FEI Quanta 250 ESEM Basic instructions - OSCAR...FEI Quanta 250 ESEM Basic Operation Manual For...

8
1 DSU Imaging Facility FEI Quanta 250 ESEM Basic Operation Manual For Assistance with SEM Please Contact: Dr. Mukti Rana Associate Professor & Chair Department of Physics and Engineering SC Room 216 302-857-6588, [email protected]

Transcript of FEI Quanta 250 ESEM Basic instructions - OSCAR...FEI Quanta 250 ESEM Basic Operation Manual For...

Page 1: FEI Quanta 250 ESEM Basic instructions - OSCAR...FEI Quanta 250 ESEM Basic Operation Manual For Assistance with SEM Please Contact: Dr. Mukti Rana Associate Professor & Chair Department

1

DSUImagingFacilityFEIQuanta250ESEMBasicOperationManual

ForAssistancewithSEMPleaseContact:Dr.MuktiRanaAssociateProfessor&ChairDepartmentofPhysicsandEngineeringSCRoom216302-857-6588,[email protected]

Page 2: FEI Quanta 250 ESEM Basic instructions - OSCAR...FEI Quanta 250 ESEM Basic Operation Manual For Assistance with SEM Please Contact: Dr. Mukti Rana Associate Professor & Chair Department

2

The chamber of the SEM is kept under vacuum.

Under Normal condition the system is logged on.

If you do not see that the system is logged on, double click on the xT Microscope icon on the desktop

Then you will see the status bar of xT microscope.

Click on start UI.

Page 3: FEI Quanta 250 ESEM Basic instructions - OSCAR...FEI Quanta 250 ESEM Basic Operation Manual For Assistance with SEM Please Contact: Dr. Mukti Rana Associate Professor & Chair Department

3

Log-intotheMicroscopeusingthefollowing- Username:supervisorPassword:Quanta250-9923724

Makesurethatyouareinthebeamcontroltabontherightsideofthescreen.VentthechamberbyclickingontheVENTbuttontorightsideofthewindow

Insert you sample into the chamber by using one of the sample holders and double sidedtapes

ifnecessary.Usingthe“Elephant Trunk”makesurethatyoursample is lessthan10mm. Ifyouneedtochangethebaseofthesampleholderyoucandothatbyunscrewingtheboltandconelikestructure.

Page 4: FEI Quanta 250 ESEM Basic instructions - OSCAR...FEI Quanta 250 ESEM Basic Operation Manual For Assistance with SEM Please Contact: Dr. Mukti Rana Associate Professor & Chair Department

4

Determinethetypeofsampleyouhaveandselecttheappropriatevacuumconditionsyouwilluse.Usehighvacuummodefornormalconductivesampleandlowvacuummodeforsoftoranduncoatednon- conductivesamples.YoumayalsousetheESEMmodewithPeltierStageandGSEDdetectorforsampleswhichcontainliquid.Vacuumbuttons

Next,selecttherightaccelerationvoltagefromthepulldowntablocatedjustbelowthemenubarorfromtherightsideofthebeamcontroltab.

Oncetheproperpressurehasbeenachieved,thechamberICONinthelowerrightsideofthescreenwillbefullygreen.

Page 5: FEI Quanta 250 ESEM Basic instructions - OSCAR...FEI Quanta 250 ESEM Basic Operation Manual For Assistance with SEM Please Contact: Dr. Mukti Rana Associate Professor & Chair Department

5

TurnontheBeambyclickingontheBEAMONbutton

Thedisplayshouldlooklikethis. Scanspeedcontrol Pausereleasebutton

TopleftSEimage,ToprightBackscatter,BottomLeftisthecombinedSEandBackscatterimageandtheBottomrightisthechamberviewscope.Activateuptothreedisplaysatonetimebyselectingthedisplayandtheclickonthepausereleasebutton.Thedisplaywillstarttoscan.

Toincreaseanddecreasethescanspeedsusethearrowupanddown.

Page 6: FEI Quanta 250 ESEM Basic instructions - OSCAR...FEI Quanta 250 ESEM Basic Operation Manual For Assistance with SEM Please Contact: Dr. Mukti Rana Associate Professor & Chair Department

6

Whileviewingthechamberscope(bottomrightscreen)pressandholdthescrollwheelbuttonandpushthemouseupwardtopositionyoursamplenearthe10mmmarkonthescreen.Itmaybenecessarytounpausethechamberscopescreenfirst.

IncreasetheMagnificationandfocustheSEimagebyusingtheUSERinterfacepanelshownbelow.F7displaysareducedscreenforeaserrealtimeadjustments.F5enlargestheactivescreentofittheentirewindow.

OncetheimageisfocusedlinktheZtoFWDbyclickingontheICONonthetoolbar

Youmayalsogotonavigationtabtoseethezdistance.

FormostoperationsitisrecommendedthataZvalueof10mmisnowsetastheworkingdistance.

Page 7: FEI Quanta 250 ESEM Basic instructions - OSCAR...FEI Quanta 250 ESEM Basic Operation Manual For Assistance with SEM Please Contact: Dr. Mukti Rana Associate Professor & Chair Department

7

Increasemagnificationtoatleast1k(higherthebetter)andfocustheimageaswellasyoucan.Nextfromthemainmenu,selecttheBeam>DirectadjustmentsandthenstarttheLensalignbyclickingontheLENSALIGN button.

Whileobservingtheimage,placeyoumouseoverthesmallsquareboxintheLensAlignandholditdown.Movethemouseandboxwithinthefieldtoreducemovementintheimage.

TurnoftheLensAlignbycrossingitout.RefocusyouimageandincreasetheMagnificationtoabove20k(thehigherthebetter)andusingtheUserInterfacePanel,StigmatetheimageusingtheXandYcontrolsforsharpestimagepossible.F7displaysareducedscreenforeaserrealtimeadjustments.F5enlargestheactivescreentofittheentirewindow.

WhenusingtheLowVacmode,itmaybenecessarytoadjustthepressureinthechamberforahighcontrastimage.Simplyadjustthepressureuntiltheappropriatecontrasthasbeenachieved.Inmostcasespressurebetween0.83and0.53Torrwillworkwell.

Page 8: FEI Quanta 250 ESEM Basic instructions - OSCAR...FEI Quanta 250 ESEM Basic Operation Manual For Assistance with SEM Please Contact: Dr. Mukti Rana Associate Professor & Chair Department

8

Oncethedesiredimageisachieved,pressF2orgotoscan>Photoorscan>snapshotfortakingandsavingimage.OncethescanhasstoppedadefaultdirectorywillappearintheSharedDataFolder.Storeyour imagesintheSharedFolderonly.Datawillbestoreonthecomputerontheleftsideofthe operator’sconsol.

Turnoffthebeambyclickingthebeamonbutton(itgoesfromyellowbackgroundtopuregrayonceitisturnedoff)

NeverleavetheESEMintheLOWVACorESEMmode,alwaysreturntheSEMtotheHIGHVACMode.