What is AQR Color-Tec® ?What is AQR Color-Tec® ?

Field test kit for chlorinated solvents Detects total chlorinated halocarbons (CVOHs) in

soil and water samples to less than 3 µg/L or µg/kg Provides qualitative data and estimated

concentrations Results in 3 minutes per sample Over 20,000 samples analyzed Initially developed in 1997 Tested and verified at EPA, DOD, DOE and state

chlorinated solvent site investigations Commercial production began 2007

Field test kit for chlorinated solvents Detects total chlorinated halocarbons (CVOHs) in

soil and water samples to less than 3 µg/L or µg/kg Provides qualitative data and estimated

concentrations Results in 3 minutes per sample Over 20,000 samples analyzed Initially developed in 1997 Tested and verified at EPA, DOD, DOE and state

chlorinated solvent site investigations Commercial production began 2007

Page 3: Field-Based Analysis of Chlorinated Volatile Organic Halocarbons Field-Based Analysis of Chlorinated Volatile Organic Halocarbons .

Statistical Comparison of Split-Sample CG/MS Results to AQR Color-Tec® Zero Values Only (non-detections)

(typical chlorinated solvent site data set N = 42)

Statistical Comparison of Split-Sample CG/MS Results to AQR Color-Tec® Zero Values Only (non-detections)

(typical chlorinated solvent site data set N = 42)

Parameter Concentration (µg/L)

Mean 0.362 Range 0 to 2.55Standard Deviation 0.73695% UCL of the mean 0.85790th Percentile Value 1.62695th Percentile Value 2.370

Parameter Concentration (µg/L)

Mean 0.362 Range 0 to 2.55Standard Deviation 0.73695% UCL of the mean 0.85790th Percentile Value 1.62695th Percentile Value 2.370

AQR Color-Tec® Method Detection Limits

AQR Color-Tec® Method Detection Limits

AQR Color-Tec® detects all CVOHs to less than 3 µg/L

AQR Color-Tec® detects all CVOHs to less than 3 µg/L

Page 4: Field-Based Analysis of Chlorinated Volatile Organic Halocarbons Field-Based Analysis of Chlorinated Volatile Organic Halocarbons .

AQR Color-Tec® results consistently indicate a relative correlation to split sample laboratory GC/MS (EPA 8260B) comparison data.

This graph shows the relative correlation by plotting AQR Color-Tec readings with the split sample GC/MS concentrations of total CVOHs detected in 61 groundwater samples from a chlorinated solvent site.

AQR Color-Tec® results consistently indicate a relative correlation to split sample laboratory GC/MS (EPA 8260B) comparison data.

This graph shows the relative correlation by plotting AQR Color-Tec readings with the split sample GC/MS concentrations of total CVOHs detected in 61 groundwater samples from a chlorinated solvent site.

AQR Color-Tec® Comparison to GC/MS data

AQR Color-Tec® Comparison to GC/MS data

0.01

0.1

1

10

100

1000

10000

100000

GC

/MS

Co

ncen

trati

on

(ug

/L T

ota

l C

VO

Hs)

0.01

0.1

1

10

100

1000

AQ

R C

olo

r-Tec V

alu

e

(Un

its)

GC/MS Concentrations AQR Color-Tec Readings

Page 5: Field-Based Analysis of Chlorinated Volatile Organic Halocarbons Field-Based Analysis of Chlorinated Volatile Organic Halocarbons .

Pre-filtered ambient air is purged through the

sample, stripping the volatile compounds into

the headspace

Pre-filtered ambient air is purged through the

sample, stripping the volatile compounds into

the headspace

Hand-Operated Vacuum Pump

(100cc)

Hand-Operated Vacuum Pump

(100cc)

Pump HandlePump

HandleColorimetric Tube

Colorimetric Tube

Tygon TubingTygon Tubing

Pump StandPump Stand

Carbon Filter

Carbon Filter

Purge Needle (long)

Purge Needle (long)

Headspace (~30%)Headspace (~30%)

Water SampleWater

Sample

40 ml VOA Vial40 ml VOA Vial

Ambient air enters through carbon

filter

Ambient air enters through carbon

filter

Air FlowAir

Flow

Air Bubble

s

Air Bubble

s

AQR Color-Tec® Open-Loop Stripping Analysis Procedure

AQR Color-Tec® Open-Loop Stripping Analysis Procedure

Page 6: Field-Based Analysis of Chlorinated Volatile Organic Halocarbons Field-Based Analysis of Chlorinated Volatile Organic Halocarbons .

Purged volatile compounds pass through the headspace, needle,

and tubing into the colorimetric tube where they react and cause a distinct color change

Purged volatile compounds pass through the headspace, needle,

and tubing into the colorimetric tube where they react and cause a distinct color change

Hand-Operated Vacuum Pump

(100cc)

Hand-Operated Vacuum Pump

(100cc)

Pump HandlePump

HandleColorimetric Tube

Colorimetric Tube

Tygon TubingTygon Tubing

Pump StandPump Stand

Carbon Filter

Carbon Filter

Purge Needle (long)

Purge Needle (long)

Headspace (~30%)Headspace (~30%)

Water SampleWater

Sample

Ambient air enters through carbon

filter

Ambient air enters through carbon

filter

Air Bubble

s

Air Bubble

s

Air FlowAir

Flow

40 ml VOA Vial40 ml VOA Vial

AQR Color-Tec® Open-Loop Stripping Analysis Procedure

AQR Color-Tec® Open-Loop Stripping Analysis Procedure

Page 7: Field-Based Analysis of Chlorinated Volatile Organic Halocarbons Field-Based Analysis of Chlorinated Volatile Organic Halocarbons .

Analyze liquid or solid samplesAnalyze liquid or solid samples

Headspace

Headspace

Water SampleWater

Sample

Added Water

(de-ionized)

Added Water

(de-ionized)

HeadspaceHeadspace

Soil Sample

Soil Sample

Water Set-up

Water Set-up

Soil Set-upSoil Set-up

AQR Color-Tec® Sampling Media AQR Color-Tec® Sampling Media

Samples and tubes are heated before AnalysisSamples and tubes are heated before Analysis

Page 8: Field-Based Analysis of Chlorinated Volatile Organic Halocarbons Field-Based Analysis of Chlorinated Volatile Organic Halocarbons .

AQR Color-Tec® Conversion TableAQR Color-Tec® Conversion Table

Converts AQR Color-Tec® tube readings to µg/L or µg/kg

Based on comparison of Color-Tec data to GC/MS data using over 5000 water samples collected from 152 chlorinated solvent sites.

A “Median” estimated value for each Color-Tec reading was obtained using statistical comparison of Color-Tec data and GC/MS (EPA Method 8260B) data.

The “Range” value reflects the potential deviation from the Median value based on the Color-Tec and EPA Method 8260B comparison results.

Converts AQR Color-Tec® tube readings to µg/L or µg/kg

Based on comparison of Color-Tec data to GC/MS data using over 5000 water samples collected from 152 chlorinated solvent sites.

A “Median” estimated value for each Color-Tec reading was obtained using statistical comparison of Color-Tec data and GC/MS (EPA Method 8260B) data.

The “Range” value reflects the potential deviation from the Median value based on the Color-Tec and EPA Method 8260B comparison results.

Page 9: Field-Based Analysis of Chlorinated Volatile Organic Halocarbons Field-Based Analysis of Chlorinated Volatile Organic Halocarbons .

48 Color-Tec Samples @ $25 each

48 Color-Tec Samples @ $25 each

6 Laboratory Samples @ $100 each

6 Laboratory Samples @ $100 each

Definitive analysis provides high analytical accuracy, but sampling quantity is often limited to control costs, resulting in data gaps, sampling uncertainty, and low overall data quality

Definitive analysis provides high analytical accuracy, but sampling quantity is often limited to control costs, resulting in data gaps, sampling uncertainty, and low overall data quality

Laboratory Sampling Only

Laboratory Sampling Only

AQR Color-Tec® with Laboratory sampling

AQR Color-Tec® with Laboratory sampling

Total Cost = $1,800

Total Cost = $1,800

24 Samples @ $100 each24 Samples @ $100 each

The low per-sample cost of AQR Color-Tec® allows for more samples, resulting in increased coverage and reduced uncertainty.

The AQR Color-Tec® data is used to focus definitive analysis sampling onto critical areas.

Combining AQR Color-Tec® with laboratory sampling increases overall data quality at a lower cost by providing improved sampling coverage with high analytical accuracy.

The low per-sample cost of AQR Color-Tec® allows for more samples, resulting in increased coverage and reduced uncertainty.

The AQR Color-Tec® data is used to focus definitive analysis sampling onto critical areas.

Combining AQR Color-Tec® with laboratory sampling increases overall data quality at a lower cost by providing improved sampling coverage with high analytical accuracy.

Total Cost = $2,400

Total Cost = $2,400

AQR Color-Tec® Increases Sampling Coverage

AQR Color-Tec® Increases Sampling Coverage

Page 10: Field-Based Analysis of Chlorinated Volatile Organic Halocarbons Field-Based Analysis of Chlorinated Volatile Organic Halocarbons .

AQR Color-Tec® Comprehensive Instructions AQR Color-Tec® Comprehensive Instructions

Detailed Procedures Manual Kit Set-up Method principles Sample Collection Reading the tubes QA/QC procedures Trouble shooting guide Conversion Table References Method Applications

Detailed Procedures Manual Kit Set-up Method principles Sample Collection Reading the tubes QA/QC procedures Trouble shooting guide Conversion Table References Method Applications

Page 11: Field-Based Analysis of Chlorinated Volatile Organic Halocarbons Field-Based Analysis of Chlorinated Volatile Organic Halocarbons .

EPA Guidance FeaturingAQR Color-Tec® EPA Guidance FeaturingAQR Color-Tec®

Site Characterization Technologies for DNAPL Investigations September 2004; EPA 542-R-04-017;

http://www.clu-in.org/download/char/542r04017.pdf

Understanding Procurement for Sampling and Analytical Services under a Triad Approach June 2005, EPA

542-R-05-022 http://www.epa.gov/swertio1/download/char/procurement.pdf

Conducting Contamination Assessments at Drycleaning Sites EPA Technology Innovation Program; State Coalition for Remediation of Drycleaners http://www.drycleancoalition.org/download/assessment.pdf

Using Dynamic Field Activities for On-Site Decision Making May 2003; OSWER No. 9200.1-40 EPA/540/R03/002; Chapter 5, http://www.epa.gov/superfund/programs/dfa/download/guidance/40r03002.pdf

Page 12: Field-Based Analysis of Chlorinated Volatile Organic Halocarbons Field-Based Analysis of Chlorinated Volatile Organic Halocarbons .

State Guidance FeaturingAQR Color-Tec® State Guidance FeaturingAQR Color-Tec®

D H E C

South Carolina Department of Health and Environmental Control; Field Inspection Guidance http://www.schouse.org/reports/dhec/scdcrtfprogramstatusreport2007.doc

North Carolina Dry-Cleaning Solvent Cleanup Program Guidance for Conducting Drycleaner Facility Prioritization Assessments http://wastenot.enr.state.nc.us/SFHOME/DryClean/Download/PAR

%20guidance.doc

Colorado Department of Public Health and Environment Drycleaner Remediation Guidance Document 2004 http://www.cdphe.state.co.us/HM/drycleanerguidance.pdf

Page 13: Field-Based Analysis of Chlorinated Volatile Organic Halocarbons Field-Based Analysis of Chlorinated Volatile Organic Halocarbons .

EPA Triad Implementations Case Studies Using AQR Color-Tec®

EPA Triad Implementations Case Studies Using AQR Color-Tec®

Best Practices in Triad Approach to Characterize TCE National Laboratory Environmental Sciences Division Argonne, IL

http://www.triadcentral.org/user/doc/TPP-Hurlburt-BestPractices.pdf

Adaptations to Triad as a Basis for Exit Strategy Development Decision Logic Flow Chart 2006 Triad Poster Session; CH2MHill http://www.triadcentral.org/user/doc/TPP-Hurlburt-TriadAdaptations.pdf Successful Triad Implementations at Federal Sites AQR Color-Tec Method locates source areas at Calloway Drum Recycling Site, Auburndale, Florida http://www.triadcentral.org/user/doc/TPP-Callaway-Field_Based_Decision_Approach.pdf

Fast Track to Reducing Conceptual Site Model Uncertainty CH2MHill; Storage Tank Site ST-123 POL Fuel Yard http://www.Triadcentral.org/user/includes/dsp_profile.cfm?Project_ID=25

Using AQR Color-Tec for Source Identification and Delineation Naval Construction Battalion Center Davisville North Kingstown, RI - 2008 Triad Conference http://www.umass.edu/tei/conferences/Triad_PDF/Anderson.pdf

Page 14: Field-Based Analysis of Chlorinated Volatile Organic Halocarbons Field-Based Analysis of Chlorinated Volatile Organic Halocarbons .

AQR Color-Tec® Featured OnEPA CLU-IN Technology PageAQR Color-Tec® Featured OnEPA CLU-IN Technology Page

EPA CLU-IN web page provides information about treatment and characterization technologies.

AQR Color-Tec is featured with other test kits in the technology descriptions and selection tools section

EPA CLU-IN web page provides information about treatment and characterization technologies.

AQR Color-Tec is featured with other test kits in the technology descriptions and selection tools section

Page 15: Field-Based Analysis of Chlorinated Volatile Organic Halocarbons Field-Based Analysis of Chlorinated Volatile Organic Halocarbons .

Hardware Kit = $1,195

Hardware Kit = $1,195

20 Sample Pack = $395

20 Sample Pack = $395

Cost per sample is $19.95 (Basic QA/QC included) Provided in 20-sample kits with all required expendables Hardware kit includes piston pump, pump stand, heating

equipment, and carrying case Professional technical support is included with any purchase Customized comprehensive training courses are available

Cost per sample is $19.95 (Basic QA/QC included) Provided in 20-sample kits with all required expendables Hardware kit includes piston pump, pump stand, heating

equipment, and carrying case Professional technical support is included with any purchase Customized comprehensive training courses are available

AQR Color-Tec® Equipment and Expendables