FOR HDR AND LDR BRACHYTHERAPY FOR HDR QUALITY ASSURANCE Accurate positioning of the HDR 192 IR- source is essential for proper treatment. The Luthmann Source Positioning QA Tool has been developed to perform required quality assurance checks for your HDR unit. Determine the distance to the most distal position. Verify the accuracy of HDR unit’s stepping motor in positioning the source for up to 18 channels at one time. Perform quick, daily constancy checks of source positioning by the HDR unit. Verify accuracy of the stepping motor in positioning the source at dwell positions in each channel. Eliminate the need for expensive and cumbersome video camera and monitor. Provide excellent hard copy documentation for QA files. BRACHYTHERAPY FOR LDR QUALITY ASSURANCE TESTS OF RIBBON ISOTOPES Upon receipt of LDR sources, the following quality assurance tests are useful to verify the integrity of your LDR sources. Confirm the relative uniformity of sources. Verify custom loading of ribbons. Visual verification of source spacing. Provide excellent hard copy documentation for QA files. Developed by Robert Luthmann, PhD Atlantic Beach, FL The Luthmann QA Tool in position for advancement of the HDR source.

Transcript of for HDr anD LDr Brachytherapy - JRT Associatesjrtassociates.com/pdfs/luthmann.pdf · for HDr anD...

Page 1: for HDr anD LDr Brachytherapy - JRT Associatesjrtassociates.com/pdfs/luthmann.pdf · for HDr anD LDr Brachytherapy ... advancement of the HDr source. This product is available through:

•for HDr anD LDr Brachytherapy

•ForHDrQualityassurance

Accurate positioning of the HDR 192IR- source is essential

for proper treatment. The Luthmann Source Positioning

QA Tool has been developed to perform required quality

assurance checks for your HDR unit.

• Determine the distance to the most distal position.

• Verify the accuracy of HDR unit’s stepping motor in

positioning the source for up to 18 channels at one time.

• Perform quick, daily constancy checks of source

positioning by the HDR unit.

• Verify accuracy of the stepping motor in positioning

the source at dwell positions in each channel.

• Eliminate the need for expensive and cumbersome

video camera and monitor.

• ProvideexcellenthardcopydocumentationforQAfiles.

Brachytherapy

•ForlDrQualityassurancetestsoFribbonisotopes

Upon receipt of LDR sources, the following quality assurance

tests are useful to verify the integrity of your LDR sources.

• Confirmtherelativeuniformityofsources.

• Verify custom loading of ribbons.

• Visualverificationofsourcespacing.

• ProvideexcellenthardcopydocumentationforQAfiles.

Developedbyrobertluthmann,phDatlantic Beach, FL theluthmannQatoolinpositionforadvancementoftheHDrsource.

Janet
Text Box
This product is available through: JRT Associates 800-221-0111
Page 2: for HDr anD LDr Brachytherapy - JRT Associatesjrtassociates.com/pdfs/luthmann.pdf · for HDr anD LDr Brachytherapy ... advancement of the HDr source. This product is available through:

For HDR QA

l u t H m a n n Q a t o o l ( r e F 9 0 0 1 1 ) s p e c i F i c a t i o n stoolDimensions

Height: 2.7cm(1.06in) Width: 22.2cm(8.75in)

Length: 32.4cm(12.75in) Weight: 1.8kg(4lbs)

leaDblockerDimensions

Height: 0.32cm(0.125in) Width: 10.16cm(4in) Length: 30.48cm(12in) Weight: 1.4kg(3lbs)

Specifications subject to change without notice.

© 2010 Standard Imaging, Inc. 1217-21, 4/10

3120 Deming Way Middleton WI 53562-1461 USa800-261-4446 . ph 608-831-0025 . fax 608-831-2202 www.standardimaging.com

Brachytherapy

Test OnePerformed Upon System Acceptance and After Source Changes

Purpose: The purpose of this test is to:

1. Determine the distance from the HDR unit to the most distal position of the source.

2. Verify accuracy of the stepping motor in positioning the source at various dwell positions in each channel.

3. Verify the integrity of each of the localization dummies used clinically for HDR source position localization.

Frequency: Test One is performed at the time of machine acceptance, after source change and at specific intervals as a QA check to verify the integrity of each localization dummy.

Test Two, Periodic Quality Assurance Tests

Purpose: The purpose of this test is to act as a constancy check for the three parameters in Test One.

Frequency: Test Two is performed on a daily basis before the HDR unit is used clinically for patient treatment.

The Luthmann Source Positioning QA Tool provides for an auto-radiograph of the sources and visual inspection of the spacing of ribbon seeds. The ribbons with multiple seeds are placed in the channels for the autoradiograph. Steel rulers placed along the outer two channels can be used for visual verification.

For LDR QA

Doc No. 1217-00 Two Pages, 09/03/2003

Test One, Performed Upon System Acceptance and After Source Changes Purpose: The purpose of this test is to:

1. Determine the distance from the HDR unit to the most distal position of the source. 2. Verify accuracy of the stepping motor in positioning the source at various dwell positions in

each channel. 3. Verify the integrity of each of the localization dummies used clinically for HDR source position

localization.

Frequency: Test One is performed at the time of machine acceptance, after source change and at specific intervals as a QA check to verify the integrity of each localization dummy.

Test Two, Periodic Quality Assurance Tests Purpose: The purpose of this test is to act as a constancy check for the three parameters in Test One.Frequency: Test Two is performed on a daily basis before the HDR unit is used clinically for patient treatment.

For LDR Quality Assurance Tests of Ribbon Isotopes Upon receipt of LDR sources, the following quality assurance tests are useful to verify the integrity of your LDR sources.

This device allows you to: • Confirm the relative uniformity of sources. • Verify custom loading of ribbons. • Visual verification of source spacing.• Provide excellent hard copy documentation for QA files.

The Luthmann Source Positioning QA Tool provides for an autoradiograph of the sources and visual inspection of the spacing of ribbon seeds. The ribbons with multiple seeds are placed in the channels for the autoradiograph. Steel rulers placed along the outer two channels can be used for visual verification.

Autoradiograph showing relative uniformity of sources.

Autoradiograph confirming custom loading of sources.

autoradiographshowingrelativeuniformityofsources.

Autoradiograph confirming custom loading of sources.

Doc No. 1217-00 Two Pages, 09/03/2003

Test One, Performed Upon System Acceptance and After Source Changes Purpose: The purpose of this test is to:

1. Determine the distance from the HDR unit to the most distal position of the source. 2. Verify accuracy of the stepping motor in positioning the source at various dwell positions in

each channel. 3. Verify the integrity of each of the localization dummies used clinically for HDR source position

localization.

Frequency: Test One is performed at the time of machine acceptance, after source change and at specific intervals as a QA check to verify the integrity of each localization dummy.

Test Two, Periodic Quality Assurance Tests Purpose: The purpose of this test is to act as a constancy check for the three parameters in Test One.Frequency: Test Two is performed on a daily basis before the HDR unit is used clinically for patient treatment.

For LDR Quality Assurance Tests of Ribbon Isotopes Upon receipt of LDR sources, the following quality assurance tests are useful to verify the integrity of your LDR sources.

This device allows you to: • Confirm the relative uniformity of sources. • Verify custom loading of ribbons. • Visual verification of source spacing.• Provide excellent hard copy documentation for QA files.

The Luthmann Source Positioning QA Tool provides for an autoradiograph of the sources and visual inspection of the spacing of ribbon seeds. The ribbons with multiple seeds are placed in the channels for the autoradiograph. Steel rulers placed along the outer two channels can be used for visual verification.

Autoradiograph showing relative uniformity of sources.

Autoradiograph confirming custom loading of sources.