FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC...

72
FAM 10 FluidAqua Mobil Operating and Maintenance Instructions English (translation of original operating instructions) Documentation no.: 3549460a

Transcript of FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC...

Page 1: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FAM 10

FluidAqua Mobil

Operating and Maintenance Instructions

English (translation of original operating instructions)

Documentation no.: 3549460a

Page 2: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Trademarks

HYDAC FILTER SYSTEMS GMBH en(us) Page 2 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Trademarks

The trademarks of other companies are exclusively used for the products of those companies.

Copyright © 2011 by HYDAC FILTER SYSTMES GMBH all rights reserved

All rights reserved. This manual may not be reproduced in part or whole without the express written consent of HYDAC Filter Systems GmbH. Contraventions are liable to compensation.

Exclusion of Liability

We have made every endeavor to ensure the accuracy of the content of this document. However, errors cannot be ruled out. Consequently, we accept no liability for such errors as may exist nor for any damage or loss whatsoever which may arise as a result of such errors. The content of the manual is checked regularly. Any corrections required will be incorporated in future editions. We welcome any suggestions for improvements.

All details are subject to technical modifications.

Technical specifications are subject to change without notice.

HYDAC Filter Systems GmbH Technische Dokumentation Postfach 12 51 66273 Sulzbach / Saar Germany

Documentation Representative

Günter Harge

c/o HYDAC International GmbH, Industriegebiet, 66280 Sulzbach / Saar

Telephone: ++49 (0)6897 509 1511

Fax: ++49 (0)6897 509 1394

E-mail: [email protected]

Page 3: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Content

HYDAC FILTER SYSTEMS GMBH en(us) Page 3 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Content

Trademarks ...............................................................................................................2

Documentation Representative...............................................................................2

Content......................................................................................................................3

Preface ......................................................................................................................6

Technical Support...................................................................................................7

Modifications to the Product ...................................................................................7

Warranty .................................................................................................................7

Using the Documentation .......................................................................................8

Safety Information and Instructions.......................................................................9

Obligations and Liability..........................................................................................9

Explanation of Symbols and Warnings, etc. .........................................................10

General Safety Precautions..................................................................................10

Proper/Designated Use ........................................................................................11

Improper Use ........................................................................................................11

Safety Devices......................................................................................................11

Informal Safety Precautions..................................................................................12

Training and Instruction of Personnel ...................................................................13

Safety Measures to Be Followed during Normal Operation..................................13

Electrical Hazards.................................................................................................14

Maintenance, Servicing and Troubleshooting.......................................................14

Modifications to the FluidAqua Mobil ....................................................................14

What to Do in Case of Emergency .......................................................................15

Unpacking the FAM................................................................................................15

Transporting the FAM ............................................................................................16

Forklift ...................................................................................................................16

Crane ....................................................................................................................16

Train/truck.............................................................................................................17

Scope of Delivery ...................................................................................................18

FAM Description.....................................................................................................19

FAM Versions .......................................................................................................19

Performance .........................................................................................................20

FAM system components .....................................................................................21

Suitable Fluids ......................................................................................................23

Dimensions .............................................................................................................24

Stationary version .................................................................................................24

Mobile version.......................................................................................................25

Page 4: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Content

HYDAC FILTER SYSTEMS GMBH en(us) Page 4 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Hydraulic diagram ..................................................................................................26

Function Description .............................................................................................27

Possible Applications............................................................................................28

Bypass purification (dewatering, filtration and degassing). ...............................28

Transfer by pumping (dewatering, filtration and degassing) .............................28

FAM Set-up and Connection .................................................................................29

Setting up the FAM ...............................................................................................29

Notes on Piping / Hosing ......................................................................................29

Connecting the inlet (IN) ...................................................................................31

Connect outlet (OUT) ........................................................................................31

Preparing the Vacuum Pump................................................................................32

Filling up the rotary vane vacuum pump........................................................33

Electrical Connection of the FAM .........................................................................34

Checking the settings on the heater (optional) .....................................................35

Starting up the FAM ...............................................................................................36

Switching on the FAM...........................................................................................36

Venting the filter housing ......................................................................................37

Shut off FAM .....................................................................................................37

Operating the FAM .................................................................................................38

Main menu ............................................................................................................38

Manual Menu 1 .................................................................................................39

Manual Menu 2 .................................................................................................40

Troubleshooting .....................................................................................................41

Electrical fuses......................................................................................................42

Vacuum chamber overfilled ..................................................................................43

Level switch 01/02(02/03).....................................................................................44

V-pump fill level min/max......................................................................................45

Replace fluid filter .................................................................................................46

Float switch...........................................................................................................47

E/F/V pump motor protection................................................................................48

Incorrect phase sequence ....................................................................................49

Dry running E-pump..............................................................................................50

Outlet (OUT) clogged............................................................................................51

Inlet (IN) clogged ..................................................................................................52

FCU plug missing (only in conjunction with accessory FCU)................................53

Reset safety thermostat on the heater..................................................................54

Performing Maintenance .......................................................................................55

Changing the Air Filter ..........................................................................................56

Vacuum pump.......................................................................................................56

Changing oil and oil filter...................................................................................56

Page 5: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Content

HYDAC FILTER SYSTEMS GMBH en(us) Page 5 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Replacing the Air De-oiling Element .................................................................57

Clean the suction strainer. ....................................................................................58

Replacing the filter element on the fluid filter ........................................................59

Checking the settings on the heater (optional) .....................................................63

Spare Parts .............................................................................................................64

FAM 10 .................................................................................................................64

Fluid filter (OLF 5/10)............................................................................................64

Vacuum pump.......................................................................................................64

Contact / Service ....................................................................................................65

Storing the FAM/Taking it out of Operation .........................................................66

Disposing of the FAM ............................................................................................66

Type label................................................................................................................66

Technical data ........................................................................................................67

Model Code .............................................................................................................68

EU declaration of conformity (only for mobile version) .....................................69

Page 6: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Preface

HYDAC FILTER SYSTEMS GMBH en(us) Page 6 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Preface

For you, as the owner of a product manufactured by us, we have produced this manual, comprising the most important instructions for its operation and maintenance.

It will acquaint you with the product and assist you in using it as intended in an optimal manner.

Keep it in the vicinity of the product so it is always available.

Note that the information on the unit's engineering contained in the documentation was that available at the time of publication.There may be deviations in technical details, figures, and dimensions as a result.

If you discover errors while reading the documentation or have additional comments or suggestions, contact us at:

HYDAC FILTER SYSTEMS GMBH Technische Dokumentation Postfach 12 51 66273 Sulzbach / Saar

Germany

We look forward to receiving your input.

“Putting experience into practice”

Page 7: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Preface

HYDAC FILTER SYSTEMS GMBH en(us) Page 7 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Technical Support Contact our technical sales department if you have any questions on our product. When contacting us, please always include the model/type designation, serial no. and part-no. of the product:

Fax: ++49 (0) 6897 / 509 - 846

E-mail: [email protected]

Modifications to the Product We would like to point out that changes to the product (e.g. purchasing additional options, etc.) may mean that the information in the operating instructions is no longer applicable or adequate.

After modification or repair work that affects the safety of the product has been carried out on components, the product may not be returned to operation until it has been checked and released by a HYDAC technician.

Please notify us immediately of any modifications made to the product whether by you or a third party.

Warranty For the warranty provided by us, please refer to the General Terms of Sale and Delivery of HYDAC Filter Systems GmbH.

Refer to these at www.hydac.com General terms and conditions.

Page 8: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Preface

HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Using the Documentation

Note that the method described for locating specific information does not release you from your responsibility of carefully reading these instructions prior to starting the unit up for the first time and at regular intervals in the future.

What do I want to know?

I determine which topic I am looking for.

Where can I find the information I’m looking for?

The documentation has a table of contents at the beginning. There, I select the chapter I'm looking for and the corresponding page number.

The documentation number with its index enables you to order another copy of the operating and maintenance instructions. The index is incremented every time the manual is revised or changed.

Chapter description

Page number

Edition date

Document language Documentation no.with index/

file name

Page 9: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Safety Information and Instructions

HYDAC FILTER SYSTEMS GMBH en(us) Page 9 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Safety Information and Instructions

These operating instructions contain the key instructions for properly and safely operating the FluidAqua Mobil.

The safety of the FAM largely depends on maintenance. Regular required maintenance is described in this manual. We will support you if the unit needs repair and will provide original spare parts.

Obligations and Liability

The basic prerequisite for the safe and proper handling and operation of the FAM is knowledge of the safety instructions and warnings.

These operating instructions in general, and the safety precautions in particular, are to be adhered by all those who work with the FluidAqua Mobil.

Adherence is to be maintained to pertinent accident prevention regulations applicable at the site where the product is used.

The safety precautions listed herein are limited solely to using the FluidAqua Mobile.

The FluidAqua Mobil has been designed and constructed in accordance with the current state of the art and recognized safety regulations. Nevertheless, hazards may be posed to the life and limb of the individual using the product or to third parties. Risk of damage may be posed to the product or other equipment and property.

The FluidAqua Mobil is to be used:

solely for its designated use

only when in a safe, perfect condition

Any faults or malfunctions which might impair safety are to be properly repaired or remedied immediately.

Our General Terms and Conditions apply. They are made available to the owner upon concluding purchase of the unit at the latest. Any and all warranty and liability claims for personal injuries and damage to property shall be excluded in the event they are attributable to one or more of the following causes:

Improper use of the FluidAqua Mobil or use deviating from its designated use

Improper commissioning, operation and maintenance of the FluidAqua Mobil

Operating the FluidAqua Mobil when the safety systems are defective

Modifications to the FluidAqua Mobil made by the user or purchaser

Improper monitoring of unit components that are subject to wear and tear

improperly performed repair work

Page 10: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Safety Information and Instructions

HYDAC FILTER SYSTEMS GMBH en(us) Page 10 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Explanation of Symbols and Warnings, etc.

DANGER DANGER denotes situations which can lead to death if safety precautions are not observed.

WARNING WARNING denotes situations which can lead to mortal injury if safety precautions are not observed.

CAUTION CAUTION denotes situations which can lead to severe injuries if safety precautions are not observed.

NOTICE NOTICE denotes situations which can lead to property damage if safety precautions are not observed.

General Safety Precautions

Maintenance work is to be carried out only by trained personnel.

The safe operation of this unit can only be ensured if it is used for the purpose it was intended. If there is any question about the use, please contact the manufacturer. The manufacturer will not accept responsibility for damages resulting from misuse of this equipment.

The following applies to all work performed using the unit: adherence to pertinent national regulations pertaining to accident prevention and safety at the workplace in addition to any applicable internal rules and regulations of the owner/operator, even though they are not specifically cited herein.

Leaks of dangerous materials must be properly collected and disposed of so as not to harm any persons or the environment. The corresponding statutory regulations are to be followed.

Page 11: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Safety Information and Instructions

HYDAC FILTER SYSTEMS GMBH en(us) Page 11 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Proper/Designated Use

The FAM was developed for the dewatering, filtration and degassing of hydraulic and lubricating oils. It removes free water, emulsified water and a large percentage of the water to be found in solution.

Any other use shall be deemed to be improper and not in keeping with the product's designated use; the manufacturer accepts no liability for any damage resulting from such use.

Proper or designated use of the product extends to the following:

Maintaining adherence to all the instructions contained herein.

Performing requisite inspection and maintenance work.

Improper Use

Any use deviating from the proper/designated use described above is prohibited.

Improper use may result in hazard to life and limb. Examples of improper use:

Operation with a non-approved fluid.

Operation under non-approved conditions.

Safety Devices

Prior to starting up the FAM unit each time, make sure that all the protective safety devices are properly installed and are in proper working order.

Safety devices may not be removed until the product has been shut down and secured against being restarted (e.g. warning sign or padlock on the main switch).

When the product is supplied in partial consignments, the safety devices are to be applied by the operator as specified by law/pertinent regulations.

Page 12: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Safety Information and Instructions

HYDAC FILTER SYSTEMS GMBH en(us) Page 12 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Informal Safety Precautions

Make sure to always keep the operating instructions in the vicinity of the FluidAqua Mobil unit.

Apart from the operating instructions, any and all general and local regulations pertaining to accident prevention and environmental protection are to be made available and observance is to be maintained to them.

Make sure to keep the safety and hazard symbols and warnings on the FluidAqua Mobil in a legible condition at all times. If necessary, replace the instruction sheets.

FluidAqua Mobil components may not be opened except in normal, non-contaminated environments. The mains cable must be pulled out before the FluidAqua Mobil is opened. Tests conducted with the housing open may only be performed by properly trained, certified electricians. This also applies to all repair work or any HYDAC-approved modifications to electrical components.

Check the hoses and connectors for leaks on a daily basis.

Check the electrical equipment of the FluidAqua Mobil on a monthly basis. Replace any loose connections or damaged cables immediately.

WARNING Hydraulic systems are under pressure

Risk of personal injury

► The hydraulic system must be depressurized before performing any work on the hydraulic system.

Page 13: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Safety Information and Instructions

HYDAC FILTER SYSTEMS GMBH en(us) Page 13 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Training and Instruction of Personnel

The FAM may only be operated by properly trained and instructed personnel.

The areas of responsibility of your staff must be established in a clear-cut manner.

Staff undergoing training may not use the FAM unless supervised by an experienced staff member.

Indi

vidu

als

Indi

vidu

als

unde

rgoi

ng tr

aini

ng

Indi

vidu

als

with

te

chni

cal

trai

ning

/eng

inee

ring

back

grou

nd

Ele

ctric

ian

Sup

ervi

sor

with

the

appr

opria

te a

utho

rity

Activity

Packing / transportation X X X

Commissioning X X X

Operation X X X X

Troubleshooting/ locating the source of malfunction

X X X

Remedying of mechanical faults X X

Remedying of electrical faults X X

Maintenance X X X X

Repair work X

Decommissioning / Storage X X X X

Safety Measures to Be Followed during Normal Operation

Do not put the FAM into operation unless all the safety devices function properly.

The unit is to be checked once a day for visible external damage and for the proper functioning of the safety devices.

Page 14: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Safety Information and Instructions

HYDAC FILTER SYSTEMS GMBH en(us) Page 14 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Electrical Hazards

DANGER Electric shock

Risk of personal injury with fatal consequences

► Any work involving electrical equipment may only

be done by a properly trained, certified electrician.

Check the FAM electrical equipment on a monthly basis. Immediately replace loose connections, damaged cables and hoses with original spare parts.

Maintenance, Servicing and Troubleshooting

The prescribed adjustment, maintenance/servicing and inspection work is to be conducted in accordance with the respective schedules.

Disconnect the FluidAqua Mobil from the power supply when performing any maintenance, inspection or repair work. Protect the FAM from being inadvertently switched back on.

Any screwed fittings which have been undone/removed are to be checked to see that they have been properly resecured.

Always check the product to see that it functions properly when performing maintenance and servicing work.

Modifications to the FluidAqua Mobil

Do not make any modifications (design modifications, extensions) to the FluidAqua Mobil without the prior consent of the manufacturer.

Any modifications require written permission from HYDAC Filter Systems GmbH.

Immediately replace any machine components which are not in perfect condition.

Only use original spare parts and consumables.

Page 15: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Unpacking the FAM

HYDAC FILTER SYSTEMS GMBH en(us) Page 15 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

What to Do in Case of Emergency

In the event of an emergency, turn the main switch by 90° to the "OFF" position. The entire system is then voltage-free downstream from this switch.

Normal pressure is restored to the vacuum column after ≈ 1 minute.

Unpacking the FAM

The FAM is inspected for leaks and proper functioning at the factory, then carefully packed for shipment. The connectors/ports are closed off so that no contamination can enter the unit while it is in transit.

When receiving and unpacking the unit, check it for damage in transit.

The packaging material should be re-used/re-cycled as appropriate for your area.

Page 16: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Transporting the FAM

HYDAC FILTER SYSTEMS GMBH en(us) Page 16 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Transporting the FAM

CAUTION High empty weight (≈ 400 … 850 kg, depending on the version)

Risk of personal injury

► Use at least 2 persons to shift the FAM.

NOTICE

Using components for pushing / pulling

The FAM will be damaged

► Never use the components to push or pull the FAM.

► Use the grips provided for shifting.

Drain the unit completely before transport. Seal off the inlets and/or outlets.

On the mobile FAM, wind the suction and pressure hose and the connection cable around the holders provided for this purpose and fasten them in place.

Shift the FAM manually using the rollers. When doing so, use only the grips provided on the FAM frame for pushing.

Before shifting the unit, make sure to release the hand brake on the steering rollers.

Once the FAM is in its new position, actuate the hand brake on the steering rollers.

Forklift

Use the lift eyes when transporting the unit with a forklift.

Crane

NOTICE

Page 17: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Transporting the FAM

HYDAC FILTER SYSTEMS GMBH en(us) Page 17 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Unsuitable lifting accessories

The unit will be damaged / Components will be damaged / destroyed

► Use only suitable lifting accessories to raise or lash the FAM.

► Take care to ensure that the lifting accessories does not cause any pressures to be brought to bear against the components on the FAM.

Train/truck

For transport by rail or truck, supports must be placed under the mobile FAM so that the rollers are subjected to no load pressure.

The FAM is to be secured with suitable belts.

Page 18: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Scope of Delivery

HYDAC FILTER SYSTEMS GMBH en(us) Page 18 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Scope of Delivery

Upon receiving the FAM check it for any damage in transit. The FAM may not be set up and installed unless it is in perfect order. Any damage in transit is to be reported to the forwarding agent or the department in charge immediately; the unit may not be commissioned until this damage is properly remedied.

The following items are included in the scope of delivery:

Item Qty. Designation

1 1 FluidAqua Mobil, ready for connection with suction and pressure hose

2 1 Tool, square 6 mm for switch box

- 1 Operating and Maintenance Instructions

- - Supplementary operating instructions (depending on FAM design)

- 1 Operating and maintenance instructions for vacuum pump

1

2

Page 19: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 FAM Description

HYDAC FILTER SYSTEMS GMBH en(us) Page 19 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

FAM Description

The FluidAqua FAM was developed for the dewatering, filtration and degassing of hydraulic and lubricating oils. It removes free water, emulsified water and a large proportion of the water in solution. The integrated fluid filter ensures an efficient separation of solid particles.

Degassing of the medium is also achieved via the vacuum created in the vacuum chamber.

FAM Versions

Stationary (FAM-xx-1-..) Mobil (FAM-xx-2-..)

In addition to the stationary version, the mobile FAM has:

- 2 casters, 2 bock rolls

- Suction hose, L= 5 m

- Return hose , L= 5 m

- Connection cable 10m

Page 20: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 FAM Description

HYDAC FILTER SYSTEMS GMBH en(us) Page 20 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Performance

The FAM is able to reduce the water content of those fluids described in the chapter "Suitable fluids" to less than 100 PPM.

Typical dewatering rates:

FAM 10 ≈ 0.8 l/h

FAM 10/15 ≈ 1.0 l/h

The dewatering speed is dependent on:

Dewatering rate

Water content

Fluid temperature

Detergent additives

Volumetric flow of the FAM

As an approximate guideline, the dimensioning of the FluidAqua Mobil can be defined in accordance with the tank volume.

Tank volume in liters Filter Size

< 2,000 FAM 10

2,000 – 7,000 FAM 10/15

Note that free water, e.g. on the bottom of the tank, can have an influence on the entire water content in the system. Drain off free water before taking a sample.

For the determination of the water content, the sample should be extracted from the system to be cleaned at a test connection during operation if at all possible.

Page 21: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 FAM Description

HYDAC FILTER SYSTEMS GMBH en(us) Page 21 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

FAM system components

(13)

1

3

2

5

4

10

11

8

69

Page 22: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 FAM Description

HYDAC FILTER SYSTEMS GMBH en(us) Page 22 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Item Designation

1 control cabinet with control panel consisting of: - main switch - alarm signal lamp - control panel with display - component diagram

2 Regulating valve to regulate the necessary vacuum in the vacuum chamber

3 Vacuum gauge

4 Motor-pump unit (OLF Compact)

5 Vacuum chamber

6 3/2-directional control valve

7 Suction strainer

8 Vacuum pump

9 Evacuation pump

10 Fluid filter for the separation of solid particles

11 Air filter and dryer

13 Heater mounted in vacuum chamber (Option)

Page 23: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 FAM Description

HYDAC FILTER SYSTEMS GMBH en(us) Page 23 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Suitable Fluids

NOTICE

Impermissible operating conditions

The unit will be damaged

► The unit may be used only in connection with the media mentioned below.

Depending on the version (see model code), the FAM may only be used for the following media (see model code):

FAM-xx-M-.. Mineral oils according to DIN 50524 Gear oils according to DIN 51517, 51524 Other hydraulic and lubrication oils which require or are

compatible with NBR seals

FAM-xx-B-.. Synthetic esters (HEES) DIN 51524/2 Vegetable oils (HETG, HTG) Hydraulic and lubrication oils which require or are

compatible with Viton seals.

FAM-xx-X-.. HFD fluids (not for pure phosphate esters which require EPDM seals).

Page 24: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Dimensions

HYDAC FILTER SYSTEMS GMBH en(us) Page 24 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Dimensions

Stationary version 17

13

550

620

620

1081

G 1

/2"

G 1

"10

0O

UT

IN

510

Page 25: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Dimensions

HYDAC FILTER SYSTEMS GMBH en(us) Page 25 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Mobile version

620

175

6

1081

G 1

/2"

G 1

"

100

OU

TIN

510

552

Page 26: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Hydraulic diagram

HYDAC FILTER SYSTEMS GMBH en(us) Page 26 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Hydraulic diagram

Item Designation

1 control cabinet with control panel consisting of: - main switch - alarm signal lamp - control panel with display - component diagram

2 Regulating valve to regulate the necessary vacuum in the vacuum chamber

3 Vacuum gauge

4 Motor-pump unit (OLF Compact)

5 Vacuum chamber

6 3/2-directional control valve

7 Suction strainer

8 Vacuum pump

9 Evacuation pump

10 Fluid filter for the separation of solid particles

11 Air filter and dryer

12 Transport eyelets

13 Heater mounted in vacuum chamber (Option)

10

7

4 2 11 5

9 6

3

(13)

Page 27: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Hydraulic diagram

HYDAC FILTER SYSTEMS GMBH en(us) Page 27 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Function Description

After the FAM has been switched on, the filter pump unit 1 starts to draw the fluid from the tank to be cleaned and transfers it with a flow rate of 10 l/min into the vacuum chamber 3. At the same time, the vacuum pump 7 creates a vacuum inside the vacuum chamber. The fluid to be cleaned trickles downward through special filling materials. These filling materials are responsible for forming a thin fluid film and reducing the speed of the trickle. At the same time, a stream of air flowing upward from below against this current assimilates the water from the fluid and conveys it to the outside by means of the vacuum pump. Once level 02 is reached on the level switch, emptying pump 5 is switched on in order to convey around 15 l/min to the outlet. The filling level falls until level 01 is reached, after which solenoid valve 6 switches on and the two pumps convey into the vacuum chamber. The filling level continues to rise until level 02 is reached, after which the solenoid valve shuts back off again, the filling level falls again, etc.

After actuating the stop switch, the filter-pump unit stops and the evacuation pump empties the vacuum chamber.

Switching the main switch to "OFF" renders all engines and valves voltage-free.

The vacuum in the vacuum chamber is regulated with the regulating valve 8.

Page 28: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Hydraulic diagram

HYDAC FILTER SYSTEMS GMBH en(us) Page 28 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Possible Applications

Bypass purification (dewatering, filtration and degassing).

The FAM is connected with suction and pressure lines to the tank and cleans the medium to be found there in continuous operation.

Transfer by pumping (dewatering, filtration and degassing)

The FAM is connected to a contaminated oil tank by means of a suction line and pumps the fluid into the clean oil tank.

However, better purification results are achieved with bypass purification. Therefore, the FAM should be operated after pumping in the bypass flow.

To avoid overfilling the purified oil tank, permanently monitor the oil level.

Contaminated oil

Clean oil

Page 29: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 FAM Set-up and Connection

HYDAC FILTER SYSTEMS GMBH en(us) Page 29 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

FAM Set-up and Connection

Setting up the FAM

The following points are to be observed when setting up the FAM:

Place the unit on a horizontal and level surface. Special mounting is not required.

The locking brakes on the wheels must be locked to avoid unintended movement of the unit (only on the mobile version).

Position the FAM in the immediate vicinity of the tank to be cleaned. Observe the maximum permitted suction pressure. See page 31.

Observe the ambient temperature on the type label of the FAM.

Access to the main switch has to be unimpeded at all times.

An area of at least 0.8 m must be kept free for maintenance purposes on both the left hand and right hand sides of the FAM.

Take care to ensure free access to the control panel and that the control cabinet can be opened completely.

Notes on Piping / Hosing

NOTICE

Non-permitted pressure at the inlet IN / outlet OUT

Failure malfunction

► Determine the pressure to be anticipated at the inlet / outlet with the prescribed values.

Note that the cross-section of the connected hoses/piping, must be at least as large as the cross-section of the inlet/outlet port sizes.

In order to keep the pressure loss as low as possible, use as few threaded connections as possible.

The pressure at the inlet/outlet depends on the height differential between the FAM fluid surface in the tank port (aspiration height P(height)) and the line losses (P(line)).

If the FAM is above the fluid surface in the tank, determine the pressure as follows: P = 1 bar - P(height) - P(line)

If the FAM is below the fluid surface in the tank, determine the pressure as follows: P = 1 bar + P(height) - P(line)

Page 30: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 FAM Set-up and Connection

HYDAC FILTER SYSTEMS GMBH en(us) Page 30 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Determine the pressure loss P(height) as follows: P(height)[bar] = h/10. h stands here for the distance between the suction connection on the FAM and the fluid surface in the tank.

The pressure loss in a hydraulic line (P(line)) depends on the following:

Flow rate

Kinematic viscosity

Pipe dimensions

Fluid density

The pressure loss in straight pipes (P(line)) can be calculated as follows:

Δp ≈ 6.8 * L / d4 * Q * V * D Δp = Pressure differential in [bar]

L = Pipe length [m]

d = Internal pipe diameter [mm]

Q = Flow rate [l/min]

V = Kinematic viscosity [mm²/s]

D = Density [kg/dm³] Mineral oil based hydraulic fluid has a density of ≈ 0.9 kg/dm³.

The density (D) is:

Mineral hydraulic oil HLP 0.85 – 0.90 kg / dm³

Phosphate ester HFDR 1.00 – 1.10 kg / dm³

Insulating oil 0.90 – 1.00 kg / dm³

Lubrication oil 0.90 – 0.95 kg / dm³

Additional threaded connections and pipe bends increase the pressure differential and must be taken into account.

Keep the height difference between the pump and the oil level in the tank as low as possible.

Avoid constrictions in the connected hoses. They compromise output and increase the risk of cavitation.

Make sure that no tension or vibrations are carried over to the pump or filter housing when the pipes are connected. Use hoses or expansion joints if necessary.

Page 31: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 FAM Set-up and Connection

HYDAC FILTER SYSTEMS GMBH en(us) Page 31 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Connecting the inlet (IN)

The suction pressure at the FAM inlet must be in the range of -0.4 bar.

Use a negative pressure-resistant, flexible hose or a pipe for the suction-side connection.

Note that the cross-section of the connected hoses/piping, must be at least as large as the cross-section of the inlet/outlet port sizes. The cross-section of the connection line (FAM 75-95) must be at least DN50.

The shape of the tank connection should be set up in such a way that it will always be lower than the level of the oil.

NOTICE

Severe contamination

The unit will be damaged

► Do not prime directly at the bottom of the tank

► Do not prime in the sump

► Never prime without a built-in suction screen

The greatest contamination is found on the bottom of the tank. All impurities and other particles are deposited on the bottom of the tank.

Connect outlet (OUT)

Take care to ensure that the maximum pressure of 2 bar is not exceeded at the outlet.

NOTICE

OUT connection closed off

The unit will be destroyed

► Check to be sure that all of the locking fixtures at the outlet are in "open" position each time before start-up.

To prevent air from entering the medium, make sure that the pressure hose with lance is always below the oil level in operation.

Page 32: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 FAM Set-up and Connection

HYDAC FILTER SYSTEMS GMBH en(us) Page 32 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Preparing the Vacuum Pump

As a basic rule, the rotary vane vacuum pump is not filled with oil at the time the FAM is delivered. A sufficient amount of vacuum pump oil for the initial filling is included in the scope of delivery.

NOTICE

Operation without vacuum pump oil

The rotary vane vacuum pump will be destroyed.

► The rotary vane vacuum pump requires oil as its operating medium.

► Check the oil level before start-up, topping up with vacuum pump oil VE101 if necessary.

The maximum filling amount for vacuum pump oil VE101 is 0.3 liters.

The oil level is monitored automatically by the integrated float switch while the FAM is in operation. A message will appear on the control panel when the MIN MAX filling level has been reached.

Synthetic vacuum pump oil in accordance with DIN 51506, Group VDL, ISO VG100 must be used for topping up.

The vacuum pump oil can be ordered from HYDAC:

Description Part no.:

Vacuum pump oil VE101 DIN 51506, Group VDL, ISO VG100, synthetic

1 liter 06018128

Vacuum pump oil VE101 DIN 51506, Group VDL, ISO VG100, synthetic

5 liters 06018129

Page 33: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 FAM Set-up and Connection

HYDAC FILTER SYSTEMS GMBH en(us) Page 33 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Filling up the rotary vane vacuum pump

Remove the safety screw for the plug at the level switch (1).

Remove the plug (2) for the level switch.

Unscrew the level switch (3) from the vacuum pump.

Fill up with vacuum pump oil VE101 through the filling nozzle (4).

Check the oil level through the inspection glass (5).

Fill up the vacuum pump oil no higher than the MAX marking.

We recommend filling up to a level between the MIN and MAX markings.

The suction strainer is re-mounted in reverse order.

12

34

5

Page 34: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 FAM Set-up and Connection

HYDAC FILTER SYSTEMS GMBH en(us) Page 34 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Electrical Connection of the FAM

Check that the voltage and frequency information on the type label coincides with the voltage supply present.

The FAM has two shockproof plugs. Each is marked with its respective function.

Item Description Supply voltage Performance

(A) FAM 220 … 240 V DC, 50 Hz, 1 Ph 2000 Watt

(B) Heater 220 … 240 V DC, 50 Hz, 1 Ph 2900 Watt

Plug each of the FAM's power plugs into a power outlet that has been fitted with a 16A fuse.

NOTICE

Heater without voltage

The operating indicator lamp "Heater on" is not illuminated

► Plug in the heater's power plug or check the voltage available to the heater.

Page 35: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 FAM Set-up and Connection

HYDAC FILTER SYSTEMS GMBH en(us) Page 35 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Checking the settings on the heater (optional) If necessary the desired fluid temperature can now be set.

For this, the upper cover of the heater must be opened with appropriate equipment and the desired operating temperature set. (The working temperature is preset to 55°C; if 80°C is exceeded, then the safety thermostat will shut the device off.)

(A) Fluid temperature 55°C

(B) Safety thermostat 80°C

Page 36: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Starting up the FAM

HYDAC FILTER SYSTEMS GMBH en(us) Page 36 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Starting up the FAM

NOTICE

Connection IN / OUT closed

The unit will be destroyed

► Check to be sure that all of the locking fixtures at the inlet / outlet are in "open" position each time before start-up.

Switching on the FAM

Move main switch to the "ON" position and wait until the "Fault" lamp lights up briefly (lamp test) and this message: appears on the display.

Ready for operation

--> Press start

Press the key. Now the system starts with the starting phase:

FluidAqua operating

Starting

1. Vacuum pump starts.

2. Inlet valve opens and the vacuum chamber is filled.

3. When the vacuum chamber is filled to filling level 02, the evacuation pump is started up and the starting period is finished.

4. Automatic mode is active:

FluidAqua operating

Automatic mode

Now regulate the vacuum pressure with the throttle valve on the right of the vacuum gauge, depending on the operating viscosity.

Operating viscosity Pressure setting (absolute pressure)

50 mm2/s hydraulic oil 300 mbar / 30 kPa

300 mm2/s lubrication oil 400 mbar / 40 kPa

50 mm2/s ester oil 400 mbar / 40 kPa

Some oils begin to foam intensively when they enter the vacuum chamber. This can especially be the case when the water content is very high, but also when the oil is old or high viscosity media are used. If too much foam is caused, it is drawn in by the vacuum pump.

In this case, for example, reduce the vacuum from 300mbar to 400mbar.

Page 37: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Starting up the FAM

HYDAC FILTER SYSTEMS GMBH en(us) Page 37 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Venting the filter housing

WARNING Hot fluid

Risk of burns

► Note that hot fluid may exit during bleeding.

NOTICE

Air in the fluid filter

The filter element is not utilized completely

► Bleed the fluid filter after each shut-down / filter element replacement.

The fluid filter is to be bled through the air-vent plug until fluid emerges. Using an Allen wrench carefully open the air-vent plug in the cover.

Shut off FAM

Press the key. Now the unit starts the run-down phase:

FluidAqua operating

Run-down

1. The inlet valve closes

2. The vacuum pump is switched off

3. The filter-pump unit runs until the vacuum chamber is emptied.

Wait until the message appears on the display. Now move the main switch to the position "0". The system is now switched off.

Ready for operation

--> Press start

Page 38: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Operating the FAM

HYDAC FILTER SYSTEMS GMBH en(us) Page 38 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Operating the FAM

Except for the regulation of the vacuum pressure, each function is assigned to a key or a key combination on the control panel. The functions on the upper part of the keys (or those on the keys with only one assignment) can be triggered directly by pressing the respective key. To select the function on the lower part of the keys

(double allocation of keys), you must press beforehand.

The key assignments and the descriptions of the Main and Manual Menus are described below.

Main menu

Function key Function

Start FAM. Start and transition in Automatic mode.

Stop FAM (run-down and evacuation of the vacuum chamber)

Acknowledge all remedied faults.

Button for special functions (depending on version)

+

Select German language (or other languages depending on version).

+

Select English language (or other languages depending on version).

Text line 1

Text line 2

Page 39: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Operating the FAM

HYDAC FILTER SYSTEMS GMBH en(us) Page 39 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Manual Menu 1

In order to access Manual Menu 1 from the Main Menu, press the following keys in succession.

/

The key assignment in the Manual Menu 1 is as follows:

Evacuation pump off / on

Vacuum pump off / on

3/2 directional valve close / open

Key Key assignment in Manual Menu 1

Function

Manual Menu 2 Switches in the Manual Menu 2

"E-Pump off" Switches evacuation pump off

"E-Pump on" Switches evacuation pump on

(Press the buttons one after the other)

"V-pump off" Switches vacuum pump off

"V-pump on" Switches vacuum pump on

(Press the buttons one after the other)

"Valve close“ Closes the 3/2-directional control valve.

"Valve open“ Opens the 3/2 directional valve.

(Press the buttons one after the other)

Back With this key combination you can exit the Manual Menu

This menu also switches itself off automatically after 15 minutes.

Page 40: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Operating the FAM

HYDAC FILTER SYSTEMS GMBH en(us) Page 40 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Manual Menu 2

To change from Manual Menu 1 to the Manual Menu 2 press the key:

The key assignment in the Manual Menu 1 is as follows:

Feed pump filter unit on / off

Key Key assignment in Manual Menu 1

Function

"Feed pump-filter unit off"

Switches the Feed pump-filter unit off

"Feed pump-filter unit on"

Switches the Feed pump-filter unit on

(Press the buttons one after the other)

Back -> Manual Menu 1

Pressing these keys in succession allows you to exit Manual Menu 2 to go back to Manual Menu 1

Back ->

Exit

Pressing these keys in succession allows you to exit Manual Menu 2 to go back to the Main Menu

The manual menu switches itself off automatically after 15 minutes.

Page 41: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Troubleshooting

HYDAC FILTER SYSTEMS GMBH en(us) Page 41 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Troubleshooting

Error message Cause Remedy (also see following detailed description)

Vacuum chamber overfilled

Maximum filling level 03 in the vacuum chamber exceeded.

Empty vacuum chamber and restart FAM 10.

Level switch 02/03 Level switch defective Check sensor and replace if necessary

Level switch 01/02 Level switch defective Check sensor and replace if necessary

Filling pos. V-Pump Min Lack of oil in the vacuum pump

Fill oil into the vacuum pump according to manual

Filling pos. V-Pump Max.

Too much oil in vacuum pump

Drain oil so that the level is in between max and min

Change fluid filter Maximum differential pressure at the filter reached

Change the filter element

Float switch Oil pan of the FAM filled Empty oil pan, check FAM 10 for leakage

E/F/V pump motor protection

Electric motor of the evacuation or feed or vacuum pump overloaded

Switch on the motor protection in the E controller, check operating viscosity

Incorrect phase sequence

Engines have incorrect direction of rotation

Rotate two phase wire on power cord 180°.

Dry running E-pump Vacuum chamber is being filled too slowly, flow into vacuum chamber is smaller than flow rate of the evacuation pump.

Clean the suction strainer in front of the inlet valve, check the viscosity of the operating medium and the cross-section of the length of suction hose, check the filling pump, check the dirt indicator of the fluid filter

Outlet blocked No operating fluid is pumped out of the FAM 10

Check shut-off devices in return line, check 3/2 valve

Inlet clogged No operating fluid is pumped into the FAM 10

Check shut-off devices in suction line, check pump-filter unit

Plug FCU missing The FCU operating mode was selected without connecting the Sub D plug

Connect plug or change operating mode.

Page 42: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Troubleshooting

HYDAC FILTER SYSTEMS GMBH en(us) Page 42 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Error message Cause Remedy (also see following detailed description)

of the FCU.

Heater fault (only with option heater)

Oil too hot due to interruption of oil flow or faulty thermostat.

Reset the safety thermometer. Remove the cover of the heater to do so. Clean the suction strainer, length of the suction hose may need to be shortened.

No parameterblock Setup settings of the TD200 deleted.

Reenter setup values.

(Please contact our Technical Sales Office. For contact information, see the chapter "Foreword")

Electrical fuses Automatic safety devices have been installed into the control cabinet for the security of the operating voltage transformer.

Fuse

F1 The primary side of the transformer has 2A 400V

F2 The primary side of the transformer has 2A 400V

F3 The secondary side of the transformer has 3A 250V

F4 The secondary side of the transformer (control current circuit) has 1A 250V

F5 The secondary side of the transformer (control current circuit) has 1A 250V

F6 Contamination Sensor Display CSD (Optional)

F7 Contamination Sensor CS (Optional)

Page 43: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Troubleshooting

HYDAC FILTER SYSTEMS GMBH en(us) Page 43 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Vacuum chamber overfilled

Ste

p

Description

Malfunction:

"Reactor overfilled"

1. Press RESET key

2. Select manual menu 1, start evacuation pump

3. Can vacuum chamber be emptied within 3.5°minutes from level 03 to level 01?

->a. Yes -> proceed to Step 5. 4.

->b. No -> proceed to Step 6.

5. Excessive foaming -> vacuum in the vacuum chamber too high – reduce vacuum (e.g. from 300 mbar(abs) -> 400 mbar(abs))

6. Check the evacuation pump for proper functioning

7. The failure is repaired.

Page 44: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Troubleshooting

HYDAC FILTER SYSTEMS GMBH en(us) Page 44 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Level switch 01/02(02/03)

Ste

p

Description

Malfunction:

Level switch 01/02(02/03)

1. Switch off the FAM at the main switch.

2 Disconnect plug at level switch and remove level switch.

3. Manually check the functioning of the switching points

->a. Switching points are available -> proceed to Step 5. 4.

->b. Switching points not available -> proceed to Step 6.

5. Inspect the plug and the cable to and in the control cabinet for the PLC. Functioning?

-> proceed to Step 7.

6. Replace level switch

->a. Function available -> proceed to Step 8. 7.

->b. Function not available -> proceed to Step 9.

8. The PLC is defective.

9. Replace the damaged parts

Page 45: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Troubleshooting

HYDAC FILTER SYSTEMS GMBH en(us) Page 45 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

V-pump fill level min/max

Ste

p

Description

Malfunction:

"Filling level V-pump min/max"

1. Press RESET key

->a. Error message lapsed -> proceed to Step 6. 2.

->b. Error message still illuminated -> proceed to Step 3.

3. Check oil level; fill up or drain oil as necessary

Check level sensor of the vacuum pump and the supply lines

->a. The fault recurs -> proceed to Step 5. 4.

->b. The fault does not recur -> proceed to Step 6.

5. Replace damaged parts

6. The failure is repaired.

Page 46: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Troubleshooting

HYDAC FILTER SYSTEMS GMBH en(us) Page 46 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Replace fluid filter

Ste

p

Description

Malfunction:

"Replace fluid filter"

1. Press RESET key

2. Switch off the FAM at the main switch

3. Close ball faucets on the IN and OUT connections

4. Empty filter housing via the drain ball valve

5. Open filter housing, remove filter elements from housing

6. Insert new filter elements, close filter housing

7. Open ball faucets on the IN and OUT connections

8. Switch the FAM on with the main switch

9 Does the fault recur after ≈ 4 minutes?

->a. Yes -> go to step 11. 10.

->b. No -> proceed to Step 12.

11. Check differential pressure switch and supply lines

12. The failure is repaired.

Page 47: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Troubleshooting

HYDAC FILTER SYSTEMS GMBH en(us) Page 47 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Float switch

Ste

p

Description

Malfunction:

"Float switch"

1. Empty oil pan

2. Press RESET key

3. Start FAM

4. Carry out leakage check, remedy leakages

5. Does the fault recur after ≈ 2 seconds?

->a. Yes -> proceed to Step 7. 6.

->b. No -> proceed to Step 8.

7. Check float switch and supply lines

8. The failure is repaired.

Page 48: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Troubleshooting

HYDAC FILTER SYSTEMS GMBH en(us) Page 48 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

E/F/V pump motor protection

Ste

p

Description

Malfunction:

"E/F/V pump motor protection"

1. Switch off FAM at the main switch and pull out mains plug

2 Open the switch cabinet door, switch the motor protection switch and/or fuse back on "I" that has tripped and check the setting of the motor protection switch and/or the voltage capacity of the fuse in accordance with the motor type plate and/or the heater type plate.

->a. The fault recurs -> proceed to Step 4. 3.

->b. The failure does not recur -> proceed to Step 8.

4. Measure the current consumption of the engines whose motor protection switch or fuse has tripped.

5. Close switch cabinet door, main switch to "On", start FAM

->a. The fault recurs -> proceed to Step 7. 6.

->b. The failure does not recur -> proceed to Step 9.

7. Check the operating viscosity of the medium (only for motor protection switch), perform maintenance work on the vacuum pump

8. The failure is repaired.

Page 49: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Troubleshooting

HYDAC FILTER SYSTEMS GMBH en(us) Page 49 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Incorrect phase sequence

Ste

p

Description

Malfunction:

"Wrong motor rotation"

1. Switch off FAM, pull out mains plug

2 Replacement of two phases on the mains plug, check mains voltage

3. Plug mains plug back in, switch on main switch

4. Check direction of rotation in accordance with the direction of rotation arrow on the pumps

->a. The fault recurs -> proceed to Step 6. 5.

->b. The failure does not recur -> proceed to Step 7.

6. Pull out mains plug, check mains cable and plug, replacing as necessary

7. The failure is repaired.

NOTICE

Direction of rotation, electric motor

The unit will be damaged

► After replacing electric motors, check the direction of rotation in manual operation before commissioning.

Page 50: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Troubleshooting

HYDAC FILTER SYSTEMS GMBH en(us) Page 50 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Dry running E-pump

Ste

p

Description

Fault "Dry running, E-pump"

1. Cause: Insufficient inflow into vacuum chamber

2. The suction-side pressure loss is too great. Possible remedies:

- Clean the suction strainer.

- Reduce the suction height

- Reduce absolute pressure in the vacuum chamber e.g. from 500 mbar(abs) -> 400 mbar(abs).

- Increase fluid temperature

- Increase diameter of the suction hose

- Shorten length of the suction hose

- Check the 2/2-directional solenoid valve at the inlet

3. Suction port connection in the tank sucks in air:

- Position the suction port connection lower

- Tank to be cleaned is too small

4. Check the functioning of vacuum pump in manual mode

5. Find and remedy corresponding cause

6. To reset the failure, press the RESET key

7. The failure is repaired.

Page 51: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Troubleshooting

HYDAC FILTER SYSTEMS GMBH en(us) Page 51 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Outlet (OUT) clogged

Ste

p

Description

1. Fault: "Outlet clogged"

Cause: Medium cannot flow out of the vacuum chamber (emptying pump pressure limit = 5 bar)

2. Check locking features in the return-flow line. Open if blocked.

3. To reset the failure, press the RESET key

4. Does the fault recur after ≈ 4 minutes?

->a. Yes -> proceed to Step 6. 5.

->b. No -> proceed to Step 7.

6. Check the pressure relief valve of the evacuation pump for proper functioning.

7. The failure is repaired.

Page 52: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Troubleshooting

HYDAC FILTER SYSTEMS GMBH en(us) Page 52 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Inlet (IN) clogged

Ste

p

Description

1. Malfunction: "Inlet clogged"

2. Cause: Medium cannot make its way into the vacuum chamber

3. Check locking features in the suction line. Open if blocked.

Check suction strainer and clean if necessary

4. To reset the failure, press the RESET key

5. Does the fault recur after ≈ 4 minutes?

->a. Yes -> proceed to Step 7. 6.

->b. No -> proceed to Step 8.

7. Check the following points:

- Check the suction pipe for cross-section constriction

- Check the functioning of vacuum pump in manual mode

- Check the position of the shut-off device in the suction line

8. The failure is repaired.

Page 53: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Troubleshooting

HYDAC FILTER SYSTEMS GMBH en(us) Page 53 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

FCU plug missing (only in conjunction with accessory FCU)

Ste

p

Description

1. Fault: "FCU plug missing"

2. Cause: Connection cable to the FluidControl Unit FCU is not connected.

3. Establish the connection with a connection cable

4. To reset the failure, press the RESET key

5. Fault appears again?

->a. Yes -> proceed to Step 7. 6.

->b. No -> go to step 10.

7. Check connection cable and plug; replace if necessary

8. Fault appears again?

->a. Yes -> proceed to Step 4. 9.

->b. No -> go to step 10.

10. The failure is repaired.

Page 54: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Troubleshooting

HYDAC FILTER SYSTEMS GMBH en(us) Page 54 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Reset safety thermostat on the heater

The safety thermostat is triggered when the flow of oil through the heater is too low or if the thermostat is damaged.

After the fault has been remedied, the safety thermostat can be reset by pressing the green key (C) in the head. To protect yourself from burns, wait until the heater has cooled down before doing so.

Page 55: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Performing Maintenance

HYDAC FILTER SYSTEMS GMBH en(us) Page 55 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Performing Maintenance

The safety of all persons coming into contact with the FAM and the availability of the unit for use are extremely dependent on service and maintenance.

Routine service and maintenance work is described in the documentation included:

Regular visual inspection of all hoses and conduits for leakage.

Regular visual inspection of the electrical apparatus for possible damage to lines, plugs, sensors or cables and/or devices in the control cabinet.

Regular visual inspections of filling level, vacuum pump

Before carrying out any maintenance work the FAM must be shut down safely. This means:

1. Main switch in position "0",

2. Power cord disconnected.

3. Mains cable wound around cable drum apparatus present on the FAM. (This last as a protection against switching back on again!)

Page 56: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Performing Maintenance

HYDAC FILTER SYSTEMS GMBH en(us) Page 56 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Changing the Air Filter

Replace the air filter (1) every six months.

Change the air filter (1) more frequently if you operate the FAM in a very dusty/humid environment.

Unscrew the air filter (1) by hand.

Screw in the new air filter (1) by hand and tighten it only by hand.

Vacuum pump

Changing oil and oil filter

The first oil change must take place after 100 operating hours. Later oil changes after ≈ 3000 hours or every six months.

The vacuum pump must be at operating state temperature when the oil is changed:

Put the FAM into operation for 10 minutes.

Switch off the FAM at the main switch.

Wait until the pressure in the vacuum column has reached atmospheric pressure (≈ 1000 mbar abs).

To perform these tasks, you will need the following tools and equipment:

Compressor oil VE101 (for part-no., see replacement parts list)

Maintenance set, comprising oil filter, air deoiling element, ventilation cover seal (for part-no., see replacement parts list).

Fork wrenches SW 27 / SW 32

To change the oil, proceed as follows:

1. Put a suitable container in place to catch the used oil.

2. Open the oil drain (c) and allow the used oil to drain into the container. Dispose of the used oil properly.

3. Close the screw plug when the flow of oil diminishes. Let the pump run again briefly for a few seconds,

4. Open the screw plug once again and drain off the remaining oil.

5. Screw in the screw plug and tighten it firmly.

6. Remove the oil filter and dispose of it properly.

7. Install the new oil filter.

8. Fill in new oil through the oil filling screw plug (a) until the MAX marking (d) has been reached.

Page 57: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Performing Maintenance

HYDAC FILTER SYSTEMS GMBH en(us) Page 57 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Replacing the Air De-oiling Element

Oil mist or increased electricity consumption by the drive engine (response of the engine protection switch) are signs of a soiled air de-oiling element.

To replace the air de-oiling element, proceed as follows:

1. Remove the ventilation cover (i) from the oil separator (g).

2. Undo the screw in the middle of the filter spring, but do not remove this.

3. Press the filter spring down and rotate it out of the depression.

4. Remove the filter spring from the oil separator (g).

5. Pull the air de-oiling element out of the oil separator (g).

6. Make sure that the new air deoiling element is provided with a new O-ring.

7. Install the air de-oiling element in such a way that the opening is situated correctly in the holder in the oil separator (g).

8. Make sure that the tip of the screw in the middle of the filter spring protrudes ≈ 2 – 5 thread rounds out of the filter spring.

9. Install the filter spring in such a way that the ends are secured against slipping by lugs in the receptacles in the oil separator (g) and the tip of the screw engages in the depression of the air de-oiling element.

10. Tighten the screw in the filter spring until the head of the screw impacts against the spring steel sheet.

11. Make sure that the seal under the ventilation cover is clean and undamaged. Replace the seal if necessary.

12. Fasten the ventilation cover (i) with seal and hexagon-head screws to the oil separator (g).

The air de-oiling element becomes saturated with oil during operation. A slight lowering of the oil filling level after the air de-oiling element has been replaced is therefore quite normal.

Page 58: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Clean the suction strainer.

HYDAC FILTER SYSTEMS GMBH en(us) Page 58 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Clean the suction strainer.

NOTICE

Operation without a suction strainer

The pump will be destroyed

► The FCU may not be operated without the suction strainer.

► Clean the suction strainer regularly.

To protect the pump from coarse contamination particles and other foreign bodies, a dirt trap with a strainer insert is fitted at the pump inlet.

Clean the suction strainer at regular intervals.

To disassemble/install the suction strainer, proceed as follows:

1. Unscrew the screw plug (1) counterclockwise with an open-jaw wrench.

2. Remove strainer insert (3) and clean it.

3. Clean the strainer insert, flush it out and then blow-flush it using compressed air.

3. Check strainer insert (3) and sealing ring (2) on the screw plug for damage and replace if necessary.

4. Insert the strainer insert (3).

5. Manually screw in the screw plug (1) clockwise.

Tighten the screw plug (1) with a open-jaw wrench.

6. Check the suction strainer fitting for leaks after commissioning.

12

3

Page 59: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Clean the suction strainer.

HYDAC FILTER SYSTEMS GMBH en(us) Page 59 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Replacing the filter element on the fluid filter

As soon as the error message "change fluid filter" appears on the display the filter should be taken out of the housing and be replaced by a new one.

The contaminated filter must to be disposed of according to the local regulations as hazardous waste.

To change the filter element, proceed as follows:

1. Switch off the system and close the locking devices at the inlet and outlet ports

Make sure that the unit cannot be inadvertently switched back on during maintenance work.

2. Put a suitable container (1) in place to catch escaping operating medium.

Depressurize the filter housing.

Carefully open the drain plug (2) on the filter bowl and empty the filter housing completely using the drain screw (6mm) on the bottom of the filter housing.

The volume is ≈2 liters.

After draining, replace the drain plug in the filter bowl

3. Undo the tensioning clamp (1) on the filter housing and then remove the housing by moving it downwards (2).

Use a 6 mm Allen wrench (1).

Page 60: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Clean the suction strainer.

HYDAC FILTER SYSTEMS GMBH en(us) Page 60 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

4. Pull the filter bowl including filter element downwards.

5. Remove (2) the filter element from the filter bowl by turning slightly (1).

Page 61: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Clean the suction strainer.

HYDAC FILTER SYSTEMS GMBH en(us) Page 61 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

6. Clean the inside of the filter bowl of dirt and deposits

7. Check the O-ring on the filter bowl for damage.

Replace it if necessary.

Moisten the O-ring on the filter bowl with the operating medium.

8. Moisten the O-rings on the filter element with the operating medium.

Page 62: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Clean the suction strainer.

HYDAC FILTER SYSTEMS GMBH en(us) Page 62 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

9. Turning it slightly, press the new filter element down into the mount on the filter bowl.

10. Push the filter bowl with the filter element up into the mount on the filter head.

Page 63: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Clean the suction strainer.

HYDAC FILTER SYSTEMS GMBH en(us) Page 63 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

11. Push the tensioning clamp (1) up around the filter bowl over the bead on the filter bowl / filter head.

Tighten the tensioning clamp with a 6 mm Allen wrench in a clockwise direction.

The tightening torque is 5 Nm.

12. Check that the drain plug on the filter bowl is secure.

13. The filter element change is now complete.

After commissioning, check the filtration unit for leaks.

Checking the settings on the heater (optional)

Under normal operating conditions the heater element is maintenance-free.

Page 64: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Spare Parts

HYDAC FILTER SYSTEMS GMBH en(us) Page 64 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Spare Parts

FAM 10

Quantity

Designation Part no.

1 Air filters 0160 MU 003 M 0126 5765

1 Drain pump incl. motor *

1 Suction strainer 1" assy., consisting of: 0063 5624

1x strainer insert, 250µm 0601 6797

1x sealing ring 0603 3969

1 Vacuum pressure gauge 0063 9989

1 3/2-directional control valve 0601 4177

1 Pressure hose, Length = 5 m 0601 3308

1 Suction hose, Length = 5 m 0600 4781

1 Level sensor (vacuum chamber) 0120 4801

*) available on request

Fluid filter (OLF 5/10)

Quantity

Designation Part no.

1 Filter pumps unit OLF 5/10 *

1 Filter element, filtration rating 2 µm N5DM002 0034 9494

1 Filter element, filtration rating 5 µm N5DM005 0306 8101

1 Filter element, filtration rating 10 µm N5DM010 0310 2924

1 Filter element, filtration rating 20 µm N5DM020 0302 3508

*) available on request

Vacuum pump

Quantity

Designation Part no.

1 Vacuum pump *

1 Service kit vacuum pump consisting of exhaust filter and seals 0601 4161

1 Vacuum pump oil VE101, 1 liter 0601 8128

1 Vacuum pump oil VE101, 5 liter 0601 8129

*) available on request

Page 65: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Contact / Service

HYDAC FILTER SYSTEMS GMBH en(us) Page 65 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Contact / Service

For product information, technical support or if you have comments or suggestions concerning this manual, please contact:

HYDAC FILTER SYSTEMS GMBH

Telephone: ++49 (0) 6897 509 1174

Fax: ++49 (0) 6897 509 846

E-mail: [email protected]

Regular inspection and maintenance work is indispensable for ensuring trouble-free operation and long service life for your FluidAqua Mobil.

Our HYDAC Servicenter offers you these tasks within agreed-upon time frames and at fixed prices.

HYDAC Service GmbH Rehgrabenstrasse 3 66128 Saarbrücken - Dudweiler

Germany

Telephone: ++49 (0)681 509 883

Fax: ++49 (0)681 509 324

E-mail: [email protected]

Page 66: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Storing the FAM/Taking it out of Operation

HYDAC FILTER SYSTEMS GMBH en(us) Page 66 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Storing the FAM/Taking it out of Operation

Drain the unit completely, including all of its components, the same way as is done before putting it into storage.

Pull out the power plug and securely fasten the hoses and power cord to the unit.

Store the unit in a clean, dry (non-condensing) space.

Observe the storage temperature range of: –20C … 60C.

Disposing of the FAM

When decommissioning and/or disposing of the unit, observe all local guidelines and regulations pertaining to occupational safety and environmental protection. This applies in particular to the oil in the unit, components covered with oil and electronic components.

After disassembling the unit and separating the various materials, reuse them or dispose of them properly in accordance with local regulations.

Type label

The type label is located on the front panel, where it is clearly visible. Copy the data from the type label into the following, in order to have at hand when required.

Page 67: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Technical data

HYDAC FILTER SYSTEMS GMBH en(us) Page 67 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Technical data

Refer to the type label for specific data of the unit.

FAM 10/10 FAM10/15

Filter size OLF5 OLF5

Filter Element N5DMxxx N5DMxxx

Dirt absorption capacity as per ISO 4572

200 g 200 g

Clogging indicator VM 2 C.0 VM 2 C.0

Pump types Vane pump (feed pump), gear pump (drain pump

Vane pump (feed pump), gear pump (drain pump

Flow rate in l/min at 50Hz 10 15

Operating pressure in bar 3.5 3.5

Viscosity range in mm²/s 15 - 800 15 - 800

Electrical power consumption

1500 W without heater 4400 W with heater

2000 W without heater 4900 W with heater

Length of electronic cable in m

10 10

IP class IP 54 IP 54

Length of hoses in m 5 5

Hoses, material NBR NBR

Connection Inlet/outlet G 1 / G 1/2 G 1 / G 1/2

Material of sealings NBR (FPM) NBR (FPM)

Fluid temperature in°C 10 - 80 10 - 80

Ambient temperature in °C 10 - 40 10 - 40

Typical dewatering speed l/h

0.8 l/h (without heater at 40°C)

1 l/h (without heater at 40°C)

Minimum water content in ppm

< 100 < 100

Page 68: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 Model Code

HYDAC FILTER SYSTEMS GMBH en(us) Page 68 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

Model Code

FAM - 10 - M - 1 - A/1900 - 060 - DM - 02 - S - z - 1Product FAM = FluidAqua Mobil Filter Size 10 = 10 l/min, 50 Hz / 12 l/min, 60 Hz 10/15 = 15 l/min, 50 Hz / 18 l/min, 60 Hz

Operating fluid M = Mineral oil – NBR seals, checked with

mineral oil I = Insulation oil – NBR seals, checked with

insulation oil X = HFD-R phosphoric acid ester fluids –

FPM seals checked with HFD-R fluid* P = Aviation phosphoric acid esters e.g.

Skydrol® or Hyjet®

checked with Skydrol®* B = Readily biologically degradable (on ester

base) – FPM seals *Residues of test liquid will remain in the unit after testing.

Mechanical type 1 = stationary 2 = mobile

Voltage / Frequency / Power supply A = 400 V, 50 Hz, 3 Ph+PE B = 415 V, 50 Hz, 3 Ph+PE C = 200 V, 50 Hz, 3 Ph+PE D = 200 V, 60 Hz, 3 Ph+PE E = 220 V, 60 Hz, 3 Ph+PE F = 230 V, 60 Hz, 3 Ph+PE G = 380 V, 60 Hz, 3 Ph+PE H = 440 V, 60 Hz, 3 Ph+PE J = 230 V, 50 Hz, 3 Ph+PE K = 480 V, 60 Hz, 3 Ph+PE L = 220 V, 50 Hz, 3 Ph+PE M = 230 V, 50 Hz, 1 Ph+PE N = 575 V, 60 Hz, 3 Ph+PE O = 460 V, 60 Hz, 3 Ph+PE X = other voltage

Filter size fluid filter 05 = OLF-5 Filter material of element DM = Dimicron® z = without filter element

Filtration rating 02 = 2 µm 05 = 5 µm 10 = 10 µm 20 = 20 µm z = without filter element Type of vacuum pump S = Standard – rotary vane vacuum pump Instrumentation z = Without A = AquaSensor (AS) C = ContaminationSensor (CS) AC = AquaSensor + ContaminationSensor (CS only with measured value

display)

Modification number 1 Supplementary details Without details = standard V = Sealing material Viton

Page 69: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

FluidAqua Mobil - FAM 10 EU declaration of conformity (only for mobile version)

HYDAC FILTER SYSTEMS GMBH en(us) Page 69 / 72

BEWA FAM10 3549460a en-us 2011-05-05.doc 2011-05-05

EU declaration of conformity (only for mobile version)

HYDAC FILTER SYSTEMS GMBH

Postfach 12 51 66273 Sulzbach / Saar Germany

Industriegebiet 66280 Sulzbach / Saar Germany

Telephone: ++49 (0) 6897 509 01

Internet: www.hydac.com

EC declaration of conformity FS / 13 / 09 No.

We hereby declare that the following designated product, on the basis of its design and construction, and in the version which we have brought to market, corresponds to the fundamental safety and health requirements contained in the standards listed below.

Any modification of this product that is not coordinated with us in writing will cause this declaration to lose its validity.

Designation FluidAqua Mobil

Type FAM - series

Part no. -

Serial-no. -

EU-Machinery Directive 98/37/EG

EU Electrical Equipment Regulations 2006/95/EC

EU Directive on Electromagnetic Compatibility 2004/108/EC

Applied standards EN 292-1/2, EN982

DIN EN 61000-6-2:2001

2009-09-11 Dr. Andreas Schunk Date Name (CE-authorized person)

Executive director:

Mathias Dieter, Dipl.Kfm. Wolfgang Haering

Registered seat of company: 66280 Sulzbach / Saar - Germany

Registry Court: Saarbrücken, HRB 17216

Value added tax identification number : DE 815001609

Tax number : 040/110/50773

Page 70: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72
Page 71: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72
Page 72: FAM 10 - HYDAC · FAM Description ... Manual Menu 1 ... FluidAqua Mobil - FAM 10 Preface HYDAC FILTER SYSTEMS GMBH en(us) Page 8 / 72

HYDAC FILTER SYSTEMS GMBH Industriegebiet Postfach 1251 66280 Sulzbach/Saar 66273 Sulzbach/Saar Germany Germany Phone: +49 (0) 6897 509 01 Central Fax: +49 (0) 6897 509 846 (Technical Department) Fax: +49 (0) 6897 509 577 (Sales Department) Internet: www.hydac.com E-mail: [email protected]