Functional Brain Signal Processing: EEG & fMRI Lesson 11 Kaushik Majumdar Indian Statistical Institute Bangalore Center kmajumdar@isibang.ac.in M.Tech.](https://reader031.fdocuments.in/reader031/viewer/2022013012/5697bfcf1a28abf838ca9b67/html5/thumbnails/2.jpg)
T1 and T2 Relaxation Time
Buxton, 2009
Clearly, T2 << T1. Let T2*
be partly due to T2 of homogeneous magnetic property of the tissue and partly due to inhomogeneous magnetic property of the same. So
*2 2 2
1 1 1
T T T
where T’2 is time delay due to pure inhomogeneity. So
T2* < T2 << T1.
Buxton, 2009, p. 148
![Page 3: Functional Brain Signal Processing: EEG & fMRI Lesson 11 Kaushik Majumdar Indian Statistical Institute Bangalore Center kmajumdar@isibang.ac.in M.Tech.](https://reader031.fdocuments.in/reader031/viewer/2022013012/5697bfcf1a28abf838ca9b67/html5/thumbnails/3.jpg)
Free Induction Decay (FID)
Buxton, 2009
About one million oscillations during the T2
*.
![Page 4: Functional Brain Signal Processing: EEG & fMRI Lesson 11 Kaushik Majumdar Indian Statistical Institute Bangalore Center kmajumdar@isibang.ac.in M.Tech.](https://reader031.fdocuments.in/reader031/viewer/2022013012/5697bfcf1a28abf838ca9b67/html5/thumbnails/4.jpg)
Repetition Time (TR)
Buxton, 2009
![Page 5: Functional Brain Signal Processing: EEG & fMRI Lesson 11 Kaushik Majumdar Indian Statistical Institute Bangalore Center kmajumdar@isibang.ac.in M.Tech.](https://reader031.fdocuments.in/reader031/viewer/2022013012/5697bfcf1a28abf838ca9b67/html5/thumbnails/5.jpg)
Spin Echo and Echo Time (TE)
Buxton, 2009
![Page 6: Functional Brain Signal Processing: EEG & fMRI Lesson 11 Kaushik Majumdar Indian Statistical Institute Bangalore Center kmajumdar@isibang.ac.in M.Tech.](https://reader031.fdocuments.in/reader031/viewer/2022013012/5697bfcf1a28abf838ca9b67/html5/thumbnails/6.jpg)
Fast fMRI
In conventional phase encoding if TR = 2 s and 256 lines are to be drawn, it will take 8 minutes 32 seconds for generating one image. If we reduce the number of lines to 64 the time required is 2 minutes 8 seconds.
Our ultimate goal is to get as much information from fMRI as possible from LFP.
One way to address the problem is to go for ‘fast’ fMRI, at least, as fast as we can. EPI is one paradigm for fast fMRI.
![Page 7: Functional Brain Signal Processing: EEG & fMRI Lesson 11 Kaushik Majumdar Indian Statistical Institute Bangalore Center kmajumdar@isibang.ac.in M.Tech.](https://reader031.fdocuments.in/reader031/viewer/2022013012/5697bfcf1a28abf838ca9b67/html5/thumbnails/7.jpg)
Fast Spin Echo Technique
Mezrich, 1995
![Page 8: Functional Brain Signal Processing: EEG & fMRI Lesson 11 Kaushik Majumdar Indian Statistical Institute Bangalore Center kmajumdar@isibang.ac.in M.Tech.](https://reader031.fdocuments.in/reader031/viewer/2022013012/5697bfcf1a28abf838ca9b67/html5/thumbnails/8.jpg)
Echo Planar Imaging (EPI)
Deichmann et al., 2010
K space in EPI
![Page 9: Functional Brain Signal Processing: EEG & fMRI Lesson 11 Kaushik Majumdar Indian Statistical Institute Bangalore Center kmajumdar@isibang.ac.in M.Tech.](https://reader031.fdocuments.in/reader031/viewer/2022013012/5697bfcf1a28abf838ca9b67/html5/thumbnails/9.jpg)
EPI (cont)
Deichmann et al., 2010
![Page 10: Functional Brain Signal Processing: EEG & fMRI Lesson 11 Kaushik Majumdar Indian Statistical Institute Bangalore Center kmajumdar@isibang.ac.in M.Tech.](https://reader031.fdocuments.in/reader031/viewer/2022013012/5697bfcf1a28abf838ca9b67/html5/thumbnails/10.jpg)
Safety Issues in EPI
Specific absorption rate (SAR). Up to 130 dB acoustic noise due to rapid switching
(about 1000 Hz) of gradient magnetic field. (A major issue in EPI.)
Nerve stimulation. Y gradient is generally not used for frequency encoding in EPI. (A major issue in EPI.)
Generation of heat in body tissues due to RF pulses is always there in MR imaging including in EPI.
Buxton, 2009, p. 249
![Page 11: Functional Brain Signal Processing: EEG & fMRI Lesson 11 Kaushik Majumdar Indian Statistical Institute Bangalore Center kmajumdar@isibang.ac.in M.Tech.](https://reader031.fdocuments.in/reader031/viewer/2022013012/5697bfcf1a28abf838ca9b67/html5/thumbnails/11.jpg)
References
R. B. Buxton, Introduction to Functional Magnetic Resonance Imaging, 2e, Cambridge University Press, Cambridge, UK, 2009.