Fourier Holography

download Fourier Holography

of 3

Transcript of Fourier Holography

  • 8/3/2019 Fourier Holography

    1/3

    Fourier Holography

    Sahand NoorizadehSchool of Electrical and Computer Engineering

    Portland State University Portland, Oregon 97201

    Email: [email protected]

    AbstractIn this paper the limitations of direct imaging andthe need for holography is explained. The fundamental conceptof holography is presented and two types of Fourier holographytechniques with and without lens are discussed in detail withmathematical derivations. Furthermore, a brief comparison ofthe available recording media and their limitations are outlined.Lastly, an application of the Fourier holography in the biomedicalmicroscopy is explored.

    I. INTRODUCTION

    All the available optical recording media such as film or

    CCD are only capable of recording the intensity of the light.The intensity being a time-averaged quantity does not carry

    any phase information. Rather it is proportional to the power of

    the optical wave incident on the recording medium from which

    only the amplitude can be obtained. A direct recorded image

    of an object only has information about the amplitude of the

    light wave received from the object and the phase information

    is lost. In most practical applications of optics, it is the phase

    information that is of interest. The phase of a traveling wave

    with wavelength is proportional to the distance traveled bythe wave. As shown in Figure 1, two identical waves originated

    from the same point, traveling in two different directions, and

    observed at the distance x have two different amplitudes. If

    multiple observations of the amplitude of the two waves atpoint x are made over a long period of time as the waves traveland then these values are averaged, the resulting amplitude

    values for both points on the observation line would be the

    same. That is the reason why the phase information is lost in

    the intensity measurements.

    x

    o

    Fig. 1. Phase vs. traveling distance of a wave.

    Holography is a method of recording optical interference

    of light from an object with a reference light to be able

    to reconstruct the image of the recorded object. Until the

    invention of coherent light sources such as laser, holography

    was not entirely feasible and practical because in order to be

    able to form well-defined and measurable fringe patterns by

    the means of interference, the wavelength of the light source

    needs to be stable and coherent.

    Assuming two waves a(x, y) and A(x, y) represented bytheir phasor, their interference (superposition) is given by Eq.

    1 and the intensity of their interference is given by Eq. 2

    which is the magnitude squared of the amplitude interference

    function.

    B(x, y) = |a(x, y)| ei(x,y) + |A(x, y)| ei(x,y) (1)

    I(B)=|a(x, y)|2 + |A(x, y)|2 + 2|a(x, y)||A(x, y)|

    cos[(x, y) (x, y)] (2)

    The interference allows the phase difference of the waves to be

    preserved. If the phase of one of the interfered waves is known,

    the phase of the other wave can be found. In holography, the

    behavior of one of the two waves is known (the reference

    wave) and the other is the scattered wave from a subject whose

    bahavior will be measured.

    The options of arrangement of the recording setup (i.e.

    position of the object with respect to the recording medium

    and the reference wave) has led to a wide range of classes

    of holography. Fore example, depending on the distance of

    the object from the recording medium, the propagation of

    light waves from the object could be best characterized by

    the Fresnel (near-field) propagation law or by the Fraunhofer

    (far-field) propagation law. The different effects of each of

    the preceding arrangements on the interference pattern at the

    recording plane has been the cause of different classification

    of holography systems. Another type of holography is defined

    by the angle of illumination: on-axis and off-axis. One other

    type of holography that is the subject of the remaining of this

    paper is Fourier holography.

    I I . FOURIER HOLOGRAPHY

    In Fourier holography, the Fourier transform (FT) of the

    objects amplitude transmittance is recorded. To achieve this,

    there are two methods used: a) Fourier Transform Holography

    with Lens and b) Lens-Less Fourier Holography.

    A. Fourier Transform Hologram with Lens

    In the first method, a lens is used to place the object and

    the reference wave at the back focal plane of the lens and

    record the FT of the interference of the the reference and the

    objects transmittance function. Figure 2 shows a setup of a

    FT hologram where the object transmittance function O(x, y)

  • 8/3/2019 Fourier Holography

    2/3

    is illuminated by a coherent plane wave which is also incident

    on a smaller lens, L0, separated by its focal length from theobject plane to convert the incident plane wave into a point

    source, (x a, y b), that is located at point (a, b) on the(x, y) plane. The lens L1 performs the FT operation. The field

    Recording

    Medium

    Illumination

    ff

    Fourier Plane

    L1

    L0

    z

    Fig. 2. Fourier holography with lens.

    distribution on the (, ) plane is given by Eq. 3.

    U(fX , fY)= F{O(x, y) + (x a, y b)}

    = F{O(x, y)} + F{(x a, y b)}

    = Q(fX , fY) + ei2(afX+bfY) (3)

    Where Q(fX , fY) is the FT of O(x, y), fX = /f, fY =/f, and f is the focal length of L1. The recorded intensityis given by Eq. 4.

    I(fX , fY)=1 + |Q(fX , fY)|2 + Q(fX , fY) e

    i2(afX+bfY)

    + Q(fX , fY) ei2(afX+bfY) (4)

    The recording of this intensity (whether on a film or with

    a CCD) will produce a transmittance function that can beassumed is linearly proportional to the intensity of Eq. 4.

    Therefore, for reconstructing the image of the object, a plane

    wave of the same wavelength can be used to illuminate

    this transmittance function which will in turn generate a

    wavefront, W whose complex amplitude immediately passedthe transparency (the zero propagation length) is the same

    as the transmittance function. In the Fourier transform of W(done either numerically using a computer or with a lens) the

    first two terms in Eq. 4 will produce zero-order (DC) terms

    and the last two terms will reproduce two inverted images

    of the original object centered at (a,b) and (a, b). Theimages are inverted because a double FT had to be performed

    and the Fourier transform of the Fourier transform of a

    function returns the inverted-domain version of that function:

    F{F{f(x)}} = f(x).

    B. Lensless Fourier Hologram

    In the previous section, the Fourier transforming properties

    of the lens were exploited to perform the FT operation.

    However it is possible to FT the transmittance function of an

    object without employing a lens. Figure 3 shows a holography

    system in which the object is illuminated with a plane wave

    and a reference point source is located on the same plane d unitdistance away from the recording medium at point (a, b) ofthe object plane as the object. It is necessary that the reference

    wave and the object be on the same plane.

    Recording

    MediumIllumination

    z

    Point Source d

    Fig. 3. Lensless Fourier hologram.

    Since the objects illumination is a plane wave, the ampli-

    tude of the light distribution to the immediate right hand side

    of the object is simply the transmittance function of the object.

    The distance d is chosen so that the propagation of the lightfrom the object can be expressed by Fresnel diffraction given

    by Eq 6. Where the first term is constant phase factor, the

    second term is a quadratic phase exponential, and the integral

    is the FT of the product of the object transmittance function

    O(x, y) and a quadratic phase exponential.

    Ui(, )=eikd

    ideik

    2d(2+2)

    O(x, y)e

    ik

    2d(x2+y2) eik/d(x+y)dxdy (5)

    = C eik

    2d(2+2) Foe(fX , fY); (6)

    Eq. 6 is the compact form of Eq. 5 where Foe(fX , fY) =F{O(x, y) eik

    2d(x2+y2)}, fX = /d, and fY = /d.

    From the reference point source a spherical wave propagates

    towards the (, ) plane. The propagation of this wave is givenby Eq. 7.

    Ur(, )= eik

    2d(2+2) eik/d(a+b)

    = eik

    2d(2+2) ei2(fXa+fYb) (7)

    The field distribution at the holograms plane is the super-

    position of the diffracted object field and the reference wave

    and the intensity of this superposition is given by Eq. 8.

    I(fX , fY)=ADC + C Foe e

    i2(fXa+fYb)

    + C Foe ei2(fXa+fYb) (8)

    Where ADC is the sum of all zero-order terms. The quadraticphase factor at the hologram plane e

    ik

    2d(2+2), that was

    common in both Ur and Ui, was cancelled in the intensity.Eq. 8 is very similar to Eq. 4 except that there is an additional

    constant phase factor which can be dropped and the image

    contains a quadratic phase exponential.

    In reconstructing the original object, a plane wave can be

    used as in the Fourier hologram with lens but a lens will be

  • 8/3/2019 Fourier Holography

    3/3

    required to remove the quadratic phase exponential. A more

    common way is to reconstruct with a point source similar to

    the one used for the reference in the recording process which

    will make the virtual reconstructed image coincide with the

    object.

    III. RECORDING MEDIUM

    The recording medium mentioned so far was a generic term.

    The common options are film and charged coupled device

    sensors (CCD). Films require to be developed to be used as

    transparency for reconstruction of the image of the object. The

    process of preparing the film for reconstruction is often tedious

    and time-consuming. This can be specially a disadvantage if

    an application requires multiple and fast exposures. Also the

    reconstruction process is a manual and analog process which

    complicates data processing. The advantage of films are their

    very high angular resolution compared to CCDs.

    A CCD is a two-dimensional array of NM square sensorsand they also can only record the intensity. The resolution of

    CCD sensors is a function of the array size and the pixel

    dimension. Each pixel samples the intensity in its coveragearea. For a fixed boundary, the more pixels the higher the

    sampling rate. Therefore, smaller pixels are desirable. The

    limiting resolution of the CCD camera is determined by its

    Nyquist limit. This is defined as being one half of the sampling

    frequency (i.e. # pixels/mm).

    The angle between the reference beam and the object beam

    in the holographic setup is limited because the holographic

    fringe structures in the hologram plane need to be sampled

    by the CCD sensor. The sampling theorem requires that the

    angle between the object beam and the reference beam at any

    point of the CCD sensor be limited in such a way that the

    microinterference fringe spacing is larger than double the pixel

    size [1].The use of CCD allows the numerical reconstruction using

    computers. This provides easy data processing capability such

    as filtering. Also, the holograms of different object states in

    holographic interferometry can be reconstructed with different

    wavelengths and still interfere numerically. This is of particular

    interest for multiple-wavelength techniques that are used for

    holographic contouring [2].

    IV. AN APPLICATION OF FOURIER HOLOGRAPHY

    The digital Fourier hologram shown in Figure 4 can be used

    to measure the angular spectrum of the elastically scattered

    light at many spatial locations covering a large field of view

    based on a single capture or a few image captures [3].In this hologram, the beam splitter B1 splits the laser beam

    and polarizes it using P1 for a uniform illumination that isapplied to the sample at an angle in as reflected by the mirrorM1. Since the samples are placed inside a medium with adifferent index of refraction, there will be a change in the

    angles of entrance and exit beam from the sample container.

    Lens L1 is a focal length (of L1) away from the sample toFourier transform the backscattered light from the sample and

    L2 and L3 transfer the image of this spectrum to the CCD

    Fig. 4. Schematic of the setup for the spatially resolved Fourier holographiclight scattering angular spectroscopy[3].

    sensor. The other split beam from B1 is expanded by thetelescopic system T and routed by the mirror M3 to interferewith the image of the spectrum of backscattered light from

    the sample at B2 to form a holographic interference patternat the CCD plane. This reference beam is superimposed with

    the image of the spectrum at an angle of 2.3 so that the twin

    images become separable during reconstruction.

    The complex spectrum obtained by lens L1 is proportionalto the size of the scatterers and the refractive index of

    their container. In the Fourier plane there is a one-to-one

    correspondence between spatial position and scattering angle

    [4]. The holographic technique provides a reference beam

    to interfere with this spectrum so that it is the interference

    that is recorded not the spectrum itself. This way, very smallscattering angles beyond the spatial resolution limits of the

    CCD can be encoded into the interference pattern and then

    numerically be reconstructed. The analysis of the spectrum

    of the backscattered light waves can reveal information about

    the features of the sample and this method has been used

    on biological samples to deduce morphological information at

    all points in the field of view. Combining the Mie or other

    scattering theories will extract scatterer sizes and refractive

    index contrasts [3].

    REFERENCES

    [1] U. Schnars, Direct phase determination in hologram interferometry with

    use of digitally recorded holograms, J. Opt. Soc. Am. A 11, 20112015(1994).

    [2] Christoph Wagner, Snke Seebacher, Wolfgang Osten, and Werner Jptner,Digital Recording and Numerical Reconstruction of Lensless FourierHolograms in Optical Metrology, Appl. Opt. 38, 4812-4820 (1999).

    [3] Sergey A. Alexandrov, Timothy R. Hillman, and David D. Sampson,Spatially resolved Fourier holographic light scattering angular spec-troscopy, Opt. Lett. 30, 3305-3307 (2005).

    [4] M. T. Valentine, A. K. Popp, D. A. Weitz, and P. D. Kaplan, Microscope-based static light-scattering instrument, Opt. Lett. 26, 890-892 (2001)