F AC 31 08 [a] Gas Chromatogrphy
-
Upload
krushnasamy-suramaniyan -
Category
Documents
-
view
219 -
download
0
Transcript of F AC 31 08 [a] Gas Chromatogrphy
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 1/68
Chapter 31
Dong-Sun Lee/ CAT-Lab / SWU 2012-Fall version
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 2/68
Gas Chromatography (GC)
Introduction
Gas chromatography is a chromatographic technique that can be used to
separate volatile organic compounds.
Two types of GC are encountered: gas-solid chromatography(GSC) and
gas-liquid chromatography(GLC). GLC is finds widespread use in all
fields of science, where its name is usually shortened to GC.
A gas chromatograph consists of a flowing mobile phase, an injection port, a
separation column containing the stationary phase, and a detector.
GSC is based on a solid stationary phase on which retention of analytes is the
consequence of physical adsorption. GLC is based on partitioning behavior of the analyte between the mobile gas phase and the liquid stationary phase in
the column.
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 3/68
Characteristics of GLC
1. Sensitivity : mg ~ pg (10 – 3 ~ 10 – 9 g)
2. Versatility : from rare gases to liquids and solids in solution
with 800~1000 MW
3. Speed of analysis : typically 5 ~ 30 min ,
complex 3 ~ 30 mixture separation
4. Reproducibility : qualitativeAccuracy : quantitative
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 4/68
GC-14A gas chromatograph (Shimadzu)with an integrator.
HP 5890 gas chromatograph (Hewlett
Packard) with an integrator
GC-MS : GC-Q plus ion-trap GC-MSn ( Thermoquest - Finnigan ) Xcalibur software
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 5/68
Fuel
gas
Carrier
gas
Regulator
Trap Make-up gas
Splitvent
Injector Detector
Column oven
Column
Fuel
gas
Electrometer
Recorder Integrator
Computer
Valve/ Gage
Schematic diagram of a gas chromatograph system.
Basic components of Gas Chromatograph
Carrier gas supply
Sample introduction inlet
Column and controlled-temperature oven
Detector & oven
Recorder
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 6/68
Carrier gas system in gas chromatograph
The purpose of the carrier is to transport the sample through the column to the detector.
The selection of the proper carrier gas is very important because it affects both column and
detector performance. The detector that is employed usually dictates the carrier to be used.
From a column performance point of view a gas having a small diffusion coefficient is
desirable (high molecular weight, e.g., N2, CO2, Ar) for low carrier velocities while large
diffusion coefficients (low molecular weight, e.g., H2, He) are best at high carrier velocities.
The viscosity dictates the driving pressure. For high-speed analysis, the ratio of viscosity to
diffusion coefficient should be as small as possible. H2 would be the best choice, followed
by helium.
The purity of the carrier should be at least 99.995% for best results. Impurities such as air
or water can cause sample decomposition and column and detector deterioration. In
temperature programmed runs, impurities in the carrier gas such as water can be retained at
low temperatures but are then eluted at higher temperatures impairing the baseline. Manyinstrument problems have been traced to contaminated carrier gases.
The carrier must also be inert to the components of the sample and the column.
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 7/68
Properties of common carrier gases
Gas molecular weight Thermal conductivity Viscosity
× 105 at 100oC × 10 – 6 at 100oC
(g-cal/sec-cm- oC ) (P)
Ar 39.95 5.087 270.2 *
CO2 44.01 5.06 197.2
He 4.00 39.85 234.1
H2 2.016 49.94 104.6**
N2 28.01 7.18 212.0
O2 32.00 7.427 248.5***
* at 99.6oC ** at 100.5oC **** at 99.74oC
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 8/68
Using the correct carrier and detector gases are an important factor in installing
a new GC. The five gases commonly used as carrier gas and detector fuels in
capillary gas chromatography are helium, hydrogen, nitrogen, argon-methane,
and air. The types of gases necessary are partly determined by the detectionsystem used. Factors to consider for each individual gas are discussed below.
Carrier Gas Choice
Carrier gases that exhibit a broad minimum on a van Deemter profile are
essential in obtaining optimum performance. Volumetric flow through a
capillary column is affected by temperature. When temperature programming
from ambient to 300oC, the flow rate can decrease by 40 percent. A carrier gas
that retains high efficiency over a wide range of flow rates and temperatures is
essential in obtaining good resolution throughout a temperature programmed run.Figure 1 shows the van Deemter profile for hydrogen, helium, and nitrogen
carrier gases.
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 9/68
Van Deemter curves for GC of n-C17H36
at 175oC, using N2, He, or H2 in a 0.25
mm diameter × 25 m long wall coated
column with OV-101 stationary phase
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 10/68
Hydrogen is the fastest carrier gas (uopt), with an optimum linear velocity of
40cm/sec, and exhibits the flattest van Deemter profile. Helium is the next
best choice, with an optimum linear velocity of uopt = 20cm/sec. Nitrogen's
performance is inferior with capillary columns because of its slow linear velocity, uopt = 12cm/sec. Argon-methane has a slower optimum linear
velocity than nitrogen and is not recommended for use as a carrier gas with
capillary columns. Air is not recommended as a carrier gas because it can
cause stationary phase oxidation.
With hydrogen and helium as carrier gases, the minimum H.E.T.P. valuescan be maintained over a broader range of linear velocities than with nitrogen,
and high linear velocities can be used without sacrificing efficiency. Nitrogen
is beneficial only when analyzing highly volatile gases under narrow
temperature ranges where increasing stationary phase interaction is desirable.
Otherwise, the use of N2 results in longer analysis times and a loss of resolution for compounds analyzed on a wide temperature range.
http://www.restekcorp.com/gcsetup/gcsetup3.htm
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 11/68
Exert Caution when Using Hydrogen as a Carrier Gas
Hydrogen is explosive when concentrations exceed 4% in air . Proper safety precautions
should be utilized to prevent an explosion within the column oven. Most gas
chromatographs are designed with spring loaded doors, perforated or corrugated metalcolumn ovens, and back pressure/flow controlled pneumatics to minimize the hazards
when using hydrogen carrier gas. Additional precautions include:
• Frequently checking for leaks using an electronic leak detector.
• Using electronic sensors that shut down the carrier gas flow in the event of pressure loss.
• Minimizing the amount of carrier gas that could be expelled in the column oven if a leak
were to occur by installing a flow controller (needle valve) prior to the carrier inlet
bulkhead fitting to throttle the flow of gas (for head pressure controlled systems only) as
shown Fig. 2.
• Fully open the flow controller (needle valve) and obtain the proper column head pressure,
split vent flow, and septum purge flow rates. Decrease the needle valve flow rate until the
head pressure gauge begins to drop (throttle point). Next, increase the flow controller
(needle valve) setting so that the right amount of flow is available to the system. Should a
leak occur, the flow controller will throttle the flow, preventing a large amount of
hydrogen from entering the oven.
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 12/68
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 13/68
Make-up and Detector Fuel Gases
Gas added to the stream after the column is called makeup gas. Choosing the correct
make-up and detector gases will depend on both the detector and application. Most GC
detectors operate best with a total gas flow of approximately 30ml/min. to ensure high
sensitivity and excellent peak symmetry. Refer to your GC manual for optimum flowrates on different instruments. Carrier gas flows for capillary columns range from 0.5 to
10ml/min. which are well below the range where most detectors exhibit optimal
performance. To minimize detector dead volume, make-up gas is often added at the
exit end of the column to increase the total flow entering the detector. Make-up gas
helps to efficiently sweep detector dead volume thereby enhancing detector
sensitivity.
Make-up gas can be added directly to the hydrogen flame gas for flame ionization
detectors (FID), nitrogen phosphorous detectors (NPD), and flame photometric detectors
(FPD) or added to the column effluent by an adaptor fitting. However, GCs such as
Perkin-Elmer and Fisons do not require make-up gas.
Combustion type detectors (FID, NPD, FPD) use three gases: make-up, hydrogen (fuelgas), and air (combustion/oxidizing gas). For non-combustion detectors, such as the
thermal conductivity detector (TCD), electron capture (ECD), and photo ionization
detector (PID), only carrier and make-up gases are required. In the case of the
electrolytic conductivity detector (ELCD), the make-up gas is hydrogen, as a reaction gas
in the halogen and nitrogen mode or air in the sulfur mode. Table I shows recommended
gases for various detectors.
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 14/68
Carrier gases and detector fuel gases for use with various GC detectors
TCD ECD FID NPD FPD ELCD
PID
Carrier Gases He O O O O O O O
H2 O - O - O O O
N2 O O O O O - O
Combustion/
Reaction Gases
H2 - - O O O O -
Air - - O O O - -
Make-up Gases N2 O O O O O - O
He O - O O O - O
ArCH2 - O - - - - -
http://www.restekcorp.com/gcsetup/gcsetup3.htm
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 15/68
Effect of impurities
- Impurities such as hydrocarbon, oxygen, water
contribute to unwanted noise levels, excessive baseline drift .
- Molecular sieve --- Moisture trap, Oxygen trap, Chemical filter
Effect of water on column efficiency
- Carrier gas dryness is very important !! (use anhydrous sodium sulfate)
- Water can and usually does react with some portion of the column.
This results in loss of resolution and tends to produce asymmetric
or tailing peaks.
Unwanted components or ghost peaks may also appear.
Another effect is a net loss of sensitivity.
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 16/68
Gas purifiers
The trap will remove any water vapor or oils that may have been introduced inthe filling process since a number of gases are water pumped. The contaminants
removed by the trap could otherwise interact with the column packing material
to produce spurious peaks. In addition the contaminants can cause increased
detector noise and drift.
The traps should be reconditioned (about twice a year ) by heating to 300 oC
for 4~8 hr with a stream of gas passing through it or in a vacuum oven.
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 17/68
Carrier gas purifiers
GC1 2 3 4
Gas
cylinder
1. Hydrocarbon trap
2. Moisture trap
3. Oxygen trap
4. Indicating oxygen trap
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 18/68
Gas purifier recommendation for GC applications
Capillary column GC Carrier Hydrocarbon, Moisture, Oxygen
with any detector Make-up None - all detector but
ECD moisture & oxygen
Air for FID Hydrocarbon
H2
for FID None
ELCD reaction gas Hydrocarbon
Packed column GC Carrier Hydrocarbon, Moisture, Oxygen
with FID or TCD
Packed column GC Carrier Hydrocarbon, Moisture, Oxygen
with ECD, FPD, NPD,
MSD
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 19/68
Flow requirements
1. Stable
2. Reproducible
3. Convenient
The more constant the flow rates,
the more precise and accurate the results.
Flow controller ( Pressure controller )
--- to maintain precise and accurate flow rates
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 20/68
Effect of decreased flow rate or lower temperature
- All peaks have shifted to longer retention times
- Apparent loss of peak height- The base of each peak is wider ,
however, individual peak area remain constant.
Effect of increased flow rate
- Sample components are squeezed toward
the injection point
- Cause two components to elute together,
appearing as single peak
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 21/68
Regulations of carrier gas
Carrier cylinder bottled at about 2500 psi(150-160 atm)
Two stage pressure regulator :
- first stage : high inlet pressure
- second stage : low outlet pressure
( set at 40~100psi)
Gas generators
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 22/68
Gas flow rate control
A 1 % change in carrier gas flow rate will cause a 1% change in retention time.
For all these reasons it is important to keep the flow of the carrier gas constant.
1. Control of carrier gas inlet pressure
2. Control of carrier gas flow rate
In isothermal operation the means of regulation is immaterial because both
means provide constant inlet pressure as well as constant flow rate. In
temperature programmed runs, however, the situation is quite different. If onemaintains the inlet pressure constant the flow rate will change. Therefore, with
temperature programming of the column, the flow rate must be controlled.
Pressure controllers
1. The second stage regulator on the cylinder
2. A pressure regulator mounted in the GC
3. A needle valve(variable restrictor) mounted in the GC
4. A fixed restrictor mounted in the GC
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 23/68
Flow measurement
1) Rotometer
The column flow rate is typically indicated by a rotometer .
( Calibrate equilibrium position indicating the flow )
Rotometer is operated by the volume of gas passing a ball in a tapered cell.
2) Bubble meter
3) Electronic flow sensor
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 24/68
Relationships between inside diameter, column length, mesh size, and carrier
gas flow for packed column
Inside diameter Mesh size for Mesh size for Carrier flow
mm length up to 3m length over 3m N2, ml/min He or H2, ml/min
2 100~120 80~100 8~15 15~30
3 100~120 80~100 15~30 30~60
4 80~100 60~80 30~60 60~100
John A. Dean, Analytical Chemistry Handbook, McGraw-Hill, 1995, p. 4.31 .
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 25/68
Comparison of 1/8-in(0.316 cm) packed, wide bore, and WCOT columns
1/8 in packed Wide bore WCOT
Inside diameter, mm 2.2 0.53 0.25
Film thickness, m 5 1~5 0.25
Phase volume ratio() 15~30 130~250 250
Column length, m 1~2 15~30 15~60
Flow rate, ml /min 20 5 1
Effective plates(Neff ) per meter 2000 1200 3000
Effective plate height (Heff ),mm 0.5 0.6 0.3
Typical sample size 15 g 50 ng
John A. Dean, Analytical Chemistry Handbook, McGraw-Hill, 1995, p. 4.31 .
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 26/68
Recommended ranges of gas flow rates
Detector Gases Range Capillary Packed
FID Carrier 2 ml/min 40 ml/min
Hydrogen 30~50 ml/min 35 ml/min 40 ml/min
Air 300~600 ml/min 350 ml/min 500 ml/min
Make-up(N2) 10~60 ml/min 30 ml/min not used
NPD Hydrogen 2~4 ml/min
Air 40~80 ml/min
Make-up(N2, He) 10~20 ml/min not necessary
PID Make-up 5~10 ml/min
Sheath 30~40 ml/min
FPD Carrier 1~3 ml/min 30~50 ml/min
Hydrogen 85~100 ml/min 100~120 ml/min
Air 100~120 ml/min 110~135 ml/min
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 27/68
Inlet requirements
1. Temperature controlled
2. Low volume (total swept by carrier )
3. Inert construction
Column Overload
If too large an sample were allowed to enter small bore(capillary) columns
column overload and a loss of resolving power would like occur.
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 28/68
Liner
All liners help protect the vaporized sample form contacting the metal wall of the inlet as
sample flows onto the column. Deactivated glass wool may be used as an aid for sample
vaporization, to minimize discrimination based on boiling point, and to provide a surface on
which non0volatiles can be trapped. The simplest liner is a straight tube, which gives all-
around good performance at low cost. Single-taper liners improve on a straight tube byminimizing sample vapor contact with metal at the bottom of the injection port, although
they are somewhat more expensive. Liners are deactivated borosilicate glass, except quartz
where noted. Liners are guaranteed inert for phenols, organic acids and bases.
Why is Glass Wool Added to an Injector Liner? - General GC
The Glass wool serves three major purposes.
The Glass wool will prevent the small pieces of septa from reaching the column.
The presence of Glass wool will help the injected sample stay in the liner a little
longer which will help the sample to vaporize and mix more thoroughly with the
carrier gas.
If positioned properly it will wipe the outer surface of the syringe needle andim rove the recision of the li uid in ection.
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 29/68
Inlet configuration
1. Direct column inlet --- 1/8 " OD or larger column
sampling syringe is actually inserted into the end of the column
needle guide / cap / spring / septum
mounting holes / carrier gas in / inlet body column
Swagelock ferrules / Swagelock nut
2. Splitter inlet --- open tubular column or less than 1/8" OD column
Because of the limited capacity for sample of these small bore columns and
the difficulty of injecting extremely small volume samples, a large portion of
the injected sample is vented to atmosphere by the inlet.
septum / preheated carrier gas / mixing tube / restrictor
buffer volume / tapered needle / gold gasket / column fitting
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 30/68
Injection port for split injection into an open tubular column. The glass liner is slowly
contaminated by nonvolatile and decomposed samples and must be replaced periodically.
For splitless injection, the glass liner is a atraight tube with no mixing chamber. For dirty
samples, split injection is used and a packing material can be replaced inside the liner to
adsorb undesirable components of the sample.
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 31/68
Common injection techniques
1) Hot flash vaporization
Direct
Cold-trap
Split
Splitless
2) Direct cold : on column
Split or on-column
Split
1) Simple
2) High column efficiency3) Column may be protected
On-column
1) Best accuracy
2) Thermolabile compounds
3) Trace analysis
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 32/68
Representative injection conditions for split, splitless, and on-column injection into an
open tubular column.
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 33/68
Direct injector1) Good sensitivity
2) Low column efficiency
3) Best for thick films, widebore column ( 0.53 mm )
Hot on-column injectors1) Reduced column efficiency
2) Best with thick films, widebore columns
3) Nonvolatiles may damage column4) Cold on-column injector may be used with 0.1 to 0.53 mm i.d. columns
Advantage of on-column1) Best reproducibility : Quantitative results
2) No split, no loss of high boilers3) "Cold" on-column injection available
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 34/68
Advantage of splitless
1) High sensitivity ( 95 % of sample on column )2) Solvent effect produces narrow sample bands
3) Same hardware as split injection
Disadvantage of splitless
1) Slow sample transfer to column
2) Must dilute sample with volatile solvent
3) Time consuming : must cool column
4) Poor for thermolabile compounds
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 35/68
Split and splitless injections of a solution containing 1 vol % methyl isobutyl ketone
(bp 118 oC) and 1 vol % p-xylene (bp 138 oC) in dichloromethane (bp 40 oC) on a BP-
10 moderately polar cyanopropyl phenyl ,ethyl silicone open tubular column(0.22
mm I.d., 10 m long, 0.25 m, column temperature =75 oC).
C i j i
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 36/68
Common injection methods
Syringe injection
Valve injection
Sampling Syringe
0 ~ 1 μL --- the sample is totally
confined to the needle
0 ~ 5 / 0 ~ 10 μLneedle / barrel / plunger
Gas sampling valve
Sample inSample loop --- 1/4 or 10 mL loop size
compatible with needs
Sample vent
Carrier gas to column
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 37/68
Solvent effect
t-1 t-2
time t-1 --- just after injection, solvent and sample are condensed in a
long plug at the front of the column.
The column temperature must be cold enough to condense the
solvent.
time t-2 --- after some time, the column temperature has been raised,
most of the solvent has evaporated, and the solvent effect
has left the sample molecules concentrated in a narrow band.
As the column is further heated, the remaining solvent andsample molecules are rapidly vaporized
--- resulting in high column efficiency and narrow peak.
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 38/68
Syringe for solid phase microextraction.
Sampling by SPME and desorption of analyte from the coated fiber into a gas
chromatograph.
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 39/68
Purge and trap apparatus for extractingvolatile substances from a liquid or solid
by flowing gas.
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 40/68
GC column
Parts of Column
1) Tubing material
Stainless steel--- reactive ( steroids, amines, free acids )
Glass ------------ can be made inert, difficult handling
Fused silica ---- flexible
most inertmost popular
high resolution
2) Stationary phase
Solid support --- carefully sized granular
Liquid phase --- active portion of the column
Porous polymers
Adsorbents
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 41/68
Important column parameters
1) Inside diameter
2) Length
3) Film thickness
4) Stationary phase composition
5) Flow rate
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 42/68
Column diameter
i.d. Resolution Speed Capacity Ease
100 micrometer +++ +++ + +
(narrow bore )
250, 320 ++ ++ ++ ++
(mid bore)530 + ++ +++ +++
(wide bore)
Column length
N œ L R œ L1/2 t R œ L
Col mn
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 43/68
24-foot 1/8" packed column
wound on 6" coil 6-foot 1/4" packed column
wound on 5" coil
60-meter 0.53mm metal
wide bore
capillary column wound on
3.5" coil
15-meter 0.53mm fused silica
wide bore capillary column
wound on 7" cage30-meter .25mm metal narrow
bore capillary column wound
on 3.5" coilhttp://www.srigc.com/catalog/columns.htm
ColumnGlass wool --- both ends of the column
1- ½” (inlet side)
1/4” (detector side)
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 44/68
Fused silica surface made by the reaction of SiCl
4and water vapor
in a flame
- SiO2 contains 0.1 % – OH groups
- Very inert
- Uniform chemical surface
Fused silica - High tensile strength
- Flexible
- Sheath of polyimide
- Very inert
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 45/68
Fused Silica Capillary Columns A fused silica capillary column is comprised of three major parts (Figure 1).
Polyimide is used to coat the exterior of fused silica tubing. The polyimide protects
the fused silica tubing from breakage and imparts the amber-brown color of columns.
The stationary phase is a polymer that is evenly coated onto the inner wall of thetubing. The predominant stationary phases are silicon based polymers
(polysiloxanes), polyethylene glycols (PEG, Carbowax) and solid adsorbents.
F igure 1.
Capillary columns have to be properly
installed to maximize their performance
and lifetime. You can obtain enhanced
column performance and lifetime by
following these recommended installation
guidelines. More detailed installation,
operational and troubleshootinginformation can be found in the following
references
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 46/68
WCOT = Wall Coated Open Tubular
invented and patented by Dr Marcel Golay
Tubing - Fused silica
- Glass
- Stainless steel
Liquid phase coating
WCOT - - - High resolution
Film thickness : 0.5 to 5.0 micrometer
i.d. : 0.10, 0.25, 0.32, 0.53 mm
Length : 10 to 60 m
Open tubular GC column
O ti l id li f t b l GC l
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 47/68
Operational guideline for open tubular GC columns
WCOT
narrow intermediate wide bore
Column inner diameter, mm 0.25 0.32 0.53
Maximum sample volume, l 0.5 1 1
Maximum amount for
one component, ng 2~50 3~75 5~100
Effective plates(Neff ) per meter 3000~5000 2500~4000 1500~2500
Trennzahl(separation) number
per 25 m 40 35 25
Optimum flow for N2, ml /min * 0.5~1 0.8~1.5 2~4
Optimum flow for He, ml /min ** 1~2 1~2.5 5~10
Optimum flow for H2, ml /min *** 2~4 3~7 8~15
* Optimum velocity is 10 to 15 cm/s for each column
** Optimum velocity is 25 cm/s for each column
*** Optimum velocity is 35 cm/s for each column
Oth t f ill l
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 48/68
Other types of capillary columns
SCOT = Support Coated Open Tubular
Solid support : Celite
Liquid phase
Not available fused silica tubing
PLOT = Porous Layer Open Tubular
Porous adsorbent : alumina or molecular sieve
* Molecular sieve --- efficient for H2, Ne, Ar, O2, N2, CO, CH4.
Porous carbon stationary phase ( 2 m
thick) on inside wall of fused silica
open tubular column.
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 49/68
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 50/68
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 51/68
Capillary column vs Packed column
Capillary Packed
Length 60 m 2 m
Theoretical plates( N/m) 3000-5000 2000
Total plates
length×( N/m)
180000-300000 4000
Advantages of capillary column
1) Not packed : long lengths : high resolution
2) Thin film ; efficient, fast
3) Used silica --- inert surface, better results
Disadvantages of capillary column
1) More expensive
2) Limited liquid phase
Requires very small samples
3) Dedicated instrumentation ---
capillary inlets, septum purge
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 52/68
(Left) Gas chromatogram of alcohol mixture at 40o
C using packed column ( 2mmI.D., 76 cm long containing 20 % Carbowax 20 M on a Gas-Chrom R support and
FID.
(Right) Chromatogram of vapors from headspace of beer can, obtained with 0.25 mm
diameter, 30 m long porous carbon column oerated at 30 oC for 2 min and then
ramped up to 160 oC at 20 oC/min.
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 53/68
Component Separation with the Column
< The process of separation >
A series of partitions : Dynamic In-and Out (or Stop-and-Go)
All differential migration process.
The most volatile components usually pass through the column
first, the least volatile or highest boiling emerges last.
Mobile phase( Driving force)
Stationary phase ( Resistive force)
Analytes
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 54/68
Capillary Column Installation Steps
1. Check traps, carrier gas, septum, liner 2. Place the nut and ferrule on the column and carefully cut the column end
3. Install the column into the injector
4. Turn on the carrier gas
5. Install the column into the detector
6. Inspect for leaks7. Confirm carrier gas flow and proper column installation
8. Condition the column
9. Accurately set the carrier gas velocity
10. Bleed test
11. Run test mixClick here for a complete listing of tools available from J&W, including magnifiers and cutting tools.
http://www.jandw.com/gccolumn.htm#Fused Silica Capillary Columns
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 55/68
Recommended Installation Tools and Supplies
1. Cutting tool such as a diamond or carbide tipped pencil, sapphire tipped pencil or ceramic cleaving wedge
2. Magnifier (10-20X)
3. Ruler
4. Wrench
5. Ferrules
6. Vial of solvent7. Clean syringe
8. Supply of an appropriate non-retained compound
9. Column test mixture
10. Flow meter
11. Other supplies: septa, clean injector liners, liner ferrules/O-rings, etc.
C diti i f th C l
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 56/68
Conditioning of the Column
Once the column has been checked for proper installation and the absence of
leaks, it is ready for conditioning. Heat the column to its isothermal, upper
temperature limit (temperature limits listed below) or a temperature 10-20 oC
above the highest operating temperature of your particular method. Do not exceedthe upper limit or column damage will result. Heat the column rapidly - slow
temperature programming is not necessary. After the column has reached the
conditioning temperature, plot the baseline. Keep the baseline on scale so that it
can be observed. The baseline should be elevated at first then start to drop after 5-
10 minutes at the conditioning temperature. The baseline will continue to drop for 30-90 minutes then stabilize at a constant value. If the baseline does not stabilize
after 2-3 hours or does not start to significantly decrease after 15-20 minutes,
either a leak is present or a contamination problem exists. In either case,
immediately cool the oven down below 40 oC and resolve the problem. Continued
conditioning will result in column damage or the inability to obtain a stable
baseline. Excessive conditioning of the column may result in a shortened lifetime.
In general, polar stationary phases and thick film columns usually require longer
times to stabilize than less polar and thinner film columns. GS PLOT columns
require a different conditioning procedure than liquid stationary phase columns.
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 57/68
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 58/68
Temperature Limits
The temperature limits define the range over which the column can be safelyused. If the oven is operated below the lower temperature limit, poor
separation and peak shape problems will be evident, but no column damage
will occur.
Upper temperature limits are usually given as two numbers. The first or
lower temperature of the two is the isothermal limit. The column can bemaintained at this temperature for indefinite periods of time. The second or
higher temperature is the program limit. The column can be heated to this
temperature for short periods of time (<10 minutes). Exceeding the upper
temperature limits will significantly reduce column lifetime.
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 59/68
Liquid Phase(=Stationary phase) Classes
1. Non-polar phase gives boiling point order separation
2. Selective phase separates components that have close b.p. and
small structural differences
3. Polar phase depends on internal functional groups to separate
compounds that have reactive – OH, – NH2 or other polar radicals
4. Each stationary phase retains solutes in its own class best
Raising percentage of stationary phase leads to
1) Greater capacity for solute
2) Longer retention time
3) Increased HETP
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 60/68
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 61/68
Column Temperature
Dimethylpolysiloxane
DB-1; 0.1, 0.25 and 1.0 m -60 to 325/350 oC
DB-1; 3.0 and 5.0 m -60 to 280/300
o
CDB-1; Megabore 0.1 m -60 to 360 oC
DB-1; Megabore 1.5 m -60 to 300/320 oC
DB-1; Megabore 3.0 and 5.0 m -60 to 260/280 oC
DB-1ht -60 to 400 oC
(5%-Phenyl) Methylpolysiloxane
DB-5 -60 to 325/350 oC
DB-5; Megabore -60 to 300/320 oC
DB-5; Megabore 5.0 m -60 to 260/280 oC
DB-5ht -60 to 400 oC
DB-5ms -60 to 325/350 oC
DB-5ms; Megabore -60 to 300/320 oC
DB-5.625 -60 to 325/350 oC
Polyethylene Glycol
DB-WAX 20 to 250/260 oC
DB-WAX; Megabore 20 to 230/240 oC
Base Deactivated Polyethylene Glycol
CAM 60 to 220/240 oC
Carbowax 20M
Carbowax 60 to 220/240 oC
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 62/68
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 63/68
-Cyclodextrin for chiral column.
Column oven and temperature control
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 64/68
Column oven and temperature control
Oven size : column sits in an oven
Inner volume : laboratory GC ; about 22-25 L
process GC ; about 40 L
Isothermal
The oven temperature is kept constant during the entire analysis
Practical temperature < 250oC
Maximum temperature < 400oC
Temperature programming
The oven temperature is varied during the analysis
Linear
Non linear
High temperature GC (HTGC)
Working limit : conventional GC --- 330oC
HTGC --- 450oC
Masses of analysed substrates: 600-1000
Temperature programming
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 65/68
Temperature programming
With homologues, the retention time increases exponentially with the number
of carbon.
As retention time increases, width increase and the height decreases, making
detection impossible after a few peaks have eluted.
Since solubility of gas in a liquid decreases as temperatures goes up, we can
reduce the retention of a compound by increasing column temperature.
General steps to create a program assuming that the separation is possible
1) Determine initial temperature and time based on best possible separation
offirst few peaks
2) Report for the last few peaks to find the best final temperature and time
3) Experiment with various ramps to account for the rest of the components
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 66/68
Temperature programming
Factors to consider :
Variations in solubility of solutes
Changes in volatility of solutes
Stability of solutesFlow rate changes
Stability of stationary phase
Must stay within Tmin/Tmax of column
Other factors are found experimentally
A temperature program
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 67/68
A temperature program
Ex.
40 oC(5 min) – 10oC/min – 250oC(10 min)
A: initial temperature and holding time
B: ramp (oC/min)
C: final temperature and holding time
Some GCs will allow for a more complex program.
A
B
C
Raising column temperature
1) Decrease retention time
2) Decrease resolution
3) Sharpens peaks
7/27/2019 F AC 31 08 [a] Gas Chromatogrphy
http://slidepdf.com/reader/full/f-ac-31-08-a-gas-chromatogrphy 68/68
Comparison of isothermal and programmed temperature chromatography. Each sample contains linear
alkanes run on a 1 6 mm × 6 m column containing 3% Apiezon L (liquid phase) on a 100/200 mesh