O. Siegmund, J. McPhate UCB, SSL 1
First Preconditioning Test of New Batch of 33mm 20µm Pore MCPs
• ANL MCPs 164 and 163, 20µm, 60:1, 8° bias.• INCOM serial #s, 11863-99, 11863-95• Resistances 159MΩ, 138MΩ (NiCr over) • XDL detector with MCPs gapped by 0.001” spacer.• Completed 350°C bake, with RGA scans.• Scrub in progress, with RGA scans.• Have gone from 1µA to 2µA, now 3µA >0.65 C
cm-2 extracted.
O. Siegmund, J. McPhate UCB, SSL 2
MCP 164-163 Scrub Record
Minimal gain drop for 1µA to 0.06 C cm-2, some small H2 outgas (10-10 Torr).Full functional test showedno substantive performance changes except for slightly higher background.
Faster drop @ 2µA but stablized quickly with someH2 outgas (4 x 10-10 Torr).
At 3µA since then, with smalldrop then recovery to >100% gain – leveling off now @0.65 C Cm-2.Scrub @ 3 x 105 gain, 700v per MCP
164-163
Up to 2µA Up to 3µA1µA