Foundation GPC Part 2 – Basic Gel Permeation Chromatography.
-
date post
19-Dec-2015 -
Category
Documents
-
view
250 -
download
0
Transcript of Foundation GPC Part 2 – Basic Gel Permeation Chromatography.
Foundation GPC Part 2 – Basic Gel Permeation Chromatography
2
Introduction
This presentation introduces gel permeation chromatography (GPC)
The mechanism of GPC shall be discussed
The method of analysing data shall be introduced
3
Types of Liquid Chromatography
Interactive adsorption, partition, ion exchange, etc
Non-interactive GPC, SEC, GFC
4
Nomenclature
Gel Permeation Chromatography GPC
Size Exclusion Chromatography SEC
Gel Filtration Chromatography GFC
5
1. Reciprocating piston pump delivers eluent from reservoir at a constant volumetric flow rate. GPC is always an isocratic separation
2. Two position injection valve permits introduction of sample solution without interrupting solvent flow. GPC tends to use larger injection volumes than HPLC (typically up to 200ul)
3. GPC columns perform a separation based on the molecular size of polymer molecules in solution. Resolution and/or resolving range is increased by use of multiple column systems
4. Detector responds to concentration of polymer molecules eluting from the GPC columns
What is a GPC System?
6
Additional Components Used in GPC
Concentration detectors
Differential refractometer (RI)
Ultraviolet absorbance (UV)
Evaporative light scattering or mass detector (ELS, EMD)
Infra-red (IR)
Molecular weight sensitive detectors
Viscometry
Light scattering
Additional systems
Online degasser
Autosampler
Column oven
Additional specific detectors
7
Polymer Molecules in Solution
GPC is based on the behaviour of polymer molecules in solution
In the solid state polymers can be considered like spaghetti – a confusing mass of intertwined chains
In solution, polymer molecules are discrete entities
Due to entrophic effects all but the most rigid of polymer chains curls up in solution to form a ball like shape
8
GPC Column Packings
GPC columns are packing with cross-linked, insoluble beads, typically co-polymers of styrene and divinyl benzene for organic GPC
These beads have a rigid pore structure that remains intact in the presence of solvent
9
Synthesis of Porous Beads
High cross-link content gives a rigid, low swelling product with a well-defined pore structure
10
Permeation of Polymer Molecules
Polymer coils in solution can permeate the pores on GPC packing materials
Exclusion, partial permeation and total permeation are possible
11
GPC Separation Mechanism
Polymer is prepared as a dilute solution in the eluent and injected into the system
The GPC column is packed with porous beads of controlled porosity and particle size
Large molecules are not able to permeate all of the pores and have a shorter residence time in the column
Small molecules permeate deep into the porous matrix and have a long residence time in the column
Polymer molecules are separated according to molecular size, eluting largest first, smallest last
12
Elution Profiles
As a result of the GPC separation mechanism, polymer molecules elute from the column in order of size in solution
Largest elute first, smallest elute last
The separation is purely a physical partitioning, there is no interaction or binding
The separation is isocratic
If polymer molecules have the same molecular dimensions, they will co-elute by GPC and may not be separated by this technique
The calibration curve describes how different size molecules elute from the column
13
GPC Column Technology
Columns are packed with porous particles, controlled pore size and particle size
Columns are produced by slurry packing technique, packed at pressures in excess of 2000psi
Column dimensions typically 7-8mm i.d., 250-600mm in length
Exclusion volume (Vo) - Upper MW limit (also known as void volume)
Total permeation volume (Vt) – Lower MW limit
Pore volume (Vp) – Working resolving range of MW
Vp = Vt - Vo
14
Typical Calibration Curves for PLgel Individual Pore Size Columns
15
Determination of Polymer Molecular Weight Distribution by GPC
Produce a GPC calibration curve for the column set relating log M to retention time (RT)
Chromatograph the polymer sample
Normalise and integrate the GPC response versus retention time plot for the polymer sample
Convert retention time to logM via the GPC calibration curve
Present a logM distribution plot and calculate molecular weight averages (Mn, Mw) for the distribution
16
Peak Integration in GPC
The sample peak when integrated can be assumed to be a histogram consisting of a number of individual “slices”
For each slice, i, the molecular weight (Mi) can be derived from the RT and the number of molecules (Ni) from the detector response
17
Calibration of GPC Columns Using Narrow Standards
Chromatograph a series of well characterised, narrow polydispersity polymer standards
Plot peak retention time (RT) versus peak log molecular weight (logM)
Fit the data using a mathematical function
The calibration curve will be characteristic of the GPC column set used
18
Performing Narrow Standard GPC Calibration
Injections of multiple narrow standards reduces the time taken to calibrate the system
Injection 1 Injection 2
Standards peaks in each chromatogram must be fully resolved to obtain repeatable retention times
19
EasiCal Pre-prepared Calibrants
EasiCal PS-1 separation on 3 x PLgel 10µm MIXED-B
Spatula A Spatula B
20
Narrow Standards Available from PL
21
Polymer Calibrants for GPC
Most commonly used polymer calibrants
Polystyrene - THF, toluene, chloroform, TCB
Polymethyl methacrylate - MEK, ethyl acetate, acetone, DMF
Polyethylene oxide/glycol - aqueous eluents, DMF, DMSO
Mn - number average molecular weight
Mw - weight average molecular weight
Mv - viscosity average molecular weight
Mp - peak molecular weight
Mw/Mn - polydispersity by GPC
Must be extremely well characterised
22
Individual standard EasiVial
Example Certificates of Analysis
23
Calibration Methods for Conventional GPC
Narrow standards
Broad standards
Hamielec
Broad on Narrow
Integral
Aim : to produce a mathematical model for log M versus retention time
24
Curve Fitting for Narrow Standards Calibration
Polynomial
All data points fitted with one function of the form
Log M = A + B(t) Linear (1st order)
Log M = A + B(t) + C(t2) Quadratic (2nd order)
Log M = A + B(t) + C(t2) + D(t3) Cubic (3rd order)
Cubic spline
Sets of points fitted with a series of cubic equations
Point to point
Points fitted with a series of linear equations
25
Calibration of GPC Columns Using Narrow Standards
Chromatograph a series of well characterised, narrow polydispersity polymer standards
Plot peak retention time (RT) versus peak log molecular weight (logM)
Fit the data using a mathematical function (e.g. polynomial order 1,2,3, etc)
The calibration curve will be characteristic of the GPC column set used
26
Errors Due to Limited Calibration Region
The column calibration should cover the full elution time region of the sample to avoid errors due to extrapolation
27
Broad Standard Calibration - Hamielec Method
Requires a single polydisperse sample where Mn and Mw are known
Assumes a linear calibration Log M = A + B(t)
Requires only a single injection to calibrate
1. Chromatograph broad standard
2. Input actual Mn and Mw
3. Computer searches for suitable values of A and B that when back calculated give the correct values for Mn and Mw
28
Broad Standard Calibration - Broad on Narrow Method
Requires a series of narrow standards plus a broad standard of known Mn and Mw
Valid for non-linear calibration curves
Requires several injections to calibrate
1. Chromatograph the narrow standards and fit the data with a mathematical function log Mns = F(t)
2. Input the Mn and the Mw values for the broad standard
3. Chromatograph the broad standard
29
Broad Standard Calibration - Broad on Narrow Method
Universal Calibration theory predicts that
log Mbs = log (Kns/Kbs) + (1+ans) log Mns
(1+abs) (1+abs)
As Kns, Kbs, ans and abs are constants this can be simplified to
log Mbs = A log Mns + B
The computer searches for suitable values of A and B that results in the correct values of Mn and Mw for the broad standards when back calculated.
30
Broad Standard Calibration - Integral Method
Requires a broad standard which has been fully characterised to give a cumulative molecular weight table, e.g.
Log M Wt %
2.800 0.000 Accuracy is improved
2.865 0.005 with increased number
2.929 0.020 of data points, usually
….. 50-60 points are used
5.705 99.80
5.789 99.99
5.870 100.00
There are few broad standards available with this level of characterisation
31
Broad Standard Calibration - Integral Method
The broad standard is chromatographed and integrated, the log M values are assigned to the distribution based on the wt % values quoted in the table. This results in a calibration table with as many points as there are entries in the table.
The Log M versus retention time data is then fitted with a mathematical function as usual.
32
Use of Mark-Houwink Constants in GPC
There are few well characterised polymer standards for GPC calibration. Therefore “unknown” polymer molecular weight can be derived from a GPC calibration curve obtained using polymer standards and applying the following theory:
Molecular size = hydrodynamic volume (HV)
HV = M [n], where M is molecular weight and [n] is intrinsic viscosity
In GPC molecules with the same HV elute at the same retention time
Therefore for different polymer types M1 [n]1=M2 [n]2
For a given polymer system Mark-Houwink equation applies : [n] = K M
Rearranging these relationships
logM2 = log(K1/K2) + (1+ 1) logM1
(1+ 2) (1+ 2)
where K1, 1 for polymer standards and K2, 2 for polymer under investigation are known
33
Interpreting Chromatograms
The data obtained in a GPC experiment will be in the form of a chromatogram showing detector response as a function of retention time
There are fundamental parameters that are present on all chromatograms
34
Peak Separation
Peak separation in GPC is dependent upon resolution and on molecular size
If two samples have different molecular sizes, then they will be separated to baseline assuming there is sufficient resolution
However, if samples are the same molecular size, then they cannot be separated by GPC as the mechanism of SEC is based upon size
35
If two peaks are not baseline resolved in GPC, they cannot be analysed as two separate peaks
Although the calculations will give you results, the numbers will be meaningless as not all of the peak is defined and so the distribution will not be correct
36
Excluded Peaks
37
Partial Exclusion
The dead space of the separation will be around half of the total elution volume
Peaks eluting close to this volume may be partially excluded
Look for sharp peaks at the front of your chromatograms
38
Oligomeric Resolution
In GPC, the relationship between molecular weight and retention time is logarithmic
As a result, peaks equidistant in molecular weight elute closer together with increasing molecular weight
This is a classic way to tell a separation is based on SEC
39
With some columns it is possible to calibrate the column using the oligomers
The molecular weights of the initiator fragment and the repeat unit of the polymer must be known
40
Interpreting Molecular Weight Distributions
The molecular weight distribution shows the amount of material present as a function of the molecular weight
The MWD looks a bit like a ‘mirror image’ of the chromatogram
41
Effect of Baseline Position
42
The whole peak should be analysed to get a true reflection of the sample
The peak should go down to the baseline on either side
Leaving out components of the peak will leave an ‘incomplete’ MWD