FD910 Density Meter
-
Upload
saidrahmansyah4750 -
Category
Documents
-
view
249 -
download
0
Transcript of FD910 Density Meter
-
7/27/2019 FD910 Density Meter
1/2ISSUE: FD910/MAY2001/UK
continuous on-line density measurement
accuracy to 0.0001g/cm3 reading
straight through flow
smart headmount electronics option
local display option
insensitive to plant / pipeline vibration
316L stainless steel wetted parts
intrinsically safe and flameproof design
wide operating temperature span
options to suit extensive applications
compact and lightweight design
no on-site calibration required
secondary containment
Sarasota
Industrial Density Meter
FD910
Features
The Sarasota FD910 density meter is designed for highaccuracy measurement of general process liquids andslurries.
Dimensional drawing
Applications process control quality control productinterface detection process monitoring
206.5mm
530mm
198mm
128mm
271mm
89mm
Detailed specification and ordering information: see overleafThermo Measurement reserves the right to alter specifications without notice.
-
7/27/2019 FD910 Density Meter
2/2
Specification
Thermo Electron CorporationKings Worthy
Winchester, HampshireSO23 7QA, United Kingdom
Tel: +44 (0) 1962 625000Fax: +44 (0) 1962 885530email: [email protected]
Thermo Electron Corporation9303 W. Sam Houston Parkway South
HoustonTexas77099 5298 USA
Tel: +1 (713) 272 0404Fax: +1 (713) 272 2273
Ordering Information
Transducer:
Operating range: 0 to 2.1g/cm3
Factory calibration range: 0.65 to 1.6g/cm3
Accuracy: 0.0001g/cm3
Repeatability: 0.00002g/cm3
Temperature effect:
(corrected) see Note 1 0.000005g/cm3/C
Process temperature: -50 to +180C
Pressure effect:
(corrected) see Note 1 0.000003g/cm3/bar
Operat ing pressure: Flange pressure/temperature rating
Materials of construction: Sensor/wetted material 316L stainless steel
Case 316L stainless steel
Electronics housing copper free
aluminium grey epoxy finish
Secondary containment: Full pipeline specification at 40 bar
NACE MR0175 conformity: NACE specification option
Temperature measurement: High accuracy 13 DIN
Integral 4 wire PT100
General:
Environmental protection: IP65
Hazardous area certification: FD910F EEx ia IIC T6
FD910H EEx ia IIC T4
EEx d IIC T4/T3
Weight: Nett 11kg max
Shipping 20kg max
Installation: Vertical as standard
Optionally horizontal
Flow range: 0-300 l/min
5-300 l/min when installed horizontally
Outputs:
F option: Frequency related to density on 2 wire current
modulated loop 6mA to 18mA
4 wire PT100
H option: Analogue 4-20mA related to density or
density derived variable
HART protocol
System: Accuracy better than 0.2% reading plus
0.05% of span at reference conditions
HART
Local display option
Electrical:
Power supply:
F option: 13-28 VDC 10mA average (peak 18mA)
H option: 2 x 13-28 VDC 25mA
Connections: Screw terminals
Cable entry: 2 x 34 NPT
Operating temperature(ambient): -20 to +70C
Example: FD910-H-2-G-B0-I-L
FD910 industrial density meter with smartheadmount electronics, 0.00025g/cm3,standard temperature range of -20 to +120C,ANSI Class 150 RF flanges, intrinsically safeapproval with local display.
Density meter model code format
FD910 Sarasota Industrial Density Meter
Signal output
F Frequency output
H Smart headmount electronics
Transducer accuracy
2 0.00025g/cm3
1 0.0001g/cm3
Temperature range
G -20 to +120C
H -50 to +180C
Process connections
B0 1 ANSI B16.5 Class 150 RF flanges
B1 1 ANSI B16.5 Class 150 Duplex flanges
F0 1 ANSI B16.5 Class 300 RF flanges
F1 1 ANSI B16.5 Class 300 Duplex flanges
A0 1 ANSI B16.5 Class 600 RF flanges
A1 1 ANSI B16.5 Class 600 Duplex flanges
D BS4504 PN40 DN25 RF flanges (Form D)
X Other flange type (see note)
Certification
S Non hazardous / safe area application
I Intrinsically safe EEx ia IIC T4/T6
D Flameproof EEx d IIC T4/T3
Options
L Local display (with FD910H only)
M Wetted parts material certificates (DIN 50049 BS EN 10204 3.1b)
N NACE conformance
T Traceable calibration certificate
Note: Where an X option is chosenthe option must be described fully,in writing. Please contact salesoffice for pricing.
Note 1 - Corrected temperature and pressure effects after
applying correction coefficients
HART is a registered trademark of the Hart Foundation