Image Capture

Image Receptor

3

Page 4: Film/Screen Imaging Computed Radiography Digital Radiography IMAGE RECEPTOR SYSTEMS 1.

Image Receptor Types Film/Screen Systems

Computed Radiography Systems

Direct Digital Systems

4

Page 5: Film/Screen Imaging Computed Radiography Digital Radiography IMAGE RECEPTOR SYSTEMS 1.

Image Receptor Types

5

Page 6: Film/Screen Imaging Computed Radiography Digital Radiography IMAGE RECEPTOR SYSTEMS 1.

Film/Screen, CR, DR systems

6

Page 7: Film/Screen Imaging Computed Radiography Digital Radiography IMAGE RECEPTOR SYSTEMS 1.

Terminology•I

mage receptor•I

mage capture•B

eam absorption•A

ttenuation•P

rimary radiation•S

econdary radiation•S

catter radiation•R

emnant radiation•E

xit radiation•L

atitude•L

atent image•M

anifest image

7

Page 8: Film/Screen Imaging Computed Radiography Digital Radiography IMAGE RECEPTOR SYSTEMS 1.

Film/Screen Imaging

Film constructionCassette constructionFilm captureFilm developingFilm Interpretation/Storing

8

Page 9: Film/Screen Imaging Computed Radiography Digital Radiography IMAGE RECEPTOR SYSTEMS 1.

FIRST “FILM”

•Early Years:• Glass plates • Break easily• Difficulty in storage• Expensive

•Developments:• Cellulose acetate• Highly flammable• Easily torn

• Polyester Base• Highly flammable• Easily torn

9

Page 10: Film/Screen Imaging Computed Radiography Digital Radiography IMAGE RECEPTOR SYSTEMS 1.

Today’s Modern Film

10

Page 11: Film/Screen Imaging Computed Radiography Digital Radiography IMAGE RECEPTOR SYSTEMS 1.

Film Sizes

Standard:

•8” x 10”

•10” x 12”

•11” x 14”

•14” x 17”

Metric:

•18cm x 24cm

•24cm x 30cm

•30cm x 35cm

•35cm x 43cm

11

Page 12: Film/Screen Imaging Computed Radiography Digital Radiography IMAGE RECEPTOR SYSTEMS 1.

X-ray Film •T

wo basic parts:

1. Base

2. Emulsion and Topcoat

12

Page 13: Film/Screen Imaging Computed Radiography Digital Radiography IMAGE RECEPTOR SYSTEMS 1.

13

Page 14: Film/Screen Imaging Computed Radiography Digital Radiography IMAGE RECEPTOR SYSTEMS 1.

Film Construction - BASE

• Made of a polyester plastic

• Must be clear, strong, consistent thickness

• Tinted pale blue or blue-gray (reduces eye strain)

• Uniform lucency

• COATED ON 1 OR 2 SIDES WITH EMULSION

14

Page 15: Film/Screen Imaging Computed Radiography Digital Radiography IMAGE RECEPTOR SYSTEMS 1.

Film Construction - EMULSION

• Film emulsion can be on one side or both sides of base (single emulsion / double emulsion)

• Protective overcoat layered on top of emulsion-topcoat, supercoat

• Emulsion is a gelatin containing the film crystals---THE HEART OF THE FILM

15

Page 16: Film/Screen Imaging Computed Radiography Digital Radiography IMAGE RECEPTOR SYSTEMS 1.

Film Emulsion• Made of mixture of

gelatin & silver halide crystals

• Silver halide crystals made of silver bromide (90%) and silver iodide (10%)

• Photographically active layer – activated by light & radiation to create image

16

Page 17: Film/Screen Imaging Computed Radiography Digital Radiography IMAGE RECEPTOR SYSTEMS 1.

Latent Image•I

maged Formed by:• Direct x-ray interaction with

film

• Indirect interaction with light from cassette

•Image is not seen until developed chemically developed by processor.

17

Page 18: Film/Screen Imaging Computed Radiography Digital Radiography IMAGE RECEPTOR SYSTEMS 1.

Film Storage•C

lean, dry location

•40 – 60 % Humidity 70 º Fahrenheit

•Away from chemical fumes

•Safe from radiation exposure

•Standing on edge

•Expiration date clearly visible

18

Page 19: Film/Screen Imaging Computed Radiography Digital Radiography IMAGE RECEPTOR SYSTEMS 1.

CASSETTES FUNCTION

Protect film from exposure to light

Protect film from bending and scratching during use.

May contain intensifying screens, keeps film in close contact to screen during exposure.

19

Page 20: Film/Screen Imaging Computed Radiography Digital Radiography IMAGE RECEPTOR SYSTEMS 1.

FILM CASSETTE or FILM HOLDER

•The CASSETTE is used to hold the film during examinations.

•It consist of front and back intensifying screens, and has a lead (Pb) backing.

•The cassette is light tight

20

Page 21: Film/Screen Imaging Computed Radiography Digital Radiography IMAGE RECEPTOR SYSTEMS 1.

Cardboard Cassettes

Direct x-ray exposure to film:

•R

equires 25 to 400 times more radiation to create an image on the film

•Better detail than film screen (no blurring of image from light)

•All exposure made from x-ray photons

• Very large dose to the patient

21

Page 22: Film/Screen Imaging Computed Radiography Digital Radiography IMAGE RECEPTOR SYSTEMS 1.

CASSETTE CONSTRUCTION

•Exposure side of cassette is the “front”.

•Has the ID blocker (patient identification)

•Made of radiolucent material

•Intensifying screen mounted to inside of front

•Back made of metal or plastic

•Inside back is a layer of lead foil – prevents backscatter that could fog the film

•Inside foil layer is a layer of padding – maintains good film/screen contact

•Back intensifying screen mounted on padding

CASSETTE FRONT CASSETTE BACK

22

Page 23: Film/Screen Imaging Computed Radiography Digital Radiography IMAGE RECEPTOR SYSTEMS 1.

Cassette Construction

23

Page 24: Film/Screen Imaging Computed Radiography Digital Radiography IMAGE RECEPTOR SYSTEMS 1.

Screen Construction

•Polyester plastic base – support layer

•Phosphor layer – active layer

•Reflective layer – increases screen efficiency by redirecting light headed in other directions

•Protective coating

24

Page 25: Film/Screen Imaging Computed Radiography Digital Radiography IMAGE RECEPTOR SYSTEMS 1.

Intensifying Screens Phosphors

•RARE EARTH – (emits green light)• Developed in 1980’s• Most efficient • Most common in use today

•CALCIUM TUNGSTATE (blue light)• Not as efficient • Not used as often

25

Page 26: Film/Screen Imaging Computed Radiography Digital Radiography IMAGE RECEPTOR SYSTEMS 1.

INTENSIFYING SCREENS

DISADVANTAGES:• less detail than direct exposure

ADVANTAGES:

• Reduce patient exposure• Increase x-ray tube life

26

Page 27: Film/Screen Imaging Computed Radiography Digital Radiography IMAGE RECEPTOR SYSTEMS 1.

•The light photons are emitted by phosphor crystals.

•These crystals are significantly larger than the silver halide crystals in the film

•Screen reduces image sharpness

• Exams requiring extremely fine detail use screens with small crystals.

27

Page 28: Film/Screen Imaging Computed Radiography Digital Radiography IMAGE RECEPTOR SYSTEMS 1.

28

Page 29: Film/Screen Imaging Computed Radiography Digital Radiography IMAGE RECEPTOR SYSTEMS 1.

29

Page 30: Film/Screen Imaging Computed Radiography Digital Radiography IMAGE RECEPTOR SYSTEMS 1.

30

Page 31: Film/Screen Imaging Computed Radiography Digital Radiography IMAGE RECEPTOR SYSTEMS 1.

Image creationLatent Image

•1% of x-ray photons that leave patient

•Interact with phosphors of intensifying screens

•100’s of light photons created to make image on film

•Light photons expose silver halide crystals in the film emulsion –

•Turn black metallic silver after procession

31

Page 32: Film/Screen Imaging Computed Radiography Digital Radiography IMAGE RECEPTOR SYSTEMS 1.

Film Development

32

Page 33: Film/Screen Imaging Computed Radiography Digital Radiography IMAGE RECEPTOR SYSTEMS 1.

Process of X-ray Image

33

FILMX

-RA

Y

LATENT IMAGE

PRO

CE

SSO

R

MANIFEST IMAGE

Page 34: Film/Screen Imaging Computed Radiography Digital Radiography IMAGE RECEPTOR SYSTEMS 1.

34

Developing is the stage where the latent image is converted to the manifest (visible)Image.

Phenidone Developing AgentHydroquinone Developing AgentSodium Carbonate ActivatorPotassium Bromide Restrainer

Developing Agents help change ionic silver to metallic silver (black). Silver ion is said to be reduced. Only crystals that contain the latent image are reduce tometallic silver.

Page 35: Film/Screen Imaging Computed Radiography Digital Radiography IMAGE RECEPTOR SYSTEMS 1.

35

The stage of processing where the film gets treated so that the image will not fadebut will remain permanently is called fixing.

Acetic acid activatorAmmonium thiosulfate(hypo) clearing agentPotassium alum hardenerSodium sulfite preservativeWater solvent

Fixing the film CLEARS the film of all undeveloped silver halides.It is important to stop the developing process because with time, all the ionic silverwill be reduced to metallic silver, thus ruining our image. Our latent image mustbe processed in a timely fashion or we will have image degradation.

Page 36: Film/Screen Imaging Computed Radiography Digital Radiography IMAGE RECEPTOR SYSTEMS 1.

36

The function of the wash cycle is to remove ANY residual chemicalsfrom the film. Water is used as the washing agent. Failure to completelywash the residual chemicals could result in film fog, hypo retention, yellowing of films, fading of films,

The final step of the processing is to dry the film before it leaves theprocesser. Failure to do so occasionally results in sticky films which eventually cause artifacts.

Page 37: Film/Screen Imaging Computed Radiography Digital Radiography IMAGE RECEPTOR SYSTEMS 1.

37

ReplenishmentReplenishment : As chemistry is used up by the processing of image, the proper chemical balance is needed to be maintained.As a film enters the first stages of processing, the replenishmentsystem turns on and the chemistry in both developer and the fixerare added to as needed according to prescribed rates.

Silver Recovery: Silver Recovery: As the latent image is developed into the manifestimage, the unaffected silver halide crystals will be washed off. Eachprocessor will have a silver recovery system in place. This is economically soundand environmentally helpful. It also helps prevent the processor from breaking down due to clogging of the system by metallic silver.

$

Page 38: Film/Screen Imaging Computed Radiography Digital Radiography IMAGE RECEPTOR SYSTEMS 1.

Film Storage

38

Page 39: Film/Screen Imaging Computed Radiography Digital Radiography IMAGE RECEPTOR SYSTEMS 1.

You will need to bring to class an object or objects to image during week 12 and week 13. Each student will produce one image.

YOUR XRAY IMAGE

39

Page 40: Film/Screen Imaging Computed Radiography Digital Radiography IMAGE RECEPTOR SYSTEMS 1.

Any Questions ??????