FRACTION COLLECTION FOR ISOLATING … COLLECTION FOR ISOLATING IMPURITIES IN FORCED DEGRADATION...

1
FRACTION COLLECTION FOR ISOLATING IMPURITIES IN FORCED DEGRADATION STUDIES Paula Hong and Patricia R. McConville Waters Corporation, Milford, MA 01757 TO DOWNLOAD A COPY OF THIS POSTER, VISIT WWW.WATERS.COM/POSTERS ©2017 Waters Corporation INTRODUCTION Forced degradation studies are typically performed to understand the degradation pathway of pharmaceuticals. One of the specific challenges includes determining the response factor of impurities relative to the active pharmaceutical ingredient (API). Incorrectly identifying the relative response factors (RRFs) could lead to over or under quantification of the impurity, which can in turn lead to mass imbalance. In this presentation the impurity will be collected from the forced degradation analysis by small scale fraction collection and subsequently used to determine RRFs. The RRFs will be also evaluated by established methodologies, specifically comparison of the calibration curve of both the API and impurity standard. METHODS Conditions for forced degradation studies: System: ACQUITY UPLC H-Class with PDA and QDa Detector Column: ACQUITY UPLC BEH C18 1.7μm, 2.1 x 50 mm Column Temperature: 30 °C Injection volume: 4 μL (or 10 μL for scale up experiment) Mobile phase (prepared using AutoBlend): B125 mM Ammonium hydroxide CWater; DAcetonitrile Flow rate: 0.6 mL/min Wavelength: 254 nm MS Settings: Mode: Electrospray (+) Mass range: 50-500 m/z Capillary Voltage: 1.5 kV Cone Voltage:15 V Single Ion Recording channels (SIR): 383.3, 399.3 Make up pump (ISM): 0.3 mL/min, 0.1% formic acid in methanol Conditions for collection scaleup: System: ACQUITY UPLC H-Class with PDA Detector and WFMA Column: XBridge BEH C18 2.5 μm, 3.0 x 75 mm Column Temperature: 30 °C Injection volume: 30.6 μL Mobile phase (same as described for forced degradation) Flow rate: 0.833 mL/min Sample preparation: Oxidation of loratadine drug substance Loratadine and related impurities (n-oxide and epoxide) were purchased from the Toronto Research Chemicals. All standards were dissolved in 1:1 methanol:water and sonicated. The loratadine drug substance was exposed to oxidative conditions (3% H 2 O 2 ) at 70 ˚C for up to 90 minutes. Sample preparation for collected fractions Approximately 70 injections (over a period of 23 hours) of the forced degradation sample were run on a 3.0 x 75 mm column. Both impurity 1 (n- oxide) and impurity 2 (epoxide) were collected and pooled (per peak). The samples were then dried down to remove the organic portion, frozen and then lyophilized. The dried samples were reconstituted with 500 μL of methanol. References 1. Chapter <621> CHROMATOGRAPHY United States Pharmacopeia and National Formulary (USP 37-NF 32 S1) Baltimore, MD: United Book Press, Inc.; 2014. p. 6376-85. CONCLUSION Fraction collection of stressed drug substance can be performed on an analytical scale for multiple peaks in a single analysis Collection and pooling of multiple injections can be used to acquire micrograms of material Collected samples can be used to assess relative response factors by comparison of the standard curves to that of the drug substance. RELATIVE RESPONSE FACTORS To evaluate the relative response factors (RRF) for both related impurities, the standard curves were used. This approach uses the ratio of the slopes of the impurity and API calibration curves. 1 This requires a known amount of sample. For the lyophilized samples, weighing the small amounts of the impurities was challenging. Therefore, the amount of each impurity collected was calculated using calibration curves of the readily available standards. This also allowed for confirmation of the n-oxide and epoxide impurities. RESULTS AND DISCUSSION Time %A %B %C %D 0.00 0 10 60 30 5.00 0 10 35 55 7.00 0 10 10 80 FORCED DEGRADTAION ANALYSIS AND FRACTION COLLECTION Figure 2. Pathway for oxidative degradation of loratadine. Figure 5. Software used to set up fraction collection. Collection Event Table (upper right) allows events to be imported from a processed chromatogram (or result) using Fill From Chromatogram (upper left). In this window a peaks table is populated based on selected result. User has the option to select which peaks to import into collection table. Simulation option allows analyst to visualize peak collection as compared to a chromatogram (lower views). Figure 4 Chromatograms on both a 2.1 x 50mm and 3.0 x 75 mm column, the latter of which was used for fraction collection. Scaling to the 3.0 mm ID column allowed a larger injection volume per injection. Multiple injections were pooled to collect adequate amount of sample. PDA Spectrum PDA 254.0 nm (PDA Spectrum (210-400)nm) 440.2 415.2 399.2 399.2 383.2 AU 0.00 1.00 2.00 3.00 TIC, Smoothed by 5 point Mean Filter., Time offset by -0.010 mins. (QDa Positive(+) Scan (50.00-500.00)Da, Centroid, CV=15) Intensity 0 2x10 7 4x10 7 6x10 7 8x10 7 Minutes 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50 Figure 3. Stacked view of UV and mass total ion (TIC) chromatogram of forced degradation of loratadine drug substance with base mass labels. The drug substance was exposed to oxidative conditions at 70 °C for 90 min. A major impurity peak (76% area) was observed at 2.616 min, while a minor peak (0.88% area) was observed at 3.828 min. Both peaks had a base mass of 399.2 suggesting oxidation of the API at a single site. UV @ 254 nm TIC Figure 6. UV chromatogram of lyophilized fractions 1 and 2. Samples were reconstituted in 500 μL of methanol. Quantification of the fractions showed collection of 188 μg of N-oxide and 12 μg of epoxide. Presence of additional peaks in fraction 2 may be attributed to degradation or conversion of the epoxide impurity during lyophilization. Table 1. Calibration curves and relative response factors (RRF) for loratadine and degradation products. Upper limit of calibration curve range for collected samples was limited by the amount of samples. RRF values for both sets of samples were comparable and within acceptable range. Compound Range (μg/mL) R 2 Slope RRF Loratadine 1-500 0.996 16332632 1.0 Standards N-Oxide 1-500 0.998 16150615 1.1 Epoxide 1-500 1.000 3794973 0.2 Collected Fractions N-Oxide 1377 0.998 19771229 1.2 Epoxide 0.8-23 1.000 5021606 0.3 Time Total Area (μV*sec) Apparent Mass Balance Total Adjusted Area *(μV*sec) Corrected Mass Balance* Reference 534488 100.0 N/A N/A 30 min 596388 112 582878 109 60 min 604194 113 581550 109 90 min 629781 118 592461 111 Table 2. Comparison of mass balance calculations without and with (*) RRF corrections. Use of RRF resulted in a decrease in mass balance with values closer to 100% than the apparent mass balance. Figure 7. Oxidation of loratadine at various time points over 90 minutes. With increasing time, the drug substance peak area decreased and was accompanied by a significant increase of noxide (49% at 90 minutes). The epoxide impurity increased minimally from 0.6 to 0.9 % (see inset). MASS BALANCE To assess the impact of RRF, oxidative degradation of loratadine drug substance was performed at 70 °C. The samples were analyzed after 30, 60 and 90 minutes. Mass balance calculations were performed using the peak area. Peak areas for n-oxide and epoxide impurities were adjusted using the RRF values (*). In general, RRF > 1.0 result in a decrease in peak area, while RRF < 1.0 result in an increase in peak area. Given the relative amount of n-oxide formed in the degradation and its RRF of >1.0, a decrease in mass balance was observed using the RRF values. N- Oxide - 2.616 - QDa Positive Scan Apex 211.9 390.7 399.2 401.2 Epoxide - 3.828 - QDa Positive Scan Apex 266.9 396.3 399.2 401.2 API - 4.854 - QDa Positive Scan Apex 248.5 373.9 383.2 385.2 Figure 1. Waters Fraction Manager - Analytical (WFMA) N-oxide epoxide loratadine loratadine epoxide N-oxide H 2 O 2 H 2 O 2 2.1 x 50 mm, 1.7 μm 10 μL injection Flow rate: 0.6 mL/min 3.0 x 75 mm, 2.5 μm 30.6 μL injection Flow rate: 0.833 mL/min loratadine N-oxide epoxide AU 0.00 0.20 0.40 0.60 0.80 Minutes 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00 11.00 12.00 13.00 AU 0.00 0.20 0.40 0.60 0.80 1.00 Minutes 0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50 6.00 N-oxide epoxide loratadine Ref 30 min 60 min 90 min AU -0.002 0.000 0.002 0.004 0.006 Minutes 3.70 3.75 3.80 3.85 3.90 3.95 AU -0.02 0.00 0.02 0.04 0.06 0.08 0.10 0.12 0.14 0.16 0.18 0.20 0.22 0.24 Minutes 2.20 2.40 2.60 2.80 3.00 3.20 3.40 3.60 3.80 4.00 4.20 4.40 4.60 4.80 5.00 5.20 5.40 5.60 Fraction 1 n-oxide Fraction 2 epoxide UV @ 254 nm UV @ 254 nm AU 0.00 0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 AU 0.000 0.002 0.004 0.006 0.008 0.010 0.012 0.014 0.016 Minutes 1.40 1.60 1.80 2.00 2.20 2.40 2.60 2.80 3.00 3.20 3.40 3.60 3.80 4.00 4.20 4.40 4.60 4.80 5.00 5.20 5.40 5.60 5.80 1.80 0.018

Transcript of FRACTION COLLECTION FOR ISOLATING … COLLECTION FOR ISOLATING IMPURITIES IN FORCED DEGRADATION...

Page 1: FRACTION COLLECTION FOR ISOLATING … COLLECTION FOR ISOLATING IMPURITIES IN FORCED DEGRADATION STUDIES ... of forced degradation of loratadine drug …

FRACTION COLLECTION FOR ISOLATING IMPURITIES IN FORCED DEGRADATION STUDIES

Paula Hong and Patricia R. McConville Waters Corporation, Milford, MA 01757

TO DOWNLOAD A COPY OF THIS POSTER, VISIT WWW.WATERS.COM/POSTERS

©2017 Waters Corporation

INTRODUCTION

Forced degradation studies are typically performed to understand

the degradation pathway of pharmaceuticals. One of the specific

challenges includes determining the response factor of impurities

relative to the active pharmaceutical ingredient (API). Incorrectly

identifying the relative response factors (RRFs) could lead to over

or under quantification of the impurity, which can in turn lead to

mass imbalance. In this presentation the impurity will be collected

from the forced degradation analysis by small scale fraction

collection and subsequently used to determine RRFs. The RRFs

will be also evaluated by established methodologies, specifically

comparison of the calibration curve of both the API and impurity

standard.

METHODS Conditions for forced degradation studies: System: ACQUITY UPLC H-Class with PDA and QDa Detector Column: ACQUITY UPLC BEH C18 1.7µm, 2.1 x 50 mm Column Temperature: 30 °C Injection volume: 4 µL (or 10 µL for scale up experiment) Mobile phase (prepared using AutoBlend):

B– 125 mM Ammonium hydroxide C– Water; D– Acetonitrile Flow rate: 0.6 mL/min Wavelength: 254 nm

MS Settings: Mode: Electrospray (+) Mass range: 50-500 m/z Capillary Voltage: 1.5 kV Cone Voltage:15 V Single Ion Recording channels (SIR): 383.3, 399.3 Make up pump (ISM): 0.3 mL/min, 0.1% formic acid in methanol

Conditions for collection scale– up:

System: ACQUITY UPLC H-Class with PDA Detector and WFMA Column: XBridge BEH C18 2.5 µm, 3.0 x 75 mm Column Temperature: 30 °C Injection volume: 30.6 µL Mobile phase (same as described for forced degradation) Flow rate: 0.833 mL/min

Sample preparation: Oxidation of loratadine drug substance

Loratadine and related impurities (n-oxide and epoxide) were purchased from the Toronto Research Chemicals. All standards were dissolved in 1:1 methanol:water and sonicated. The loratadine drug substance was exposed to oxidative conditions (3% H2O2) at 70 ˚C for up to 90 minutes.

Sample preparation for collected fractions

Approximately 70 injections (over a period of 23 hours) of the forced degradation sample were run on a 3.0 x 75 mm column. Both impurity 1 (n-oxide) and impurity 2 (epoxide) were collected and pooled (per peak). The samples were then dried down to remove the organic portion, frozen and then lyophilized. The dried samples were reconstituted with 500 µL of methanol.

References 1. Chapter <621> CHROMATOGRAPHY United States Pharmacopeia and National Formulary (USP 37-NF 32 S1)

Baltimore, MD: United Book Press, Inc.; 2014. p. 6376-85.

CONCLUSION

Fraction collection of stressed drug substance can be performed on an analytical scale for multiple peaks in a single

analysis

Collection and pooling of multiple injections can be used to

acquire micrograms of material

Collected samples can be used to assess relative response factors by comparison of the standard curves to that of the

drug substance.

RELATIVE RESPONSE FACTORS

To evaluate the relative response factors (RRF) for both related impurities, the standard curves were used. This approach uses the ratio of the slopes of the impurity and API calibration curves.

1 This

requires a known amount of sample. For the lyophilized samples, weighing the small amounts of the impurities was challenging. Therefore, the amount of each impurity collected was calculated using calibration curves of the readily available standards. This also allowed for confirmation of the n-oxide and epoxide impurities.

RESULTS AND DISCUSSION

Time %A %B %C %D

0.00 0 10 60 30

5.00 0 10 35 55

7.00 0 10 10 80

FORCED DEGRADTAION ANALYSIS AND FRACTION COLLECTION

Figure 2. Pathway for oxidative degradation of loratadine.

Figure 5. Software used to set up fraction collection. Collection Event Table (upper right) allows events to be imported from a processed chromatogram (or result) using Fill From Chromatogram (upper left). In this window a peaks table is populated based on selected result. User has the option to select which peaks to import into collection table. Simulation option allows analyst to visualize peak collection as compared to a chromatogram (lower views).

Figure 4 Chromatograms on both a 2.1 x 50mm and 3.0 x 75 mm column, the latter of which was used for fraction collection. Scaling to the 3.0 mm ID column allowed a larger injection volume per injection. Multiple injections were pooled to collect adequate amount of sample.

PDA Spectrum PDA 254.0 nm (PDA Spectrum (210-400)nm)

440.2

415.2

399.2

399.2

383.2

AU

0.00

1.00

2.00

3.00

QDa Positive Scan MS TIC, Smoothed by 5 point Mean Filter., Time offset by -0.010 mins. (QDa Positive(+) Scan (50.00-500.00)Da, Centroid, CV=15)

Inte

nsity

0

2x107

4x107

6x107

8x107

Minutes

1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50

Figure 3. Stacked view of UV and mass total ion (TIC) chromatogram of forced degradation of loratadine drug substance with base mass labels. The drug substance was exposed to oxidative conditions at 70 °C for 90 min. A major impurity peak (76% area) was observed at 2.616 min, while a minor peak (0.88% area) was observed at 3.828 min. Both peaks had a base mass of 399.2 suggesting oxidation of the API at a single site.

UV @ 254 nm

TIC

Figure 6. UV chromatogram of lyophilized fractions 1 and 2. Samples were reconstituted in 500 µL of methanol. Quantification of the fractions showed collection of 188 µg of N-oxide and 12 µg of epoxide. Presence of additional peaks in fraction 2 may be attributed to degradation or conversion of the epoxide impurity during lyophilization.

Table 1. Calibration curves and relative response factors (RRF) for loratadine and degradation products. Upper limit of calibration curve range for collected samples was limited by the amount of samples. RRF values for both sets of samples were comparable and within acceptable range.

Compound Range (µg/mL)

R2 Slope RRF

Loratadine 1-500 0.996 16332632 1.0

Standards

N-Oxide 1-500 0.998 16150615 1.1

Epoxide 1-500 1.000 3794973 0.2

Collected Fractions

N-Oxide 1– 377 0.998 19771229 1.2

Epoxide 0.8-23 1.000 5021606 0.3

Time Total Area(μV*sec)

Apparent Mass Balance

Total Adjusted Area *(μV*sec)

Corrected Mass Balance*

Reference 534488 100.0 N/A N/A

30 min 596388 112 582878 109

60 min 604194 113 581550 109

90 min 629781 118 592461 111

Table 2. Comparison of mass balance calculations without and with (*) RRF corrections. Use of RRF resulted in a decrease in mass balance with values closer to 100% than the apparent mass balance.

Figure 7. Oxidation of loratadine at various time points over 90 minutes. With increasing time, the drug substance peak area decreased and was accompanied by a significant increase of n–oxide (49% at 90 minutes). The epoxide impurity increased minimally from 0.6 to 0.9 % (see inset).

MASS BALANCE

To assess the impact of RRF, oxidative degradation of loratadine drug substance was performed at 70 °C. The samples were analyzed after 30, 60 and 90 minutes.

Mass balance calculations were performed using the peak area. Peak areas for n-oxide and epoxide impurities were adjusted using the RRF values (*). In general, RRF > 1.0 result in a decrease in peak area, while RRF < 1.0 result in an increase in peak area. Given the relative amount of n-oxide formed in the degradation and its RRF of >1.0, a decrease in mass balance was observed using the RRF values.

N- Oxide - 2.616 - QDa Positive Scan

Apex

211.9

390.7

399.2

401.2

Epoxide - 3.828 - QDa Positive Scan

Apex

266.9396.3

399.2

401.2

API - 4.854 - QDa Positive Scan

Apex

248.5

373.9

383.2

385.2

Figure 1. Waters Fraction Manager - Analytical (WFMA)

N-oxide epoxide loratadine

loratadine

epoxide N-oxide

H2O2 H2O2

2.1 x 50 mm, 1.7 µm 10 µL injection Flow rate: 0.6 mL/min

3.0 x 75 mm, 2.5 µm 30.6 µL injection Flow rate: 0.833 mL/min

loratadine

N-oxide

epoxide

AU

0.00

0.20

0.40

0.60

0.80

Minutes

1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00 11.00 12.00 13.00

AU

0.00

0.20

0.40

0.60

0.80

1.00

Minutes

0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50 6.00

N-oxide

epoxide

loratadine Ref 30 min 60 min 90 min

AU

-0.002

0.000

0.002

0.004

0.006

Minutes

3.70 3.75 3.80 3.85 3.90 3.95

AU

-0.02

0.00

0.02

0.04

0.06

0.08

0.10

0.12

0.14

0.16

0.18

0.20

0.22

0.24

Minutes

2.20 2.40 2.60 2.80 3.00 3.20 3.40 3.60 3.80 4.00 4.20 4.40 4.60 4.80 5.00 5.20 5.40 5.60

Fraction 1 n-oxide

Fraction 2 epoxide

UV @ 254 nm

UV @ 254 nm

AU

0.00

0.20

0.40

0.60

0.80

1.00

1.20

1.40

1.60

AU

0.000

0.002

0.004

0.006

0.008

0.010

0.012

0.014

0.016

0.018

Minutes

1.40 1.60 1.80 2.00 2.20 2.40 2.60 2.80 3.00 3.20 3.40 3.60 3.80 4.00 4.20 4.40 4.60 4.80 5.00 5.20 5.40 5.60 5.80

1.80

0.018